Reducing Radiation in Chronic Total Occlusion Percutaneous Coronary Interventions

Antonis N Pavlidis 1,*, Daniel A Jones 1, Alex Sirker 2, Anthony Mathur 1, Elliot J Smith 1
PMCID: PMC4807712  PMID: 25847013

Abstract

The field of percutaneous intervention for chronic total occlusion (CTO) has enjoyed significant innovations in the recent years. Novel techniques and technologies have revolutionized the field and have resulted in considerably higher success rates even in patients with high anatomical complexity. Successful CTO recanalization is associated with significant clinical benefits, such as the improvement of angina and quality of life, reduced rates of surgical revascularization, improvement of left ventricular function and decreased mortality rates. However, complex CTO procedures often require prolonged x-ray exposure which have been associated with adverse long term outcomes.

Keywords: Coronary artery disease, chronic total occlusions, coronary intervention, radiation, skin injury

INTRODUCTION

Radiation exposure is higher during percutaneous coronary interventions (PCI) for chronic total occlusions (CTO) compared to non-CTO interventions because of prolonged fluoroscopic time and repeated cine angiography [1]. Although radiation skin injury is rare, the risks of radiation-related complications are greater in CTO procedures. Moreover, operator and lab staff exposure can result in longterm adverse outcomes, such as cataracts and malignancies, therefore reducing radiation exposure is a key factor in CTO interventions.

RADIATION DOSES IN CTO PROCEDURES

There are three different values that are currently measured by modern interventional fluoroscopic equipment: (a) the entrance surface air kerma (ESAK), measured in Gray (Gy), which represents the radiation energy released at the point where the X-ray beam enters the patient’s skin surface and includes both the incident air kerma and radiation backscattered from the tissue, (b) the dose area product (DAP), measured in Gy.cm2, which represents the product of the dose in air within the X-ray beam and the beam area, and is therefore a measure of all the radiation that enters the patient and (c) the fluoroscopic time (FT), measured in minutes, which is the time during a procedure that fluoroscopy is used. The ESAK is used to measure the deterministic risk to the patient such as skin injury, while the DAP is used to measure the stochastic risk of the patient, which involves the likelihood of developing malignancies or genetic defects in the future. FT does not include cine acquisition imaging and is therefore inadequate to assess patient radiation.

A plain chest radiography produces a DAP of 0.08 Gy.cm2 with a background equivalent of three days, while the equivalents for a non-CTO PCI with one stent are 36 Gy.cm2 and 3.7 years [2]. According to a study by Suzuki et al. [1] the median ESAK for a CTO PCI was 4.6 Gy, compared to 2.4 Gy, 1.5 Gy and 1.2 Gy for multivessel, single-vessel multiple stenosis and single stenosis PCI respectively.

Several lesion- and patient-related risk factors have been shown to affect radiation dose during percutaneous interventions. In a study of 1933 PCI procedures Fetterly et al. [3] found that lesion complexity, PCI of left circumflex artery, previous coronary artery bypass grafting (CABG), body mass index (BMI) and the number of treated lesions correlated to an increased ESAK. Similar results were found in a larger study by Delewi et al. [4] which included 9850 PCI procedures. They demonstrated that high BMI, previous history of coronary artery bypass grafting, the number of treated lesions and CTO interventions were associated with the highest patient radiation exposure.

DETERMINISTIC EFFECTS

Radiation-induced skin injury is an infrequent complication during PCI, but appears more often in CTO interventions as a result of prolonged fluoroscopy times. Radiation toxicity is rare with <5 Gy but patients with higher doses should be followed up 2-3 weeks after the procedure and assessed for development of new skin changes. At Grade I radiation-induced skin injury a faint erythema can be seen during the first 48 hours after exposure. Following a latent phase that can last up to 5 weeks, moderate to brisk erythema with oedema can be observed. Larger doses of radiation can result in Grade IV injury with skin necrosis or ulceration within 2 weeks after exposure (Fig. 1), (Table 1) [5].

Fig. (1).

Fig. (1)

Radiation-induced skin injury.

Table 1.

Staging of radiation skin injury [31].

Grade Skin appearance Radiation dose Time after radiation exposure
1 Faint erythema or desquamation > 2 Gy First 48 hours
2 Moderate to brisk erythema or moist desquamation. Moderate swelling. > 15 Gy 2-5 weeks
3 Confluent, moist desquamation > 1.5 cm diameter, which is not confined to the skin folds. Pitting oedema > 40 Gy 6-7 weeks
4 Skin necrosis or ulceration of full thickness dermis > 550 Gy 2 weeks

Radiation-induced skin injury appears to be the least frequently reported complication following CTO intervention. In a meta-analysis of 65 studies Patel et al. [6] showed an incidence of <0.01% with only 3 reported cases among 2,857 patients. However, radiation skin injury was the least frequently reported CTO complication, with only 11% of the studies reporting on its occurrence. In another metanalysis of retrograde CTO interventions, El Sabbagh et al. [7] reported an incidence of 0.5%, although only 2 out of the 26 studies (0.08%) reported this complication. Morino et al. [8] published the outcomes for 498 patients and 528 CTO lesions included in the J-CTO registry (Multicenter CTO Registry of Japan) and no incidents of radiation-induce skin injuries were reported. The incidence of radiation skin in the most recent CTO registries and meta-analyses is summarised in Table 2.

Table 2.

Radiation doses in recent CTO registries and meta-analyses.

Study type Characteristics Year CTO lesions (n) Radiation skin injury (%) Fluoroscopy time
Lin et al. [32] Retrospective study 2014 516 NR 43 ± 27 # 42 ± 24 ##
El Sabbagh et al. [7] Metanalysis Retrograde only 2014 3493 0.5% 82 ± 34
Christopoulos et al. [33] Registry 2014 496 NR 41 (26-65) *
Michael et al. [34] Registry 2013 1361 NR 42 ± 29
Karmpaliotis et al. [35] Registry Retrograde only 2012 462 NR 61 ± 40
Tsuchikane et al. [36] Registry Retrograde only 2013 801 NR 95 ± 52
Patel et al. [37] Metanalysis 2013 18941 <0.01 NR
Galassi et al. [38] Registry 2011 1983 NR 42 ± 47
Morino et al. [8] Registry 2010 528 0 45 (1-301) *
Aguiar-Souto et al. [39] Retrospective study 2010 227 NR 32 (19-47) *

#CI-AKI group, ## non CI-AKI group, CI-AKI: contrast induced acute kidney injury, * Median (range)

STOCHASTIC EFFECTS

Radiation dose exposure has been related to an additional risk of developing solid tumours [9]. The Biological Effects of Ionizing Radiation (BEIR) VIII risk model suggests that the risk of malignancy increases proportionally to the dose of the radiation, with no low-dose threshold [10]. At low doses of exposure, the risk of developing a malignancy is naturally low, but not zero. Minimal data on the risk of malignancy related to radiation exposure during PCI are available. Godino et al. [11] estimated the malignancy risk due to radiation exposure in patients undergoing PCI for acute ST-elevation myocardial infarction (STEMI) and patients undergoing CTO PCI by incorporating the effective radiation dose into the Biological Effects of Ionizing Radiation (BEIR) VII model [12]. They found that the number of estimated additional lung and bone marrow malignancy cases were on average two times higher in patients treated for CTOs compared to STEMI patients. Nevertheless, the above observations have not yet been confirmed in epidemiological studies and there is therefore insufficient evidence to defer a CTO intervention based on concerns over radiation exposure [13].

RADIATION EXPOSURE EFFECTS TO OPERATORS

Chronic exposure to low doses of ionising radiation has shown to cause DNA damage in interventional cardiologists, which appears to correlate with the number of years of catheterization laboratory experience [14]. Venneri et al. [15] used the BEIR VII model to show that interventionalists had an increased cancer risk caused by professional radiation exposure. Recently published case clusters of interventional cardiologists with left sided brain neoplasms have raised the existing concerns, since radiation exposure to the left side is higher during PCI [16-18]. Despite the above, the risks related to radiation exposure in operators remains uncertain and further studies are required.

Radiation-induced cataracts represents another occupational hazard to interventional cardiologists. The RELID (Retrospective evaluation study of lens injuries and dose) study showed that they have a three-fold higher rate of posterior subcapsular lens opacities compared to unexposed individuals. Although the risk of developing cataracts is dose-dependent [19], it appears to be lower for regular users of protective lead glasses [20]. The International Commission on Radiological Protection (ICRP) have suggested a threshold dose to the lens of 20 mSv per year, averaged over 5 consecutive years, with a maximum of 50 mSv in a single year [21].

METHODS FOR REDUCING RADIATION DURING CTO INTERVENTIONS

A). Pre-procedural Strategies

Careful selection of patients and early assessment of the risk factors that are associated with high risk for radiation injury is of primary importance. Patients with recent radiation exposure are at particularly high risk of radiation skin injury [22]. Every patient should be consented on the risks of radiation-related complications and careful examination of the skin should be performed prior to starting a CTO procedure. Moreover, the ‘CTO team’, including physicians, nurses and technicians, should always review the angiographic images prior to the procedure in order to understand the anatomy and plan the interventional strategy. Specific radiographic views that are most likely to be useful should be identified early in order to avoid unnecessary radiation exposure.

Computed tomography coronary angiography (CTCA) is a useful tool for CTO pre-procedural planning. Although the contribution of multislice CT (MSCT) is approximately 19 mSv [23] the total radiation dose can be decreased significantly with successful CTO road mapping, based on the additional information on lesion characteristics. Incorporation of ECG-pulsed modulation of the tube current [24] and the use of new generation MSCT equipment can lower the effective radiation dose significantly [2].

Finally, each cardiac lab should have an established radiation safety program and operators should undergo compulsory training on radiation dose management and safety. Studies have shown that radiation doses can be reduced up to 34% if operators have recently attended an informative conference on appropriate use of radiation and changes in x-ray delivery settings [25].

B). Intra-procedural Strategies

Staff radiation dose should be closely monitored with personal dose monitors and dosimeter records should be provided to operators regularly. The ICRP recommends the use of two dosimeters [21]: one under the protective garment, usually at waist height, and a second outside the thyroid collar. If unusually high doses are recorded a review of staff practice patterns and adoption of further safety measures should be applied.

Protective 0.5 mm lead aprons, thyroid shielding, shin leg covers and radiation-specific glasses can stop up to 95% of the scattered radiation and should be worn by all CTO operators [22]. Apart from the commonly used radiation shielding, additional protection could be achieved during CTO interventions with below table mounted shielding and the recently developed Trinity Radiation Protection system [26]. The latter consists of a combination of fixed shields, radiation drapes and interconnecting flexible radiation resistant materials that create a complete radiation protection environment for the operators.

All CTO operators should be familiar with and apply the ALARA (As Low As Reasonably Achievable) principle, which means using all relevant methods and strategies in order to minimize radiation dose. Radiation exposure should be closely monitored at any time during the procedure. The operator should be alerted by the cardiac lab team when radiation levels exceed certain limits in order to balance the risks and benefits of discontinuing the procedure. A dose of 10 Gy ESAK has been suggested as a threshold at which a CTO operator should discontinue the procedure provided it is safe to do so, unless lesion crossing has occurred and the procedure is expected to be completed within a short period of time [27].

Increasing the distance between the patient and the X-ray tube by positioning the table at a higher level can result in significant reduction of radiation dose, although this should never affect the operator’s comfort [28]. Higher magnification increases the patient’s dose and should only be utilized in special circumstances. Moreover, all CTO operators should be familial with undergoing procedures at lower framing rates per second (fps) (6.0-7.5 fps instead of 15 fps) and using pulsed fluoroscopy mode rather than the digital cine mode storage. The number of acquisition runs should be held for optimising the strategy and assessing possible complications. Altering the beam angulation during the procedure by rotating the x-ray tube more than 40° can reduce the patient's skin dose and minimize irradiation of a particular portion of the patient’s skin [29, 30]. Steep angles have been linked to higher radiation doses due to penetration through more layers of tissue and should therefore be avoided [28]. Collimation decreases scatter radiation and the overall dose received by the patient. The use of additional copper filters reduces primary beam exposure and can enhance focused visualization.

CTO SPECIFIC TECHNIQUES

Adoption of the hybrid approach with early switch from a failing strategy maximizes the chance of procedural success, reduces procedure time and minimises radiation [27]. Although there are no specific time limits for each of the algorithm steps, operators should stop pursuing a technique that has not resulted in any significant progress during a reasonable period of time [27].

The use of certain techniques during CTO interventions can result in significant reduction of total radiation exposure:

  • The trapping technique for equipment exchange (balloon inflation inside the guiding catheter to fix the wire).

  • During dual injections, the donor vessel is injected first to allow time to fill the distal vessel. Fluoroscopy or cine acquisition begins 1-2 seconds later and it is followed by injection of the occluded vessel.

  • The use of intravascular ultrasound (IVUS) for proximal cap identification, re-entry guidance, assessment of retrograde wire position in reverse CART and stent optimisation.

  • Marking the length of the wire that can be advanced safely without exiting the microcatheter during wire exchanges or when modifying the wire’s tip bend. A pre-attached torquer at the end of the inserted wire or a stable marker on the table can be used.

The methods for reducing radiation during CTO interventions are summarised in Table 3.

Table 3.

Methods for reducing radiation during CTO interventions.

Pre-procedure Intra-procedure Post-procedure
Patient selection and risk assessment Dosimeters Dose documentation
Consent Protection clothing Follow up
Review films Shielding
CTCA ALARA principle
Radiation safety program Alert operator when radiation exceeds limits
Compulsory training on radiation safety and management Table position at higher level
Lower magnification
Lower frame rates
Changing beam angulation
Collimation
Procedure techniques

CTO: chronic total occlusion, CTCA: computed tomography coronary angiography, ALARA: As Low As Reasonably Achievable

C). Post-procedural Strategies and Follow Up

Post-procedure dose analysis is important in order to determine further management and follow up, especially if a repeat procedure is planned and the initial procedure resulted in high radiation exposure. The CTO procedure report should include all available radiation dose parameters, such as the fluoroscopy time, ESAK and DAP.

Post-procedure follow up should be guided by the Air Kerma Dose that the patient received during the CTO intervention [22]:

  • 5 Gy: patients should be educated regarding potential skin changes on their back. A thorough examination of the skin should be performed 1 month following the procedure and if there is evidence of radiation-induced skin injury an appropriate specialist referral should be considered.

  • 10 Gy: patients should be educated accordingly and followed up after 2-4 weeks. A qualified physicist should promptly be asked to calculate peak skin dose.

  • 15 Gy: dosed above this level are identified by the Joint Commission as a sentinel event, therefore hospital risk management and regulatory authorities need to be contacted within 24 hours after the procedure. No interventional procedure should reach this level unless there is a life threatening complication that necessitates obligatory percutaneous fluoroscopic reversal.

CONCLUSIONS

The field of CTO-PCI has evolved significantly in recent years and the hybrid approach to CTO offers the opportunity to treat more complex anatomy successfully and meet the needs of a wider patient population. Prevention of complications related to use of radiation represents a major component of a successful CTO intervention. Operator awareness and use of all the required precautions improves patient, staff and physician safety.

FUNDING

This research received no grant from any funding agency in the public, commercial or not-for-profit sectors.

CONFLICT OF INTEREST

The authors confirm that this article content has no conflict of interest.

ACKNOWLEDGEMENTS

Declared none.

REFERENCES

  • 1.Suzuki S., Furui S., Kohtake H., Yokoyama N., Kozuma K., Yamamoto Y., Isshiki T. Radiation exposure to patient’s skin during percutaneous coronary intervention for various lesions, including chronic total occlusion. Circ. J. 2006;70(1):44–48. doi: 10.1253/circj.70.44. [DOI] [PubMed] [Google Scholar]
  • 2.Sianos G., Werner G.S., Galassi A.R., Papafaklis M.I., Escaned J., Hildick-Smith D., Christiansen E.H., Gershlick A., Carlino M., Karlas A., Konstantinidis N.V., Tomasello S.D., Di Mario C., Reifart N., EuroCTO Club Recanalisation of chronic total coronary occlusions: 2012 consensus document from the EuroCTO club. EuroIntervention. 2012;8(1):139–145. doi: 10.4244/EIJV8I1A21. [DOI] [PubMed] [Google Scholar]
  • 3.Fetterly K.A., Lennon R.J., Bell M.R., Holmes D.R., Jr, Rihal C.S. Clinical determinants of radiation dose in percutaneous coronary interventional procedures: influence of patient size, procedure complexity, and performing physician. JACC Cardiovasc. Interv. 2011;4(3):336–343. doi: 10.1016/j.jcin.2010.10.014. [DOI] [PubMed] [Google Scholar]
  • 4.Delewi R., Hoebers L.P., Råmunddal T., Henriques J.P., Angerås O., Stewart J., Robertsson L., Wahlin M., Petursson P., Piek J.J., Albertsson P., Matejka G., Omerovic E. Clinical and procedural characteristics associated with higher radiation exposure during percutaneous coronary interventions and coronary angiography. Circ. Cardiovasc. Interv. 2013;6(5):501–506. doi: 10.1161/CIRCINTERVENTIONS.113.000220. [DOI] [PubMed] [Google Scholar]
  • 5.Slovut D.P. Cutaneous radiation injury after complex coronary intervention. JACC Cardiovasc. Interv. 2009;2(7):701–702. doi: 10.1016/j.jcin.2009.04.015. [DOI] [PubMed] [Google Scholar]
  • 6.Patel V.G., Michael T.T., Mogabgab O., Fuh E., Banerjee A., Brayton K.M., Cipher D.J., Abdullah S.M., Brilakis E.S. Clinical, angiographic, and procedural predictors of periprocedural complications during chronic total occlusion percutaneous coronary intervention. J. Invasive Cardiol. 2014;26(3):100–105. [PubMed] [Google Scholar]
  • 7.El Sabbagh A., Patel V.G., Jeroudi O.M., Michael T.T., Alomar M.E., Mogabgab O., Fuh E., Roesle M., Rangan B.V., Abdullah S., Hastings J.L., Grodin J., Kumbhani D.J., Alexopoulos D., Fasseas P., Banerjee S., Brilakis E.S. Angiographic success and procedural complications in patients undergoing retrograde percutaneous coronary chronic total occlusion interventions: a weighted meta-analysis of 3,482 patients from 26 studies. Int. J. Cardiol. 2014;174(2):243–248. doi: 10.1016/j.ijcard.2014.04.004. [DOI] [PubMed] [Google Scholar]
  • 8.Morino Y., Kimura T., Hayashi Y., Muramatsu T., Ochiai M., Noguchi Y., Kato K., Shibata Y., Hiasa Y., Doi O., Yamashita T., Morimoto T., Abe M., Hinohara T., Mitsudo K., J-CTO Registry Investigators In-hospital outcomes of contemporary percutaneous coronary intervention in patients with chronic total occlusion insights from the J-CTO Registry (Multicenter CTO Registry in Japan). JACC Cardiovasc. Interv. 2010;3(2):143–151. doi: 10.1016/j.jcin.2009.10.029. [DOI] [PubMed] [Google Scholar]
  • 9.Wrixon A.D. New ICRP recommendations. J. Radiol. Prot. 2008;28(2):161–168. doi: 10.1088/0952-4746/28/2/R02. [DOI] [PubMed] [Google Scholar]
  • 10.Einstein A.J., Henzlova M.J., Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298(3):317–323. doi: 10.1001/jama.298.3.317. [DOI] [PubMed] [Google Scholar]
  • 11.Godino C., Maccagni D., Pavon A.G., Viani G., Cappelletti A., Margonato A., Colombo A. Estimating incidence of organ cancer related to PCI radiation exposure in patients treated for acute and chronic total occlusions. J. Invasive Cardiol. 2013;25(9):441–445. [PubMed] [Google Scholar]
  • 12.Committee to Assess Health Risks from Exposure to Low Level of Ionizing Radiation. Health risks from exposure to low levels of ionizing radiation: BEIR VII phase 2. Washington, DC: National Academies; National Research Council; [PubMed] [Google Scholar]
  • 13.Shah N., Deshmukh A., Sachdeva R. Radiation exposure after percutaneous coronary intervention: is the cancer risk real? J. Invasive Cardiol. 2013;25(9):447–448. [PubMed] [Google Scholar]
  • 14.Andreassi M.G., Cioppa A., Botto N., Joksic G., Manfredi S., Federici C., Ostojic M., Rubino P., Picano E. Somatic DNA damage in interventional cardiologists: a case-control study. FASEB J. 2005;19(8):998–999. doi: 10.1096/fj.04-3287fje. [DOI] [PubMed] [Google Scholar]
  • 15.Venneri L., Rossi F., Botto N., Andreassi M.G., Salcone N., Emad A., Lazzeri M., Gori C., Vano E., Picano E. Cancer risk from professional exposure in staff working in cardiac catheterization laboratory: insights from the National Research Council’s Biological Effects of Ionizing Radiation VII Report. Am. Heart J. 2009;157(1):118–124. doi: 10.1016/j.ahj.2008.08.009. [DOI] [PubMed] [Google Scholar]
  • 16.Roguin A., Goldstein J., Bar O. Brain tumours among interventional cardiologists: a cause for alarm? Report of four new cases from two cities and a review of the literature. EuroIntervention. 2012;7(9):1081–1086. doi: 10.4244/EIJV7I9A172. [DOI] [PubMed] [Google Scholar]
  • 17.Roguin A., Goldstein J., Bar O. Brain malignancies and ionising radiation: more cases reported. EuroIntervention. 2012;8(1):169–170. doi: 10.4244/EIJV8I1A26. [DOI] [PubMed] [Google Scholar]
  • 18.Roguin A., Goldstein J., Bar O., Goldstein J.A. Brain and neck tumors among physicians performing interventional procedures. Am. J. Cardiol. 2013;111(9):1368–1372. doi: 10.1016/j.amjcard.2012.12.060. [DOI] [PubMed] [Google Scholar]
  • 19.Ciraj-Bjelac O., Rehani M.M., Sim K.H., Liew H.B., Vano E., Kleiman N.J. Risk for radiation-induced cataract for staff in interventional cardiology: is there reason for concern? Catheter. Cardiovasc. Interv. 2010;76(6):826–834. doi: 10.1002/ccd.22670. [DOI] [PubMed] [Google Scholar]
  • 20.Jacob S., Boveda S., Bar O., Brézin A., Maccia C., Laurier D., Bernier M.O. Interventional cardiologists and risk of radiation-induced cataract: results of a French multicenter observational study. Int. J. Cardiol. 2013;167(5):1843–1847. doi: 10.1016/j.ijcard.2012.04.124. [DOI] [PubMed] [Google Scholar]
  • 21.International Commission on Radiological Protection. Avoidance of radiation injuries from medical interventional procedures. Ann. ICRP. 2000:25–43. doi: 10.1016/S0146-6453(01)00004-5. [DOI] [PubMed] [Google Scholar]
  • 22.Chambers C.E., Fetterly K.A., Holzer R., Lin P.J., Blankenship J.C., Balter S., Laskey W.K. Radiation safety program for the cardiac catheterization laboratory. Catheter. Cardiovasc. Interv. 2011;77(4):546–556. doi: 10.1002/ccd.22867. [DOI] [PubMed] [Google Scholar]
  • 23.García-García H.M., van Mieghem C.A., Gonzalo N., Meijboom W.B., Weustink A.C., Onuma Y., Mollet N.R., Schultz C.J., Meliga E., van der Ent M., Sianos G., Goedhart D., den Boer A., de Feyter P., Serruys P.W. Computed tomography in total coronary occlusions (CTTO registry): radiation exposure and predictors of successful percutaneous intervention. EuroIntervention. 2009;4(5):607–616. doi: 10.4244/EIJV4I5A102. [DOI] [PubMed] [Google Scholar]
  • 24.Abada H.T., Larchez C., Daoud B., Sigal-Cinqualbre A., Paul J.F. MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. AJR Am. J. Roentgenol. 2006;186(6) Suppl. 2:S387–S390. doi: 10.2214/AJR.05.0216. [DOI] [PubMed] [Google Scholar]
  • 25.Azpiri-López J.R., Assad-Morell J.L., González-González J.G., Elizondo-Riojas G., Dávila-Bortoni A., García-Martínez R., Treviño-Frutos R. Effect of physician training on the X-ray dose delivered during coronary angioplasty. J. Invasive Cardiol. 2013;25(3):109–113. [PubMed] [Google Scholar]
  • 26.Fattal P., Goldstein J.A. A novel complete radiation protection system eliminates physician radiation exposure and leaded aprons. Catheter. Cardiovasc. Interv. 2013;82(1):11–16. doi: 10.1002/ccd.24625. [DOI] [PubMed] [Google Scholar]
  • 27.Brilakis E.S., Grantham J.A., Rinfret S., Wyman R.M., Burke M.N., Karmpaliotis D., Lembo N., Pershad A., Kandzari D.E., Buller C.E., DeMartini T., Lombardi W.L., Thompson C.A. A percutaneous treatment algorithm for crossing coronary chronic total occlusions. JACC Cardiovasc. Interv. 2012;5(4):367–379. doi: 10.1016/j.jcin.2012.02.006. [DOI] [PubMed] [Google Scholar]
  • 28.Brilakis E.S., editor. Manual of Coronary Chronic Total Occlusion Interventions. A Step-By-Step Approach. Waltham, MA: Elsevier; 2013. [Google Scholar]
  • 29.Abdelaal E., Plourde G., MacHaalany J., Arsenault J., Rimac G., Déry J.P., Barbeau G., Larose E., De Larochellière R., Nguyen C.M., Allende R., Ribeiro H., Costerousse O., Mongrain R., Bertrand O.F., Interventional Cardiologists at Quebec Heart-Lung Institute Effectiveness of low rate fluoroscopy at reducing operator and patient radiation dose during transradial coronary angiography and interventions. JACC Cardiovasc. Interv. 2014;7(5):567–574. doi: 10.1016/j.jcin.2014.02.005. [DOI] [PubMed] [Google Scholar]
  • 30.Bashore T.M., Balter S., Barac A., Byrne J.G., Cavendish J.J., Chambers C.E., Hermiller J.B., Jr, Kinlay S., Landzberg J.S., Laskey W.K., McKay C.R., Miller J.M., Moliterno D.J., Moore J.W., Oliver-McNeil S.M., Popma J.J., Tommaso C.L. 2012 American College of Cardiology Foundation/Society for Cardiovascular Angiography and Interventions expert consensus document on cardiac catheterization laboratory standards update: A report of the American College of Cardiology Foundation Task Force on Expert Consensus documents developed in collaboration with the Society of Thoracic Surgeons and Society for Vascular Medicine. J. Am. Coll. Cardiol. 2012;59(24):2221–2305. doi: 10.1016/j.jacc.2012.02.010. [DOI] [PubMed] [Google Scholar]
  • 31.Bernier J., Bonner J., Vermorken J.B., Bensadoun R.J., Dummer R., Giralt J., Kornek G., Hartley A., Mesia R., Robert C., Segaert S., Ang K.K. Consensus guidelines for the management of radiation dermatitis and coexisting acne-like rash in patients receiving radiotherapy plus EGFR inhibitors for the treatment of squamous cell carcinoma of the head and neck. Ann. Oncol. 2008;19(1):142–149. doi: 10.1093/annonc/mdm400. [DOI] [PubMed] [Google Scholar]
  • 32.Lin Y.S., Fang H.Y., Hussein H., Fang C.Y., Chen Y.L., Hsueh S.K., Cheng C.I., Yang C.H., Chen C.J., Hang C.L., Yip H.K., Wu C.J. Predictors of contrast-induced nephropathy in chronic total occlusion percutaneous coronary intervention. EuroIntervention. 2014;9(10):1173–1180. doi: 10.4244/EIJV9I10A198. [DOI] [PubMed] [Google Scholar]
  • 33.Christopoulos G., Menon R.V., Karmpaliotis D., Alaswad K., Lombardi W., Grantham J.A., Michael T.T., Patel V.G., Rangan B.V., Kotsia A.P., Lembo N., Kandzari D.E., Lee J., Kalynych A., Carlson H., Garcia S., Banerjee S., Thompson C.A., Brilakis E.S. Application of the “hybrid approach” to chronic total occlusions in patients with previous coronary artery bypass graft surgery (from a Contemporary Multicenter US registry). Am. J. Cardiol. 2014;113(12):1990–1994. doi: 10.1016/j.amjcard.2014.03.039. [DOI] [PubMed] [Google Scholar]
  • 34.Michael T.T., Karmpaliotis D., Brilakis E.S., Fuh E., Patel V.G., Mogabgab O., Alomar M., Kirkland B.L., Lembo N., Kalynych A., Carlson H., Banerjee S., Lombardi W., Kandzari D.E. Procedural outcomes of revascularization of chronic total occlusion of native coronary arteries (from a multicenter United States registry). Am. J. Cardiol. 2013;112(4):488–492. doi: 10.1016/j.amjcard.2013.04.008. [DOI] [PubMed] [Google Scholar]
  • 35.Karmpaliotis D., Michael T.T., Brilakis E.S., Papayannis A.C., Tran D.L., Kirkland B.L., Lembo N., Kalynych A., Carlson H., Banerjee S., Lombardi W., Kandzari D.E. Retrograde coronary chronic total occlusion revascularization: procedural and in-hospital outcomes from a multicenter registry in the United States. JACC Cardiovasc. Interv. 2012;5(12):1273–1279. doi: 10.1016/j.jcin.2012.06.025. [DOI] [PubMed] [Google Scholar]
  • 36.Tsuchikane E., Yamane M., Mutoh M., Matsubara T., Fujita T., Nakamura S., Muramatsu T., Okamura A., Igarashi Y., Oida A., Retrograde Summit Investigators Japanese multicenter registry evaluating the retrograde approach for chronic coronary total occlusion. Catheter. Cardiovasc. Interv. 2013;82(5):E654–E661. doi: 10.1002/ccd.24823. [DOI] [PubMed] [Google Scholar]
  • 37.Patel V.G., Brayton K.M., Tamayo A., Mogabgab O., Michael T.T., Lo N., Alomar M., Shorrock D., Cipher D., Abdullah S., Banerjee S., Brilakis E.S. Angiographic success and procedural complications in patients undergoing percutaneous coronary chronic total occlusion interventions: a weighted meta-analysis of 18,061 patients from 65 studies. JACC Cardiovasc. Interv. 2013;6(2):128–136. doi: 10.1016/j.jcin.2012.10.011. [DOI] [PubMed] [Google Scholar]
  • 38.Galassi A.R., Tomasello S.D., Reifart N., Werner G.S., Sianos G., Bonnier H., Sievert H., Ehladad S., Bufe A., Shofer J., Gershlick A., Hildick-Smith D., Escaned J., Erglis A., Sheiban I., Thuesen L., Serra A., Christiansen E., Buettner A., Costanzo L., Barrano G., Di Mario C. In-hospital outcomes of percutaneous coronary intervention in patients with chronic total occlusion: insights from the ERCTO (European Registry of Chronic Total Occlusion) registry. EuroIntervention. 2011;7(4):472–479. doi: 10.4244/EIJV7I4A77. [DOI] [PubMed] [Google Scholar]
  • 39.Aguiar-Souto P., Ferrante G., Del Furia F., Barlis P., Khurana R., Di Mario C. Frequency and predictors of contrast-induced nephropathy after angioplasty for chronic total occlusions. Int. J. Cardiol. 2010;139(1):68–74. doi: 10.1016/j.ijcard.2008.10.006. [DOI] [PubMed] [Google Scholar]