Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1964 Sep;17(5):531–535.

Control of anticoagulant therapy: Postal comparison between five centres

PMCID: PMC480809  PMID: 14207787

Abstract

Laboratory techniques used in the control of oral anticoagulant therapy were compared at five centres. The methods studied were: (a) prothrombin-proconvertin (`P and P') technique with the centres using their own reagents. This resulted in major discrepancies between the centres. (b) P and P technique with all the centres using centrally supplied reagents. The discrepancies noted under (a) disappeared when (b) was employed. (c) Thrombotest: the centres used the same batch of reagent and the specimens were transmitted in glass or polystyrene containers. The centres obtained closely comparable results, whether transportation was in glass or polystyrene. (d) Onestage `prothrombin' time with all the centres using the same reagents. When the results were read from a dilution curve there were discrepancies between the centres.

For comparability of results between centres using anticoagulant therapy, centrally supplied P and P reagents or Thrombotest should be used. Since only the latter was commercially available as a centrally standardized reagent at the initiation of this trial, it emerged as the method of practical choice at the time.

Full text

PDF
531

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. ALLINGTON M. J. Owren's method for the control of anticoagulant therapy. J Clin Pathol. 1958 Jan;11(1):62–68. doi: 10.1136/jcp.11.1.62. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. OWREN P. A., AAS K. The control of dicumarol therapy and the quantitative determination of prothrombin and proconvertin. Scand J Clin Lab Invest. 1951;3(3):201–208. doi: 10.3109/00365515109060600. [DOI] [PubMed] [Google Scholar]
  3. OWREN P. A. Thrombotest. A new method for controlling anticoagulant therapy. Lancet. 1959 Nov 7;2(7106):754–758. doi: 10.1016/s0140-6736(59)90857-8. [DOI] [PubMed] [Google Scholar]
  4. TOOHEY M. Comparison of Quick's, Owren's, and Ware's techniques for the control of anticoagulant therapy. J Clin Pathol. 1958 Jan;11(1):56–61. doi: 10.1136/jcp.11.1.56. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES