Skip to main content
British Heart Journal logoLink to British Heart Journal
. 1984 Oct;52(4):377–384. doi: 10.1136/hrt.52.4.377

Internal transvenous low energy cardioversion for the treatment of cardiac arrhythmias.

A W Nathan, R S Bexton, R A Spurrell, A J Camm
PMCID: PMC481645  PMID: 6477776

Abstract

Low energy endocardial cardioversion was attempted in 23 patients with 30 arrhythmias, of whom only four were receiving additional drug treatment. Four had atrial flutter, five atrial fibrillation, three intra-atrioventricular nodal tachycardia, two atrioventricular re-entrant tachycardia, 13 ventricular tachycardia, and three ventricular fibrillation. A pacing lead with special large surface area electrodes--the active electrode positioned either in the right atrium or in the right ventricular apex and the indifferent electrode in the right atrium, superior vena cava, or inferior vena cava--was used together with a low energy defibrillator. A total of 114 shocks was delivered, 26 of which were atrial. One episode of atrial flutter was terminated, but atrial fibrillation and atrioventricular nodal tachycardia were not terminated in any of the patients. Both patients with atrioventricular tachycardia were successfully treated, as were eight of the patients with ventricular tachycardia. Atrial fibrillation was produced in three patients and non-sustained ventricular tachycardia in one, ventricular tachycardia was accelerated in two, and ventricular fibrillation induced in five. Fourteen patients experienced severe discomfort and seven mild or moderate discomfort, and only one found the procedure painless. One patient was anaesthetised throughout the procedure. Low energy endocardial cardioversion is not universally successful even at the highest energies tolerable, and with the present electrode and pulse waveforms some patients may suffer considerable discomfort.

Full text

PDF
377

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Jackman W. M., Zipes D. P. Low-energy synchronous cardioversion of ventricular tachycardia using a catheter electrode in a canine model of subacute myocardial infarction. Circulation. 1982 Jul;66(1):187–195. doi: 10.1161/01.cir.66.1.187. [DOI] [PubMed] [Google Scholar]
  2. Mirowski M., Mower M. M., Gott V. L., Brawley R. K. Feasibility and effectiveness of low-energy catheter defibrillation in man. Circulation. 1973 Jan;47(1):79–85. doi: 10.1161/01.cir.47.1.79. [DOI] [PubMed] [Google Scholar]
  3. Mirowski M., Mower M. M., Reid P. R., Watkins L., Langer A. The automatic implantable defibrillator. New Modality for treatment of life-threatening ventricular arrhythmias. Pacing Clin Electrophysiol. 1982 May;5(3):384–401. doi: 10.1111/j.1540-8159.1982.tb02246.x. [DOI] [PubMed] [Google Scholar]
  4. Mirowski M., Mower M. M., Staewen W. S., Denniston R. H., Mendeloff A. I. The development of the transvenous automatic defibrillator. Arch Intern Med. 1972 May;129(5):773–779. [PubMed] [Google Scholar]
  5. Mirowski M., Reid P. R., Mower M. M., Watkins L., Gott V. L., Schauble J. F., Langer A., Heilman M. S., Kolenik S. A., Fischell R. E. Termination of malignant ventricular arrhythmias with an implanted automatic defibrillator in human beings. N Engl J Med. 1980 Aug 7;303(6):322–324. doi: 10.1056/NEJM198008073030607. [DOI] [PubMed] [Google Scholar]
  6. Naccarelli G. V., Zipes D. P., Rahilly G. T., Heger J. J., Prystowsky E. N. Influence of tachycardia cycle length and antiarrhythmic drugs on pacing termination and acceleration of ventricular tachycardia. Am Heart J. 1983 Jan;105(1):1–5. doi: 10.1016/0002-8703(83)90269-7. [DOI] [PubMed] [Google Scholar]
  7. Peters R. W., Shafton E., Frank S., Thomas A. N., Scheinman M. M. Radiofrequency-triggered pacemakers: uses and limitations. A long-term study. Ann Intern Med. 1978 Jan;88(1):17–22. doi: 10.7326/0003-4819-88-1-17. [DOI] [PubMed] [Google Scholar]
  8. Rosen K. M., Sinno M. Z., Gunnar R. M., Rahimtoola S. H. Failure of rapid atrial pacing in the conversion of atrial flutter. Am J Cardiol. 1972 Apr;29(4):524–528. doi: 10.1016/0002-9149(72)90443-2. [DOI] [PubMed] [Google Scholar]
  9. Roy D., Waxman H. L., Buxton A. E., Marchlinski F. E., Cain M. E., Gardner M. J., Josephson M. E. Termination of ventricular tachycardia: role of tachycardia cycle length. Am J Cardiol. 1982 Dec;50(6):1346–1350. doi: 10.1016/0002-9149(82)90473-8. [DOI] [PubMed] [Google Scholar]
  10. Ward D. E., Camm A. J., Spurrell R. A. The response of regular re-entrant supraventricular tachycardia to right heart stimulation. Pacing Clin Electrophysiol. 1979 Nov;2(6):586–595. doi: 10.1111/j.1540-8159.1979.tb04277.x. [DOI] [PubMed] [Google Scholar]
  11. Waspe L. E., Kim S. G., Matos J. A., Fisher J. D. Role of a catheter lead system for transvenous countershock and pacing during electrophysiologic tests: an assessment of the usefulness of catheter shocks for terminating ventricular tachyarrhythmias. Am J Cardiol. 1983 Sep 1;52(5):477–484. doi: 10.1016/0002-9149(83)90011-5. [DOI] [PubMed] [Google Scholar]
  12. Winkle R. A. The implantable defibrillator in ventricular arrhythmias. Hosp Pract (Off Ed) 1983 Mar;18(3):149-54, 156, 162 passim. doi: 10.1080/21548331.1983.11702498. [DOI] [PubMed] [Google Scholar]
  13. Yee R., Zipes D. P., Gulamhusein S., Kallok M. J., Klein G. J. Low energy countershock using an intravascular catheter in an acute cardiac care setting. Am J Cardiol. 1982 Nov;50(5):1124–1129. doi: 10.1016/0002-9149(82)90430-1. [DOI] [PubMed] [Google Scholar]
  14. Zipes D. P., Jackman W. M., Heger J. J., Chilson D. A., Browne K. F., Naccarelli G. V., Rahilly G. T., Jr, Prystowsky E. N. Clinical transvenous cardioversion of recurrent life-threatening ventricular tachyarrhythmias: low energy synchronized cardioversion of ventricular tachycardia and termination of ventricular fibrillation in patients using a catheter electrode. Am Heart J. 1982 May;103(5):789–794. doi: 10.1016/0002-8703(82)90390-8. [DOI] [PubMed] [Google Scholar]

Articles from British Heart Journal are provided here courtesy of BMJ Publishing Group

RESOURCES