Abstract
A longstanding problem in quadrupolar NMR of semi-solids is the selection of signals originating from ordered nuclei, i.e. those that experience a non-vanishing quadrupolar coupling. Established techniques, such as for example multiple-quantum filters are not adequate in situations when the radio frequency power is on the order of the quadrupolar coupling or the quadrupolar relaxation rates, such as may be the case on an MRI scanner, or in ex situ applications. In this manuscript we show a new method for the selective excitation of ordered spin-3/2 nuclei, which produces the desired results when the radio frequency power is approximately equal or smaller than quadrupolar frequency. Using a combination of simulations and experiments with 23Na in NaCl solution, Pf1-solutions, and bovine patellar cartilage samples we further show how the value of the quadrupolar frequency and global features of a quadrupolar coupling distribution can be extracted from these experiments.
Graphical abstract
Introduction
Sodium ions play important roles in human physiology. In the body, sodium ions can either exist in their free form, or restricted in their motion. Typically, the extracellular compartment contains freely moving sodium ions, with a few exceptions, such as for example cartilage tissue, where sodium ions are associated with macromolecular constructs, and motion is restricted via binding to negatively charged entitites.1,2 Anisotropic motion, or motion with partial averaging leads to the appearance of a residual quadrupolar coupling.3,4 Biexponentional relaxation is observed when sodium ions are tumbling relatively slowly, such as in the crowded intra-cellular environment5. Although the concentration of the extracellular sodium is about 10 times larger than the one of intra-cellular sodium, monitoring the latter and sodium in cartilage tissue is of greater interest for indicating physiological changes2,6. Several methods exist for diagnosing either restricted or slow motion, which include the multiple quantum filtered sequence7,18, sequences based on mixed selective and non-selective pulse sequences19,20 and a quadrupolar jump-and-return sequence21. The multiple-quantum filtered sequences, although very powerful, require strong pulses and a relatively long phase cycle for these to be useful. In cases where the quadrupolar frequency νQ is not significantly smaller than the Rabi frequency νrf, large losses and artifacts are observed, which prevent these methods from being used reliably in vivo7,22.
In this article, we describe a new method, called Quadrupole Sensitive Pulse (QSP), for the selective excitation of ordered quadrupolar spins only (i.e. those that display residual quadrupolar couplings) when the radio frequency fields (rf-field, νrf) of the pulses are approximately equal or smaller than a given quadrupolar frequency. One feature of QSP is that it is a central transition (CT) measurement and therefore it can also be used as either a preparation module or an excitation module in other sequences, such as for example in spin echo and inversion recovery sequences (the 90° pulse can be replaced with QSP for excitation of ordered spins). The second feature of this method is the dependence of the intensity of the CT on the quadrupolar frequency − CT(νQ). By performing this experiment as a function of the total duration of the sequence, we are able to determine the value of the quadrupolar frequency or global features of a quadrupolar coupling distribution.
Theory
Experiments in which the internal interactions are much smaller or much larger than νrf are well known and most of them have analytical solutions for the action of the rf-field. We consider here in particular the quadrupolar coupling, which arises from the interaction between the quadrupole moment of the nucleus (S > ½) and a gradient of the electric field23. The value of the quadrupolar frequency quantifies the electric charge anisotropy surrounding the nucleus. The quadrupole Hamiltonian in the high field approximation is also anisotropic and it depends on orientation with respect to the Zeeman field. The first order quadrupole Hamiltonian in a static solid24 can be written as:
| (1) |
in units of Hz, where we have assumed for simplicity ηQ = 0. Here β is the angle between the magnetic field and the principal axis frame of the quadrupole tensor.
Investigating the sample with an orientational parallel to the magnetic field, Eq. (1) can be simplified to
| (2) |
where we have taken β = 0.
The observed signal is the trace of the multiplication of the appropriate detection operator and the density matrix,
| (3) |
where A = S−, and it can in many cases be reduced to just Sy (e.g. in the absence of offset effects) ; ρ(t, νQ) is the density matrix.
In the case of a random distribution of principal axis frames, the signal as given by Eq. (3) is averaged over all orientations,
| (4) |
In biological tissues, the dependence of the quadrupolar coupling distribution is more complicated because of potentially anisotropic orientational distributions25. To encompass these situations, we write the more general form using averaging over a quadrupolar frequency distribution in addition to orientational averaging9,26,
| (5) |
where is the weight factor. This weight factor is often described well by a Gaussian distribution,9 and occasionally by a parabolic function12.
The spectra of sodium with parallel orientation, in a powder or in biological tissues are well known3. For our purpose, we will consider the behavior of central transitions only.
The density matrix can be written as:
| (6) |
where
| (7) |
If νrf ≫ νQ the quadrupole interaction can be neglected during the pulse and the value of the central transition depends on the flip angle of the pulse (Fig. 1, black peaks; the flip angle is π/4). If νrf ≪ νQ, the pulse influences the central transition only and therefore the quadrupole interaction can be neglected during the pulse also (Fig. 1, blue peaks). If νrf~νQ the behavior of the central transition is more complicated and depends on the quadrupolar frequency (Fig. 1, red peak).
Fig. 1.
Dependence of the intensity of the central transition (CT) on the ratio of νrf/νQ : νrf ≫ νQ (black peaks) νrf = 100 kHz; νQ = 1, 2, 3, 4, 5 (kHz); νrf~νQ (red peaks), νrf = 1 kHz; νQ = 0.5, 2, 4, 6, 8 (kHz); νrf ≪ νQ (blue peaks), νrf = 1.1 kHz; νQ = 110, 120, 130, 140, 150 (kHz). The flip angle of the pulse is π/4.
Theoretical solutions in the regime where νrf~νQ are difficult to derive, and when they exist, they are very bulky and do not provide much insight into the physics of the experiments. On the other hand, the quadrupolar couplings in biological tissues are frequently in the range of 500–3000 Hz4 and therefore, the regime νrf~νQ can be used for the selective excitation and observation of ordered sodium. The required conditions for such a pulse sequence are:
Excitation of spins only when they have a certain minimum quadrupolar coupling;
Small dependence of performance on the inhomogeneity of the external field;
Small dependence of performance on the inhomogeneity of the rf-field.
The first condition can be achieved using two adjacent pulses with the same duration and alternating phases (0° and 180°). The second and third conditions can be achieved using a π-pulse and phase cycling of the π-pulse. This approach is similar to the quadrupolar jump-and-return sequence21, except that now we are focusing on a case where the pulses cannot be used at powers that significantly exceed the quadrupolar couplings (such as for sodium imaging in vivo, for example).
The follow sequence, called Quadrupole Sensitive Pulse (QSP), satisfies all these conditions (Fig. 2): two pairs of adjacent pulses with alternating phases, and a π – pulse between them with a two-step phase cycle (0°, 180°). The durations of the pulses (except the π-pulse) are chosen as
| (8) |
Fig. 2.
The QSP sequence consisting of 5 pulses: a π-pulse in the middle and four pulses with the same lengths tp,x on both sides of the π-pulse. The phases of the pulses are: ϕ1 = 0°; ϕ2 = 180°; ϕ3 = 0°, 180°. ϕrec = 270°.
The condition in Eq. (8) was found experimentally. All experimental points under κ = 1 (in these cases the flip angles of the four pulses surrounding the 180° pulse have a flip angle of 90° or below) diverged from simulations.
The total length of the QSP is
| (9) |
As will be shown further, the value κ can be used for scaling the duration of the sequence. Acquiring the signal as a function of this scaling parameter, one can obtain the value of the quadrupolar frequency and an estimate of the quadrupolar coupling distribution.
Experimental methods
Calculations and Simulations
All simulations were performed using MATLAB. Powder integration was performed using 50 angles according to the ZCW algorithm27. Since we considered the behavior of the CT only, 50 β angles were sufficient.
Materials
Pf1 bacteriophage was purchased from ASLA Biotech and was divided among three samples: sample I was pure; samples II and III were diluted with NaCl solution (0.4 M).
Bovine patellar cartilage was received from Dickson’s Farmstand Meats (NYC). The cartilage were cut with a scalpel into the strips. The strips were incubated in phosphate buffered saline (PBS) for 3 hours and were afterwards arbitrarily placed into a quartz tube (sample IV) or a glass tube (sample V). The tubes were filled with fluorinated oil solution (FC – 84) to minimize susceptibility artifacts. After acquiring spectra from sample IV, NaCl solution was added above the fluorinert solution (fluorinert is heavier than water and hydrophobic). The aim of the adding of NaCl solution was to prove experimentally the insensitivity of the QSP to the atoms without quadrupolar coupling.
NMR experiments
90° and QSP experiments were performed using a broad-brand direct observe probe tuned to sodium on a Bruker Avance, 500 MHz spectrometer. Nutation frequencies for 23Na NMR were calibrated using a 0.4 M solution of NaCl. Table 1 lists the experimental parameters used.
Table 1.
Experimental parameters used in the 90° and QSP experiments.
| Materials | NS | SW1(kHz) | AQ2(s) | D13(s) |
|---|---|---|---|---|
| NaCl | 64 | 25 | 0.15 | 1.2 |
| Pf1 | 512 | 25 | 0.15 | 1.2 |
| Cartilage | 512 | 25 | 0.05 | 1.0 |
Spectral Width;
Acquisition time;
Recycle delay.
Simulations
In this section we show the excitation of the ordered spins with a single quadrupolar coupling, or a distribution of quadrupolar couplings under with different settings of νrf, νQ, κ. We also show here the influence of the inhomogeneities of the rf-field and the external field (the model of inhomogeneity is uniform distribution in both cases) on the excitation of these atoms, and how the nature of the quadrupolar coupling distribution can be investigated.
Single quadrupolar coupling frequency
In Fig. 3 we show the dependence of the CT on κ (Eq. (8)) for different values of the quadruple frequency [0:1500] (Hz) for parallel orientation (Eq. (2)); νrf(Hz) = 1200 (a), 500 (b) 100 (c). At νQ = 0 the CT does not depend on the values of νrf, κ and equals zero. From all three Figures we can see that with increasing quadrupolar frequency the intensity of the CT grows also. From Figs. 3b and Fig. 3c we can see that the behavior of the CT depends on κ and the ratio νQ/νrf. For example, the absolute value of the CT (the absolute value of the complex signal) for some κ when νQ = 1500, νrf = 500 (Fig. 3b, the last red thick peaks) and νQ = 300, νrf = 100 (Fig. 3c, second green thin peak) are the same. The difference between them is the influence of T2 relaxation during the pulse. In the insert of that figure we show two spectra of a 3/2 spin with 1500 Hz quadrupolar frequency which were obtained with an ideal 90° pulse (black line) and QSP (red line) under the conditions νrf = 500, κ = 2, (νQ/νrf = 3). The intensities of the CT in both cases are the same. When νQ > 6νrf the influence of the quadrupole interaction during the pulse decreases and the intensities of the CT are small (Fig. 5, thick peaks). From this Figure we can conclude that during the QSP there appear terms of first rank in-phase single quantum (T1,±1) and third rank anti-phase single quantum (T3,±1) coherences28,29. If νQ < νrf the first rank coherences contribute the most to the signal, regardless of κ. For example, the first four columns of peaks in Fig. 3b have positive signs. In the regime where νQ ≥ νrf the contribution of the third rank coherences grows and can be larger than the in-phase terms. For example, the last two columns of peaks have negative signs. Consequently, the satellite transitions are anti-phase with respect to CT (insert of Fig. 3b).
Fig. 3.
The dependence of the CT on κ [1.25; 1.5; 1.75; 2; 2.25; 2.5] (y-axis) and νQ (Hz) [0; 150; 300; 450; 600; 750; 900; 1050; 1200; 1350; 1500] (x-axis) at different values of νrf. (a): νrf = 1200 Hz. Insert: QSP curves: νQ = 150 Hz (black line); νQ = 450 Hz (red line); νQ = 600 Hz (black line with circles); (b): νrf = 500 Hz. Insert: red – QSP spectrum under conditions νrf = 500 Hz, νQ = 1500 Hz, κ = 2; black – spectrum of 1500 Hz quadrupolar frequency after ideal 90° – pulse; (c): νrf = 100 Hz.
Fig. 5.
The behaviors of the CT(κ = 1.25) and (inserts) without (blue line) and with (red line) inhomogeneity of the rf – field (Δνrf): (a) νrf = 200 Hz, νQ = 600 Hz; (b) νrf = 200 Hz, νQ = 150 Hz; (c) νrf = 500 Hz, νQ = 1200 Hz (d) νrf = 500 Hz, νQ = 600 Hz.
As described above, QSP can be used for excitation of ordered spins and can be used for obtaining the values of the quadrupolar frequencies also. Under the condition νQ < νrf the behavior of the CT for different κ is similar. It can clearly be seen in Fig. 3a for three couplings of νQ = 150; 300; 450 Hz and in Fig. 3b for two couplings with νQ = 150; 300 Hz. In the insert of Fig. 3a we show the behaviors of the CT as a function of κ for νQ = 150 Hz (black curve), νQ = 450 Hz (red curve), and νQ = 600 Hz (black curve with circles). The points are normalized by the intensity of the first point, which has the largest intensity, as represented by
| (10) |
If the curve with νQ = 150 Hz and the curve with νQ = 450 Hz are hardly distinguished, the curve with 600 Hz quadrupolar frequency can be distinguished from the curve with 450 Hz of quadrupolar frequency (Fig. 3a, insert), but the quadrupolar frequency resolution is low. When increasing the ratio of νQ/νrf, the quadrupolar frequency resolution decreases. The quadrupolar frequency resolution means the minimal value δνQ, when two curves CT(νQ; κ) and CT(νQ + δνQ; κ) can be distinguished. Under some experimental conditions this value can be about δνQ ≈ 2 – 5 Hz.
Inhomogeneity of B0 could produce some residual signal of atoms without quadrupolar couplings from the QSP sequence. In Fig. 4a we show the influence of ∓ 20 Hz of ΔB0 on a S=3/2 spectrum with νQ = 0 (black dotted line) and S=3/2 spectrum with νQ = 400 Hz (red thick line). Also we show the spectrum without inhomogeneity (blue thin line) for comparison.
Fig. 4.
The dependence of the CT (κ = 1.25) on the inhomogeneity of the external field (ΔB0) for different rf – field strengths: (a) νrf = 600 Hz; (b) νrf = 200 Hz (νQ/νrf = 3). Blue thin spectrum − νQ = 400 Hz, ΔB0 = 0; red thick spectrum − νQ = 400 Hz, ΔB0(Hz) ∈[−20:20]; black dotted spectrum νQ = 0, ΔB0(Hz) ∈[−20:20]. The intensity of the blue peaks in (a) is about 10 times larger than blue peaks in (b).
It is seen that under the conditions investigated, the inhomogeneity has a negligible influence on isotropic spins (black dotted spectrum). On the other hand, it has a significant influence on the signals from ordered spins: the value of CT under inhomogeneity is smaller by about a factor of 4 compared to CT without inhomogeneity. Interestingly, using the very weak rf-fields such as 200 Hz one can significantly decrease the influence of ΔB0 (Fig. 4b). This effect occurs because the intensity of the blue peak in Fig. 4b has no maximum value (the intensity of the blue peaks in (a) is about 10 times larger than the blue peaks in (b)). In the presence of inhomogeneity, the intensity of CT is a little increased for certain offsets, and at other offsets it is decreased, but the intensity of the total signal is not changed.
The inhomogeneity of the rf-field Δνrf does not excite isotropic spins because of the structure of QSP, but it influences the signal of the spins when quadrupolar coupling is present and the ability of obtaining accurate measures of the quadrupolar frequency. In Fig. 5 we show the influence of 10% of Δνrf (red thick lines) on CT(νQ; κ = 1.25) for νrf = 200 Hz (a, b) and νrf = 500 Hz (c, d) and compare it with CT(νQ; κ = 1.25) without inhomogeneity (blue thin lines). In the inserts of these Figures we show the behavior of the CT as a function of κ with Δνrf and without, respectively. When νrf < νQ like in Fig. 5a and in Fig.5c, the Δνrf has a significant influence on CT(νQ; κ). In the first case, inhomogeneity leads to 45% of reduction of the CT and in the second it leads to an increase by 130%.
On other hand, it is important to point out that Δνrf does not have a significant influence on curves, except on the first points (insert, Fig. 5a) and (insert, Fig. 5c). When νrf > νQ like in Fig. 5b and νrf ≈ νQ in Fig. 5d, Δνrf has a very small influence on CT and on the curves, accordingly. Therefore B1 inhomogeneity will not have a significant influence on the ability of obtaining the quadrupolar frequencies.
Distribution of quadrupolar couplings
The behavior of the CT of a powder follows the broad trends of the CT with a single coupling. In Figs. 6 and 7 we show the dependence of the absolute value of CT on κ for different values of quadrupolar frequency [25:1000] (Hz), νrf = 100 Hz (Fig. 6) and [200:4600] (Hz), νrf = 600 Hz (Fig. 7). Like in the case of a single quadrupolar coupling frequency, during the QSP there appear terms of first rank in-phase single quantum (T1,±1) and third rank anti-phase single quantum (T3,±1) coherences. Under νQ ≲ 2νrf the first rank coherences contribute the most to the signal (in Figs. 6 and 7, the first four couplings νQ = 25, 50, 100, 200 Hz and νQ = 200, 600, 1000, 1400 Hz, respectively). Under νQ ≳ 5.5νrf the third rank coherences contribute the most to the signal (in Figs. 6 and 7, the last five couplings νQ = 600, 700, 800, 900, 1000 Hz and νQ = 3000, 3400, 3800, 4200, 4600 Hz, respectively). Between these two conditions, the contributions of the first and third rank coherences are approximately equal and therefore the intensities are small (in Figs. 6 and 7, the middle three couplings νQ = 300, 400, 500 Hz and νQ = 1800, 2200, 2600 Hz, respectively). A dependence of on νQ is seen under the conditions νQ ≥ νrf (insert of Fig. 6) and the maximum of the intensity of CT under the conditions νQ/νrf = 1.25 and νQ/νrf = 7.6 can be achieved at an intensity of 58% and 29% of the original one only (insert of Fig. 7).
Fig.6.
The dependence of the CT at νrf = 100 Hz as function of κ [1.25; 1.5; 1.75; 2; 2.25; 2.5] (y-axis) and νQ (Hz) [25; 50; 100; 200; 300; 400; 500; 600;700;800; 900; 1000] (x-axis). Insert: QSP curves: νQ = 25 Hz (magenta line); νQ = 50 Hz (blue line); νQ = 100 Hz (green line).
Fig. 7.
The dependence of the CT at (νrf = 500 Hz) as a function of κ [1.25; 1.5; 1.75; 2; 2.25; 2.5] (y-axis) and νQ (Hz) [200; 600; 1000; 1400; 1800; 2200; 2600; 3000; 3400; 3800; 4200; 4600] (x-axis). Insert: black – spectrum of 750 Hz quadrupolar frequency after ideal 90° – pulse; red – QSP spectrum under conditions νrf = 600 Hz, νQ = 750 Hz, κ = 1.25; blue – QSP spectrum under conditions νrf = 600 Hz, νQ = 4600 Hz, κ = 1.75. The ratio of intensities is 1:0.58:0.29.
In Fig. 8 we show the influence of 10% of inhomogeneity of rf-field (Δνrf in the case when νQ < νrf (Fig. 8a) and in the case when νQ > νrf (Fig. 8b and c). In the first case, the influence of the inhomogeneity is negligible. In the case where νrf = 600 Hz, νQ = 1200 Hz (Fig. 10b), the inhomogeneity has only a very small influence on the curve (insert of Fig. 10b) and can lead to a small error in the extracted quadrupolar frequency (about 20–30 Hz). In the third case (Fig. 10c), the inhomogeneity has significant influence on the CT(νQ; κ). As in the case of the single quadrupolar coupling, Δνrf has a moderate influence on curves on the first points only (insert of Fig. 10c).
Fig. 8.
The behaviors of the CT(κ = 1.25) and (inserts) without (blue line) and with (red line) inhomogeneity of the rf-field (Δνrf): (a) νrf = 100 Hz, νQ = 50 Hz; (b) νrf = 600 Hz, νQ = 1200 Hz; (c) νrf = 300 Hz, νQ = 1110 Hz. The number of orientations which was used in (a) is 50 and 538 in the case of (b) and (c). For obtaining curves the number of orientations was 50.
Fig. 10.
23Na spectrum in NaCl (0.4 M) solution. Black line 90° – pulse (νrf = 1200 Hz). Other QSP(κ = 1.25): blue line (νrf = 1200 Hz, Δν = 0); magenta line (νrf = 1200 Hz, Δν = 0.5 ppm); green line (νrf = 1200 Hz, Δν = 1 ppm); blue dotted line (νrf = 200 Hz, Δν = 0 ppm). The ratio of intensities of these spectra are 100:1:1.2:1.5:2.
As mentioned above, using this method we can investigate the nature of the quadrupolar coupling distribution. We focus here on two models: uniform orientational distribution (Eq. (4)) or Gaussian quadrupolar coupling distribution (Eq. (5)). The last figure of this section show the dependence of on the width of the quadrupolar distribution [0;50;100;150;200;250;300] under different values of νrf(Hz) : 600 (a); 500 (b); 400 (c). Under 600 Hz of rf-field strength (Fig. 9a) has a very small dependence on . Under 500 Hz of rf-field strength (Fig. 9b) this dependence grows and under 400 Hz of rf-field (Fig. 9c) there is a visual difference of for different . Therefore, the experiments, in which the first point CT(νQ; κ = 1.25) has the maximum value (Fig. 9a and b) and the curves drop continuously, can be investigated with a model of a random distribution of principal axis frames (Eq. (4)) for obtaining the value of νQ. Other experiments, where the behaviors of the QSP curves are more complicate (for example, like in Fig. 9c) can be used for investigation of the nature of the quadrupolar coupling distribution.
Fig. 9.
The dependence of on : (a) νrf = 600 Hz; (b) νrf = 500 Hz; (c) νrf = 400 Hz. .
In this section we showed that QSP excites the ordered spins with non-vanishing quadrupolar coupling only. When νrf ≤ νQ the CT has a different behavior as function of κ when the quadrupolar frequencies differ, and therefore, the value of the quadrupolar frequency can be extracted from these curves. When νrf ≲ 2.5νQ the CT curves begin to depend on the nature of the quadrupolar coupling distribution also and therefore, these can also be characterized by QSP experiments. The moderate inhomogeneity of the external field does not influence isotropic spins, but it has a strong influence on the excitation of the ordered spins. The moderate inhomogeneity of the rf-field does not influence the isotropic spins significantly and does not influence the ability to determine the values of quadrupolar frequencies of the ordered spins. As shown here, in experiments where the curves drop continuously, one can use a uniform orientational model, When the behavior of the QSP curves are more complicated (for example, Fig. 9c) one can use a Gaussian quadrupolar coupling distribution.
Experimental results
We now turn to experiments where we explore the properties of the QSP experiments for their ability to selectively observe odered sodium and characterize the quadrupolar coupling distributions. Here, we use samples 23Na in Pf1, and in cartilage and contrast that to the behavior of isotropic sodium in NaCl solution. Further experiments are shown, which demonstrate that the quadrupolar couplings can be characterized even in the presence of strong background signals from glass tubes.
In Fig. 10, we show 23Na spectra of a NaCl solution (0.4 M), which demonstrate that the QSP does not excite isotropic spins under different experimental conditions. The black line is the 23Na spectrum after a regular 90° hard-pulse excitation. All four QSP spectra (color lines) are very small compared to the 90° spectrum. The ratio of intensities of the 90° spectrum and QSP spectra are 100:1:1.2:1.5:2.
In Fig. 11 we show three different 90° 23Na spectra of Pf1 bacteriophage (a) pure sample (I) and two samples diluted with NaCl (0.4 M) solution (b) – II, (c) – III. The measured quadrupolar frequency of the first sample (a) is approximately 806 Hz, of the second sample (d) it is approximately 498 Hz and of the third (c) it is approximately 152 Hz.
Fig. 11.
23Na 90° pulse spectra of Pf1 bacteriophage in different concentrations. (a) – pure sample (sample I), νrf = 1315 Hz. Diluted samples with NaCl: (b) (sample II) νrf = 1200 Hz; (c) (sample III) νrf = 200 Hz.
Fig. 12 shows the results from 6 series of QSP experiments as a function of κ : (a–c) sample I; (d–e) sample II; (f) sample III. The obtained quadrupolar frequencies are shown in Table 2.
Fig. 12.
curves of I sample (a, b, c), sample II (d, e) and sample III (f). (a) νrf = 600 Hz; (b) νrf = 310 Hz; (c) νrf = 200 Hz; (d) νrf = 600 Hz; (e) νrf = 400 Hz; (f) νrf = 200 Hz. The inserts show the logarithm of the residual from least squares fitting (Levenberg-Marquardt method) as function of quadrupolar frequency.
Table 2.
The obtained quadrupolar frequencies with direct measurements (spectrum) and QSP experiments with different νrf (Figs.11 and 12).
| sample I | spectrum | νrf = 600 Hz | νrf = 315 Hz | νrf = 200 Hz |
| νQ (Hz) | 806 | 805 | 510/815 | 815 |
| sample II | spectrum | νrf = 600 Hz | νrf = 400 Hz | |
| νQ (Hz) | 498 | 520 | 485 | |
| sample III | spectrum | νrf = 200 Hz | ||
| νQ (Hz) | 152 | 168 |
Five of the six inserts in Fig. 12 (a, c–f) show only one unique minimum of the fitting residual, log[χ2], as function νQ of and they are close to the quadrupolar frequencies which were obtained with the direct measurements. The insert of Fig. 12b shows two minima of log[χ2] at the quadrupolar frequencies of 510 Hz and 815 Hz. This effect is seen because at κ ≥ 4 the behaviors of CT(νQ = 510 Hz ; k) and CT(νQ = 815 Hz ; k) are similar. Since the first point serves as a normalized point, only one point CT(νQ; κ = 3) differentiates them. On the one hand, that problem can be solved using smaller values of κ (κ = 1.25, 1.5, 1.75), because the strongest dependence of on κ is in the range κ ∈(1:2). On the other hand, under the condition when the total time of the pulse τp~T2, the relaxation has a strong influence on the magnetization during the pulse and it leads to results which diverge from simulations30. The most natural solution of the relaxation problem would be changing the rf-field strength of adjacent pulses and not the total length of QSP by changing κ. But as Fig. 13 shows, this strategy does not work: the simulation does not fit the experimental points and the obtained quadrupolar frequency, 700 Hz, is far from the directly measured value of 498 Hz. This is an indication that biexponential relaxation, where the central and satellite transitions have different relaxation times T2,CT and T2,ST is the reason for the observed effects (Fig. 12c)31:. The central transition of the first order Hamiltonian does not depend on the quadrupolar frequency and therefore the dependence of CT on νQ occurs because of appearance satellite transitions during the QSP, which contribute to the value of CT. T2,ST of Pf1 is less than T2,CT: 8 and 45 ms, approximately32. Therefore, when the length of the QSP is too long, like in the first three points of Fig. 12c, the contribution of the satellite transitions decreases because of the influence of T2,ST. The problem of relaxation in QSP experiments when τp~T2,ST can be avoided, however, by acquiring more points between κ ∈(2:4].
Fig. 13.
curves of sample II. The length of the pulses (except the π – pulse in the middle) is 0.5 ms and the total time of the QSP is 2.83 ms. The rf-field of the π- pulse is 600 Hz.
We now investigate cartilage samples, which are known to display broad residual quadrupolar coupling distributions. Fig. 14 shows a 90° spectrum of two prepared samples: sample IV in a quartz tube (a), sample V in a glass tube (b). In the first case the width of the background spectrum is about 2 kHz and in the second case it is about 3 kHz. The difference in the widths of the spectra arises due to the contributions of the sodium signal from the glass. It can clearly be seen in the insert of Fig. 14b that the width of the background spectrum after the QSP sequence is approximately 8 kHz. As is shown below, despite of significant background noise, QSP is still able to extract quadrupolar couplings.
Fig. 14.
The 90° spectra of 23Na in Cartilage in a quartz tube (a, sample IV) and in a glass tube (b, sample V). In both cases νrf is 1300 Hz. Inserts: QSP spectrums of these samples under n=4 and νrf = 400 Hz.
In Fig. 15 we show 6 QSP curves for sample IV (a, b, c) and V (d, e, f). The experimental points were compared to the simulations with a uniform orientational distribution (Eq. (4)). The obtained quadrupolar frequencies are shown in Table 3.
Fig. 15.
curves of sample IV (a, b, c) and sample V (d, e, f). (a) νrf = 600 Hz; (b) νrf = 500 Hz; (c) νrf = 400 Hz; (d) νrf = 500 Hz; (e) νrf = 400 Hz; (f) νrf = 300 Hz. The inserts show the logarithm of the residual from least squares fitting as a function of quadrupolar frequency. The number of the powder orientations for the simulations is 50.
Table 3.
The obtained quadrupolar frequencies with QSP – experiments under different νrf (Fig. 15).
| sample IV | νrf = 600 Hz | νrf = 500 Hz | νrf = 400 Hz | average (Hz) |
| νQ (Hz) | 940 | 1000 | 920 | 953 |
| sample V | νrf = 500 Hz | νrf = 400 Hz | νrf = 300 Hz | average |
| νQ (Hz) | 1080 | 1010 | 1110 | 1067 |
The behaviors of the first experimental points in Fig. 15f differ from simulation, which is likely due to relaxation effects (T2,ST), which were described in the case of Pf1. The average quadrupolar frequencies of sample IV (quartz) and sample V (glass) are close to each other: 953 Hz and 1067 Hz, respectively. The reason of the small divergence is due to signal of sodium in glass.
It is seen that averaging over a random distribution of principal axis frames produces acceptable results (Eq. (4)). Cartilage tissues often consist of compartments among which the ions exchange very slowly9 and the measured signal is described by Eq. (5). Because of the way the samples were prepared (three weeks storage of the cartilage in the refrigerator before experiments, cutting of the cartilage with a scalpel into strips and random packing into tubes), it is reasonable to assume such a description of the distribution of couplings. Therefore, these results suggest that the distribution of orientations can be characterized by this pulse sequence.
In Fig. 16a, we show the investigation of a sample of cartilage with additional added NaCl solution. We show the 90° spectrum of sample IV with a streak of NaCl over a fluorinated oil solution (red spectrum) and the spectrum without NaCl (black spectrum). The ratio of the black peak to the red peak is about 1:4. Comparing the intensities of CT(κ = 1.25) of sample IV, sample IV + NaCl (not shown here) and Fig. 16a, we find that the ratio of ordered sodium in cartilage and free sodium in NaCl in that sample was ~1:6.5. In Fig. 16b we show measured QSP curves of this sample at an rf-field of 600 Hz. The extracted quadrupolar frequency with and without added NaCl solution are close to each other: 1050 Hz and 940 Hz, respectively. The reasons of the small divergence are could be due to sodium diffusion in NaCl solution and inhomogeneity of the rf-field.
Fig. 16.
(a) The 90° spectra of 23Na in Cartilage in a quartz tube (sample I) with added NaCl (red line) and comparison with a sample without NaCl (black line). (b) curve of sample I + NaCl. νrf = 600 Hz.
In this section we showed QSP experiments with 23Na in NaCl solution, Pf1 solutions and cartilage samples. The QSP has negligible influence on the isotropic spins. The quadrupolar frequencies of all Pf1 samples, which were obtained by QSP experiments are close to the frequencies with direct measurements. The quadrupolar frequencies of the all cartilage samples are close to each other. The experiments with cartilage samples in a glass tube and cartilage with NaCl showed that this sequence does not require special conditions for conducting the experiments. Also the quantitative explanation is provided about influence of the spin-spin relaxation on obtaining of the quadrupolar couplings. The relaxation effects have significant influences under τp~T2,ST and they can be reduced using more short pulses κ > 2 and taking more points in the area of κ ∈ (2:4].
Conclusion
Currently, all known methods for quantifying quadrupolar couplings of sodium in tissues require the use of relatively hard pulses and often extensive phase cycles. On an MRI scanner, or in ex situ applications, one often cannot use such ideal pulses. As a result, the sensitivity in double-quantum filtered experiments is relatively low and these sequences do not perform according to theory in these situations. Furthermore, broad background signals from more solid components could contaminate spectra and prevent accurate quantification. In this article, we have demonstrated a new method for the selective excitation of ordered sodium signals using weak pulses only. Under conditions where νQ = 3νrf and κ = 2 the intensity of the CT after QSP can reach the same value as the intensity of the CT after a 90° pulse in the case of a single quadrupolar coupling frequency. For a random distribution of principal axis frames under conditions where νQ = 1.25νrf and κ = 1.25, the intensity of the CT after QSP can achieve about 58% of the intensity of the CT after 90° pulse. Because of the structure of the sequence, the spatial inhomogeneity of the rf-field does not lead to the excitation of the isotropic spins and has a small influence on ordered spins (within 10%). Furthermore, QSP could be used as a readout step in inversion recovery or as an excitation module in spin echo sequences, or simply for the selective excitation of ordered spins only. We have outlined here how the parameters of the sequence would be adjusted in order to find optimal conditions (νrf; κ) for a specific range of quadrupolar frequencies. Since all this can be achieved with low power, these techniques could become useful under in vivo conditions, or in ex situ applications.
Highlights.
We present a new method for the selective excitation of ordered quadrupolar spins only and for the measurement of the quadrupole frequency when the radio frequency fields of the pulses are approximately equal or smaller than quadrupole frequency.
The Quadrupole Sensitive Pulse does not excite nuclei with vanishing quadrupolar interaction under moderate inhomogeneity of the radio frequency and external fields.
The performance is experimentally verified on samples of 23Na in NaCl solution, Pf1, and cartilage.
Acknowledgments
The Bruker Avance – 500 NMR Spectrometer was acquired through the support of the National Science Foundation under Award Number CHE 01162222. We acknowledge funding from the U.S. National Science Foundation, award No. CHE 1412064 for methodology development, and NIH R01AR060238 and NIH R01EB016045 for tissue-related work. We also thank Dr. Uzi Eliav from Tel Aviv University for pointing us to a number of helpful references.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- 1.Linzell JL, Peaker M. Intracellular concentrations of sodium, potassium and chloride in the lactating mammary gland and their relation to the secretory mechanism. J. Physiol. 1971;216:683–700. doi: 10.1113/jphysiol.1971.sp009547. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Madelin G, Lee J, Regatte RR, Jerschow A. Sodium MRI : Methods and applications. Program. 2014;79:14–47. doi: 10.1016/j.pnmrs.2014.02.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Rooney WD, Springer J. A Comprehensive Approach to the Analysis and Interpretation of the Resonances of Spins 3 / 2 from-Living Systems. NMR Biomed. 1991;4:209–226. doi: 10.1002/nbm.1940040502. [DOI] [PubMed] [Google Scholar]
- 4.Eliav U, Navon G. Sodium NMR/MRI for anisotropic systems. NMR Biomed. 2015 doi: 10.1002/nbm.3331. [DOI] [PubMed] [Google Scholar]
- 5.Chung C, Wimperis S. Optimum Detection of Spin-3/2 Biexponential Relaxation Using Multiple-Quantum Filtration Techniques. J. Magn. Reson. 1990;88:440–447. [Google Scholar]
- 6.Keinan-Adamsky K, et al. 23Na and 2H magnetic resonance studies of osteoarthritic and osteoporotic articular cartilage. Magn. Reson. Med. 2010;64:653–661. doi: 10.1002/mrm.22479. [DOI] [PubMed] [Google Scholar]
- 7.Kemp-Harper R, Brown SP, Styles P, Wimperis S. In vivo NMR of sodium ions in ordered environments. J. Magn. Reson. B. 1994;105:199–203. doi: 10.1006/jmrb.1994.1124. [DOI] [PubMed] [Google Scholar]
- 8.Kemp-Harper R, Wimperis S. Detection of the Interaction of Sodium Ions with Ordered Structures in Biological Systems. Use of the Jeener-Broekaert Experiment. J. Magn. Reson. B. 1993;102:326–331. [Google Scholar]
- 9.Eliav U, Navon G. Analysis of double-quantum-filtered NMR spectra of 23Na in biological tissues. J. Magn. Reson. B. 1994;103:19–29. doi: 10.1006/jmrb.1994.1003. [DOI] [PubMed] [Google Scholar]
- 10.Shinar H, Knubovets T, Eliav U, Navon G. Sodium interaction with ordered structures in mammalian red blood cells detected by Na-23 double quantum NMR. Biophys. J. 1993;64:1273–1279. doi: 10.1016/S0006-3495(93)81492-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Eliav U, Navon G. Quadrupole-Echo Techniques in Multiple-Quantum-Filtered NMR Spectroscopy of Heterogeneous Systems. J. Magn. Reson. A. 1995;115:241–253. [Google Scholar]
- 12.Brown SP, Wimperis S. Extraction of Homogeneous 23 Na NMR Linewidths from Two-Dimensional Jeener – Broekaert Spectra*. J. Magn. Reson. 1995;109:291–300. [Google Scholar]
- 13.Pekar J, Leigh JS. Detection of biexponential relaxation in sodium-23 facilitated by double-quantum filtering. J. Magn. Reson. 1986;69:582–584. [Google Scholar]
- 14.Brown SP, Wimperis S. NMR measurement of spin-3/2 transverse relaxation in an inhomogeneous B1 field. Chem. Phys. Lett. 1994;224:508–516. [Google Scholar]
- 15.Hughes CE, Kemp-Harper R, Styles P, Wimperis S. NMR Spectroscopy and Imaging of Sodium in Ordered Environments. The Return of the Central Transition. J. Magn. Reson. B. 1996;111:189–193. doi: 10.1006/jmrb.1996.0081. [DOI] [PubMed] [Google Scholar]
- 16.Lee J, Regatte RR, Jerschow A. Optimal control NMR differentiation between fast and slow sodium. Chem. Phys. Lett. 2010;494:331–336. [Google Scholar]
- 17.Tsang A, Stobbe RW, Beaulieu C. Triple-quantum-filtered sodium imaging of the human brain at 4.7 T. Magn. Reson. Med. 2012;67:1633–1643. doi: 10.1002/mrm.23147. [DOI] [PubMed] [Google Scholar]
- 18.Eykyn TR, et al. Multiple quantum filtered 23Na NMR in the Langendorff perfused mouse heart: Ratio of triple/double quantum filtered signals correlates with [Na]i. J. Mol. Cell. Cardiol. 2015;86:95–101. doi: 10.1016/j.yjmcc.2015.07.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Choy J, Ling W, Jerschow A. Selective detection of ordered sodium signals via the central transition. J. Magn. Reson. 2006;180:105–109. doi: 10.1016/j.jmr.2006.01.011. [DOI] [PubMed] [Google Scholar]
- 20.Eliav U, Keinan-Adamsky K, Navon G. A new method for suppressing the central transition in I=3/2 NMR spectra with a demonstration for 23Na in bovine articular cartilage. J. Magn. Reson. 2003;165:276–281. doi: 10.1016/j.jmr.2003.09.002. [DOI] [PubMed] [Google Scholar]
- 21.Lee J, Regatte RR, Jerschow A. Selective detection of ordered sodium signals by a jump-and-return pulse sequence. J. Magn. Reson. 2009;200:126–129. doi: 10.1016/j.jmr.2009.06.015. [DOI] [PubMed] [Google Scholar]
- 22.Reddy R, Bolinger L, Shinnar M, Noyszewski E, Leigh JS. Detection of residual quadrupolar interaction in human skeletal muscle and brain in vivo via multiple quantum filtered sodium NMR spectra. Magn. Reson. Med. 1995;33:134–139. doi: 10.1002/mrm.1910330121. [DOI] [PubMed] [Google Scholar]
- 23.Pound RV. Nuclear Electric Quadrupole Interactions in Crystals. Phys. Rev. 1950;79:685–702. [Google Scholar]
- 24.Slichter CP. Principles of Magnetic Resonance. 1996:485–502. [Google Scholar]
- 25.Halle B. Theory of intramolecular spin relaxation by translational diffusion in locally ordered fluids. Mol. Phys. 1987;60:319–370. [Google Scholar]
- 26.Kemp-Harper R, Wickstead B, Wimperis S. Sodium ions in ordered environments in biological systems: analysis of 23Na NMR spectra. J. Magn. Reson. 1999;140:351–362. doi: 10.1006/jmre.1999.1834. [DOI] [PubMed] [Google Scholar]
- 27.Conroy H. Molecular Schrödinger Equation. VIII. A New Method for the Evaluation of Multidimensional Integrals. J. Chem. Phys. 1967;47:5307–5318. [Google Scholar]
- 28.Kemp-Harper R, Brown SP, Hughes CE, Wimperis S. 23Na NMR methods for selective observation of sodium ions in ordered environments. Prog. Nucl. Magn. Reson. Spectrosc. 1997;30:157–181. [Google Scholar]
- 29.Vega S, Naor Y. Triple quantum NMR on spin systems with I = 3/2 in solids. J. Chem. Phys. 1981;75:75–86. [Google Scholar]
- 30.Shekar SC, Tang JA, Jerschow A. Dynamics of I=3/2 Nuclei in Isotropic Slow Motion, Anisotropic and Partially Ordered Phases. Concepts Magn. Reson. Part A. 2010;36A:362–387. [Google Scholar]
- 31.Sobieski DN, Krueger NR, Vyas S, Augustine MP. Nuclear spin relaxation of sodium cations in bacteriophage Pf1 solutions. J. Chem. Phys. 2006;125:244509. doi: 10.1063/1.2399523. [DOI] [PubMed] [Google Scholar]
- 32.Lee J, Regatte RR, Jerschow A. Optimal excitation of N 23 a nuclear spins in the presence of residual quadrupolar coupling and quadrupolar relaxation. J. Chem. Phys. 2009;131:1–8. doi: 10.1063/1.3253970. [DOI] [PubMed] [Google Scholar]

















