Abstract
Accurate determination of in-vivo light fluence rate is critical for preclinical and clinical studies involving photodynamic therapy (PDT). This study compares the longitudinal light fluence distribution inside biological tissue in the central axis of a 1 cm diameter circular uniform light field for a range of in-vivo tissue optical properties (absorption coefficients (μa) between 0.01 and 1 cm−1 and reduced scattering coefficients (μs’) between 2 and 40 cm−1). This was done using Monte-Carlo simulations for a semi-infinite turbid medium in an air-tissue interface. The end goal is to develop an analytical expression that would fit the results from the Monte Carlo simulation for both the 1 cm diameter circular beam and the broad beam. Each of these parameters is expressed as a function of tissue optical properties. These results can then be compared against the existing expressions in the literature for broad beam for analysis in both accuracy and applicable range. Using the 6-parameter model, the range and accuracy for light transport through biological tissue is improved and may be used in the future as a guide in PDT for light fluence distribution for known tissue optical properties.
Keywords: MC, PDT, semi-infinite medium, analytical expression
1. INTRODUCTION
When treating the surface lesions with PDT, it is often necessary to treat a lesion size as small as 1-cm in diameter, e.g., all our mice studies for treatment response to PDT use a collimator field of such a circular field [1, 2]. The light fluence rate for 1 cm circular field reduces significantly from that for an infinite large field because of the reduced photon scattering by tissue [3, 4]. The purpose of this study is to determine the longitudinal distribution of light fluence rate for the 1 cm diameter field and expressed the results as an analytical function of tissue optical properties (μa, μs’). We used Monte-Carlo simulation because the diffusion theory is not valid when the lateral dimension of beam geometry becomes comparable to the mean-free-path of the photons or when the region of interest is near the air-tissue interface. Literature search reveals only analytical expression for broad beams, when the lateral dimension is much larger than the mean-free path [5]. For broad beams, dependence on the two dimensional tissue optical properties (μa, μs’) can be simplified as a one-dimensional function of albedo, a’ = μs’/(μa+μs’), or diffuse reflectance, Rd, so long as all spatial dimensions are scaled by the mean-free path (l = 1/(μa+μ’s)) [5].
2. MATERIALS AND METHODS
2.1 MC simulation
The setup geometry to be calculated by Monte-Carlo simulation for a semi-infinite medium with uniform optical properties, i.e. the absorption coefficient μa, the scattering coefficient μs, the scattering anisotropy g (= 0.9), and the index of refraction n (1.4 for tissue) is shown in Fig. 1a. The outside medium is air (n0 =1). The light field is parallel and uniform inside a circle with radius R and incident toward the air-tissue interface. (R = 0.5 cm or 10 cm for broad beam except for when μa = 0.01 and μs’= 2 cm−1, where R = 30 cm was used.) Most tissues have a preferential scattering in the forward direction with a scattering anisotropy g = 0.90 [6]. In the diffusion approximation, the tissue scattering is converted to isotropic conditions (g = 0) and the tissue scattering property is described by a reduced scattering coefficient μs’ = μs (1−g). To improve the scoring efficiency, Monte-Carlo simulation is performed for a pencil-beam incident on a semi-infinite tissue phantom (Fig. 1b). Details of the Monte-Carlo simulation process are described elsewhere [3]. The range of tissue optical properties (μa between 0.01 and 1 cm−1 and μs’ between 2 and 40 cm−1) is based on a review on the in-vivo tissue optical properties [7].
Figure 1.
(a) Setup geometry for Monte-Carlo simulation for a semi-infinite turbid medium. The light fluence rate is calculated along the central-axis on the center of the field. (b) Pencil beam setup is actually used for the Monte Carlo simulation. The ring scoring voxel has radius width dr and thickness dz at depth z and radius r in a cylindrical geometry.
2.2 Analytical expression and fitting algorithm
For 1-cm diameter circular fields, we found that Eq. (1) can fit the resulting MC data very well, typically to a depth of at least 0.5 cm or a few mean-free-path, l, whichever is larger. This fitting equation is termed 6-parameter fit since it includes 6 fitting parameters: λ1, λ2, λ3, b, C2, C3:
| (1) |
where ϕair = P/A, A = πR2 is called in-air fluence rate that can be calculated using the total power, P (in mW), and the treatment area, A, of the collimated beam. For a broad beam, existing literature shows that a 4-parameter fit (λ1, λ2, b, C2) is sufficient to fit the MC data [5]:
| (2) |
We have modified the original form (ϕ / ϕair = C'2 e−λ2d − C'1e−λ1d) [5] to be Eq. (2) because the term (1 − b · e−λ1d) that fits the buildup region only can be taken out of the final results, and one will automatically get the analytical expression for diffusion approximation, i.e., C2 is the backscattering coefficient and λ2 = μeff is the effective attenuation coefficient.
The non-linear optimization routine was performed in Matlab by minimizing a standard deviation according to:
| (3) |
Where ϕfit is based on either Eq. 1 and/or Eq. 2 to result in either 6 or 4 parameters as the results of optimization. ϕMC is the MC simulation result as a function of discrete depth zi, i = 1, 2, …, N. Typical value of zi is between 0 and 2 cm in 0.01 cm steps, i.e., N = 200. Multi-variable optimization algorithm uses a deferential convolution algorithm described elsewhere [8].
2.3 Relationship between fitting parameters and the tissue optical properties
For broad beams, we have found that all four fitting parameters are a function of Rd (or a’) as used by Jacques [5]. The two fitting parameters (λ1 and λ2) should be proportional to μeff. We have found the following relationships:
| (4) |
| (5) |
| (6) |
| (7) |
For 1-cm diameter beam, we found that that it is still possible to express the parameters λ1 as a function of Rd and the parameters λ2/μeff as a function of μeff but the λ2 value is larger than the corresponding values for the broad beam. It is not possible to express the other fitting parameters as a function of Rd alone. We have found the following relationship between them to optical properties (μa, μs’) as a two-dimensional function:
| (8) |
| (9) |
| (10) |
| (11) |
| (12) |
| (13) |
3. RESULTS AND DISCUSSIONS
Tables 1 and 2 show the summary results of fitting the MC simulation results for a board beam (10 cm diameter for all tissue optical properties except for μa = 0.01 cm−1 and μs’ = 2 cm−1, where a 30 cm diameter circular beam was used) and a 1-cm diameter beam, respectively, for depth up to 3.5 cm deep. For 1-cm diameter circular beam, C3 approaches zero for a subset of data (μa ≥ 0.1 cm−1) so that only a 4-parameter fit is necessary. In this case, the value of λ3 is unnecessary and its value is expressed as “ —” in Table 2.
Table 1.
Summary of the 4 fitting parameters (b, λ1, λ2, C2) using Eq. 2 to the MC calculated ϕ/ϕair for broad beam for optical properties (μa, μs’). MC calculated diffuse reflectance Rd is also listed.
| μa | Rd | b | λ1 | λ2 | C2 | |
|---|---|---|---|---|---|---|
| 01 | 2 | 0.6907 | 0.23 | 4.82 | 0.25 | 7.10 |
| 0.01 | 10 | 0.8317 | 0.16 | 16 | 0.55 | 7.89 |
| 0.01 | 23.5 | 0.8777 | 0.16 | 30 | 0.85 | 8.2 |
| 0.01 | 34.5 | 0.8936 | 0.15 | 38 | 1.03 | 8.24 |
| 0.01 | 40 | 0.8991 | 0.14 | 42 | 1.12 | 8.27 |
| 0.1 | 2 | 0.3589 | 0.29 | 4.21 | 0.79 | 5.1 |
| 0.1 | 10 | 0.6042 | 0.2 | 22.3 | 1.74 | 6.54 |
| 0.1 | 23.5 | 0.7084 | 0.2 | 52.6 | 2.67 | 7.19 |
| 0.1 | 34.5 | 0.7475 | 0.18 | 80.6 | 3.23 | 7.36 |
| 0.1 | 40 | 0.7611 | 0.18 | 89.7 | 3.48 | 7.46 |
| 0.5 | 2 | 0.1334 | 0.45 | 3.81 | 1.86 | 3.69 |
| 0.5 | 10 | 0.3589 | 0.29 | 22.1 | 3.93 | 5.08 |
| 0.5 | 23.5 | 0.4947 | 0.25 | 52.9 | 5.98 | 5.93 |
| 0.5 | 34.5 | 0.5518 | 0.24 | 77.4 | 7.23 | 6.30 |
| 0.5 | 40 | 0.5730 | 0.24 | 90.62 | 7.78 | 6.43 |
| 1 | 2 | 0.0710 | 0.52 | 3.74 | 2.78 | 3.25 |
| 1 | 10 | 0.2520 | 0.36 | 21.66 | 5.63 | 4.42 |
| 1 | 23.5 | 0.3848 | 0.3 | 52.95 | 8.5 | 5.29 |
| 1 | 34.5 | 0.4463 | 0.29 | 76.26 | 10.28 | 5.71 |
| 1 | 40 | 0.4694 | 0.25 | 89.21 | 11.04 | 5.77 |
Table 2.
Summary of the 6 fitting parameters (b, λ1, λ2, C2, λ3, C3) using Eq. 1 to the MC calculated ϕ/ϕair for 1-cm diameter circular beam for optical properties (μa, μs’).
| μa | b | λ1 | λ2 | λ3 | C2 | C3 | |
|---|---|---|---|---|---|---|---|
| 0.01 | 2 | 0.844 | 3.23 | 2.74 | 0.684 | 5.27 | 0.494 |
| 0.01 | 10 | 0.4271 | 16.8 | 2.67 | 0.749 | 6.591 | 0.467 |
| 0.01 | 23.5 | 0.33785 | 31.05 | 2.76 | 0.964 | 7.849 | 0.496 |
| 0.01 | 34.5 | 0.2486 | 42.3 | 2.85 | 1.11 | 8.163 | 0.505 |
| 0.01 | 40 | 0.2293 | 46.9 | 2.89 | 1.162 | 8.254 | 0.49 |
| 0.1 | 2 | 0.6785 | 2.22 | 2.55 | 0.811 | 5.149 | 0.178 |
| 0.1 | 10 | 0.2904 | 16.6 | 3.22 | 0.98 | 5.927 | 0.118 |
| 0.1 | 23.5 | 0.2302 | 33.8 | 3.86 | 1.149 | 7.231 | 0.0735 |
| 0.1 | 34.5 | 0.19 | 44.4 | 4.27 | — | 7.535 | 0.00122 |
| 0.1 | 40 | 0.1803 | 50.7 | 4.4 | — | 7.614 | 0.00117 |
| 0.5 | 2 | 0.6638 | 2.85 | 3.14 | — | 4.26 | 0.00485 |
| 0.5 | 10 | 0.357 | 17.4 | 4.87 | — | 5.19 | 0.00115 |
| 0.5 | 23.5 | 0.2618 | 34 | 6.57 | — | 6.175 | 0.002819 |
| 0.5 | 34.5 | 0.2354 | 48.2 | 7.66 | — | 6.525 | 0.000385 |
| 0.5 | 40 | 0.2252 | 55.2 | 8.15 | — | 6.64 | 0.000172 |
| 1 | 2 | 0.6986 | 3.25 | 3.79 | — | 3.876 | 0.000543 |
| 1 | 10 | 0.4081 | 16.5 | 6.33 | — | 4.699 | 0 |
| 1 | 23.5 | 0.3068 | 40 | 8.84 | — | 5.472 | 0 |
| 1 | 34.5 | 0.2812 | 58.1 | 10.5 | — | 5.845 | 0 |
| 1 | 40 | 0.2513 | 69.8 | 11.2 | — | 5.887 | 0 |
Figure 2 show the fitting results for the 4-parameters used for broad beam and Eqs. 4 – 7. The fitting is performed using the curve fitting tool from Matlab global optimization toolbox (cftool.m). R2 of the fitting results are all very good (R2 > 0.98). For broad beam, as pointed out in previous publication [5], the diffuse reflectance can be used in place of the pairs of optical properties (μa, μs’). The Rd values are obtained from MC simulation [3, 4] and are listed in Table 1 and Figure 3. An analytical expression based on the same form as a diffusion approximation [9] can be determined between Rd and the albedo, a’ as:
| (14) |
| (15) |
Figure 2.
Fitting results between the 4 parameters (b, C2, λ1/λeff, λ2/μeff) and the diffuse reflectance (Rd) for broad beams. Solid lines are Eqs. 4–7 for their respective parameters, black circles represent fitting results using 4 parameters from Table 1, and red triangles represent parameters using formulas from Ref. [5]. R2 refers to the goodness of fit between the fitting results of each parameter to their respective equation.
Figure 3.
Relationship between Rd and a’/(1−a’), where the albedo a’ = μs’/(μa+μs’). R2 refers to the goodness of fit between the Rd and Eq. 14.
Figure 3 shows the goodness of the fit between Rd and a’/(1−a’) using Eq. 14.
Figure 4 shows the fitting results for the 6-parameters used for the 1-cm diameter circular beam and Eqs. 8–13. Notice that the scaling theorem is no longer in existence so that most of the parameters have to be expressed as a function of two dimensional function of μa and μs’. R2 of the fitting results are also very good (R2 > 0.97). It is also found that 6 parameters are necessary to fit the data instead of 4 parameters for some tissue optical properties, particularly those with low abosrption coefficients (μa < 0.1 cm−1). We actually started with 4 parameter fitting to the MC data and then introduce the two additional parameters (C3, λ3) in Eq. 1 to improve fitting results to the MC data.
Figure 4.
Fitting results between the 6 parameters (b, C2, C3, λ1/μeff, λ2, λ3) and the optical properties (μa and μs’) for 1-cm circular beam. Solid lines are Eqs. 8–13 for their respective parameters, black circles represent fitting results using 6 parameters from Table 2. R2 refers to the goodness of fit between the fitting results of each parameter to their respective equation.
A comparison between the 4 parameters between 1-cm diameter beam and the broad beam for the same optical properties shows that λ2 is always larger for 1-cm beam than that for the broad beam. C2 is quite similar between 1-cm beam and broad beam for most tissue optical properties except for μa = 0.01 cm−1. We used a different expresion (Eq. 10) for the 1-cm diameter circular beam than that for the broad beam (Eq. 6).
Figure 5 shows the comparison between MC simulation results of ϕ/ϕair for (a) broad beam and (b) 1-cm diameter beam using the fitting parameters in Tables 1 and 2, respectively, as well as Eqs. 4–7 and Eqs. 8–13. For comparison, we have also included the fitting results from Ref.[5] for the broad beam.
Figure 5.
ϕ/ϕair for (a) broad beam and (b) 1-cm diameter beam for three tissue optical properties (μa, μs’) = (i) (1, 40), (ii) (0.1,10), and (iii) (0.01, 2) cm−1.
Figure 6 shows the relative deviation of the fitting and the MC simulation results for (a) broad beam and (b) 1-cm diameter beam, defined as the difference between the fit and the MC result relative to the maximum fluence rate.
Figure 6.
Relative deviation of the fitting results to the MC simulation data for (a) broad beam and (b) 1-cm beam for three tissue optical properties (μa, μs’) = (i) (1, 40), (ii) (0.1, 10), and (iii) (0.01, 2) cm−1. Data for (i) and (ii) are vertically shifted for clarity: +25% for (ii) and +30% for (i) for (a) and +25% for (ii) and +30% for (i) for (b), respectively.
For broad beam, Eqs. 4–7, fits all MC data to a maximum error of ±2% for all depths, while parameters from Table 1 fits the MC data to within 1% for all depths. Parameters from Ref. [5] fit most of the data except for μa = 0.01 cm−1, where 20% error is observed. Among the difference between our MC data and Jacque’s fit, 3% can be attributed to difference in the value of index of refractions used for the MC simulation: 1.4 and 1.38 between us and Jacques, as well as other subtle differences in MC simulation. For the 1-cm diameter circular beam, the 6 parameter fitting using Eqs. 8–13, can fit the MC data to a maximum error of ±8% for depth up to 3.5 cm for all optical properties considered in this study, while parameters from Table 2 fits the MC data to a maximum error of ±2%. It is clear that 4 parameter fits (Eqs. 4 – 7) cannot fit the MC data very well for 1-cm diameter beams for low absorption coefficients, yielding a maximum error of over 22% for μa = 0.01 cm−1 and μs’ = 2 cm−1.
One additional advantage of our expression (Eq. 1) is that the fitting based on diffusion approximation is automatically obtained by dividing it by (1-be−λ1d). The results are shown in Fig. 7 for (a) broad beam and (b) 1-cm diameter circular beam, respectively. For broad beam, Eqs. 4 – 7 without the buildup region fits the data very well. For 1-cm circular beam, the buildup region term have a larger impact on the light fluence rate distribution on the central axis, extending up to 0.5 cm depth. The fit is not good for low absorption coefficients (μa = 0.01 cm−1, see Fig. 7b)
Figure 7.
(a) ϕ/ϕair for broad beam with MC simulation and 4-parameter fitting using Eqs. 4 – 7 without the buildup term and (b) ϕ/ϕair for 1-cm diameter beam with MC simulation and 6-parameters fitting using Eqs. 8 – 13 without the buildup region for three tissue optical properties (μa, μs’) = (i) (1, 40), (ii) (0.1, 10), and (iii) (0.01,2) cm−1.
Figure 8 shows the relative deviation between the fit and the MC simulation, normalized to the maximum MC calculated fluence rate. Beyond the buildup region, the maximum error is 1% for the broad beam and 12% for the 1-cm circular beam for depth up to 1 cm. Within the buildup region, the maximum error is 23% for the broad beam and 120% for the 1-cm circular beam.
Figure 8.
(a) Relative deviation of the 4-parameter fitting using Eq. 2 without the buildup region to the MC simulation data for broad beam, (b) Relative standard deviation of the 6-parameter fitting using Eq. 1 without the buildup region to the MC simulation data for 1-cm beam.
4. CONCLUSION
We have performed MC simulation for air-tissue interface and provided an expression for the longitudinal light fluence rate distribution on the central axis with an accuracy of better than a relative standard deviation of 2% for broad beams and 8% for the 1-cm diameter circular beams for the full range of the in-vivo tissue optical properties studied. An analytical expression is provided to determine the parameters as a function of tissue properties. We have also independently validated the analytical fitting result for broad beams.
Acknowledgments
This work is supported by grants from the National Institute of Health (NIH) R01 CA154562 and P01 CA87971.
REFERENCES
- 1.Kim MM, Penjweini R, Zhu TC. In vivo outcome study of BPD-mediated PDT using a macroscopic singlet oxygen model. Proc. SPIE. 2015;9308:93080A. doi: 10.1117/12.2077803. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Penjweini R, Kim MM, Zhu TC. In-vivo outcome study of HPPH mediated PDT using singlet oxygen explicit dosimetry (SOED) Proc. SPIE. 2015;9308:93080N. doi: 10.1117/12.2076441. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Zhu TC, Dimofte A, Hahn SM, Lustig RA. Light Dosimetry at Tissue Surfaces for Small Circular Fields. Proc. SPIE. 2003;4952:56–67. doi: 10.1117/12.474144. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Zhu TC, Finlay J, Dimofte CA, Hahn SM. Light Dosimetry at Tissue Surfaces for Oblique Incident Circular Fields. Proc. SPIE. 2004;5315:113–124. doi: 10.1117/12.529827. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Jacques SL. Light Distributions from Point, Line and Plane Sources for Photochemical Reactions and Fluorescence in Turbid Biological Tissues. Photochem. Photobiol. 1998;67(1):23–32. [PubMed] [Google Scholar]
- 6.Cheong WF, Prahl SA, Welch AJ. A review of the optical properties of biological tissues. IEEE J. Quant. Electron. 1990;26:2166–2185. [Google Scholar]
- 7.Sandell JL, Zhu TC. A review of in-vivo optical properties of human tissues and its impact on PDT. J Biophotonics. 2011;4(11):773–787. doi: 10.1002/jbio.201100062. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Zhu TC, Bjarngard BE, Xiao Y, Bieda MR. Output ratio in air for irregular fields shaped by MLC. Mecl. Phys. 2004;31:2480–2490. doi: 10.1118/1.1767052. [DOI] [PubMed] [Google Scholar]
- 9.Dimofte A, Zhu TC, Finlay J, Cullighan CM, Edmonds CE, Friedberg JS, Cengel K, Hahn SM. In-vivo Light Dosimetry for Pleural PDT. Proc. SPIE. 2009;7164:71640A. doi: 10.1117/12.809548. [DOI] [PMC free article] [PubMed] [Google Scholar]








