Predictive value of ST segment elevation in cardiac pacing.

A G Arnold
PMCID: PMC482421  PMID: 7426204

Abstract

In 366 pacemaker patients the influence of intracardiac ST segment elevation on the fate (displacement or perforation) of permanent cardiac pacemaker electrodes was studied. In one group of patients a retrospective analysis was carried out, the operator being aware of the degree of ST segment elevation and positioning the electrode to avoid very high or vey low values wherever possible. In the second group a prospective trial was carried out. The final electrode position was determined solely by fluoroscopy and electrical threshold, and the ST segment shift measured afterwards. The combined rates of electrode displacement and perforation did not differ between these groups. High levels of ST segment elevation, greater than 10 mV, were found to predispose to electrode perforation. On the other hand, low ST segments did not predict a displacement. It is concluded that the ideal positioning technique should seek a low electrical threshold, a stable position, and avoid high ST elevation.

Full text

PDF
416

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Brewster G. M., Evans A. L. Displacement of pacemaker leads--a 10-year survey. Br Heart J. 1979 Sep;42(3):266–270. doi: 10.1136/hrt.42.3.266. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Chatterjee K., Sutton R., Layton C. A., Edwards A. The cavity electrocardiogram in emergency artificial pacing. Postgrad Med J. 1969 Nov;45(529):713–717. doi: 10.1136/pgmj.45.529.713. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Dancy M., Jackson G., Lau O. J., Farnsworth A. Successful treatment of myocardial perforation and tamponade after temporary endocardial pacing. Br Med J. 1978 Jan 14;1(6105):79–80. doi: 10.1136/bmj.1.6105.79-a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Furman S., Hurzeler P., DeCaprio V. The ventricular endocardial electrogram and pacemaker sensing. J Thorac Cardiovasc Surg. 1977 Feb;73(2):258–266. [PubMed] [Google Scholar]
  5. Gialafos J., Kolettis M., Hadijgeorge C., Georgiou V., Avgoustakis D. The significance of intraventricular electrocardiogram in artificial heart pacing. Observations made on 232 patients. Acta Cardiol. 1975;30(6):487–497. [PubMed] [Google Scholar]
  6. Gulotta S. J. Transvenous cardiac pacing. Technics for optimal electrode positioning and prevention of coronary sinus placement. Circulation. 1970 Oct;42(4):701–718. doi: 10.1161/01.cir.42.4.701. [DOI] [PubMed] [Google Scholar]
  7. Parsonnet V., Bilitch M., Furman S., Fisher J. D., Escher D. J., Myers G., Cassady E. Early malfunction of transvenous pacemaker electrodes. A three-center study. Circulation. 1979 Sep;60(3):590–596. doi: 10.1161/01.cir.60.3.590. [DOI] [PubMed] [Google Scholar]
  8. Rubenfire M., Anbe D. T., Drake E. H., Ormond R. S. Clinical evaluation of myocardial perforation as a complication of permanent transvenous pacemakers. Chest. 1973 Feb;63(2):185–188. doi: 10.1378/chest.63.2.185. [DOI] [PubMed] [Google Scholar]

Articles from British Heart Journal are provided here courtesy of BMJ Publishing Group

RESOURCES