. 1975 Dec;37(12):1311–1313. doi: 10.1136/hrt.37.12.1311

Unusual fracture of cardiac pacing electrode.

F Hepburn, M K Meeran
PMCID: PMC482958  PMID: 1225346

Abstract

This report describes a patient with an epicardial demand pacing system in whom pacing failure two months after implantation was found to be due to a fracture of the helical wire electrode within its supporting structure. We know of no report of this site of failure of a Cordis 323-451 or similar electrode.

Full text