Removal of AC interference from the electrocardiogram.

H Moseley, J A Kennedy
PMCID: PMC483341  PMID: 901687

Abstract

A method is described for the removal of AC interference from the electrocardiogram in monitors which do not have isolated inputs. The patient's leads are connected to an isolation amplifier which provides isolation between the input and the output and has a high common mode rejection ratio. The output from the isolation amplifier is taken to the electrocardiographic monitor. The isolation amplifier not only provides additional patient safety but also reduces the level of AC interference without interfering with any frequency components in the electrocardiogram.

Full text

PDF
907

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Monks P. S. Safe use of electro--medical equipment. Anaesthesia. 1971 Jul;26(3):264–280. doi: 10.1111/j.1365-2044.1971.tb04784.x. [DOI] [PubMed] [Google Scholar]
  2. Raftery E. B., Green H. L., Gregory I. C. Disturbances of heart rhythm produced by 50 Hz leakage currents in dogs. Cardiovasc Res. 1975 Mar;9(2):256–262. doi: 10.1093/cvr/9.2.256. [DOI] [PubMed] [Google Scholar]

Articles from British Heart Journal are provided here courtesy of BMJ Publishing Group