Skip to main content
Heart logoLink to Heart
. 1996 Jan;75(1):83–88. doi: 10.1136/hrt.75.1.83

Transcatheter closure of secundum atrial septal defects with the atrial septal defect occlusion system (ASDOS): initial experience in children.

G Hausdorf 1, M Schneider 1, B Franzbach 1, C Kampmann 1, K Kargus 1, B Goeldner 1
PMCID: PMC484230  PMID: 8624881

Abstract

OBJECTIVE--To report initial experiences with transcatheter occlusion of atrial septal defects using a new occlusion device. SUBJECTS--10 children aged 1.1 to 14.9 years. INCLUSION CRITERIA--Patients with a body weight above 10 kg, normal pulmonary resistance and an indication for surgical closure of a secundum atrial septal defect, a residual tissue rim of interatrial septum surrounding the defect of more than 5 mm, and a maximum defect diameter of 20 mm. METHODS--The defects were closed by a transcatheter device (ASDOS) consisting of two umbrellas which are introduced over a guidewire loop. Both umbrellas consist of a central body and five arms formed from preshaped nitinol wire covered with a thin polyurethane patch. The central body of the distal umbrella contains a thread, the proximal umbrella contains a bolt. The two umbrellas are connected by screwing the bolt on the thread using a screwdriver catheter. RESULTS--The implantation was performed under echocardiographic guidance; in six of 10 patients, transoesophageal echocardiography was necessary. The "stretched" diameter of the defect evaluated during balloon sizing ranged from 10 to 20 mm, and the pulmonary to systemic blood flow ratio from 1.5:1 to 2.8:1. Transcatheter closure was successfully performed in 9/10 patients using devices with a diameter of 25 mm to 40 mm. No severe complications occurred. However, in one patient with a pre-existing prolonged PR interval brief periods of second and third degree atrioventricular block occurred after the implantation but normalised within 3 d. During a follow up period of 21 to 29 weeks no device embolisation, thromboembolic complications, fractures of the implanted device, atrial perforations, pericardial effusions, obstructions of systemic or pulmonary veins, atrioventricular valve dysfunction, or other complications occurred. CONCLUSIONS--The new device is a promising transcatheter approach for the occlusion of secundum atrial septal defects in children. However, further evaluation and long term data are needed before this transcatheter technique can be recommended.

Full text

PDF
83

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Babic U. U., Grujicic S., Popovic Z., Djurisic Z., Vucinic M., Pejcic P. Double-umbrella device for transvenous closure of patent ductus arteriosus and atrial septal defect: first experience. J Interv Cardiol. 1991;4(4):283–294. doi: 10.1111/j.1540-8183.1991.tb00810.x. [DOI] [PubMed] [Google Scholar]
  2. Boutin C., Musewe N. N., Smallhorn J. F., Dyck J. D., Kobayashi T., Benson L. N. Echocardiographic follow-up of atrial septal defect after catheter closure by double-umbrella device. Circulation. 1993 Aug;88(2):621–627. doi: 10.1161/01.cir.88.2.621. [DOI] [PubMed] [Google Scholar]
  3. Das G. S., Voss G., Jarvis G., Wyche K., Gunther R., Wilson R. F. Experimental atrial septal defect closure with a new, transcatheter, self-centering device. Circulation. 1993 Oct;88(4 Pt 1):1754–1764. doi: 10.1161/01.cir.88.4.1754. [DOI] [PubMed] [Google Scholar]
  4. Ferreira S. M., Ho S. Y., Anderson R. H. Morphological study of defects of the atrial septum within the oval fossa: implications for transcatheter closure of left-to-right shunt. Br Heart J. 1992 Apr;67(4):316–320. doi: 10.1136/hrt.67.4.316. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Hellenbrand W. E., Fahey J. T., McGowan F. X., Weltin G. G., Kleinman C. S. Transesophageal echocardiographic guidance of transcatheter closure of atrial septal defect. Am J Cardiol. 1990 Jul 15;66(2):207–213. doi: 10.1016/0002-9149(90)90590-w. [DOI] [PubMed] [Google Scholar]
  6. Ishii M., Kato H., Inoue O., Takagi J., Maeno Y., Sugimura T., Miyake T., Kumate M., Kosuga K., Ohishi K. Biplane transesophageal echo-Doppler studies of atrial septal defects: quantitative evaluation and monitoring for transcatheter closure. Am Heart J. 1993 May;125(5 Pt 1):1363–1368. doi: 10.1016/0002-8703(93)91008-3. [DOI] [PubMed] [Google Scholar]
  7. Ishii M., Kato H., Inoue O., Takagi J., Maeno Y., Sugimura T., Miyake T., Kumate M., Kosuga K., Ohishi K. Biplane transesophageal echo-Doppler studies of atrial septal defects: quantitative evaluation and monitoring for transcatheter closure. Am Heart J. 1993 May;125(5 Pt 1):1363–1368. doi: 10.1016/0002-8703(93)91008-3. [DOI] [PubMed] [Google Scholar]
  8. King T. D., Thompson S. L., Mills N. L. Measurement of atrial septal defect during cardiac catheterization. Experimental and clinical results. Am J Cardiol. 1978 Mar;41(3):537–542. doi: 10.1016/0002-9149(78)90012-7. [DOI] [PubMed] [Google Scholar]
  9. King T. D., Thompson S. L., Steiner C., Mills N. L. Secundum atrial septal defect. Nonoperative closure during cardiac catheterization. JAMA. 1976 Jun 7;235(23):2506–2509. [PubMed] [Google Scholar]
  10. Lock J. E., Rome J. J., Davis R., Van Praagh S., Perry S. B., Van Praagh R., Keane J. F. Transcatheter closure of atrial septal defects. Experimental studies. Circulation. 1989 May;79(5):1091–1099. doi: 10.1161/01.cir.79.5.1091. [DOI] [PubMed] [Google Scholar]
  11. Rao P. S., Langhough R., Beekman R. H., Lloyd T. R., Sideris E. B. Echocardiographic estimation of balloon-stretched diameter of secundum atrial septal defect for transcatheter occlusion. Am Heart J. 1992 Jul;124(1):172–175. doi: 10.1016/0002-8703(92)90937-q. [DOI] [PubMed] [Google Scholar]
  12. Rao P. S., Sideris E. B., Hausdorf G., Rey C., Lloyd T. R., Beekman R. H., Worms A. M., Bourlon F., Onorato E., Khalilullah M. International experience with secundum atrial septal defect occlusion by the buttoned device. Am Heart J. 1994 Nov;128(5):1022–1035. doi: 10.1016/0002-8703(94)90602-5. [DOI] [PubMed] [Google Scholar]
  13. Rashkind W. J. Transcatheter treatment of congenital heart disease. Circulation. 1983 Apr;67(4):711–716. doi: 10.1161/01.cir.67.4.711. [DOI] [PubMed] [Google Scholar]
  14. Rome J. J., Keane J. F., Perry S. B., Spevak P. J., Lock J. E. Double-umbrella closure of atrial defects. Initial clinical applications. Circulation. 1990 Sep;82(3):751–758. doi: 10.1161/01.cir.82.3.751. [DOI] [PubMed] [Google Scholar]
  15. Sideris E. B., Sideris S. E., Fowlkes J. P., Ehly R. L., Smith J. E., Gulde R. E. Transvenous atrial septal defect occlusion in piglets with a "buttoned" double-disk device. Circulation. 1990 Jan;81(1):312–318. doi: 10.1161/01.cir.81.1.312. [DOI] [PubMed] [Google Scholar]

Articles from Heart are provided here courtesy of BMJ Publishing Group

RESOURCES