Ex vivo evaluation of new 2D and 3D dental radiographic technology for detecting caries

Laurence Gaalaas 1,, Donald Tyndall 2, André Mol 2, Eric T Everett 3, Ananta Bangdiwala 4
PMCID: PMC4846148  PMID: 26670605

Abstract

Objectives:

Proximal dental caries remains a prevalent disease with only modest detection rates by current diagnostic systems. Many new systems are available without controlled validation of diagnostic efficacy. The objective of this study was to evaluate the diagnostic efficacy of three potentially promising new imaging systems.

Methods:

This study evaluated the caries detection efficacy of Schick 33 (Sirona Dental, Salzburg, Austria) intraoral digital detector images employing an advanced sharpening filter, Planmeca ProMax® (Planmeca Inc., Helsinki, Finland) extraoral “panoramic bitewing” images and Sirona Orthophos XG3D (Sirona Dental) CBCT images with advanced artefact reduction. Conventional photostimulable phosphor images served as the control modality. An ex vivo study design using extracted human teeth, ten expert observers and micro-CT ground truth was employed.

Results:

Receiver operating characteristic analysis indicated similar diagnostic efficacy of all systems (ANOVA p > 0.05). The sensitivity of the Schick 33 images (0.48) was significantly lower than the other modalities (0.53–0.62). The specificity of the Planmeca images (0.86) was significantly lower than Schick 33 (0.96) and XG3D (0.97). The XG3D showed significantly better cavitation detection sensitivity (0.62) than the other modalities (0.48–0.57).

Conclusions:

The Schick 33 images demonstrated reduced caries sensitivity, whereas the Planmeca panoramic bitewing images demonstrated reduced specificity. XG3D with artefact reduction demonstrated elevated sensitivity and specificity for caries detection, improved depth accuracy and substantially improved cavitation detection. Care must be taken to recognize potential false-positive caries lesions with Planmeca panoramic bitewing images. Use of CBCT for caries detection must be carefully balanced with the presence of metal artefacts, time commitment, financial cost and radiation dose.

Keywords: dental caries; radiography, dental; bitewing radiography; cone-beam CT; MicroCT

Introduction

The radiographic detection rate of proximal caries is far from ideal as standard bitewing radiographs detect only about 60% of proximal lesions.13 The relatively recent introduction of digital dental radiographic imaging has so far failed to demonstrate any increase in caries detection rates.411 Overall, there does not yet appear to be an image enhancement algorithm that delivers clear gains in caries diagnosis.12

Compared with intraoral imaging, extraoral imaging using panoramic X-ray units promises improved patient comfort and efficiency. These modalities have been evaluated for their caries detection potential with mostly inferior performance compared with intraoral imaging owing to superimposition of additional structures, increased image blurriness and inconsistent opening of posterior proximal contacts.1315 At this time, the authors are aware of only one published study that has evaluated panoramic bitewing images (Planmeca Promax®; Planmeca Inc., Helsinki, Finland) for proximal caries detection, which found inferior detection rates compared with standard intraoral bitewing radiography.16

Many studies show that CBCT caries detection rates are approximately equivalent to intraoral modalities for non-restored teeth.11,1723 A recent review of the literature summarises that CBCT is equivalent to intraoral techniques at detecting clinically relevant caries lesions in minimally restored teeth; however, beam-hardening and streak artefacts from metal objects and dense tooth structure (enamel) are limiting factors.24,25 Furthermore, increased dose, cost, time and artefact concerns dictate that bitewing radiographs are still the preferred modality for proximal caries detection.26

Broadly, caries diagnosis is not limited to simple lesion detection. Our understanding of the caries disease process has identified lesion depth, activity and cavitation status as significant indicators for the likelihood of lesion progression.27,28 Assessment of lesion cavitation is also important because cavitated lesions demonstrate a much higher likelihood of progression.2830 Ultimately, the approach to identifying and treating incipient and early caries lesions non-restoratively places a maximum demand on radiographic imaging systems.3133 With regard to lesion detection, dentistry needs increased sensitivity from new diagnostic systems with no corresponding compromise in the already high specificity rates. Dentistry also needs improved ability to accurately identify lesion depth and lesion cavitation.34

A variety of dental imaging technologies have been recently introduced, each having developments that may increase caries detection. These new technologies include the Schick 33 (Schick 33) intraoral direct digital sensor (Sirona Dental, Salzburg, Austria), the Planmeca ProMax® (PanBW) panoramic unit in panoramic bitewing mode (Planmeca Inc.) and the Sirona Orthophos XG3D CBCT (XG3D) in high definition (HD) mode with metal artefact reduction software (MARS) (Sirona Dental).

The National Institutes of Health has called for continued research on diagnostic methods, including new devices and techniques.31 All three of these imaging systems are available, yet there has been limited evaluation of their caries diagnostic performance. Accordingly, the aim of this project was to establish the diagnostic efficacy of these three new dental radiographic imaging technologies for assessing the presence, depth and cavitation status of proximal caries in non-restored teeth using an ex vivo study design.

Methods and materials

The Schick33 intraoral sensor has a dynamic image sharpening filter which may aid caries detection by improving lesion edge visualization.35,36 PanBW images are acquired using the image projection geometry designed to open posterior dental contacts and consequently increase caries visibility. XG3D images with HD mode and MARS employ artefact reduction algorithms which may diminish beam-hardening and streak artefacts, both of which are limitations in CBCT caries imaging for restored as well as non-restored teeth.

Institutional review board approvals were obtained to collect deidentified extracted human teeth from existing specimen repositories and to perform observer sessions. Tooth selection criteria included premolar or molar teeth with an unrestored or minimally restored status with cervical (non-coronal, non-proximal) restoration involvement only. Selection criteria also included proximal caries status of sound/no-lesion or small-to-moderate-sized lesion as estimated by visual, tactile and bitewing radiographic means. Teeth with large, cavitated coronal lesions were excluded. A total of 29 extracted teeth were selected for the sample. 3 teeth were used twice giving a total sample size of 32 teeth (64 proximal surfaces). The three teeth were selected to be used twice because they demonstrated simultaneously classic proximal lesion morphology and challenging lesion visibility when screened with bitewing radiography. A power calculation for receiver operating characteristic (ROC) analysis provided an estimated sample size of 51 surfaces for this “paired-case, paired-reader” study design.3739

A dry human mandible served as the ex vivo phantom for this study. Extracted teeth were held in residual extraction sockets with wax. Premolar–molar pairs were randomly selected from the sample to establish an anatomically appropriate pair of teeth for placement in the mandible. Non-restored and non-carious premolars and molars were placed anteriorly and posteriorly to the sample teeth. The mandible was restored bilaterally to best simulate the attenuation properties of scanning a normal dental arch. Efforts were made to make contacts as parallel as possible in order to minimize the confounding effect of closed contacts on the task of interpretation.

A 1-cm layer of dental “boxing” wax was placed around the mandible as a suitable simulation of soft-tissue attenuation. For the extraoral imaging modalities (XG3D and PanBW), an additional water balloon and a 3-mm thick aluminium cylinder with a 15.5-cm diameter were placed inside and around the mandible/wax phantom, respectively, to further simulate the attenuation properties of the human head relevant to extraoral imaging.

The control modality consisted of photostimulable phosphor (PSP) bitewing radiographs: Gendex Size 2 PSP plates (Gendex, Hatfield, PA) were exposed with a Focus™ intraoral source (Instrumentarium Dental, Tuusula, Finland) at 70 kVp, 7 mA, 0.2 s and 40 cm source to image distance (SID) with standard rectangular collimation. The exposed plates were processed with a ScanX IO ILE scanner (Air Techniques, Melville, NY) through MiPACS Dental Enterprise Viewer 3.1 operating ScanX Plugin v. 1.2.8 (Medicore Imaging, Charlotte, NC). The Schick33 images were taken with the same Focus intraoral source at 70 kVp, 0.05 s and 40 cm SID standard rectangular collimation in the same manner as the PSP modality. The software interface for the Schick33 sensor was CDR DICOM for Windows® v. 4 (Sirona Dental). PanBW images were taken at 72 kVp and 11 mA with a square average jaw shape. The mandible was placed in the machine according to the standard anatomic positioning. Dexis v. 9.0.5 imaging software (Dexis LLC, Hatfield, PA) was used to acquire and store the images. XG3D scans were taken at 85 kVp and 6 mA with a 5 × 5-cm field of view (FOV) and 0.1-mm voxel size. Both HD mode and MARS were used. Scans were reconstructed and stored using Sidexis XG v. 2.56 software (Sirona Dental).

Caries ground truth status was established by micro-CT (microCT). The microCT scans were acquired with a SCANCO Medical µCT 40 scanner (Scanco Medical AG, Bruttisellen, Switzerland) operating at 70 kVp, 0.115 mA, 200-s scan time, with 0.5-mm aluminium filtration. Scans were reconstructed with a 20 µm voxel size using Scanco v. 1.2a software (SCANCO Medical AG). Teeth were scanned individually to avoid any pronounced beam-hardening artefacts due to adjacent teeth and/or restorations.

A total of 10 observers were recruited from the institution's Division of Oral and Maxillofacial Radiology. All observers had specialty-level training in oral and maxillofacial radiology and familiarity with diagnosing caries using intraoral, extraoral and three-dimensional imaging modalities. All observers participated in an orientation session prior to the interpretation sessions. The specific tasks asked of the observers for each proximal surface of each sample tooth were as follows. Task 1: rate the likelihood of caries presence on a 5-point scale where 1 = caries definitely not present, 2 = caries probably not present, 3 = unsure, 4 = caries probably present and 5 = caries definitely present. Task 2: rate the caries lesion depth on a 5-point scale where 1 = caries not present, 2 = caries involving the outer half of enamel, 3 = caries involving the inner half of enamel, 4 = caries involving the outer half of dentin and 5 = caries involving the inner half of dentin. Task 3: rate the likelihood of lesion cavitation on a 5-point scale where 1 = cavitation definitely not present, 2 = cavitation probably not present, 3 = unsure, 4 = cavitation probably present and 5 = cavitation definitely present. Observers used dual-monitor workstations with Lenovo LT2252P monitors (Lenovo, Beijing, China) as the primary diagnostic displays. All diagnostic displays passed TCG-18 test pattern quality control checks of the 5% and 95% contrast levels prior to the sessions. Ambient lighting in the room was subdued to appropriate interpretation levels.

MiPACS Dental Enterprise Viewer 3.1 (Medicore Imaging) was used as the PSP bitewing interpretation software. CDR DICOM for Windows® v. 5.4 was used for the Schick33 interpretation software. The general dentistry task with default 35% sharpening was selected for initial image display. DEXview v. 10.0.2 (Dexis LLC, Hatfield, PA) was used to view the panoramic bitewing images. Galaxis v. 1.9 (Sirona Dental) was used for the CBCT interpretation software. Observers were allowed to use brightness/contrast adjustment tools for their interpretation. Observers were also allowed to adjust the level of sharpening of the Schick33 images with the CDR software.

Following interpretation and discussion of the calibration set of images, the observers completed interpretation of images of all 32 subject teeth imaged by all four modalities. The sequence of image and modality interpretation was randomized so that no modality was biased towards the beginning or end of the session and no set of the same sample teeth were interpreted in succession. After a washout period of 2–3 weeks, observers returned to complete a second session. The observers interpreted images of exactly half of the subject teeth imaged by all four modalities (16 of 32).

To complete the caries ground truth analysis, each microCT scan was reviewed by the principal investigator and coauthor experienced with caries ground truth assessment to establish caries status, lesion depth and cavitation status of every proximal surface in the sample. Decision discrepancies were resolved by consensus following a discussion of the image findings.

ROC curves were constructed and fitted area-under-the-curve (AZ) scores were recorded using a web-based ROC analysis tool (www.jrocfit.org). Caries presence detection and cavitation detection sensitivity and specificity scores were calculated for each observer–modality combination where caries presence ratings 4 and 5 were considered a positive diagnosis and caries ratings 1, 2 and 3 were considered a negative diagnosis. Observer–modality combination AZ scores were compared with a fixed-effects main-effects two-way ANOVA. Sensitivity and specificity scores were compared with a fixed-effects Friedman's two-way non-parametric ANOVA using observer and modality as main factors. A non-parametric approach was chosen based on the non-normal behaviour of sensitivity and specificity scores. For both AZ and sensitivity/specificity tests, a p-value <0.05 was considered statistically significant for overall test of effects. Wilcoxon rank-sum pairwise comparisons with Bonferroni adjustment were made when statistical significance was observed between modalities (p < 0.05/6 = 0.0083 was considered statistically significant).

Only true-positive observations were considered for depth accuracy analysis. For each observer's depth observation on each modality, the level of agreement between the depth observation and depth assessment by microCT ground truth were calculated. Weighted kappa statistics were calculated using Cicchetti–Allison weights based on calculated scores. Weighted kappa coefficients for each modality were compared with one-way ANOVA. A p-value <0.05 was considered statistically significant for overall test of effects. Post-hoc t-test pairwise comparisons with Bonferroni adjustment were made when statistical significance was observed between modalities (p < 0.05/6 = 0.0083 was considered statistically significant).

The observation scores from the second session were used to determine the intraobserver agreement. For each of every observer's tooth surface assessment (presence, depth and cavitation), the level of agreement between the first and second session observations were calculated. It was assumed that images were independent and that ratings were independent. Weighted kappa statistics were calculated using Cicchetti–Allison weights based on scores. The median weighted kappa for each modality was determined, and coefficients for each observer were compared with one-way ANOVA. Analyses were performed using SAS® software v. 9.3 (SAS® Institute Inc., Cary, NC).

Results

Example images of extracted teeth and caries can be seen in Figure 1. A breakdown of proximal surface lesion status as determined by microCT is provided in Table 1. ROC curves for compiled caries presence observations by modality are provided in Figure 2. For the ROC analysis, fixed-effects main effects two-way ANOVA of AZ scores indicated no significant difference between observers (p = 0.15) and between modalities (p = 0.10). Fixed-effects Friedman's two-way non-parametric ANOVA of caries presence sensitivity scores for the full data set indicated significant differences between observers (p < 0.0001) and between modalities (p = 0.0002). Similar analysis of specificity scores indicated significant differences between observers (p = 0.0005) and between modalities (p = 0.0016). Subsequent Wilcoxon rank-sum pairwise comparisons between modalities indicated that PSP (0.57) demonstrated a higher sensitivity than Schick33 (0.48) (p = 0.0018), XG3D demonstrated a higher sensitivity than Schick33 (0.48) (p < 0.0001) and XG3D (0.62) demonstrated a higher sensitivity than PanBW (0.53) (p = 0.0071). Schick33 (0.96) demonstrated higher specificity than PanBW (0.86) (p = 0.0005) and XG3D (0.97) demonstrated a higher specificity than PanBW (0.86) (p = 0.0013).

Figure 1.

Figure 1.

Representative images of extracted teeth and caries for photostimulable phosphor (a), Schick 33 (Sirona Dental, Salzburg, Austria) (b), Planmeca panoramic bitewing (Planmeca Inc., Helsinki, Finland) (c), Sirona Orthophos XG3D CBCT (Sirona Dental) (d) and micro-CT ground truth (e). The lower right first molar is the same tooth in all images. Calibrations rods visualized in images are relevant to a separate study.

Table 1.

Proximal surface ground truth status as assessed by micro-CT

Caries Surfaces Deptha Surfaces Cavitation Surfaces
Sound 30 1 30 Absent 52
Lesion 34 2 9 Present 12
    3 9    
    4 14    
    5 2    
Total 64   64   64
a

Lesion depth scale according to depth definitions in text.

Figure 2.

Figure 2.

Receiver operating characteristic curves for each modality's compiled observations. PanBW from Planmeca Inc., Helsinki, Finland; PSP from Gendex, Hatfield, PA; and Schick33 and XG3D have been obtained from Sirona Dental, Salzburg, Austria.

The median accuracy of the observer-assigned depth scores were moderately high for PSP (κ = 0.40), Schick 33 (κ = 0.43) and XG3D, with XG3D having the best accuracy (κ = 0.63). The PanBW modality was considered fairly inaccurate (κ = 0.27). ANOVA of kappa coefficients indicated significant differences between modalities (p = 0.003). Pairwise comparisons indicated that XG3D had significantly better depth agreement than PanBW (p = 0.002).

For cavitation sensitivity scores, ANOVA indicated significant differences between observers (p = 0.0063) and between modalities (p < 0.0001). In cavitation specificity scores, there were similar significant differences between modalities (p = 0.0025) but no significant differences between observers (p = 0.1829). Pairwise comparisons indicated that XG3D (0.83) had higher sensitivity than PSP (0.54), Schick33 (0.47) and PanBW (0.38) (p = 0.0011, <0.0001 and <0.0001, respectively). XG3D (0.96) had statistically significant lower specificity than PSP (0.99) and Schick33 (0.99).

Overall, based on median weighted kappa, intraobserver agreement of caries presence was moderately high for each modality. Intraobserver agreement for caries depth was also moderately high for each modality. Intraobserver agreement of caries cavitation was moderately high as well for each modality. One-way ANOVA of kappa coefficients indicated no significant differences between modalities for caries presence, depth and cavitation ratings (p = 0.411, 0.376, 0.197, respectively). A summary of all statistical findings is provided in Table 2.

Table 2.

Summary of statistical findings comparing caries presence AZ, caries sensitivity/specificity, depth observation weighted kappa correlation, cavitation sensitivity/specificity and intraobserver agreement

Data set, modality Diagnostic measure Test effect p-value Pairwise comparison p-value
AZ scoresa Mean AZ ± 95% CI        
PSP 0.820 ± 0.037        
Schick33 0.780 ± 0.032        
PanBW 0.791 ± 0.035        
XG3D 0.828 ± 0.020        
    Observer 0.15 N/A N/A
    Modality 0.10 N/A N/A
Caries sensitivityb Mean sensitivity ± 95% CI        
PSP 0.57 ± 0.07        
Schick33 0.48 ± 0.09        
PanBW 0.53 ± 0.08        
XG3D 0.62 ± 0.07        
    Observer <0.0001c    
    Modality 0.0002c    
        PSP vs PanBW 0.23
        PSP vs Schick33 0.0018c
        PSP vs XG3D 0.10
        PanBW vs Schick33 0.033
        PanBW vs XG3D 0.0071c
        Schick33 vs XG3D <0.0001c
        Schick33 vs XG3D 0.73
Caries specificitya Mean specificity ± 95% CI        
PSP 0.90 ± 0.08        
Schick33 0.96 ± 0.04        
PanBW 0.86 ± 0.09        
XG3D 0.97 ± 0.02        
    Observer 0.0005c    
    Modality 0.0016c    
        PSP vs PanBW 0.10
        PSP vs Shick33 0.034
        PSP vs XG3D 0.07
        PanBW vs Schick33 0.0005c
        PanBW vs XG3D 0.0013c
Lesion depth correlation with micro-CT (kappa)d Median kappa (IQR)        
PSP 0.40 (0.30, 0.53)        
Schick33 0.43 (0.40, 0.50)        
PanBW 0.27 (0.23, 0.38)        
XG3D 0.63 (0.44, 0.72)        
    Modality 0.003c    
        PSP vs PanBW 0.62
        PSP vs Shick33 1.00
        PSP vs XG3D 0.15
        PanBW vs Schick33 0.32
        PanBW vs XG3D 0.002c
        Schick33 vs XG3D 0.30
Cavitation sensitivitye Mean sensitivity ± 95% CI        
PSP 0.57 ± 0.14        
Schick33 0.48 ± 0.10        
PanBW 0.53 ± 0.11        
XG3D 0.62 ± 0.06        
    Observer 0.0063c    
    Modality <0.0001c    
        PSP vs PanBW 0.043
        PSP vs Schick33 0.16
        PSP vs XG3D 0.0011c
        PanBW vs Schick33 0.50
        PanBW vs XG3D <0.0001c
        Schick33 vs XG3D <0.0001c
Cavitation specificitye Mean specificity ± 95% CI        
PSP 0.90 ± 0.01        
Schick33 0.96 ± 0.01        
PanBW 0.86 ± 0.01        
XG3D 0.97 ± 0.02        
    Observer 0.18    
    Modality 0.0025c    
        PSP vs PanBW 0.14
        PSP vs Schick33 0.78
        PSP vs XG3D 0.0015c
        PanBW vs Schick33 0.077
        PanBW vs XG3D 0.056
        Schick33 vs XG3D 0.0007c
Intraobserver agreement caries presencef Median kappa (IQR)        
PSP 0.62 (0.57, 0.74)        
Schick33 0.73 (0.62, 0.81)        
PanBW 0.62 (0.52, 0.66)        
XG3D 0.65 (0.58, 0.74)        
    Modality 0.411 N/A N/A
Intraobserver agreement caries depthf Median kappa (IQR)        
PSP 0.76 (0.59, 0.81)        
Schick33 0.74 (0.66, 0.80)        
PanBW 0.66 (0.53, 0.740)        
XG3D 0.73 (0.70, 0.75)        
    Modality 0.376 N/A N/A
Intraobserver agreement caries cavitationf Median kappa (IQR)        
PSP 0.69 (0.60, 0.75)        
Schick33 0.70 (0.63, 0.81)        
PanBW 0.60 (0.41, 0.69)        
XG3D 0.72 (0.62, 0.80)        
    Modality 0.197 N/A N/A

CI, confidence interval; IQR, interquartile range.

PanBW from Planmeca Inc., Helsinki, Finland; PSP from Gendex, Hatfield, PA; and Schick33 and XG3D have been obtained from Sirona Dental, Salzburg, Austria.

a

Statistically significant difference.

b

Fixed-effects main effects two-way ANOVA.

c

Fixed-effects Friedman's two-way non-parametric ANOVA and Wilcoxon rank-sum pairwise comparison, Bonferroni adjustment (p < 0.05, p < 0.0083, respectively).

d

One-way ANOVA with t-test pairwise comparison, Bonferroni adjustment (p < 0.05, p < 0.0083, respectively).

e

Fixed-effects Friedman's two-way non-parametric ANOVA and Wilcoxon rank-sum pairwise comparison, Bonferroni adjustment (p < 0.05, p < 0.0083, respectively).

f

One-way ANOVA.

Discussion

ROC analysis was chosen because it removes the effect of observers' different diagnosis decision-making and provides the best overall indication of diagnostic accuracy to be compared between modalities.4042 ROC analysis of caries diagnosis indicated no significant differences between modalities' in terms of their AZ scores. This suggests that all four modalities performed equally with respect to identifying lesions. One key nuance of ROC analysis, however, is that two separate ROC curves may have the exact same AZ but different curve shapes, depending on the trade-off between sensitivity and specificity. This trade-off does not affect the overall diagnostic accuracy indication, but it is clinically relevant since it is much more beneficial to the patient and provider to maximize specificity because a false-positive lesion can result in unnecessary surgical treatment of the tooth at additional cost to the patient and significant detriment to the tooth's future restorative prognosis.1 To reconcile these points with our previously stated desires for increased sensitivity, the authors reiterate that we desire improved sensitivity from new diagnostic systems with no corresponding penalty in the already high specificities.

Subsequent analysis of modality sensitivity and specificity scores elucidated potentially relevant clinical differences in performance. PSP (0.57) demonstrated a higher sensitivity than Schick33 (0.48), XG3D demonstrated a higher sensitivity than Schick33 (0.48) and XG3D (0.62) demonstrated a higher sensitivity than PanBW (0.53). Schick33 (0.96) demonstrated higher specificity than PanBW (0.86) and XG3D (0.97) demonstrated a higher specificity than PanBW (0.86). Finding significant differences between observers in the caries and cavitation sensitivity/specificity analyses was not surprising because these measures of diagnostic accuracy are dependent on the individual observer's decision threshold; therefore the specific ANOVA analyses for sensitivity/specificity results were chosen because they address observer variation before making a determination of significant modality effects.

The lower sensitivity of Schick33 may be attributable to the dynamic image-sharpening filter. Based on observer comments, it was found that different observers had different preferences for the level of sharpness. The immediate adjustment of image sharpness during interpretation may have distracted or otherwise influenced the observers into missing very subtle, difficult-to-detect lesions resulting in somewhat decreased sensitivity. Alternatively, the filter may have truly obscured lesions. However, this study did not test the influence of specific levels of image sharpening in the Schick33 system; therefore no definitive conclusions about the sharpening system itself can be made at this time.

Decreased specificity with the panoramic bitewing images may be attributable to image “ghost” artefacts from the contralateral aspect of the mandible formed as a result of image acquisition geometry, similar to that in traditional panoramic imaging.1315 Observers commented that horizontal tomographic streak artefacts from the contralateral body of the phantom mandible were superimposed over the teeth of interest, resulting in the perception that image interpretation was more challenging. It is possible that these artefacts may have mimicked lesions in a number of cases, resulting in an elevated false-positive rate. The authors are concerned that an elevated false-positive rate could have detrimental effects associated with caries treatment decisions.

The XG3D system demonstrated higher sensitivity than other modalities while maintaining high specificity. This finding is different than previous studies which cited beam-hardening artefacts from nearby enamel as a source of decreased specificity (increased false-positive diagnoses).18,43 One possible explanation of reduced false-positive diagnoses is that the XG3D was operating in the HD mode with the MARS reconstruction algorithm. An alternative explanation is the investigators' demonstration of beam-hardening artefacts vs real caries lesions to the observers during orientation. This demonstration may have educated the observers to more consistently avoid false-positive diagnoses of artefacts as lesions much in the same way that well-trained dentists are able to differentiate caries from cervical burnout on two-dimensional images. Unfortunately, this study was designed to test the overall efficacy of the XG3D system compared with other modalities, therefore definitive conclusions cannot be drawn at this time. Regarding true-positive depth accuracy, the XG3D modality had significantly more accurate depth scores than the PanBW modality.

The XG3D system demonstrated an approximately 30% increase in detection of cavitated lesions compared with other modalities. This finding is highly consistent with other recent studies. Our statistical analysis indicated that the XG3D (0.96) had significantly lower cavitation specificity than PSP (0.99) and Schick (0.99); however, these observed specificities are all extremely high. Therefore, it can be argued that the statistically significant cavitation specificity differences are clinically not relevant. Based on the findings in this study and other recent studies, it seems apparent that CBCT affords better detection of cavitated proximal lesions.20,26,29

Importantly, the authors reiterate that tooth selection criteria included premolar or molar teeth with an unrestored or minimally restored status with cervical (non-coronal, non-proximal) restoration involvement only. For these imaging systems, the effect of caries diagnosis on restored teeth on non-restored teeth with restorations within the scan FOV is undetermined at this time and warrants additional investigation. Also, the detrimental effect of patient motion on scan quality was not evaluated by this ex vivo study design and similarly warrants additional investigation.44,45

Finally, CBCT imaging incurs much greater radiation dose, financial cost and acquisition/interpretation time than other imaging modalities. The typical radiation dose for a CBCT scan varies widely from several micro Sieverts (µSv) to >1000 µSv, depending primarily on selected scan resolution and FOV.4648 The XG3D protocol used in this study had an effective dose of 53 µSv per mandibular posterior quadrant scan.48 Comparatively, the dose from a standard set of four PSP or digital bitewing images is about 5 µSv, and dose from a standard panoramic image (approximately equal to a panoramic bitewing dose) is about 15 µSv.46 The financial cost of CBCT imaging is significantly more than conventional bitewing or panoramic/panoramic bitewing examinations. The total acquisition and interpretation time for a CBCT is also significantly more than conventional bitewing or panoramic modalities. These costs must be carefully balanced with differences in diagnostic efficacy before selecting the imaging modality to address a relevant diagnostic question. With regard to caries detection, it is the authors' opinion that the marginal increase in lesion sensitivity, marginal increase in lesion depth assessment and even the substantial improvement in lesion cavitation detection may not be worth the significant increases in dose, cost and time. Of course, if a CBCT is taken for other clinical indications, it appears prudent to assess the visible regions of teeth for caries.

Almost all observer–modality combination weighted-kappa correlation scores were at least moderately agreeing (κ >  0.40). A review of similar multireader ROC studies cites a range of intraobserver agreement from 0.35 to 0.59,49 confirming typical results in this study. No statistically significant differences in intraobserver agreement scores were observed between modalities for the caries presence, caries depth and caries cavitation questions. This finding indicates that our observers performed uniformly throughout the study without a modality effect.

A limitation of this study was the ideal conditions established for proximal caries diagnosis. These ideal conditions were chosen to best control confounding variables and isolate the actual differences in detection efficacy between modalities, but they may not represent clinically relevant interpretation scenarios. Interpretation took place under ideal viewing conditions with subdued ambient lighting, quiet surroundings and contrast calibrated monitors. Additionally, the observers were qualified experts in dental radiographic diagnosis, each having specialty training in oral and maxillofacial radiology and a detailed knowledge of radiographic caries diagnosis beyond the general dental level. Finally, all images were acquired with ideal image geometry resulting in open proximal contacts, whereas clinical imaging of real patients dictates that it is not always possible to open all posterior proximal contacts.

The traditional gold-standard method to validate dental caries status is microscopic examination of thin sections; however, in recent years, microCT has proven to be an effective alternative. Numerous studies have employed microCT as a ground truth caries validation tool for the purpose of caries detection.39,50 MicroCT has also been shown to agree with transverse microradiography, another accepted gold-standard technique for the evaluation of caries.51 The selection of 70 kVp tube potential and system calibration efforts in this study's microCT analysis are well supported by the literature.52

Conclusions

  • This study found equivalent overall diagnostic efficacy for posterior proximal caries detection between three new and one control dental radiographic systems.

  • The Schick 33 intraoral sensor demonstrated reduced caries sensitivity compared with other modalities.

  • The Planmeca panoramic bitewing modality demonstrated an elevated false-positive rate with potentially clinically relevant implications for treatment decisions.

  • The CBCT with artefact reduction demonstrated promising sensitivity/specificity for caries detection, somewhat improved depth accuracy and substantially improved cavitation detection.

  • Practitioners must carefully balance caries detection on CBCT with the presence of metal artefacts, time commitment, financial cost and radiation dose.

  • This study was performed with expert observers under ideal viewing conditions with ideal image geometry. These features of study design were chosen to best evaluate the diagnostic capabilities of the specific experimental modalities. Accordingly, the results may not apply to all clinical scenarios, environments and modalities. More research is needed to further clarify caries detection with modern imaging systems.

Acknowledgments

Acknowledgments

The authors thank Dr Thomas Schaaf for generously offering his time and access to the Planmeca unit; Dr John Preisser, Jr and Adane Wogu for their statistical support; Sirona Dental Inc. USA for their donation of a Schick33 intraoral sensor and imaging software; Dr Mansur Ahmad for his assistance with figure preparation; and the Small Animal Imaging Core facility at the UNC Biomedical Imaging Research Center for providing the micro-CT imaging service.

Contributor Information

Laurence Gaalaas, Email: gaal0017@umn.edu.

Donald Tyndall, Email: don_tyndall@unc.edu.

André Mol, Email: Andre_Mol@unc.edu.

Eric T Everett, Email: everette@DENTISTRY.UNC.EDU.

Ananta Bangdiwala, Email: ananta.bangdiwala@gmail.com.

Conflicts of interest

The author Dr Donald Tyndall is a consultant for Sirona Dental Inc. This entity had no input into study design, collection/analysis/reporting of results or decision of publication.

References

  • 1.Dove SB. Radiographic diagnosis of dental caries. J Dent Educ 2001; 65: 985–90. [PubMed] [Google Scholar]
  • 2.Douglass CW, Valachovic RW, Wijesinha A, Chauncey HH, Kapur KK, McNeil BJ. Clinical efficacy of dental radiography in the detection of dental caries and periodontal diseases. Oral Surg Oral Med Oral Pathol 1986; 62: 330–9. doi: 10.1016/0030-4220(86)90017-4 [DOI] [PubMed] [Google Scholar]
  • 3.Bader JD, Shugars DA, Bonito AJ. Systematic reviews of selected dental caries diagnostic and management methods. J Dent Educ 2001; 65: 960–8. [PubMed] [Google Scholar]
  • 4.White SC, Yoon DC. Comparative performance of digital and conventional images for detecting proximal surface caries. Dentomaxillofac Radiol 1997; 26: 32–8. doi: 10.1038/sj.dmfr.4600208 [DOI] [PubMed] [Google Scholar]
  • 5.Khan EA, Tyndall DA, Ludlow JB, Caplan D. Proximal caries detection: Sirona Sidexis versus Kodak Ektaspeed Plus. Gen Dent 2005; 53: 43–8. [PubMed] [Google Scholar]
  • 6.Wenzel A. Digital radiography and caries diagnosis. Dentomaxillofac Radiol 1998; 27: 3–11. doi: 10.1038/sj.dmfr.4600321 [DOI] [PubMed] [Google Scholar]
  • 7.Wenzel A. Bitewing and digital bitewing radiography for detection of caries lesions. J Dental Res 2004; 83: C72–5. doi: 10.1177/154405910408301S14 [DOI] [PubMed] [Google Scholar]
  • 8.Wenzel A. Digital imaging for dental caries. Dent Clin North Am 2000; 44: 319–38. [PubMed] [Google Scholar]
  • 9.Syriopoulos K, Sanderink GC, Velders XL, van der Stelt PF. Radiographic detection of approximal caries: a comparison of dental films and digital imaging systems. Dentomaxillofac Radiol 2000; 29: 312–18. doi: 10.1038/sj.dmfr.4600553 [DOI] [PubMed] [Google Scholar]
  • 10.Dove SB, McDavid WD. A comparison of conventional intra-oral radiography and computer imaging techniques for the detection of proximal surface dental caries. Dentomaxillofac Radiol 1992; 21: 127–34. doi: 10.1038/dmfr.21.3.1397467 [DOI] [PubMed] [Google Scholar]
  • 11.Senel B, Kamburoglu K, Uçok O, Yüksel SP, Ozen T, Avsever H. Diagnostic accuracy of different imaging modalities in detection of proximal caries. Dentomaxillofac Radiol 2010; 39: 501–11. doi: 10.1259/dmfr/28628723 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Wenzel A. A review of dentists' use of digital radiography and caries diagnosis with digital systems. Dentomaxillofac Radiol 2006; 35: 307–14. doi: 10.1259/dmfr/64693712 [DOI] [PubMed] [Google Scholar]
  • 13.Akkaya N, Kansu O, Kansu H, Cagirankaya LB, Arslan U. Comparing the accuracy of panoramic and intraoral radiography in the diagnosis of proximal caries. Dentomaxillofac Radiol 2006; 35: 170–4. doi: 10.1259/dmfr/26750940 [DOI] [PubMed] [Google Scholar]
  • 14.Flint DJ, Paunovich E, Moore WS, Wofford DT, Hermesch CB. A diagnostic comparison of panoramic and intraoral radiographs. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1998; 85: 731–5. [DOI] [PubMed] [Google Scholar]
  • 15.Akarslan ZZ, Akdevelioğlu M, Güngör K, Erten H. A comparison of the diagnostic accuracy of bitewing, periapical, unfiltered and filtered digital panoramic images for approximal caries detection in posterior teeth. Dentomaxillofac Radiol 2008; 37: 458–63. doi: 10.1259/dmfr/84698143 [DOI] [PubMed] [Google Scholar]
  • 16.Kamburoglu K, Kolsuz E, Murat S, Yüksel S, Ozen T. Proximal caries detection accuracy using intraoral bitewing radiography, extraoral bitewing radiography and panoramic radiography. Dentomaxillofac Radiol 2012; 41: 450–9. doi: 10.1259/dmfr/30526171 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Park YS, Ahn JS, Kwon HB, Lee SP. Current status of dental caries diagnosis using cone beam computed tomography. Imaging Sci Dent 2011; 41: 43–51. doi: 10.5624/isd.2011.41.2.43 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Rathore S, Tyndall D, Wright J, Everett E. Ex vivo comparison of Galileos cone beam CT and intraoral radiographs in detecting occlusal caries. Dentomaxillofac Radiol 2012; 41: 489–93. doi: 10.1259/dmfr/57329547 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Valizadeh S, Tavakkoli MA, Karimi Vasigh H, Azizi Z, Zarrabian T. Evaluation of cone beam computed tomography (CBCT) system: comparison with intraoral periapical radiography in proximal caries detection. J Dent Res Dent Clin Dent Prospects 2012; 6: 1–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Wenzel A, Hirsch E, Christensen J, Matzen LH, Scaf G, Frydenberg M. Detection of cavitated approximal surfaces using cone beam CT and intraoral receptors. Dentomaxillofac Radiol 2013; 42: 39458105. doi: 10.1259/dmfr/39458105 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Tsuchida R, Araki K, Okano T. Evaluation of a limited cone-beam volumetric imaging system: comparison with film radiography in detecting incipient proximal caries. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007; 104: 412–16. doi: 10.1016/j.tripleo.2007.02.028 [DOI] [PubMed] [Google Scholar]
  • 22.Kalathingal SM, Mol A, Tyndall DA, Caplan DJ. In vitro assessment of cone beam local computed tomography for proximal caries detection. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007; 104: 699–704. doi: 10.1016/j.tripleo.2006.08.032 [DOI] [PubMed] [Google Scholar]
  • 23.Kayipmaz S, Sezgin ÖS, Saricaoğlu ST, Can G. An in vitro comparison of diagnostic abilities of conventional radiography, storage phosphor, and cone beam computed tomography to determine occlusal and approximal caries. Eur J Radiol 2011; 80: 478–82. doi: 10.1016/j.ejrad.2010.09.011 [DOI] [PubMed] [Google Scholar]
  • 24.Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol 2011; 40: 265–73. doi: 10.1259/dmfr/30642039 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Draenert FG, Coppenrath E, Herzog P, Muller S, Mueller-Lisse UG. Beam hardening artefacts occur in dental implant scans with the NewTom cone beam CT but not with the dental 4-row multidetector CT. Dentomaxillofac Radiol 2007; 36: 198–203. doi: 10.1259/dmfr/32579161 [DOI] [PubMed] [Google Scholar]
  • 26.Wenzel A. Radiographic display of carious lesions and cavitation in approximal surfaces: advantages and drawbacks of conventional and advanced modalities. Acta Odontol Scand 2014; 72: 251–64. doi: 10.3109/00016357.2014.888757 [DOI] [PubMed] [Google Scholar]
  • 27.Baelum V. What is an appropriate caries diagnosis? Acta Odontol Scand 2010; 68: 65–79. doi: 10.3109/00016350903530786 [DOI] [PubMed] [Google Scholar]
  • 28.Kidd EA, Fejerskov O. What constitutes dental caries? Histopathology of carious enamel and dentin related to the action of cariogenic biofilms. J Dent Res 2004; 83: C35–8. [DOI] [PubMed] [Google Scholar]
  • 29.Sansare K, Raghav M, Sontakke S, Karjodkar F, Wenzel A. Clinical cavitation and radiographic lesion depth in proximal surfaces in an Indian population. Acta Odontol Scand 2014; 72: 1084–8. doi: 10.3109/00016357.2014.926025 [DOI] [PubMed] [Google Scholar]
  • 30.Ferreira Zandoná A, Santiago E, Eckert GJ, Katz BP, Pereira de Oliveira S, Capin OR, et al. The natural history of dental caries lesions: a 4-year observational study. J Dent Res 2012; 91: 841–6. doi: 10.1177/0022034512455030 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Diagnosis and management of dental caries throughout life. National Institutes of Health Consensus Development Conference statement, March 26-28, 2001. J Dent Educ 2001; 65: 1162–8. [PubMed] [Google Scholar]
  • 32.Selwitz RH, Ismail AI, Pitts NB. Dental caries. The Lancet 2007; 369: 51–9. doi: 10.1016/S0140-6736(07)60031-2 [DOI] [PubMed] [Google Scholar]
  • 33.Braga MM, Mendes FM, Ekstrand KR. Detection activity assessment and diagnosis of dental caries lesions. Dent Clin North Am 2010; 54: 479–93. doi: 10.1016/j.cden.2010.03.006 [DOI] [PubMed] [Google Scholar]
  • 34.Wenzel A. Current trends in radiographic caries imaging. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1995; 80: 527–39. doi: 10.1016/S1079-2104(05)80152-0 [DOI] [PubMed] [Google Scholar]
  • 35.Kashima I. Computed radiography with photostimulable phosphor in oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1995; 80: 577–98. doi: 10.1016/S1079-2104(05)80155-6 [DOI] [PubMed] [Google Scholar]
  • 36.Huda W, Rill LN, Benn DK, Pettigrew JC. Comparison of a photostimulable phosphor system with film for dental radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1997; 83: 725–31. doi: 10.1016/S1079-2104(97)90327-9 [DOI] [PubMed] [Google Scholar]
  • 37.Obuchowski NA. Multireader receiver operating characteristic studies: a comparison of study designs. Acad Radiol 1995; 2: 709–16. doi: 10.1016/S1076-6332(05)80441-6 [DOI] [PubMed] [Google Scholar]
  • 38.Obuchowski NA. How many observers are needed in clinical studies of medical imaging? AJR Am J Roentgenol 2004; 182: 867–9. doi: 10.2214/ajr.182.4.1820867 [DOI] [PubMed] [Google Scholar]
  • 39.Park YS, Bae KH, Chang J, Shon WJ. Theory of X-ray microcomputed tomography in dental research: application for the caries research. J Korean Acad Conservative Dentistry 2011; 36: 98–107. doi: 10.5395/JKACD.2011.36.2.98 [DOI] [Google Scholar]
  • 40.Metz CE. Basic principles of ROC analysis. Semin Nucl Med 1978; 8: 283–98. doi: 10.1016/S0001-2998(78)80014-2 [DOI] [PubMed] [Google Scholar]
  • 41.Metz CE. ROC methodology in radiologic imaging. Invest Radiol 1986; 21: 720–33. doi: 10.1097/00004424-198609000-00009 [DOI] [PubMed] [Google Scholar]
  • 42.ten Bosch JJ, Angmar-Månsson B. Characterization and validation of diagnostic methods. Monogr Oral Sci 2000; 17: 174–89. [DOI] [PubMed] [Google Scholar]
  • 43.Tyndall DA, Rathore S. Cone-beam CT diagnostic applications: caries, periodontal bone assessment, and endodontic applications. Dent Clin North Am 2008; 52: 825–41. doi: 10.1016/j.cden.2008.05.002 [DOI] [PubMed] [Google Scholar]
  • 44.Hanzelka T, Dusek J, Ocasek F, Kucera J, Sedy J, Benes J, et al. Movement of the patient and the cone beam computed tomography scanner: objectives and possible solutions. Oral Surg Oral Med Oral Pathol Oral Radiol 2013; 116: 769–73. doi: 10.1016/j.oooo.2013.08.010 [DOI] [PubMed] [Google Scholar]
  • 45.Spin-Neto R, Mudrak J, Matzen LH, Christensen J, Gotfredsen E, Wenzel A. Cone beam CT image artefacts related to head motion simulated by a robot skull: visual characteristics and impact on image quality. Dentomaxillofac Radiol 2013; 42: 32310645. doi: 10.1259/dmfr/32310645 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Ludlow JB, Davies-Ludlow LE, White SC. Patient risk related to common dental radiographic examinations: the impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation. J Am Dent Assoc 2008; 139: 1237–43. doi: 10.14219/jada.archive.2008.0339 [DOI] [PubMed] [Google Scholar]
  • 47.Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008; 106: 106–14. doi: 10.1016/j.tripleo.2008.03.018 [DOI] [PubMed] [Google Scholar]
  • 48.Ludlow JB, Timothy R, Walker C, Hunter R, Benavides E, Samuelson DB, et al. Effective dose of dental CBCT-a meta analysis of published data and additional data for nine CBCT units. Dentomaxillofac Radiol 2015; 44: 20140197. doi: 10.1259/dmfr.20140197 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Rockette HE, Campbell WL, Britton CA, Holbert JM, King JL, Gur D. Empiric assessment of parameters that affect the design of multireader receiver operating characteristic studies. Acad Radiol 1999; 6: 723–9. doi: 10.1016/S1076-6332(99)80468-1 [DOI] [PubMed] [Google Scholar]
  • 50.Soviero VM, Leal SC, Silva RC, Azevedo RB. Validity of MicroCT for in vitro detection of proximal carious lesions in primary molars. J Dent 2012; 40: 35–40. doi: 10.1016/j.jdent.2011.09.002 [DOI] [PubMed] [Google Scholar]
  • 51.Hamba H, Nikaido T, Sadr A, Nakashima S, Tagami J. Enamel lesion parameter correlations between polychromatic micro-CT and TMR. J Dent Res 2012; 91: 586–91. doi: 10.1177/0022034512444127 [DOI] [PubMed] [Google Scholar]
  • 52.Zou W, Hunter N, Swain MV. Application of polychromatic µCT for mineral density determination. J Dent Res 2011; 90: 18–30. doi: 10.1177/0022034510378429 [DOI] [PMC free article] [PubMed] [Google Scholar]