Abstract
This study was aimed to establish a standard protocol and to quantitatively assess the reliability of harmonic analysis of the radial pulse wave measured by a harmonic wave analyzer (TD01C system). Both intraobserver and interobserver assessments were conducted to investigate whether the values of harmonics are stable in successive measurements. An intraclass correlation coefficient (ICC) and a Bland–Altman plot were used for this purpose. For the reliability assessments of the intraobserver and the interobserver, 22 subjects (mean age 45 ± 14 years; 14 males and 8 females) were enrolled. The first eleven harmonics of the radial pulse wave presented excellent repeatability (
and
) for the intraobserver assessment and high reproducibility (ICCs range from 0.83 to 0.96 and
) for the interobserver assessment. The Bland–Altman plot indicated that more than 90% of harmonic values fell within two standard deviations of the mean difference. Thus, we concluded that the harmonic analysis of the radial pulse wave using the TD01C system is a feasible and reliable method to assess a hemodynamic characteristic in clinical trial.
Keywords: Harmonic analysis, radial pulse wave, blood pressure, reliability
This study aimed to establish a standard protocol and to quantitatively assess the reliability of harmonics analysis of the radial pulse wave for monitoring hemodynamic status. Clinical trials in 22 subjects demonstrated that harmonic analysis of the radial pulse wave using the TD01C system is a feasible and reliable method to assess hemodynamic characteristics for further usage in cardiovascular diagnosis and intervention. Furthermore, the harmonic analysis of radial pulse wave can be viewed as a tool to evaluate the development of drug or therapy.

I. Introduction
Pulse wave analysis (PWA), extracting characteristic features of pressure pulse either from time domain or from frequency domain, has been extensively used in monitoring cardiovascular disease [1], [2]. Traditionally, the assessment of circulatory function relies on the brachial blood pressure measurement, where systolic pressure [3], diastolic pressure [4], [5], and pulse pressure [6] has been proved correlated with cardiovascular event. In recent years, the clinical interest has increasingly focused on arterial stiffness. Therefore, the markers of arterial stiffness such as pulse wave velocity (PWV) and augmentation index (AIx) were studied in aging [7], [8], hypertension [9], [10], coronary artery disease [11], [12], and renal disease [13], [14].
However, those variables mentioned above, extracting time-domain features from pulse, are only a partial description of the pressure wave [2]. To obtain a complete quantitative expression of the pressure wave, harmonic analysis (HA) is one of the most satisfactory approaches [2]. HA is a basic frequency analysis in periodic signal, using Fourier series calculation to decompose the arterial pulse wave into harmonic components, also abbreviated harmonics [15].
One branch of HA is generalized transfer function, where scaling relationships between two measurement sites were calculated in terms of amplitude and phase as a function of harmonics of pressure wave. O’rourke et al. applied the generalized transfer function to reconstruct the waveform of central aortic pressure wave from either the brachial or radial blood pressure wave [16].
Wang et al. used HA in another perspective, where those harmonic components were deemed as a set of basic characteristics of whole arterial system [17]. Thus, the variation of harmonics reveals variation of status of the arterial system, which has been proved a useful method in cardiovascular researches [18]. According to the observation of the previous researches, Lin Wang et al. developed a PS wave model to simulate the low damped oscillation of arterial pressure wave [19]–[24] and to elaborate the meaning of the harmonic components in cardiovascular physiology [25], [26].
Based on the PS wave model, Jan et al. proposed that HA can reveal the function of the organs [27]. Wang et al. found that harmonic components are correlated with liver function [28]–[30], with acute uncomplicated myocardial infection [31], and with hypertension [32], [33]. Furthermore, harmonic components are also used in examining the effects of Chinese herb [34]–[39] and acupuncture [40]–[42].
In spite of the various clinical application of HA were proposed, the reliability of HA in clinical study has not confirmed. The accuracy of harmonics can be affected by several measurement factors such as effective size of sensor, control of tensile force, and the net force on the tissue surface. Those measurement factors might cause potential errors and threaten the credibility of HA.
Therefore, validating the reliability of the harmonic components of the arterial pulse wave are necessary to ensure valid interpretation of HA, to determine the correlation between HA and cardiovascular disease, and to quantify the effects of Chinese herb and of mechanical therapies. In other word, a reliable instrument and a standard protocol should be applied to provide assurance that each time an individual is observed, the harmonic changes are due to the different arterial condition rather than inconsistent measurement capabilities of the instrument.
To achieve the goals mentioned above, the objective of the study was to investigate that, in successive measurements, whether the harmonic components are repeatable by same observer and are reproducible by different observers. Both of the measurements will calculate the intra-class correlation coefficients (ICC) and use Bland-Altman plot to examine the degree of the agreement.
II. Material and Methods
A. Harmonic Wave Analyzer TD01C
TD01C (MII-ANN Technology, Taiwan) is a noninvasive system that detects the blood pressure pulse of the radial artery and analyzed the pressure pulse in frequency domain. TD01C has an effective and continuous sensing surface (>2.0 cm
. TD01C has architectures of dynamic range adjustment and of automatic pulse-sensing such that it can analyze pressure data within minutes and reach resolution of the 11 harmonic waves. TD01C should operate within working condition:
PSI (peak to peak pressure), 23°C-35°C, and 15% to 85% Relative humidity. The instrument reliability of TD01C had been validated before the measurements were performed [43].
B. Study Protocol
Each assessment was performed after 10-minute rest in a quiet room at controlled temperatures of ranging from 23-25 °C. Drinking alcohol, tea, and coffee was not allowed for 4 hour before assessment. All the observers needed to read the instruction manual of harmonic wave analyzer TD01C. The standard process listed below for each assessment:
-
1.
Require the subject to sit up straight.
-
2.
Measure the wrist circumference of the subject.
-
3.
Find the skin surface over the radial artery with maximum pulse pressure, and mark it.
-
4.
Attach the TD01C sensor on the skin surface mentioned above.
-
5.
Strap the wristband with appropriate length corresponding to the wrist circumference.
-
6.
Check that signal exceeds 1/3 Full scale on the screen.
-
7.
Adjust the angle between the surface of the sensor and the skin, so that amplitude of the signal is maximized.
-
8.
Start the formal assessment.
C. Data Acquisition and Harmonic Analysis
In each measurement, the 12-second data were recorded by harmonic wave analyzer TD01C at sampling rate of 400 Hz (Figure 1). All pulses were separated sequentially by the minimum points from pulse peak to pulse peak. Each pulse was transformed into Fourier series coefficients (
and then normalized by the mean value of the pulse (
, where i indicated the ith pulse in twelve-second data and n indicated the nth harmonic components of the pulse. The representative amplitude of harmonic component (
within one measurement was defined by following equation, where N is the total number of pulses within the 12-second measurement.
![]() |
In this study, we focused on the first eleven harmonic components (
). For each harmonic component, the within-measurement standard deviation
and coefficient of variation (WCV
were calculated by following equation
![]() |
FIGURE 1.
A typical 12-second radial pulse wave recorded by harmonic wave analyzer TD01C at sampling rate of 400 Hz.
were representative parameters that described the characteristic of sequential pulses, and WCV
WCVc11 represented the degree of variation among sequential pulses within the measurement.
D. Assessment of Intra-Observer Repeatability
Twenty two subjects (aged 27-70; mean age 45 ± 14 years; 14 males and 8 females) were enrolled for the assessment of intra-observer repeatability from staffs and students of National Taiwan University. Observer 1 measured the harmonic components (
) of radial pulse wave twice for each participant with 5 minutes between measurements.
E. Assessment of Inter-Observer Reproducibility
Twenty two subjects (aged 27-70; mean age 45 ± 14 years; 14 males and 8 females) were enrolled for the assessment of intra-observer repeatability from staffs and students of National Taiwan University. For each subject, the harmonic components (
) of radial pulse wave were measured by Observer 1 and Observer 2 in random order. A minimum of 5 minutes was allowed between measurements, and wristband was retied at the second measurements.
Volunteer recruitment was after receiving approval from the institutional review board of the RenAi Branch of Taipei City Hospital (IRB number: TCHIRB1010710).
F. Statistics
1). Bland-Altman Analysis
The intra-observer and inter-observer reliability assessment was evaluated using the Bland-Altman analysis [44], which is widely accepted and used for reliability validation. For each harmonic component, the difference between successive measurements was plotted against their mean value. Then the mean (
and the standard deviation (SD) of differences for all participants were calculated. Limits of agreement (
SD) were marked by dash lines (Figure 2).
FIGURE 2.
The Bland-Altman plot of the harmonic components (C9, C10, C11) of the arterial pulse wave. The left part is the intra-observer repeatability assessment (
) and the right part is the inter-observer reproducibility assessment (
).
2). Intraclass Correlation Coefficient
For the intra-observer and inter-observer reliability assessment, two-way random single-measure intraclass correlation coefficients, ICC
, were calculated with Matlab 2008, USA. According to the suggestion of Fleiss [45], the value
was deemed as excellent reliability. Furthermore, Portney and Watkins [46] recommend that
is appropriate for clinical application to ensure valid interpretation (Table 1).
TABLE I. the intra-observer and the inter-observer
reliability assessment of the harmonic components of the arterial pulse waveusing intraclass correlation coefficient. For each harmonic component,SD is the standard deviation of differences (
between two of successive measurements.
| Intra-observer assessment | Inter-observer assessment | |||||
|---|---|---|---|---|---|---|
| SD | ICC | P value | SD | ICC | P value | |
| C1 | 0.027 | 0.97 | 3.1E-14 | 0.040 | 0.94 | 3.9E-11 |
| C2 | 0.022 | 0.96 | 2.6E-13 | 0.053 | 0.93 | 6.9E-11 |
| C3 | 0.032 | 0.95 | 1.9E-12 | 0.039 | 0.96 | 1.6E-13 |
| C4 | 0.014 | 0.97 | 1.2E-14 | 0.021 | 0.94 | 6.0E-12 |
| C5 | 0.010 | 0.96 | 6.1E-13 | 0.022 | 0.88 | 2.3E-08 |
| C6 | 0.016 | 0.98 | 1.1E-16 | 0.019 | 0.89 | 6.3E-09 |
| C7 | 0.013 | 0.95 | 5.6E-12 | 0.009 | 0.94 | 2.4E-11 |
| C8 | 0.007 | 0.91 | 6.1E-10 | 0.006 | 0.92 | 3.4E-10 |
| C9 | 0.005 | 0.97 | 6.0E-14 | 0.005 | 0.92 | 2.7E-10 |
| C10 | 0.005 | 0.95 | 6.1E-12 | 0.005 | 0.88 | 1.7E-08 |
| C11 | 0.004 | 0.97 | 1.0E-14 | 0.004 | 0.83 | 4.2E-07 |
III. Result
A. Result of Intra-Observer Assessment
Bland-Altman plots were constructed for all harmonics of intra-observer assessment. Three examples (C9, C10, C11) of the plot for pairs of measurements assessed by observer1 are shown in Figure 2. For all harmonics the mean differences were near to the zero. Intraclass correlation coefficient for repeatability, ICC (2, 1), were summarized in Table 1, which presents that all the ICCs exceeded 0.9 for all harmonics in the intra-observer case.
B. Result of Inter-Observer Assessment
Bland-Altman plots were used to visualize all harmonics of inter-observer assessment. Figure 2 shows three plots (C9, C10, C11) for successive measurements assessed by observer1 and observer 2 in a random order. For all harmonics the mean differences were all near to zero. Intraclass correlation coefficients for reproducibility, ICC (2, 1), were summarized in Table 1. ICCs of some harmonic components (C5, C6, C10, C11) reach 0.88, 0.89, 0.88, 0.83 respectively, and ICCs were above 0.9 for all the other harmonics case.
IV. Discussion
From the result of intra-observer assessment (Table 1), the ICC (2, 1) of the first five harmonics were all greater than 0.9 with p-value less than 0.001, which significantly confirmed that the variation of the harmonics in repeatability study were mostly caused by population variation rather than variation between two successive measurement. The reliability of repeated measurement reached the suggested criteria of the clinical application level [46].
The result of inter-observer assessment (Table 1) showed that the ICC (2,1) of the first five harmonics were all greater than 0.83 with p-value less than 0.001, which significantly confirmed that the variation of the harmonics in reproducibility study were mostly caused by population variation rather than variation caused by two observer. The reliability of reproducible measurement was good enough for Fleiss recommended level [45] but still had room for improvement.
In other studies, repeatability of AIx were examined, where intra-observer ICC ranged from 0.75 to 0.96 [47]–[49] and inter-observer ICC ranged from 0.92 to 0.93 [47]; Intra-observer ICC of PWV ranged from 0.75 to 0.92 [50]–[52] and inter-observer ICC of PWV was 0.88 [53]; The intra-observer ICC of central systolic blood pressure reconstruct from transfer function was 0.74 in resting state [48]. Comparing to other index of PWA, the harmonic components are competitive and reliable.
From aspect of signal analysis, the radial pulse wave is an almost periodic signal, which conserves the most energy in its fundamental frequency and harmonics. Furthermore, the harmonic components are orthogonal basis which completely represent the whole pulse signal. Therefore, to examine stability of the radial pulse wave, assessing reliability of its harmonic components is satisfactory and necessary. We suggest that all measurements of radial pulse wave should assess the reliability of harmonics before transforming the pulse into specific parameters.
HA has been used in cardiovascular study for many clinical applications. Jan et al. proposed that HA can reveal the function of the organs in pressure pulse spectrum [27], and Hsu et al. found that harmonic components could be a real time biomarker for liver function [28]. Since the TD01C is a noninvasive system that detects the radial artery pulse and has good reliability of HA, it reveals the good potential to assess the function of the organ in real time. Nevertheless, HA is suitable for the effects of taking Chinese medicine by percentage change of harmonics of radial artery pulse [54].
The effects of acupuncture at Tai-Tsih (K-3) [40], at Tsu San Li (St-36) [41], and at Hsien-Ku (St-43) [42] were studied using HA in health human. Furthermore, Hsiu proved that increments of fifth and sixth harmonics were related to increment of brain perfusion during acupuncture intervention in stroke patient [55]. Hence, harmonic amplitude increment has potential to be a real-time and noninvasive biomarker during or after intervention.
To estimate the effect of acupuncture or Chinese Herb, the differences of harmonic components between pre- and post-intervention should take intra-observer errors into consideration. According to the intra-observer result, we suggested that the change of
C11 should exceed SDs
respectively (Table 1) to interpret the change of specific harmonic component; roughly, to verify the change before and after intervention,
C2 should exceed about 5% change;
C5 should exceed about 10% change;
C8 should exceed about 20% change;
C11 should exceed about 40% change. If a smaller minimal detectable difference of harmonic was needed, the sample size should be re-estimated. Therefore, in order to overcome the intra-observer error and to validate the effective change of harmonics between measurements, this report could be used as a guideline and reference to estimate the number of subjects needed.
In addition, if a meta-analysis such as comparing acupuncture effect at K-3, at St-36, and at St-43 was performed, the inter-observer error should take into account. Thus, more minimal detectable difference is needed to overcome the effect inter-observer error. The methodology and criteria could be a guideline to select the results appropriate for further clinical interpretation. The criteria could also be a reference to estimate the appropriate sample size.
This report focused on harmonics amplitude between intra-observer and inter-observer measurements. The harmonic amplitudes were used as representative characteristics to estimate reliability of HA in arterial pulse wave. There are still two parameters need to be further investigated in the future. One is the coefficient of variation of harmonic amplitude among pulses within one measurement (WCV
. Since the WCV
were less than 3% using TD01C in the phantom study [43], the WCV
beyond 3% revealed that the variation were partially introduced by the arterial system of the subject. The previous reports had shown that this within-measurement variation of the harmonics reflected the stage of severity in dying patients [56], [57]. Therefore, the WCV
could also been meaningful biomarkers that reveals function or status of the arterial system and need more studies to prove the conjecture. Second parameter is the phase of harmonic components, which still need more exploratory studies.
In summary, the report confirmed that HA of radial pulse wave using TD01C system was a feasible and reliable method to assess hemodynamic characteristic in clinical study. Harmonic components has been proved correlated with some cardiovascular function and disease [18]. Since TD01C measures radial pulse wave noninvasively and takes only minutes to practice the standard protocol, HA can be used in monitoring cardiovascular disease in real time. The noninvasive nature of TD01C allows repeated measurements over time to study the effectiveness of various interventions that may affect cardiovascular system. This report is a beginning and a bridge to discover what the pressure wave tells us and to utilize the information in clinical practice.
V. Conclusion
The primary findings of this research demonstrate that, following the standard protocol, TD01C system is a reliable instrument for HA of radial pulse wave. TD01C system has not only the good repeatability of intra-observer measurement but also high reproducibility of inter-observer measurement.
The good reliability of harmonic analysis, combining with previous clinical finding in reference, showed the great niche and potential for further usage in cardiovascular diagnosis and intervention. The research built up the protocol and criteria for intra-observer and for inter-observer measurements to validate the effective change of harmonic component before and after intervention. This report could help bridging laboratory studies to clinical applications.
Acknowledgment
W.-K. Wang was with the Biophysics Laboratory, Institute of Physics, Academia Sinica, Taipei, Taiwan.
Biographies

Chi-Wei Chang received the B.S. degree in electrical engineering and the M.S. degree in biomedical electronics and bioinformatics from National Taiwan University, Taipei, Taiwan, in 2008 and 2010, respectively, where he is currently pursuing the Ph.D. degree with the Graduate Institute of Biomedical Electronics and Bioinformatics.

Jiang-Ming Chen received the B.S. degree in physics from National Dong Hwa University, Hualien, Taiwan, in 2008, and the M.S. degree in physics from National Taiwan Normal University, Taipei, Taiwan, in 2010.

Wei-Kung Wang received the B.S. degree in physics from National Taiwan University, Taipei, Taiwan, in 1966, the M.S. degree from National Tsing Hua University, Hsinchu, Taiwan, and the Ph.D. degree in biophysics from John Hopkins University, Baltimore, MD, in 1973. He served as a Professor and the Director of the Institute of Biomedical Engineering with National Yang Ming University, Taipei, from 1985 to 1987. He has been a Professor with the Department of Electrical Engineering, National Taiwan University, and a Research Fellow with the Biophysics Laboratory, Institute of Physics, Academia Sinica (joint appointment) since 1988.
References
- [1].O’Rourke M. F., Pauca A., and Jiang X. J., “Pulse wave analysis,” Brit. J. Clin. Pharmacol., vol. 51, no. , pp. 507–522, 2001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Milnor W. R., Hemodynamics. Baltimore, MD, USA: Williams & Wilkins, 1982. [Google Scholar]
- [3].Postel-Vinay N., A Century of Arterial Hypertension. New York, NY, USA: Wiley, 1996, pp. 1896–1996. [Google Scholar]
- [4].MacMahon S., et al. , “Blood pressure, stroke, and coronary heart disease: Part 1, prolonged differences in blood pressure: Prospective observational studies corrected for the regression dilution bias,” Lancet, vol. 335, no. 8692, pp. 765–774, Mar. 1990. [DOI] [PubMed] [Google Scholar]
- [5].Stamler J., Neaton J. D., and Wentworth D. N., “Blood pressure (systolic and diastolic) and risk of fatal coronary heart disease,” Hypertension, vol. 13, no. 5, pp. 1–12, 1989. [DOI] [PubMed] [Google Scholar]
- [6].Staessen J. A., et al. , “Risks of untreated and treated isolated systolic hypertension in the elderly: Meta-analysis of outcome trials,” Lancet, vol. 355, no. 9207, pp. 865–872, Mar. 2000. [DOI] [PubMed] [Google Scholar]
- [7].McEniery C. M., et al. , “Normal vascular aging: Differential effects on wave reflection and aortic pulse wave velocity: The anglo-cardiff collaborative trial (ACCT),” J. Amer. College Cardiol., vol. 46, no. 9, pp. 1753–1760, 2005. [DOI] [PubMed] [Google Scholar]
- [8].Schillaci G., et al. , “Age-specific relationship of aortic pulse wave velocity with left ventricular geometry and function in hypertension,” Hypertension, vol. 49, no. 2, pp. 317–321, 2007. [DOI] [PubMed] [Google Scholar]
- [9].Matsui Y., Ishikawa J., Shibasaki S., Shimada K., and Kario K., “Association between home arterial stiffness index and target organ damage in hypertension: Comparison with pulse wave velocity and augmentation index,” Atherosclerosis, vol. 219, no. 2, pp. 637–642, Dec. 2011. [DOI] [PubMed] [Google Scholar]
- [10].McCall D. O., McGartland C. P., Woodside J. V., Sharpe P., McCance D. R., and Young I. S., “The relationship between microvascular endothelial function and carotid-radial pulse wave velocity in patients with mild hypertension,” Clin. Experim. Hypertension, vol. 32, no. 7, pp. 474–479, 2010. [DOI] [PubMed] [Google Scholar]
- [11].Covic A., Haydar A. A., Bhamra-Ariza P., Gusbeth-Tatomir P., and Goldsmith D. J., “Aortic pulse wave velocity and arterial wave reflections predict the extent and severity of coronary artery disease in chronic kidney disease patients,” J. Nephrol., vol. 18, no. 4, pp. 388–396, 2005. [PubMed] [Google Scholar]
- [12].Koyoshi R., Miura S.-I., Kumagai N., Shiga Y., Mitsutake R., and Saku K., “Clinical significance of flow-mediated dilation, brachial intima-media thickness and pulse wave velocity in patients with and without coronary artery disease,” Circulat. J., vol. 76, no. 6, pp. 1469–1475, 2012. [DOI] [PubMed] [Google Scholar]
- [13].Hsu P.-F., et al. , “Differential effects of age on carotid augmentation index and aortic pulse wave velocity in end-stage renal disease patients,” J. Chin. Med. Assoc., vol. 71, no. 4, pp. 166–173, Apr. 2008. [DOI] [PubMed] [Google Scholar]
- [14].Kis É., et al. , “Pulse wave velocity in end-stage renal disease: Influence of age and body dimensions,” Pediatric Res., vol. 63, no. 3, pp. 95–98, 2008. [DOI] [PubMed] [Google Scholar]
- [15].Nichols W., O’Rourke M., and Vlachopoulos C., McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles. London, U.K.: Hodder Arnold, 2011. [Google Scholar]
- [16].Hirata K., Kawakami M., and O’Rourke M. F., “Pulse wave analysis and pulse wave velocity: A review of blood pressure interpretation 100 years after Korotkov,” Circulat. J., vol. 70, no. 10, pp. 1231–1239, 2006. [DOI] [PubMed] [Google Scholar]
- [17].Wang Y. Y., Chang S. L., Wu Y. E., Hsu T. L., and Wang W. K., “Resonance. The missing phenomenon in hemodynamics,” Circulat. Res., vol. 69, no. 1, pp. 246–249, 1991. [DOI] [PubMed] [Google Scholar]
- [18].Wang Y.-Y. L., Hsu T.-L., Jan M.-Y., and Wang W.-K., “Review: Theory and applications of the harmonic analysis of arterial pressure pulse waves,” J. Med. Biol. Eng., vol. 30, no. 3, pp. 125–131, 2010. [Google Scholar]
- [19].Wang Y.-Y. L., Chang C. C., Chen J. C., Hsiu H., and Wang W. K., “Pressure wave propagation in arteries. A model with radial dilatation for simulating the behavior of a real artery,” IEEE Eng. Med. Biol. Mag., vol. 16, no. 1, pp. 51–54, Jan-Feb 1997. [DOI] [PubMed] [Google Scholar]
- [20].Wang Y.-Y. L., Jan M.-Y., Wang G.-C., Bau J.-G., and Wang W.-K., “Pressure pulse velocity is related to the longitudinal elastic properties of the artery,” Physiol. Meas., vol. 25, no. 6, pp. 1397–1403, 2004. [DOI] [PubMed] [Google Scholar]
- [21].Wang Y.-Y. L., et al. , “The ventricular-arterial coupling system can be analyzed by the eigenwave modes of the whole arterial system,” Appl. Phys. Lett., vol. 92, no. 15, p. 153901, 2008. [Google Scholar]
- [22].Wang Y.-Y. L., “Analysis of transverse wave as a propagation mode for the pressure pulse in large arteries,” J. Appl. Phys., vol. 102, no. 6, p. 064702, 2007. [Google Scholar]
- [23].Wang Y.-Y. L. and Wang W.-K., “A hemodynamics model to study the collective behavior of the ventricular-arterial system,” J. Appl. Phys., vol. 113, no. 2, pp. 024702-1–024702-6, 2013. [Google Scholar]
- [24].Wang Y.-Y. L., et al. , “Examining the response pressure along a fluid-filled elastic tube to comprehend Frank’s arterial resonance model,” J. Biomech., vol. 48, no. 6, pp. 907–910, Apr. 2015. [DOI] [PubMed] [Google Scholar]
- [25].Wang Y.-Y. L. and Wang W.-K., “Anatomy of arterial systems reveals that the major function of the heart is not to emit waves associated with the axial blood motion,” J. Physiol., vol. 592, no. 3, p. 409, Jan. 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Wang Y.-Y. L. and Wang W.-K., “From a basic principle of evolution to the heart rate of mammals,” J. Physiol., vol. 593, no. 9, pp. 2241–2242, May 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [27].Jan M.-Y., Hsiu H., Hsu T.-L., Wang W.-K., and Wang Y.-Y. L., “The physical conditions of different organs are reflected specifically in the pressure pulse spectrum of the peripheral artery,” Cardiovascular Eng., vol. 3, no. 1, pp. 21–29, Mar. 2003. [Google Scholar]
- [28].Hsu T.-L., Chiang Y., Wang W. K., Chao P. T., Bao J. G., and Wang Y.-Y. L., “Pulse analysis as a possible real-time biomarker complementary to SGPT and SGOT for monitoring acute hepatotoxicity,” Toxicol. Mech. Methods, vol. 13, no. 3, pp. 181–186, 2003. [DOI] [PubMed] [Google Scholar]
- [29].Wang W. K., Tsuei J., Chang H. C., Hsu T. L., and Wang Y.-Y. L., “Pulse spectrum analysis of chemical factory workers with abnormal blood test,” Amer. J. Chin. Med., vol. 24, no. 3, pp. 199–203, 1996. [DOI] [PubMed] [Google Scholar]
- [30].Lu W. A., Wang Y.-Y. L., and Wang W. K., “Pulse analysis of patients with severe liver problems. Studying pulse spectrums to determine the effects on other organs,” IEEE Eng. Med. Biol. Mag., vol. 18, no. 1, pp. 73–75, Jan-Feb 1999. [DOI] [PubMed] [Google Scholar]
- [31].Chen C.-Y., Wang W.-K., Kao T., Yu B. C., and Chiang B. C., “Spectral analysis of radial pulse in patients with acute, uncomplicated myocardial infarction,” Jpn. Heart J., vol. 34, no. 2, pp. 131–143, 1993. [DOI] [PubMed] [Google Scholar]
- [32].Wang S. H., Jan M. Y., Wang W. K., and Wang Y.-Y. L., “Effects of antihypertensive drugs on specific harmonic indices of the pulse waveform in normotensive Wistar Kyoto rats,” Clin. Experim. Hypertension, vol. 34, no. 1, pp. 74–78, 2012. [DOI] [PubMed] [Google Scholar]
- [33].Wang S.-H., Wang W.-K., Hsu T.-L., Jan M.-Y., and Wang Y.-Y. L., “Effects of captopril on specific harmonic indexes of the peripheral pressure pulse waveform,” in Proc. 4th Int. Conf. Bioinformat. Biomed. Eng. (iCBBE), Jun. 2010, pp. 1–3. [Google Scholar]
- [34].Wang Y.-Y. L., Sheu J. I., and Wang W.-K., “Alterations of pulse by chinese herb medicine,” Amer. J. Chin. Med., vol. 20, no. 3, pp. 181–190, 1992. [DOI] [PubMed] [Google Scholar]
- [35].Wang W. K., Hsu T. L., Chiang Y., and Wang Y.-Y. L., “Pulse spectrum study on the effect of sie-zie-tang and radix aconiti,” Amer. J. Chin. Med., vol. 25, nos. 3–4, pp. 357–366, 1997. [DOI] [PubMed] [Google Scholar]
- [36].Wang W. K., Hsu T. L., and Wang Y.-Y. L., “Liu-Wei-Dihuang: A study by pulse analysis,” Amer. J. Chin. Med., vol. 26, no. 3, pp. 73–82, 1998. [DOI] [PubMed] [Google Scholar]
- [37].Wang W. K., Bau J. G., Hsu T. L., and Wang Y.-Y. L., “Influence of spleen meridian herbs on the harmonic spectrum of the arterial pulse,” Amer. J. Chin. Med., vol. 28, no. 3, pp. 279–289, 2000. [DOI] [PubMed] [Google Scholar]
- [38].Wang W.-K., Hsu T.-L., Bau J.-G., and Wang-Lin Y.-Y., “Evaluation of herbal formulas by pulse analysis method,” Acta Pharmacol. Sinica, vol. 24, no. 2, pp. 145–151, 2003. [PubMed] [Google Scholar]
- [39].Wang W. K., Hsu T. L., Huang Z. Y., and Wang Y. Y. L., “Collective effect of a Chinese formula—A study of Xiao-Jian-Zhong-Tang,” Amer. J. Chin. Med., vol. 23, nos. 3–4, pp. 299–304, 1995. [DOI] [PubMed] [Google Scholar]
- [40].Wang W. K., Hsu T. L., Chang H. C., and Wang Y. Y. L., “Effect of acupuncture at Tai–Tsih (K-3) on the pulse spectrum,” Amer. J. Chin. Med., vol. 24, nos. 3–4, pp. 305–313, 1996. [DOI] [PubMed] [Google Scholar]
- [41].Wang W. K., Hsu T. L., Chang H. C., and Wang Y.-Y. L., “Effect of acupuncture at Tsu San Li (St-36) on the pulse spectrum,” Amer. J. Chin. Med., vol. 23, no. 3, pp. 121–130, 1995. [DOI] [PubMed] [Google Scholar]
- [42].Wang W. K., Hsu T. L., Chang H. C., and Wang Y.-Y. L., “Effect of acupuncture at Hsien-Ku (St-43) on the pulse spectrum and a discussion of the evidence for the frequency structure of Chinese medicine,” Amer. J. Chin. Med., vol. 28, no. 1, pp. 41–55, 2000. [DOI] [PubMed] [Google Scholar]
- [43].Chang C.-W. and Wang W.-K., “Reliability assessment for pulse wave measurement using artificial pulse generator,” J. Med. Eng. Technol., vol. 39, no. 3, pp. 177–184, 2015. [DOI] [PubMed] [Google Scholar]
- [44].Bland J. M. and Altman D. G., “Statistical methods for assessing agreement between two methods of clinical measurement,” Lancet, vol. 327, no. 8476, pp. 307–310, Feb. 1986. [PubMed] [Google Scholar]
- [45].Fleiss J. L., “Front matter,” in The Design and Analysis of Clinical Experiments. New York, NY, USA: Wiley, 1999, pp. 1–14. [Google Scholar]
- [46].Portney L. G. and Watkins M. P., Foundations of Clinical Research: Applications to Practice. Englewood Cliffs, NJ, USA: Prentice-Hall, 2009. [Google Scholar]
- [47].Crilly M., Coch C., Clark H., Bruce M., and Williams D., “Repeatability of the measurement of augmentation index in the clinical assessment of arterial stiffness using radial applanation tonometry,” Scandin. J. Clin. Lab. Invest., vol. 67, no. 4, pp. 413–422, 2007. [DOI] [PubMed] [Google Scholar]
- [48].Holland D. J., Sacre J. W., McFarlane S. J., Coombes J. S., and Sharman J. E., “Pulse wave analysis is a reproducible technique for measuring central blood pressure during hemodynamic perturbations induced by exercise,” Amer. J. Hypertension, vol. 21, no. 3, pp. 1100–1106, 2008. [DOI] [PubMed] [Google Scholar]
- [49].Frimodt-Møller M., “Reproducibility of pulse-wave analysis and pulse-wave velocity determination in chronic kidney disease,” Nephrol. Dialysis Transplantation, vol. 23, no. 3, pp. 594–600, 2008. [DOI] [PubMed] [Google Scholar]
- [50].Miyatani M., Masani K., Moore C., Szeto M., Oh P., and Craven C., “TestŰretest reliability of pulse wave velocity in individuals with chronic spinal cord injury,” J. Spinal Cord Med., vol. 35, no. 3, pp. 400–405, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].Vivodtzev I., et al. , “Arterial stiffness by pulse wave velocity in COPD: Reliability and reproducibility,” Eur. Respirat. J., vol. 42, no. 3, pp. 1140–1142, 2013. [DOI] [PubMed] [Google Scholar]
- [52].Chae Y. M. and Park J. K., “The relationship between brachial ankle pulse wave velocity and complement 1 inhibitor,” J. Korean Med. Sci., vol. 24, no. 5, pp. 831–836, 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [53].Li P., et al. , “Determinants of Brachial–Ankle pulse wave velocity in Chinese patients with rheumatoid arthritis,” Clin. Develop. Immunol., vol. 2013, p. 342869, Jul. 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [54].Wang W. K., Chen H. L., Hsu T. L., and Wang Y.-Y. L., “Alteration of pulse in human subjects by three Chinese herbs,” Amer. J. Chin. Med., vol. 22, no. 2, pp. 197–203, 1994. [DOI] [PubMed] [Google Scholar]
- [55].Hsiu H., Hsu C. L., Chen C. T., Hsu W. C., and Lin F. C., “Effects of acupuncture on the harmonic components of the radial arterial blood-pressure waveform in stroke patients,” Biorheology, vol. 50, nos. 3–4, pp. 69–81, 2013. [DOI] [PubMed] [Google Scholar]
- [56].Kuo Y.-C., et al. , “Losing harmonic stability of arterial pulse in terminally ill patients,” Blood Pressure Monitor., vol. 9, no. 5, pp. 255–258, 2004. [DOI] [PubMed] [Google Scholar]
- [57].Scolletta S., et al. , “Assessment of left ventricular function by pulse wave analysis in critically ill patients,” Intensive Care Med., vol. 39, no. 6, pp. 1025–1033, Jun. 2013. [DOI] [PubMed] [Google Scholar]




