Abstract
We present here a microfabricated, multi-functional neural interface with the ability to selectively apply electrical and chemical stimuli, while simultaneously monitoring both electrical and chemical activity in the brain. Such a comprehensive approach is required to understand and treat neuropsychiatric disorders, such as major depressive disorder (MDD), and to understand the mechanisms underlying treatments, such as pharmaceutical therapies and deep brain stimulation (DBS). The polymer-based, multi-functional neural interface is capable of electrical stimulation and recording, targeted drug delivery, and electrochemical sensing. A variety of different electrode and fluidic channel arrangements are possible with this fabrication process. Preliminary testing has shown the suitability of these neural interfaces for in vivo electrical stimulation and recording, as well as in vitro chemical sensing. Testing of the in vitro drug delivery and combined in vivo functionalities this neural interface are currently underway.
I. Introduction
Recently, deep brain stimulation (DBS), a method of electrically stimulating specific brain regions using an implanted electrode, has been approved by the FDA for the treatment of symptoms of Parkinson’s disease, essential tremor, and dystonia. More recently, DBS has also been applied to a variety of neuropsychiatric disorders, such as obsessive-compulsive disorder (OCD) and major depressive disorder (MDD) [1–4]. Ongoing studies of DBS targeting the subcallosal cingulate white matter have shown sustained, long-term, antidepressant response without appreciable side effects in otherwise treatment resistant patients [1, 3–4]. Despite these encouraging results, DBS mechanisms are unknown, slowing refinement of surgical targeting and stimulation parameters that might further optimize clinical outcomes.
To understand a subject as complex as brain function and to treat an illness as elusive as treatment-resistant depression, a comprehensive approach is required. We are developing a multifunctional neural interface to selectively apply electrical and pharmacological treatments to a behaving animal, while simultaneously monitoring electrical and neurochemical brain activity.
Some of the most common neural interfaces are thin-film micromachined probes, utilizing either silicon or polymer backbones [ 5–18]. In general, these devices have limited functionality, allowing for only electrical stimulation and/or recording [ 5–6, 11–15] or only chemical sensing [ 16]. Initial work has concentrated on incorporating fluidic channels for drug delivery into these neural interfaces [7–10, 17–18]. Recently, Frey et al. have demonstrated a silicon-based neural interface with electrical stimulation/recording, drug delivery, and chemical sensing [10]. However, there are concerns regarding the suitability of silicon-based neural interfaces for long-term in vivo studies [19]. Further, the continuous micro-motion of the neural tissue can induce strain between the neural tissue and the implanted interface, inducing chronic injury and glial scarring at the implant site. Unlike silicon-based, polymer-based neural interfaces are fully flexible. Thus, the strain between the neural tissue and the implanted probe is minimized, thereby minimizing injury and scarring at the implantation site [20].
We present here a microfabricated, multifunctional, polymer-based neural interface (Fig 1.). The multifunctional interface is capable of electrical stimulation/recording, targeted drug delivery, and electrochemical sensing. The fabrication process is extremely versatile, allowing a variety of different electrode and fluidic channel geometries to be produced.
Figure 1.
Model of the multifunctional neural interface presented here.
II. Multifunctional Neural Interface Fabrication
A. Neural Interface Fabrication
The fabrication process depicted here can be used for a variety of different electrode quantities (8 – 120), sizes (10 μm to 2 mm), and materials (e.g. platinum, iridium, iridium-oxide, gold). This process can also be used for a variety of different fluidic channel dimensions: height (1 μm – 100 μm), width (10 μm – 1mm), and length (5 mm – 100 mm). In addition, multiple fluidic channels can be included on a single neural interface.
A model of the multifunctional neural interface is shown in Fig. 1. This multifunctional device contains: (1) electrodes for both electrical stimulation and recording, (2) fluidic channels for targeted drug delivery, and (3) post-fabrication, surface-modified electrodes for electrochemical sensing. The fabrication process is versatile, allowing for a variety of different orientations of electrodes and fluidic channels (Fig. 2). The fabrication steps are as follows:
Figure 2.
Fabrication process of the multifunctional neural interface.
Etch pillars into the silicon at the electrode and electrical connector locations (Fig. 2A) using a standard Bosch process. These pillars allow the electrodes to be recessed from the surface of the parylene, which creates more uniform electric fields during electrical stimulation [21]. The height of the pillars defines the recess depth of the electrodes. The first layer of parylene is then deposited and O2 plasma is used to remove the parylene at the pillar locations.
The electrode and trace metals are then deposited and patterned (Fig. 2B). A variety of electrode metals can be used (platinum, iridium, and iridium-oxide are the most common). The trace metal is typically gold. The metals can be patterned using either liftoff or etching (wet or dry) techniques.
The next layer of parylene is then deposited (Fig. 2C) and interconnection vias are etched through this plasma (Fig. 2D) at both the parylene layer using O2 electrode and electrical connector locations. The second layer of trace metal is then deposited and patterned. The process is then repeated for a third layer of parylene and a third layer of trace metal (Fig. 2E).
The parylene is then through-etched using O2 plasma to create the fluidic inlet (Fig. 2F). This places the fluidic inlet on the same side of the device as the electrodes and the electrical connector. Although not shown in the fabrication process depicted here, if the fluidic outlet is desired on the same side of the device as the electrodes, this would also be etched at this time. Similarly, if the fluidic inlet is desired on the opposite side of the device, this step can be omitted.
The sacrificial photoresist used to define the fluidic channel is patterned and the final layer of parylene is then deposited (Fig. 2G). The fluidic outlet is then etched through the parylene using O2 plasma (Fig. 2H). The location of this fluidic outlet on either side of the neural interface (as opposed to the tip of the device) is crucial to prevent occlusion during insertion [22].
Finally, the sacrificial photoresist in the fluidic channel is removed and the devices are released from the silicon (Fig. 2I). Individual electrodes are then surface-modified to create the chemical sensors.
The process used here can be used with varying numbers of trace metal layers (up to 4 layers of trace metal have been demonstrated [23]) and varying orientations of electrodes and fluidic channels. With this process it is possible to put electrodes on both sides of the devices and/or inside the fluidic channels.
B. Electrochemical Sensor
Post-fabrication chemical modifications are performed on individual electrodes. These electrodes can be selectively modified to detect multiple analytes. As a representative chemical sensor, an amperometric dopamine (DA) sensor was developed. Platinum electrodes were chemically modified with permselective polymers to allow for the detection of dopamine and rejection of interferents [24]. Fig. 3 shows the results of an in vitro calibration of a DA sensor. The sensor shows a step response to each injection of 10μm of DA but no response to injections of 250 μm of ascorbic acid, a common interferent present in extracellular fluid. Additionally, the sensor responds strongly to 1 μM hydrogen peroxide, making the device suitable for use as an enzymatic sensor when modified with appropriate enzymes.
Figure 3.
In vitro calibration showing a step response to 5 μM dopamine (DA) and 1 μM hydrogen peroxide (H2O2), but no response to the interferent ascorbic acid (AA) at 250 μM per injection
C. Electrical and Fluidic Connectors
A variety of different electrical connectors can be used with this neural interface, depending upon the application. Most standardly used are the Omnetics (Omnetics Connector Co., Minneapolis, MN) and the zero-insertion-force (ZIF) connectors. This neural interface can also be used with specially-designed, percutaneous connectors for long-term use, such as that designed by Shah et al. [25]. Alternately, specialized connectors, with on-board electronics and wireless telemetry, can also be used [ 15]. All of these methods allow for reliable and repeatable electrical connections.
The most commonly used fluidic connectors for drug delivery consist of tubing epoxied to the neural interface [8–9, 17]. The neural interface presented here can be used with this type of interface.
D. Polymer Neural Probe Insertion Stiffeners
A major challenge with these polymer-based neural interfaces is their flexibility as the probes tend to buckle upon insertion, thereby preventing successful penetration into neural tissue. We have developed a method for temporarily attaching the neural interface to a silicon stiffener with biodissolvable polyethylene glycol (PEG) [26]. The silicon can easily penetrate the neural tissue, and once inserted, the PEG dissolves, the silicon stiffener can be extracted, and the polymer probe regains its intended flexibility. The removable silicon stiffener has been tested with devices that have electrical stimulation and recording capabilities and no loss of functionality has been seen due to the PEG dissolution or the removal of the stiffener. Plans are underway to test the removal silicon stiffeners with the chemical sensing and delivery functionalities.
III. Neural Interface Characterization
Images of the multifunctional neural interfaces are shown in Fig. 4. The images show both multi-shank and single-shank interfaces, with both top (fluidic channel is on top) and bottom views (electrodes are on top). As can be seen, the fabrication process is compatible with a variety of different probe, fluidic channel, and electrode designs.
Figure 4.
Images of three different multifunctional neural interfaces. The top image shows a probe with 4 separate shanks, where each shank has a fluidic channel. The middle image shows a single-shank design. The bottom image shows a single-shank design with a tetrode array. The top and middle images show top-down views of the neural interfaces (i.e. the fluidic channel is on top). The bottom image shows a bottom-up view of the neural interface (i.e. the electrodes are on top).
A. In Vitro Electrical Characterization
Both iridium and platinum electrodes were tested. For the iridium electrodes, activation was performed using biphasic potential pulsing in phosphate-buffered saline to form an activated iridium oxide film (AIROF). All electrodes were characterized to determine charge storage capacity (CSC) and impedance. Cyclic voltammetry (CV) and electrochemical impedance measurements were made with a Princeton Applied Research (PAR) potentiostat using vendor-supplied software. All measurements were made in a three-electrode cell using a Pt counter electrode, an Ag/ AgCl reference electrode, and phosphate-buffered saline (pH 7.4) as the electrolyte. Potential cycling for the CVs was performed between −600 m V and +800 m V at a scan rate of 100 mV/s. The CSC for a typical 100 μm AIROF electrode is 21.8 mC/cm2, with an impedance of 3.3 kΩ (at f= 1 kHz). Other electrode sizes exhibit similar results. The CSC for a typical 100 μm platinum electrode is 1.72 mC/cm2, with an impedance of 42.2 kΩ (at f= 1 kHz).
B. Preliminary In Vivo Testing
Preliminary in vivo testing of these neural interfaces was performed to demonstrate their efficacy in both neural electrical stimulation and recording. For these experiments, two neural interfaces were used: one inserted into the basolateral amygdala (BLA) and the other inserted into the medial prefrontal cortex (mPFC).
1) Neural Recording
For this test, one of the neural interfaces presented here was implanted in the BLA and used as the recording array. A commercial monopolar electrode was implanted in the mPFC and used for stimulation. Fig. 5 shows the evoked response potentials (ERPs) recorded using the neural interface in the BLA.
Figure 5.
Data recorded in the BLA by the multifunctional neural interface (following stimulation in the mPFC).
2) Neural Stimulation
For this test, data was recorded from the BLA using a commercial microwire array during stimulation from the mPFC using a neural interface presented here. Fig. 6 shows the data recorded after stimulation.
Figure 6.
Data recorded in the BLA following stimulation in the mPFC by the multifunctional neural interface.
IV. Conclusion
We present a complete microelectrode array designed for electrical stimulation and recording and chemical delivery and sensing. We have demonstrated a versatile fabrication process for multifunctional neural interfaces. The neural interface was designed to be implanted in soft tissue using a removable stiffener. Preliminary testing has shown the suitability of these neural interfaces for in vivo electrical stimulation and recording and in vitro chemical sensing. In vitro testing of the fluidic channels for drug delivery is currently underway. Additional in vivo testing of the drug delivery and chemical sensing functions of this neural interface is also planned.
Acknowledgments
The authors would like to thank Terri Delima, Heeral Sheth, and William Benett at Lawrence Livermore National Laboratory for their lab support. This work has been funded by Lawrence Livermore National Laboratory. This work performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under Contract DE-AC52-07NA27344. LLNL-ABS-612192
Contributor Information
Angela Tooker, Email: tookerl@llnl.gov, Lawrence Livermore National Laboratory, Livermore, CA 94550 USA phone: 925-422-2326; fax: 925-422-2373.
Teresa E. Madsen, Email: tmadsen@emory.edu, Emory University School of Medicine, Atlanta, GA
Allison Yorita, Email: ayorita@ucla.edu, Chemical and Biomolecular Engineering Department, University of California, Los Angeles, CA.
Andrea Crowell, Email: andrea.crowell@emory.edu, Emory University School of Medicine, Atlanta, GA.
Kedar G. Shah, Email: shah22@llnl.gov, Lawrence Livermore National Laboratory, Livermore, CA 94550 USA
Sarah Felix, Email: felix5@llnl.gov, Lawrence Livermore National Laboratory, Livermore, CA 94550 USA.
Helen S. Mayberg, Email: hmayber@emory.edu, Department of Psychiatry, Emory University, Atlanta, GA
Satinderpall Pannu, Email: pannul@llnl.gov, Lawrence Livermore National Laboratory, Livermore, CA 94550 USA.
Donald G. Rainnie, Email: drainni@emory.edu, Emory University School of Medicine, Atlanta, GA
Vanessa Tolosa, Email: tolosal@llnl.gov, Lawrence Livermore National Laboratory, Livermore, CA 94550 USA.
References
- 1.Mayberg H, Lozano A, Voon V, McNealy H, Seminowicz D, Hamani C, Schwalb J, Kennedy S. Deep Brain Stimulation for Treatment-Resistant Depression. Neuron. 2005;45:651–660. doi: 10.1016/j.neuron.2005.02.014. [DOI] [PubMed] [Google Scholar]
- 2.Nuttin B, Cosyns P, Demeulemeester H, Gybels J, Meyerson B. Electrical stimulation in anterior limbs of internal capsules in patients with obsessive-compulsive disorder. Lancet. 1999 Oct; doi: 10.1016/S0140-6736(99)02376-4. [DOI] [PubMed] [Google Scholar]
- 3.Greenberg BD, Malone DA, Friehs GM, Rezai AR, Kubu CS, Malloy PF, Salloway SP, Okun MS, Goodman WK, Rasmussen SA. Three-year outcomes in deep brain stimulation for highly resistant obsessive-compulsive disorder. Neuropsychopharm. 2006 Nov;31(11):2384–93. doi: 10.1038/sj.npp.1301165. [DOI] [PubMed] [Google Scholar]
- 4.Kennedy SH, Giacobbe P, Rizi SJ, Placenza FM, Nishikawa Y, Mayberg HS, Lozano AM. Deep Brain Stimulation for Treatment-Resistant Depression: Follow-up after 3 to 6 years. Am J Psych. 2011;168:502–510. doi: 10.1176/appi.ajp.2010.10081187. [DOI] [PubMed] [Google Scholar]
- 5.Sharma A, Rieth L, Tathireddy P, Harrison R, Oppermann H, Klein M, Topper M, Jung E, Normann R, Clark G, Solzbacher F. Long term in vitro functional stability and recording longevity of fully integrated wireless neural interfaces based on the Utah Slant Electrode Array. J Neural Eng. 2011 Aug;8(4) doi: 10.1088/1741-2560/8/4/045004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Chestek C, Gilja V, Nuyujukian P, Foster J, Fan J, Kaufman M, Churchland M, Rivera-Alvidrez Z, Cunningham J, Ryu S, Shenoy K. Long-term stability of neural prosthetic control signals from silicon cortical arrays in rhesus macaque motor cortex. J Neural Eng. 2011 Aug;8(4) doi: 10.1088/1741-2560/8/4/045005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Li Y, Baek K, Gulari M, Wise KD. A Drug-Delivery Probe with an In-Line Flowmeter Based on Trench Refill and Chemical Mechanical Polishing Techniques. IEEE Sensors 2007 Conf; 2007; pp. 1144–1147. [Google Scholar]
- 8.Cheung KC, Djupsund K, Dan Y, Lee LP. Implantable Multichannel Electrode Array Based on SOI Technology. JMEMS. 2003 Apr;12(2):179–184. [Google Scholar]
- 9.Pellinen DS, Moon T, Vetter RJ, Miriani R, Kipke DR. Multifunctional Flexible Parylene-Based Intracortical Microelectrodes. Proc. 27th IEEE Eng. Med. Biol. Soc. Ann. Int. Conf. 2005; Shanghai China. September 2005; pp. 5272–5275. [DOI] [PubMed] [Google Scholar]
- 10.Frey O, van der Wall PD, Spieth S, Brett O, Seidl K, Paul O, Ruther P, Zengerle R, de Rooij NF. Biosensor microprobes with integrated microfluidic channels for bi-directional neural chemical interaction. J Neural Eng. 2011;8 doi: 10.1088/1741-2560/8/6/066001. [DOI] [PubMed] [Google Scholar]
- 11.Lee S, Jung J, Chae Y, Suh J-K, Kang J. Fabrication and characterization of implantable flexible nerve cuff electrodes with Pt., Ir, and IrOx films deposited by RF sputtering. J Micromech Microeng. 2010 Mar;20(3) [Google Scholar]
- 12.Rubehn B, Bosman C, Oostenveld R, Fries P, Stieglitz T. A MEMS-based flexible multi-channel ECoG-electrode array. J Neural Eng. 2009 Jun;6(3) doi: 10.1088/1741-2560/6/3/036003. [DOI] [PubMed] [Google Scholar]
- 13.Seymour J, Langhals N, Anderson D, Kipke D. Novel multi-side, microelectrode arrays for implantable neural applications. Biomed Microdevices. 2011 Feb;13:441–451. doi: 10.1007/s10544-011-9512-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Mercanzini A, Cheung K, Buhl D, Boers M, Maillard A, Colin P, Bensadoun JC, Bertsch A, Renaud P. Demonstration of cortical recording using novel flexible polymer neural probes. Sens Actuators A. 2008;143:90–96. [Google Scholar]
- 15.Tooker A, Shah KG, Tolosa V, Sheth H, Felix S, Delima T, Pannu S. Chronically Implantable, 121-Channel, Polymer Microelectrode Array with Hermetically-Sealed Wireless Interface. Proc. Transducers 2012; Hilton Head, SC, USA. June 2012. [Google Scholar]
- 16.Wassum KM, Tolosa VM, Wang J, Walker E, Monbouquette HG, Maidment NT. Silicon Wafer-Based Platinum Microelectrode Array Biosensor for Near Real-Time Measurement of Glutamate in Vivo. Sensors. 2008;8:5023–5036. doi: 10.3390/s8085023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Ziegler D, Suzuki T, Takeuchi S. Fabrication of Flexible Neural Probes with Built-In Microfluidic Channels by Thermal Bonding of Parylene. J MEMS. 2006 Dec;15(6):1477–1482. [Google Scholar]
- 18.Metz S, Bertsch A, Bertrand D, Renaud P. Flexible polyimide probes with microelectrodes and embedded microfluidic channels for simultaneous drug delivery and multi-channel monitoring of bioelectric activity. Biosens Bioelectron. 2004;19:1309–1318. doi: 10.1016/j.bios.2003.11.021. [DOI] [PubMed] [Google Scholar]
- 19.Biran R, Martin D, Tresco P. Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays. Exp Neurology. 2005;195:115–126. doi: 10.1016/j.expneurol.2005.04.020. [DOI] [PubMed] [Google Scholar]
- 20.Polikov V, Tresco P, Reichert W. Response of brain tissue to chronically implanted neural electrodes. J Neurosci Meth. 2005;148:1–18. doi: 10.1016/j.jneumeth.2005.08.015. [DOI] [PubMed] [Google Scholar]
- 21.Suesserman M, Spelman F, Rubinstein J. In Vitro Measurement and Characterization of Current Density Profiles Produced by Nonrecessed, Simple Recessed, and Radially Varying Recessed Stimulating Electrodes. IEEE Trans Biomed Eng. 1991;38(5):401–408. doi: 10.1109/10.81558. [DOI] [PubMed] [Google Scholar]
- 22.Neeves KB, Lo CT, Foley CP, Saltzman WM, Olbricht WL. Fabrication and characterization of microfluidic probes for convection enhanced drug delivery. J Cont Release. 2006 Feb;111:252–262. doi: 10.1016/j.jconrel.2005.11.018. [DOI] [PubMed] [Google Scholar]
- 23.Tooker A, Tolosa V, Shah KG, Sheth H, Felix S, Delima T, Pannu S. Optimization of Multi-Layer Metal Neural Probe Design. Proc. 34th IEEE Eng. Med. Biol. Soc. Ann. Int. Conf. 2012; California, USA. August 2012; [DOI] [PubMed] [Google Scholar]
- 24.Tseng TT, Monbouquette HG. Implantable Microprobe with Arrayed Microsensors for Combined Amperometric Monitoring of the Neurotransmitters, Glutamate and Dopamine. J Electroanal Chem. 2012 Aug;682:141–145. doi: 10.1016/j.jelechem.2012.07.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Shah KG, Benett W, Delima T, Felix S, Sheth H, Tolosa V, Tooker A, Pannu S. Percutaneous Electrical Connector for Long-term Electrical Recording and Stimulation with Microelectrode Arrays. presented at the Neural Interfaces Conference; June 2012. [Google Scholar]
- 26.Felix S, Shah KG, George D, Tolosa V, Tooker A, Sheth H, Delima T, Pannu S. Removable Silicon Insertion Stiffeners for Neural Probes using Polyethylene Glycol as a Biodissolvable Adhesive. Proc. 34th IEEE Eng. Med. Biol. Soc. Ann. Int. Conf. 2012; California, USA. August 2012; [DOI] [PubMed] [Google Scholar]