Abstract
Great deal of research is still going on in the field of orthopedic and craniofacial implant development to resolve various issues being faced by the industry today. Despite several disadvantages of the metallic implants, they continue to be used, primarily because of their superior mechanical properties. In order to minimize the harmful effects of the metallic implants and its by-products, several modifications are being made to these materials, for instance nickel-free stainless steel, cobalt-chromium and titanium alloys are being introduced to eliminate the toxic effects of nickel being released from the alloys, introduce metallic implants with lower modulus, reduce the cost of these alloys by replacing rare elements with less expensive elements etc. New alloys like tantalum, niobium, zirconium, and magnesium are receiving attention given their satisfying mechanical and biological properties. Non-oxide ceramics like silicon nitride and silicon carbide are being currently developed as a promising implant material possessing a combination of properties such as good wear and corrosion resistance, increased ductility, good fracture and creep resistance, and relatively high hardness in comparison to alumina. Polymer/magnesium composites are being developed to improve mechanical properties as well as retain polymer’s property of degradation. Recent advances in orthobiologics are proving interesting as well. This paper thus deals with the latest improvements being made to the existing implant materials and includes new materials being introduced in the field of biomaterials.
Keywords: metal alloys, implants, orthopedic, craniofacial, mechanical properties, ceramics, polymers
INTRODUCTION
In 2007, “The Economist” reckoned that humanity is on the verge of biology’s “Big Bang,” meaning that the most important problems being faced by current society and their solutions are biological. Till date, always the numbers of young children have outnumbered the elderly population; however, in about five years from now, the elderly population will outnumber the children younger than five years old.1 One of the major chronic conditions reported by aging population is arthritis; it is a fact that one in every five adults suffer from it.2 One of the other major causes of the increased arthritis rates found amongst patients is obesity; about 54% of adults who suffer from arthritis are obese.3 Treatment for arthritis depends on the type of arthritis, its location, severity, and the medical condition of the patient. Therefore, the treatment procedure can vary from medication, joint injections to surgical operation being the last resort. Osteoporosis and trauma are the other major reasons for joint replacement.
Musculoskeletal disorders are thus the most widespread human health issue, with knee and hip replacements costing around $42.3 billion within US in 2009.4 Surgical implantation of artificial biomaterials have allowed surgeons to ameliorate the lives of patients by reducing the pain and restoring function to the otherwise functionally compromised structure. Hip, knee, and spine are the body parts being replaced most frequently.
Apart from the increased number of the replacement surgery, there has been a simultaneous and parallel increase in the revision surgery of hip and knee implants.5 From the year 2000 through 2011, Food and Drug Administration (FDA) gave a premarket approval for more than 150 new high-risk medical devices, and an additional 600 devices were cleared through the 510(k) process.6 The public directly relates increased flooding of the market with new devices to the demand rate. In 2006, the sales of orthopedic implants and trauma products totaled $12.2 billion, and by 2012, it has increased at a compound annual rate of $18.1 billion.6 The total cost for hip and knee replacement surgery has increased to 10.2% annually since 2000 and this value outpaces the overall U.S. healthcare expenditure growth of 8%.7 Various implantable medical devices being used for these applications are constructed from materials like stainless steel (SS), titanium alloys, and cobalt base alloys. But these materials have demonstrated strong tendencies of failure after long-term use thus recommending the need for revision. There are a number of reasons for revision as depicted in Table I.8 Due to advancement in medical technology, people live longer along with the combined benefit of the increased population of active young adults traumatized by sports injuries, exertive exercise habits, or road accidents appropriating more comfortable living conditions. If this is to be so, the implants have to serve a much longer period without failure or revision surgery. Therefore, it has become necessary to develop implants with superior longevity and biocompatibility.
TABLE I.
Major Reasons for Revisions in Orthopedic Surgeries
| Reasons for Revision | Number According to Kaiser Permanente Total Joint Replacement Registry | % |
|---|---|---|
| Infection | 464 | 40.4 |
| Instability | 233 | 20.3 |
| Pain | 212 | 18.5 |
| Aseptic loosening | 143 | 12.5 |
| Anthrofibrosis | 124 | 10.8 |
| Fracture | 33 | 2.9 |
| Others (include extensor mechanical failure, osteolysis, hematoma, polylinear wear, ingrowth failure) | 78 | 6.9 |
BIOMATERIALS AND THEIR PROPERTIES
In order to attain optimal osseointegration, the material properties of these biomaterials are of paramount significance. Biocompatibility and mechanical endurance are the most crucial properties for both temporary and permanent implants.9 Resistance to corrosion also plays a role especially when different implants are combined such as plate and screw for internal fixation of bone fractures.10 In order for bones to sustain pressurized loads, they must be stiff and able to resist deformation. They must also be flexible in order to absorb energy from possible deformation, shorten and widen when compressed, and to lengthen and narrow through tension without cracking. Implant materials must also be light to facilitate motion. Thus, the implant materials chosen should feature characteristics, which serve these contradictory needs of stiffness yet flexibility and, lightness yet, strength. Apart from these mechanical properties, the material must also be biologically stable because of its interaction with soft and hard tissues, blood, and intra- and extracellular fluids of the human body. Thus, for an implant to serve for an extended period of time, it should possess biomechanical compatibility, biocompatibility, high corrosion and wear resistance, and osseointegration.
These properties and requirements evolved gradually with time leading to the development of innovative devices for improved solutions to clinical issues of implant failure due to infection, corrosion, dislocation etc.11 The first generation of biomaterials was used with the only requirement to achieve physical properties to match those of the replaced tissue with minimal toxic response to the host.12 The second generation of biomaterials was defined by their ability to interact with biological environment to enhance tissue bonding as well as progressively degrade while the new tissue regenerates and heals. Today, we are in the era of third generation of biomaterials, which have the additional ability to stimulate specific cellular responses at the molecular level.13 Despite the tremendous advancement, there is still a need in the industry to develop orthopedic and craniofacial implants, which last longer. There is a need to focus more on the physical, mechanical, and biological properties of the materials being used today to understand their behavior in vivo in order to improve their characteristics (Fig. 1).13 Fig. 1 indicates the mechanical properties of various materials being used for manufacturing orthopedic implants in relation with bone. It is clear that there are very few materials that satisfy all the mechanical properties exhibited by bone. This paper focuses on the various implant materials being used today, especially for orthopedic and craniofacial implants, the problems being faced by these materials and what remedies are being taken to resort those issues.
FIGURE 1.

Comparison of (a) stiffness, (b) strength, (c) fracture toughness for metals, technical ceramics, composites, and fiber reinforced plastic with respect to bone. CF- carbon fibre, GF – glass fibre, PA12: polyamide12, PC – polycarbonate, PE – polyethylene, PEX – poly ether ether ketone, PLGA – poly(l-lactic-co-glycolic acid), PLLA (poly(1-lactic acid), PP – polypropylene, PSU – polysulfone, PTTE – polytetrafluoroethylene, PUR – polyurethane. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]
Metals and their composites
Stainless steel
Majority of the internal fixation devices are currently made from SS. These devices can be roughly classified into few major categories, which include wires, pins, screws, plates, and intramedullary nails or rods (Fig. 2).14 The advantages associated with the use of SS are the fact that they are relatively low priced, easily available, have excellent fabrication properties, are biocompatible, and have great strength.15 SS is an alloy consisting mainly of nickel, chromium, and molybdenum. SUS 316 is the only austenitic steel being used for biomedical applications.
FIGURE 2.

a) Transverse patellar fracture (knee) transfixed with a tension band wire b) Subcapital proximal femoral fracture (hip) transfixed with screws c) Distal fibular shaft fracture (ankle) transfixed with plate and screws d) Intertrochanteric fracture (hip) transfixed with intramedullary femoral rod with an interlocking screw.
Studies have shown that 90% of implant failures of 316L SS are due to pitting and crevice corrosion arising by the presence of chloride ions and reduced sulfur compounds in bodily fluids leading to the release of elements like nickel and chromium, which have a toxic effect.16–22 Corrosive properties of these alloys have been worked on, and they have increased the levels of chromium, molybdenum, and nitrogen in order to increase its corrosion resistance.23,24 SS are also being coated with transition metal nitrides like TiN, VN, TiAlN etc. which act as protective coating against wear and corrosion thus increasing life expectancy of the implants.25 The major concern in using SS for biomedical applications is the presence of large amounts of nickel content. Nickel has been reported as a toxic substance to human body. Therefore, great deal of research is being focused on developing nickel-free SS, one of the approaches being the addition of large amounts of nitrogen, which is an austenite stabilizing element instead of nickel (which is also an austenite stabilizing element). Studies have compared the cytocompatibility of thus produced nickel-free SS (one containing Fe–Cr–Mo–N and the other containing Fe–Cr–Mo) with 316L SS, and results indicated that cell viability of nickel-free SS Fe–Cr–Mo–N is the best.26,27
The primary cause of response of immune system of the host against the biomaterial is the adhesion of proteins to its surface through interactions such as electrostatic or hydrophobic forces. Therefore, surface modification to generate protein-resistant surface is being accepted as a promising technology. Surfaces are being treated with chemical groups like epoxides and then grafted with various polymers such as polyethylene glycol, poly (dimethylsiloxane) etc., which increase surface hydrophilicity, thereby decreasing protein adsorption.28,29
Infection caused at the site of surgery can also lead to implant failure. Decreasing antibacterial activity exhibited by different materials is given in the following order gold> titanium>cobalt>vanadium>aluminium>chromium>iron showing little resistance of SS to microbial attack.30 The most extensive remedy for this problem is being sort by implanting SS with silver/copper ions, which show strong antimicrobial properties to both Staphylococcus aureus and Aspergillus niger.31,32 Titanium based biomaterials have shown better behavior in all the above-mentioned aspects.33 But still SS continues to be used for manufacturing implants due to their cost aspect and will be used in future as well, and thus, its fatigue corrosion behavior should be studied well and resolved.34,35 But, when a situation arises where a larger implant is required, such as joint replacements due to high elastic modulus possessed by SS as shown in Table II, the material is stiffer, and thus demands the use of other appropriate implant materials.36
TABLE II.
Mechanical Properties of Metallic Materials used for Orthopedic Implants. The Range of Values is Dependent on the Composition of the Material and Testing Conditions
| Material | Elastic Modulus (GPa) | Fracture Toughness (MPa) | Hardness (MPa) | Compressive Strength (MPa) | Density (g cm−3) |
|---|---|---|---|---|---|
| Human cortical bone | 3–20 | 3–6 | 300–480 | 90–120 | 1.8–2.1 |
| Stainless steel | 190 | 50–200 | 130–180 | 170–310 | 7.6 |
| Co–Cr alloys | 200–300 | N/A | 300–400 | 450–1000 | 8.9 |
| Ti and Ti alloys | 110–116 | 55–115 | 310 | 758–1117 | 4.5 |
| Tantalum and alloys | 3 | 96–124 | 240–393 | 42–78 | – |
| Zirconium | 96–100 | – | 210–235 | 276–345 | 6.51–6.64 |
| Magnesium | 41–45 | 15–40 | – | 65–345 | 1.74–1.84 |
Cobalt based alloys
Cobalt–chrome alloy is ubiquitously used in knee and hip replacement implants, particularly in the ball and socket joint where the motion occurs. In hip anthroplasty, it is generally the femoral head component that is hard and Co-Cr is used as a very smooth and scratch resistant surface. Other applications of this alloy include implants like tibial trays, acetabular cups, dental parts, pacemaker lead casings, as well as cardiovascular stents. All of these alloys are primarily composed of cobalt and chromium, but they may also include other metals like molybdenum to refine grain size, enhance solid solution strength, as well as increase corrosion resistance while nickel increases plasticity, fatigue crack growth resistance, and malleability of the alloy.37,38 These alloys are known to possess high wear resistance and good mechanical properties under static loadings.39 In the beginning stellite series of Co–Cr, alloys were developed composed of base material cobalt, 27–32% Cr, 4–6% W, 0.9–1.4% C and later vitallium composed of 62 % cobalt, 30% chromium, 6% molybdenum, 2% other elements (silicon, manganese, carbon) were patented. Since then, there has not been a significant change in the composition of the alloy, except for the carbon content. There has been a strict regulation on the percent of carbon and the homogeneous distribution of hard carbide grains, which increases the abrasion resistance of the alloy. Cast and wrought alloys are primarily used with identical composition but different microstructure. The cast iron exhibits a heterogeneous, large-grained, and cored microstructure, while the later one has a face-centered cubic structure with an austenitic microstructure and a finely distributed small block of carbides.40 These properties make it one of the toughest materials available (Table II).41
Despite the corrosion resistance provided by both cobalt and chromium, biocorrosion is one of the major issues this alloy possesses.42 Metal on metal implants are known to create 1012–1014 nanoparticles of cobalt–chromium (Co–Cr) alloy per year in the size range of 20–60 nm. The presence of these nanoparticles near the implant site has an antibacterial characteristic preventing formation of biofilms. Research has shown that metal on metal wear debris and metal on polymer wear debris accelerate the growth of common organisms that infect prosthetic hip joints, but the clinical significance of this result is still uncertain. In isolation, the heavy metals contained within the wear debris (cobalt and nickel) have shown to retard the growth of bacteria, particularly nickel. Therefore, nanoparticles in the wear debris do not possess the toxic effects of its constituent metals on bacteria. This reinforces the safety of wear debris but also shows the potential for antibacterial effects to be harnessed.43 Studies have shown that increased levels of cobalt and chromium in the serum can also be related to implant loosening.44,45 Inflammatory responses are observed near the implant showing tissue necrosis, ulceration, and allergic reactions.46 Numerous animal studies have also shown carcinogenicity.47–48 Therefore, nowadays, Co–Cr–Mo alloy is being used in collaboration with ultrahigh molecular weighted polyethylene (UHMWPE) to reduce corrosion and wear.
Nickel is also often employed as an alloying element as it increases the stacking fault energy of the alloy and thus stabilizes the γ phase rather than the ε phase, thereby achieving higher ductility.49 However, it is well known that nickel is a very toxic element. Therefore, Ni-free cobalt alloys are being developed for biomedical applications. In order to maintain good mechanical properties as well as to remove Ni for cobalt alloys, N is being used as a substitute under hot compression.50,51
Lack of integration of the Co–Cr alloy into the bone also leads to implant failure. Efforts are being made to modify its surface including covalent protein grafting and self-assembled monolayers of certain chemicals, to mention few. Studies have shown that osteoblasts attach, proliferate and differentiate better on these surfaces thus providing a potential solution for tissue integration.52,53
Titanium and its alloys
The application of titanium and its alloys in the medical area is truly astonishing as through its utilization in craniofacial implants, dental implants, joint replacement parts for hip, knee, shoulder, spine, elbow, and wrist, and bone fixation materials like nails, screws, nuts, and plates have been possible. This wide variety of applications can be attributed to its excellent characteristics as indicated in Table II such as high strength, low density, high specific strength (strength per density), good resistance to corrosion due to formation of an adhesive titanium oxide layer at the surface, inertness to body environment, enhanced biocompatibility, moderate elastic modulus, and high rate of integrity with the bone.54–56 Titanium is commonly used in either grade 4 (G4Ti) or 5 (Ti-6Al-4V) forms because of their excellent chemical and mechanical properties. There are four types of titanium alloys-α, near α, α+β, and β alloys based on their chemical composition, the content and nature of alloying elements, and the resulting microstructure.57 Of the different types of titanium alloys found, β phase titanium alloy tends to satisfy most of the requirements for an orthopedic implant application in contrast to an α phase alloy.58
The third generation titanium alloys are focused on developing porous structures to allow penetration of the vascular system for total integration of metal and bone.59,60 Titanium fiber meshes with a porosity of 86% with an average pore size of 250 μm have been used in ex vivo studies in rats.61 Titanium foams were used in vitro with human osteoblasts showing that the cells colonize and differentiate into mature bone cells and also possess osteoinductive properties, the only metal to possess this property.62,63 Titanium is one of the few materials that do not contain calcium phosphate and has been shown to calcify when exposed to simulate body fluids and therefore is expected to calcify in vivo also.64
Despite all these advantages, long-term performance of these alloys has raised some concerns. This is due to the release of metals like aluminum and vanadium from Ti64 alloys that have been shown to be associated with health problems like Alzheimer’s disease (although recent finding do not support this), neuropathy, and osteomalacia.65,66 Vanadium alone is toxic in both elemental and oxide states at the surface of the implant.67 Therefore, V-free Ti alloys such as Ti-6Al-7Nb and Ti-5Al-2.5Fe were developed. Then, V-and Al-free Ti alloys such as Ti-Zr and Ti-Sn alloys were developed.68–70
Titanium alloy also undergoes severe wear when it comes in contact with other metal surfaces due to a high coefficient of friction. This leads to the inflammation of the surrounding tissues and as a consequence loosening of implants.71 Therefore, in order to increase wear resistance, thermomechanical processing of titanium microstructure is being studied as temperature is found to play a crucial role in the evolution of microstructures.72,73
Stress shielding is another major issue with metallic implants given the difference in the Young’s modulus of the implants and bone. Young’s modulus or the tensile modulus or elastic modulus is a measure of inherent stiffness of the material. It defines as a ratio of uniaxial stress to strain. This leads to bone absorption followed by loosening of the implant or refracture after removal of the implant. Therefore, it becomes necessary to lower the modulus of metallic implants to match the modulus of the bone. Studies showed that the additions of non-toxic elements like Nb, Zr, Mo, and Ta to titanium decreases the modulus of elasticity without compromising its strength.74–76 Very recently, elements such as Fe, Cr, Mn, Sn, and Al are being proposed instead of rare metals like Nb, Zr, Mo, and Ta in order to lower the cost of the implant.77–80 The lowest Young’s modulus thus obtained for β-type Ti alloy is 35 GPa and is thus similar to the top range of bone which is 30 GPa.81
Spine fixation implants need to be bent to reproduce the curvature of patient’s spine. In such cases, it is necessary to have a Ti alloy that has Young’s modulus low enough to avoid stress shielding but high enough to suppress spring back that will cause trouble in patients. Therefore, a novel type of Ti alloy with low Young’s modulus before deformation but increasing during deformation (self-tuning Young’ modulus) has been developed and includes Ti–Cr, Ti–Mo, Ti–Zr–Cr, Ti–Zr–Mo, and Ti–Zr–Mo–Cr.82,83
Among several types of shape memory alloys, the NiTi alloy is considered the best biomaterial for orthopedic implants. The austenite form is quite strong and hard like titanium.84 NiTi has an elastic modulus much similar to that of bone, it also has unique fatigue resistance and ductile properties while wear resistance is high and comparable to the Co–Cr–Mo alloy.7 But, due to the toxicity of Ni, various studies are being carried out to develop Ni-free Ti-shape memory alloy. Currently, there are four categories of Ni-free Ti-shape memory alloys: Ti–Cr alloy system, Ti–Nb alloy system, Ti–Ta alloy system, and Ti–Mo alloy system.85–88 Addition of small amounts of oxygen and nitrogen have been found to enhance super-elastic behavior of Ni-free Ti alloys as they lower the starting temperature of the martensite transformation. Temperature has been found to play an important role in functionality and strength of shape memory alloys.
Another approach being used for improving biocompatibility of Ti is provided by the surface modification approach based on controlling surface topography, ceramic coating, physicochemical, or inorganic modification and very recently biochemical modification of titanium surfaces.89–91 In this recent approach, proteins, enzymes, glycosaminoglycans, chondroitin sulfate, collagen, and hyaluronic acid have been found to influence bone remodeling by improving osteointegration.92,93 In the above-mentioned materials, the adjuvant remains only superficially associated with Ti. However, it is difficult to make these coatings remain adherent to Ti due to differences in mechanical moduli with eventual failure either at bone–implant interface or implant–coating interface.94,95 There have also been attempts in which materials are impregnated into the core material, like calcium phosphate crystals within titanium and the results showed that there was an improved cell differentiation, bone bonding to implant, and osseoactivity.96 Modification of topography of the Ti surface has also been shown to be a promising method.97 Roughness has been established as an important character that determines and enhances osteoblast proliferation, adhesion, alkaline phosphatase activity, and mineralization.98 However, principle contents of bone either organic (collagen) or inorganic [hydroxyapatite (HA)] are nanophase materials.99,100 Thus, nanostructured materials are being studied as they provide a completely different sort of interaction between implant and cell, increasing surface area for cell attachment, cell proliferation, and mineralization.101,102
Knee joints operate similarly to a dynamically loaded bearing with about 108 cycles of loading in a 70-year lifespan. The average coefficient of friction of a load bearing synovial joints like hips and knees is about 0.02 with an average wear factor of about 106 mm3/N. (wear properties of knee joint are always greater than that of the hip, given its polyaxial movement with flexion extension rotation and translation).7 On the contrary, the coefficient of friction lies in between 0.16 and 0.05 depending on implant material and lubricant used. Commonly, prosthetic hip joints consist of a femoral head articulating against an UHMWPE. Of the metals used for femoral head construction, titanium was found to possess maximum wear followed by 316L SS, with the least observed in Co–Cr–Mo alloys.
Tantalum and its alloys
With the increasing number of young, active population undergoing arthroplasty, the need for a material, which has adequate biological and mechanical properties, is apparent. Porous tantalum has been identified as a potential alternative given its ability to integrate with the tissue allowing bone–tissue ingrowth, biomechanical properties, inert nature in vivo, and excellent chemical stability.103 As indicated in Table II, the microhardness value for porous tantalum ranged from 240 to 393, the compressive strength was 60 ± 18 MPa, the tensile strength was 63 ± 6 MPa, the bending strength was 110 ± 14 MPa, and modulus of elasticity on the order of 3 GPa (similar to trabecular bone).104 The ductility of this material is less in comparison with the metal implants but is high in comparison with other naturally occurring substances, ceramics, composites, and bone itself. The compressive fatigue endurance was 23 MPa at 5 × 106 cycles.105 Tantalum implant possesses porosity of 400–600 μm with a volume porosity of 75–85% in comparison with Co–Cr (30–35%) and fiber metal (40–50%).106 Distal femoral and proximal femoral components and patella have been designed using porous tantalum. Studies have revealed that there have not been any early failures in clinical performance of porous tantalum even in some harsh mechanical circumstances in cases of hip and knee revisions. But it still remains ascertained whether the theoretically advantageous mechanical compatibility of porous tantalum with bone will cause less-stress shielding and disuse atrophy of the surrounding bone.107 Studies have indicated excellent stability during early bone development with bone ingrowth and minimal stress shielding problems.108
Niobium and zirconium alloys
Niobium alloys are very recently being proposed for biomedical applications. Nb-2Zr that has been recently introduced has been reported to exhibit excellent corrosion resistance, fatigue strength, and crack propagation in simulated body fluids.109
Zirconium belongs to the same group as titanium and exhibits similar properties as indicated in Table II.110 But, zirconium is found to possess lower magnetic susceptibility (1.3 × 10−6 cm3g−1) than paramagnetic Ti (3.2 × 10−6 cm3 g−1). This property has been found to be important, because an insertion of metallic implants into patients undergoing magnetic resonance imaging (MRI) may increase probability of defects and distortions of this analysis. Therefore, Zr-based alloys have been developed like Zr–Nb and Zr–Mo.111
Ceramification of zirconium niobium alloy (Zr-2.5Nb) is being considered to improve the alloys wear and scratch resistance. Roughened-oxidized zirconium (OxZr) surface resulted in 61% less wear than roughened CoCr surface.112 OxZr is also found to possess increased hardness and wettability, which supports formation of a lubrication film. It also supports higher adhesion in comparison with TiN coating.113
Mg alloys
Magnesium alloys are being sorted as a promising candidate for fracture fixation because magnesium possesses mechanical properties very similar to those of the bone (Table II). Its tensile strength in wet conditions is 230 MPa and Young’s modulus is 45 GPa.114 Some of the commercially available magnesium alloys include WE43 [Mg-Y-RE (rare earth)-Zr system], AZ91 (Mg–Al–Zn alloy system), AZ 31 (Mg–Al–Zn–Mn alloy system) and LAE442 (Mg–Li–Al–RE–Mn alloy system).115–117 Magnesium is the fourth most abundant element in the human body; therefore, degraded products of this alloy can either be stored in fracture callus or new bone or eliminated into blood to be excreted by urine, without causing hyper-magnesium.118 However, its corrosion resistance is far less than Ti and Co–Cr alloys, and therefore, magnesium is being considered as biodegradable material. Because of rapid degradation, magnesium alloys lead to formation of gaps at the interface of bone and implant; therefore, significant effort is applied to develop methods preventing fast degradation in vivo. Addition of alloying elements was one of the first approaches, but due to restrictions on usage of certain elements like Ni, other alternatives were sought. Surface modification of magnesium by chemical, physical, and a combination of both were used for treating it. Chemical modification included procedures like acid etching, fluoride treatment, alkaline treatment, anodizing coating, ion implantation etc. Physical modifications included apatite coating, implantation of metal oxides, and laser surface processing. Chemical and physical modification processes were also combined to improve corrosion resistance of magnesium alloys.119
Ceramics and their composites
Inert Ceramics
Alumina and zirconia are the two most widely used inert ceramics in hip and knee arthroplasty, maxillofacial reconstructions, ossicular bone substitutes, keratoprostheses, blade screws, bone screws, and post dental implants. The wide usage of these biomaterials can be attributed to their inertness (no formation of direct bone-material interface) and desirable mechanical properties, as they possess high compressive strength, elastic modulus, and hardness. They also possess excellent corrosion resistance, biocompatibility, low friction, high wear resistance, and stability in a physiological environment. The surface groups promote interaction with bodily fluids.120 Alumina has been used for nearly 20 years now owing to its favorable characteristics. Polycrystalline α-Al2O3 (alumina) when mixed with small amounts of MgO has exceptionally low coefficients of friction making it suitable as a bearing surface. Alumina also has excellent biocompatibility with tissue aging not affecting the material structure or strength.121 The current fracture rates of ceramics is negligible in comparison with overall failure rates of implants, though it must be kept in mind that the current use of ceramics has been restricted to limited number of which demand less mechanical stress. For femoral heads, fine-grained alumina fitted with Ti–Al–V stem is theoretically established as the best material combination for hip replacement surgeries.122
Despite these advantages, alumina femoral heads and acetabular liners are found to be associated with catastrophic failures.123,124 The surface of the bearing must be perfectly spherical and congruent for hip prosthesis, thus emphasizing on the finalizing steps of its manufacturing. Due to its bioinert characteristic, there is always a soft tissue interlayer between the bone and the implant leading to stress-shielding problems causing troubles amongst the elderly population with osteoporosis and rheumatoid arthritis leading to loosening of the alumina acetabular cap. Thus, only under high compression, the fibrous tissue growth at the interface can be restricted with successful bone ingrowth. Therefore nowadays, ceramics are being fixed inside a taper lining of titanium acetabular shell rather than directly interfacing with the bone.
In response to the concerns regarding Al2O3 femoral fractures, yttria-stabilized zirconia (Y-TZP) was introduced. Due to the excellent mechanical properties of zirconia, it is called “cement steel”.125 Zirconia has the characteristic property of changing its crystalline reticulation when force is applied to its surface. This force on the crystal surface causes a volumetric change thus sealing the crack caused by the stress.126 Y-TZP has about two to three times larger fracture toughness and flexural strength than Al2O3, making it one of the strongest ceramics for medical use. Zirconia has been shown to be non-cytotoxic, with no mutagenic effect.127–129 Zirconia has been shown to support growth of blood vessels, but in comparison with titanium based implants, it shows lower inflammatory infiltrate, micro vessel density, and vascular endothelial growth factor characteristics.130 Zirconia was also found to influence the gene expression of the cells, thus acting as a self-regulatory material.131 Zirconia heads are used widely against UHMWPE bearings in Total Hip Arthroplasty and Total Knee Arthroplasty. But, despite these characteristics, recent studies have shown low temperature degradation for Y-TZP, including the temperature used for steam sterilization (140°C) and body temperature (37°C).132 In Y-TZP, the presence of numerous vacancies in its lattice (trivalent) makes the diffusion of water higher than in normal zirconia, which ultimately leads to micro cracking.133 Studies have also shown aging on the surface of the Y-TZP implants and predicting failures from low temperature degradation in the large number of zirconia implants used in the last decade (>500,000).134 To overcome these limitations, two kinds of materials are considered: (1) alumina-zirconia composites – doping zirconia with alumina strongly reduces the number of vacancies in the lattice thus preventing the uptake of water molecules and improving crack resistance (there are already some implants in the market with SrO and Cr2O3 addition to zirconia-alumina grains); and (2) ceria and magnesia doped zirconia (Ce-TZP, Mg-PSZ) – these dopants were found to increase the toughness of the material.
Bioactive ceramics
This category of ceramics include calcium phosphate, glass, glass–ceramic, and composites, as these materials have the capability to interact with the biological environment to improve host response as well as bind the injured tissue. These materials also undergo progressive degradation while the new tissue regenerates.135
Natural bone can be viewed as a composite made from biopolymer (mainly collagen) and bioceramic [calcium phosphate (CaP)]. A total of 69% of the mineral content of the natural bone is composed of CaP in the form of carbonated apatite. The organic matrix of the bone consist of proteins (22%), type I collagen (90% of organic matrix) with some non-collagenous proteins (proteoglycans), lipids, and osteogenic factors.99 Though CaP itself is not an osteoinductive material, it can be combined with materials like growth factors, bioactive proteins, or osteogenic drugs, and osteoinductivity can be induced.136–139 Depending on temperature, impurities, and the presence of water, they exhibit different physical and chemical properties.140–142 CaP has always showed excellent biocompatibility, bioactivity, biodegradability, and osteoinductive ability.143–146 Of the different types of calcium phosphate, HA and β-tricalcium phosphate (β-TCP) are the most commonly used biomaterials because of their osteogenic properties, degradation properties, and their ability to form strong bond with bones. HA is a naturally occurring mineral form of calcium apatite. β-TCP is more soluble than HA and thus is called bioresorbable ceramic.147,148 Some studies have shown that β-TCP supports cells better when compared to HA, while other studies have shown HA to be a better supporter.149–151 The biological property of CaP can be attributed to the liberation of Ca2+, PO43−, HPO42− to the surrounding environment which causes super saturation of the nearby fluid causing precipitation of carbonated apatite thus incorporating ions like Mg2+, Na+, CO32− as well as proteins and other inorganic compounds.142,152 They also provide nucleation sites for deposition of apatite layer containing organics. Porous ceramic scaffolds have shown to better support bone formation in comparison with dense sintered scaffolds. This is because the pores help in the invasion of blood vessels into the scaffold which brings nutrients and oxygen supply very essential for metabolism of cells.153–155 The mechanical properties of this material depend on a number of factors including crystallinity, porosity, composition, and grain size. When comparing the mechanical properties of bone (Table III), we find that HA has higher compressive strength than bone, but bone has higher fracture toughness than HA. Despite the excellent biological and mechanical properties, HA cannot be used for all bone–tissue growth, especially at load bearing sites due to its low mechanical strength. They have also shown slow biodegradation.
TABLE III.
Mechanical Properties of Ceramics. The Range of Values is Dependent on the Composition of the Material and Testing Conditions
| Ceramics | Compressive Strength (MPa) | Tensile Strength (MPa) | Elastic Modulus (GPa) | Fracture Toughness (MPa) |
|---|---|---|---|---|
| Hydroxylapatite | >400 | ~40 | ~ 100 | ~ 1 |
| 45S5 | ~500 | 42 | 35 | 0.5–1 |
| Glass–ceramic | 1080 | 215 | 118 | 2.0 |
| Porous bioactive glass 70S30C (82%) | 2.25 | – | – | – |
| Porous bioglass derived glass ceramic (>90%) | 0.2–0.4 | – | – | – |
| Porous HA (82–86%) | 0.21–0.41 | – | 0.002–0.83 | – |
| Non-oxide ceramic | 4000–4500 | 450–1200 | 175–250 | 2.8–12 |
Glass-based scaffolds can be classified into two main groups: (1) glass/glass–ceramic porous scaffolds (2) glass–polymer porous composites. Silicon found in glass has been found to greatly upregulate the gene expression in osteoblasts that regulate osteogenesis and production of growth factors.156 Silicon has also been found to play a role in bone mineralization.157 Recent work has shown that bioglass can also induce neovascularization promoting the formation of blood vessels in vitro.158 Studies have shown that silicon substituted HA granules have higher bone ingrowth when compared to pure HA.159 Phase transformations occur during thermal treatment leading to a high versatility of the final component which cannot be obtained in CaP ceramics. One of the main drawbacks of bioactive glass is its low fracture toughness and mechanical strength, especially in porous forms (Table III). Thus, similar to HA, these materials also have limitations in usage as substitutes in load bearing positions.
Non-oxide ceramics
This category of materials includes silicon nitride (Si3N4) and silicon carbide (SiC). They can be considered as glass–ceramic composites as they contain partially crystalline grain and amorphous boundary phases. They possess a combination of properties such as good wear and corrosion resistance, increased ductility, good fracture and creep resistance, and relatively high hardness in comparison to Al2O3.160
Despite these advantages, studies have found that silicon oxide layer, which forms by super oxidation of Si3N4 and SiC surface, may chip off over time resulting in significant increase in wear, as indicated in Table III.161,162 Also despite high fracture toughnesss and strength, these materials have shown catastrophic in vivo failure. Therefore, efforts are being made to overcome these issues by working on their microstructure development and design and fabrication techniques. In the microstructure development, β-Si3N4 can be formed with a perfect circular cross-section and aspect ratio which are major factors contributing to the toughness and strength of a material. Designing of a concentric ring-structured implant consisting of alternate layers of Si3N4 and partially porous interface, which eliminates the direction of easy crack propagation and offers high damage tolerance.163
Apart from the high mechanical strength, these materials have also exhibited excellent cytocompatibility in vitro.164 Studies have also found that Si3N4 showed better resistance to bacterial film formation in comparison with Ti and polyether-ether ketone (PEEK). This difference can be attributed to the surface chemistry and surface nanostructure properties of the material.165
Polymers and their composites
For orthopedic and craniofacial applications, polymers currently used are acrylics, polyamides, silicones, polyurethanes, PEEK, UHMWPE, and polypropylene (PP).166 The main advantages of these materials include the possibility to tailor their composition and structure according to the specific needs. In addition, bioresorbable polymers are being used in both orthopedic and craniofacial bone regeneration. Polymethyl methacylate (PMMA) commonly referred to as bone cement, continues to play a vital role as synthetic biomaterial used in orthopedic surgery due to a comparable elastic modulus to that of the bone. It is the only implant that is manufactured in the operating room. It functions as a supportive material that forms a mechanical bond between the cement and bone ad cement and prosthesis. Another important polymer to consider is poly-hydroxyethylmethacrylate (poly-HEMA), as it is the only polymer possessing osteoinductive capability in vivo till date. However, composites of polymers and HA possess osteoinductive property.101,167 Even in the absence of osteoinductive factors like bone morphogenic proteins (BMPs), these materials were successful in bone formation in pigs.168 This phenomenon was explained by the fact that bone formation can be induced by tendons and arteries only if they were first calcified and the CaP formed plays a role in osteoinduction.169 Thus polymers with growth factors incorporated are being evaluated as promising material for tissue regeneration.170 Polymers can also be fused with bioactive peptides like arginine–glycine–aspartic acid (RGD), which help in cell migration onto the surface. Despite these efforts, polymers still face the problem of interfacial adhesion at tissue and implant junction.171
The main advantage of using polymers is that they can be injected into the defect site thus reducing the cost and pain that follows surgery. One approach to do this is to design photopolymerizable materials, such as Poly(sebacic acid) (PSA), poly [1,3-bis(p-carboxyphenoxy) propane], and poly[1,6-bis(p-carboxyphenoxy] hexane, which can be injected as liquid and once in the bone defect site; they can be solidified by exposure to light. Since these materials can absorb water, they undergo degradation at their surface. In order to avoid these troubles, scientists are exploring polymers which can cross-link in vivo such as poly (propylene fumarate) and polyanhydrides. These polymers have showed the ability to control polymerization and form three-dimensional structures, along with the ability of controlled degradation rate.172–176
New processing techniques and composite materials are being used such as polyvinyl alcohol (PVA)/polyvinylpyrrolidone (PVP) hydrogels and PVA melted on titanium mesh to resolve the issues of mechanical instability.177 Composite consisting of polyethylene (PE), younger cross-linked variant of polyethylene (XPE), and CoCr grafted with 2-methacryloxyloxethyl phosphorylcholine polymer are being designed which mimic the neutral phospholipids of biomembranes having coefficient of friction in the range of cartilage–cartilage articulation thus minimizing the wear debris, thereby reducing the chances of implant failure due to implant loosening.178
Polymers also have been found to lack mechanical stability as shown in Table IV.179 To overcome this, fiber-reinforced polymer composites are being used widely in orthopedics.180 These materials have the required strength (Fig. 1) to match the properties of bone which include low elastic modulus as well as high strength, corrosion, and fatigue resistance. Also the properties of the composites can be adapted according to the requirements by varying the arrangement or volume fraction of the fibers.181
TABLE IV.
Mechanical Properties of some Polymers
| Polymer | Young’s Modulus (GPa) | Compressive or Tensile Strength (MPa) |
|---|---|---|
| ultrahigh molecular weighted polyethylene (UHMWPE) | 0.495 | 231.1 |
| Poly(glycolic acid) – PGA | 7–14 (FIBRE) | 340–920 (FIBRE) |
| poly(L-lactic acid) – PLLA | 10–16 (FIBRE) | 870–2300 (FIBRE) |
| poly(L-lactic-co-glycolic acid) – PLGA | 1.4–2.8 | 41.4–55.2 |
| Poly(anhydrides) | 0.14–1.4 | 25–27 |
A number of studies have been conducted on ceramic–polymer composites and polymer–polymer composites. But, despite tremendous efforts to improve their mechanical properties, disastrous failures have been reported. Therefore, efforts are being made to design polymer/magnesium composites.182 The polymer matrix will benefit from magnesium which possesses high mechanical strength and fracture toughness, while the magnesium can benefit from polymer which will prevent its degradation and will give the strength required by the composite to reach the values of cortical bone.183,184 Magnesium and polymer being degradable can degrade with time as the tissue heals.
Orthobiologics
Orthobiologics include resorbable calcium salt bone void fillers, few of which contain demineralized bone matrix (DBM), which is processed from human bone tissue and contains a number of bone growth stimulators, BMPs, stem cells, gene therapy etc. They are among the cutting edge technologies for musculoskeletal treatments. They possess tissue regeneration characteristics which include osteoconduction, osteoinduction, osteogenesis, and osteocompatibility. Apparent differences in the mechanisms of osteoinduction have been shown between BMPs and inorganic biomaterials: (1) bone induced by biomaterials is always intramembranous, that is, direct bone formation without cartilage intermediate at heterotropic sites, while bone formed by BMP induction is mostly endochondrial; (2) in small animals, bone formation is rarely induced by biomaterials in contrast to BMPs; (3) in larger animals, bone induction by biomaterials is relatively slow and requires few months to be accomplished, whereas for BMPs, it takes 2–3 weeks; (4) in biomaterials, bone formation is seen mostly in the porous area of the scaffold, whereas BMPs-induced bone formation is seen in the periphery as well as in the soft tissue away from the carrier.137,185–195 Due to these advantages, several studies have been focused on growth factor incorporating biomaterials. CaP matrices and scaffolds have been used to deliver growth factors locally, in the defect site, which is more effective as compared to the methods of systemic delivery used for treatment of majority musculoskeletal disorders. CaP cements, CaP coated metallic implants, and several other custom-designed implants with growth factors incorporated are being used in clinics.196–199
An optimum concentration level of DBM and carrier is required in order to resist the migration from graft site. Cortical and cancellous bone chips are used to provide the three-dimensional scaffold for osteoconduction and mechanical strength. The orthobiologics are stored in dry form in sterilized condition, and when they need to be used, they are mixed with blood or autogenous bone grafts and delivered into the bone defect using a syringe or can be molded into a shape and fixed at that site. These substances are made water insoluble to prevent flowing off from the graft site.
Type I collagen along with bone mineral has been proven to be osteoconductive. The limitations of using these new technologies are the delivery of these materials to the specific site of implantation and to study the specific response from these products. Fibrous textile materials have been designed for use as composites with the orthobiologics to bridge this gap. The fabric structures can be formed from polyglycolides, poly-L-lactic acid, and similar materials. And they are given a structural design by either of the technologies: weaving, braiding, and knitting to obtain a unique design feature that allows resorption of the polymer within the structure to coincide with the repair rate of the biological material.
RECENT IMPROVEMENTS IN COMMERCIALLY AVAILABLE IMPLANTS
Knee replacements
Knee joint is composed of tibia, femur, and patella. In order to provide pain-free motion, the ends of bones are covered with cartilage, which is a lubricating tissue that prevents friction between the bones. The most common form of cartilage wear-osteoarthritis results in irritation, swelling, stiffness, and discomfort urging them to resort to cartilage replacements.
Several other alternatives are thought before considering knee replacement. Minimally invasive surgery, arthroscopy, replacing the worn out cartilage with either tissue engineered cartilage or cartilage transplant, and microfracture arthroplasty are some of the methods. But, if the surgeon recommends knee replacement, then the implants are typically made of metal alloy on the end of thighbone and PE on top of the tibia and underneath the kneecap. The word replacement does not mean removing the entire joint, but it includes removal of the damaged bone and cartilage at the end of the joints.
When designing a knee implant, the most advantageous method is to consider the individual’s joint anatomy. Indeed, many manufacturing companies are nowadays offering customized personal implants by utilizing MRI or CT scan technology to create three-dimensional joint reconstruction. This allows for better implant positioning and alignment with other joint components.
Partial knee replacement is considered when only one compartment of the three compartments in the knee is affected by arthritis and the other two are healthy. The advantages of this procedure are smaller incision, lesser bone removal, better natural motion, and faster recovery of the patient. The disadvantage of this procedure is that partial knee replacement does not last as long as total knee replacement surgery. The New Zealand National Joint Registry reviewed the rate of revision of unicompartmental knee replacements (UKR) between 1999 and 2008 and found that the rate of revision for the UKR to a total knee replacement (TKR) was four times higher than that for primary TKR.200
Complete knee replacement needs to take many factors into consideration before designing the implant for an individual like their gender, age, weight, medical condition, activity levels etc. But, generally, for the manufacturing companies, the important consideration is how the kneecap moves against the femoral component. The kneecap moves in the groove provided in the implant during the flexion and extension movement. The angle of movement has to be crafted individually because it differs among individuals.
Hip replacements
The hip is a ball and socket joint with the femoral head (ball) sitting in the acetabulum (socket) with a cartilage in-between (Fig. 3). Constant motion and pressure on the joints can result in continuous wear and tear of the cartilage.
FIGURE 3.

Total Hip Replacement implant with metal on polyethylene.
Customized implants are also available to ensure better implant-bone match. Also in implants where highly cross-linked polyethylene is being used, they are preparing vitamin E induced polyethylene, as it is found that polyethylene gets oxidized in presence of polyethylene and with time becomes weak. Vitamin E acts like an antioxidant and protects the polyethylene from oxidation thus increasing its life.
Future development
The promise of the biomedical implants in eliminating some of the intricate issues of middle-and old-age population has led to outpouring of demands for new procedures. The major reason for the torrent of demand is prolonged average life expectancy of the population. However, the current generation of implants still faces certain issues during the long-term performance. By the extensive studies carried out till date, we can provide a theoretical answer to the properties required by these biomaterials, but still we are unable to design materials that can provide us desired results in vivo. The main reason is that properties of the end product greatly depend on the process parameters, which greatly influence various characters like macroporosity, grain size, surface roughness etc, which in turn determine the mechanical and biological properties of that material. These parameters are often missing in the publications and thus should be standardized and studied carefully. There is also an increasing need for understanding the basic interaction of the biomaterials with the implant surface, the host, and the biological environment at atomic levels as well as to know all types of micromotions executed by the implant inside the host, in order to develop implants which can last longer in the human body. Intense studies should be carried out to understand aspects like wear of the implant under pressurized loads, biologic response to wear and corrosion debris, the effect of biologic medium surrounding the implant etc. In the end, one has to accept the fact that humans have been diligently working and in many ways have succeeded in relieving many people from the sufferings and have increased their longevity, but the mission is still unaccomplished.
Acknowledgments
This work is supported by National Science Foundation (NSF) grant #0652024 and National Institute of Health (NIH) grant #DE019508.
Footnotes
CONFLICT OF INTEREST STATEMENT
None of the authors have a financial interest in any of the products or implants cited in this manuscript.
References
- 1.Hootman JM, Helmick CG, Brady TJ. A Public Health Approach to Addressing Arthritis in Older Adults: The Most Common Cause of Disability. Am J Pub Health. 2012;102:426–433. doi: 10.2105/AJPH.2011.300423. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Helmick C, Felson D, Lawrence R, Gabriel S. Estimates of the prevalence of arthritis and other rhematic conditions in the United States. Arthritis Rheum. 2008;58:15–25. doi: 10.1002/art.23177. [DOI] [PubMed] [Google Scholar]
- 3.Lee J, Song J, Hootman JM, Semanik PA, Chang RW, Sharma L, Van Horn L, Bathon JM, Eaton CB, Hochberg MC, Jackson R, Kwoh CK, Mysiw WJ, Nevitt M, Dunlop DD. Obesity and other modifiable factors for physical inactivity measured by accelerometer in adults with knee osteoarthritis. Arthritis Care Res (Hoboken) 2013;65:53–61. doi: 10.1002/acr.21754. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Murphy L, Helmick CG. The impact of osteoarthritis in the United States: A population-health perspective: A population-based review of the fourth most common cause of hospitalization in U.S. adults. Orthop Nurs. 2012;31:85–91. doi: 10.1097/NOR.0b013e31824fcd42. [DOI] [PubMed] [Google Scholar]
- 5.Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projection of primary and revision hip and knee arthroplasty in the United States from 2005–2030. J Bone Joint Surg Am. 2007;89:780–785. doi: 10.2106/JBJS.F.00222. [DOI] [PubMed] [Google Scholar]
- 6.Resnic FS, Normand ST. Postmarketing surveillance of medical devices-filling in the gaps. N Engl J Med. 2012;366:875–877. doi: 10.1056/NEJMp1114865. [DOI] [PubMed] [Google Scholar]
- 7.Geetha M, Singh AK, Asokamani R, Gogia AK. Ti based biomaterials, the ultimate choice for orthopaedic implants – A review. Prog Mat Sci. 2009;54:397–425. [Google Scholar]
- 8.Paxton EW, Fumes O, Namba RS, Inacio MCS, Fenstad AM, Havelin LI. Comparison of the Norwegian knee anthroplasty register and a United States anthroplasty registry. J Bone Joint Surg Am. 2011;93E:20–30. doi: 10.2106/JBJS.K.01045. [DOI] [PubMed] [Google Scholar]
- 9.Mavrogenis AF, Dimitriou R, Parvizi J, Babis GC. Biology of implant osseointegration. J Musculoskelet Neuronal Interact. 2009;9:61–71. [PubMed] [Google Scholar]
- 10.Niinomi M. Metallic biomaterials. J Artif Organs. 2008;11:105–110. doi: 10.1007/s10047-008-0422-7. [DOI] [PubMed] [Google Scholar]
- 11.Sakka S, Coulthard P. Implant failure: Etiology and complications. Med Oral Patol Oral Cir Bucal. 2011;16:42–44. doi: 10.4317/medoral.16.e42. [DOI] [PubMed] [Google Scholar]
- 12.Hench LL. Biomaterials. Science. 1980;208:826–831. doi: 10.1126/science.6246576. [DOI] [PubMed] [Google Scholar]
- 13.Scholz MS, Blanchfield LD, Bloom LD, Coburn BH, Elkington M, Fuller JD, Gilbert ME, Muflahi SA, Pernice MF, Pae SI, Trevarthen JA, White SC, Weaver PM, Bond IP. The use of composite materials in modern orthopedic medicine and prosthetic devices: A review. Compos Sci Technol. 2011;71:1791–1803. [Google Scholar]
- 14.Taljanovic MS, Jones MD, Ruth JT, Benjamin JB, Sheppard JE, Hunter TB. Joint arthroplasties and prostheses. Frac Fix Radio. 2003;23:1569–1590. doi: 10.1148/rg.235035059. [DOI] [PubMed] [Google Scholar]
- 15.Ratner BD, Hoffman AS, Schoen FJ, Lemons JE. Biomaterials science An introduction to Materials in Medicine. California: Elsevier Academic Press; 2008. p. 864. [Google Scholar]
- 16.Manivasagam G, Dhinasekaran D, Rajamanickam A. biomedical implants: corrosion and its prevention – A review. Rec Pat Corros Sci. 2010;2:40–54. [Google Scholar]
- 17.Amel-Farzad H, Peivandi MT, Yusof-Sani SMR. In-body corrosion fatigue failure of a stainless steel orthopaedic implant with a rare collection of different damage mechanisms. Eng Fail Anal. 2007;14:1205–1217. [Google Scholar]
- 18.Fini M, Aldini NN, Torricelli P, Giavaresi G, Borsari V, Lenger H. A new austenitic stainless steel with negligible nickel content: An in vitro and in vivo comparative investigation. Biomaterials. 2003;24:4929–4939. doi: 10.1016/s0142-9612(03)00416-2. [DOI] [PubMed] [Google Scholar]
- 19.Lin DJ, Chuang CC, Lin JHC, Lee JW, Ju CP, Yin HS. Bone formation at the surface of low modulus Ti–7 5Mo implants in rabbit femur. Biomaterials. 2007;28:2582–2589. doi: 10.1016/j.biomaterials.2007.02.005. [DOI] [PubMed] [Google Scholar]
- 20.Müller R, Abke J, Schnell E, Macionczyk F, Gbureck U, Mehrl R. Surface engineering of stainless steel materials by covalent collagen immobilization to improve implant biocompatibility. Biomaterials. 2005;26:6962–6972. doi: 10.1016/j.biomaterials.2005.05.013. [DOI] [PubMed] [Google Scholar]
- 21.Okazaki Y, Gotoh E. Comparison of metal release from various metallic biomaterials in vitro. Biomaterials. 2005;26:11–21. doi: 10.1016/j.biomaterials.2004.02.005. [DOI] [PubMed] [Google Scholar]
- 22.Takemoto S, Hattori M, Yoshinari M, Kawada E, Oda Y. Corrosion behavior and surface characterization of titanium in solution containing fluoride and albumin. Biomaterials. 2005;26:829–837. doi: 10.1016/j.biomaterials.2004.03.025. [DOI] [PubMed] [Google Scholar]
- 23.Begum Z, Poonguzhali A, Basu R, Sudha C, Shaikh H, Rao RVS. Studies of the tensile and corrosion fatigue behaviour of austenitic stainless steels. Corros Sci. 2011;53:1424–1432. [Google Scholar]
- 24.Walter MJ. Stainless Steel for Medical: The high level of nitrogen in BioDur 108 stainless steel provides anhanced mechanical and physical properties for medical implants. Adv mater proc. 2006 Apr;:84–86. [Google Scholar]
- 25.Braic M, Balaceanu M, Braic V, Vladescu A, Pavelescu G, Albulescu M. Synthesis and characterization of TiN, TiAIN and TiN/TiAIN biocompatible coatings. Surf Coat Technol. 2005;200:1014–1017. [Google Scholar]
- 26.Yamamoto A, Kohama Y, Kuroda D, Hanawa T. Cytocompatibility evaluation of Ni-free stainless steel manufactured by nitrogen adsorption treatment. Mater Sci Eng C. 2004;24:737–743. [Google Scholar]
- 27.Wang P, Ren Y, Zhang B, Yang K. Effect of nitrogen on blood compatibility of nickel-free high nitrogen stainless steel for biomaterial. Mater Sci Eng C. 2010;30:1183–1189. [Google Scholar]
- 28.Kang CK, Lee YS. The surface modification of stainless steel and the correlation between the surface properties and protein adsorption. J Mater Sci: Mater in med. 2007;18:1389–1398. doi: 10.1007/s10856-006-0079-9. [DOI] [PubMed] [Google Scholar]
- 29.Zhang F, Kang ET, Neoh KG, Wang P, Tan KL. Surface modification of stainless steel by grafting of poly(ethylene glycol) for reduction in protein adsorption. Biomaterials. 2001;22:1541–1548. doi: 10.1016/s0142-9612(00)00310-0. [DOI] [PubMed] [Google Scholar]
- 30.Berry CW, Moore TJ, Safar JA, Henry CA, Wagner MJ. Antibacterial activity of dental implant metals. Implant Dent. 1992;1:59–65. doi: 10.1097/00008505-199200110-00006. [DOI] [PubMed] [Google Scholar]
- 31.Dong Y, Li X, Sammons R, Dong H. The generation of wear resistant antimicrobial stainless steel surface by active screen plasma alloying with N and nanocrytaslline Ag. J Biomed Mat Res. 2010;93:185–193. doi: 10.1002/jbm.b.31573. [DOI] [PubMed] [Google Scholar]
- 32.Van Der Borden AJ, Maathuis PG, Engels E, Rakhorst G, Van Der Mei HC, Busscher HJ, Sharma PK. Prevention of pin tract infection in external stainless steel fixator frames using electric current in a goat model. Biomaterials. 2007;28:2122–2126. doi: 10.1016/j.biomaterials.2007.01.001. [DOI] [PubMed] [Google Scholar]
- 33.Lin CW, Ju CP, Lin JHC. A comparison of the fatigue behavior of cast Ti–7.5Mo with c.p. titanium, Ti–6Al–4V and Ti–13Nb–13Zr alloys. Biomaterials. 2005;26:2899–2907. doi: 10.1016/j.biomaterials.2004.09.007. [DOI] [PubMed] [Google Scholar]
- 34.Long M, Rack HJ. Titanium alloys in total joint replacement – A materials science perspective. Biomaterials. 1998;19:1621–1639. doi: 10.1016/s0142-9612(97)00146-4. [DOI] [PubMed] [Google Scholar]
- 35.Samandari SS, Berndt CC, Gross KA. Selection of the implant and coating materials for optimized performance by means of nanoindentation. Acta Biomater. 2011;7:874–881. doi: 10.1016/j.actbio.2010.09.023. [DOI] [PubMed] [Google Scholar]
- 36.Kamachi MU, Baldev R. Corrosion science and technology: Mechanism, mitigation and monitoring. UK: Taylor & Francis; 2008. pp. 283–356. [Google Scholar]
- 37.Novikova GE. Introduction to corrosion of bioimplants. Prot Met Phys Chem Surf. 2011;47:372–380. [Google Scholar]
- 38.Zhuang L, Langer EW. Effects of alloy additions on the fatigue properties of cast Co–Cr–Mo alloy used for surgical implants. J Mater Sci. 1990;25:683–689. [Google Scholar]
- 39.Song YC, Park C-H, Moriwaki T. Mirror finishing of Co–Cr–Mo alloy using elliptical vibration cutting. Precis Eng. 2010;34:784–789. [Google Scholar]
- 40.Williams DF. Biocompatibility of Clinical Implant Materials. Michigan: CRC Press; 1981. p. 288. [Google Scholar]
- 41.Navarro M, Michiardi A, Castano O, Planell JA. Biomaterials in orthopaedics. J R Soc Interf. 2008;5:1137–1158. doi: 10.1098/rsif.2008.0151. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Antunes RA, Oliveira MC. Corrosion fatigue of biomedical metallic alloys: Mechanisms and mitigation. Acta Biomaterialia. 2012;8:937–962. doi: 10.1016/j.actbio.2011.09.012. [DOI] [PubMed] [Google Scholar]
- 43.Anwar HA, Hart AJ, Visuvanathan S, Aldam CH. The effects of the heavy metals from total hip replacement wear debris on bacterial growth? J Bone Joint Surg Br. 2009;91:300. doi: 10.1302/0301-620X.89B12.19714. [DOI] [PubMed] [Google Scholar]
- 44.Hosman AH, Mei HC, Bulstra SK, Kuijer R, Busscher HJ, Neut D. Influence of Co-Cr particles and Co-Cr ions on the growth of staphylococcal biofilms. Int J Artif Organs. 2011;34:759–765. doi: 10.5301/ijao.5000031. [DOI] [PubMed] [Google Scholar]
- 45.Lhotka C, Szekes T, Stefan I. Four years study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacement. J of Orthopedic Res. 2003;21:189–195. doi: 10.1016/S0736-0266(02)00152-3. [DOI] [PubMed] [Google Scholar]
- 46.Keegan GM, Leatmonth ID, Case CP. A systemic comparison of the actual potential and theoretical health effects of cobalt and chromium exposures from industry and surgical implants. Crit Rev Toxicol. 2008;38:645–674. doi: 10.1080/10408440701845534. [DOI] [PubMed] [Google Scholar]
- 47.McAfolaranmi GA, Tettey J, Meek RMD, Grant MH. Release of chromium from orthopedic anthroplasties. The Open Ortho J. 2008;2:10–18. doi: 10.2174/1874325000802010010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Polyzois I, Nikolopoulos D, Michos I, Patsouris E, Theocharis S. Local and systemic toxicity of nanoscale debris particles in total hip anthroplasty. J Appl Toxicol. 2012;32:255–269. doi: 10.1002/jat.2729. [DOI] [PubMed] [Google Scholar]
- 49.Mori M, Yamanaka K, Matsumoto H, Chiba A. Evolution of cold-rolled microstructures of biomedical Co–Cr–Mo alloys with and without N doping. Mater Sci Eng A. 2010;528:614–621. [Google Scholar]
- 50.Yamanaka K, Mori M, Chiba A. Mechanical properties of as-forged Ni-free Co–29Cr–6Mo alloys with ultrafine-grained microstructure. Mater Sci Eng A. 2011;528:5961–5966. [Google Scholar]
- 51.Yamanaka K, Mori M, Kurosu S, Matsumoto H, Chiba A. Ultrafine grain refinement of biomedical Co–29Cr–6Mo alloy during conventional hot-compression deformation. Metall Mater Trans A. 2009;40:1980–1994. [Google Scholar]
- 52.Bhure R, Fattah TM, Bonner C, Hall F, Mahapatro A. Stability of phosphonic self-assembled monolayers (SAMs) on cobalt chromium (Co-Cr) alloy under oxidative conditions. Appl Surf Sci. 2011;257:5605–5612. doi: 10.1016/j.apsusc.2011.01.055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Poh CK, Shi Z, Tan XW, Liang ZC, Foo XM, Tan HC, Neoh KG, Wang W. Cobalt chromium alloy with immobilized BMP peptide for enhanced bone growth. J Orthop Res. 2011;29:1424–1430. doi: 10.1002/jor.21409. [DOI] [PubMed] [Google Scholar]
- 54.Levine BR, Sporer S, Poggie RA, Della Valle CJ, Jacobs JJ. Experimental and clinical performance of porous tantalum in orthopedic surgery. Biomaterials. 2006;27:4671–4681. doi: 10.1016/j.biomaterials.2006.04.041. [DOI] [PubMed] [Google Scholar]
- 55.Navarro M, Michiardi A, Castano O, Planell JA. Biomaterials in orthopaedics. J R Soc Inter. 2008;5:1137–1158. doi: 10.1098/rsif.2008.0151. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Palmquist A, Jarman T, Emanuelsson L, Branemark R, Engqvist H, Thomsen P. Forearm bone-anchored amputation prosthesis: a case study on the osseointegration. Acta Orthop. 2008;79:78–85. doi: 10.1080/17453670710014806. [DOI] [PubMed] [Google Scholar]
- 57.Boyer RR. An overview on the use of titanium in the aerospace industry. Mater Sci Eng A. 1996;213:103–114. [Google Scholar]
- 58.Nag S, Sonia S, Anantha P, Rajarshi B. Characterization of novel borides in Ti–Nb–Zr–Ta+ 2B metal-matrix composites. Mat charac. 2009;60:106–113. [Google Scholar]
- 59.Brailovski V, Prokoshkin S, Gauthier M, Inaekyan K, Dubinskiy S, Petrzhik M. Bulk and porous metastable beta Ti–Nb–Zr(Ta) alloys for biomedical applications. Mater Sci Eng C. 2011;31:643–657. [Google Scholar]
- 60.Li JP, De Wijn JR, Van Blitterswijk CA, De Groot K. Porous Ti6Al4V scaffold directly fabricating by rapid prototyping: preparation and in vitro experiment. Biomaterials. 2006;27:1223–1235. doi: 10.1016/j.biomaterials.2005.08.033. [DOI] [PubMed] [Google Scholar]
- 61.Van den Dolder J, Farber E, Spauwen PH, Jansen JA. Bone tissue reconstruction using titanium fiber mesh combined with rat bone marrow stromal cells. Biomaterials. 2003;24:1745–1750. doi: 10.1016/s0142-9612(02)00537-9. [DOI] [PubMed] [Google Scholar]
- 62.Mendonca G, Mendonca DBS, Simoes LGP, Araujo AL, Leite ER, Duarte WR, Aragao FJL, Cooper LF. The effects of implant surface nanoscale features on osteoblast-specific gene expression. Biomaterials. 2009;30:4053–4062. doi: 10.1016/j.biomaterials.2009.04.010. [DOI] [PubMed] [Google Scholar]
- 63.Muller U, Imwinkelried T, Horst M, Sievers M, Graf- Hausner U. Do human osteoblasts grow into open-porous titanium? Eur Cells Mater. 2006;11:8–15. doi: 10.22203/ecm.v011a02. [DOI] [PubMed] [Google Scholar]
- 64.Barradas AMC, Yuan H, Blitterswijk CA, Habibovic P. Osteoinductive biomaterials: current knowledge of properties, experimental models and biological mechanisms. Eur Cell Mater. 2011;21:407–429. doi: 10.22203/ecm.v021a31. [DOI] [PubMed] [Google Scholar]
- 65.Perl DP, Brody AR. Alzheimer’s disease: X-ray spectrometer evidence of aluminum accumulation in neurofibrillary tangle-bearing neurons. Science. 1980;208:297–299. doi: 10.1126/science.7367858. [DOI] [PubMed] [Google Scholar]
- 66.Japan Aluminium Association. Aluminum and health. Tokyo: Japan Aluminium Association; 2002. pp. 1–6. [Google Scholar]
- 67.Eisenbarth E, Velten D, Muller M, Thull R, Breme J. Biocompatibility of beta-stabilizing elements of titanium alloys. Biomaterials. 2004;25:5705–5713. doi: 10.1016/j.biomaterials.2004.01.021. [DOI] [PubMed] [Google Scholar]
- 68.Semlitsch M. Classic and new titanium alloys for production of artificial hip joints. Titan. 1986;2:721–740. [Google Scholar]
- 69.Browy K-H, Kramer KH. On the properties of a new titanium alloy (TiAl5Fe2.5) as implant material. In: Lütjering G, Zwicker U, Bunk W, editors. Titanium science and technology. Vol. 2. Stuttgart: Deutsche Gesellshaft fur Metallkunde; 1985. pp. 1381–1386. [Google Scholar]
- 70.Okazaki Y, Ito Y, Tateishi T, Ito A. Effect of heat treatment on microstructure and mechanical properties of new titanium alloys for surgical implantation. J Jpn Inst Met. 1995;59:108–115. [Google Scholar]
- 71.Mouhyi J, Dohan Ehrenfest DM, Albrektsson T. The Peri-implantitis: implant surfaces, microstructure and physico-chemical aspects. Clin Implant Dent Relat Res. 2012;14:170–183. doi: 10.1111/j.1708-8208.2009.00244.x. [DOI] [PubMed] [Google Scholar]
- 72.Boehlert CJ, Cowen CJ, Quast JP, Akahori T, Niinomi M. Fatigue and wear evaluation of Ti–Al–Nb alloys for biomedical applications. Mater Sci Eng C. 2008;28:323–330. [Google Scholar]
- 73.Ghonem H. Microstructure and fatigue crack growth mechanisms in high temperature titanium alloys. Int J Fatigue. 2010;32:1448–1460. [Google Scholar]
- 74.Mishra AK, Davidson JA, Poggie RA, Kovacs P, Fitzgerald TJ. Mechanical and tribological properties and biocompatibility of diffusion hardened Ti–13Nb– 13Zr – a new titanium alloy for surgical implants. In: Brown SA, Lemons JE, editors. Medical applications of titanium and its alloys, ASTM STP 1272. West Conshohocken, PA: ASTM International; 1996. pp. 96–116. [Google Scholar]
- 75.Wang KK, Gustavson LJ, Dumbleton JH. Micro structure and properties of a new beta titanium alloy, Ti–12Mo–6Zr–2Fe, developed for surgical implants. In: Brown SA, Lemons JE, editors. Medical applications of titanium and its alloys, ASTM STP 1272. West Conshohocken, PA: ASTM International; 1996. pp. 76–87. [Google Scholar]
- 76.Zardiackas LD, Mitchell DW, Disegi JA. Characterization of Ti–15Mo beta titanium alloy for orthopedic implant. In: Brown SA, Lemons JE, editors. Medical applications of titanium and its alloys, ASTM STP 1272. West Conshohocken, PA: ASTM International; 1996. pp. 60–75. [Google Scholar]
- 77.Hatanaka S, Ueda M, Ikeda M, Niinomi M. Isothermal aging behavior in Ti– 10Cr–Al alloys for medical applications. Adv Mater Res. 2010;89–91:232–237. [Google Scholar]
- 78.Ikeda M, Ueda M, Kinoshita T, Ogawa M, Niinomi M. Influence of Fe content of Ti–Mn–Fe alloys on phase constitution and heat treatment behavior. Mater Sci Forum. 2012;706–709:1893–1898. [Google Scholar]
- 79.Ikeda M, Ueda M, Matsunaga R, Niinomi M. Phase constitution and heat treatment behavior of Ti–7 mass%Mn–Al alloys. Mater Sci Forum. 2010;654–656:855–858. [Google Scholar]
- 80.Kasano Y, Inamura T, Kanetaka H, Miyazaki S, Hosoda H. Phase constitution and mechanical properties of Ti–(Cr, Mn)–Sn biomedical alloys. Mater Sci Forum. 2010;654–656:2118–2121. [Google Scholar]
- 81.Tane M, Akita S, Nakano T, Hagihara K, Umakoshi Y, Niinomi M, et al. Peculiar elastic behavior of Ti–Nb–Ta–Zr single crystals. Acta Mater. 2008;56:2856–2863. [Google Scholar]
- 82.Nakai M, Niinomi M, Zhao XF, Zhao XL. Self-adjustment of Young’s modulus in biomedical titanium alloy during orthopaedic operation. Mater Lett. 2011;65:688–690. [Google Scholar]
- 83.Zhao XF, Niinomi M, Nakai M, Hieda J. Beta type Ti–Mo alloys with changeable Young’s modulus for spinal fixation applications. Acta Biomater. 2012;8:1990–1997. doi: 10.1016/j.actbio.2012.02.004. [DOI] [PubMed] [Google Scholar]
- 84.Alvarado J, Maldonado R, Marxuach J, Otero R. Biomechanics of hip and knee prostheses. Appl Eng Mech Med. 2003:1–20. [Google Scholar]
- 85.Hosoda H, Hosoda N, Miyazaki S. Mechanical properties of Ti–Mo–Al biomedical shape memory alloys. Trans MRS-J. 2001;26:243–246. [Google Scholar]
- 86.Ikeda M, Komatsu S, Nakamura Y. Effects of Sn and Zr additions on phase constitution and aging behavior of Ti–50 mass%Ta alloys quenched from b single phase region. Mater Trans. 2004;45:1106–1112. [Google Scholar]
- 87.Ikeda M, Sugano D, Masuda S, Ogawa M. The influence of aluminum content on shape memory effect of Ti–7Cr–Al alloys fabricated using low grade sponge titanium. Mater Trans. 2005;46:1604–1609. [Google Scholar]
- 88.Borsari V, Fini M, Giavaresi G. Osteointegration of titanium and hydroxyapatite rough surfaces in healthy and compromised cortical and trabecular bone: in vivo comparative study on young, aged, and estrogen-deficient sheep. J Orthop Res. 2007;25:1250–1260. doi: 10.1002/jor.20413. [DOI] [PubMed] [Google Scholar]
- 89.Cooper LF, Zhou Y, Takebe J. Fluoride modification effects on osteoblast behavior and bone formation at TiO2 grit-blasted c.p. titanium endosseous implants. Biomaterials. 2006;27:926–936. doi: 10.1016/j.biomaterials.2005.07.009. [DOI] [PubMed] [Google Scholar]
- 90.Rupp F, Scheideler L, Olshanska N. Enhancing surface free energy and hydrophilicity through chemical modification of micro structured titanium implant surfaces. J Biomed Mater Res. 2006;76:323–334. doi: 10.1002/jbm.a.30518. [DOI] [PubMed] [Google Scholar]
- 91.Morra M. Biochemical modification of Ti surfaces: peptides and ECM proteins. Eur Cell Mater. 2006;12:1–15. doi: 10.22203/ecm.v012a01. [DOI] [PubMed] [Google Scholar]
- 92.Rammelt S, Illert T, Bierbaum S. Coating of titanium implants with collagen, RGD peptide and chondroitin sulphate. Biomaterials. 2006;27:5561–5572. doi: 10.1016/j.biomaterials.2006.06.034. [DOI] [PubMed] [Google Scholar]
- 93.Kinov P, Leithner A, Radl R, Bodo K, Khoschsorur GA, Schauenstein K, Windhager R. Role of free radicals in aseptic loosening of hip arthroplasty. J Orthop Res. 2006;24:55–62. doi: 10.1002/jor.20013. [DOI] [PubMed] [Google Scholar]
- 94.Bucci-Sabattini V, Cassinelli C, Coelho P, Minnici A, Trani A, Ehernfest DD. Effect of titanium implant surface nanoroughness and calcium phosphate low impregnation on bone cell activity in vitro. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010;109:217–224. doi: 10.1016/j.tripleo.2009.09.007. [DOI] [PubMed] [Google Scholar]
- 95.McNamara LE, Sjostrom T, Burgess KE, Kim JJ, Liu E, Gordonov S, Moghe PV, Meek RM, Oreffo RO, Su B, Dalby MJ. Skeletal stem cell physiology on functionally distinct titania nanotopographies. Biomaterials. 2011;32:7403–7410. doi: 10.1016/j.biomaterials.2011.06.063. [DOI] [PubMed] [Google Scholar]
- 96.Bjursten LM, Rasmusson L, Oh S, Smith GC, Brammer KS, Jin S. Titanium dioxide nanotubes enhance bone bonding in vivo. J Biomed Mater Res A. 2010;92:1218–1224. doi: 10.1002/jbm.a.32463. [DOI] [PubMed] [Google Scholar]
- 97.Shalabi MM, Gortemaker A, Van’t Hof MA, Jansen JA, Creugers NH. Implant surface roughness and bone healing: A systematic review. J Dent Res. 2006;85:496–500. doi: 10.1177/154405910608500603. [DOI] [PubMed] [Google Scholar]
- 98.Bourne GH, Herring GM. The biochemistry and physiology of bone. New York: Academic Press; 1972. The organic matrix of bone; pp. 127–189. [Google Scholar]
- 99.Porter JR, Ruckh TT, Popat KC. Bone tissue engineering: a review in bone biomimetics and drug delivery strategies. Biotechnol Prog. 2009;25:1539–1560. doi: 10.1002/btpr.246. [DOI] [PubMed] [Google Scholar]
- 100.Barbieri D, Renard AJ, de Bruijn JD, Yuan H. Heterotopic bone formation by nano apatite containing poly(D,L-lactide) composites. Eur Cell Mater. 2010;19:252–261. doi: 10.22203/ecm.v019a24. [DOI] [PubMed] [Google Scholar]
- 101.Oh S, Daraio C, Chen LH, Pisanic TR, Finones RR, Jin S. Significantly accelerated osteoblast cell growth on aligned TiO2 nanotubes. J Biomed Mater Res Part A. 2006;78:97–103. doi: 10.1002/jbm.a.30722. [DOI] [PubMed] [Google Scholar]
- 102.Findlay DM, Welldon K, Atkin GJ, Howie DW, Zannettino ACW, Bobyn D. The proliferation and phenotypic expression of human osteoblasts on tantalum metal. Biomaterials. 2004;25:2215–2227. doi: 10.1016/j.biomaterials.2003.09.005. [DOI] [PubMed] [Google Scholar]
- 103.Levine BR, Sporer S, Poggie RA, Valle CJD, Jacobs JJ. Experimental and clinical performance of porous tantalum in orthopedic surgery. Biomaterials. 2006;27:4671–4681. doi: 10.1016/j.biomaterials.2006.04.041. [DOI] [PubMed] [Google Scholar]
- 104.Sevilla P, Aparicio C, Planell JA, Gil FJ. Comparison of the mechanical properties between tantalum and nickel–titanium foams implant materials for bone ingrowth applications. J Alloys Compd. 2007;439:67–73. [Google Scholar]
- 105.Levine B, Della Valle CJ, Jacobs JJ. Applications of porous tantalum in total hip arthroplasty. J Am Acad Orthop Surg. 2006;14:646–655. doi: 10.5435/00124635-200611000-00008. [DOI] [PubMed] [Google Scholar]
- 106.Sheth NP, Lonner JH. Clinical use of porous tantalum in complex primary total knee anthroplasty. Am J Orthop. 2009;38:526–530. [PubMed] [Google Scholar]
- 107.Bobyn JD, Poggie RA, Krygier JJ, Lewallen DG, Hanssen AD, Lewis RJ, Unger AS, O’Keefe TJ, Christie MJ, Nasser S, Wood JE, Stulberg SD, Tanzer M. Clinical validation of a structural porous tantalum biomaterial for adult reconstruction. J Bone Joint Surg. 2004;86:123–129. doi: 10.2106/00004623-200412002-00017. [DOI] [PubMed] [Google Scholar]
- 108.Rubitschek E, Niendorf T, Karaman I, Maier HJ. Corrosion fatigue behavior of a biocompatible ultrafine-grained niobium alloy in simulated body fluid. J Mech Behav Biomed Mater. 2012;5:181–192. doi: 10.1016/j.jmbbm.2011.08.023. [DOI] [PubMed] [Google Scholar]
- 109.Kondo R, Nomura N, Suyalatu, Tsutsumi Y, Doi H, Hanawa T. Microstructure and mechanical properties of as-cast Zr–Nb alloys. Acta Biomater. 2011;7:4278–4284. doi: 10.1016/j.actbio.2011.07.020. [DOI] [PubMed] [Google Scholar]
- 110.Basso A, Roy S, Mermoud A. Biocompatibility of an xshaped zirconium implant in deep sclerectomy in rabbits. Graefe’s Archive for Clinical and Experimental Ophthalmology. 2008;246:849–855. doi: 10.1007/s00417-008-0782-y. [DOI] [PubMed] [Google Scholar]
- 111.Suyalatu, Kondo R, Tsutsumi Y, Doi H, Nomura N, Hanawa T. Effects of phase constitution on magnetic susceptibility and mechanical properties of Zr-rich Zr–Mo alloys. Acta Biomater. 2011;7:4259–4266. doi: 10.1016/j.actbio.2011.07.005. [DOI] [PubMed] [Google Scholar]
- 112.Galetz MC, Fleischmann EW, Konrad CH, Schuetz A, Glatzel U. Abrasion resistance of oxidized zirconium in comparison with CoCrMo and titanium nitride coatings for artificial knee joints. J Biomed Mater Res B. 2010;93B:244–251. doi: 10.1002/jbm.b.31581. [DOI] [PubMed] [Google Scholar]
- 113.Niinomi M. Mechanical biocompatibilities of titanium alloys for biomedical applications. J Mech Behav Biomed Mater. 2008;1:30–42. doi: 10.1016/j.jmbbm.2007.07.001. [DOI] [PubMed] [Google Scholar]
- 114.Witte F. The histry of biodegradable magnesium: a review. Acta Biomater. 2010;6:1680–1692. doi: 10.1016/j.actbio.2010.02.028. [DOI] [PubMed] [Google Scholar]
- 115.Gu XN, Li N, Zhou WR, Zheng YF, Zhao X, Cai QZ, et al. Corrosion resistance and surface biocompatibility of a microarc oxidation coating on a Mg–Ca alloy. Acta Biomater. 2011;7:1880–1889. doi: 10.1016/j.actbio.2010.11.034. [DOI] [PubMed] [Google Scholar]
- 116.Kannan MB, Raman RKS. Evaluating the stress corrosion cracking susceptibility of Mg–Al–Zn alloy in modified-simulated body fluid for orthopaedic implant application. Scripta Mater. 2008;59:175–178. [Google Scholar]
- 117.Vormann J. Magnesium: Nutrition and metabolism. Mol Aspects Med. 2003;24:27–37. doi: 10.1016/s0098-2997(02)00089-4. [DOI] [PubMed] [Google Scholar]
- 118.Wang J, Tang J, Zhang P, Wang J, Lai Y, Qin L. Surface modification of magnesium alloys developed for bioabsorbable orthopedic implants: a general review. J Biomed Mat Res Part B. 2012;100B:1691–1701. doi: 10.1002/jbm.b.32707. [DOI] [PubMed] [Google Scholar]
- 119.Rahaman MN, Yao A. Ceramics for prosthetic hip and knee joint replacement. J Am Ceram Soc. 2007;90:1965–1988. [Google Scholar]
- 120.Urban A, Garvin KL, Boese CK, Bryson L, Pedersen DR, Callaghan JJ, Miller RK. Ceramic-on-Polyethylene Bearing Surfaces in Total Hip Arthroplasty. Seventeen to Twenty-One-Year Results. J Bone Joint Surg Am. 2001;83:1688–1694. doi: 10.2106/00004623-200111000-00011. [DOI] [PubMed] [Google Scholar]
- 121.Sarkar R, Banerjee G. Ceramic based bio-medical implants. Interceram. 2010;59:98–102. [Google Scholar]
- 122.Campbell P, Shen FW, McKellop H. Biologic and tribologic considerations of alternative bearing surfaces. Clin Orthop. 2004;418:98–111. doi: 10.1097/00003086-200401000-00017. [DOI] [PubMed] [Google Scholar]
- 123.Garvie RC, Hannink RH, Pascoe RT. Ceramic steel? Nature. 1975;258:703–704. [Google Scholar]
- 124.Garvie RC, Nicholson PS. Structure and thermodynamical properties of partially stabilized zirconia in the CaO-ZrO2 system. J Am Ceram Soc. 1972;55:152–157. [Google Scholar]
- 125.Torricelli P, Verne E, Brovarone CV, Appendino P, Rustichelli F, Krajewski A. Biological glass coating on ceramic materials: in vitro evaluation using primary osteoblast cultures from healthy and osteopenic rat bone. Biomaterials. 2001;22:2535–2543. doi: 10.1016/s0142-9612(00)00444-0. [DOI] [PubMed] [Google Scholar]
- 126.Degidi M, Artese L, Scarano A, Perrotti V, Gehrke P, Piattelli A. Inflammatory infiltrate, microvessel density, nitric oxide synthase expression, vascular endothelial growth factor expression, and proliferative activity in peri-implant soft tissues around titanium and zirconium oxide healing caps. J Periodont. 2006;77:73–80. doi: 10.1902/jop.2006.77.1.73. [DOI] [PubMed] [Google Scholar]
- 127.Lohmann CH, Dean DD, Koster G, Casasola D, Buchhorn GH, Fink U. Ceramic and PMMA particles differentially affect osteoblast phenotype. Biomaterials. 2002;23:1855–1863. doi: 10.1016/s0142-9612(01)00312-x. [DOI] [PubMed] [Google Scholar]
- 128.Silva VV, Lameiras FS, Lobato ZI. Biological reactivity of zirconia-hydroxyapatite composites. J Biomed Mater Research. 2002;63:583–590. doi: 10.1002/jbm.10308. [DOI] [PubMed] [Google Scholar]
- 129.Carinci F, Pezzetti F, Volinia S, Francioso F, Arcelli D, Farina E. Zirconium oxide: analysis of MG63 osteoblast-like cell response by means of a microarray technology. Biomaterials. 2004;25:215–228. doi: 10.1016/s0142-9612(03)00486-1. [DOI] [PubMed] [Google Scholar]
- 130.Chevalier L, Gremillard, Deville S. Low-Temperature Degradation of Zirconia and Implications for Biomedical Implants. Annu Rev Mater Res. 2007;37:1–32. [Google Scholar]
- 131.Clarke IC, Manaka M, Green DD, Williams P, Pezzotti G, Kim YH, Ries M, Donaldson NT, Gustafson GA. Current Status of Zirconia Used in Total Hip Implants. J Bone Joint Surg Am. 2003;85:73–84. doi: 10.2106/00004623-200300004-00009. [DOI] [PubMed] [Google Scholar]
- 132.Schubert H, Frey F. Stability of Y-TZP during hydrothermal treatment: neutron experiments and stability considerations. J Eur Ceram Soc. 2005;25:1597–1602. [Google Scholar]
- 133.Willmann G, Fru HJ, Pfaff HG. Wear Characteristics of Sliding Pairs of Zirconia (Y-TZP) for Hip Endoprostheses. Biomaterials. 1996;17:2157–2162. doi: 10.1016/0142-9612(96)00042-7. [DOI] [PubMed] [Google Scholar]
- 134.El Gannham A. Bone reconstruction: from bioceramics to tissue engineering. Exp Rev Med Dev. 2005;2:87–101. doi: 10.1586/17434440.2.1.87. [DOI] [PubMed] [Google Scholar]
- 135.Ikawa N, Kimura T, Oumi Y, Sano T. Amino acid containing amorphous calcium phosphates and the rapid transformation into apatite. J Mater Chem. 2009;19:4906–4913. [Google Scholar]
- 136.Liu Y, de Groot K, Hunziker EB. BMP-2 liberated from biomimetic implant coatings induces and sustains direct ossification in an ectopic rat model. Bone. 2005;36:745–757. doi: 10.1016/j.bone.2005.02.005. [DOI] [PubMed] [Google Scholar]
- 137.Weir MD, Xu HH. High-strength, in situ-setting calcium phosphate composite with protein release. J Biomed Mater Res. 2008;85:388–396. doi: 10.1002/jbm.a.31347. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Jayasuriya AC, Shah C. Controlled Release of Insulin like Growth Factor-1 and Bone Marrow Stromal Cell Function of Bone-Like Mineral Layers Coated PLGA Scaffolds. J Tissue Eng Regen Med. 2008;2:43–49. doi: 10.1002/term.65. [DOI] [PubMed] [Google Scholar]
- 139.Bandyopadhyay A, Bernard S, Xue W, Bose S. Calcium phosphate-based resorbable ceramics: influence of MgO, ZnO, and SiO2 dopants. J Am Cer Soc. 2006;89:2675–2688. [Google Scholar]
- 140.Hench LL, Wilson J. Surface-active biomaterials. Science. 1984;226:630–636. doi: 10.1126/science.6093253. [DOI] [PubMed] [Google Scholar]
- 141.LeGeros RZ. Calcium phosphate-based osteoinductive materials. Chem Rev. 2008;108:4742–4753. doi: 10.1021/cr800427g. [DOI] [PubMed] [Google Scholar]
- 142.Jayasuriya AC, Kibbe S. Rapid biomineralization of chitosan microparticles to apply in bone regeneration. J Mater Sci Mater Med. 2010;21:393–398. doi: 10.1007/s10856-009-3874-2. [DOI] [PubMed] [Google Scholar]
- 143.Jayasuriya AC, Asaad M, Jayatissa AH, Ebraheim NA. Dissolution Behavior of Biomimetic Minerals on 3D PLGA Scaffold. Surf Coat Tech. 2006;200:6336–6339. [Google Scholar]
- 144.Shin K, Jayasuriya AC, Kohn DH. Effect of Ionic Activity Products on the Structure and Composition of the Mineral formed on 3-D poly(lactic-co-glycolide) Scaffolds. J Biomed Mater Res A. 2007;83:1076–1086. doi: 10.1002/jbm.a.31437. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 145.Jayasuriya AC, Shah C, Ebraheim NA, Jayatissa AH. Acceleration of biomimetic mineralization to apply in bone regeneration. Biomed Mater. 2008;3:015003. doi: 10.1088/1748-6041/3/1/015003. [DOI] [PubMed] [Google Scholar]
- 146.Li Y, Weng W, Tam KC. Novel highly biodegradable biphasic tricalcium phosphates composed of [alpha]-tricalcium phosphate and [beta]-tricalcium phosphate. Acta Biomater. 2007;3:251–254. doi: 10.1016/j.actbio.2006.07.003. [DOI] [PubMed] [Google Scholar]
- 147.Matsushita N, Terai H, Okada T, Nozaki K, Inoue H, Miyamoto S. A new bone-inducing biodegradable porous b-tricalcium phosphate. J Biomed Mater Res. 2004;70:450–458. doi: 10.1002/jbm.a.30102. [DOI] [PubMed] [Google Scholar]
- 148.Yuan H, van Blitterswijk CA, de Groot K, de Bruijn JD. A comparison of bone Formation in biphasic calcium phosphate (BCP) and hydroxyapatite (HA) implanted in muscle and bone of dogs at different time periods. J Biomed Mater Res A. 2006;78:139–147. doi: 10.1002/jbm.a.30707. [DOI] [PubMed] [Google Scholar]
- 149.Yuan H, van Blitterswijk CA, de Groot K, de Bruijn JD. Cross-species comparison of ectopic bone formation in biphasic calcium phosphate (BCP) and hydroxyapatite (HA) scaffolds. Tissue Eng. 2006;12:1607–1615. doi: 10.1089/ten.2006.12.1607. [DOI] [PubMed] [Google Scholar]
- 150.Kurashina K, Kurita H, Wu Q, Ohtsuka A, Kobayashi H. Ectopic osteogenesis with biphasic ceramics of hydroxyapatite and tricalcium phosphate in rabbits. Biomaterials. 2002;23:407–412. doi: 10.1016/s0142-9612(01)00119-3. [DOI] [PubMed] [Google Scholar]
- 151.Habibovic P, Juhl MV, Clyens S, Martinetti R, Dolcini L, Theilgaard N, van Blitterswijk CA. Comparison of two carbonated apatite ceramics in vivo. Acta Biomater. 2010;6:2219–2226. doi: 10.1016/j.actbio.2009.11.028. [DOI] [PubMed] [Google Scholar]
- 152.Klein C, de Groot K, Chen W, Li Y, Zhang X. Osseous substance formation induced in porous calcium phosphate ceramics in soft tissues. Biomaterials. 1994;15:31–34. doi: 10.1016/0142-9612(94)90193-7. [DOI] [PubMed] [Google Scholar]
- 153.Muschler GF, Nakamoto C, Griffith LG. Engineering principles of clinical cell-based tissue engineering. J Bone Joint Surg Am. 2004;86:1541–1558. doi: 10.2106/00004623-200407000-00029. [DOI] [PubMed] [Google Scholar]
- 154.Yuan H, Fernandes H, Habibovic P, de Boer J, Barradas AM, de Ruiter A, Walsh WR, van Blitterswijk CA, de Bruijn JD. Osteoinductive ceramics as a synthetic alternative to autologous bone grafting. Proc Natl Acad Sci. 2010;107:13614–13619. doi: 10.1073/pnas.1003600107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 155.Xynos ID, Hukkanen MVJ, Buttery LDK, Hench LL, Polak JM. Bioglasss 45S5 stimulates osteoblast turnover and enhances bone formation in vitro: implications and applications for bone tissue engineering. Calcif Tissue Int. 2000;67:321–329. doi: 10.1007/s002230001134. [DOI] [PubMed] [Google Scholar]
- 156.Yuan H, de Bruijn JD, Zhang X, van Blitterswijk CA, de Groot K. Bone induction by porous glass ceramic made from bioglass (45s5) J Biomed Mater Res. 2001;58:270–276. doi: 10.1002/1097-4636(2001)58:3<270::aid-jbm1016>3.0.co;2-2. [DOI] [PubMed] [Google Scholar]
- 157.Banchet V, Jallot E, Michel J, Wortham L, Laurent-Maquin D, Balossier G. X-ray microanalysis in STEM of short-term physico-chemical reactions at bioactive glass particle/biological fluid interface. Determination of O/Si atomic ratios. Surf Interface Anal. 2004;36:658–665. [Google Scholar]
- 158.Rokkanen PU. Bioabsorbable fixation devices in orthopaedics and traumatology. Ann Chir Gyn. 1998;87:13–20. [PubMed] [Google Scholar]
- 159.Mazzocchi M, Bellosi A. On the possibility of silicon nitride as a ceramic for structural orthopaedic implants. Part I: Processing, microstructure, mechanical properties, cytotoxicity. J Mater Sci Mater Med. 2008;19:2881–2887. doi: 10.1007/s10856-008-3417-2. [DOI] [PubMed] [Google Scholar]
- 160.Klemm H. Silicon nitride for high-temperature applications. J Am Ceram Soc. 2010;93:1501–1522. [Google Scholar]
- 161.Rahaman MN, Yao A, Bal BS, Garino JP, Ries MD. Ceramics for prosthetic hip and knee joint replacement. J Am Ceram Soc. 2007;90:1965–1988. [Google Scholar]
- 162.Krstic Z, Krstic VD. Young’s modulus, density and phase composition of pressureless sintered self-sealed Si3N4/BN laminated structures. J Eur Ceram Soc. 2008;28:1723–1730. [Google Scholar]
- 163.Neuss S, Jochen S, Rainer T, Ruth K. Cytocompatibility of high strength non-oxide ceramics. J Biomed Mater Res Part A. 2010;93A:67–76. doi: 10.1002/jbm.a.32527. [DOI] [PubMed] [Google Scholar]
- 164.Puckett S, Ercan B, Webster TJ, Rahaman M. Decreased bacteria activity on Si3N4 surfaces compared with PEEK or titanium. Inter J nanomed. 2012;7:4829–4840. doi: 10.2147/IJN.S35190. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 165.Rezwan K, Chen QZ, Blaker JJ, Boccaccini AR. Biodegradable and bioactive porous polymer/inorganic composite scaffolds for bone tissue engineering. Biomaterials. 2006;27:3413–3431. doi: 10.1016/j.biomaterials.2006.01.039. [DOI] [PubMed] [Google Scholar]
- 166.Hasegawa S, Neo M, Tamura J, Fujibayashi S, Takemoto M, Shikinami Y, Okazaki K, Nakamura T. In vivo evaluation of a porous hydroxyapatite/poly-DLlactide composite for bone tissue engineering. J Biomed Mater Res A. 2007;81:930–938. doi: 10.1002/jbm.a.31109. [DOI] [PubMed] [Google Scholar]
- 167.Winter GD, Simpson BJ. Heterotopic bone formed in a synthetic sponge in the skin of young pigs. Nature. 1969;223:88–90. doi: 10.1038/223088a0. [DOI] [PubMed] [Google Scholar]
- 168.De Groot K. Letter: Some considerations about bone-induction. Calcif Tissue Res. 1973;13:335–337. doi: 10.1007/BF02015425. [DOI] [PubMed] [Google Scholar]
- 169.Bleach NC, Nazhat SN, Tanner KE, Kelloma M. Effect of filler content on mechanical and dynamic mechanical properties of particulate biphasic calcium phosphate-polylactide composites. Biomaterials. 2002;23:1579–1585. doi: 10.1016/s0142-9612(01)00283-6. [DOI] [PubMed] [Google Scholar]
- 170.Dorozhkin SV. Biocomposites and hybrid biomaterials based on calcium orthophosphates. Biomater. 2011;1:3–56. doi: 10.4161/biom.1.1.16782. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 171.Zhu Y, Gao C, Guan J, Shen J. Promoting the cytocompatibility of polyurethane scaffolds via surface photo-grafting polymerization of acrylamide. J Mater Sci Mater Med. 2004;15:283–289. doi: 10.1023/b:jmsm.0000015489.27261.f0. [DOI] [PubMed] [Google Scholar]
- 172.Zhu Y, Gao C, He T, Shen J. Endothelium regeneration on luminal surface of polyurethane vascular scaffold modified with diamine and covalently grafted with gelatin. Biomaterials. 2004;25:423–430. doi: 10.1016/s0142-9612(03)00549-0. [DOI] [PubMed] [Google Scholar]
- 173.Zhu Y, Sun Y. The influence of polyelectrolyte charges of polyurethane membrane surface on the growth of human endothelial cells. Colloids Surf B. 2004;36:49–55. doi: 10.1016/j.colsurfb.2004.04.008. [DOI] [PubMed] [Google Scholar]
- 174.Jayasuriya AC, Bhat A. Fabrication and characterization of novel hybrid organic/inorganic microparticles to apply in bone regeneration. J Biomed Mater Res Part A. 2010;93:1280–1288. doi: 10.1002/jbm.a.32623. [DOI] [PubMed] [Google Scholar]
- 175.Jayasuriya AC, Bhat A. Mesenchymal stem cell function on hybrid organic/inorganic microparticles in vitro. J Tissue Eng Regen Med. 2010;4:340–348. doi: 10.1002/term.244. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 176.Bhat A, Dreifke M, Gomez C, Kandimalla Y, Ebraheim NA, Jayasuriya AC. Evaluation of cross-linked chitosan microparticles for bone regeneration. J Tissue Eng Regen Med. 2010;4:532–542. doi: 10.1002/term.270. [DOI] [PubMed] [Google Scholar]
- 177.Kyomoto M, Moro T, Iwasaki Y, Miyaji F, Kawaguchi H, Takatori Y, et al. Superlubricious surface mimicking articular cartilage by grafting poly(2- methacryloyloxyethyl phosphorylcholine) on orthopaedic metal bearings. J Biomed Mater Res A. 2009;91A:730–741. doi: 10.1002/jbm.a.32280. [DOI] [PubMed] [Google Scholar]
- 178.Fu T, Zhao J-L, Xu K-W. The designable elastic modulus of 3-D fabric reinforced biocomposites. Mater Lett. 2007;61:330–333. [Google Scholar]
- 179.Zhou H, Lawrence JG, Bhaduri SB. Fabrication aspects of PLA-CaP/PLGA=CaP composites for orthopedic applications: A review. Acta Biomat. 2012;8:1999–2016. doi: 10.1016/j.actbio.2012.01.031. [DOI] [PubMed] [Google Scholar]
- 180.Sen N. Composites: use in saucepan handles, artificial limbs and the AGNI missile. Curr Sci. 2004;86:372–375. [Google Scholar]
- 181.Cifuentes SC, Frutos E, González-Carrasco JL, Muñoz M, Multigner M, Chao J, Benavente R, Lieblich M. Novel PLLA/magnesium composite for orthopedic applications: A proof of concept. Mat Let. 2012;74:239–242. [Google Scholar]
- 182.Müller WD, Nascimento ML, Zeddies M, Córsico M, Gassa LM, Lorenzo de Mele MAF. Mater Res. 2007;10:5–10. [Google Scholar]
- 183.Lacroix D. In: Bone repair biomaterials. Planell JA, Best SM, Lacroix D, Merolli A, editors. Cambridge, UK: Woodhead Publishing Limited; 2009. pp. 106–118. [Google Scholar]
- 184.Yuan H, Van Den Doel M, Li S, Van Blitterswijk CA, De Groot K, De Bruijn JD. A comparison of the osteoinductive potential of two calcium phosphate ceramics implanted intramuscularly in goats. J Mater Sci Mater Med. 2002;13:1271–1275. doi: 10.1023/a:1021191432366. [DOI] [PubMed] [Google Scholar]
- 185.Reddi AH. Cell biology and biochemistry of endochondral bone development. Coll Relat Res. 1981;1:209–226. doi: 10.1016/s0174-173x(81)80021-0. [DOI] [PubMed] [Google Scholar]
- 186.Ohgushi H, Dohi Y, Tamai S, Tabata S. Osteogenic differentiation of marrow stromal stem cells in porous hydroxyapatite ceramics. J Biomed Mater Res. 1993;27:1401–1407. doi: 10.1002/jbm.820271107. [DOI] [PubMed] [Google Scholar]
- 187.Reddi AH. Regulation of cartilage and bone differentiation by bone morphogenetic proteins. Curr Opin Cell Biol. 1992;4:850–855. doi: 10.1016/0955-0674(92)90110-x. [DOI] [PubMed] [Google Scholar]
- 188.Habibovic P, Yuan H, van der Valk CM, Meijer G, van Blitterswijk CA, de Groot K. 3D microenvironment as essential element for osteoinduction by biomaterials. Biomaterials. 2005;26:3565–3575. doi: 10.1016/j.biomaterials.2004.09.056. [DOI] [PubMed] [Google Scholar]
- 189.Habibovic P, Yuan H, van den Doel M, Sees TM, van Blitterswijk CA, de Groot K. Relevance of osteoinductive biomaterials in critical-sized orthotopic defect. J Orthop Res. 2006;24:867–876. doi: 10.1002/jor.20115. [DOI] [PubMed] [Google Scholar]
- 190.Ripamonti U, Klar RM, Renton LF, Ferretti C. Synergistic induction of bone formation by Hop-1, HTGbeta3 and inhibition by zoledronate in macroporous coralderived hydroxyapatites. Biomaterials. 2010;31:6400–6410. doi: 10.1016/j.biomaterials.2010.04.037. [DOI] [PubMed] [Google Scholar]
- 191.Kato M, Toyoda H, Namikawa T, Hoshino M, Terai H, Miyamoto S, Takaoka K. Optimized use of a biodegradable polymer as a carrier material for the local delivery of recombinant human bone morphogenetic protein-2 (RHBMP-2) Biomaterials. 2006;27:2035–2041. doi: 10.1016/j.biomaterials.2005.10.007. [DOI] [PubMed] [Google Scholar]
- 192.Habibovic P, Kruyt MC, Juhl MV, Clyens S, Martinetti R, Dolcini L, Theilgaard N, van Blitterswijk CA. Comparative in vivo study of six hydroxyapatite-based bone graft substitutes. J Orthop Res. 2008;26:1363–1370. doi: 10.1002/jor.20648. [DOI] [PubMed] [Google Scholar]
- 193.Le Nihouannen D, Saffarzadeh A, Gauthier O, Moreau F, Pilet P, Spaethe R, Layrolle P, Daculsi G. Bone tissue formation in sheep muscles induced by a biphasic calcium phosphate ceramic and fi brin glue composite. J Mater Sci Mater Med. 2008;19:667–675. doi: 10.1007/s10856-007-3206-3. [DOI] [PubMed] [Google Scholar]
- 194.Yuan H, de Bruijn JD, Zhang X, van Blitterswijk CA, de Groot K. Use of an osteoinductive biomaterial as a bone morphogenetic protein carrier. J Mater Sci Mater Med. 2001;12:761–766. doi: 10.1023/a:1013957431372. [DOI] [PubMed] [Google Scholar]
- 195.Dellinger JG, Eurell JAC, Jamison RD. Bone response to 3D periodic hydroxyapatite scaffolds with and without tailored microporosity to deliver bone morphogenetic protein 2. J Biomed Mater Res. 2006;76A:366–376. doi: 10.1002/jbm.a.30523. [DOI] [PubMed] [Google Scholar]
- 196.Lan Levengood SK. The effect of BMP-2 on micro- and macroscale osteointegration of biphasic calcium phosphate scaffolds with multiscale porosity. Acta Biomater. 2010;6:3283–3291. doi: 10.1016/j.actbio.2010.02.026. [DOI] [PubMed] [Google Scholar]
- 197.Polak SJ, Levengood SKL, Wheeler MB, Maki AJ, Clark SG, Johnson AJW. Analysis of the roles of microporosity and BMP-2 on multiple measures of bone regeneration and healing in calcium phosphate scaffolds. Acta Biomater. 2011;7:1760–1771. doi: 10.1016/j.actbio.2010.12.030. [DOI] [PubMed] [Google Scholar]
- 198.Woodard JR. The mechanical properties and osteoconductivity of hydroxyapatite bone scaffolds with multi-scale porosity. Biomaterials. 2007;28:45–54. doi: 10.1016/j.biomaterials.2006.08.021. [DOI] [PubMed] [Google Scholar]
- 199.Jayasuriya AC, Ebraheim NA. Evaluation of Bone Matrix and Demineralized Bone Matrix Incorporated PLGA Matrices for Bone Repair. J Mater Sci Mater Med. 2009;20:1637–1644. doi: 10.1007/s10856-009-3738-9. [DOI] [PubMed] [Google Scholar]
- 200.Pearse AJ, Hooper GJ, Roothwell A, Frampton C. Survival and functional outcome after revision of a unicompartmental to a total knee replacement: the New Zealand National Joint Registry. J Bone Joint Surg Br. 2010;92:508–512. doi: 10.1302/0301-620X.92B4.22659. [DOI] [PubMed] [Google Scholar]