Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2017 May 15.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2016 Jan 2;1021:91–100. doi: 10.1016/j.jchromb.2015.12.055

Use of Protein G Microcolumns in Chromatographic Immunoassays: A Comparison of Competitive Binding Formats

Erika L Pfaunmiller 1, Jeanethe A Anguizola 1, Mitchell L Milanuk 1, NaTasha Carter 1, David S Hage 1,*
PMCID: PMC4862902  NIHMSID: NIHMS750081  PMID: 26777776

Abstract

Affinity microcolumns containing protein G were used as general platforms for creating chromatographic-based competitive binding immunoassays. Human serum albumin (HSA) was used as a model target for this work and HSA tagged with a near infrared fluorescent dye was utilized as the label. The protein G microcolumns were evaluated for use in several assay formats, including both solution-based and column-based competitive binding immunoassays and simultaneous or sequential injection formats. All of these methods were characterized by using the same amounts of labeled HSA and anti-HSA antibodies per sample, as chosen for the analysis of a protein target in the low-to-mid ng/mL range. The results were used to compare these formats in terms of their response, precision, limits of detection, and analysis time. All these methods gave detection limits in the range of 8–19 ng/mL and precisions ranging from ± 5% to ± 10% when using an injection flow rate of 0.10 mL/min. The column-based sequential injection immunoassay provided the best limit of detection and the greatest change in response at low target concentrations, while the solution-based simultaneous injection method had the broadest linear and dynamic ranges. These results provided valuable guidelines that can be employed to develop and extend the use of protein G microcolumns and these competitive binding formats to other protein biomarkers or biological agents of clinical or pharmaceutical interest.

Keywords: Chromatographic immunoassay, Affinity microcolumn, Protein G, Competitive binding immunoassay, Human serum albumin

1. Introduction

Antibodies and related agents, such as Fab fragments, have been used for many years for the selective and sensitive detection of analytes in a variety of samples [14]. Over the last few decades there has been ongoing interest in combining these selective binding agents with chromatographic systems, giving a method known as a chromatographic immunoassay or flow-injection immunoanalysis [511]. Chromatographic immunoassays have been explored in many previous studies for the analysis of proteins, drugs, hormones, and other targets [5,6,811]. These assays have been utilized in numerous formats and have been combined with detection methods that have included absorbance, fluorescence, chemiluminescence and electrochemical detection, as well as enzyme-based detection and mass spectrometry [611]. Chromatographic immunoassays can be fast and relatively easy to automate when they are used as part of an HPLC system [5,6,8], and these methods can be employed with samples such as serum, plasma, urine, cerebrospinal fluid, and food [8,10,11].

Many chromatographic immunoassays involve the covalent immobilization of an antibody or a related binding agent to a support that is suitable for use in a column [5,811,12]. However, a chromatographic immunoassay can also be carried out by adsorbing antibodies to a support that contains protein G as a secondary binding agent [6,911]. Protein G is a bacterial cell wall protein produced by groups C and G streptococci and that has the ability to bind at the Fc region of many immunoglobulins [1315]. This results in adsorbed antibodies that have a good orientation and high activity for binding to their target compounds [9,1416]. The ability of protein G to bind many types of immunoglobulins also makes this agent of interest in creating general platforms that can be used with a wide range of antibodies, targets and samples [6,911]. For instance, protein G has already been used in a number of chromatographic immunoassays to measure analytes in complex samples such as serum, food, and cell lysates [1723]. In addition, antibodies that are adsorbed to protein G can later be released under relatively mild acidic conditions, and the protein G can be regenerated by placing it back into a neutral buffer [9,14]. This last feature makes protein G columns valuable in assays where good run-to-run reproducibility is required and/or frequent replacement of the adsorbed antibodies is desired [6,9,10,14].

Another area of interest in chromatographic immunoassays is the use of miniaturized columns, capillaries or microchips to carry out such methods [1721,2432]. This type of work has included the utilization of protein G, or the related agent protein A, in such techniques [17,1923,3237]. One approach for creating small-scale chromatographic immunoassays is to use affinity microcolumns, which are columns that contain biologically-related binding agents (e.g., immobilized antibodies) and which have volumes in the low microliter range [29,38]. Advantages to using affinity microcolumns in chromatographic immunoassays include the small amounts of antibodies or binding ligands that are required, the low non-specific binding of these columns for sample components, and their short residence times, with the latter often allowing analysis times of only a few minutes [29,38].

This study examined the development of affinity microcolumns containing protein G and the use of these microcolumns in chromatographic immunoassays based on several competitive binding formats (see Figs. 13). Human serum albumin (HSA) was used as a model protein and target for this work and a near infrared fluorescent tag was used as the label [39,40]. Competitive binding formats that were considered included both solution- and column-based simultaneous injection assays and a column-based sequential immunoassay format [8,10]. These formats were used under similar chromatographic conditions and compared in terms of their response, precision, limits of detection, and total assay time. The results were used to provide general guidelines regarding the behavior of these methods, as well as information on the relative advantages and disadvantages of each approach. This information should be useful in selecting assay formats in future work with protein G microcolumns and in extending this approach to the analysis of other protein biomarkers, biological agents, or target compounds.

Figure 1.

Figure 1

Scheme for the use of a protein G microcolumn in a solution-based simultaneous injection immunoassay.

Figure 3.

Figure 3

Scheme for the use of a protein G microcolumn in a column-based sequential injection immunoassay

2. Experimental

2.1. Reagents

Nucleosil Si-1000 (1000 Å pore size, 7 μm particle size) was purchased from Macherey-Nagel (Bethlehem, PA, USA). The HSA (Cohn fraction V, essentially fatty acid free, ≥ 96% pure), sodium cyanoborohydride (94%, a mild reducing agent), sodium borohydride (98%, a strong reducing agent), periodic acid (> 99%, an oxidizing agent), Tween-20 (polyethylene glycol sorbitan monolaurate), and rabbit immunoglobulin G (rabbit IgG, > 95%) were from Sigma-Aldrich (St. Louis, MO, USA). IRDye 800CW N-hydroxysuccinimide (NHS) ester was obtained from LI-COR (Lincoln, NE, USA). Polyclonal anti-HSA antibodies (rabbit) were obtained from Rockland Immunochemicals (Gilbertsville, PA, USA). Reagents for the bicinchoninic acid (BCA) protein assay and the protein G (recombinant, albumin binding domains removed) were from Pierce (Rockford, IL, USA). All buffers and aqueous solutions were prepared using water from an EMD MILLI-Q system and were filtered using 0.2 μm GNWP nylon filters, both of which were from Millipore (Billerica, MA, USA).

2.2. Apparatus

A Jasco 2000 HPLC system (Easton, MD, USA) was used in these experiments. This system consisted of three DG-2080–53 solvent degassers, two PU-2080 isocratic pumps, an AS-2057 autosampler equipped with a 100 μL sample loop (operated in the partial loop injection mode), a UV-2075 absorbance detector, and a FP-2020 fluorescence detector. This system also included a custom-built near infrared fluorescence detector [26,39] that was provided by LI-COR. The temperature of the columns was controlled by using an on-line column heater. A Rheodyne Advantage PF six-port switching valve (Cotati, CA, USA) was used for alternating passage of the analyte and buffer solutions through the columns during the frontal analysis studies. The system components were controlled by a Jasco LC-Net II/ADC system and ChromNAV software v1.18.03 (Easton, MD, USA). Chromatographic data were collected using ChromNAV and processed using PeakFit 4.12 (SeaSolve Software, San Jose, CA, USA). Purification of the labeled HSA was performed using Zeba spin columns (7 kDa MW cutoff, 0.7 to 4 mL sample capacity) from Pierce, along with a 5702RH temperature-controlled centrifuge from Eppendorf (New York, NY, USA) and a fixed-angle centrifuge rotor from VWR (West Chester, PA, USA). The microcolumns were packed using an HPLC slurry packing system from ChromTech (Apple Valley, MN, USA).

2.3. Preparation of protein G microcolumns

Nucleosil Si-1000 silica was converted into a diol-bonded form, with this material then being oxidized with periodic acid and placed into an aldehyde-activated form for use in the Schiff base immobilization method, as described previously [5,12]. Prior to its use in activation or immobilization, the silica was placed into the appropriate reaction buffer and sonicated under vacuum for 5 min. Approximately 3 mg of protein G per 100 mg of aldehyde-activated silica was placed into an immobilization buffer that contained pH 6.0, 0.10 M potassium phosphate and 100 mg/mL sodium cyanoborohydride. This protein/silica slurry was then shaken at 4°C for 3 days. After completion of the immobilization reaction, the protein G support was washed using pH 8.0, 0.10 M potassium phosphate buffer. A 2 mg/mL solution of sodium borohydride in pH 8.0, 0.10 M potassium phosphate buffer was then added to this support, and the mixture was shaken at room temperature for 90 min. The final support was washed with pH 7.4, 0.067 M potassium phosphate buffer and stored in this buffer at 4°C until use. A control support was prepared in the same manner, but with no protein G being added during the immobilization step.

The amount of immobilized protein on each support was determined in triplicate by a BCA protein assay [41], using protein G as the standard and the control support as the blank. The protein G silica and control support were downward slurry packed at 4000 psi (28 MPa) into separate 2.1 mm i.d. × 5 mm stainless steel columns using pH 7.4, 0.067 M potassium phosphate buffer as the packing solution. These columns were stored in the same buffer at 4°C when not in use and were stable for up to one year, with some columns being used over approximately 200 application and elution cycles.

2.4. Preparation of labeled HSA

A 0.10 mg portion of IRDye 800CW NHS ester was dissolved in 20 μL water and a 10 μL aliquot of this solution was combined with 1 mL of a 1 mg/mL HSA solution in pH 8.5, 0.10 M potassium phosphate buffer. This solution was mixed and allowed to shake for 2 h in the dark at room temperature. A Zeba spin column was utilized with pH 7.4, 0.067 M potassium phosphate buffer to remove any unreacted dye from the labeled HSA. Absorbance measurements were made at 780 and 280 nm to determine the dye/protein ratio and protein concentration of the final labeled HSA solution. The labeled HSA solution had a final concentration of 0.8–0.9 mg HSA/mL and a dye/protein ratio of 1.0–2.0, as determined over five batches of this conjugate. The labeled HSA solutions were stored at 4°C in pH 7.4, 0.067 M potassium phosphate buffer when not in use. The labeled HSA was stable for up to 2 weeks when protected from light and stored under these conditions. The near infrared fluorescent labeled HSA was detected in the chromatographic systems by using an excitation wavelength of 774 nm and an emission wavelength of 789 nm.

2.5. Evaluation of protein G microcolumns

Rabbit IgG was used in frontal analysis experiments to determine the active amount of protein G that was in the affinity microcolumns [9,1315,38]. In these experiments, pH 7.4, 0.067 M potassium phosphate buffer was applied to a protein G microcolumn at 0.50 mL/min for 1 min, followed by application of 0.05 mg/mL rabbit IgG in the same buffer at 0.10 mL/min. The elution of the rabbit IgG was monitored at 280 nm. After a breakthrough curve had been obtained, a pH 2.5, 0.067 M potassium phosphate buffer was passed through the protein G microcolumn at 0.50 mL/min for 10 min to elute the retained rabbit IgG [42]. This cycle was repeated as needed. The mean location of each breakthrough curve was determined by using a Savitzky-Golay first derivative algorithm for smoothing, followed by fitting of the first derivative to an exponentially-modified Gaussian curve and determination of the central moment. The same approach was employed with a control column to correct for the void time of the system and non-specific binding of rabbit IgG to the support or system components. The level of non-specific binding was negligible for rabbit IgG, as has been noted for the same analyte on comparable protein G columns [15,42].

The binding of the protein G microcolumns to small amounts of antibodies was evaluated by using samples containing various concentrations of anti-HSA antibodies (e.g., 1.67 μg/mL) that were mixed in a 15:1 (v/v) ratio with a fixed amount of labeled HSA (initial concentration, 1200 ng/mL; concentration in mixture, 80 ng/mL) and incubated for 30 min prior to injection. A series of 50 μL injections for each mixture were then made onto a protein G microcolumn at 0.10 mL/min and in the presence of a pH 7.4, 0.067 M potassium phosphate buffer containing 0.01% Tween 20. The relative capture efficiency was determined by comparing the areas of these peaks to the total areas that were obtained for the same samples on a control column.

2.5. Chromatographic immunoassays

The final application buffer for each assay format used in this study was pH 7.4, 0.067 M potassium phosphate buffer that contained 0.01% Tween 20. The flow rate used for sample application was 0.10 mL/min. The elution buffer consisted of pH 2.5, 0.067 M potassium phosphate buffer, and the elution flow rate was 0.50 mL/min. The elution step for each assay was 10 min in length, after which the application buffer was reapplied for 20 min to regenerate the column prior to the next sample analysis [42]. The near infrared fluorescent label was detected by using an excitation wavelength of 774 nm and an emission wavelength of 789 nm. All chromatographic studies were conducted at room temperature.

In the solution-based simultaneous injection immunoassay, the sample was mixed and incubated with a fixed amount of antibodies and a labeled analog of the target (i.e., labeled HSA, or the “label”) prior to injection onto a protein G microcolumn. The general format that was used for this assay is shown in Fig. 1. The concentration of the anti-HSA antibodies prior to mixing was 5.0 μg/mL, the concentration of the labeled HSA was 240 ng/mL, and the samples contained 0–400 ng/mL HSA. These reagents were combined in a 1:1:1 volume ratio, using 0.5 mL of each reagent. This mixture was allowed to incubate at room temperature for at least 30 min. A series of 50 μL injections of this mixture were then made onto the protein G microcolumn, with an elution step being conducted after every three injections.

In the column-based simultaneous injection immunoassay, only the sample and labeled analog of the target (i.e., labeled HSA) were pre-mixed prior to injection, as illustrated in Fig. 2. In this format, the antibodies were applied to the protein G microcolumn before the sample/labeled analog mixture was injected. In this study, a 50 μL injection containing 1.67 μg/mL of anti-HSA antibodies was first made onto a protein G microcolumn at 0.10 mL/min. A 0.75 mL portion of 160 ng/mL labeled HSA was then combined with a 0.75 mL sample containing 0–400 ng/mL HSA and 50 μL injections of this mixture were applied to the protein G column that contained the adsorbed anti-HSA antibodies. An elution and regeneration step was carried out either after each sample injection or after three replicate injections of a sample had been applied to the column.

Figure 2.

Figure 2

Scheme for the use of a protein G microcolumn in a column-based simultaneous injection immunoassay

In the column-based sequential injection format, the antibodies, sample, and labeled analog were injected separately and in series, as shown in Fig. 3. For each analysis, a 50 μL injection of 1.67 μg/mL anti-HSA antibodies was made at 0.10 mL/min, followed at the same flow rate by a 50 μL injection of a sample containing 0–400 ng/mL HSA and a 50 μL injection containing 80 ng/mL of the labelled HSA. This sequence was followed by an elution and regeneration step and was repeated in triplicate for each sample.

3. Results and Discussion

3.1. Preparation and characterization of protein G support

The protein G microcolumns were prepared by first converting HPLC-grade 7 μm diameter silica into a diol-bonded form. This type of support was employed because of its good mechanical stability and efficiency, the ease with which this material can be modified for protein attachment, and the low non-specific binding of diol-bonded silica for many biological compounds [9,15,29]. The Schiff base immobilization method was used with this material because it has been shown to provide higher activities for protein G and similar ligands (e.g., protein A) when compared to other common amine-based coupling methods [12,14,15,33]. Silica with a pore size of 1000 Å was employed to allow sufficient room for the immobilization of protein G, the binding of this ligand to antibodies, and the interaction of these antibodies with HSA. However, supports with smaller pore sizes (e.g., 50–500 Å) can allow up to two-fold more IgG-class antibodies to be bound to immobilized protein G per gram of the support [15].

The protein G supports that were made in this report contained 6.1 (± 0.4) mg protein G/g silica. Based on a molecular weight for protein G of roughly 32 kDa [14], this content corresponded to 0.19 (± 0.01) μmol protein G/g silica and a surface coverage of 0.24 (± 0.02) mg protein G/m2 silica, or 0.011 (± 0.005) μmol protein G/m2 silica. These results were in same general range as has been observed previously for comparable materials (e.g., a maximum protein G content of 20.6 (± 1.1) mg/g silica, as noted in Ref. [15]).

Frontal analysis using rabbit IgG as the applied target indicated that the protein G columns had an overall binding capacity of 8.7 (± 1.0) mg IgG/g silica. This corresponded to a total binding capacity of approximately 0.067 (± 0.020) mg IgG for a 2.1 mm i.d. × 5 mm protein G microcolumn. The specific activity of these microcolumns (i.e., the relative amount of active protein G) was 31%. This result agreed with previous work utilizing similar protein G supports, which have had specific activities in the range of 30–47% [15,42]. The stability of the protein G microcolumns was also examined by using a series of frontal analysis experiments over the course of 20–50 application and elution cycles. These data indicated that only a 0.3–0.9% change in the binding capacity occurred from one application/elution cycle to the next.

3.2. Retention and elution properties of protein G microcolumns

The ability of the protein G microcolumns to bind small amounts of antibodies, such as might be used in a competitive binding chromatographic immunoassay, was next examined. This was done by injecting mixtures containing representative concentrations of anti-HSA antibodies that had been incubated with a fixed amount of labeled HSA. Of particular interest was the behavior of samples that contained around 84 ng of anti-HSA antibodies (e.g., 1.67 μg/mL in a 50 μL sample), which was the approximate amount of antibodies that were used per injection in the chromatographic immunoassays that are described later in this report. This amount of antibodies had an average capture efficiency of 92 (± 5)% on the protein G microcolumns at an application flow rate of 0.10 mL/min. No significant increase in this capture efficiency was noted at the 95% confidence level when the flow rate was reduced to 0.01 mL/min. However, the capture efficiency did decrease by 25–30% as the flow rate was increased to 1.0 mL/min.

This last effect, in which a decrease in binding is noted at higher flow rates, has been noted in prior work with small affinity columns containing immobilized protein G or protein A (i.e., a related binding agent for antibodies) [33,34,43,44]. This type of behavior has been previously shown to be due to the presence of slow adsorption kinetics and/or slow mass transfer [33]. These conditions, in turn, can lead to a lower probability for binding by the applied antibodies as a higher flow rate and smaller sample residence time are used to inject samples onto the column [33,34,43,44].

The elution time for the non-retained peak in these studies was also of interest because monitoring the amount of the non-retained target analog (e.g., labeled HSA) is one way of following the response of a competition-based chromatographic immunoassay [10,11]. The non-retained peak for the protein G microcolumns eluted within 3.5 min at 0.10 mL/min (see Fig. 4), with this value increasing to 6.8 min at 0.05 mL/min and decreasing to 0.35 min at 1.0 mL/min. The average back pressure across each protein G microcolumn was less than 100 psi for flow rates up to 1.0 mL/min. At a flow rate of 0.10 mL/min or less, the back pressure was below 50 psi. As a compromise between capture efficiency, assay speed and column back pressure, an application flow rate of 0.10 mL/min was used in the remainder of this report.

Figure 4.

Figure 4

Typical chromatograms and non-retained peaks obtained for labeled HSA in a column-based sequential injection immunoassay at an injection flow rate of 0.10 mL/min. The concentrations that are listed represented the unlabeled HSA in the samples that were injected prior to a fixed amount of the labeled HSA. Both the samples and labeled HSA were injected onto a protein G microcolumn that contained previously adsorbed anti-HSA antibodies. The time scale for these chromatograms is shown at the bottom of the figure. The set of double diagonal lines at the bottom represent the time intervals over which an elution and regeneration step, along with the application of the antibodies and the sample to the protein G microcolumn, were carried out prior to injection of the labeled HSA.

3.3. Use of protein G microcolumns in a solution-based simultaneous injection immunoassay

One format for a chromatographic competitive binding immunoassay that was examined was a solution-based simultaneous injection method. This approach was a modified version of a standard solution-phase competitive binding immunoassay [13]. In this method, a protein G column was used to separate the bound and free fractions of the labeled target analog (or “label”) after the label and target had been allowed to compete in solution for binding to a small amount of antibodies against the target [36].

The general scheme that was used for such a method is illustrated in Fig. 1. The first step involved incubation of the sample, a fixed amount of the label, and a limited amount of antibodies. After being allowed to react in solution (e.g., for 30 min in this study), the mixture was injected onto a protein G microcolumn, to which the antibody-label and antibody-target complexes were allowed to bind. Based on the competition that had already occurred between the target and label for the antibodies, the non-retained fraction for the label was then used to provide an indirect measurement of the amount of target in the original sample. After the peak due to the non-retained label had been measured for one or more injections, an elution buffer was used to release the retained components and the column was regenerated before making another series of injections.

In this report, a solution-based simultaneous injection immunoassay for HSA was carried out by first combining the anti-HSA antibodies with the labeled HSA and a sample containing HSA. A 50 μL injection of this mixture was then made onto a protein G microcolumn. A typical calibration curve for this assay is shown in Fig. 5(a), as obtained with an injection flow rate of 0.10 mL/min. The lower limit of detection was 19 ng/mL (S/N = 3), and the assay results had a precision of less than or equal to ± 10% over the entire calibration range. The linear range of this method extended up to about 225 ng/mL, and the dynamic range went up to around 400 ng/mL. Results based on the non-retained peak were obtained within 3.5–5.0 min of injecting the antibody/label/sample mixture at 0.10 mL/min. This gave an assay time, including the half-hour incubation step, of around 34–35 min.

Figure 5.

Figure 5

Calibration plots for (a) a solution-based simultaneous injection immunoassay, (b) a column-based simultaneous injection immunoassay, and (c) a column-based sequential injection immunoassay for HSA that were each conducted on 2.1 mm i.d. × 5 mm protein G microcolumns using anti-HSA antibodies. The conditions are given in the text. These results are based on the measured height, or amplitude, due to the labeled HSA (or “label”) that was present in the non-retained peak. The error bars represent ± 1 standard error of the mean (n = 3).

It was possible in this format to make multiple injections onto a protein G column prior to elution of the retained components. This feature can be useful in helping to increase the throughput of this method in work with large batches of samples. However, care must be exercised in using this approach on a protein G microcolumn and with a solution-based competitive binding immunoassay because the capture efficiency will decrease and the response based on the non-retained peak will increase slightly when no elution step is used between injections, as is illustrated in Fig. 6. The decrease in capture efficiency occurs because a greater total load of the antibody is being applied to the column with each additional injection [43,44]. This lower capture efficiency, in turn, leads to a greater response based on the non-retained peak because more of the antibody-label complex is now eluting in the non-retained fraction. In this study, up to three replicate injections of a given sample/reagent mixture were made prior to applying the elution buffer; this provided stable results while also minimizing the overall analysis time.

Figure 6.

Figure 6

Effects of antibody concentration on the response of a solution-based simultaneous injection immunoassay using anti-HSA antibodies and a 2.1 mm i.d. × 5 mm protein G microcolumn. These samples contained various concentrations of anti-HSA antibodies (1.67 μg/mL, ∘; 16.7 μg/mL, ▲; 134 μg/mL, ●) that were mixed with a fixed amount (1200 ng/mL) of near infrared fluorescent labeled HSA in a 15:1 (v/v) ratio and incubated for 30 min prior to injection. A series of 50 μL injections were then made for each mixture at 0.10 mL/min, with no elution step being used between injections.

The response of this format is affected by the same factors that determine the behavior of a traditional solution-phase competitive binding immunoassay [13,36]. These factors include the concentrations of the label and antibodies that are combined with the sample, the affinity of these antibodies for the target and the label, and the incubation time that is allowed for these components to react [13]. For instance, Fig. 6 shows how an increase in the amount of antibodies led to greater binding by these antibodies to the label and a lower signal for the non-retained portion of the label. However, the use of more antibodies also increased the cost of the assay and required a greater concentration of the target to produce a change in the response, which lead to a higher limit of detection. For the sake of comparison, the amounts of the label and antibodies that were used in this report were kept the same from one assay format to the next and were chosen to provide a usable assay response in the low-to-mid ng/mL range for the target.

Other parameters that were important in this assay format when using a protein G microcolumn were the relative binding capacity of this column and its capture efficiency for antibodies [10,33,34]. These parameters were initially evaluated in Sections 3.1–3.2 in terms of the overall behavior of the protein G microcolumns. As an example, the measured binding capacity of 0.067 mg IgG for a 2.1 mm i.d. × 5 mm protein G microcolumn was about 800-fold greater than the 84 ng of antibodies that were applied during one injection in the solution-based simultaneous injection immunoassay. It was again noted that the capture efficiency of 92% for such a sample at 0.10 mL/min was affected by a change in the application flow rate. For instance, if the elution time for the non-retained peak in the solution-based simultaneous injection immunoassay was reduced by five-fold in going from an injection flow rate of 0.10 to 0.50 mL/min, the higher flow rate also lead to a loss of about 15% in the capture efficiency. This decrease in capture efficiency produced a higher response for standards containing only the labeled HSA and anti-HSA antibodies (i.e., due to less efficient capture of the antibody-label complex), as well as a calibration curve with a smaller slope. The result was a higher limit of detection. The same relative changes in the capture efficiency and response, as well as a shift to higher limits of detection, were also noted at higher flow rates in the other assay formats that were examined in this report.

3.4. Use of protein G microcolumns in a column-based simultaneous injection immunoassay

The second assay format that was evaluated for use with the protein G microcolumns was a column-based simultaneous injection immunoassay [811,22,23,3537]. This method is illustrated in Fig. 2, and had the antibodies being adsorbed to the protein G microcolumn before the label and the target were applied. The label and target were pre-mixed prior to injection and applied to the column simultaneously, with no incubation step being required. The label and target in this format competed for the limited number of binding sites that were present on the adsorbed antibodies within the column. The rest of the steps in this format were the same as those in the solution-based simultaneous injection immunoassay.

A column-based simultaneous injection immunoassay was conducted by using the same amounts of antibodies and labeled HSA as utilized in Section 3.3 for the solution-based method. An example of a calibration plot that was generated with this assay at 0.10 mL/min is given in Fig. 5(b). The lower limit of detection was 13 ng/mL (S/N = 3), and the precision obtained for triplicate injections at each sample concentration was ± 5% over the entire concentration range that was examined. Some curvature in the response was seen above 75–80 ng/mL HSA, and the dynamic range extended to a little over 250 ng/mL. Both this linear region and dynamic range were narrower than noted in Fig. 5(a) for a solution-based simultaneous injection immunoassay that made use of the same reagents. This was likely due to the much shorter time that was allowed for competition and binding of the HSA and labeled HSA with the anti-HSA antibodies in the column-based method. In this column-based format, the incubation time was less than or equal to the microcolumn's void time at 0.10 mL/min, which was only 8–9 s. However, Fig. 5(b) demonstrates that even these conditions resulted in a calibration curve that could be used for the measurement of HSA.

Results were obtained in this method in about 3.5–5.0 min after injection of the label and sample. However, the total assay time up to this point was now only 7–10 min at 0.10 mL/min. This was the time required for two sets of injections, spaced 3.5–5.0 min apart, for the application of the antibodies and labeled HSA/sample mixture. Unlike the solution-based simultaneous injection immunoassay, no pre-incubation step was required in the column-based method; however, an elution and regeneration step was now needed more frequently between sample injections. To obtain the most reproducible results, the elution and regeneration step was done after each sample injection; however, only a 3–5% difference in the results were noted when up to three injections for the same sample were made prior to such a step. The elimination of the pre-incubation step and the manual procedures involved in such a step were important factors in leading to the improved precision of this column-based method over the solution-based approach. The better limit of detection that was obtained in the column-based method was further aided by this improvement in precision, along with the steeper response that was seen in this format at low target concentrations.

There are several parameters which can be varied to adjust the response in a column-based simultaneous injection immunoassay [10,22,23,3537,45,46]. These factors include the relative amounts of the label and antibodies that are used, the binding capacity of the column, the rate of the antibody-target/label interaction, and the injection flow rate [10,45,46]. For instance, as the injection flow rate was increased from 0.10 mL/min to 0.50 mL/min for the application of the label and HSA, the change in the capture efficiency by the anti-HSA antibodies decreased by approximately 10%. Another factor to consider in such an assay when using a protein G microcolumn is the antibody capture efficiency. When the flow rate for adsorption of the anti-HSA antibodies to the protein G microcolumn was increased from 0.10 to 0.50 mL/min, the capture efficiency for these antibodies decreased by 12%.

3.5. Use of protein G microcolumns in a column-based sequential injection immunoassay

The third format evaluated for use with the protein G microcolumns was a sequential injection immunoassay, as shown in Fig. 3. In this method, the sample and each of the reagents were applied to the protein G microcolumn independently. This format has not previously been reported in work with antibodies that have been adsorbed to protein G or protein A supports; however, it has been used with antibodies that have been covalently immobilized to silica or other materials [811,45,4751]. An important advantage of the sequential injection method is that the label and target/sample never come into contact with each other, which minimizes or eliminates any matrix effects the sample may have on the final response due to the label [10,51].

An example of a calibration plot that was generated in this format at an injection flow rate of 0.10 mL/min is given in Fig. 5(c). The lower limit of detection under these conditions was 7.9 ng/mL (S/N = 3), and the results had a typical relative precision of ± 5%. The linear range for this method went up to about 50 ng/mL and the dynamic range extended up to around 150 ng/mL. These features gave this approach the best limit of detection but the smallest linear and dynamic ranges seen in all of the assay formats that were examined. The better detection limit of this method was the combined result of its improved precision over the solution-based simultaneous injection immunoassay and its steeper response at low HSA concentrations versus both the solution- and column-based simultaneous injection immunoassays. The narrower linear and dynamic ranges seen in this format versus those for the solution-based simultaneous injection immunoassay were probably a result of the shorter time allowed in the column-based sequential injection method for competition of the HSA and labeled HSA for the anti-HSA antibodies. The steeper response and smaller linear/dynamic ranges versus the column-based simultaneous injection immunoassays were due to the target being allowed to contact and bind to the adsorbed antibodies before the label had been applied. The same trends have been noted when comparing these latter two methods for covalently immobilized antibodies [45,46].

Results were obtained within about 3.5–5.0 min of injection for the label, and within 10.5–15.0 min of the application of antibodies to the protein G column. These times reflect the fact that each component of this method had to be independently applied to the protein G microcolumn. There was no need to have a pre-incubation period in this method, which helped to minimize the total analysis time per sample; however, elution and column regeneration steps were needed between each sample injection.

3.6. Comparison of assays based on adsorbed antibodies versus immobilized antibodies

A comparison was next made between the responses seen in Figs. 5(b–c) for antibodies adsorbed to protein G microcolumns and in prior work with column-based simultaneous or sequential injection immunoassays that used covalently immobilized anti-HSA antibodies [45,46,51]. One difference in these two groups of methods was in the amount of antibodies that was utilized. This issue in important to consider in that it can significantly affect the overall cost of the assay. With the protein G microcolumns, only 84 ng (or 0.56 pmol) of antibodies were used per sample injection. In the prior work with immobilized antibodies, the equivalent of about 0.41 nmol of active antibodies were present in 2.1 mm i.d. × 6.34 mm columns [45,46,51]. This was roughly a 730-fold difference in the amount of antibodies that was used per injection for the protein G microcolumns versus the columns containing covalently immobilized antibodies. Even though the immobilized antibody columns from the previous studies were each used to examine many samples and standards, sometimes for up to hundreds of injections per column [45,46,51], the net result was still a large decrease in the amount of antibodies that were required in the assays that employed protein G microcolumns.

The response obtained with the adsorbed versus covalently immobilized antibodies was also considered. This was accomplished by comparing the calibration curves in Figs. 5(b–c) with the results that would be expected when using an equivalent amount of immobilized antibodies under the same flow rate conditions and using comparable amounts of the label [45,46]. These conditions included the use of a column that contained 0.56 pmol of immobilized antibodies, an injection flow rate of 0.10 mL/min, and a label amount equal to 0.11 mol label/mol antibody (i.e., the ratio obtained when mixing 4 ng of labeled HSA with 84 ng of anti-HSA antibodies). An association rate constant of 4.0 × 104 M−1 s−1, as measured previously for HSA with immobilized anti-HSA antibodies [45,51], was also used to make this comparison.

It was found in this comparison that the responses seen in Fig. 5(b–c) were much narrower than those expected for covalently immobilized antibodies. During the use of antibodies that were adsorbed to protein G microcolumns, the dynamic ranges that were obtained for the column-based simultaneous and sequential injection formats covered about a 20-fold range in concentration (i.e., in going from the lower limit of detection to the upper end of the dynamic range). In contrast to this, a dynamic range covering roughly a 60- to 250-fold change in concentration (i.e., in going from a 5% to 95% change in the response) would be expected for the same assay formats and under comparable conditions when using covalently immobilized antibodies [45].

Another difference in these two groups of methods was in the limits of detection that were obtained. The antibodies that were adsorbed to the protein G microcolumns gave detection limits that were more than 8- to 10-fold lower than would be expected when using covalently immobilized antibodies [45,46,51]. This difference, as well as the narrower dynamic ranges that were seen with the adsorbed antibodies, is most likely related to the differences in the distribution of the antibodies within the two types of columns. In the prior studies with covalently immobilized antibodies, these antibodies were spread uniformly throughout the length of the column. However, in the protein G microcolumns most of the antibodies were adsorbed in a narrow band at or near the column's entrance. The latter situation would give the label and target a much shorter contact time with the antibodies and may have decreased the apparent amount of antibodies that were encountered by these two competing agents in the column. Such an effect could have produced a sharper increase in the response at low target concentrations. The expected result would be similar to the differences in response that were observed earlier when comparing the calibration curves for the solution- and column-based simultaneous injection assays in Fig. 5.

4. Conclusion

This study examined the use of affinity microcolumns that contained immobilized protein G as general platforms for creating chromatographic-based competitive binding immunoassays for protein biomarkers. HSA was used as a model target for this work and the label was HSA that contained a near infrared fluorescent dye. Protein G microcolumns were made with dimensions of 2.1 mm i.d. × 5 mm that had binding capacities of around 0.067 mg IgG and that could be used at flow rates of 0.10–1.0 mL/min with back pressures of 50 psi or less. In addition, these microcolumns had an average capture efficiency of 92% at 0.10 mL/min for the typical amounts of antibodies that were later used with these columns in various competitive binding assays.

The assay formats that were examined included a solution-based simultaneous injection method, as well as column-based simultaneous injection and sequential injection formats. The same amounts of label and antibodies were used per sample in each technique, as chosen for the analysis of a protein target in the low-to-mid ng/mL range. The analytical properties that were obtained in these assays are summarized in Table 1, with all of these particular methods giving detection limits in the range of 8–19 ng/mL. Although the responses and detection limits that were noted in this study may vary as other types of antibodies, targets and labeling methods are used in these formats, the results in Table 1 do demonstrate several general trends that should apply to all such assays. For instance, the column-based sequential injection immunoassay provided the best limit of detection and the greatest change in response at low target concentrations. The precision of the two column-based methods was slightly better than the precision of the solution-based method (i.e., ± 5% versus ± 10%), but the solution-based method had broader linear and dynamic ranges. The solution-based method required a pre-incubation step for the antibodies with the label and sample, which was not needed in the column-based methods. However, the solution-based method also had the possibility of being used for multiple injections during each analysis cycle. It was further noted that the assays based on antibodies adsorbed to protein G microcolumns gave lower detection limits than would be expected for the same types of assays when using columns with a uniform distribution of covalently immobilized antibodies.

Table 1.

Comparison of the analytical properties of several competitive binding immunoassay formats carried out on protein G microcolumns

Type of Assay Assay Range (ng/mL)a Assay Time (min)b Precision Advantages Disadvantages
Solution-based simultaneous injection immunoassay Linear range: 19 to ~225 Dynamic range: 19 to ~400 34–35 ± 10% Can allow injection of multiple samples per cycle; relatively broad linear & dynamic ranges Requires pre-incubation step; highest limit of detection for these methods
Column-based simultaneous injection immunoassay Linear range: 13 to ~75–80 Dynamic range: 13 to >250 7–10 ± 5% Does not require a pre-incubation step; relatively fast for individual samples Moderate linear and dynamic ranges; intermediate limit of detection
Column-based sequential injection immunoassay Linear range: 7.9 to ~50 Dynamic range: 7.9 to ~150 10.5–15 ± 5% Does not require pre-incubation step; moderately fast for single samples; lowest limit of detection Narrow linear and dynamic range; elution & regeneration required between samples
a

The first value given in each range is the lower limit of detection at S/N = 3.

b

These are the elapsed times at an application flow rate of 0.10 mL/min that span from the initial contact of the sample with the antibodies, or the application of the antibodies to the protein G column, to the time when the non-retained peak for the injection of the labeled target analog is detected and measured. Additional time in each method is required for the elution and regeneration of the protein G column after each series of injections.

These results provide valuable guidelines that can be used in future work to choose between these various formats and in the use of protein G microcolumns for these assays. Various factors that can affect and be controlled in these assays were also identified, which should be useful in the development of new applications for these techniques. This present study builds on prior work that has used immobilized antibodies in affinity microcolumns for other assay formats (e.g., a one-site immunometric assay) [29], and further extends the range of applications that can be assessed by using such columns in chromatographic immunoassays. The use of protein G microcolumns in these and other formats is not limited to HSA but could be extended to other proteins, biomarkers or targets of clinical or pharmaceutical interest. All that is required is the use of an appropriate antibody for the given target and a labeled analog that can compete with this target for its antibodies. It is expected that protein G microcolumns and these methods should be of great future interest with the growing demand for methods and miniaturized devices in* which antibodies and immunoassays can be employed for the rapid analysis of specific target compounds.

Highlights.

  • Protein G microcolumns were developed for use in chromatographic immunoassays.

  • Human serum albumin was used as a model protein target for this work.

  • Several solution- and column-based competitive binding formats were compared.

  • These methods were compared in terms of their response, speed, and precision.

  • The guidelines that were created can be used to extend this approach to other targets.

Acknowledgements

This work was supported, in part, by the National Institutes of Health under grant R01 GM044931 and by the National Science Foundation/EPSCoR program under grant EPS-1004094. Support for the remodeled facilities that were used to perform these experiments was provided under NIH grant RR015468-001.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • [1].Wild D, editor. The Immunoassay Handbook: Theory and Applications of Ligand Binding, ELISA and Related Techniques. 4th ed. Elsevier; Amsterdam: 2013. [Google Scholar]
  • [2].Darwish IA. Immunoassay methods and their applications in pharmaceutical analysis: basic methodology and recent advances. Int. J. Biomed. Sci. 2006;2:217–235. [PMC free article] [PubMed] [Google Scholar]
  • [3].Hage DS. Immunoassays. Anal. Chem. 1999;71:294R–304R. doi: 10.1021/a1999901+. [DOI] [PubMed] [Google Scholar]
  • [4].Butler JE, editor. Immunochemistry of Solid-Phase Immunoassay. CRC Press; Boca Raton: 1991. [Google Scholar]
  • [5].Phillips TM. High performance immunoaffinity chromatography. An introduction, LC Mag. 3(1985):962–972. [Google Scholar]
  • [6].de Frutos M, Regnier FE. Tandem chromatographic-immunological analyses. Anal. Chem. 1993;65:17A–25A. doi: 10.1021/ac00049a001. [DOI] [PubMed] [Google Scholar]
  • [7].Weller MG. Immunochromatographic techniques – a critical review. Fres. J. Anal. Chem. 2000;366:635–645. doi: 10.1007/s002160051558. [DOI] [PubMed] [Google Scholar]
  • [8].Hage DS, Nelson MA. Chromatographic immunoassays. Anal. Chem. 2001;73:198A–205A. [PubMed] [Google Scholar]
  • [9].Hage DS, Phillips TM. Immunoaffinity chromatography. In: Hage DS, editor. Handbook of Affinity Chromatography. 2nd ed. Chapter 6. Taylor & Francis; New York: 2006. [Google Scholar]
  • [10].Moser A, Hage DS. Chromatographic immunoassays. In: Hage DS, editor. Handbook of Affinity Chromatography. 2nd ed. Chapter 29. Taylor & Francis; New York: 2006. [Google Scholar]
  • [11].Moser AC, Hage DS. Immunoaffinity chromatography: an introduction to applications and recent development. Bioanalysis. 2010;4:769–790. doi: 10.4155/bio.10.31. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Hermanson GT, Mallia AK, Smith PK. Immobilized Affinity Ligand Techniques. Academic Press; Boca Raton: 1992. [Google Scholar]
  • [13].Bjorck L, Kronvall G. Purification and some properties of streptococcal protein G, a novel IgG-binding reagent. J. Immunol. 1984;133:969–974. [PubMed] [Google Scholar]
  • [14].Hage DS, Bian M, Burks R, Karle E, Ohnmacht C, Wa C. Bioaffinity chromatography. In: Hage DS, editor. Handbook of Affinity Chromatography. 2nd ed. Chapter 5. Taylor & Francis; New York: 2006. [Google Scholar]
  • [15].Jackson AJ, Karle EM, Hage DS. Preparation of high-capacity supports containing protein G immobilized to porous silica. Anal. Biochem. 2010;406:235–237. doi: 10.1016/j.ab.2010.07.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Turkova J. Oriented immobilization of biologically active proteins as a tool for revealing protein interactions and function. J. Chromatogr. B. 1999;722:11–31. doi: 10.1016/s0378-4347(98)00434-4. [DOI] [PubMed] [Google Scholar]
  • [17].Cole LJ, Kennedy RT. Selective preconcentration for capillary zone electrophoresis using protein G immunoaffinity capillary chromatography. Electrophoresis. 1995;16:549–556. doi: 10.1002/elps.1150160190. [DOI] [PubMed] [Google Scholar]
  • [18].Shen H, Aspinwall CA, Kennedy RT. Dual microcolumn immunoassay applied to determination of insulin secretion from single islets of Langerhans and insulin in serum. J. Chromatogr. B. 1997;689:295–303. doi: 10.1016/s0378-4347(96)00336-2. [DOI] [PubMed] [Google Scholar]
  • [19].de Frutos M, Paliwal SK, Regnier FE. Liquid chromatography based enzyme-amplified immunological assays in fused-silica capillaries at the zeptomole level. Anal. Chem. 1993;65:2159–2163. doi: 10.1021/ac00063a040. [DOI] [PubMed] [Google Scholar]
  • [20].Vlakh EG, Platonova GA, Tennikova TB. Affinity chromatography of proteins on monolithic columns. Meth. Mol. Biol. 2014;1129:303–324. doi: 10.1007/978-1-62703-977-2_23. [DOI] [PubMed] [Google Scholar]
  • [21].Tennikova TB, Freitag R. An introduction to monolithic disks as stationary phases for high performance biochromatography. J. High Res. Chromatogr. 2000;23:27–38. [Google Scholar]
  • [22].Turiel E, Fernandez P, Perez-Conde C, Gutierrez AM, Camara C. Flow-through fluorescence immunosensor for atrazine determination. Talanta. 1998;47:1255–1261. doi: 10.1016/s0039-9140(98)00206-9. [DOI] [PubMed] [Google Scholar]
  • [23].Meyer UJ, Zhi ZL, Meusel M, Spener F, Loomans E. Automated stand-alone flow injection immunoanalysis system for the determination of cephalexin in milk. Analyst. 1999;124:1605–1610. doi: 10.1039/a907121k. [DOI] [PubMed] [Google Scholar]
  • [24].Phillips TM. Microanalytical methods based on affinity chromatography. In: Hage DS, editor. Handbook of Affinity Chromatography. 2nd ed. Chapter 28. Taylor & Francis; New York: 2006. [Google Scholar]
  • [25].Clarke W, Choudhuri AR, Hage DS. Analysis of free drug fractions by ultra-fast immunoaffinity chromatography. Anal. Chem. 2001;73:2157–2164. doi: 10.1021/ac0009752. [DOI] [PubMed] [Google Scholar]
  • [26].Ohnmacht CM, Schiel JE, Hage DS. Analysis of free drug fractions using near infrared fluorescent labels and an ultrafast immunoextraction/displacement assay. Anal. Chem. 2006;78:7547–7556. doi: 10.1021/ac061215f. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Moser AC, Willicott CW, Hage DS. Clinical applications of capillary electrophoresis-based immunoassays. Electrophoresis. 2014;35:937–955. doi: 10.1002/elps.201300421. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Schiel JE, Tong Z, Sakulthaew C, Hage DS. Development of a flow-based ultrafast immunoextraction and reverse displacement immunoassay: analysis of free drug fractions. Anal. Chem. 2011;83:9384–9390. doi: 10.1021/ac201973v. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Pfaunmiller EL, Anguizola JA, Milanuk ML, Papastavros E, Carter N, Matsuda R, Zheng X, Hage DS. Development of microcolumn-based one-site immunometric assays for protein biomarkers. J. Chromatogr. A. 2014;1366:92–100. doi: 10.1016/j.chroma.2014.09.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Phillips TM. Multi-analyte analysis of biological fluids with a recycling immunoaffinity column array. J. Biochem. Biophys. Methods. 2001;49:253–262. doi: 10.1016/s0165-022x(01)00202-0. [DOI] [PubMed] [Google Scholar]
  • [31].Jiang T, Mallik R, Hage DS. Affinity monoliths for ultrafast immunoextraction. Anal. Chem. 2005;77:2362–2372. doi: 10.1021/ac0483668. [DOI] [PubMed] [Google Scholar]
  • [32].Clarke W, Schiel JE, Moser A, Hage DS. Analysis of free hormone fractions by an ultrafast immunoextraction/displacement immunoassay: studies using free thyroxine as a model system. Anal. Chem. 2005;77:1859–1866. doi: 10.1021/ac040127x. [DOI] [PubMed] [Google Scholar]
  • [33].Hage DS, Walters RR, Hethcote HW. Split-peak affinity chromatographic studies of the immobilization-dependent adsorption kinetics of protein A. Anal. Chem. 1986;58:274–279. doi: 10.1021/ac00293a003. [DOI] [PubMed] [Google Scholar]
  • [34].Hage DS, Walters RR. Dual-column determination of albumin and immunoglobulin G in serum by high-performance affinity chromatography. J. Chromatogr. 1987;386:37–49. doi: 10.1016/s0021-9673(01)94582-0. [DOI] [PubMed] [Google Scholar]
  • [35].Ren X. Flow injection fluoroimmunoassay for human transferrin using a protein A immunoreactor. Anal. Lett. 1994;27:1067–1074. [Google Scholar]
  • [36].Palmer DA, Evans M, Miller JN, French MT. Rapid fluorescence flow injection immunoassay using a novel perfusion chromatographic material. Analyst. 1994;119:943–947. [Google Scholar]
  • [37].Rico CM, Fernandez MD, Gutierrez AM, Conde MCP, Camara C. Development of a flow fluoroimmunosensor for determination of theophylline. Analyst. 1995;120:2589–2591. [Google Scholar]
  • [38].Zheng X, Li Z, Beeram S, Matsuda R, Pfaunmiller EL, Podariu M, White CJ, II, Carter N, Hage DS. Analysis of biomolecular interactions using affinity microcolumns: a review. J. Chromatogr. B. 2014;968:49–63. doi: 10.1016/j.jchromb.2014.01.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39].Middendorf L, Amen J, Bruce R, Draney D, DeGraff D, Gewecke J, Grone D, Humphrey P, Little G, Lugade A, Narayanan N, Oommen A, Osterman H, Peterson R, Rada J, Raghavachari R, Roemer S, Daehne S, editors. Near-Infrared Dyes for High Technology Applications. Kluwer Academic; Netherlands: 1998. [Google Scholar]
  • [40].Patonay G, editor. HPLC Detection - Newer Methods. VCH; New York: 1993. [Google Scholar]
  • [41].Smith PK, Krohn RI, Hermanson GT, Mallia AK, Gartner FH, Provenzano MD, Fujmoto EK, Goeke NM, Olson BJ, Klenk DC. Measurement of protein using bicinchoninic acid. Anal. Biochem. 1985;150:76–85. doi: 10.1016/0003-2697(85)90442-7. [DOI] [PubMed] [Google Scholar]
  • [42].Anguizola JA. Ph.D. Dissertation. University of Nebraska; Lincoln: 2013. [Google Scholar]
  • [43].Hage DS, Walters RR. Non-linear elution effects in split-peak chromatography. I. Computer simulations for the cases of irreversible diffusion- and adsorption-limited kinetics. J. Chromatogr. 1988;436:111–135. [Google Scholar]
  • [44].Rollag JG, Hage DS. Non-linear elution effects in split-peak chromatography. 2. Role of ligand heterogeneity in solute binding to columns with adsorption-limited kinetics. J. Chromatogr. A. 1998;795:185–198. doi: 10.1016/s0021-9673(97)00975-8. [DOI] [PubMed] [Google Scholar]
  • [45].Nelson MA, Reiter WS, Hage DS. Chromatographic competitive binding immunoassays: a comparison of the sequential and simultaneous injection methods. Biomed. Chromatogr. 2003;17:188–200. doi: 10.1002/bmc.241. [DOI] [PubMed] [Google Scholar]
  • [46].Hage DS, Thomas DH, Chowdhuri AR, Clarke W. Development of a theoretical model for chromatographic-based competitive binding immunoassays with simultaneous injection of sample and label. Anal. Chem. 1999;71:2965–2975. doi: 10.1021/ac990070s. [DOI] [PubMed] [Google Scholar]
  • [47].Lee M, Durst RA, Wong RB. Comparison of liposome amplification and fluorophor detection in flow-injection immunoanalyses. Anal. Chim. Acta. 1997;354:23–28. [Google Scholar]
  • [48].Kaneki N, Xu Y, Kumari A, Halsall HB, Heineman WR, Kissinger PT. Electrochemical enzyme immunoassay using sequential saturation technique in a 20-μL capillary: digoxin as a model analyte. Anal. Chim. Acta. 1994;287:253–258. [Google Scholar]
  • [49].Kramer P, Schmid R. Flow injection immunoanalysis (FIIA) - a new immunoassay format for the determination of pesticides in water. Biosens. Bioelectron. 1991;6:239–243. doi: 10.1016/0956-5663(91)80009-m. [DOI] [PubMed] [Google Scholar]
  • [50].Nilsson M, Haakanson H, Mattiasson B. Process monitoring by flow-injection immunoassay: evaluation of a sequential competitive binding assay. J. Chromatogr. 1992;597:383–389. doi: 10.1016/0021-9673(92)80135-h. [DOI] [PubMed] [Google Scholar]
  • [51].Hage DS, Thomas DH, Beck MS. Theory of a sequential addition competitive binding immunoassay based on high-performance immunoaffinity chromatography. Anal. Chem. 1993;65:1622–1630. doi: 10.1021/ac00059a023. [DOI] [PubMed] [Google Scholar]

RESOURCES