. 2001 Dec 1;106(6):1071–1094. doi: 10.6028/jres.106.058

Standard Reference Material (SRM 1990) For Single Crystal Diffractometer Alignment

W Wong-Ng 1, T Siegrist 2,3, G T DeTitta 4, L W Finger 5, H T Evans Jr 6, E J Gabe 7, G D Enright 7, J T Armstrong 8, M Levenson 8, L P Cook 8, C R Hubbard 9
PMCID: PMC4865310  PMID: 27500067

Abstract

An international project was successfully completed which involved two major undertakings: (1) a round-robin to demonstrate the viability of the selected standard and (2) the certification of the lattice parameters of the SRM 1990, a Standard Reference Material® for single crystal diffractometer alignment. This SRM is a set of ≈3500 units of Cr-doped Al2O3, or ruby spheres [(0.420.011 mole fraction % Cr (expanded uncertainty)]. The round-robin consisted of determination of lattice parameters of a pair of crystals: the ruby sphere as a standard, and a zeolite reference to serve as an unknown. Fifty pairs of crystals were dispatched from Hauptman-Woodward Medical Research Institute to volunteers in x-ray laboratories world-wide. A total of 45 sets of data was received from 32 laboratories. The mean unit cell parameters of the ruby spheres was found to be a=4.7608 ű0.0062 Å, and c=12.9979 ű0.020 Å (95 % intervals of the laboratory means). The source of errors of outlier data was identified. The SRM project involved the certification of lattice parameters using four well-aligned single crystal diffractometers at (Bell Laboratories) Lucent Technologies and at NRC of Canada (39 ruby spheres), the quantification of the Cr content using a combined microprobe and SEM/EDS technique, and the evaluation of the mosaicity of the ruby spheres using a double-crystal spectrometry method. A confirmation of the lattice parameters was also conducted using a Guinier-Hägg camera. Systematic corrections of thermal expansion and refraction corrections were applied. These rubies are rhombohedral, with space group R3¯c. The certified mean unit cell parameters are a=4.76080±0.00029 Å, and c=12.99568 ű0.00087 Å (expanded uncertainty). These certified lattice parameters fall well within the results of those obtained from the international round-robin study. The Guinier-Hägg transmission measurements on five samples of powdered rubies (a=4.7610 ű0.0013 Å, and c = 12.9954 ű0.0034 Å) agreed well with the values obtained from the single crystal spheres.

Keywords: alignment standard, Cr-content, international round robin, NIST SRM 1990, ruby spheres, single crystal x-ray diffractometers

1. Introduction

In order to provide industrial, academic and government laboratories with a Standard Reference Material® (SRM) for the alignment of single crystal x-ray diffractometers, an international project was completed which involved two major undertakings: (1) an international round-robin to demonstrate the viability of the selected standard and (2) the certification of the lattice parameters of the SRM.

A lattice parameter standard is essential for the single crystal x-ray diffraction community for two principal reasons. First, x-ray structural determinations using automatic x-ray diffractometer data collection and automatic structure solution schemes require accurate initial cell parameter data. The unit cell metric gives a good indication of the Bravais Lattice, and therefore should be known as precisely as possible. Accurate cell parameters can only be obtained with well-aligned x-ray diffractometers; a standard crystal is critical for diffractometer calibration. Second, a standard is important for both intra- and inter-laboratory comparison of data. Much structural work reported in literature today is based on single crystal x-ray diffraction methods, and there have been claims of six digit accuracy in lattice parameters. The editors of Acta Crystallographica considered these data unrealistic and mostly unsupported. The Commission on Crystallographic Apparatus and Standards Committee of the International Union of Crystallography (IUCr) was requested to investigate the accuracy and precision of lattice parameters measured in the industrial, academic and government x-ray laboratories. The project was initially organized by Professor Ludmilla Malakhova of the Institute of Crystallography in the Soviet Union and passed into the hands of George DeTitta of the Hauptman-Woodward Medical Research Institute. An international round- robin project was conceived during the 1981 IUCr meeting in Ottawa, Canada. As various other institutions became interested in the project the goals of the project expanded and matured so that a number of important issues were identified. The American Crystallographic Association (ACA) also joined in this effort.

Presently, other than a few commercially available crystals from various manufacturers and materials that have been prepared locally at individual laboratories, no certified standard material is available for widespread use in diffractometer alignment. The leadership of NIST in this project is crucial to the development and the future distribution of the SRM for a number of reasons. First, it is the mission of NIST to take an active role in developing measurement standards and techniques, and NIST has a strong internal interest in producing SRMs for in-house research instruments as well. Second, NIST has the experience and expertise, in addition to the storage and distribution facilities required to make SRMs available for a wide range of users. It also has the necessary personnel to handle the business aspects of marketing and selling the products. Third, neither the IUCr nor the ACA has the resources to support a long term commitment to produce and maintain an SRM to supply a broad community of commercial, academic, and government needs.

This paper summarizes the round-robin and the SRM certification projects. The following discussion of the round-robin project includes the goals, procedures, results of statistical analysis, and errors in diffractometer alignment. Although the particular diffractometers used do not include all the types of diffractometers in use worldwide, this study gives a reasonable set of data for inter-laboratory comparison. The discussion of the SRM certification project includes descriptions of several important aspects, such as physical characteristics of the spheres, experimental procedures used for determining the lattice parameters, the Cr content of the ruby, and factors affecting accuracy of single-crystal diffractometer alignment and lattice parameter determination.

2. International Round-Robin

2.1 Goal of Study

The goal of this international round-robin project has four parts: (1) to determine realistic limits on the precision and accuracy of lattice parameters using various commercial diffractometers; (2) to assess the x-ray technique and the state of the instrument employed at each local laboratory, (3) to evaluate data collection method at each laboratory, and (4) to evaluate the usefulness of the ruby spheres as a NIST standard reference material (SRM) for diffractometer alignment.

2.2 Procedures

The round-robin project involves the use of single crystal x-ray diffractometers to determine the lattice parameters of a standard crystal (alignment standard) and an “unknown” reference crystal (representing a typical laboratory sample). Preliminary studies of the structural and physical properties of a batch of five hundred ruby spheres which were purchased by the ACA and IUCr from the Arcanum Corporation,1 Michigan, indicated these spheres to be stable, relatively homogeneous, easy to handle, and safe to use. A single crystal boule was prepared using the Verneuil technique (flame fusion) [1]. Cubes were cut from this boule by a diamond saw and then ground between two disks rotating in opposite directions to produce small spheres. These spheres were reported by the manufacturer to have a diameter of 0.152 mm and sphericity of 0.0013 mm. The sphere contains of the order of 3.4 × 1014 Al2O3 formula units. The high crystal quality and the high hardness give rise to small thermal motion parameters, and therefore produce strong reflections at high angles for MoKα radiation as well as for the copper. In addition, this radius of ruby sphere is appreciably smaller than a typical incident beam of a diffractometer and therefore can satisfy beam uniformity conditions. Figure 1 shows a SEM micrograph of a typical ruby sphere. These small crystals are nearly perfect spheres which allow accurate optical centering. From well-centered reflections at high angles, high precision lattice parameters can be obtained [25]. Therefore, they can be used as a round-robin standard and a potential Standard Reference Material® for alignment of single crystal x-ray diffractometers.

Fig. 1.

Fig. 1

SEM micrograph showing the morphology of a typical ruby sphere (SRM 1991), with a diameter of 150 μm.

During the early stage of study, an organic crystal (Raffinose) was chosen as an unknown reference crystal which has cell dimensions ranging from moderately short to long (≈8×12×23 Å). This material, however, was found to be unstable for shipping and has been replaced by synthetic ferrierite zeolite crystals, which are stable under operating conditions. These “giant” zeolite molecular sieve crystals crystallized with a primitive orthorhombic unit cell, with a formula Al2Si34O72 · 2(C5H5N) · 2HF. These crystals exhibit a plate-like morphology, posing a typical challenge to crystal alignment. The cell parameters are estimated to be a = 18.8430(51) Å, b = 14.0981(33) Å and c = 7.4383(24) Å[6]. Figure 1 shows a SEM micrograph of a typical ruby sphere. These small crystals are nearly perfect spheres which allow accurate optical centering. Figure 2 shows a SEM micrograph of the morphology of a selected zeolite crystal. The zeolite crystals were obtained from the Chemistry Department of the University of Toronto, and were grown using non-aqueous solvent [6].

Fig. 2.

Fig. 2

SEM micrograph showing the morphology of the zeolite crystal.

Optical examination of crystals of both ruby and zeolite was followed by mounting approximately 100 of each type. Both the zeolite and ruby spheres were mounted on the tips of fibers of approximate diameter of 0.1 mm which were secured in blocks. The zeolite crystals were mounted along the face diagonals. Preliminary characterization, both by x-ray photography and by diffractometry methods (using Siemens P21 diffractometers) allowed identification of the most suitable candidates for the round-robin project. Orientations were measured for the selected round-robin crystals. Reflections were measured on the positive and negative sides of 2θ and in eight positions described by Hamilton in the International Table for Crystallography [2], by King and Finger [3] and by Hazen and Finger [4]. Typically 48 reflections were used (6 independent reflections in 8 quadrants). A total of 100 crystals (50 sets of ruby spheres and zeolite crystals) were used to prepare the round-robin kits. Ten kits each containing 5 rubies and 5 unknowns (Fig. 3) were assembled and shipped to participants from the Hauptman-Woodward Medical Research Institute, Buffalo. Mounts of the crystals were carefully evaluated as to mechanical strength. Special boxes were constructed for the shipment of the crystals. Several severe baggage-handling simulations were undertaken (with drops of 2 to 3 meters) to evaluate the likelihood that the samples would survive the shipping process. The samples were assembled in their enclosures package and mailed out. Notebooks were prepared with necessary data on the crystals, literature references, instruction sheets, etc. and were dispatched along with the samples for the evaluation of diffractometry. Instructions were designed to ensure the collection of data pertinent to the evaluation of the crystal centering algorithm and the alignment of the instrument. This procedure was based on the method developed by King and Finger [3].

Fig. 3.

Fig. 3

Interior of the crystal kit.

The main working hypotheses of the project is that the main errors associated with obtaining accurate lattice parameters are due to the misalignment of the diffractometer and of the diffracting samples. Therefore participants were expected to measure data pertinent to evaluation of: the crystal centering algorithm, the alignment of the instrument, and the lattice parameter determination software. The procedure included in the round-robin instructions also called for measurement of auxiliary data in order to determine the mechanical and optical conditions of sample and diffractometer.

In summary, scientists were asked to perform the following:

  1. Determine the lattice parameters of the zeolite crystals by standard laboratory procedures in use in their facilities.

  2. Determine additional diffraction data (i.e., orientation matrix [7]) as instructed, providing information concerning the state of their equipment and sample.

  3. Measure data on the ruby standard pertinent to the evaluation of the centering algorithm, alignment of the instrument, and lattice parameter determination software.

  4. Submit results for statistical analysis.

After receiving the data, the design team which included scientists from the Geophysical Laboratory, the Hauptman-Woodard Medical Research Institute and NIST, then analyzed results statistically.

The orientation matrices used for the evaluation were determined with respect to the following orthonormal coordinate frame (shown in Fig. 4), where x, y, and z are the crystal Cartesian axes, and a*, b*, and c* is the reciprocal lattice. This orthonormal coordinate frame is valid for diffractometer types such as Syntex, Nicolet, Siemens P3, etc.

UB=(a*xb*xc*xa*yb*yc*ya*zb*zc*z)

where UB is the orientation matrix

(hokolo)=UB(hkl)

where h, k, l are the Miller indices, and ho, ko, lo are the coordinates of a reflection in the ϕ axial frame.

Fig. 4.

Fig. 4

Orthonormal coordinate frame used for analysis.

The definition of diffractometer angles used is that given by Busing and Levy [5], which is the bisecting mode in which the incident, diffracted beams and the counter all lie on a horizontal plane, and ω = 0°.

Various types of systematic errors can affect the positions of the diffracted beam, which in turn affect the lattice parameters of the crystal being studied. These errors include diffractometer zero position, errors in crystal centering, and misalignment of the instrument (including error in counter or tube height). In the procedure by King and Finger [3], the measured angles for a single reflection, or Friedel pair of reflections in eight different orientations (quadrants) are used to determine the values for various errors associated with the mounting of the crystal and alignment of the diffractometer.

2.3 Participants

A survey of structural crystallographers identified about 50 laboratories worldwide who were interested and willing to participate in the round-robin project, and preliminary information provided by this survey has also identified the kinds of equipment that were in use in the community for this work.

The 50 international participants are all active crystallographers who make frequent use of diffractometers for their research, the areas of macromolecules, small molecules, inorganic, organic, intermetallic, pharmaceutical, and ceramics. More than 10 different types of diffractometers were employed by these participants, including CAD4 (Enraf Nonius), AFCS6 (Rigaku), P3 (Syntex), SMART (Siemens CCD), R3m (Nicolet), P21 (Syntex), Huber, Kumar, StoeAED, Stoe-4C, R-4Cir, and P4. Molybdenum radiation (MoKα) was used by most of the laboratories, followed in frequency of use by copper radiation (CuKα). Silver radiation (Ag Kα was used in one laboratory.

2.4 Results and Discussion

2.4.1 Lattice Parameter Measurement

A total of 45 (44 complete) sets of reports for the ruby spheres and zeolite crystals have been received from 32 laboratories. Information obtained from these reports includes type of diffractometer(s) and wavelength used, identification of crystal sets, measured cell parameters, and relevant angles of the diffractometers including in the Eulerian system: χ, ϕ, ω, and 2θ [3]; and in the case of a Kappa-type diffractometer, the angles are expressed in the Kappa system. Among these data sets, 15 laboratories used the eight-quadrant method, which allowed for further detailed evaluation by using the routine by King and Finger [3].

Most of the participants collected their data within the ambient temperature range. A correction of the lattice parameters to nominal room temperature of 298 K was made before comparison. Belyaev [8] and Campbell and Grain [9] reported the thermal expansion coefficient of Al2O3. Within the range of 0 °C to 100 °C the α-Al2O3 was found to expand approximately linearly. The expansion is anisotropic and the corrections can be calculated according to the following expression:

a(25°C)=a[1+αa(25T)],whereαa=5.0×106c(25°C)=c[1+αc(25T)],whereαc=6.66×106.

One participant collected the data at a low temperature of 153 K, and the data were not used.

We have also applied a refraction correction [10] in two parts. One corresponds to the Snell’s law correction and is too small to be included (δd/d = −v cotθ = (1−n) cotθ). Another part is due to the change of wavelength: 1−n = cρλ2Z/Σa), where ρ is the density (taken as 3.98 g/cm3 [8]), λ is the wavelength, Z is the atomic number, a is the atomic weight, and c is a constant. In the equation, δd/d = (1−n)/n, where n is the refractive index, we found that for Mo radiation (1−n) = 2.69 × 10−6, and for Cu radiation, the value of (1−n) = 1.27 × 10−5. This correction is a relatively small quantity.

For the ruby crystals, most least-squares lattice refinements were constrained to obey the hexagonal setting. For those that were not constrained, results indicated that the parameters are very close to hexagonal. Table 1 summarizes the lattice parameters of the ruby spheres after the application of thermal expansion and refraction corrections. For those laboratories that have reported more than one set of data, the values were averaged and the mean value obtained was then computed with other data sets for obtaining the global mean. Table 1 also gives the mean lattice parameter values for several laboratories. Table 2 gives the corresponding data for the zeolite crystals. The mean values of the lattice parameters and the 95 % intervals on the grand mean and on the population of laboratory means are:

  1. ruby spheres:
    a=4.7608ű0.0011Å(95%intervals on grand mean)±0.0062Å(95%intervals of laboratory means)b=4.7609ű0.0010ű0.0057Åc=12.9979ű0.0035ű0.020Å
  2. zeolite crystals:
    a=18.8338ű0.0051Å(95%intervals on grand mean)±0.014Å(95%intervals of laboratory means)b=14.1036ű0.0054ű0.014Åc=7.4366ű0.0026ű0.0070Å

The mean value of the lattice parameters in general agrees well with the certified SRM data (discussed later), which is (at 25°C): a=4.76080ű0.00029Å (expanded uncertainty), and c=12.99568ű0.00087Å, despite the large spread in reported values.

Table 1.

Lattice parameters of the ruby spheres after the application of the thermal and refraction corrections. Identification of each laboratory is not given. The total number of experiments are greater than the number of the laboratory because of multiple experiments performed in some laboratories

Lab No. Expt. No. Rad. Ty. Diff. Ty. a (Å) b (Å) c (Å) T (°C) Ave. (each laboratory)
a(Å) b(Å) c(Å)
1 1 Mo CAD4 4.7569 4.7574 12.9848 25.00
2 2 Mo AFC6S 4.7593 4.7595 13.0018 −145.00
3 3 Cu CAD4 4.7573 4.7622 12.9942 21.00
4 4 Mo CAD4 4.7595 4.7595 12.9873 23.00 4.7601 4.7601 12.9899
5 Mo CAD4 4.7608 4.7608 12.9924 23.00
5 6 Mo CAD4 4.7612 4.7612 12.9960 25.00 4.7612 4.7612 12.9985
7 Mo CAD4 4.7612 4.7612 13.0010 25.00
6 8 Mo CAD4 4.7599 4.7599 12.9924 25.00 4.7608 4.7608 12.9957
9 Mo CAD4 4.7618 4.7618 12.9989 −145.0
7 10 Mo P3 4.7597 4.7591 12.9915 22.00
8 11 Mo CAD4(1) 4.7601 4.7601 12.9950 20.00 4.7617 4.7617 12.9974
12 Mo CAD4(2) 4.7602 4.7602 12.9946 20.00
13 Cu CAD4(3) 4.7616 4.7616 12.9956 20.00
14 Mo CAD4(4) 4.7599 4.7599 12.9924 20.00
15 Mo SMART 4.7640 4.7668 13.0095 20.00
9 16 Mo CAD4 4.7609 4.7609 12.9951 20.00
10 17 Mo R3m 4.7619 4.7619 12.9925 23.00
11 18 Mo P3 4.7658 4.7661 13.0097 25.00 4.7623 4.7622 12.9991
19 Cu P3 4.7589 4.7584 12.9885 25.00
12 20 Mo CAD4 4.7606 4.7614 12.9965 23.00
13 21 Mo AFC6 4.7642 4.7637 13.0067 23.00
14 22 Mo AFC6 4.7654 4.7647 13.0161 26.00
15 23 Mo Huber 4.7696 4.7696 13.0207 20.00
16 24 Cu CAD4 4.7603 4.7595 12.9959 20.00
17 25 Ag CAD4 4.7603 4.7610 12.9947 20.00
18 26 Cu CAD4 4.7588 4.7588 12.9926 20.00
19 27 Mo Kuma K 4.7602 4.7602 12.9925 22.00
20 28 Mo Stoe AED 4.7614 4.7614 12.9952 24.00 4.7623 4.7622 12.9991
29 Mo Stoe AED 4.7621 4.7621 12.9977 24.00
30 Mo P4 4.7634 4.7634 13.0004 24.00
31 Mo P4 4.7629 4.7629 13.0024 24.00
21 32 Mo CAD4 4.7594 4.7577 12.9755 20.00 4.7586 4.7595 12.9881
33 Mo CAD4 4.7606 4.7605 12.9905 20.00
34 Mo CAD4 4.7557 4.7602 12.9985 20.00
22 35 Mo CAD4 4.7559 4.7548 12.9843 23.00
23 36 Mo CAD4 4.7592 4.7596 12.9932 23.00
24 37 Mo P21 4.7662 4.7642 13.0032 25.00
25 38 Cu Russia4C 4.7638 4.7639 13.0062 25.00
26 39 Mo CAD4 4.7555 4.7558 12.9870 25.00
27 40 Mo Stoe 4C 4.7614 4.7614 12.9983 22.00
28 41 Mo AED 2 4.7639 4.7639 13.0293 22.00
29 42 Mo CAD4 4.7621 4.7621 12.9983 22.00
30 43 Cu CAD4 4.7585 4.7588 12.9933 20.00
31 44 Mo CAD4 4.7609 4.7605 12.9945 25.00
32 45 Mo Rigaku 4.7570 4.7570 13.0020 25.00
Table 2.

Round-Robin Results of the reference zeolite crystals, Al2Si34O72 · 2(C5H5O) · 2HF. The identification of each laboratory is not given. The total number of experiments are greater than the number of the laboratory because of multiple experiments performed in some laboratories.

Lab No. Expt. No. Rad. Ty. Diff. Ty. a (Å) b (Å) c (Å) T(°C) Ave. (each laboratory)
a(Å) b(Å) c(Å)
1 1 Mo CAD4 18.8286 14.0963 7.4316 25.20
2 2 Mo AFC6S 18.7568 14.0915 7.4125 −145.00
3 3 Cu CAD4 18.8320 14.1060 7.4345 21.00
4 4 Mo CAD4 18.8309 14.1016 7.4362 23.00 18.8306 14.1014 7.4353
5 Mo CAD4 18.8303 14.1012 7.4344 23.00
5 6 Mo CAD4 18.8360 14.1060 7.4340 25.00 18.8356 14.1055 7.4350
7 Mo CAD4 18.8352 14.1050 7.4360 25.00
6 8 Mo CAD4 18.8257 14.1014 7.4323 25.00
7 9 Mo P3 18.8270 14.1030 7.4300 22.00
8 10 Mo CAD4(1) 18.8326 14.0977 7.4341 20.00 18.8290 14.1023 7.4341
11 Mo CAD4(2) 18.8265 14.1020 7.4333 20.00
12 Mo CAD4(3) 18.8318 14.1032 7.4351 20.00
13 Mo CAD(4) 18.8250 14.1073 7.4340 20.00
9 14 Mo CAD4 18.8249 14.1037 7.4361 20.00
10 15 Mo R3m 18.8410 14.1140 7.4360 23.00
11 16 Mo P3 18.8500 14.1095 7.4910 25.00 18.8350 14.1006 7.4606
17 Cu P3 18.8199 14.0918 7.4303 25.00
12 18 Mo CAD4 18.8300 14.1060 7.4329 22.00
13 19 Mo AFC6 18.8503 14.1459 7.4400 23.00
14 20 Mo AFC6 18.8632 14.1284 7.4501 26.00
15 21 Mo Huber 18.8500 14.1160 7.4420 20.00
16 22 Cu CAD4 18.8348 14.1019 7.4350 20.00
18 23 Cu CAD4 18.8343 14.0924 7.4305 20.00
19 24 Mo Kuma K 18.8336 14.0979 7.4344 22.00
20 25 Mo Stoe AED 18.8394 14.1118 7.4395 24.00 18.8404 14.1028 7.4381
26 Mo Stoe AED 18.8332 14.1192 7.4398 24.00
27 Mo P4 18.8442 14.0905 7.4352 24.00
28 Mo P4 18.8450 14.0896 7.4378 24.00
21 29 Mo CAD4 18.8050 14.0800 7.4240 20.00 18.8070 14.0887 7.4280
30 Mo CAD4 18.8150 14.0870 7.4370 20.00
31 Mo CAD4 18.8010 14.0990 7.4230 20.00
22 32 Mo CAD4 18.8210 14.0940 7.4313 23.00
23 33 Mo CAD4 18.8000 14.0900 7.4270 23.00
24 34 Mo P21 18.8513 14.0757 7.4407 25.00
25 35 Cu Russia4C 18.8450 14.1050 7.4387 25.00
26 36 Mo CAD4 18.8220 14.0950 7.4340 25.00
27 37 Mo Stoe 4C 18.8400 14.1050 7.4390 22.00
28 38 Mo AED 2 18.8609 14.1283 7.4507 22.00
29 39 Mo CAD4 18.8430 14.0981 7.4383 22.00
30 40 Cu CAD4 18.8269 14.0758 7.4303 22.00
31 41 Mo CAD4 18.8235 14.1027 7.4339 25.00
32 42 Mo Rigaku 18.8280 14.1230 7.4380 25.00

The fact that the round-robin results show a large spread (as compared to the certified values) of data indicates that uncertainty of measurements in general laboratories is relatively large unless great care is taken about the diffractometer alignment. The closest match of the lattice parameter results between the round-robin data and ruby SRM value is that submitted from laboratory No. 9 (a = 4.7609 Å, c = 12.9951 Å).

2.4.2 Histograms

Figure 5 displays histograms of the results of lattice parameters a and c of 30 laboratories participating in the round-robin project. Note that results of two laboratories were not used due to problems with the measurements. For the lattice parameter a, the results of the laboratories are centered on the certified value and are symmetric around this certified value. For the lattice parameter c, there are three results that are somewhat separated from the other results. The spread of the results for the c parameter (relative standard deviation = 0.0035) is greater than that for the a parameter (relative standard deviation = 0.001).

Fig. 5.

Fig. 5

Histograms of the results of (a) lattice parameters a, and (b) lattice parameter c of the ruby spheres from 30 laboratories participating in the round-robin project.

2.4.3 Youden Plots

Those laboratories that reported ruby lattice parameters with relatively large deviations from the mean also reported corresponding significant deviations for the zeolite crystals (for example, laboratory Nos. 14, 24, and 28). This similarity strongly indicates that the large deviation of the ruby lattice parameters can be used as an indicator of the alignment condition of the diffractometer. Once the diffractometer is re-aligned using the ruby spheres, the accuracy of the lattice parameters of future determinations will hopefully be improved.

One technique to analyze results from a round-robin exercise is the use of the Youden plot [11]. In the Youden plot, two related measurements are plotted versus each other. In the present study, the lattice parameter a from the reference material is plotted against the lattice a of the standard reference material (see Fig. 6). The plot and auxiliary calculations can be used to quantify between-laboratory variation (laboratory biases) and within-laboratory variation. A laboratory is biased if it tends to be higher or lower than the true value. The presence of laboratory biases appears in the Youden plot as a positive linear pattern.

Fig. 6.

Fig. 6

Joint confidence regions on the round-robin results (Youden analysis). Plot of lattice parameter a from the reference material (zeolite crystals) against the lattice a of the standard reference material (ruby spheres).

Using robust measures of the mean and standard deviation of the results, outlier laboratories can be identified [12]. The two dotted lines in Fig. 6 represent the robust estimates of the means of the two sets of measurement results. The solid-lined ellipse defines a 95 % confidence region for the pairs of results. Four data sets (laboratories Nos. 14, 21, 23 and 28) are well outside the ellipse.

The dotted-line ellipse in Fig. 6 defines a 10 % confidence region. Using this region, the best data sets determined at eight laboratories are determined. Figure 7 repeats the Youden analysis based solely on these eight laboratories. One of the eight is just outside the 95|jpercnt| region, which is not indicative of outliner behavior. Even for these “best” data sets, between-laboratory variation exists. The sum of the between- and within-laboratory variation can be used as the basis of a measure of the standard uncertainty for these laboratories. The resulting relative standard uncertainty is 0.0018. Thus, the third decimal place is the limit of the accuracy for these “best” data sets.

Fig. 7.

Fig. 7

Youden analysis based solely on the results of eight best laboratories.

2.4.4 Errors of Diffractometer Alignment

As mentioned above, various types of errors could be evaluated for the data sets that were collected using the eight-quadrant technique, namely, diffractometer zero position, errors in crystal centering, and misalignment of the instrument including error in counter or tube height. In general, for data that fall well within the region of the 95 % interval, these errors appear to be relatively small. On the other hand, those data sets, which show great deviations from the mean also exhibit significantly large errors due to one or more types of misalignment error.

On many occasions, deviations of measured results from standard values could be explained in terms of the errors in the diffractometer alignment. For example, one can plot the values of δd/d versus various types of errors. Estimation of the angles can be used to correct for the errors from the measurements of the ruby spheres. The corrected values can then be used to refine the lattice parameters, and much higher precision can be obtained. Examples of results of analysis pertaining to various types of corrections are shown in Figs. 812. Fig. 8 shows the Δd/d values (where Δd = dstddobs) versus 2θ. The round-robin data are shown as filled squares. The α1−α 2 doublet (with latest values of Mo and Cu wavelength reported by Haertwig et al. [13], i.e., λ for CuKα is 0.154059292(45) nm, λ for MoKα is 0.070931631(84) nm) was used for the fitting. It is seen that the fitted doublet values (filled circles) almost form a straight line at the value of Δd/d = 0.0. It can be seen that most of the measured reflection data have relatively low 2θ values which are lower than the fitted doublet values.

Fig. 8.

Fig. 8

Plot of δd/d vs 2θ.

Fig. 9.

Fig. 9

Plot of δd/d vs Δx crystal offset.

Fig. 10.

Fig. 10

Plot of δd/d vs Δy crystal offset.

Fig. 11.

Fig. 11

Plot of δd/d vs Δz crystal offset.

Fig. 12.

Fig. 12

Plot of δd/d vs Δh, an error in the height of the counter aperture [3].

Figures 912 contain the plots of Δd/d versus Δx, Δy, Δz, and Δh which are the errors in centering the crystal on the diffractometer. For ruby spheres, these deviations should be no larger than ±0.005 mm; however, the value was found to be as large as 0.5 mm for Δx. Clearly, such large apparent errors are indicative of severe problems with the diffractometer, including large inaccuracies and/or non-uniformity in the gears, or failure of the goniometer axes to intersect in a point, or lack of centering of the crystal.

There were a few examples of “bad” lattice parameter data sets that we have examined and we were able to identify the source of the “problem.” Three examples of measurement of the ruby spheres can be used to illustrate this point (laboratory Nos. 14, 15, and 28). Each of these laboratories used the eight-quadrant routine. For each set of data, the maximum deviation between dcal and corrected dobs is −0.002 Å. It appears that the corrections succeeded in removing the main errors. For No.28, the value of Δ(h) ranged from 0.125 to 0.142. This large value indicates a serious misalignment of the diffractometer, due to the counter aperture offset. The corresponding Δ(h) value of No. 14 was from 0.069 to 0.118, for No. 15 is from −0.009 to −0.037, which is approximately the resolution of the measurement.

2.5 Summary of the Round-Robin Study

The results of the round-robin project re-emphasized that well-aligned diffractometers are essential for obtaining accurate lattice parameters, and confirmed that the ruby spheres satisfy the criteria required of a standard reference material.

  1. The ruby spheres are a stable material that possesses high symmetry. They are easy to handle and readily used to perform both optical and diffractometer alignment, and are a good standard for alignment, and for inter-laboratory comparison of data. They can be useful to diagnose various possible sources of diffractometer errors (including misalignment of the crystal in x,y,z coordinates, zero settings for the instrument, error in counter or tube height, etc.), therefore enabling accurate alignment of the instrument.

  2. The magnitude of the deviation between the measured ruby lattice parameters and the SRM values can be used as an indicator for the condition of the diffractometer. Once the diffractometer is re-aligned using the ruby spheres, the accuracy of the lattice parameters from determinations of unknown crystals will be improved.

  3. The mean value of the lattice parameters of the ruby data sets agree with the SRM data and that reported by Kuperman et al. [6], respectively, despite the large spread of round-robin data.

  4. For data sets with large deviations of the measured ruby cell from that of the SRM value, the corresponding cell data of the zeolite crystals also show a similar magnitude of deviations from the known values, indicating that improved alignment using the ruby will increase the accuracy of a laboratory sample.

  5. With the exception of a few outlier results, the distributions for both the ruby and zeolite data sets are symmetric. 6. To obtain accurate cell parameters, high angle reflections must be used (i.e., 2θ > 60° (Mo), peak separation of α1, α2 ≈ 0.4°) for least-squares refinements, otherwise the lattice parameters are too small.

  6. The eight-quadrant algorithm is a reliable method to be used for single crystal diffractometer alignment, and the King and Finger method [3] can be used to estimate angle corrections.

  7. In some data sets, correction of diffractometer or crystal alignment errors resulted in much better agreement with the SRM value. In other cases, no single variable can adequately account for the variations, and there may be inherent diffractometer defects.

  8. The standard uncertainty from the results of eight “best” data sets is 0.0018. Thus, the third decimal place is the limit of the accuracy for these “best” laboratories. A realistic limit for uncertainty obtainable in laboratories is estimated to be Δa/a > 2 × 10−5. Any value smaller than this reported in the literature should be regarded as questionable.

3. Certification of the Ruby Spheres (SRM 1990)

3.1 Technical Objective

The technical objective of this project is to provide industry, academic and government laboratories with a standard reference material (SRM) for the alignment of single crystal diffractometers. This SRM is intended to improve the accuracy of lattice parameter determinations, and can be used to evaluate the x-ray technique and the state of the instrument employed at each local laboratory. The auxiliary data on the chromium content will also be useful for microanalytical calibrations.

3.2 International Collaborations

The development of the SRM has been carried out as a team effort, involving NIST, Lucent Technologies, Woodward-Hoffmann Medical Research Institute, Geophysical Laboratory, National Research Council (NRC) of Canada (Ottawa), U.S. Geological Survey, and Oak Ridge National laboratory.

Bell Laboratories (Lucent Technologies) and NRC of Canada have state-of-the-art, well-aligned serial diffractometers equipped with advanced data collection and analysis software necessary for the completion and certification of SRM 1990. A Guinier-Hägg camera was available at the U.S. Geological Survey and was used to perform secondary measurements by crushing the spheres into powder. NIST has a well-aligned single crystal x-ray diffractometer for orientation determinations of the spheres, as well as electron microprobe and scanning electron spectroscopy/energy dispersive x-ray (SEM/EDS) equipment for the determination of the Cr-content of these spheres.

3.3 Characteristics of the Ruby Spheres

An additional 3000 ruby spheres were purchased from the Arcanum corporation for the SRM certification process. Crystals for the round-robin and the SRM projects were obtained from the same boule to ensure maximum homogeneity.

In order to confirm that the Cr3+ ion substitutes for an Al3+ ion in ruby and to understand the local structural arrangement of ions around a substituted Cr3+ ion, Kizler et al. studied extended x-ray absorption fine structure (EXAFS) in the vicinity of the Cr absorption edge [14]. The findings were compared by Mott-Littleton [15] with an ionic model using two sets of pairwise potentials. Both the EXAFS results and the computations reveal that when Cr3+ ion substitutes for an Al3+, which is smaller, the surrounding ions relax to an arrangement similar to that for Cr in α-Cr2O3. For example, the octahedra of oxygen ions surrounding the Cr3+ ion is expanded, becoming similar in size to that characteristic of α-Cr2O3.

These ruby spheres fulfill the requirements for a standard, namely, they are a readily available material that has long term chemical stability and no phase transformation over a wide range of temperature. Ruby is insoluble in most solvents and not subject to radiation damage. It is also non- toxic, adequately homogeneous (less than 0.02 % mole fraction variation) and with a very small mosaic spread of 0.005° to 0.015° full-width-half-maxima (FWHM), which will give rise to properly shaped reflection profiles (discussed later). These spheres also possess high symmetry (rhombohedral) with the space group of R3¯c (Al and Cr in position 12c: 0,0,z; and O in position 18e: x,0,1/4 [16]). High symmetry will allow comparison of many symmetry-equivalent reflections. In addition to many reflections of high intensity which can be used for relatively fast measurements, ruby has a relatively low absorption coefficient of 124 cm−1 (for Cu radiation), which enables valid comparison of intensities of symmetry equivalent reflections.

3.4 Experimental

The lattice parameters of SRM 1990 were studied and certified by using four well-aligned commercial single crystal diffractometers (three Enraf-Nonius CAD4 and one Picker). A second method, the Guinier-Hägg transmission technique, was employed to support the diffractometer single crystal lattice parameter data. Statistical analysis of the resulting data was carried out in collaborations with statistician Mark Levenson of the Statistical Division of NIST. Auxiliary data such as the content of the chromium in these crystals were analyzed using the electron microprobe, and also by the energy dispersive x-ray technique (EDS).

3.4.1 Measurements using Single Crystal Diffractometers

The ruby crystals were mounted with a minimum amount of epoxy on the tip of ≈0.1 mm glass fibers or Lindemann capillary tubes. Among them, two sets of measurements (11 and 15 spheres each) were studied using Enraf-Nonius CAD4 diffractometers equipped with graphite monochromatized Mo and Cu radiation, respectively, at Bell Laboratories (Lucent Technologies). At NRC, fifteen spheres were studied using a CAD4 diffractometer equipped with monochromatized CuKα radiation, and four with a Picker Diffractometer equipped with monochromatized MoKα radiation. The diffractometer control program, DIFRAC [17], which has state-of-the-art data collection and data reduction schemes was employed for all data collection and reduction. This software package was developed at the National Research Council (NRC) of Ottawa and was described at the 1994 ACA Summer Meeting, Abstract M14 [18]. This program can be adapted to machines with different geometry, including the Kappa geometry used by the CAD4 diffractometers. All angles are specified in terms of the Eüler geometry with Eüler angles, 2θ, ω, χ, and ϕ.

On a CAD4 machine all precise centering of peaks is achieved by optimizing 2θ, ω, χ, at fixed ϕ, with continuous slow scans in the following sequences: (1) an ω/2θ scan with a variable slit, (2) a 2θ scan with −45° slit, and (3) a 2θ scan with the +45° slit. From the centroids of these scans the optimum values 2θ, ω, χ, and ϕ which will center the peak in the detector are calculated. If the initial position of the peak is vastly displaced from the center of the detector the routine performs an initial step scan in 2θ, ω, and χ to improve the starting position for the final precise adjustment. Peaks from an unknown crystal are located by rotating ϕ through 180° at each point on a specified grid of locations in 2θ, and χ, until the required number of coarse peak positions has been found and saved. These positions are then subjected to the precise centering described above and the final positions are used by the indexer to find the orientation matrix and the best reduced cell. The program also includes 2θ0, ω0, and χ0 corrections. The centroids of the peaks were found by determining the high and low-angle half-heights of fully resolved a1 peaks, followed by integrating between the two actual positions used to find the median. In this way, peak asymmetry will also be accounted for. Crucial to Kα1–α2 doublet is that the dispersion from the monochromator is at right angles to the scattering plane (vertical dispersion from monochromator and horizontal scattering plane for the CAD4 and the Picker diffractometers). Then, the Friedel pairs are symmetrical and the splitting of the doublet is identical, giving Friedel centroids relative to 2θ0 = 0 is determined unambiguously. The resolution of the doublet depends on the beam divergence and the size and mosaicity of the sample. The primary beam divergence is tied to the monochromator mosaic and the collimation of the beam. The ruby then samples 0.15 mm of the beam at the center. Therefore, the diffracted beam profile is a convolution of the monochromator mosaic (≈0.05° to 0.3°), the collimator, the ruby diameter and the ruby mosaic. The ruby mosaic is much smaller than the monochromator mosaic and therefore does not contribute. For each input reflection +h, +k, +l and −h, −k, −l are centered together with symmetry equivalents. The influence of absorption can be further reduced by measuring the difference in±for the Friedel reflections (Bond method), which is related to 2θ.

Because of the Kα1–α2 doublet, in order to obtain accurate d-spacing values, high 2θ angles were employed for the determination of lattice parameters in order to avoid α1 and α2 overlapping. For CuKα radiation, the 2θ values should be greater than 120° (at 120°, assuming FWHM of a peak profile of 0.3°, the α1–α2 splitting is 0.51°. This is a good separation and a valid determination of the α1 peak position by using the Busing and Levy method [19] (step to half-intensity on both sides of peak top)). For MoKα radiation, at 120° 2θ, α12 separation is 1.2°; at 60°, of 0.4° separation (α1- of 59.936°, and α2 of 60.336°). Therefore, for Mo radiation, it would be important to use reflections with 2θ> 60° for obtaining accurate lattice parameters.

For precise instrument alignment, at Bell Laboratories, equivalent settings of selected reflections were obtained in all octants for data sets collected by using Mo radiation in order to establish zero corrections on 2θ, ω, and χ. Using the diffractometer equipped with Cu radiation, not all equivalent reflections were accessible at high angles, therefore 2θ0 corrections were applied by using the software UNITCELL [20] which determines the unit cell parameters by incorporating 2θ0 as a parameter. In the case where “error-free” 2θ-values were obtained, the final lattice parameter refinement used the 2θ’s only.

X-ray wavelengths used for all calculations were taken from Haertwig et al. [13] and from Cohen and Taylor [21]. The values for CuKα1 maximum is 1.54059292 (45)Å or (8047.8264(24) eV, and for MoKα1 radiation the maximum is 0.70931631 (84) Å, or 17479.401 (21) eV.

3.4.2 Measurement of the Cr Content

The homogeneity and the quantity of the chromium content of these spheres were investigated by using both electron microprobe and quantitative SEM/EDS techniques. A total of 15 spheres were studied. One of the spheres was randomly selected to study in detail using the electron microprobe technique for the Cr concentration, and to study whether Cr is relatively uniformly distributed in the sphere. After the Cr content was obtained, this sphere was in turn used as a secondary standard for the rest of the 14 spheres by using an SEM/EDS broad-beam technique. These spheres were prepared for analysis by potting in epoxy and polishing and carbon coating. The final polishing step was completed using 0.1 μm diamond abrasive.

3.4.2.1 Electron Microprobe Study of the Secondary Standard
Sample Preparation

Polished grains of chromite and Cr-bearing pyroxene mineral standards [22] were used for the evaluation of the Cr content. These two standards and the ruby sphere were mounted on a one- inch diameter holder. The sample block was carbon-coated for analysis using standard laboratory procedures. The sample block was mounted in the electron microprobe stage holder along with laboratory reference standard blocks.

Analytical Method Summary

The samples and laboratory reference standards were placed in a JEOL 8600 electron microprobe and examined by reflected light optical microscopy, secondary and back-scattered electron imaging, qualitative x-ray microanalysis using an energy dispersive x-ray detector, and quantitative x-ray microanalysis using a wavelength dispersive x-ray detector (WDS). Standard methods of examination and analysis were employed [23]. Analyses were performed at a electron beam accelerating potential of 15 keV and a current of ≈30 nA. The electron microprobe was calibrated to perform quantitative DS analysis using the Al Kα and Cr Kα x-ray lines. Peaks and background positions for the Al and Cr x-ray lines were determined by performing wavelength scans over the peak and background regions in laboratory standards of Cr2O3 ad Al2O3 (primary standard). A hypersthene sample from the Smithsonian was analyzed for Cr and used as a secondary standard. Three replicate analyses were performed on standard hypersthene and 16 replicate analyses were performed on the ruby pellet. Data was processed using the ϕ(ρ z) correction of Armstrong [23] with the correction program CITZAF [24].

3.4.2.2 EDS Analysis of the Ruby Spheres

The secondary standard was used for analysis in the remaining 14 ruby spheres. For this, samples were carbon coated and analyzed in an AMRAY 1400 SEM, with the accelerating potential set at 15.8 kV, as confirmed by measurements of the upper energy limit of the continuum. Beam current, measured with a Faraday cup, was maintained at 1.0 nA. Sample inclination was 45° and x-ray take-off angle was 41°. The secondary standard, an analyzed ruby sphere with 0.44 mass fraction % Cr2O3 was used. X-ray data were collected with an HNU detector coupled to a 4 Pi Analysis digital beam control and data acquisition interface. During analysis, the beam was rastered over an area ≈100μm × 100 μm in size. Data was reduced using the conventional methods [25] with the aid of the DTSA software package [26].

3.4.3 Mosaic Spread of the Ruby Spheres

One ruby was chosen randomly to analyze for the diffraction peak width using a double crystal diffractometer equipped with a Ge monochromator and analyzer that has an intrinsic crystal resolution of 0.005° (determined from the full width half maximum (FWHM) of a piece of float- zoned silicon). 2θ/θ-scans and rocking curves of the (300), (006), and (104) reflections of the rubies are recorded. The rocking curve scans measure the mosaic spread of the spheres, while the 2θ/θ scans show lattice strains.

3.4.4 Guinier-Hägg Transmission Powder Technique

The Guinier-Hägg transmission technique was used as a second method for determining the lattice parameters. The principle of this technique was reported elsewhere, and details of this study will be reported separately. The Guinier-Hägg focussing x-ray powder camera [27] with the asymmetric geometry installed at the U.S. Geological Survey was manufactured by Incentive Research and Development AB (Stockholm, Sweden). It has been in continuous use until now, yielding x-ray powder data of high quality with trouble-free maintenance and utility and providing great sensitivity for recording weak lines. Prior to the measurements of the samples, the camera was aligned and calibrated. The powder pattern of SRM 640b was measured to check the cylindrical form and the adjustment of the camera. The measured cell parameter was 5.4307(3) Å, which agrees with the reported certified value of a = 5.43094(5) Å to within 2 standard deviations.

Five samples were prepared for this study. In each of five preparations, about 12 spheres (recovered from the original mounts) were crushed in a small agate mortar under toluene, but not ground. The powdered material (about 5 mg) was transferred to the planchet and mixed with silicon powder (NIST SRM 640b [28], untreated) as an internal standard. In all, about 60 spheres were consumed in the procedure. Samples are prepared by sticking the ruby powder onto a piece of Scotch “Magic Tape” placed over a 6 mm hole in a 21 mm diameter, thin aluminum plate or planchet. With the sample rotated, rather coarsely ground powder gives sharp, uniform diffraction lines on the film. The tape gives a weak, diffuse background, but no distinct lines on the film.

The 8 in × 1/2 in films used were Kodak Type SB5 and were measured with a NONIUS film viewer. In order to correct for the film shrinkage problem, Hägg et al. (1947) [29] has designed an automatic correction procedure. Before development the films are exposed in a special light box through a millimeter scale on a photographic glass plate (made by Zeiss), so that a scale graduated in tenth millimeters is recorded directly on the film parallel to the x-ray pattern. The film measurements were converted to Bragg angles (2θ) by multiplying the millimeter scale readings (after subtracting the measured zero reading), by a factor k, which is determined at six different angles by measurement of the Si lines. To test the trend of k, a sample composed of Si plus quartz was exposed, giving 6 Si lines and 27 quartz lines. Using the central 4 Si lines to standardize k, the quartz unit cell was determined by least-squares analysis of the determined 20 values. This led to unit cell parameters for quartz a = 4.9121(3) Å, c = 5.4033(6) Å, with the standard deviations for 2θ of 0.012°. These parameters were then used to calculate the expected 2θ values for quartz, and from these the corresponding k value for each line on the observed pattern. From these data it is clear that k remains constant within the standard error of measurement, except above 85°. Using this method, for each ruby sample k was determined by averaging k values obtained from the four central lines of Si. The resulting set of 2θ values for 9 to 11 lines of ruby was used to obtain the best unit cell parameters by using the least-squares program of Appleman and Evans [30]. This program automatically indexes at each cycle the input 2θ data (starting with an approximate unit cell) and yields a set of unit cell parameters (including cell volume) with their standard errors, the standard error of observed data of unit weight, and the variance-covariance matrix.

3.4.5 Crystallographic Structural Parameters

Relevant information on the ruby spheres such as the structural parameters have been obtained using four ruby spheres and the Picker diffractometer at NRC, Canada. For this study, data was collected with the θ/2θ scan technique at a rate of 4°/min and profile analysis was applied. Graphite monochromatized MoKα radiation was used with the x-ray tube operated at 50 kV, 30 mA and 2θmax = 120° for the ruby spheres and 90° for Al2O3. Five sets of intensity data were collected using the data collection routine DIFRAC [17].

3.5 Results

3.5.1 Chromium Content of the Ruby Spheres

Using the electron microprobe technique, it was found that the measured value for Cr in the Smithsonian hypersthene was 0.53±0.02 (standard deviation) mass fraction % (compared to the nominal value of 0.51 %). The mean concentration and standard deviation of the 16 replicate measurements for Cr in the secondary standard (ruby pellet) was 0.44±0.04 mass fraction % (Table 3). The measurement precision of Cr (for both sample and secondary standard), based on the counting statistics is 0.02 mass fraction %. Considering the degree of surface relief and a small amount of charging that occurred on the ruby sample, the variation in measured Cr concentration is consistent with the sample being homogenous in composition. This sphere was found to have relatively large areas of microscopic surface roughness. The amount of variability of the measured concentration of the major Al in the sample is indicative of the surface roughness. Figure 13 shows the analysis points of the secondary ruby sphere standard.

Table 3.

Results of analysis points on the standard sphere

No. Mass fraction, % Mole fraction, %
Al Cr Al Cr
1 50.95 0.51 99.48 0.52
2 51.28 0.43 99.57 0.43
3 51.09 0.52 99.47 0.53
4 52.11 0.49 99.51 0.49
5 50.71 0.38 99.61 0.39
6 51.72 0.41 99.59 0.41
7 51.39 0.40 99.60 0.40
8 52.25 0.41 99.59 0.41
9 51.27 0.42 99.58 0.42
10 51.48 0.44 99.56 0.44
11 51.53 0.44 99.56 0.44
12 51.78 0.45 99.55 0.45
13 51.88 0.43 99.57 0.43
14 51.62 0.43 99.57 0.43
15 54.26 0.48 99.54 0.46
16 51.05 0.40 99.60 0.40

Average 51.65 0.44 99.56 0.44
Standard dev. 0.41 0.04 0.41 0.04
Fig. 13.

Fig. 13

Analysis points of the secondary ruby sphere standard.

Table 4 shows the measured values of the Cr content of 15 spheres using the SEM/EDS method. Among them, sphere No. 15 is the secondary standard. The chromium composition was found to be relatively ho mogenous and was estimated to be 0.42 mole fraction %±0.011 % (expanded uncertainty). Figure 14 shows a typical EDS spectrum of one of the measured spheres. The peaks corresponding to Cr Kα1+Cr Kα2, and Cr Kβ are shown.

Table 4.

Chromium content of the ruby spheres (mole fraction, %)

No. Al Cr
1 99.59 0.41
2 99.58 0.42
3 99.55 0.45
4 99.60 0.40
5 99.57 0.43
6 99.59 0.41
7 99.55 0.45
8 99.58 0.42
9 99.61 0.39
10 99.57 0.43
11 99.56 0.44
12 99.59 0.41
13 99.60 0.40
14 99.60 0.40a
15 99.56 0.44

Average 0.42
Standard deviation 0.019
Expanded uncertainty 0.011
a

Secondary standard.

Fig. 14.

Fig. 14

Energy dispersive spectrum showing the Cr Kα1, Kα2, and Kβ lines of a typical ruby sphere.

3.5.2 Mosaic Spread of the Ruby Spheres

The x-ray rocking curves of the (006), (300) and (104) reflections were rather narrow, with FWHM from 0.007° to 0.012°, indicating a very small mosaic spread within the crystal. The variation of peak width indicates a small anisotropic mosaicity. The 2θ/θ scans of the (300), (104) and (006) reflections show narrow FWHMs of 0.027°, 0.0113°, and 0.0106°. From the Scherer formula [31], an average coherent crystalline size can be estimated, giving approximately 3600 Å for the distance perpendicular to the (300) and about twice this number, 8100 Å for distances perpendicular to (006). Therefore crystal growth of the ruby crystal appears to be anisotropic, with corresponding mosaicity due to the shape of these grains. Figures 1517 show the rocking curves of reflections (300), (104) and (006) of a typical ruby sphere using a double crystal spectrometer. In general, negligible residual strain was found in the sample, which is indicative of the good quality of these spheres, fulfilling another requirement for being a good standard.

Fig. 15.

Fig. 15

Rocking curves of a typical ruby crystal using a double crystal diffractometer. The reflection (300) is shown.

Fig. 16.

Fig. 16

Rocking curves of a typical ruby crystal using a double crystal diffractometer. The reflection (104) is shown.

Fig. 17.

Fig. 17

Rocking curves of a typical ruby crystal using a double crystal diffractometer. The reflection (006) is shown.

3.5.3 Ruby Spheres as a SRM

Statistically, the lattice parameter measurements of the four groups of data (41 sets together) agree with each other within an acceptable range. The structural parameters of four ruby spheres determined using the Picker diffractometer are shown in Table 5. Small residual factors (RF < 2 %, and RW < 3 %) indicate correctness of the structure model and good quality of crystals. All atomic positions and thermal parameters are very close to each other, with a difference in general less than 2σ’s. For crystal C2-2, although the extinction and scale parameters differ somewhat, the other parameters are unaffected; the range of scale factors is consistent with the range of diameters. This result shows the corrections applied for the different measurements result in consistent structural parameters.

Table 5.

Structural parameters for four ruby spheres and one alumina crystal (auxiliary information, not certified values), measured with a Picker diffractometer at NRC Canada using Mo radiation (space group, R3¯c.)

Sphere No. C2-1 C2-2 C2-3 C2-4 Al2O3
No. refls. measured 880 886 1039 942 2844
No. observed/unique 414/446 415/446 414/446 410/446 237/242
Al z 0.35227(2) 0.35229(3) 0.35225(2) 0.35226(2) 0.35129(1)
u11 0.00312(8) 0.00309(9) 0.00306(8) 0.00300(9) 0.00338(5)
u13 0.00355(10) 0.00361(11) 0.00343(10) 0.00349(11) 0.00335(5)
O x 0.69374(11) 0.69373(13 0.69380(13) 0.69382(12) 0.69367(5)
u11 0.00369(12) 0.00364(14) 0.00362(12) 0.00356(13) 0.00359(6)
u33 0.00369(13) 0.00364(15) 0.00352(13) 0.00366(14) 0.00394(8)
u12 0.00173(13) 0.00166(16) 0.00170(14) 0.00164(15) 0.00166(7)
u13 0.00029(6) 0.00030(7) 0.00029(6) 0.00029(7) 0.00035(3)
Scale 0.2673(1) 0.2472(1) 0.2820(1) 0.2792(1) 0.1496(4)
Ext. (μm) 0.76(3) 0.32(2) 0.75(3) 0.66(3) 0.52(3)
RF 0.0158 0.0178 0.0152 0.0162 0.0134
Rw 0.0257 0.0318 0.0262 0.0284 0.0068

Before we present the result of lattice parameter determinations, various systematic errors that are associated with these measurements need to be addressed first.

3.5.3.1 Systematic Error Corrections

The systematic corrections investigated include thermal expansion, absorption and eccentricity, horizontal divergence, vertical divergence, and refraction. Among them, only thermal expansion and refraction corrections were applied. Other corrections were too small to be considered.

Using the capabilities of single crystal diffractometer to bring any (hkl)-plane into the reflection position in more than just one setting, systematic errors due to zero-point offsets, sphere of confusion, sample displacements and absorption can be minimized. Such procedures depend on the diffractometer geometry and have been described elsewhere [3].

Thermal effect

Among all factors affecting the lattice parameters of the rubies, thermal expansion appears to have the most significant effect. Because of the small amount of Cr, the thermal expansion coefficient of ruby was assumed to be the same as Al2O3. The anisotropic thermal expansion coefficient of Al2O3 was reported by Campbell and Grain [9], and was more recently reviewed by Munro [32]. The thermal expansion coefficient curve is quadratic in nature (up to 1800 °C). For data near the room temperature range, it can be assumed to be linear. Within the range of 0 °C to 100 °C the a and c axes of α-Al2O3 were found to expand linearly but anisotropically [32]. Under these conditions, the corrections can be calculated according to the following expression:

a(25°C)=a[1+αa(T25)],whereαa=5.0×106c(25°C)=c[1+αc(T25)],whereαc=6.66×106

[8]The data collection of these experiments took place in a range of 19 °C to 26 °C. Based on the above equations, corrections to 25 °C have been applied.

Refraction

There are two contributions to the refraction correction [10]. One corresponds to the Snell’s law correction: δd/d = −v cotθ = (1−n) cotθ. This part of correction is too small to be included. The other part, refraction due to change of wavelength is 1−n = (2.71 × 10−6) ρλ2 (ΣZ/Σa). In this equation, ρ is the density (taken as 3.98 g/cm3 [7], λ is the wavelength, Z is the atomic number and a is the atomic weight. In the expression δd/d = (1−n)/n, (1−n) = 2.69×10−6 for Mo radiation, and (1−n) = 1.27×10−5 for Cu radiation.

Absorption

According to Hubbard and Mauer [10], the absorption correction for the Si crystal that they studied is about the value of 0.000079 Å in 5.43 Å, or 15×10−6 for Si at 2θ of 80°. An estimation shows that this value will be much smaller in the case of the ruby due to the following two reasons. (1) The Si spheres were 0.25 mm in diameter, whereas the rubies are only about 0.15 mm, therefore the volume of the rubies are about 0.14 that for Si, and (2) for Cu radiation, the μ value is 141 cm−1 for Si, but 124 cm−1 for ruby. This indicates that in the θ range of 0° to 80°, the transmission factor ranges from approximately 0.1 to 0.2 for Si, and from 0.32 to 0.39 for Al2O3. Absorption was estimated to affect the lattice parameters of the ruby in the order of ≈10−5 using Cu radiation, and negligible for Mo radiation. Accurate absorption correction is a non-trivial process. But because of the small magnitude, the absorption correction was not applied.

Divergence of x-ray beams

In the equation δd/d = (−2cosθ/λ) δθ = Av2/6 [10], since the collimator used has a small aperture, the Av value is of an approximate value of 0.05°, and the δd value is ≈10−6 Å. The axial divergence is a negligible source of error compared with the other causes. These values are too small in magnitude to be measured.

3.5.3.2 Lattice Parameter Determinations

A total of 45 sets of 4-circle diffractometer data were measured and analyzed. Measurement results are listed in Table 6. There are a total of four subsets of data. Consistency of diffractometer measurements were performed by cross comparison of data measured with these diffractometers using the same spheres. Satisfactory data were obtained for these comparisons (sample pairs of 2a,2b; 5a,5b; 6a,6b; 29a,29b; L1–17, L2–17; L1–22, L2–22). The first set was measured using CuKα radiation at the NRC of Canada on a CAD4 diffractometer (C1). The second set of data was also measured at NRC Canada using Mo Kα radiation, but with a Picker diffractometer (C2). The third subset was measured at Bell Laboratories with Mo radiation (L2), and the fourth one was also measured at Bell Laboratories using Cu radiation (L1).

Table 6.

Lattice parameters for 39 ruby spheres (SRM 1990). A total of 45 measurements were performed using four units of single crystal diffractometers (C1: Enraf-Nonius at NRC Canada, C2: Picker at NRC Canada, L1 and L2: Enraf-Nonius at Lucent Technologies), after applying both thermal expansion and refraction corrections

No. ID Rad Type a (Å) c (Å) T (°C)
1 C1-1 Cu 4.760686(51) 12.99513(15) 20.4
2 C1-2a Cu 4.760303(65) 12.99450(17) 20.2
3 C1-2b Cu 4.760400(54) 12.99462(14) 20.3
4 C1-3 Cu 4.760597(63) 12.99478(17) 20.7
5 C1-4 Cu 4.760561(51) 12.99500(14) 20.2
6 C1-5a Cu 4.760584(67) 12.99468(19) 22.6
7 C1-5b Cu 4.760577(60) 12.99492(17) 22.6
8 C1-6a Cu 4.760836(58) 12.99572(16) 19.5
9 C1-6b Cu 4.760830(67) 12.99601(19) 21.0
10 C1-7 Cu 4.760592(58) 12.99525(16) 20.4
11 C1-8 Cu 4.760521(59) 12.99469(16) 19.2
12 C1-9 Cu 4.760559(54) 12.99496(15) 19.0
13 C1-10 Cu 4.760608(63) 12.99473(17) 20.0
14 C1-11 Cu 4.760669(74) 12.99459(23) 19.9
15 C1-12 Cu 4.760406(62) 12.99460(14) 19.9
16 C2-1 Mo 4.760924(30) 12.99609(16) 26.2
17 C2-2 Mo 4.761024(60) 12.99633(16) 26.2
18 C2-3 Mo 4.760924(60) 12.99573(16) 26.2
19 C2-4 Mo 4.761044(60) 12.99655(14) 26.2
20 L2-1 Mo 4.760795(70) 12.99575(20) 20.6
21 L2-3 Mo 4.760705(59) 12.99567(23) 20.6
22 L2-17 Mo 4.760809(62) 12.99578(17) 19.4
23 L2-18 Mo 4.760805(62) 12.99578(16) 19.4
24 L2-22 Mo 4.760695(79) 12.99576(27) 19.3
25 L2-24 Mo 4.760655(66) 12.99568(18) 19.3
26 L2-25 Mo 4.760897(83) 12.99507(30) 19.0
27 L2-27 Mo 4.760487(80) 12.99501(16) 19.0
28 L2-28 Mo 4.760640(98) 12.99579(20) 19.2
29 L2-29a Mo 4.760700(81) 12.99564(17) 19.2
30 L2-29b Mo 4.760730(80) 12.99574(17) 19.2
31 L1-2 Cu 4.760881(98) 12.99597(26) 25.8
32 L1-4 Cu 4.761081(94) 12.99629(28) 26.0
33 L1-5 Cu 4.760788(75) 12.99572(20) 25.8
34 L1-7 Cu 4.761045(71) 12.99629(20) 25.8
35 L1-8 Cu 4.760853(75) 12.99581(21) 25.8
36 L1-9 Cu 4.761005(84) 12.99608(22) 25.8
37 L1-10 Cu 4.760944(67) 12.99628(18) 25.8
38 L1-11 Cu 4.760879(80) 12.99607(22) 22.3
39 L1-12 Cu 4.760953(79) 12.99618(21) 22.5
40 L1-14 Cu 4.761009(82) 12.99613(22) 25.8
41 L1-15 Cu 4.760749(110) 12.99564(32) 22.2
42 L1-16 Cu 4.760839(65) 12.99573(19) 22.5
43 L1-17 Cu 4.761067(84) 12.99633(18) 25.8
44 L1-22 Cu 4.760841(64) 12.99555(18) 25.8
45 L1-23 Cu 4.760998(112) 12.99612(30) 25.8
3.5.3.3 Statistical Analysis of the Lattice Parameter Results

The certified value and its uncertainty represents a consensus of the four sets of measurements. Each of the four sets of measurements for a particular lattice parameter (a and c) is reduced to a mean value. The certified value for the lattice parameter is the mean of the resulting four mean values. The standard uncertainty of the certified value is the standard error of the mean, which is equal to the sample deviation of the four mean values divided by the square root of 4. The associated degree of freedom is 3 and the corresponding coverage factor is equal to the standard uncertainty times the coverage factor. The interval contains the lattice value with 95 % level of confidence. Readers are to refer to reference [37] for the expression of uncertainty in measurement for details on the procedure and terminology.

The certified value of the lattice parameters (mean value) are a = 4.76080 ± 0.00029 Å (expanded uncertanties), and c = 12.99568 ± 0.00087 Å. The interval defined by the estimate the expanded uncertainty provides an interval with an approximate 95 % level of confidence. It is equal to the standard uncertainty times a coverage factor, k (k = 3.18). The standard uncertainty represents the standard deviation of the estimate. The coverage factor k accounts for the degrees of freedom in the estimation of the uncertainty. These values fall well within the results obtained from the international round-robin study (a = 4.7608 Å ± 0.0062 Å [expanded uncertainties], c = 12.9979 Å ± 0.020 Å). However, the expanded uncertainties are significantly smaller in our certified values. Therefore the round-robin data can not be used as certified data, only as a comparison. The five significant digits reported for these certified lattice parameters is more precise than the four obtained in typical laboratories. This batch of ruby spheres therefore will serve well as a diffractometer alignment standard.

Figures 18 and 19 display the four sets of measurements, the certified value, and the expanded uncertainty for the lattice parameters a and c, respectively. For both lattice parameters, the expanded uncertainty interval contains the means of the four sets of measurements.

Fig. 18.

Fig. 18

The four sets of measurements, the certified value, and the expanded uncertainty for the lattice parameters a using single crystal diffractometers. The expanded uncertainty interval contains the means of the four sets of measurements.

Fig. 19.

Fig. 19

The four sets of measurements, the certified value, and the expanded uncertainty for the lattice parameters c using single crystal diffractometers. The expanded uncertainty interval contains the means of the four sets of measurements.

The measurement procedure described above strives to compensate for differences between diffractometer. However, total elimination of hardware influences cannot be achieved under any circumstances. Examination of the data from the similar ENRAF-NONIUS CAD-4 diffractometers shows a small residual offset of the unit cell parameters measured, even though a similar procedure using the same high-level software was used. Since the control over the two experiments in different locations was not absolute, small differences are expected. We can speculate on the influence of some differences: for instance, the diffractometers have slight differences in their respective low-level positioning and automation system, affecting the positioning feedback system. The quality of the x-ray optical elements and their alignment will also influence the peak profiles of the diffracted intensities. Environmental variables such as air temperature and air pressure are also going to affect the results, even though care was taken to minimize their influences. However, the temperature variamtions in the rooms and in radiation safety enclosures could not be fully taken into account due to varying conditions during the measurements and physical distance between the diffractometers used.

A comparison of the ruby cell parameters with the alumina data [33-35] indicates that the rubies have larger lattice constants. The increase in cell volume of the ruby spheres over that for pure corundum is expected since the effective ionic radius of Al3+ (0.535 Å) is smaller than that of Cr3+(0.615 Å) [36]. Table 7 shows reported cell parameters for alumina. The lattice parameters of the alumina from Ref. [18] have a small error associated because there was only one sample.

Table 7.

Literature lattice parameters of alumina (and of rubies for comparison)

Material Reference a (Å) c (Å)
Alumina Ishizawa et al. [33] 4.754(1) 12.99(2)
Alumina Morris, et al. [34] 4.7588(1) 12.992(1)
Alumina Newnham et al. [35] 4.7589 12.991
Alumina This work [18] 4.75999(3) 12.99481(7)
SRM rubies This work 4.76080(29) 12.99568(87)
SRM rubies This work (Guinier-Hägg) 4.76093(31) 12.9959(23)

3.5.4 Guinier-Hägg Transmission Camera Data

The Guinier-Hägg powder data of the five ruby samples show good agreement with the single crystal data, and thus confirm the certified lattice parameters are free from systematic errors (Table 8). After thermal expansion and refraction corrections were applied (discussed below), the mean values of these lattice parameters are a = 4.7610 ± 0.0013 Å (expanded uncertainty), and c = 12.9954 ± 0.0034 Å.

Table 8.

Crystallographic Data for the ruby spheres using the Guinier-Hägg camera technique, data taken at 22 °C

Sample No.
I II III IV V
a (Å)  4.76091(47)  4.76073(20)  4.76117(48)  4.76053(67)  4.76065(44)
c (Å)   12.9977(14)   12.99577(53)   12.9967(13)   12.9937(16)   12.9934(15)
V3) 255.14(4) 255.08(5) 255.15(6) 255.02(7) 255.03(5)

Average with thermal and refraction correction:

a = 4.7610 ± 0.00013 Å (expanded uncertainty)

c = 12.9954 ± 0.0034 Å (expanded uncertainty)

Various possible errors involved with the Guinier-Hägg technique that may affect the accuracy of cell parameters include errors due to x-ray beam divergence, film shrinkage and thermal expansion. Other problems caused by specimen displacement, transparency effect, and absorption can mostly be compensated by using the internal standard, Si 640b [28].

Thermal Expansion Effect

The same corrections as those for the single crystal diffractometer data were applied here. The corrections were calculated according to the following expression:

a(25°C)=a[1+αa(25T)],whereαa=5.0×106c(25°C)=c[1+αc(25T)],whereαc=6.66×106

[8]The data collection of these experiments took place at 21°C. Based on the above equations, a correction of 4° (21° to 25°C) was applied.

Refraction

Similar corrections as that for the single crystal diffractometers was applied here. In the expression δd/d = (1−n)/n, (1−n) = 2.69 × 10−6 for Mo radiation, and (1−n) = 1.27 × 10−5 for Cu radiation. The corrections are also of relatively small quantity.

Absorption

The absorption coefficient of Al2O3 (124 cm−1) and Si (141 cm−1) is of similar order of magnitude, and the difference between the Cr-doped Al2O3 is expected to be < 17 cm−1. Therefore the difference in the d-spacing aberration of the rubies due to absorption can be largely compensated for by mixing in with the Si standard, 640b [28].

Sample Displacement

A condition for the camera to give sharp interference lines on a film is that the powder sample is situated on the cylinder defined by the film. Since in the actual procedure, the sample is adhered to a piece of thin scotch tape which is in turn attached to a flat metal ring rotated in a plane tangent to the focusing cylinder, a minor deviation of the sample from the correct location causes a slight broadening of the interference lines and displacements of their positions on the film. This problem can again be largely corrected for by mixing the sample with the Si standard [28].

Sample Transparency

The extraordinary focusing property of the Guinier-Hägg transmission camera also eliminates to a large extent the influence of the thickness of the specimen. As pointed out by Jenkins and Snyder [38], this error can be corrected for by intimate mixing with Si as an internal standard [28].

Divergence of the X-ray Beam

Axial divergence is expected to be a negligible source of error compared with the other causes. Klug and Alexander [31] have shown that because of the axial divergence of the beam, the center of the blackening of a line in a powder camera will be shifted by a small angular amount, δθ= −[(1+x)/96] (H/R)2 cot 2θ, where R is the radius of the film, H is the axial divergence of the beam in a camera of radius R. If the beam is homogenous, x = 0. Since the collimator used has a small aperture, the H/R value is small and the correction is trivial. For example, with a H/R value of 1/50, a δθ of 0.0004° is obtained at 2θ at 30°, and a δθ of 0.00004° is obtained and at 80° 2θ. These values are much too small in magnitude to be measured, and can be ignored.

3.6 Discussion

In recent years, due to the advances of computer and x-ray diffractometry technology, x-ray diffractionists have a great number of venues of obtaining data with adequate accuracy. In order to achieve high accuracy in the range of 0.001 % to 0.0001 %, extra effort is needed. In the following, a list of factors that may affect the accuracy of lattice parameter determination using a single crystal diffractometer is given.

Based on the results of the round-robin project and the current cell parameter certification project, the use of SRM 1990 is expected to enhance the alignment capability of single crystal x-ray diffractometers in industrial, government and academic x-ray laboratories. The recommended use of the SRM will also be discussed.

3.6.1 Factors Affecting Accuracy of Single-crystal Diffractometer Alignment and Lattice Parameter Determination (Serial Diffractometers Only)

  1. The precision of diffractometer gears, spindle pitch, encoders, etc. are important for achieving accurate measurements, and diffractometers must be carefully adjusted to avoid mechanical problems.

  2. The goniostat must be well aligned, with a sphere of confusion as small as possible. Typical values for a modern diffractometer are of the order of 10 μm to 20 μm. This is the maximum wobble of the sample when it is rotated about all the axes (When its reflections are aligned in the scattering plane). For a κ-axis system, the sphere of confusion indicates how close the four axes intersect in one point. For an Eüler system, the sphere of confusion describes how close to a circle the Eüler cradle is, and how well the three remaining axes intersect in one point. Usually the κ-axis system tends to have a smaller sphere of confusion. Calibration constants are supplied with the goniometer and the aperture system, and should be checked carefully.

  3. The major task of the alignment procedure is to direct the primary x-ray beam through the center of the goniometer and through the center of the receiving aperture when positioned at θ = 0°. The beam must be carefully centered. The routine to center a reflection is important. It is recommended from the international round-robin result that the alignment routine should incorporate the King and Finger’s algorithm [3, 4, 39] for calculating various alignment corrections such as offsets of detector, x-ray tube in both horizontal and vertical direction, the angular offsets of the monochromator, and the offsets of the crystals from the center of the primary beam, X,Y, and Z direction.

  4. Reflections with high angles are important to achieve accuracy. As in the Bragg equation, δd = (−n/2)(cotθ/sinθθ, it can be seen that the error in δd is relatively small when higher angle reflections are used. In the case of the ruby spheres, there are a great number of strong reflections at high angles, therefore the usual concerns of disadvantages of back reflections, namely, low intensity, lower peak-to-background ratio and broadening by wavelength dispersion will not be a problem here. Table 9 lists the high angle reflections that can be used for initial orientation, and for accurate alignment of the SRM.

  5. Most diffractometer programs assume the intensity ratio in the α12 doublet to be 2:1. This is true only if the diffractometer optical system is achromatic (slit/collimators). If a monochromator is used, then a deviation from the 2:1 intensity is likely [40]. Applying the wrong ratio will therefore give an incorrect value for the position.

  6. The profile shape of reflections chosen has a strong influence on the accuracy of angle measurement. Therefore a profile analysis routine is critical for determining the correct peak positions. Different software may use different methods, i.e., centroid of the peak, full-width-half-maximum, and mid-chord method, etc. The profile-fitting program must be chosen and used carefully.

  7. If the routine of eight-reflection is not used, then systematic errors such as the 2θ0 correction should be applied for obtaining accurate 2θ values.

  8. Different computer programs that perform least-squares lattice parameter refinements may use different routines. One should understand these routines and how the associated errors are calculated. This is important for inter-laboratory comparison of lattice parameters.

Table 9.

High-angle 2θ reflections for obtaining both initial orientation of the spheres and for accurate alignment of diffractometers. M is the multiplicity, the h k l values are the Miller indices and Fc is the calculated structure amplitude.

h k l M Fc 2 θ (CuKα1) (°) 2θ(MoKα1) (°)
(A) For initial orientation
 0 1 2 6 47 25.567 11.694
 1 0 4 6 83 35.139 15.978
 1 1 0 6 61 37.762 17.137
 0 0 6 2 13 41.665 18.848
 3 0 0 6 141 68.180 29.910
 0 0 12 2 57 90.678 38.232
(B) For accurate alignment
 1 3 4 12 51 91.144
 2 2 6 12 74 95.203
 0 4 2 6 38 98.342
 2 1 10 12 64 101.031
 4 0 4 6 38 103.262
 3 1 8 12 36 110.931
 2 2 9 12 30 114.010
 3 2 4 12 64 116.030
 0 1 14 6 52 116.553
 4 1 0 6 47 117.777
 4 1 3 12 26 121.956
 1 3 10 12 66 127.605
 3 0 12 6 41 129.802
 2 0 14 6 58 131.031
 4 1 6 12 56 135.971
 1 1 15 12 29 142.225
 4 0 10 6 69 145.049
 0 5 4 6 56 149.060
 1 2 14 12 52 149.986
 1 0 16 6 32 150.300
 3 3 0 6 79 152.239
 0 4 20 6 42 80.363
 5 1 16 12 28 80.798
 7 1 0 6 26 80.998
 7 0 10 6 50 82.767
 3 3 18 12 36 83.234
 1 7 6 12 29 84.097
 1 1 24 12 41 84.397
 5 4 4 12 32 85.789
 6 3 0 6 42 86.120
 0 8 4 6 32 88.336
 4 2 20 12 26 90.584
 2 2 24 12 37 92.038
 2 6 14 12 34 93.525
 8 1 4 12 32 95.978
 3 4 20 12 38 98.243
 6 1 20 12 30 103.425
 4 6 10 12 36 105.883
 7 3 10 12 41 108.546
 0 7 20 6 39 108.721
 0 0 30 2 44 109.913
 5 2 24 12 29 115.794
 3 0 30 6 40 118.280
 1 9 10 12 26 119.713

3.6.2 Recommended Use of SRM 1990

In order to obtain reliable data, it is recommended that users perform regular alignment checks of the diffractometers using the SRM ruby spheres. The spheres should be mounted using a minimal amount of adhesive material and on the center of the tip of a fiber or preferably a capillary with a diameter slightly smaller than that of the spheres (i.e., <0.1 mm). The crystal should be kept mounted permanently on a goniometer head (i.e., devote one goniometer head to that purpose). The orientation matrix should be known and the results of alignment should be kept and used for comparison later. A complete set of intensity data should be collected and the structure refinement results should yield a low residual value equivalent to those in Table 5.

It is easy to center the ruby sphere so that the misalignment is minimal. The center of the ruby sphere then defines the center of the diffractometer. Maximizing the intensity of a given reflection ensures that the highest intensity part of the beam intercepts the ruby sphere.

During the alignment process, high angle reflections should be used. Relatively low angle reflections such as 006, 0012, and 300 can be used for initial orientation of the spheres [(Table 9, part (A)]. Further accurate alignment can be carried out with the reflections as indicated in Table 9 [part (B)]. In this table, the Miller indices, multiplicity of the reflection, the calculated structure amplitudes, Fc, and the 2θ (CuKα1) and 2θ (MoKα1) are listed. The 2θ and Fc values were calculated based on the certified lattice parameters and structure of sample C1-1. The Fc values are not certified, but are for reference purposes.

In addition to being excellent standards for alignment of conventional diffractometers, these ruby spheres will also be a valuable standard for instrument calibration for diffractometers equipped with CCD detectors. CCDs and other area detectors are a recent addition to the field of crystallography. With these instruments, the diffracted intensities are measured over a large solid angle. The diffracted plane then varies and is no longer well defined as is the case for a serial diffractometer. For an achromatic system, the doublet splitting is symmetrical with respect to the center of the diffractometer. If an incident beam monochromator is used, then the system is asymmetrical, giving sharp reflection on one side where the dispersion is compensated (non-dispersive side), while giving broad reflections the opposite side (dispersive side). For a calibration of the flat CCD, the distance of the detector to the sample has to be determined accurately, then the flat field is unwarped and mapped onto a spherical detector. This process will benefit from an accurate lattice parameters standard, since both the distance, the beam offsets, roll and yaw of the detector can be determined using the ruby standard.

4. Summary

X-ray structural determinations using automatic data collection and structure solution schemes require accurate initial cell parameter data. Until now, no certified standard was available for the evaluation of the diffractometer condition, alignment and inter-laboratory comparison of data. The result of this work is expected to enhance the alignment capability of single crystal x-ray diffractometers in industrial, government and academic x-ray laboratories. Therefore the success of this project will have a significant impact on accurate scientific investigations using single crystal diffractometers.

The lattice parameter is being certified as a = 4.76080 ± 0.00029 Å (expanded uncertainty), and c = 12.99568 ± 0.00087 Å. Five different samples of powdered rubies were measured on a Guinier-Hägg transmission camera. The values of a=4.7610 ± 0.0013 Å (expanded uncertainty), and c = 12.9954 ± 0.0034 Å give good agreement with the values obtained from the single crystal spheres. Among all systematic errors, only the thermal acorrection and refraction corrections were applied. the auxiliary data on the Cr-content will also be useful for microanalytical calibrations.

Acknowledgments

The authors would like to thank the Office of Standard Reference Materials of the National Institute of Standards and Technology, American Crystallographic Association, and International Union of Crystallography for partial financial support. We would also like to thank Juan Graces of the University of Toronto for his genrous contribution of zeolite crystals.

Biography

About the Authors: Winnie Wong-Ng is a senior research chemist in the Ceramics Division of the NIST Materials Science and Engineering laboratory. Her research interests include crystallography and phase equilibria of high-temperature oxides. Theo Siegrist is a senior member of the Technical Staff at Lucent Technologies. His crystallographic interests include inorganic structures, structural chemistry, thin films, and crystal growth. George T. DeTitta is a senior research scientist of the Hauptman-Woodward Medical Research Institute (formerly the Medical Foundation of Buffalo). His main research interests lie in the area of macromolecular crystal growth, charge density studies, and cryocrystallography. Larry W. Finger is a crystallographer who retired from the Geophysical Laboratory, Carnegie Institution of Washington about 2 years ago. Howard T. Evans, Jr. passed away last year. He was a research chemist with the U.S. Geological Survey. Eric J. Gabe has been retired from the Steacie Institute for Molecular Science, National Research Council (NRC) of Canada. He is still active in crystallographic research and remains as a guest scientist at the NRC. Gary D. Enright is an associate research officer of the Steacie Institute for Molecular Science, NRC of Canada. His research interests include x-ray diffractometry and structure of small molecules. John T. Armstrong is a research chemist in the Surface and Microanalysis Science Division of NIST Chemical Science and Technology Laboratory. His research interests include electron microprobe analysis, scanning electron microscopy, and Auger electron spectroscopy. Mark S. Levenson was a mathematical statistician at NIST for 7 years. Currently he is working in hazard analysis for the Federal Government. Lawrence P. Cook is a research scientist with the Ceramics Division of NIST. He is interested in phase equilibria and thermal analysis of electronic materials. Camden R. Hubbard is the leader of the Diffraction and Thermophysical Properties Group of the Metals and Ceramics Division, Oak Ridge National Laboratory. His current research includes x-ray and neutron diffraction, residual stress, and thermophysical properties of high-temperature materials.

Footnotes

1

Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

Contributor Information

W. Wong-Ng, Email: winnie.wong-ng@nist.gov.

J. T. Armstrong, Email: john.armstrong@nist.gov.

L. P. Cook, Email: lawrence.cook@nist.gov.

6. References

  • 1.Voitsekhovskii VN. Evaporation Figures of Corundum. Kristallografiya. 1968;13:563–565. [Google Scholar]
  • 2.Hamilton WC. Angle settings for four-circle diffractometers. International Tables for x-ray Crystallography. 1974;4:275–284. [Google Scholar]
  • 3.King HE, Finger LW. Diffracted Beam Crystal Centering and its Application to High- Pressure Crystallography. J Appl Crystallogr. 1979;12:374–378. [Google Scholar]
  • 4.Hazen RM, Finger LW. Comparative Crystal Chemistry, Temperature, Pressure, Composition and the variation of Crystal Chemistry. John Wiley & Sons; New York: 1982. pp. 46–56. [Google Scholar]
  • 5.Busing WR, Levy HA. Angle Calculations for 3- and 4-circle x-ray and neutron diffractometers. Acta Cryst. 1967;22:457–464. [Google Scholar]
  • 6.Kuperman A, Nadimi S, Oliver S, Ozin GA, Garceees JM, Olken MM. Non-aqueous synthesis of giant crystals of zeolite and molecular sieves. Nature. 1993;365:239. [Google Scholar]
  • 7.DeTitta G. Ruby Spheres Round-Robin Project Instruction Maunal. Hauptman-Woodward Medical Research Institute; Buffalo, NY 14203: [Google Scholar]
  • 8.Belyaev LM. In: Ruby and Sapphire. Russian, National Bureau of Standards, national Science Foundation, translator. Nauka; Moscow: Amerind Publishing Co. Pvt. Ltd; New Delhi, India: 1974. 1980. pp. 1–11. [Google Scholar]
  • 9.Campbell WC, Grain C. Thermal Expansion of Alpha Alumina. US Bur Mines Rep Invest. 1961:5757. [Google Scholar]
  • 10.Hubbard CR, Mauer FA. Precision and Accuracy of the Bond Method as Applied to Small Spherical Crystals. J Appl Cryst. 1976;9:1–8. [Google Scholar]
  • 11.Youden WJ. Graphical Diagnosis of Interlaboratory test results. 1969. (National Bureau of Standards Special Publication 300). [Google Scholar]
  • 12.Guide to NATA Proficiency Testing. National Association of Testing Authorities; Australia: 1997. [Google Scholar]
  • 13.Härtwig J, Hölzer G, Förster E, Goetz K, Wokulska K, Wolf J. Remeasurement of characteristic x-ray emission lines and their application to line profile analysis and lattice parameter determination. Phys Stat Sol (a) 1994;143:23–33. [Google Scholar]
  • 14.Kizler P, He J, Clarke JDR, Kenway PR. Structural relaxation around substituted Cr3+ ions in sapphire. J Am Ceram Soc. 1996;79(1):3–11. [Google Scholar]
  • 15.Mott NF, Littleton MJ. Conduction in polar crystals. Trans Faraday Soc. 1938;34(485):499. [Google Scholar]
  • 16.Wyckoff RWG. Crystal Structure. II. Interscience Publishers Inc; New York: Chap. V, 4. [Google Scholar]
  • 17.DIFRAC . software suite for data collection and reduction of x-ray single crystal diffractometers, developed by E Gabe et al. National Research Council (NRC); Ottawa, Canada: [Google Scholar]
  • 18.Calvert LD, Gabe EJ, Le Page Y. Ruby Spheres for Aligning Single-Crystal Diffractometers, IUCr Abstract 16. 2-02 (M-14) Ottawa, Canada: 1981. [Google Scholar]
  • 19.Busing WR, Henri AL. Angle Calculations for 3- and 4-circle x-ray and neutron diffractometers. Acta Cryst. 1967;22:457–464. [Google Scholar]
  • 20.Toraya H. The determination of unit-cell parameters from Bragg reflections data using a standard reference material but without a calibration curve. J Appl Cryst. 1993;26:583–590. [Google Scholar]
  • 21.Cohen ER, Taylor BN. The 1986 Adjustment of the fundamental Physical Constants. Rev Mod Phys. 1987;59:1121–1148. [Google Scholar]
  • 22.Jaroeswich E, Nelen JA, Noberg A. Reference samples for electron microprobe analysis. Geostand Newslett. 4:43–47. [Google Scholar]
  • 23.Armstrong JT. quantitative elemental analysis of individual microparticle with electron beam instruments. In: Heinrich KFJ, Newbury DE, editors. Electron Probe Quantification. Plenum Press; New York: 1991. pp. 261–315. [Google Scholar]
  • 24.Armstrong JT. CITZAF: A Package of correction programs for the quantitative electron microbeam x-ray analysis of thick polished materials, thin films and particles. Microbeam Anal. 1995;4:177–200. [Google Scholar]
  • 25.Heinrich KFJ. Electron Beam x-ray Microanalysis. Van Nostrand Reinhold Co; New York: 1981. p. 578. [Google Scholar]
  • 26.Fiori CE, Swyt CR, Myklebust RL. NIST/NIH Desktop Spectrum Analyzer Program and x-ray Database. 1991. (NIST Standard Reference Database No. 36). [Google Scholar]
  • 27.Hägg G, Ersson NO. (1960), An Easily Adjustable Guinier Camera of Highest precision. Acta Cryst. 1960;A25(Sppl):S64. [Google Scholar]
  • 28.Hubbard CR, Robbins C, Wong-Ng W. Standard Reference Material 640b (1987), Silicon Powder x-ray Diffraction Standard, obtainable from the NIST Office of Standard Reference Materials. Gaithersburg, MD 20899: Current price will be quoted on request. [Google Scholar]
  • 29.Hägg G. Measurement of x-ray Powder Diffraction Films with automatic Correction for Shrinkage. Rev Sci Instrum. 1947;18(371):377. [Google Scholar]
  • 30.Appleman DE, Evans HT., Jr . US Dept of Commerce, Natl Tech Inform Serv. National Technical Information Service; 5285 Port Royal Rd., Springfield, VA 22151: 1973. (Publ. No. PB-216-188). [Google Scholar]
  • 31.Klug HP, Alexander LE. X-ray Diffraction Procedures. For Polycrystalline and Amorphous Materials. John Wiley & Sons, Inc; New York: 1972. [Google Scholar]
  • 32.Munro RG. Evaluated Material Properties for a Sintered α-alumina. J Am Ceram Soc. 1997;80(8):1919–28. [Google Scholar]
  • 33.Ishizawa N, Ishikawa A, Miyata T, Minato I, Marumo F, Iwai S. A Structural Investigation of α-Al2O3 at 2170 K. Acta Cryst. 1980;B36:228–230. [Google Scholar]
  • 34.Morris MC, McMurdie HF, Evans EH, Paretzkin B, Parker HS, Panbgiotpoulos NC. Standard x-ray Diffraction Powder Patterns. Sect. 1981. p. 61. (Natl Bur Standards Monograph 25). [Google Scholar]
  • 35.Newnham RE, de Haan YM. Refinement of the α Al2O3, Ti2O3, V2O3, and Cr2O3 structures. Z Kristallogr. 1962;117(8):235–237. [Google Scholar]
  • 36.Shannon RD. Revised Effective Ionic Radii and Systematic Studies of Interatomic Distances in Halides and Chalcogenides. Acta Cryst. 1976;A32:751–767. [Google Scholar]
  • 37.The American National Standard for Expressing Uncertainty—U.S. Guide to the Expression of Uncertainty of Measurement (ANSI/NCSL Z540-2-1997).
  • 38.Jenkins R, Snyder RS. Introduction to x-ray Powder Diffractometry. John-Wiley & Sons Inc; 1996. p. 177. [Google Scholar]
  • 39.Ralph RL, Finger LW. A computer program for refinement of crystal orientation matrix and lattice constants from diffractometer data with lattice symmetry constraints. J Appl Cryst. 1982;15:537–539. [Google Scholar]
  • 40.Frevel LK, Emge TJ, Kistenmacher TJ. An Error Assessment of Single-Crystal 2θ Data from Four-Circle Diffractometer. J Appl Cryst. 1983;16:126–132. [Google Scholar]

Articles from Journal of Research of the National Institute of Standards and Technology are provided here courtesy of National Institute of Standards and Technology

RESOURCES