Ultra-high-pressure liquid chromatography tandem mass spectrometry method for the determination of 9 organophosphate flame retardants in water samples

María Lorenzo a,, Julián Campo b,c, Yolanda Picó a
PMCID: PMC4865636  PMID: 27222824

Graphical abstract

graphic file with name fx1.jpg

Method name: Analysis of organophosphate flame retardants in water by SPE and UHPLC–MS/MS using a trap column.

Keywords: Organophosphate flame retardants, Interferences, Trap column, LC–MS/MS, Water, solid phase extraction

Abstract

Few methods are available for comprehensive organophosphate flame retardants (PFRs) detection in water and wastewater. Gas chromatography has been employed previously, but this approach is less selective, not amenable for use with deuterated standards and can suffer unfavorable fragmentation. Ultra-high-pressure liquid chromatography tandem mass spectrometry (UHPLC-QqQ-MS/MS) has become the most promising platform, already applied successfully for analysis of selected PFRs in some environmental matrices like water and wastewater. However, the presence of some interferences from the dissolvent, the equipment and the used materials should be taken into account. The procedure involves:

  • The first determination of PFRs by UHPLC-QqQ-MS/MS using a trap column to distinguish the interferences coming from the instrument and mobile phases.

  • The optimization of the LC separation to distinguish all target compounds and their interferences.

  • This method coupled to a solid-phase extraction (SPE) improve the detection and quantification of PFRs.

Method details

Interference analysis

Ultra-high-pressure liquid chromatography tandem mass spectrometry (UHPLC-QqQ-MS/MS) is likely the most promising platform for the determination of PFRs in water and wastewater matrices [1], [2], [3], [4], [5]. However, the presence of some interferences from the dissolvent, the equipment and the laboratory material can disturb their analytical determination. We made, up to our knowledge, the first determination of PFRs using a trap column to distinguish these interferences. Furthermore, we prove two different solvents, acetonitrile (ACN) and methanol (MeOH), observing the presence of analyte’s contamination in both cases. The system is applied to the determination of PFR in water.

Fig. 1, Fig. 2 present chromatograms of a non-spiked river sample and the same sample spiked with the analytes.

Fig. 1.

Fig. 1

Chromatogram of a river sample.

Fig. 2.

Fig. 2

Chromatogram of the previous river sample spiked at 500 ng/mL.

Reagents and samples

  • (a)

    Compounds: the nine PFRs and the four deuterated PFRs used as internal standard (IS) (Table 1) were purchased from LGC Standards (Germany).

  • (b)

    Solvents: methanol (MeOH), dichloromethane (DCM) and acetonitrile (ACN) were bought from VWR (Radnor, PA, USA) and formic acid from AMRESCO (Solon, OH, USA), all of them were of analytical quality.

  • (c)

    Ultra-pure water: prepared with Milli-Q SP Reagent Water System from Millipore (Bedford, MA, USA).

  • (d)

    Standard solutions: individual solutions (1 mg/mL) were prepared by dissolving the PFRs in MeOH. Mixed stock solutions containing 10000 ng/mL and 100 ng/mL of each of the nine PFRs were prepared by dilution of the stock solutions with MeOH. Stock and mixed standard solutions were stored in polypropylene tubes at 4 °C.

  • (e)

    Samples: water samples were obtained from the Turia River (May 2015) from the influent and the effluent of the wastewater treatment plant (WWTP) of Pinedo (March 2015), both near the city of Valencia (Spain). Influent wastewater samples were filtered under vacuum using the ADVANTEC® filters to remove the particulate matter. Analysis of spiked samples and blanks were made in triplicate.

Table 1.

List of the compounds name, acronym and formula.

Compounds Acronym Formula CAS number
Tris(2-chloroethyl) phosphate TCEP C6H12Cl3O4P 115-96-8
Tri-n-propylphosphate TPP C9H21O4P 513-08-6
Tris(2-chloroisopropyl)phosphate TClPP C9H18Cl3O4P 13674-84-5
Tris(1,3-dichloro-2-propyl)phosphate TDClPP C9H15Cl6O4P 13674-87-8
Triphenyl phosphate TPhP C18H15O4P 115-86-6
Cresyl diphenyl phosphate CDP C19H17O4P 26444-49-5
Tris(2,3-dibromopropyl)phosphate TDBPP C9H15Br6O4P 126-72-7
Tricresylphosphate TMPP C21H21O4P 1330-78-5
Tris-(2-ethylhexyl)phosphate TEHP C24H51O4P 78-42-2
Triphenyl phosphate D15 (IS) M-TPHP C18D15O4P 1173020-30-8
Tris(1,3-dichloro-2-propyl) phosphate D15 (IS) M-TDCIPP C9D15Cl6O4P Not available
Tris-(2-chloroethyl) phosphate D12 (IS) M-TCEP C6D12Cl3O4P 115-96-8
Tris(2-chloroisopropyl) phosphate D18 (IS) M-TCIPP C9D18Cl3O4P 13674-84-5

Optimization of sample extraction

  • (a)

    Put the Phenomenex Strata-X 33u Polymeric Reversed Phase (200 mg/6 mL) cartridges (Phenomenex, Torrance, Ca, USA) into a 12 port vacuum manifold Supelco Visiprep 57030-U de Sigma-Aldrich (St. Louis, MO, EEUU).

  • (b)

    Precondition the cartridge with 6 mL MeOH:DCM (1:1 v/v), 6 mL of MeOH and 6 mL of water, with 350 mba/h·PA vacuum.

  • (c)

    Pass the water samples through the cartridges under previous vacuum at a flow rate of 10 mL/min.

  • (d)

    Dry the cartridges under vacuum for 15 min.

  • (e)

    Elute the analytes on a 15 mL Falcon tube VWR (Radnor, PA, USA) with 10 mL of MeOH:DCM (50:50 v/v).

  • (f)

    Evaporate the extracts to dryness at 40 °C using a combined sample concentrator model SBHCONC/1 and a heating plate model SBH130D/3 both manufactured by Stuart (Stafford, UK).

  • (g)

    Redissolve the residue in 1 mL of methanol by agitation and ultrasonication for 1 min and pass the extract to 2 mL amber vials with 250 μL insert polypropylene 100/PK + Septum Sil/PTFE, both manufactured by Análisis Vínicos S.L. (Tomelloso, Spain).

Instrumentation

The analysis was performed using:

  • (a)

    UHPLC-QqQ-MS/MS system: 1260 Infinity ultra-high-performance liquid chromatograph combined with a 6410 triple quadrupole mass spectrometer (MS/MS) with electrospray ionization (ESI) of Agilent Technologies (Santa Clara, CA, USA).

  • (b)

    LC column: Kinetex C18 (50 × 2.1 mm, 1.7 μm) from Phenomenex (Torrance, CA, EEUU).

  • (c)

    Trap column: Zorbax Eclipse Plus C18 (30 × 4.6 mm, 3.5 μm) from Agilent Technologies (Santa Clara, CA, USA). Place this column between the pump and the auto sampler.

  • (d)

    Mobile phases: (A) water and (B) methanol, both containing 0.1% of formic acid.

  • (e)

    Gradient: 0 min (30% B), 0.5 min (30% B), 12 min (95% B), 18 min (98% B) and 25 min (98% B) and return to the initial conditions. An equilibration time of 15 min was applied to stabilize the column conditions.

  • (f)

    Flow rate: 0.2 mL/min and the sample volume injected was 5 μL.

  • (g)

    Analysis: performed in positive ionization mode.

  • (h)

    Data acquisition: carried out in selected reaction monitoring (SRM) to identify and quantify using two precursor-product ion transitions, retention times, and the ratio of intensities between the two product ions.

Fragmentor and collision energies were optimized for each compound individually (Table 2).

Table 2.

Dynamic MRM conditions for UHPLC-QqQ-MS/MS determination of PFRs.

Acronym Associated IS Precursor Ion Product ion Fragmentor Collision Energy
TCEP M-TCEP 287 99 100 15
285 223 100 10
TPP M-TPhP 225.1 140.9 84 3
225.1 98.9 84 15
TClPP M-TClPP 329 99 80 15
327 175 80 10
TDClPP M-TDClPP 432.9 99.1 80 15
430.9 99.1 80 15
TPhP M-TPhP 327.1 214.9 117 27
327.1 151.9 117 43
CDP M-TPhP 341.1 151.9 167 43
341.1 90.9 167 39
TDBPP M-TDClPP 698.6 98.9 120 25
696.6 98.9 120 25
TMPP M-TPhP 369.1 165.6 192 31
369.1 91 192 43
TEHP M-TPhP 435.4 98.9 113 7
435.4 71 113 5
M-TPHP 342 160 120 47
342 82 120 47
M-TDCIPP 448 102 120 15
446 102 120 15
M-TCEP 299 67 100 20
297 67 100 20
M-TCIPP 347 102 100 20
345 102 100 20

In italics, the compounds used as IS.

Validation of the method

The validation of the instrumental parameters (Table 3) was performed by determining recoveries, limit of detection (LOD), limit of quantification (LOQ) and linearity over the range to obtain suitable R2. The quantification was performed using the internal standard method. The mixture of the internal standards was added to water to a concentration of 200 ng/L to get a final concentration in the injected extract of 50 ng/mL. The reported parameters were calculated from distilled, river and wastewater samples spiked with standard solutions prepared in methanol at the appropriate concentrations, in a way that the volume of organic solvent added were never higher than 250 μL. The samples were processed with plastic materials as much as possible to avoid adsorption to glass. There were not river or wastewater samples free of PFRs. Then, to perform these experiments several non-spiked samples were analyzed and the peak area of those compounds were subtracted to that found in the spiked samples.

Table 3.

Recoveries and R2 values for calibration curves (n = 7), LODs and LOQs (ng/mL) and intra and inter-day precision.

Acronym Recoveries R2 LOD (ng/ml) LOQ (ng/ml) Intra-day precision Inter-day precision
TCEP 98% 0.994 0.03 0.3 7.5 6.6
TPP 95% 0.990 0.25 2.5 5.8 19.6
TClPP 102% 0.990 0.25 2.5 3.0 6.6
TDClPP 96% 0.990 0.03 0.3 4.1 8.2
TPhP 96% 0.998 0.1 1 4.2 18.9
CDP 94% 0.994 0.1 1 2.6 19.9
TDBPP 99% 0.998 0.25 2.5 11.4 17.5
TMPP 101% 0.990 0.1 1 3.2 18.6
TEHP 106% 0.992 0.25 2.5 5.9 20.0

The LOD was calculated as the mass of analyte required to produce a signal-to-noise (S/N) ratio of 3:1 and the LOQ of 10:1. S/N ratios were calculated using MassHunter Workstation Software (GL Sciences, Tokyo, Japan). LODs and LOQs were estimated from water samples spiked at 4 ng/L for river water and at 100 ng/L in wastewater. Values reported in Table 3 are LODs and LOQs in the injected extracts that would correspond to LODs ranging from 0.12 to 1 ng/L and LOQs from 1.2 to 10 ng/L in water samples. The results obtained for the LODs and LOQs show that the proposed method is sensitive enough for the determination of the PFRs in water samples.

The precision studies were carried out from the evaluation of intra-day and inter-day variations of the areas ratio between the analyte and its IS using water extracts suitably stored. The intra-day precision of peak areas ratio (five replicates at 400 ng/L), expressed by means of the percentage of relative standard deviation (%RSD (n = 5)) were lower than 11.4% and the inter-day precision (five replicates) were less than 20%. Recoveries were calculated from samples spiked at 400 ng/L analyzed in quintuplicate.

The calibration of the LC–MS/MS was conducted using seven different concentrations (from the LOQ to 300 ng/mL) of PFR standard solutions in methanol with 50 ng/mL of each deuterated compound used as IS. The use of IS prevent the matrix effects

The robustness of the method was clearly ascertained during the optimization procedure, by establishing the consequences of the deliberate introduction of minor reasonable variations (mostly different water volume analyzed, sample flow-rates during the extraction step, cartridge drying times and eluent composition) and by the similar results obtained checking different types of water. The results obtained from the variations of ±20% in the different parameters were not significantly different from those achieved by the validated method. These results proved that the proposed method was robust.

Each 10 samples, one instrumental and one procedural blank were analyzed to serve as quality control. Samples analyzed shown clearly two peaks, the first one corresponding to the TClPP found in the river water and the second to the instrumental background, indicating background contamination from the injection system and tubing of the LC–MS/MS can be successfully separated from the sample contamination. Table 4 shows the performance of the system in water samples (river, influent and effluent wastewater).

Table 4.

Analysis of 250 mL of the river water and the influent and effluent wastewater samples (ng/ml).

Acronym River water Influent wastewater Effluent wastewater
TCEP <LOQ 19.0 0.8
TPP 4.0 <LOQ <LOQ
TClPP 72.5 448.9 134.6
TDClPP <LOQ 24.4 20.5
TPhP 5.6 20.2 9.7
CDP <LOQ 298.5 23.7
TDBPP <LOQ 393.9 7.3
TMPP <LOQ 17.0 2.2
TEHP 12.5 9.7 99.9

Acknowledgements

This work has been supported by the Spanish Ministry of Economy and Competitiveness through the projects GCL2011-29703-C02-02 (http://mefturia.es) and GCL2015-64454-C2-1-R (ECO2risk-dds) and the University of Valencia through the project (UV-INV-AE15-348995). María Lorenzo also acknowledges to the Foundation “Tatiana Pérez de Guzmán el Bueno” for the grant to get the PhD.

MethodsX thanks the reviewers (anonymous) of this article for taking the time to provide valuable feedback.

Footnotes

1

Refs. [6–14] are cited in additional information.

Appendix A

Supplementary data associated with this article can be found, in the online version, at http://dx.doi.org/10.1016/j.mex.2016.04.006.

Appendix A. Supplementary data

The following are Supplementary data to this article:

mmc1.docx (45.9KB, docx)

References1

  • 1.Wang X.-w., Liu J.-f., Yin Y.-g. Development of an ultra-high-performance liquid chromatography–tandem mass spectrometry method for high throughput determination of organophosphorus flame retardants in environmental water. J. Chromatogr. A. 2011;1218(38):6705–6711. doi: 10.1016/j.chroma.2011.07.067. [DOI] [PubMed] [Google Scholar]
  • 2.Woudneh M.B. Quantitative determination of 13 organophosphorous flame retardants and plasticizers in a wastewater treatment system by high performance liquid chromatography tandem mass spectrometry. J. Chromatogr. A. 2015;1400:149–155. doi: 10.1016/j.chroma.2015.04.026. [DOI] [PubMed] [Google Scholar]
  • 3.Bacaloni A. Liquid chromatography/tandem mass spectrometry determination of organophosphorus flame retardants and plasticizers in drinking and surface waters. Rapid Commun. Mass Spectrom. 2007;21:1123–1130. doi: 10.1002/rcm.2937. [DOI] [PubMed] [Google Scholar]
  • 4.Martínez-Carballo E. Determination of selected organophosphate esters in the aquatic environment of Austria. Sci. Total Environ. 2007;388(1–3):290–299. doi: 10.1016/j.scitotenv.2007.08.005. [DOI] [PubMed] [Google Scholar]