Abstract
Simple, sensitive, precise, and specific high-performance liquid chromategraphic (HPLC) and high-performance thin-layer chromatographic (HPTLC) methods for the determination of dolutegravir sodium in bulk drug and pharmaceutical dosage form were developed and validated. In the HPLC method, analysis of the drug was carried out on the ODS C18 column (150 × 4.6 mm, 5 μm particle size) using a mixture of acetonitrile: water (pH 7.5) in the ratio of 80:20 v/v as the mobile phase at the flow rate 1 mL/min at 260 nm. This method was found to be linear in the concentration range of 5–35 μg/mL. The peak for dolutegravir sodium was observed at 3.0 ± 0.1 minutes. In the HPTLC method, analysis was performed on aluminum-backed plates pre-coated with silica gel G60 F254 using methanol: chloroform: formic acid in the proportion of 8:2:0.5 v/v/v as the mobile phase. This solvent system was found to give compact spots for dolutegravir sodium with the Rf value 0.77 ± 0.01. Densitometric analysis of dolutegravir sodium was carried out in the absorbance mode at 265 nm. Linear regression analysis showed good linearity with respect to peak area in the concentration range of 200–900 ng/spot. The methods were validated for precision, limit of detection (LOD), limit of quantitation (LOQ), accuracy, and specificity. Statistical analysis showed that both of the methods are repeatable and specific for the estimation of the said drug. The methods can be used for routine quality control analysis of dolutegravir sodium.
Keywords: Dolutegravir sodium, HPLC, Forced Degradation, HPTLC, Validation
Introduction
Dolutegravir sodium, chemically, sodium (4R,12aS)-9-[(2,4-difluorobenzyl)carbamoyl]-4-methyl-6,8-dioxo-3,4,6,8,12,12a-hexahydro-2H-pyrido[1’,2’:4,5]pyrazino[2,1-b][1,3]oxazin-7-olate (Fig. 1), is a novel integrase stand transfer inhibitor active against human immunedeficiency virus. The drug is active against HIV type 1 (HIV-1) and also has some in vitro activity against HIV type 2 (HIV-2) [1–4].
Fig. 1.

Structure of dolutegravir sodium
The literature review revealed a liquid chromatography–tandem mass spectrometry method [5] and a sensitive HPLC–MS/MS method [6] for the estimation of dolutegravir in human blood plasma. There is no HPLC and HPTLC method available yet for the quantitative determination of dolutegravir in tablet formulation. Further, no official or draft monograph of dolutegravir sodium was published in any of the pharmacopoeia for compendia applications.
The aim of the present work was to develop simple, economic, precise, accurate, specific, stability-indicating HPLC method and a simple HPTLC method for the assay determination of dolutegravir in bulk form and/or in pharmaceutical dosage form. The developed methods were validated as per International Conference on Harmonisation (ICH) guidelines [7–9].
Experimental
Chemicals and Reagents
Dolutegravir sodium bulk drug was obtained from Mylan Labs Ltd, (Hyderabad, India), the commercial tablets of dolutegravir sodium were not available in the Indian market; hence, we have prepared immediate release tablets containing dolutegravir sodium equivalent to 50 mg of dolutegravir as per cGMP guidelines. The tablets contain lactose monohydrate, microcrystalline cellulose, starch, and magnesium stearate with an average weight of 250 mg. Water was obtained from a Milli-Q UF-Plus apparatus (Millipore) and was used to prepare all solutions for the method. Other chemicals used were analytical or HPLC-grade and the glassware used was Class A grade.
Method A: HPLC
Instrumentation
Chromatographic separation was performed using the Waters 600 Controller chromatographic system equipment with a reciprocating pump UV/Visible detector and Reodyne (7725i) with a 20 μL fixed loop, and the data were analyzed by using Data Ace software. The analytical balance used for weighing the standard and sample was Shimadzu Aux 220 Uni Bloc Pat 1987. The balance (Mettler Toledo XP205, Mumbai) was used for weighing and an Ultrasonicator (Enertech Electronics Pvt. Ltd., Mumbai) was used for sonication. The Thermo Scientific Forma 3960 Series environmental chamber was used for stress testing.
Chromatographic Conditions
The analysis was performed using the Inertsil ODS C18 (150 × 4.6 mm, 5 µm particle size) column using a mobile phase consisting of acetonitrile: water (pH 7.5) in the ratio of 80:20, v/v at a flow rate of 1 mL/min. The eluent was monitored using UV detection at a wavelength of 260 nm at ambient column temperature. The injection volume 20 μL was used. The total run time was 6 min. The mobile phase was filtered through a 0.45 μm micron filter prior to use.
Preparation of Standard Stock Solution
Accurately weighed dolutegravir sodium standard equivalent to 100 mg dolutegravir was transferred into a 100-mL volumetric flask. It was dissolved in 70 mL of methanol by sonication for one minute and then diluted to volume with methanol to obtain the standard stock solution of the concentration 1000 μg/mL of dolutegravir.
Selection of Detection Wavelength
The spectra of 10 μg/mL solution of dolutegravir in methanol were recorded on a UV spectrophotometer. The wavelength of the maximum absorbance was observed.
Preparation of the Calibration Curve
The aliquots from the standard stock solution (0.5, 1, 1.5, 2.0, 2.5, 3.0, and 3.5 mL) were transferred to a series of 100-mL volumetric flasks and diluted up to the mark with mobile phase, to give the working standard solutions of the concentration 5 μg/mL to 35 μg/mL.
Analysis of the Pharmaceutical Dosage Form
To determine the content of dolutegravir in prepared tablet dosage form, 20 tablets were weighed; their average weight was determined and finely powdered. Powder equivalent to 100 mg dolutegravir was accurately weighed and transferred to a 100-mL volumetric flask and dissolved in 70 mL of methanol and the mixture was sonicated for 10 min. The volume was adjusted up to the mark with the same solvent. This solution was filtered through a 0.45 μm nylon syringe filter. From this solution, an aliquot of 1 mL was further diluted to 100 mL with the mobile phase. The chromatograms were developed using previously described chromatographic conditions. Six replicate injections of this solution were injected and the chromatogram and peak area were recorded. The concentration of the tablet solution was determined using a linear regression equation of a calibration graph and the amount of drug in the tablet was determined. The possibility of excipient interference in the analysis was studied.
Forced Degradation Studies [8, 9]
The forced degradation studies were performed to establish the stability-indicating nature and specificity of the assay method and to observe any degraded compounds. The forced degradation of dolutegravir sodium sample solution was carried out with 1 N hydrochloric acid, 1 N sodium hydroxide, 30% v/v hydrogen peroxide, UV degradation at 254 nm, and thermal degradation at 60°C in the presence of 80% RH. The forced degradation conditions are as given in Table 1. Chromatograms were recorded for all of the above solutions.
Tab. 1.
Forced Degradation Conditions

Method Validation
The HPLC method was validated as per the ICH guidelines [8, 9].
Linearity
Twenty μL from each working standard solution was injected in hexaplicate into the HPLC system. The peak areas were plotted against the corresponding concentrations to obtain the calibration graph. A linear calibration curve was generated using least-squares linear regression analysis.
Precision
Precision of the method was verified by repeatability and intermediate precision studies. Repeatability studies were performed by analyzing the solution of 20 µg/mL of the drug in hexaplicate on the same day. The % RSD of the six determinations was calculated. Intermediate precision of the method was checked by repeating studies on two different days. The % RSD of 12 determinations was calculated.
Accuracy
Accuracy of the method was determined by the standard addition method in which the known amounts of standard dolutegravir solutions were added to a pre-analyzed solution. These amounts corresponded to 80, 100, and 120% of the sample concentration. The amounts of dolutegravir were estimated by applying these values to the regression equation of the calibration curve. An accuracy study was performed for three times at each level, and the % recovery of dolutegravir was calculated.
Robustness
To determine the robustness of the method, the experimental conditions, i.e. flow rate and pH of the mobile phase, were deliberately altered. For changes in conditions, the sample was assayed in triplicate. The optimum mobile phase flow rate was 1.0 mL/min. This was changed by 0.1 units to 0.9 and 1.1 mL/min and the effects were studied. The effect of a change in pH of the mobile phase was studied at pH 7.3 and pH 7.7.
Limit of Detection and Limit of Quantitation
The LOD and LOQ were calculated by using the formula, LOD = 3.3 × σ/S and LOQ = 10 × σ/S, where σ is a residual standard deviation of the regression line and S is the slope of the corresponding regression line.
Specificity
The specificity of the developed method was established by analyzing the sample solutions containing dolutegravir standard and prepared tablets in relation to interferences from formulation ingredients.
Solution Stability [10]
The stability of standard solutions was tested after 1, 6, and 24 h of storage. The stability of the solutions was determined by analyzing the sample solution of the concentration 10 µg/mL.
System Suitability Parameters
Six replicate injections of system suitability solutions (working standard solution) were injected. The retention time, areas, theoretical plates, peak asymmetry, and resolution were calculated for standard solutions.
Method B: HPTLC
HPTLC Instrumentation and Chromatographic Conditions
The HPTLC plates were prewashed with methanol and activated at 110°C for 5 minutes prior to chromatography. The samples were spotted in the form of bands of 8 mm width with a Camag 100 microliter sample syringe (Hamilton, Bonaduz, Switzerland) on HPTLC aluminum-backed plates pre-coated with silica gel G60 F254, [(20 × 10 cm); E. Merck, Darmstadt, Germany, supplied by Anchrom Technologists, Mumbai] using a Camag Linomat V applicator (Switzerland). Linear ascending development was carried out in a 20 cm × 10 cm twin trough glass chamber (Camag, Muttenz, Switzerland) saturated with the mobile phase. The mobile phase consisted of methanol, chloroform, and formic acid in the proportion of 8:2:0.5, v/v/v and 20 mL mobile phase was used per chromatography run. The optimized chamber saturation time for the mobile phase was 30 min using saturation pads at room temperature (25 ± 2°C). The length of the chromatogram run was 8 cm. Densitometric scanning was performed using a Camag TLC Scanner III in the absorbance mode and operated by WinCATS software (V 3.15, Camag). The slit dimension was kept at 5 mm × 0.45 mm and the scanning speed was 10 mm/s. The source of radiation used was a deuterium lamp emitting a continuous UV spectrum between 190 and 400 nm. All determinations were performed at ambient temperature with a detection wavelength of 265 nm.
Preparation of Standard Solution
Dolutegravir sodium (equivalent to 100 mg of dolutegravir) was accurately weighed and transferred into 100-mL volumetric flasks, and dissolved in 50 mL of methanol and the solution was further diluted with methanol to get the final concentration of 1000 μg/mL. From this stock solution, 10 mL was diluted to 100 mL with methanol to give the final concentration of 100 μg/mL.
Preparation of Sample Solution
Twenty tablets (each tablet containing dolutegravir 50 mg) were accurately weighed; their average weight was calculated and finely powdered. The powder equivalent to 50 mg dolutegravir was accurately weighed and transferred into 50-mL volumetric flasks containing 20 mL of methanol and ultrasonicated for 15 minutes. The solution was diluted to 50 mL with methanol. The above solution was filtered through the Whatman no. 41 filter paper. From this solution, 5 mL of solution was transferred to 50-mL volumetric flasks and made up to the mark with methanol.
Analysis of Pharmaceutical Dosage Form
An aliquot of 4 μL was applied on the TLC plates followed by the development and measured at 265 nm. The analysis was repeated six times. The concentration of the sample was determined using the linear regression equation of the calibration graph and the amount of drug in each tablet was determined.
Method Validation
The HPTLC method was validated as per the ICH guidelines [8–10].
Linearity
Aliquots from the standard solution were spotted on the HPTLC plate to obtain concentrations in the range of 200–900 ng/spot. The plate was developed using the previously described mobile phase and scanned. The peak areas were plotted against the corresponding concentrations to obtain the calibration graph. A linear calibration curve was generated using least-squares linear regression analysis.
Precision
Precision of the method was verified by repeatability and intermediate precision studies. Repeatability studies were performed by analyses of the drug (400 ng/spot) in hexaplicate on the same day. Intermediate precision of the method was checked by repeating studies on two different days.
Limit of Detection and Limit of Quantitation
The sensitivity of the method was determined in terms of limit of detection (LOD) and limit of quantitation (LOQ). The LOD and LOQ were calculated by using the formula, LOD = 3.3 × σ/S and LOQ = 10 × σ/S, where σ is the residual standard deviation of the regression line and S is the slope of the corresponding regression line.
Accuracy
The accuracy of the method was determined by the standard addition method in which the known amount of standard dolutegravir solutions were added to pre-analyzed sample solutions. These amounts corresponded to 80, 100, and 120% of the sample concentration. The amounts of dolutegravir were estimated by applying these values to the regression equation of the calibration curve. The accuracy study was performed in triplicate and the % recovery of dolutegravir was calculated.
Specificity
The specificity of the developed method was established by comparing the chromatogram of the sample with the chromatograms of the standard.
Solution Stability
The stability of the standard solutions was tested after 1, 6, and 24 h of storage. The stability of the solutions was determined by comparing the peak area percentage and peak purity at 500 ng/spot.
Results and Discussion Method A: HPLC
Selection of Detection Wavelength
The spectra of dolutegravir showed the maximum absorbance at the wavelength 260.0 nm (Figure 2). So, the wavelength 260.0 nm was selected for the analysis.
Fig. 2.

UV spectrum of dolutegravir in methanol
Assay of Dolutegravir in Tablets by the HPTLC Method
The concentration of the tablet solution was determined using a linear regression equation (using the slope and Y-intercept) and the amount of drug in each tablet was determined. The results of the assay of the tablets are summarized in Table 2.
Tab. 2.
Results of the Assay of Dolutegravir in Tablets by the HPLC Method

Forced Degradation Studies of Dolutegravir
The purity of dolutegravir was unaffected by the presence of its degradation products; thus, the method can be said to be stability-indicating. The percent assay of all the degraded samples varied between 92.1 and 100.2%. The numbers of the degradation peaks observed in different stress conditions are shown in Figure 3 and are as follows:
Fig. 3.

Typical degradation chromatogram of dolutegravir sodium: A- acid-induced degradation; B- alkali-induced degradation; C- peroxide degradation; D- UV degradation; and E- thermal with humidity-induced degradation
Acid-Induced Degradation
Only minor degradation products were observed at Rt 2.36 (0.07%) in the acid degradation of sample preparation.
Base-Induced Degradation
Major degradation products were observed at Rt 2.2, 2.6, and 5.6 (9%) in 1 N sodium hydroxide for 80°C for 20 h in the sample preparation.
Oxidative Degradation
No degradation peak was observed in the peroxide degradation of the sample preparation.
UV Radiation Degradation (at 254 nm)
No degradation product was observed after exposure of the sample to UV light at 254 nm.
Thermal with Humidity Degradation (60°C/80% RH)
No degradation peak was observed in the sample preparation exposed to thermal with humidity degradation.
Method Validation
Linearity
A linear relationship was observed by plotting the drug concentration against the peak areas. Dolutegravir showed a linear response in the concentration range of 5-35 µg/mL by the HPLC method. The regression of the plot was computed by the least-squares regression method. Linearity results are presented in Table 3 and the linearity plot is shown in Figure 4.
Tab. 3.
Linearity Data

Fig. 4.

Linearity plot of dolutegravir by the HPLC method
Precision
Precision is the degree of repeatability of an analytical method under normal operational conditions. The developed method was found to be precise as the % RSD values for repeatability and intermediate precision studies were less than 2%. The results of repeatability and intermediate precision are shown in Table 4.
Tab. 4.
Results of Precision Studies by the HPLC Method

Accuracy
The results of the recovery studies showed the accuracy of the method. Satisfactory recoveries ranging from 98.82 to 99.70% were obtained by the proposed method. This indicated that the proposed method was accurate. The results obtained are given in Table 5.
Tab. 5.
Results of Recovery Studies by the HPLC Method

Robustness
The robustness study was performed by a slight modification in the flow rate of the mobile phase and pH of the water. Samples of dolutegravir at 10 μg/mL concentration were analyzed under these changed experimental conditions. It was observed that there were no marked changes in chromatograms, which demonstrated that the developed method was robust in nature. The results are given in Table 6.
Tab. 6.
Results of Robustness Studies by the HPLC Method

Limit of Detection and Limit of Quantification
The LOD and LOQ were found to be 1.91 µg/mL and 5.17 µg/mL, respectively.
Specificity
The specificity of method was performed by comparing the chromatograms of the blank, standard, and sample. It was found that there was no interference due to the excipient in the tablet formulation (Figure 5). A good correlation was also found between the retention time of the standard and sample of dolutegravir.
Fig. 5.

Typical chromatogram of dolutegravir sodium tablet sample
Solution Stability
The solution was found to be stable at ambient temperature for 24 h, and no unknown peaks were observed. The results of the stability studies by the HPLC method are shown in Table 7.
Tab. 7.
Results of Stability Studies by the HPLC Method

System Suitability
System suitability was studied by injecting six replicates of the standard solution. The system suitability parameters are given in Table 8. The number of theoretical plates was found to be 19944, which indicates efficient performance of the column.
Tab. 8.
System Suitability Parameters by the HPLC Method

Method B: HPTLC
Analysis of the Pharmaceutical Dosage Form by the HPTLC method
A single spot at the Rf value of 0.77 was observed in the chromatogram of the drug samples extracted from the tablet. There was no interference from the excipients that are commonly present in the formulations. The drug content was found to be 101.62 % as given in Table 9.
Tab. 9.
Analysis of Tablet Formulation by the HPTLC Method

Method Validation
Linearity
A linear relationship was observed by plotting the drug concentration against the peak areas. Dolutegravir showed a linear response in the concentration range of 200–900 ng/spot by the HPTLC method (Figure 6). The corresponding linear regression equation was Y = −468.1 + 3.966x with the square of the correlation co-effcient (r2) of 0.997 for dolutegravir as shown in Table 10.
Fig. 6.

Linearity plot for dolutegravir by the HPTLC method
Tab. 10.
Linear Regression Data for the HPTLC Method

Precision
The results of the repeatability and intermediate precision experiments are shown in Table 11. The developed method was found to be precise as the % RSD values for repeatability and intermediate precision studies were < 2%, respectively.
Tab. 11.
Results of Precision Studies by the HPTLC Method

Limit of Detection and Limit of Quantitation
The LOD and LOQ were found to be 28.61 ng/spot and 86.69 ng/spot, respectively.
Accuracy
The developed method showed high and consistent recoveries at all studied levels. The results obtained from the recovery studies are presented in Table 12. The mean % recovery ranged from 98.67% to 99.70%. Additionally, the obtained recoveries were found to be normally distributed with low % RSD at all concentration levels.
Tab. 12.
Results of Recovery Studies of Dolutegravir by the HPTLC Method

Specificity
The peak purity for dolutegravir was assessed by means of comparing the chromatogram of the sample with that of the mobile phase, diluents, and standard (Figure 7). No interference of excipients with the dolutegravir peak was observed in the sample chromatogram. A single peak of dolutegravir in the tablet solution was observed at Rf 0.77 (Figure 8).
Fig. 7.

Densitogram indicating specificity of the HPTLC method for dolutegravir
Fig. 8.

Densitogram of dolutegravir in tablet solution by the HPTLC method
Solution Stability
There was no indication of degradation in the sample solutions of dolutegravir as revealed by the peak purity data and from the value of % RSD (< 2%) for the peak areas of the bands of solution stored at different times. The solution was found to be stable at ambient temperature for 24 h, and no unknown peaks were observed. The results are shown in Table 13.
Tab. 13.
Results of Stability Studies

Conclusion
A simple and reliable HPLC method has been developed and successfully validated for the estimation of dolutegravir in the presence of degradation products. As the method separates the drug from its degradation products, it can be employed as a stability-indicating one. The results of the HPLC and HPTLC validation tests indicated that the method was accurate, precise, robust, and reproducible. This assay system provides an accurate, precise, and sensitive method for dolutegravir quantitation and was successfully applied to the bulk and pharmaceutical dosage form. Hence, the proposed HPLC and HPTLC method is suitable for routine determination of dolutegravir in pharmaceutical formulation in quality control laboratories, where economy and time are essential.
Acknowledgement
Authors are thankful to Mylan Laboratories Limited, Hyderabad, India, for providing the pure drug samples for this study and the Amrutvahini Sheti and ShikshanVikas Sanstha, Sangamner, India and Anchrom Lab, Mumbai, India for providing us with the research facility.
Authors’ Statement
Competing Interests
The authors declare no conflicts of interest.
References
- 1.ViiV Healthcare. Tivicay®(dolutegravir) prescribing information. NC: Research Triangle Park; 2014. May, [Google Scholar]
- 2.FDA. Food and Drug Administration. FDA approves new drug to treat HIV infection. 2013. Aug 12, http://www.fda.gov/NewsEvents/Newsroom/PressAnnouncements/ucm364744.htm .
- 3.Castellino S, Moss L, Wagner D, Borland J, Song I, Chen S, Lou Y, Min SS, Goljer I, Culp A, Piscitelli SC, Savina PM. Metabolism excretion, and mass balance of the HIV-1 integrase inhibitor dolutegravir in humans. Antimicrob Agents Chemother. 2013;57:3536–3546. doi: 10.1128/AAC.00292-13. http://dx.doi.org/10.1128/FAAC.00292-13 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Min S, Song I, Borland J, Chen S, Lou Y, Fujiwara T, Piscitelli SC. Pharmacokinetics and safety of S/GSK1349572, a next-generation HIV integrase inhibitor, in healthy volunteers. Antimicrob Agents Chemother. 2010;54:254–258. doi: 10.1128/AAC.00842-09. http://dx.doi.org/10.1128/AAC.00842-09 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Bennetto-Hood C, Tabolt G, Savina P, Acosta EP. A sensitive HPLC-MS/MS method for the determination of dolutegravir in human plasma. J Chromatogr B Analyt Technol Biomed Life Sci. 2014;945–946:225–232. doi: 10.1016/j.jchromb.2013.11.054. http://dx.doi.org/10.1016/j.jchromb.2013.11.054 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Grégoirea M, Deslandesa G, Renauda C, Bouquiéa R, Allaenac C, Raffic F, Jollieta P, et al. A liquid chromatography–tandem mass spectrometry assay for quantification of rilpivirine and dolutegravir in human plasma. J Chromatogr B Analyt Technol Biomed Life Sci. 2014;971:1–9. doi: 10.1016/j.jchromb.2014.09.006. http://dx.doi.org/10.1016/j.jchromb.2014.09.006 . [DOI] [PubMed] [Google Scholar]
- 7.International Conference on Harmonisation. ICH –Q2 (R1) (2005). Guideline on Validation of Analytical Procedure: Text and Methodology [Google Scholar]
- 8.International Conference on Harmonisation. ICH - Q1B (1996). Photo stability testing of new drug substances and products [Google Scholar]
- 9.International Conference on Harmonisation. Q1A - (R2) (2003). Stability testing of New Drug Substances and Products. [PubMed] [Google Scholar]
- 10.Singh S, Bakshi M. Guidance on conduct of stress tests to determine inherent stability of drugs. Pharmaceutical Technology Online. 2000 Apr;24:1–14. [Google Scholar]