Abstract
Absolute measurements of the thermal conductivity of a distilled and dried sample of toluene near saturation are reported. The transient hot-wire technique with an anodized tantalum hot wire was used. The thermal conductivities were measured at temperatures from 300 K to 550 K at different applied power levels to assess the uncertainty with which it is possible to measure liquid thermal conductivity over wide temperature ranges with an anodized tantalum wire. The wire resistance versus temperature was monitored throughout the measurements to study the stability of the wire calibration. The relative expanded uncertainty of the resulting data at the level of 2 standard deviations (coverage factor k = 2) is 0.5 % up to 480 K and 1.5 % between 480 K and 550 K, and is limited by drift in the wire calibration at temperatures above 450 K. Significant thermal-radiation effects are observed at the highest temperatures. The radiation-corrected results agree well with data from transient hot-wire measurements with bare platinum hot wires as well as with data derived from thermal diffusivities obtained using light-scattering techniques.
Keywords: anodized tantalum wire, liquid, thermal conductivity, thermal radiation, toluene, transient hot-wire
1. Introduction
Saturated liquid toluene has been widely studied and is recommended by the International Union of Applied Chemistry (IUPAC) as a reference standard for thermal conductivity from 189 K to 360 K [1]. Efforts to extend this temperature range to 553 K have been recently reported by Ramires et al. [2]. The barriers to obtaining reliable high-temperature reference standards for thermal conductivity are a lack of data from multiple experimental techniques, and increased uncertainty due to the effects of thermal radiation. Both transient and steady-state measurement techniques for the determination of thermal conductivity are susceptible to errors due to thermal-radiative heat transfer at high temperatures since temperature gradients are imposed during the measurement, and fluids such as toluene absorb and emit the associated thermal radiation [3–5]. Even though thermal-radiation errors may be present in data from transient and steady-state techniques, agreement between radiation-corrected data from these two different techniques would provide evidence of the accuracy of the data. Unfortunately, the relative uncertainty (at the level of 2 standard deviations) of available steady-state thermal conductivity data that have been corrected for radiation exceeds 1 %, which is desired for the development of reference standards [2]. As a result, only data from a single transient hot-wire instrument [4] were designated as primary data during the development of the previous reference standard by Ramires et al. [2].
Recently, thermal diffusivity has been measured for saturated liquid toluene from 293 K to 523 K using light scattering [6]. These light-scattering data have a relative uncertainty of 2.5 %; and because there are no significant thermal gradients in the sample, thermal radiation errors are not present. The thermal conductivity λ can be calculated from the thermal diffusivity a by using
| (1) |
where ρ is the fluid mass density and Cp is the isobaric specific heat. An accurate equation of state is available for toluene [7] to calculate ρ and Cp, but uncertainties in Cp must be considered during this process. The light-scattering data were not available during the development of the previous reference standard of Ramires et al. [2].
The present measurements are made using the transient hot-wire technique as used previously [4,5]. The previous measurements were made with bare 12.7 μm diameter platinum wires and were corrected for thermal radiation. The present measurements were made with anodized 25 μm diameter tantalum wires. Anodized tantalum wires have the advantage of being electrically insulated from the fluid under study. This anodized coating allows measurements of electrically conducting fluids such as water. The highest temperature at which toluene has been previously studied with an anodized-tantalum hot-wire instrument is 370 K [8]. The present measurements extend the temperature range at which anodized tantalum hot wires have been used to measure thermal conductivity from 370 K to 550 K. The use of anodized-tantalum hot wires allows some evaluation of the reliability of the thermal radiative correction for absorbing media since the anodized tantalum wires have a different emissivity (that of tantalum pentoxide) than those of the previous platinum wires, and the diameter of the tantalum wire is twice that of the previous platinum wires.
2. Experimental
The transient hot-wire technique is widely recognized as an accurate method to measure the thermal conductivity and thermal diffusivity of fluids. The present measurements are absolute and require only knowledge of the geometry of the hot wires, the applied power, the resistance-versus-temperature characteristics of the wires, and time. The ideal working equation is based on the heat transfer from an infinitely long line source into an infinite medium. The temperature rise of the fluid at the surface of the wire, where r = r0, is given [9] by
| (2) |
where q is the power divided by the length of the wire, t is the elapsed time, and C = eγ = 1.781 … is the exponential of Euler’s constant. The ideal temperature rise of the wire is linear with respect to the logarithm of elapsed time, as shown in Eq. (2). The thermal conductivity is obtained from the slope, and the thermal diffusivity is obtained from the intercept using linear regression [10]. The temperature associated with a given thermal conductivity data point is given by
| (3) |
where ΔTinitial and ΔTfinal are the temperature rise at the start time and the end time of the linear region selected for the regression. The thermal diffusivity is associated with the initial cell temperature T0 and is obtained from a calibrated reference-standard platinum resistance thermometer (PRT). All temperatures in this work are reported according to the 1990 International Temperature Scale (ITS 90), and all uncertainties are expanded uncertainties at the level of 2 standard deviations (coverage factor k = 2, 95 % level of confidence).
The experimental cell is designed to approximate this ideal model as closely as possible. There are, however, a number of corrections that account for deviations between the ideal line-source solution and the actual experimental heat transfer. The ideal temperature rise is obtained by adding a number of corrections δTi to the experimental temperature rise according to
| (4) |
These temperature-rise corrections are described in detail for our case of a coated wire in Refs. [3,11,12]. Our implementation of the corrections follows these references with the following exceptions. The compression work correction δT3 and the radial convection correction δT4 are set to zero following the recommendations of Assael et al. [13]. The thermal-radiation correction δT5 is described for absorbing fluids by Nieto de Castro et al. [5].
The data-acquisition system used in this work has been described previously [4], and consists of a microcomputer with a 16 bit analog-to-digital converter, three digital voltmeters, a digital power supply, and a Wheatstone bridge which contains two hot wires in opposing legs of the bridge. The two hot wires have different lengths, and the Wheatstone bridge, which is initially balanced, subtracts the resistance change of the short hot wire from the resistance change of the long hot wire. Thus, if both wires are immersed in the same fluid, the bridge response behaves as for a finite length (the difference between the wire lengths) of an infinitely long wire and the end effects arising from axial conduction are eliminated. Heating voltage is applied to the wires through the Wheatstone bridge, and the bridge imbalance is measured in 250 equal time increments. The total time for the measurements may be varied from 1 s to 40 s, allowing one to verify that the data are obtained prior to the onset of convection. The computer checks the bridge balance prior to each experiment and records the temperature of the reference thermometer and the resistance of each hot wire for calibration purposes. The cell temperature is measured with an uncertainty of 1 mK by use of a current source and a standard resistor in series with the reference-standard PRT. The cell pressure is measured with a quartz pressure transducer from 0 MPa to 70 MPa with an uncertainty of 0.007 MPa.
2.1 Hot-Wire Cell
The hot-wire cell used in these measurements was designed for measurements on corrosive solutions at temperatures from 300 K to 550 K at pressures up to 70 MPa. The design of the pressure vessel and temperature control system is the same as for our previous high-temperature cells [4], so it will only be briefly described here. The pressure vessel and the internal components of the hot-wire cell are constructed of nickel alloy UNS-N10276, which is particularly resistant to halides and halide salts. The cell contains long and short hot wires located within long and short cavities with diameters of 9.5 mm. The total volume of the cell and supporting pressure system is relatively small, about 50 cm3, to facilitate measurements on scarce or hazardous materials. There are separate voltage and current leads to each end of each hot wire to eliminate the effects of lead resistance during the calibration process. All electrical connections in the cell are spot welded. In the assembly of the tantalum hot wires (25.4 μm diameter), polytetrafluoroethane-insulated nickel/chromium alloy wire (254 μm diameter) and polyimide-insulated platinum (76 μm diameter) wire were used to make electrical connections inside the pressure system. The connections were welded, then all the bare leads were coated three times with a polyimide/polytetrafluoroethane resin, and the assembly was baked at 550 K for several hours. The baking both cured the polymer resin and annealed the tantalum hot wires from their initial hard-drawn condition. The tantalum hot wires were then anodized in aqueous citric acid with up to 50 V to produce a film of tantalum pentoxide with an estimated thickness of 70 nm. Although the temperature control and pressure systems are rated to 750 K, the upper operating temperature of the present tantalum hot-wire cell assembly is limited to 550 K because of the melting point of polytetrafluoroethane used to electrically insulate the lead wires.
2.2 Sample Purification
The toluene sample used in these measurements was prepared from spectroscopic-grade toluene. The toluene was further purified by distillation over calcium hydride. Calcium hydride reacts irreversibly with any water in the sample to form calcium hydroxide precipitate, which remains in the distillation flask. The principal impurity is benzene, which has a lower boiling temperature, so the initial condensate is discarded. The sample used for measurement is then collected when the inlet to the condenser is stable at the boiling temperature of toluene. The purified sample was analyzed by gas chromatography and found to have less than 50 ng/g of benzene and less than 100 ng/g of water. The sample preparation procedure was the same as in our previous measurements [4] using bare platinum hot wires.
2.3 Wire Calibration
Platinum is preferred for use in resistance thermometry when it is properly annealed and is free from stress because its resistance is very stable for prolonged periods of time. Tantalum has not been used widely for resistance thermometry, so the stability of its resistance must be carefully examined and characterized. The situation is further complicated since the present tantalum wires have been anodized to form a protective layer of tantalum pentoxide. The electrical resistivity of tantalum from 273 K to 1273 K has been shown to increase in proportion to the concentration of oxygen in the sample [14]. Since the concentration of oxygen is not uniform through the wire’s cross section, there is a possibility that oxygen from the anodized layer might diffuse into the bulk tantalum wire and alter its resistance. Any oxygen-diffusion process would be enhanced at higher temperatures. The present system is ideally suited for characterizing the anodized tantalum wires since the wire resistance is measured for each wire, along with the temperature from the reference PRT and the pressure from the quartz pressure transducer, during the balance cycle for each measurement. The instrument also maintains a record of the time and date of each measurement to allow examination of the stability of the wire calibration. Since the present measurements were made along the saturation line of toluene, at pressures less than 3.3 MPa, there is not enough pressure range to allow characterization of the pressure dependence. The electrical resistivity of tantalum is known to decrease in proportion to the pressure on the sample [15].
To eliminate uncertainty due to the resistance of the lead wires, there are separate current-supply and voltage-sensing leads to each end of each hot wire. The resistance is measured during the balance cycle by measuring the voltage drop across standard resistors in each leg of the Wheatstone bridge containing the hot wires, together with the voltage drop across each hot wire. The measured value is the average of five readings with a forward current of about 0.3 mA and five readings with the current reversed. This process minimizes uncertainty due to thermoelectric voltages at weld junctions and electrical connectors. The uncertainty of the resistance measurement is about 0.003 Ω. The measurements were made first with temperature increasing from 300 K to 550 K. Then, measurements were made at 550 K in the morning and evening for a period of 4 days. Finally, measurements were made with temperature decreasing from 550 K to 300 K so that hysteresis effects could be examined. The electrical resistance of the long (188.08 mm) and short (49.07 mm) hot wires are shown in Fig. 1 during this temperature cycle. It is apparent in the figure that the resistances of both the long and short hot wires increased with the elapsed time at 550 K. Although the resistance of the long wire increased more than that of the short wire, the increase was not in proportion to the wire lengths, as would be expected if the process were uniform over the entire length of each wire.
Fig. 1.
Resistance of anodized tantalum hot wires as a function of temperature. Arrows indicate the sequence of elapsed time (first increasing then decreasing temperature).
Since both the long and short hot wires come from the same sample of tantalum wire, the resistance of each wire should scale with the length of each wire. This is a requirement for use of these wires in the transient hot-wire experiment. To insure uniform heat generation over the length of each hot wire, the resistance divided by the length of the long and short hot wires respectfully must be very nearly equal. It is desirable to characterize both the uniformity of the power generation in both hot wires and the adequacy of compensation for the end effects by using two wires in different arms of the measuring bridge. This can be measured by taking the ratio σlw/σsw between the resistance divided by the length of the long wire, σlw = Rsw/Lsw, and the resistance divided by the length of the short wire, σsw = Rsw/Lsw, where R is the wire resistance, L is the wire length, and the subscripts lw and sw designate the long wire and short wire, respectively. Following a previous recommendation by Kestin and Wakeham [16], σlw/σsw must not deviate from unity by more than 2 %, in order to assume a correct end-effect compensation with nearly identical wires. With the definition
| (5) |
Figure 2 shows this percentage difference as a function of elapsed time. It can be seen that this deviation was quite stable and nearly zero during the experiments at increasing temperature from 300 K to the start of the 550 K isotherm but began to drift as the resistance of both wires increased. There was a decrease of 2 % by the end of the 550 K isotherm. This behavior is compatible with previous calculations by Kestin and Wakeham [16].
Fig. 2.
Plot of the quantity Δσ as a function of time, where Δσ = 100(σlw/σsw−1) with σlw = Rlw/Llw and σsw = Rsw/Lsw. Here R is the wire resistance, L is the wire length, and the subscripts lw and sw designate the long wire and short wire, respectively.
However, the increase in resistance during the 4 days at 550 K is quite dramatic and unexpected, based on our previous experience with pure platinum hot wires [4]. With pure platinum the normal behavior is a slight decrease in resistance due to annealing and stress release in the hot wires if the wires have not been at this temperature recently. Figure 2 clearly shows that the increase in resistance was not consistent with wire length. Thus, resistance increases at the welds must be considered a possibility. The only weld locations that can contribute to the measured resistance during a four terminal measurement occur where the ends of the 25 μm diameter tantalum wires are joined to the 254 μm diameter nickel-chromium alloy lead wires. Each hot wire has two welds which could potentially contribute to the measured resistance, but it is not possible to separate contributions due to the welds from those due to changes in the wires. The use of all-tantalum lead wires should be examined in the future to see whether this resistance increases within the welds or within the wires themselves.
Based on Fig. 1 and the requirement that Δσ ≤ 2 %, it was decided that only the data at increasing temperatures, including the first few hours at 550 K, should be considered for the wire-resistance calibration. The wire’s resistance was fit to a quadratic polynomial in temperature of the form R(T) = A1+A2T+A3T2 over four regions, 300 K to 450 K, 300 K to 480 K, 300 K to 515 K, and 300 K to 550 K. The results of these fits are given in Table 1. The resistance of a pure metal such as tantalum is known to be well approximated by such a quadratic expression [17], and the sign of A3 should be small and negative over this temperature range. Table 1 shows that the sign of the quadratic coefficients change if the resistance data above 480 K are included in the fits. This is a good indication that the increase in resistance with time becomes significant at temperatures above 480 K.
Table 1.
Calibration coefficients for the anodized tantalum hot wires
| Temp./Wire | A1 (Ω) | A2 (Ω K−1) | A3 (Ω K−2) |
|---|---|---|---|
| 450 K / Long | 6.425 77 | 0.179 655 | −0.595 598 × 10−5 |
| 450 K / Short | 1.811 64 | 0.463 252 × 10−1 | −0.102 980 × 10−5 |
| 480 K / Long | 6.642 31 | 0.178 426 | −0.424 702 × 10−5 |
| 480 K / Short | 1.902 16 | 0.458 112 × 10−1 | −0.315 381 × 10−6 |
| 515 K / Long | 9.252 54 | 0.164 042 | 0.150 564 × 10−4 |
| 515 K / Short | 2.666 52 | 0.415 993 × 10−1 | 0.533 726 × 10−5 |
| 550 K / Long | 9.471 42 | 0.162 877 | 0.165 577 × 10−4 |
| 550 K / Short | 2.697 78 | 0.414 329 × 10−1 | 0.555 170 × 10−5 |
Given that the increase in resistance with time is significant at 515 K, the best possible calibration must be determined, and the influence of uncertainty in the calibration on the thermal conductivity results must be assessed. The rise in temperature at any elapsed time during a measurement is obtained from the change in resistance of the hot wires using the derivative of the calibration curve for resistance versus temperature:
| (6) |
Thus, the uncertainty of the wire’s temperature rise, and of the measured thermal conductivity, is directly related to the uncertainty of the derivative of wire resistance with respect to temperature. This resistance derivative divided by length is plotted as a function of temperature in Fig. 3, which shows the effect of the change in sign of A3 when resistance data at temperatures above 480 K are included in the fit. Based on data below 480 K, where there is confidence that the calibration is stable, the resistance derivative decreases by 1 % for a temperature increase of 180 K. Since this temperature dependence is quite small, and the region of extrapolation (480 K to 550 K) is less than half this temperature range, it is anticipated that extrapolation errors should be less than 1 % if only data below 480 K are used for the calibration. If resistance data from temperatures above 480 K are used in the calibration, the thermal conductivity results at 550 K will be about 4 % higher at 550 K and results at 300 K will be 2 % too low. Based on these considerations, the wire calibration from 300 K to 480 K is used in the subsequent data analysis, with the estimate that the relative uncertainty of the measured thermal conductivity increases above 480 K by the additive amount of 1.0 % due to extrapolation in the wire calibration.
Fig. 3.
Values of divided by the length of anodized tantalum wire.
3. Results
The thermal conductivity results for the purified toluene sample are given in Table 2 and are shown in Fig. 4. The results in Table 2 have been corrected for thermal radiation, as were our previous results using bare platinum hot wires [4], and with the same empirical optical parameters that were found for the fluid using the previous platinum hot wires [5]. Since the power divided by length was not equal for the long and short wires after the isotherm at 550 K, as shown in Fig. 2, the data from the decreasing temperature portion of the temperature cycle are not reported. There are 184 thermal conductivity data points at temperatures from 300 K to 550 K.
Table 2.
Thermal conductivity of toluene near the saturated liquid line
| T (K) | p (MPa) | λ (W m−1 K−1) | q (W m−1) | T (K) | p (MPa) | λ (W m−1 K−1) | q (W m−1) |
|---|---|---|---|---|---|---|---|
| 301.248 | 3.246 | 0.128 18 | 0.344 40 | 381.226 | 0.092 | 0.107 47 | 0.391 08 |
| 301.150 | 3.252 | 0.128 20 | 0.317 30 | 381.078 | 0.088 | 0.107 42 | 0.359 11 |
| 301.051 | 3.253 | 0.128 24 | 0.291 35 | 380.929 | 0.087 | 0.107 76 | 0.328 51 |
| 300.950 | 3.258 | 0.128 27 | 0.266 53 | 380.804 | 0.084 | 0.107 69 | 0.299 29 |
| 300.856 | 3.260 | 0.128 40 | 0.242 81 | 380.672 | 0.081 | 0.107 71 | 0.271 43 |
| 300.762 | 3.260 | 0.128 52 | 0.220 20 | 380.551 | 0.079 | 0.107 81 | 0.244 93 |
| 300.676 | 3.263 | 0.128 49 | 0.198 70 | 380.429 | 0.078 | 0.107 73 | 0.219 80 |
| 300.595 | 3.266 | 0.128 45 | 0.178 31 | 381.304 | 0.293 | 0.107 64 | 0.424 55 |
| 301.231 | 0.077 | 0.127 25 | 0.344 42 | 381.166 | 0.292 | 0.107 50 | 0.391 11 |
| 301.134 | 0.075 | 0.127 23 | 0.317 30 | 381.024 | 0.290 | 0.107 53 | 0.359 06 |
| 301.030 | 0.067 | 0.127 35 | 0.291 32 | 380.889 | 0.290 | 0.107 76 | 0.328 46 |
| 300.934 | 0.059 | 0.127 28 | 0.266 49 | 380.759 | 0.288 | 0.107 77 | 0.299 25 |
| 300.834 | 0.053 | 0.127 36 | 0.242 78 | 380.620 | 0.288 | 0.107 85 | 0.271 38 |
| 300.745 | 0.047 | 0.127 35 | 0.220 17 | 380.509 | 0.286 | 0.107 81 | 0.244 89 |
| 300.661 | 0.044 | 0.127 49 | 0.198 68 | 380.391 | 0.285 | 0.107 94 | 0.219 76 |
| 300.580 | 0.040 | 0.127 44 | 0.178 29 | 414.154 | 0.272 | 0.100 76 | 0.224 35 |
| 370.214 | 0.101 | 0.110 17 | 0.413 57 | 414.288 | 0.271 | 0.100 71 | 0.250 66 |
| 370.078 | 0.101 | 0.110 14 | 0.381 02 | 414.416 | 0.271 | 0.100 64 | 0.278 46 |
| 369.941 | 0.101 | 0.110 13 | 0.349 85 | 414.559 | 0.270 | 0.100 60 | 0.307 76 |
| 369.820 | 0.101 | 0.110 33 | 0.320 03 | 414.709 | 0.270 | 0.100 43 | 0.338 56 |
| 369.698 | 0.101 | 0.110 38 | 0.291 55 | 414.858 | 0.269 | 0.100 23 | 0.370 81 |
| 369.578 | 0.101 | 0.110 48 | 0.264 41 | 415.036 | 0.269 | 0.100 38 | 0.404 56 |
| 369.467 | 0.101 | 0.110 49 | 0.238 59 | 415.201 | 0.269 | 0.100 37 | 0.439 75 |
| 369.368 | 0.101 | 0.110 51 | 0.214 11 | 451.768 | 0.522 | 0.093 51 | 0.494 27 |
| 370.250 | 0.324 | 0.110 24 | 0.413 58 | 451.575 | 0.522 | 0.093 69 | 0.455 48 |
| 370.115 | 0.310 | 0.110 21 | 0.381 03 | 451.393 | 0.522 | 0.093 73 | 0.418 30 |
| 369.982 | 0.297 | 0.110 11 | 0.349 82 | 451.223 | 0.522 | 0.093 81 | 0.382 68 |
| 369.859 | 0.286 | 0.110 39 | 0.320 01 | 451.044 | 0.522 | 0.093 77 | 0.348 64 |
| 369.739 | 0.276 | 0.110 39 | 0.291 52 | 450.874 | 0.522 | 0.093 76 | 0.316 18 |
| 369.611 | 0.267 | 0.110 43 | 0.264 38 | 450.720 | 0.521 | 0.093 71 | 0.285 32 |
| 369.500 | 0.259 | 0.110 43 | 0.238 57 | 450.573 | 0.521 | 0.093 68 | 0.256 05 |
| 369.393 | 0.252 | 0.110 50 | 0.214 09 | 451.732 | 0.859 | 0.093 87 | 0.494 47 |
| 381.373 | 0.093 | 0.107 52 | 0.424 40 | 451.539 | 0.858 | 0.093 74 | 0.455 50 |
| 451.364 | 0.858 | 0.093 90 | 0.418 24 | 552.555 | 2.567 | 0.077 98 | 0.382 26 |
| 451.172 | 0.857 | 0.093 45 | 0.382 64 | 552.367 | 2.567 | 0.078 05 | 0.344 96 |
| 451.017 | 0.857 | 0.093 60 | 0.348 60 | 552.182 | 2.566 | 0.078 22 | 0.309 59 |
| 450.855 | 0.854 | 0.093 90 | 0.316 16 | 552.004 | 2.565 | 0.077 92 | 0.276 13 |
| 450.711 | 0.855 | 0.093 91 | 0.285 31 | 551.852 | 2.563 | 0.078 39 | 0.244 58 |
| 450.566 | 0.855 | 0.093 93 | 0.256 03 | 551.688 | 2.562 | 0.078 45 | 0.214 95 |
| 451.752 | 0.802 | 0.093 62 | 0.494 80 | 551.547 | 2.562 | 0.078 54 | 0.187 25 |
| 451.567 | 0.802 | 0.093 91 | 0.455 83 | 553.467 | 2.628 | 0.078 57 | 0.606 40 |
| 451.382 | 0.802 | 0.093 87 | 0.418 52 | 553.200 | 2.632 | 0.078 48 | 0.558 67 |
| 451.213 | 0.803 | 0.093 79 | 0.382 85 | 552.981 | 2.635 | 0.078 35 | 0.512 97 |
| 451.037 | 0.804 | 0.093 72 | 0.348 77 | 552.771 | 2.637 | 0.078 42 | 0.469 24 |
| 450.876 | 0.803 | 0.094 02 | 0.316 29 | 552.550 | 2.640 | 0.078 25 | 0.427 50 |
| 450.711 | 0.804 | 0.093 53 | 0.285 41 | 552.365 | 2.641 | 0.078 22 | 0.387 73 |
| 450.566 | 0.805 | 0.093 72 | 0.256 14 | 552.165 | 2.643 | 0.078 15 | 0.349 90 |
| 483.230 | 0.922 | 0.088 56 | 0.525 51 | 551.991 | 2.645 | 0.078 16 | 0.314 01 |
| 483.013 | 0.917 | 0.088 51 | 0.484 20 | 551.816 | 2.646 | 0.078 32 | 0.280 08 |
| 482.805 | 0.913 | 0.088 46 | 0.444 67 | 551.656 | 2.647 | 0.078 39 | 0.248 07 |
| 482.598 | 0.910 | 0.088 76 | 0.406 78 | 551.501 | 2.648 | 0.078 38 | 0.218 03 |
| 482.413 | 0.908 | 0.088 45 | 0.370 60 | 551.367 | 2.649 | 0.078 70 | 0.189 92 |
| 482.234 | 0.906 | 0.087 86 | 0.336 12 | 553.540 | 2.657 | 0.078 63 | 0.607 76 |
| 482.064 | 0.905 | 0.088 64 | 0.303 31 | 553.293 | 2.658 | 0.078 43 | 0.559 88 |
| 481.906 | 0.903 | 0.088 08 | 0.272 20 | 553.057 | 2.658 | 0.078 34 | 0.514 02 |
| 483.220 | 1.574 | 0.088 83 | 0.525 69 | 552.830 | 2.659 | 0.078 33 | 0.470 20 |
| 483.019 | 1.574 | 0.088 93 | 0.484 21 | 552.604 | 2.659 | 0.078 27 | 0.428 39 |
| 482.808 | 1.574 | 0.088 62 | 0.444 55 | 552.393 | 2.659 | 0.078 29 | 0.388 53 |
| 482.614 | 1.572 | 0.088 99 | 0.406 68 | 552.193 | 2.659 | 0.078 15 | 0.350 62 |
| 482.433 | 1.572 | 0.088 58 | 0.370 53 | 552.001 | 2.659 | 0.078 19 | 0.314 68 |
| 482.263 | 1.571 | 0.088 94 | 0.336 08 | 551.814 | 2.659 | 0.078 22 | 0.280 69 |
| 482.086 | 1.571 | 0.088 52 | 0.303 29 | 551.652 | 2.658 | 0.078 39 | 0.248 62 |
| 481.934 | 1.572 | 0.088 45 | 0.272 17 | 551.492 | 2.659 | 0.078 58 | 0.218 50 |
| 518.628 | 1.534 | 0.083 52 | 0.562 32 | 551.342 | 2.659 | 0.078 88 | 0.190 33 |
| 518.393 | 1.533 | 0.083 38 | 0.518 02 | 553.646 | 2.635 | 0.078 65 | 0.614 86 |
| 518.162 | 1.533 | 0.083 24 | 0.475 68 | 553.388 | 2.635 | 0.078 54 | 0.566 49 |
| 518.032 | 1.533 | 0.083 08 | 0.435 19 | 553.139 | 2.635 | 0.078 37 | 0.520 16 |
| 517.748 | 1.533 | 0.082 75 | 0.396 48 | 552.896 | 2.634 | 0.078 32 | 0.475 86 |
| 517.546 | 1.532 | 0.083 05 | 0.359 60 | 552.668 | 2.635 | 0.078 28 | 0.433 54 |
| 517.348 | 1.532 | 0.082 93 | 0.324 50 | 552.445 | 2.634 | 0.078 12 | 0.393 21 |
| 517.179 | 1.532 | 0.082 96 | 0.291 20 | 552.240 | 2.634 | 0.078 02 | 0.354 85 |
| 517.010 | 1.532 | 0.082 51 | 0.259 72 | 552.047 | 2.634 | 0.078 08 | 0.318 45 |
| 516.792 | 1.532 | 0.083 03 | 0.230 05 | 551.852 | 2.634 | 0.078 20 | 0.284 03 |
| 516.638 | 1.532 | 0.082 92 | 0.202 18 | 551.684 | 2.634 | 0.078 19 | 0.251 58 |
| 516.501 | 1.532 | 0.083 14 | 0.176 11 | 551.517 | 2.634 | 0.078 32 | 0.221 11 |
| 518.578 | 2.143 | 0.083 67 | 0.562 45 | 551.367 | 2.633 | 0.078 57 | 0.192 61 |
| 518.347 | 2.144 | 0.083 66 | 0.518 04 | 553.163 | 2.619 | 0.078 43 | 0.619 33 |
| 518.147 | 2.143 | 0.083 57 | 0.475 63 | 553.346 | 2.619 | 0.078 82 | 0.571 03 |
| 517.928 | 2.142 | 0.083 73 | 0.435 12 | 553.083 | 2.620 | 0.078 42 | 0.524 29 |
| 517.670 | 2.142 | 0.083 64 | 0.396 47 | 552.840 | 2.620 | 0.078 31 | 0.479 62 |
| 517.482 | 2.141 | 0.083 50 | 0.359 57 | 552.612 | 2.620 | 0.078 34 | 0.436 95 |
| 517.275 | 2.142 | 0.083 63 | 0.324 48 | 552.388 | 2.620 | 0.078 17 | 0.396 28 |
| 517.094 | 2.142 | 0.083 67 | 0.291 19 | 552.169 | 2.621 | 0.077 89 | 0.357 63 |
| 516.915 | 2.142 | 0.083 74 | 0.259 73 | 551.974 | 2.621 | 0.078 02 | 0.320 95 |
| 516.764 | 2.143 | 0.083 71 | 0.230 05 | 551.784 | 2.621 | 0.078 20 | 0.286 26 |
| 516.606 | 2.143 | 0.083 74 | 0.202 18 | 551.607 | 2.621 | 0.078 22 | 0.253 55 |
| 516.467 | 2.143 | 0.083 80 | 0.176 10 | 551.436 | 2.622 | 0.078 24 | 0.222 84 |
| 553.594 | 2.565 | 0.077 94 | 0.597 49 | 551.285 | 2.622 | 0.078 20 | 0.194 11 |
| 553.379 | 2.567 | 0.077 93 | 0.550 59 | 552.155 | 2.642 | 0.078 09 | 0.362 33 |
| 553.178 | 2.566 | 0.078 04 | 0.505 64 | 551.949 | 2.642 | 0.077 98 | 0.325 12 |
| 552.967 | 2.567 | 0.078 10 | 0.462 60 | 551.762 | 2.642 | 0.078 15 | 0.289 94 |
| 552.740 | 2.567 | 0.077 83 | 0.421 46 | 551.579 | 2.642 | 0.078 21 | 0.256 80 |
Fig. 4.
Thermal conductivity of saturated liquid toluene as a function of temperature. The correction for thermal radiation increases with T3 to about 3 % at 550 K.
3.1 Repeatability at High Temperatures
The isotherm at 550 K is quite interesting since it includes several replications, with a wide range of power levels, over the course of four days. During this time, the resistance of the long hot wire increased by 5.6 Ω and that of the short hot wire increased by 1.9 Ω. In addition, since this increase in resistance was not proportional to the wire lengths, the power generation of the two wires differed by up to 1.6 %. Dispite these complications, there seems to be little additional scatter in the thermal conductivity results measured during this 4 day period. In Fig. 5, deviations between the results for the 550 K isotherm are plotted relative to the reference standard of Ramires et al. [2] as a function of applied power level at 550 K. The mean deviation of the data (solid line) is 1.45 % higher than the earlier reference standard, and the scatter (dashed lines) is ±0.6 % at the level of 95 % confidence. No trend is noted with respect to time throughout this four-day period. Convection in the sample occurs at shorter times for experiments with higher powers and correspondingly larger temperature rises. Consistency between thermal conductivity results at different power levels is considered a good indication that there was no significant convection during the measurements and that compensation for the wire’s end effects was achieved. Data at power levels above 0.5 Wm−1 appear to have some influence due to the onset of convection since this isotherm is relatively close to the critical point of toluene at 593.95 K. The contribution of convection on the apparent thermal conductivity appears to be less than 0.5 % at even the highest power levels.
Fig. 5.
Thermal conductivity of saturated liquid toluene as a function of applied power near 550 K. Different plot symbols designate the average elapsed time of the experiment series. The mean deviation of the data (solid line) is 1.45 % higher than the earlier reference standard, and the scatter (dashed lines) is ±0.6 % at the level of 95 % confidence.
3.2 Uncertainty Assessment
The contribution of thermal radiation to measurements of the thermal conductivity of fluids such as toluene has been a topic of debate for many years. An empirical technique has been described for correcting for thermal radiation in transient hot-wire measurements [5]. Empirical optical parameters have been reported for toluene [5] at these same conditions based on measurements with platinum hot wires 12.7 μm in diameter. If this radiation correction is valid, then these same optical parameters should apply to the present case of a tantalum wire 25 μm in diameter. The radiation correction assumes that thermal emission from the wire is small compared to emission from the expanding thermal front in the fluid, so there should be little effect due to changing the emissivity of the wire material. The contribution of thermal radiation is insignificant at 300 K, increases as T3 and is estimated to be 3 % at 550 K. The experimental thermal conductivity data with and without this radiative correction are shown in Fig. 4.
Deviations between the reference standard of Ramires et al. [2] and the present data are shown in Fig. 6. High-temperature data sets from light scattering [6] as well as transient hot-wire experiments using bare platinum [4,18] and anodized tantalum[8] hot wires are also compared in this figure. The data of Kraft et al. [6] were obtained from thermal diffusivities by light scattering using the equation of state of Goodwin [7]. These light-scattering data are free from uncertainty due to thermal radiation but are believed to have a relative uncertainty of 2.5 % in thermal diffusivity. The uncertainty in the thermal conductivity obtained from the expression λ = aρCp is estimated to vary between 3 % and 4.2 % since the uncertainties in ρ and Cp from the equation of state [7] are 0.2 % for ρ and varies from 0.2 % to 3 % for Cp. The thermal conductivity obtained from light scattering is offset from the direct thermal conductivity measurements by about 2 % to 3 %, but agreement is still within the combined uncertainty of the data sets. It is unlikely that the offset is due to thermal radiation since it is nearly constant and thermal-radiative errors should increase as T3.
Fig. 6.
Deviations between the thermal conductivity data, [4,6,8,18] and present work, and the reference standard of Ramires et al. [2].
No systematic difference between transient hot-wire measurements using bare and anodized tantalum hot wires is apparent in Fig. 6. The present tantalum hot-wire data are lower than the other transient data [4,8,18] at 300 K by about 2 %. This is partly due to larger cell fluctuations of temperature in the furnace containing the hot-wire cell near ambient temperature. The remainder of this difference is likely due to the drift in calibration of the tantalum hot wires at elevated temperatures, as shown in Fig. 3. The uncertainty of the present tantalum measurements are evaluated as 1 % at 300 K, 0.5 % from 369 K to 480 K, and 1.5 % from 480 K to 550 K. The transient hot-wire data using platinum or anodized tantalum wires agree within their combined uncertainties over the entire temperature range, with the exception of the data of Yamada et al. [18] at temperatures above 400 K. The purity of the toluene sample used in this case [18] was stated to be 99.7 % by the supplier. The purity of the toluene sample used in the light scattering study [6] was stated to be 99.9 % by the supplier. The transient hot-wire data of Perkins et al. [4] and Ramires et al. [8] were made on purified samples of toluene as described in the present work.
4. Conclusions
The present measurements demonstrate that anodized-tantalum hot wires can be used to make absolute measurements of the thermal conductivity a liquid from 300 K to 550 K. Previous studies with anodized-tantalum hot wires have been limited to temperatures below 370 K [8]. The present transient hot-wire measurements using anodized-tantalum hot wires have a larger uncertainty in the temperature extremes than our previous measurements using bare tantalum hot wires [4] over the same temperature range. This is primarily due to drift in the resistance calibration of the anodized tantalum hot wires at high temperatures. Use of tantalum lead wires may reduce or eliminate this problem in the future and allow accurate measurements of the fluid thermal diffusivity. It was also noted during the experiments that convection occurs earlier, and at lower power levels, as the wire diameter increases. Thus, experiments must be done with lower levels of applied power (smaller temperature rises) when the larger tantalum hot wires are used.
The big advantage of anodized-tantalum hot wires is for measurement of electrically conducting fluids, and this is not a problem in the case of toluene. The anodized tantalum hot wires have a geometry and emissivity different from those of platinum, so the present measurements support the validity of the thermal-radiation correction for absorbing fluids [5]. There is no significant temperature trend in deviations between the present radiation-corrected thermal-conductivity data and thermal-diffusivity data derived from light-scattering measurements of thermal diffusivity. This again supports the validity of the radiation correction [5], since the contribution of thermal radiation is expected to increase with T3.
Acknowledgments
We gratefully acknowledge the financial support of this work by the United States Department of Energy, Division of Engineering and Geosciences, Office of Basic Energy Sciences, through DE-AI02-96ER14597. Both M. L. V. Ramires and C. A. Nieto de Castro thank the Faculty of Sciences of the University of Lisbon for a leave of absence and the Physical and Chemical Properties Division of the National Institute of Standards and Technology for support as guest researchers. M. L. V. Ramires also thanks JNICT, Portugal, for an additional grant.
Biography
About the authors: Richard A. Perkins is a chemical engineer in the Physical and Chemical Properties Division of the NIST Chemical Science and Technology Laboratory. Maria L. V. Ramires is also a chemical engineer and Auxiliary Professor at the Faculdade de Ciências da Universidade de Lisboa, and was a Guest Researcher at NIST. Carlos A. Nieto de Castro is a chemical engineer, director of the Centro de Ciência e Tecnologia de Materiais and of the Laboratory of Metrology and Testing of the Instituto de Ciencia Aplicada e Tecnologia, Faculdade de Ciências da Universidade de Lisboa, and a Guest Researcher at NIST. The National Institute of Standards and Technology is an agency of the Technology Administration, U.S. Department of Commerce.
Contributor Information
R. A. Perkins, Email: richard.perkins@nist.gov.
M. L. V. Ramires, Email: mlvr@fc.ul.pt.
C. A. Nieto de Castro, Email: carlos.castro@fc.ul.pt.
6. References
- 1.Nieto de Castro CA, Li SFY, Nagashima A, Trengove RD, Wakeham WA. J Phys Chem Ref Data. 1986;15:1078. [Google Scholar]
- 2.Ramires MLV, Nieto de Castro CA, Perkins RA. High Temp High Press. 1993;25:269. [Google Scholar]
- 3.Wakeham WA, Nagashima A, Sengers JV, editors. Experimental Thermodynamics Volume III. Blackwell Scientific Publications; Oxford: 1991. Measurement of the Transport Properties of Fluids. [Google Scholar]
- 4.Perkins RA, Roder HM, Nieto de Castro CA. J Res Natl Inst Stand Technol. 1991;96:247. doi: 10.6028/jres.096.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Nieto de Castro CA, Perkins RA, Roder HM. Int J Thermophys. 1991;12:985. [Google Scholar]
- 6.Kraft K, Matos Lopes M, Leipertz A. Int J Thermophys. 1995;16:423. [Google Scholar]
- 7.Goodwin RD. J Phys Chem Ref Data. 1989;18:1565. [Google Scholar]
- 8.Ramires MLV, Fareleira JMNA, Nieto de Castro CA, Dix M, Wakeham WA. Int J Thermophys. 1993;14:1119. [Google Scholar]
- 9.Carslaw HS, Jaeger JC. Conduction of heat in solids. 2nd ed. Oxford University Press; London: 1959. [Google Scholar]
- 10.Nieto de Castro CA, Taxis B, Roder HM, Wakeham WA. Int J Thermophys. 1992;9:293. [Google Scholar]
- 11.Healy JJ, de Groot JJ, Kestin J. Physica. 1976;82C:392. [Google Scholar]
- 12.Nagasaka Y, Nagashima A. J Phys E: Sci Instrum. 1981;14:1435. [Google Scholar]
- 13.Assael MJ, Karagiannidis L, Richardson SM, Wakeham WA. Int J Thermophys. 1992;13:223. [Google Scholar]
- 14.Gebhardt E, Seghezzi HD, Metallk Z. 1957;48:430. [Google Scholar]
- 15.Bridgman PW. Proc Amer Acad Arts Sci. 1952;81:165. [Google Scholar]
- 16.Kestin J, Wakeham WA. Physica. 1978;92A:102. [Google Scholar]
- 17.Touloukian YS, editor. Thermophysical properties of high temperature solid materials—Volume 1: Elements. Macmillan; New York: 1967. [Google Scholar]
- 18.Yamada T, Yaguchi T, Nagasaka Y, Nagashima A. High Temp High Press. 1993;25:513. [Google Scholar]






