Electrical interference in non-competitive pacemakers

E Sowton 1, K Gray 1, T Preston 1
PMCID: PMC487384  PMID: 5470044

Abstract

Patients with 41 implanted non-competitive pacemakers were investigated. A variety of domestic electrical equipment, a motor-car, and a physiotherapy diathermy apparatus were each operated in turn at various ranges from the patient. Interference effects on pacemaker function were assessed on the electrocardiograph. Medtronic demand 5841 pacemakers were stopped by diathermy while Cordis Ectocor pacemakers developed a fast discharge rate. Cordis triggered pacemakers (both Atricor and Ectocor) were sensitive to interference from many items of domestic equipment and the motor car. The Elema EM153 ran at an increased rate when an electric razor was running close to the pacemaker. The Devices demand 2980 and the Medtronic demand 5841 were not affected by the domestic equipment tested. The significance of interference effects is discussed in relation to pacemaker design.

Full text

PDF
626

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. CARLETON R. A., SESSIONS R. W., GRAETTINGER J. S. ENVIRONMENTAL INFLUENCE ON IMPLANTABLE CARDIAC PACEMAKERS. JAMA. 1964 Dec 7;190:938–940. doi: 10.1001/jama.1964.03070230074028. [DOI] [PubMed] [Google Scholar]
  2. Chardack W. M., Gage A. A., Federico A. J., Schimert G., Greatbatch W. Five years' clinical experience with an implantable pacemaker: an appraisal. Surgery. 1965 Nov;58(5):915–922. [PubMed] [Google Scholar]
  3. Dodinot B. P., Petitier H. A., Gilgenkrantz J. M., Faivre G. R. Clinical experience with atrial-synchronous pacing. Ann N Y Acad Sci. 1969 Oct 30;167(2):1038–1054. doi: 10.1111/j.1749-6632.1969.tb34167.x. [DOI] [PubMed] [Google Scholar]
  4. LICHTER I., BORRIE J., MILLER W. M. RADIO-FREQUENCY HAZARDS WITH CARDIAC PACEMAKERS. Br Med J. 1965 Jun 12;1(5449):1513–1518. doi: 10.1136/bmj.1.5449.1513. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. SILVER A. W., ROOT G., BYRON F. X., SANDBERG H. EXTERNALLY RECHARGEABLE CARDIAC PACEMAKER. Ann Thorac Surg. 1965 Jul;1:380–388. doi: 10.1016/s0003-4975(10)66774-4. [DOI] [PubMed] [Google Scholar]

Articles from British Heart Journal are provided here courtesy of BMJ Publishing Group