Abstract
Background
The aim of the proposed study was to conduct a feasibility study using a flat rectangular (3 mm × 10 mm) MRI compatible transducer operating at 5.3 MHz for destroying calcified material in an in vitro model. The proposed method can be used in the future for treating atherosclerosis plaques of the coronary, carotid or peripheral arteries.
Methods
The system was tested initially on calcium rods. Another test was performed in a hydroxyapatite–polylactide model.
Results
A parametric study was performed where the mass of calcified material removed was studied as a function of intensity, pulse repetition frequency (PRF), duty factor (DF) and presence of bubbles.
Conclusions
The amount of calcified material removed is directly related to the intensity, PRF and DF. It was found that the presence of bubbles accelerates the removal of calcified material. In order to ensure that pure mechanical mode ultrasound was used, the protocols were designed so that the temperature does not exceed 1 °C.
Keywords: Ultrasound, Atherosclerotic, Plaque, Rabbit
Sommario
Scopo
Scopo dello studio è stato quello di valutare la fattibilità della distruzione di materiale calcifico, in un modello in vitro, con un trasduttore da 5.3 MHz., rettangolare, piatto (3 mm x 10 mm), compatibile con la MRI. Il metodo proposto potrebbe essere utilizzato in futuro per il trattamento di placche aterosclerosi di coronarie, carotidi o arterie periferiche.
Materiale e metodi
Il sistema è stato testato inizialmente su barre di calcio. Un altro test è stato eseguito su un modello di idrossiapatite-polylactide (HA- / PLA). E’ stato eseguito uno studio parametrico in cui la massa di materiale calcificato rimossa è stata studiata in funzione di intensità, frequenza di ripetizione degli impulsi (PRF), fattore di utilizzo (DF) e presenza di bolle.
Risultati
La quantità di materiale calcificato rimossa è direttamente legato a intensità, PRF e DF. Si è riscontrato che la presenza di bolle accelera l’asportazione di materiale calcificato.
Conclusioni
Questa tecnologia potrebbe essere utilizzata in futuro nella pratica clinica quotidiana principalmente per il trattamento di placche calcificate della carotide.
Introduction
Atherosclerosis also known as arteriosclerotic vascular disease (ASVD) is a condition in which fatty material collects along the walls of arteries. This fatty material thickens, hardens (forms calcium deposits), and may eventually block the arteries [1]. The plaque is composed of distinctly morphological features including a fibrous cap comprised of smooth muscle cells and fibrotic tissue, and lipid core containing fat-laden macrophages and extracellular lipids [1]. In the advanced stage, atherosclerotic plaque contains large amounts of calcium salt or hydroxyapatite (HA) [2], which significantly increases the size of the plaque. Histology studies have also led to the recognition that plaque structure influences the risk of plaque rupture. Specifically, a plaque with a thin fibrous cap and a large lipid core is more prone to rupture [3].
Lifestyle changes, such as eating a healthy diet and exercising, are often the best treatment for atherosclerosis. But sometimes, medication or surgical procedures may be recommended as well [4]. Over the years researchers were involved in clinical research to develop medication treatments and approaches to reduce the risk of heart attack and other medical problems caused by atherosclerosis (for example Angiotensin II receptor blockers (ARBs) [5], Calcium Channel Blockers [5], Cholesterol Medications [6], and Diuretics [7]).
Atherosclerosis treatment may require special surgical procedures such as Balloon Angioplasty [8–11], Balloon Angioplasty and Stenting [12, 13], Cutting Balloon [14–19], Atherectomy [20, 21], and Surgical Bypass [22–27] to open an artery and improve blood flow. Endarterectomy is the main treatment for the carotid artery [28–30].
A group in the University of Minnesota [31] recently developed a new HIFU technology that performs noninvasive, real-time ultrasonic imaging and localized treatment using thermal HIFU. A HIFU exposure was applied to a very small, localized treatment area, usually as small as a grain of rice.
In this paper, pulsed ultrasound was utilized for removing calcified material. In order to ensure that non-thermal ultrasound was used, the protocols were designed so that the temperature does not exceed 1 °C. Hydroxyapatite-polylactide (HA–PLA) was investigated as a possible material for atherosclerotic plaque [32]. Recently an ablation experiment was performed on an HA–PLA composite using a high-speed rotational ablation tool [32]. In this particular research, a rotational atherectomy technique was used in which a small grinder is inserted into the coronary arteries to ablate plaque and increase blood flow to the heart. However, to our knowledge, no prior studies have been reported on ablation of calcified material using mechanical mode HIFU.
This paper describes a feasibility study that investigated the effectiveness of a therapeutic protocol in removing calcified material using pulsed ultrasound with a planar unfocused transducer operating at 5.3 MHz. In order to achieve this, the effect of various parameters such as intensity, pulse repetition frequency (PRF), duty factor (DF), and the presence of bubbles were explored. Thus, in vitro experiments were conducted using mainly cylindrical chalks and HA-/PLA composites.
Materials and methods
Experimental setup
Figure 1 shows the schematic diagram of the high intensity ultrasound system. The samples under sonication were placed in an acrylic tank containing filtered, degassed, and de-ionized water. The apparatus is divided into two primary systems: (1) high intensity ultrasound generation, and (2) passive cavitation detection (a hydrophone was employed to monitor cavitation activity). A 5.3 MHz sinusoidal input was generated by a function generator (Agilent 33120 15 MHz Function/Arbitrary Waveform Generator, Englewood, CO, USA). The electrical signal is amplified by an RF amplifier (500 W, AR, Souderton, PA, USA), and then delivered through an impedance matching network designed exclusively for this particular transducer.
Fig. 1.
Diagram of the pulsed high intensity ultrasound system which includes a generator/amplifier, a temperature reader, and a passive cavitation detector
The active size of the transducer is 3 mm × 10 mm. The transducer material is P762-type PZT piezoceramic (Quartz & Silice, Nemours, France) with air backing, operating at 5.3 MHz. Figure 2a shows the transducer holder manufactured using Acrylonitrile Butadiene Styrene (ABS). The ABS parts of the transducer were designed using Computer-Aided Design (CAD) software (Microstation V8, Bentley Systems, Inc.). The files were then exported to Computer-Aided Manufacturing (CAM) software (Insight V. 6.4.1, Stratasys Inc.). The files were sent to a 3D printer (FDM400, Stratasys, 7665 Commerce Way, Eden Prairie, Minnesota, 55344, USA) for production. This structure contains two inlets for transducer cooling and two inlets for the transducer wiring. Figure 2b shows the drawing of the transducer holder indicating the water inlets and wiring inlets. Figure 2c shows the final assembly of the transducer inside the plastic holder. Figure 2d shows the pressure field for this transducer with 6 W applied using the simulation model proposed by Lafon et al. [33]. Because in future clinical trials the transducer will be incorporated in a catheter that is guided through arteries (1–3 mm wide), the transducer element must be as compact as possible. If a spherically focused technology will be used, then the size of the transducer will increase and therefore it will not fit in the catheter. Since the catheter will be inserted in the body, after the treatment, then the catheter is not reusable (i.e., it is considered consumable). Thus, a spherically focused technology has the additional disadvantage that will increase the cost substantially. Besides as it will be seen in the results section, a flat transducer produces enough plaque destruction.
Fig. 2.
a Transducer holder manufactured using ABS. This structure contains two inlets for transducer cooler and two inlets for the transducer wires. b Drawing of the transducer holder indicating the water inlets and wiring inlets. c Final assembly of the transducer inside the plastic holder. d Simulated pressure field for this transducer with 6 W applied
RF connections were made via a miniaturized 10 m long, 50 Ω coaxial cable with a 0.9-mm outer diameter. The outer ground conductor was connected to the external face of the transducer. The inner conductor reached the internal face of the transducer through the lumen of the tube. To limit energy loss during transmission of the electric signal, a connection with a capacitor–inductor network was included to match the transducer. Using the acoustic balance technique 0, the electroacoustic efficiency of the applicator was measured (55 % at 5.3 MHz). The external face of the transducer was cooled by a continuous flow of degassed water circulating within of the transducer. The cooling water serves as coupling medium between the transducer and the ablated tissue. The water cooling circuit was maintained at 15 °C and was driven by a Masterflex peristaltic pump (Cole Parmer Instrument Co., Chicago, IL) at a flow of 0.15 L/min.
Sonication parameters
The Spatial Average intensity was estimated, by dividing the power with the area of the transducer. The PRF was varied from 5 to 100 Hz. The duty factor varied from 5 to 40 %. The sonication duration was set to 30 min. The bubbles from the cylindrical chalk were removed by immersing the chalks in the water container of the degassing system for 1 h. With this method, the voids within the chalk are filled with degassed water and thus the bubble influence is minimized. If the effect of bubbles was needed, then the chalk samples were not immersed in the degassed water bath, and therefore, the bubbles in the voids of the chalk are preserved.
Cavitation detection
The cavitation activity was monitored using a hydrophone (Specialty Eng. Associates, California, USA) which is placed perpendicularly to the transducer. The signal from the hydrophone was fed to a custom made amplifier (20× amplifications), and was high-pass filtered using custom made filter. Fast Fourier Transform (FFT) spectra of the acoustic emission signals were acquired using a PC-card (Gage, Lockport, USA). The hydrophone is aligned perpendicular to the sample under investigation.
In vitro experiments
Twenty in vitro experiments were carried out initially in cylindrical chalks with dimensions of 80 mm in length and 9 mm in diameter. The efficacy of pulsed ultrasound was also investigated in vitro during experiments on artificial materials that have elastic moduli similar to those of calcified plaque, such as HA–PLA composite. HA–PLA was used in this research because of the high HA content of calcified plaque. The PLA (4032D, NatureWorks, Minnetonka, MN, USA), 40 wt%, was dissolved in Dimethylformamide (A.J. Vouros Ltd, Nicosia, Cyprus) at 60 °C, and then HA powder (Medisell, Nicosia, Cyprus), 60 wt%, was dispersed in the mechanical stirring for 1 h and sonication for 20 min. Then the solution was poured into a mold and cured for 3 days at 40 °C.
Particle size distribution
The particle size distribution after a specific sonication protocol was measured using a laser diffraction particle size analyzer (SALD-2300, SHIMADZU CORPORATION, Kyoto, Japan). At the end of the experiment, the water of the container was left in the sun until the water evaporated. Thus the debris remained in the bottom of the container and was collected.
Temperature measurement
A data acquisition board (6251 DAQ, National Instruments, Texas, USA) was used to measure the temperature in the chalk. An analog input of the data acquisition board was used to capture the temperature. An Omega (M2813-1205, OMEGA Engineering, INC. Stamford, Connecticut, USA) voltage to temperature converter was used to measure temperature using a software written in MatLab (The Mathworks Inc., Natick, MA). A thermocouple (Omega engineering) was placed inside the chalk in order to measure temperature elevation. The size of the thermocouple was chosen to be 50 μm, so that the interaction with the ultrasound field is minimized.
Results
First, the effect of the intensity on the erosion of cylindrical chalk was investigated. Various chalks were exposed to different spatial average, temporal average (SATA) intensity varying from 5.5 to 20 W/cm2, with the DF and PRF kept constant to 10 % and 100 Hz, respectively. Figure 3a shows the picture of the cavities generated on the chalk (degassed) for two different SATA intensities (5.5 and 11 W/cm2) for 30 min. We believe that erosion appeared mostly in areas with trapped bubbles. In areas, with no bubbles, the effect of erosion was minimal. Figure 3b shows the weight loss of degassed chalk versus intensity (SATA) with 10 % DF and 100 Hz PRF. At SATA intensity above 5.5 W/cm2, the temperature increases beyond 1 °C. Figure 3c shows the spectral power measured by the cavitation detector during the exposure with SATA intensity of 5.5 W/cm2, 10 % DF and 100 Hz PRF. The sub-harmonic emission indicated the appearance of stable cavitation. These harmonic emissions appeared, during all the exposures evaluated in Fig. 3b. Although the chalks were degassed, the cavitation detector showed cavitation activity. It seems even if the chalks are degassed, still some bubbles remain in the chalks. The size of the removed calcified material by the ultrasonic transducer in a plane parallel to the transducer face is slightly less (8.5 mm × 3 mm) than the transducer area (10 mm × 3 mm).
Fig. 3.
a Picture of cavities generated on chalk at two different SATA intensities (5.5 and 11 W/cm2) for 30 min. DF and PRF were kept constant to 10 % and 100 Hz, respectively. b Weight loss as a function of SATA intensity (f = 5.3 MHz, DF = 10 %, PRF = 100 Hz), and c Spectral power measured by the cavitation detector during the exposure with SATA intensity of 5.5 W/cm2, 10 % DF and 100 Hz PRF
The next step was to investigate the effect of the PRF on the erosion rate of the cylindrical chalk (degassed). Four chalks were exposed to SATA intensity of 5.5 W/cm2 for 30 min, varying the PRF from 5 to 80 Hz with the DF kept constant to 10 %. Figure 4a shows the weight loss at different PRF for intensity (SATA) of 5.5 W/cm2 and DF of 10 %. During these experiments, the temperature did not exceed 1 °C.
Fig. 4.
a Weight loss vs. PRF. (f = 5.3 MHz, ISATA = 5.5 W/cm2, DF = 10 %, 30 min), b Variation of weight loss of each chalk versus DF (f = 5.3 MHz, ISATA = 5.5 W/cm2 PRF = 100 Hz)
The next step was to investigate the effect of the DF on erosion of the chalk. The chalk was exposed to a spatial average, pulse average (SAPA) intensity of 55 W/cm2 for 30 min, with varying DF from 5 to 40 % (while PRF was kept constant to 100 Hz (Fig. 4b). With DF above 20 % the temperature exceeded 1 °C indicating, that the exposure is no longer in mechanical mode. With a DF of 10 % (SATA intensity is 5.5 W/cm2) the temperature elevation is approximately 1 °C, and increases linearly to 4 °C with a DF of 40 %. This temperature increase does not result in drastic thermal effects, that justify the increased mass reduction. Therefore, the increased mass reduction is attributed to cavitation, since with increased DF, the probability of the occurrence of cavitation possibly increases.
The effect of the presence of bubbles or not on the erosion of the chalk is shown in Fig. 5 for DF of 20 %, PRF = 100 Hz and intensity (SATA) of 5.5 W/cm2 for 30 min. With the presence of bubbles, the cavitation activity was increased as recorded by the cavitation detector.
Fig. 5.
Variation of weight loss of chalk without bubbles and with bubbles. (f = 5.3 MHz, ISATA = 5.5 W/cm2, for 30 min, PRF = 100 Hz, DF = 20 %)
In another experiment, the goal was to visualize the chalk’s debris after pulsed ultrasound exposure. A chalk was exposed to intensity (SATA) of 10 W/cm2 for 30 min, with DF of 10 % and PRF of 100 Hz. Particles of mean diameter of 35 nm (standard deviation = 19 nm) were measured.
HA–PLA composite was exposed to the intensity (SATA) of 11 W/cm2 for 30 min, with the DF and PRF kept at 10 % and 100 Hz, respectively. Figure 6a shows the HA–PLA before the sonication and Fig. 6b shows the erosion of HA–PLA after this ultrasonic exposure. With this transducer and protocol we were able to remove HA–PLA material up to a depth of 3 mm (1664 mg) in just 30 min.
Fig. 6.

a HA–PLA composite before sonication, and b HA–PLA composite after ablation (f = 5.3 MHz, ISATA = 11 W/cm2, for 30 min, DF = 10 %, PRF = 100 Hz)
Conclusions
This paper includes a feasibility study that investigates the effectiveness of pulsed ultrasound in removing calcified material. A previous study [33] utilized thermal HIFU which might cause damage to the artery. Using mechanical mode ultrasound, avoids thermal damage to the artery, and yet it is possible to remove efficiently calcified plaque. This feasibility study includes a parametric study of the effect of intensity, PRF, DF, and the presence amount of bubbles. Various in vitro experiments were carried out, in chalk pieces and HA–PLA composite. The protocols used resulted into temperatures less than 1 °C (safe temperature). It was demonstrated in a set of experiments, that if the DF is increased, then the safe temperature is exceeded. Another parameter that increases the temperature is the intensity. In these experiments, the intensity used was carefully chosen, and, therefore did not cause any temperature elevation above 1 °C.
The removed calcified material as expected increases with the intensity, PRF, and DF. It was found that the removal of calcified material increases with the presence of bubbles. This may be attributed to the effect of cavitation. During the application of any exposure, sub-harmonic emissions were detected by the cavitation detector indicating the presence of cavitation. Even when the chalk material was degassed, stable cavitation activity was recorded due to the synergy with bubbles trapped inside the chalk. Perhaps, the non-degassed chalk possibly includes more bubbles and, therefore the mass removal increases drastically compared to the degassed chalk.
Very promising results were obtained in carrying out experiments on HA–PLA composites using this planar transducer. With this transducer and protocol we were able to remove HA–PLA material up to a depth of 3 mm in just 30 min. This type of technology looks promising for the removal of atherosclerotic plaque in humans, provided that the dissolved material from the plaque is collected so that it does not flow from the blood stream to other arteries, thus causing the blockage of arteries. Although we have shown that the particle size of the removed material is very small, care should be taken to collect this material. For example in studies involving atherectomy [34, 35] suction mechanism was used to collect the removed particles. Such suction technology can also be incorporated with this ultrasound technology to remove the residual particles. Another possible complication of this technology is possible endothelial damage due to the ultrasonic pressure applied under microbubble flow. The effect of pressure on the endothelium which is well described by Li et al. [36], needs to be controlled. Therefore, further studies are needed that determines the safe acoustic pressure that can be used.
It is possible that if the particle size is too small, then there is no need to collect the residual particles and, therefore, the deployment of ultrasound technology would be more feasible. Yet clinical studies need to be done in order to reveal how small this residual particle must to be. This specific device with this size can be used probably in peripheral arteries or in the carotid. Special design of the device needs to be done so as not to block the artery. For the application in the cardiac arteries, the device must be scaled down to possibly 1 mm.
The in vitro test was performed to derive an ultrasonic protocol that completely avoids thermal effects. It seems that the presence of bubbles in the chalk is responsible for the mass removal due to stable cavitation.
This technology can be used in the future for clinical trials primarily to treat calcified plaques in the carotid. Unstable plaques in the carotid are a major source of plaques that can reach the brain and cause stroke 0. The device can be attached to a catheter of appropriate size and the catheter can be guided intravenously to the carotid for ultrasonic treatment. Care will be taken to avoid the escape of debris reaching the brain, thus causing stroke. This technology can also be used for other stenotic conditions such as the femoral artery especially for superficial targets. Currently, this condition is under clinical evaluation using other technologies such directional atherectomy [37].
This technology potentially can be used for the treatment of intracranial atherosclerosis [38] diseases that may also cause stroke. The current technology has to be scaled down to smaller dimensions so that it can reach vessels in the brain using microcatheters. Another potential application is to treat plaques that rapture from the carotid and reach the brain [39]. This can be done intracranially using a small transducer attached to a microcatheter or by focused ultrasound transducer transcranially.
Acknowledgments
This work was supported by the Research Promotion Foundation (RPF) of Cyprus (program ΕΠΙΧΕΙΡΗΣΕΙΣ/ΠΡΟΪΟΝ/0311/01, and the European regional development structural funds).
Compliance with ethical standards
Conflict of interest
C. Damianou, A. Couppis declare that they have no conflict of interest.
Ethical standard
This article does not contain any studies with human participants or animals performed by any of the authors.
Informed consent
For this type of study formal consent is not required.
References
- 1.Wagenknecht L, Wasserman B, Chambless L, Coresh J, Folsom A, Mosley T, Ballantyne C, Sharrett R, Boerwinkle E. Correlates of carotid plaque presence and composition as measured by magnetic resonance imaging: the Atherosclerosis Risk in Communities (ARIC) Study Wagenknecht, Risk factors & MRI measured carotid plaque’. Circ Cardiovasc Imaging. 2009;2(4):314–322. doi: 10.1161/CIRCIMAGING.108.823922. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Romer TJ, Brennan JF, Fitzmaurice M, Feldstein ML, Deinum G, Myles JL, Myles JR, Kramer JR, Lees RS, Feld MS. A rotational ablation tool for calcified atherosclerotic plaque removal. Biomed Microdevices Circ. 1998;97:8. [Google Scholar]
- 3.Insull W., Jr The pathology of atherosclerosis: plaque development and plaque responses to medical treatment, section of atherosclerosis and vascular medicine. AJM. 2009;122(1, Supplement):S3–S14. doi: 10.1016/j.amjmed.2008.10.013. [DOI] [PubMed] [Google Scholar]
- 4.Mercando AD, Lai HM, Aronow WS, Kalen P, Desai HV, Gandhi K, Sharma M, Amin H, Lai TM. Reduction in atherosclerotic events: a retrospective study in an outpatient cardiology practice. Arch Med Sci. 2012;8(1):57–62. doi: 10.5114/aoms.2012.27282. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Nishida Y, Takahashi Y, Nakayama T, Asai S. Comparative effect of angiotensin II type I receptor blockers and calcium channel blockers on laboratory parameters in hypertensive patients with type 2 diabetes. Cardiovasc Diabetol. 2012;11(1):53. doi: 10.1186/1475-2840-11-53. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Superko HR, Momary KM, Li Y. Statins personalized. Med Clin North Am. 2012;96(1):123–139. doi: 10.1016/j.mcna.2011.11.004. [DOI] [PubMed] [Google Scholar]
- 7.Matsui Y, O’Rourke MF, Hoshide S, Ishikawa J, Shimada K, Kario K. Combined effect of angiotensin II receptor blocker and either a calcium channel blocker or diuretic on day-by-day variability of home blood pressure: the japan combined treatment with Olmesartan and a calcium-channel blocker versus Olmesartan and Diuretics randomized efficacy study. Hypertension. 2012;59(6):1132–1138. doi: 10.1161/HYPERTENSIONAHA.111.189217. [DOI] [PubMed] [Google Scholar]
- 8.Weintraub WS, Mauldin PD, Becker E, Kosinski AS, King SB., 3rd A comparison of the costs of and quality of life after coronary angioplasty or coronary surgery for multivessel coronary artery disease. Results from the Emory Angioplasty Versus Surgery Trial (EAST) Circulation. 1995;92(10):2831–2840. doi: 10.1161/01.CIR.92.10.2831. [DOI] [PubMed] [Google Scholar]
- 9.Lan C, Chen SY, Chiu SF, Hsu CJ, Lai JS, Kuan PL. Poor functional recovery may indicate restenosis in patients after coronary angioplasty. Arch Phys Med Rehabil. 2003;84(7):1023–1027. doi: 10.1016/S0003-9993(03)00031-5. [DOI] [PubMed] [Google Scholar]
- 10.Rafajlovski S, Orozović V, Ristić-Andelkov A, Raden G. Percutaneous transluminal coronary angioplasty in the right ventricle myocardial infarction treatment. Vojnosanit Pregl. 2005;62(10):731–738. doi: 10.2298/VSP0510731R. [DOI] [PubMed] [Google Scholar]
- 11.Dake MD, Ansel GM, Jaff MR, Ohki T, Saxon RR, Smouse HB, Zeller T, Roubin GS, Burket MW, Khatib Y, Snyder SA, Ragheb AO, White JK, Machan LS. Paclitaxel-Eluting stents show superiority to balloon angioplasty and bare metal stents in femoropopliteal disease: twelve-month zilver ptx randomized study results. Circ Cardiovasc Interv. 2011;4(5):495–504. doi: 10.1161/CIRCINTERVENTIONS.111.962324. [DOI] [PubMed] [Google Scholar]
- 12.Boyle CJ, Lennon AB, Prendergast PJ. In silico prediction of the mechanobiological response of arterial tissue: application to angioplasty and stenting. J Biomech Eng. 2011;133(8):081001. doi: 10.1115/1.4004492. [DOI] [PubMed] [Google Scholar]
- 13.Kossovsky N. Intravascular stents after transluminal angioplasty. N Engl J Med. 1987;317(13):836–837. doi: 10.1056/NEJM198709243171314. [DOI] [PubMed] [Google Scholar]
- 14.Takebayashi H, Haruta S, Kohno H, et al. Immediate and 3-month follow-up outcome after cutting balloon angioplasty for bifurcation lesions. J Interv Cardiol. 2004;17:1–7. doi: 10.1111/j.1540-8183.2004.00246.x. [DOI] [PubMed] [Google Scholar]
- 15.Bergersen LJ, Perry SB, Lock JE. Effect of cutting balloon angioplasty on resistant pulmonary artery stenosis. Am J Cardiol. 2003;15:185–189. doi: 10.1016/S0002-9149(02)03107-7. [DOI] [PubMed] [Google Scholar]
- 16.Unterberg C, Buchwald AB, Barath P, Schmidt T, Kreuzer H, Wiegand V. Cutting balloon coronary angioplasty–initial clinical experience. Clin Cardiol. 1993;16(9):660–664. doi: 10.1002/clc.4960160907. [DOI] [PubMed] [Google Scholar]
- 17.Ansel GM, Sample NS, Botti III CF, Jr, et al. Cutting balloon angioplasty of the popliteal and infrapopliteal vessels for symptomatic limb ischemia. Catheter Cardiovasc Interv. 2004;61:1–4. doi: 10.1002/ccd.10731. [DOI] [PubMed] [Google Scholar]
- 18.Engelke C, Sandhu C, Morgan RA, et al. Using 6-mm Cutting Balloon angioplasty in patients with resistant peripheral artery stenosis: preliminary results. AJR. 2002;1(79):619–623. doi: 10.2214/ajr.179.3.1790619. [DOI] [PubMed] [Google Scholar]
- 19.Kasirajan K, Schneider PA. Early outcome of “cutting” balloon angioplasty for infrainguinal vein graft stenosis. J Vasc Surg. 2004;39:702–708. doi: 10.1016/j.jvs.2003.10.046. [DOI] [PubMed] [Google Scholar]
- 20.Safian RD, Grines CL, May MA, Lichtenberg A, Juran N, Schreiber TL, Pavlides G, Meany TB, Savas V, O’Neill WW (2012) Clinical and angiographic results of transluminal extraction coronary atherectomy in saphenous vein bypass grafts. Circulation 89:302–312 [DOI] [PubMed]
- 21.Mangiacapra F, Heyndrickx GR, Puymirat E, Peace AJ, Wijns W, De Bruyne B, Barbato E. Comparison of drug-eluting versus bare-metal stents after rotational atherectomy for the treatment of calcified coronary lesions. Int J Cardiol. 2012;154(3):373–376. doi: 10.1016/j.ijcard.2011.11.048. [DOI] [PubMed] [Google Scholar]
- 22.Forouzannia SK, Abdollahi MH, Mirhosseini SJ, Hosseini H, Moshtaghion SH, Golzar A, Naserzadeh N, Ghoraishian SM, Emami Meybodi T. Clinical ‘Outcome and Cost in Patients with Off-pump vs. On-Pump Coronary Artery Bypass Surgery’. Acta Med Iran. 2011;49(7):414–419. [PubMed] [Google Scholar]
- 23.Lee EJ, Choi KH, Ryu JS, Jeon SB, Lee SW, Park SW, Park SJ, Lee JW, Choo SJ, Chung CH, Jung SH, Kang DW, Kim JS, Kwon SU. Stroke risk after coronary artery bypass graft surgery and extent of cerebral artery atherosclerosis. J Am Coll Cardiol. 2011;57(18):1811–1818. doi: 10.1016/j.jacc.2010.12.026. [DOI] [PubMed] [Google Scholar]
- 24.McGinn JT, Jr, Shariff MA, Bhat TM, Azab B, Molloy WJ, Quattrocchi E, Farid M, Eichorn AM, Dlugacz YD, Silverman RA. Prevalence of dysglycemia among coronary artery bypass surgery patients with no previous diabetic history. J Cardiothorac Surg. 2011;6:104. doi: 10.1186/1749-8090-6-104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Kulik A, Ruel M. Lipid-lowering therapy and coronary artery bypass graft surgery: what are the benefits? Curr Opin Cardiol. 2011;26(6):508–517. doi: 10.1097/HCO.0b013e32834b9fb1. [DOI] [PubMed] [Google Scholar]
- 26.Zhao X, Zhou Y, Song H, Guan L, Zheng G, Jin Z, Shi D, Li Y, Guo Y, Shi GP, Cheng XW. Comparison of bypass surgery with drug-eluting stents in diabetic patients with left main coronary stenosis. Nsei Med J. 2011;52(6):923–932. doi: 10.3349/ymj.2011.52.6.923. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Bogardus GM, Beretta FF, Huff R, Payne JT. Endarterectomy for peripheral arteriosclerosis. AMA Arch Surg. 1954;68(2):222–236. doi: 10.1001/archsurg.1954.01260050224010. [DOI] [PubMed] [Google Scholar]
- 28.Warren JA, Jordan WD, Jr, Heudebert GR, Whitley D, Wirthlin DJ. Determining patient preference for treatment of extracranial carotid artery stenosis: carotid angioplasty and stenting versus carotid endarterectomy. Ann Vasc Surg. 2003;17(1):15–21. doi: 10.1007/s10016-001-0332-0. [DOI] [PubMed] [Google Scholar]
- 29.Hollenbeak CS, Bowman AR, Harbaugh RE, Casale PN, Han D. The impact of surgical specialty on outcomes for carotid endarterectomy. J Surg Res. 2010;159(1):595–602. doi: 10.1016/j.jss.2008.03.049. [DOI] [PubMed] [Google Scholar]
- 30.McDonald RJ, Cloft HJ, Kallmes DF. Intracranial hemorrhage is much more common after carotid stenting than after endarterectomy: evidence from the national inpatient sample. Stroke. 2011;42(10):2782–2787. doi: 10.1161/STROKEAHA.111.618769. [DOI] [PubMed] [Google Scholar]
- 31.Shehata IA, Ballard JR, Casper AJ, Liu D, Mitchell T, Ebbini ES. Feasibility of targeting atherosclerotic plaques by high-intensity-focused ultrasound: an in vivo study. J Vasc Interv Radiol. 2013;24(12):1880–1887. doi: 10.1016/j.jvir.2013.08.013. [DOI] [PubMed] [Google Scholar]
- 32.Kim MH, Kim HJ, Kim NN, Yoon HS, Ahn SH. A rotational ablation tool for calcified atherosclerotic plaque removal’. Biomed Microdevices. 2011;13(6):963–971. doi: 10.1007/s10544-011-9566-y. [DOI] [PubMed] [Google Scholar]
- 33.Lafon C, Chapelon JY, Prat F, Gorry F, Margonari J, Theillère Y, Cathignol D. Design and preliminary results of an ultrasound applicator for interstitial thermal coagulation. Ultrasound Med Biol. 1998;24(1):113–122. doi: 10.1016/S0301-5629(97)00203-2. [DOI] [PubMed] [Google Scholar]
- 34.Davidson F. Ultrasonic power balances. In: Preston RC, editor. Output measurements for medical ultrasound. Berlin: Springer; 1991. pp. 75–90. [Google Scholar]
- 35.Ruiz-Ares G, Fuentes B, Martínez-Sánchez P, Díez-Tejedor E (2014) A prediction model for unstable carotid atheromatous plaque in acute ischemic stroke patients: proposal and internal validation. Ultrasound Med Biol. doi:10.1016 [DOI] [PubMed]
- 36.Li ZY, Taviani V, Tang T, Sadat U, Young V, Patterson A, Graves M, Gillard JH. The mechanical triggers of plaque rupture: shear stress vs pressure gradient. Br J Radiol. 2009;1:S39–S45. doi: 10.1259/bjr/15036781. [DOI] [PubMed] [Google Scholar]
- 37.Minko P, Buecker A, Jaeger S, Katoh M. Three-year results after directional atherectomy of calcified stenotic lesions of the superficial femoral artery. Cardiovasc Intervent Radiol. 2014;37(5):1165–1170. doi: 10.1007/s00270-014-0884-3. [DOI] [PubMed] [Google Scholar]
- 38.Gao T, Yu W, Liu C. Mechanisms of ischemic stroke in patients with intracranial atherosclerosis: a high-resolution magnetic resonance imaging study. Exp Ther Med. 2014;7(5):1415–1419. doi: 10.3892/etm.2014.1600. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Kiyofuji S, Inoue T, Hasegawa H, Tamura A, Saito I. Emergent surgical embolectomy for middle cerebral artery occlusion due to carotid plaque rupture followed by elective carotid endarterectomy. J Neurosurg. 2014;6:1–6. doi: 10.3171/2014.4.JNS132441. [DOI] [PubMed] [Google Scholar]





