Skip to main content
NIST Author Manuscripts logoLink to NIST Author Manuscripts
. Author manuscript; available in PMC: 2017 Mar 28.
Published in final edited form as: Philos Trans A Math Phys Eng Sci. 2016 Mar 28;374(2064):20150049. doi: 10.1098/rsta.2015.0049

Acoustic and microwave tests in a cylindrical cavity for acoustic gas thermometry at high temperature

K Zhang 1,2, XJ Feng 2, K Gillis 3, M Moldover 3, JT Zhang 2,*, H Lin 2, JF Qu 2, YN Duan 2
PMCID: PMC4884642  NIHMSID: NIHMS784138  PMID: 26903106

Abstract

Relative primary acoustic gas thermometry determines the ratios of thermodynamic temperatures from measured ratios of acoustic and microwave resonance frequencies in a gas-filled metal cavity on isotherms of interest. When measured in a cavity with known dimensions, the frequencies of acoustic resonances in a gas determine the speed of sound, which is a known function of the thermodynamic temperature T. Changes in the dimensions of the cavity are measured using the frequencies of the cavity's microwave resonances. We explored techniques and materials for acoustic gas thermometry at high temperatures using a cylindrical cavity with remote acoustic transducers. We used gas-filled ducts as acoustic waveguides to transmit sound between the cavity at high temperatures and the acoustic transducers at room temperature. We measured non-degenerate acoustic modes in a cylindrical cavity in the range 295 K < T < 797 K. The fractional uncertainty of the measured acoustic frequencies increased from 2×10−6 at 295 K to 5×10−6 at 797 K. In addition, we measured the frequencies of several transverse magnetic (TM) microwave resonances up to 1000 K in order to track changes in the cavity's length L and radius R. The fractional standard deviation of the values of L deduced from three TM modes increased from 3×10−6 for T < 600 K to 57×10−6 at 1000 K. We observed similar inconsistencies in a previous study.

Keywords: cylindrical acoustic gas thermometer, waveguide, acoustic resonator, microwave resonator, thermodynamic temperature

1. Introduction

Thermodynamic temperature T, with the SI unit of kelvin, is a measure of the average energy per degree of freedom in a system at equilibrium. Because accurate measurements of T require sophisticated techniques that are impractical in commercial applications, approximations to the kelvin scale have been developed, of which the latest are the International Temperature Scale of 1990 (ITS-90) and the Provisional Low Temperature Scale of 2000 (PLTS-2000) for T < 1 K. ITS-90 is an empirical scale based on a set of fixed points whose temperatures were assigned by an a priori determination using primary thermometry. Estimates of the difference TT90 have been evaluated periodically for temperatures from 0.65 K to 1358 K by the Consultative Committee for Temperature (CCT). The disagreement between different measurements is still significant in the range from 25 K to 273 K and from 600 K to 1358 K [1]. Near the copper point (1358 K), Ref. [1] estimated TT90 as (0.52 ± 0.02) K. (We use k = 1 standard uncertainties throughout this paper.) The CCT encourages additional measurements of TT90 in these temperature ranges to resolve these discrepancies.

The plan to redefine the kelvin in terms of the Boltzmann constant kB stimulated a worldwide effort to more accurately determine kB that has advanced primary thermometry. [2] To date, acoustic gas thermometry (AGT) is the most accurate means for determining kB. AGT determinations of kB require absolute measurements of the speed of sound in a well-characterized monatomic gas at the temperature of the triple point of water, defined as 273.16 K [3,4,5,6,7,8,9]. The speed of sound is deduced from highly precise measurements of the frequencies of acoustic resonances in a monatomic gas (with a known average molar mass) confined in a cavity with known dimensions. In contrast, the ratio of resonance frequencies measured on two isotherms is a function of the ratio of the thermodynamic temperatures, but the ratio does not depend on the molar mass of the gas as long as the molar mass does not change. Likewise, the ratio will not depend on the dimensions of the cavity as long as the dimensions do not change. In reality, the dimensions of the cavity do depend on the temperature due to thermal expansion, so the ratio of acoustic resonance frequencies must be corrected by the ratios of the cavity's dimensions. The ratios of microwave resonance frequencies measured on the two isotherms are used to correct the ratio of acoustic resonance frequencies for thermal expansion. [6,7,8] Finally, the thermodynamic temperature of one isotherm can be determined from the corrected ratio of measured acoustic frequencies, if the other isotherm is a known fixed-point temperature. This is the basis of relative primary AGT.

Rigorous theory for waves in a cavity shows that the measured acoustic and microwave resonance frequencies must be corrected for small known perturbations to the acoustic and electromagnetic fields that occur near the cavity's walls. If the cavity expands isotropically, then the working equation for relative AGT is

TTr=limp0u2(T,p)limp0u2(Tr,p)=limp0(fa+Δfa)2/(fm+Δfm)2limp0(fa+Δfa)r2/(fm+Δfm)r2, (1)

in which u(T,p) denotes the speed of sound in the thermometric gas at temperature T and pressure p; fa and fm are the measured frequencies of acoustic and microwave resonances, respectively; Δfa and Δfm are corrections to the frequencies due to known perturbations; the subscript r labels a quantity measured on the reference isotherm Tr which is often at the triple point of water. Equation (1) relies on the ideal-gas limit, i.e. p→0. In practice the measurements are made on isotherms as a function of p and u2(p,T) is fitted by a physically-motivated function which includes the parameter u02(T), the value of the speed of sound at zero pressure. Equation (1) illustrates the principle of relative primary acoustic gas thermometers: T is determined from the measured ratios of the acoustic resonance frequencies and the microwave resonance frequencies at the unknown temperature T and at the reference temperature at Tr.

Since 1999, several laboratories have reported AGT measurements of TT90 spanning the range 7 K to 600 K. [10,11,12,13,14,15,16] In the narrower range 90 K < T < 384 K, data from two or more laboratories overlap; with minor exceptions, the AGT values of (TT90)/T are mutually consistent within the fraction 3×10−6. Stimulated by these impressive results, we are dealing with the technological challenges of extending the measurements of TT90 above 600 K. [17,18,19]

In the moderate temperature range 90 K < T < 384 K our previous acoustic transducers were mounted directly in the walls of the cavity resonators. However, these transducers were constructed of materials that cannot withstand high temperatures. Instead, following Gillis et al. [20] and Ripple et al. [18], we are now using acoustic waveguides to transmit sound between the cavity at high temperature and the transducers at room temperature. Ripple et al. designed and explored the performance of acoustic waveguides operating from 293 K to 600 K. Their novel sound source contained two identical, bending-mode, piezoelectric lead-zirconium-titanate (PZT) ceramic transducers that were driven by equal voltages at identical phase. The PZTs were mounted face-to-face on the side of the waveguide. The symmetry ensures that the vibrations of the PZTs did not move their center of mass, thereby minimizing the coupling of the transverse motions of the PZTs to the waveguides and to the walls of the resonant cavity. The coupled PZTs also generate stronger sound pressures than a single PZT. [21]

Accurately measuring the thermal expansion of the cavity resonator operating above 600 K is another technological challenge. This can be done by measuring microwave resonance frequencies. [19,22,23,24]. Feng et al. [19] explored the performance of homemade antennas and microwave cables up to 1349 K. Underwood et al. [22,23] studied the perturbations arising from microwave probes and waveguides; they reported techniques for increasing signal-to-noise ratios and for avoiding currents at the joints between the cylindrical shell and its endplates. They used looped microwave antennas to measure both transverse electric (TE) modes and transverse magnetic (TM) modes. Their antennas were located at the midline of the test cavity in order to probe TE modes. They showed that the thermal expansion from 303 K to 373 K determined by the TE0pq modes agreed with the thermal expansion determined by the TM0pq modes within three parts per million. The uncertainties of their measurements were only weakly dependent on temperature.

Guided by these earlier studies, we fabricated a cylindrical resonator cavity with built-in acoustic waveguides and microwave antennas. Here, we report the acoustic performance of the acoustic waveguides from 295 K to 797 K and measurements of the thermal expansion of the cavity's length from 377 K to 1000 K. Our objective was to test the techniques and materials at high temperatures in preparation for future thermodynamic temperature measurements.

2. Experiment setup for high temperature

2.1. Resonant frequencies in a cylindrical cavity

The acoustic resonant frequencies for non-degenerate longitudinal modes in a cylindrical cavity with length of L and radius of a are:

fa,l0=ul2L, (2)

where u is the speed of sound and l=1,2,3… is the longitudinal eigenvalue of the Helmholtz equation.

The microwave resonant frequencies can be represented as:

TM modesfTM,lmm0=c2π(nπL)2+(χlma)2 (3)
TM modesfTM,lmm0=c2π(nπL)2+(χlma)2, (4)

where c is the speed of light in vacuum, n = 0,1,2,3… is the longitudinal eigenvalue. For TM modes, the transverse eigenvalue χlm is the mth root of the Bessel function Jl(x) = 0. For TE modes, the transverse eigenvalue χlm is the mth root of dJl (x)/dx = 0.

The superscript 0 in Eqs. (2)-(4) indicates that the predicted frequency is from a model that neglects perturbations, such as, boundary-layer or skin-depth effects at the wall of the cavity, openings in the wall of the cavity, and imperfect fabrication. The analyses of acoustic and microwave measurements for AGT include corrections to account for these perturbations, however the corrections will cancel to a large degree in the ratio in Eq. (1).

2.2. Experimental system

Our previous work [19] showed the successful use of the Ni-Fe-Cr alloy HR 120 (Haynes International, Kokomo, IN, USA)1 for the fabrication of the cylindrical cavity resonators with novel home-made microwave antennas and cables operating up to 1349 K. In this work, we fabricated a new cylindrical resonator that included acoustic waveguides for acoustic measurements while retaining the same, successful coaxial cables and microwave antennas that were used in Ref. [19] for thermal expansion measurements.

As illustrated in Figure 1, the cavity resonator consists of a cylindrical shell with two endplates. The cylindrical shell, 80 mm long with inner diameter (ID) of 63 mm and outer diameter (OD) of 76 mm, was fabricated from a seamless HR120 tube. The endplates were 6.4 mm thick, 76 mm diameter, and were fabricated from a HR120 plate. The endplates were welded to the ends of the cylindrical shell along the outside of the joint. The assembled resonator cavity was placed vertically in the middle of the furnace chamber. The resonator was not in a pressure vessel. The furnace chamber had an ID of 170 mm, length of 600 mm and had three independently controlled zones that provided a region 440 mm long with a temperature uniformity of ±5 K.

Figure 1. Schematic of the furnace, the resonator, and waveguides.

Figure 1

Acoustic waveguides leading from the resonator to the acoustic source or detector were welded into each endplate (Figure 1). Each waveguide had an ID of 4.57 mm and was made from a nickel-based high temperature alloy GH747 (Ni 45 %, Cr 16 %, Al 3%, Fe balance) (MTI Corporation, Richmond, CA USA). The lengths of the acoustic source and detector waveguides were 340 mm and 500 mm, respectively, measured from the outer surface of the endplates to the transducers. Each waveguide was connected to a tee fitting at room temperature. The tee provided a tap for the acoustic transducers and a side port for a longer duct providing a terminal impedance that reduced acoustic reflections. The side port of the upper tee was connected to the argon supply manifold through a duct with an ID of 4.57 mm. BIP Plus argon from Air Products [25], with a purity specification of 99.9999 % was used during the tests. As we have done before [26], a getter, SAES model GC50 (SAES Pure Gas, Inc, San Luis Obispo, CA USA), was placed downstream of the gas supply manifold for further purifying the argon before it flowed into the cylindrical cavity. The manufacturer specified that the getter reduced the mole fractions of reactive impurities to 5×10−8. The top port of the tee fitting connected to the source transducer housing. The waveguide served two functions: it coupled sound from the source to the cavity and served as a path for flowing purified argon from the gas manifold into the cavity. We used a flow controller to maintain a flow rate of about 40 mL/min.

Figure 2 is a schematic cross-section of the acoustic source assembly. The source was a single piezoelectric transducer (PZT) mounted to face the waveguide. The PZT disk had a diameter of 6 mm and a thickness of 0.5 mm and was bonded with epoxy to a 0.3 mm thick oxygen-free copper diaphragm. The electrical wires driving the source were attached directly to the surfaces of the PZT disk.

Figure 2. Source PZT transducer and housing.

Figure 2

The detector waveguide (Figure 1) was mounted symmetrically with respect to the source waveguide and it also was mounted on a tee fitting. Like its counterpart, the side port of the tee was connected to 4.57 mm ID tubing used to exhaust the gas from the resonator and to provide a terminal impedance to reduce reflections. The bottom port of the tee connected to the detector housing (Figure 3). The detector was a 1/4 inch (6.35 mm) microphone (G.R.A.S. Sound & Vibration A/S, Holte, Denmark). The housing assured that the pressures between the two sides of the microphone diaphragm were equal. An electrical hermetic feedthrough (LEMO) rated for operation from vacuum to 6 MPa connected the preamplifier and the power module G.R.A.S 12 AA.

Figure 3. Microphone and housing.

Figure 3

The thermal expansion of the cavity's length is a key parameter for a cylindrical acoustic gas thermometer (c-AGT) operating in relative mode. Underwood et al. [23] measured both TE modes and TM modes using looped antennas. TE modes have significantly higher signal-to-noise ratios than TM modes. Their source and detector antennas were mounted at the midline of the cylindrical shell, separated by an angle of 90° on the cross section perpendicular to the cylinder's axis. The TE modes generated no current at the joint between the endplate and the cylindrical cavity body. Thus, the measurements of the TE modes are anticipated to be more accurate than the TM modes. Their measurements showed that the absolute length and diameter of the cylindrical cavity determined from the TE modes differed from the length and diameter determined from the TM modes; however, the fractional changes in the length and diameter agreed within several parts per million as the temperature was changed from 313 K to 373 K.

Feng et al. [19] designed and tested homemade coaxial cables and antennas for measuring the TM modes from 377 K to 1349 K. Figure 4 shows their design. The outer conductor of each coaxial cable was a tube (3.2mm OD; 2.2mm ID) of Ni-Cr-Fe alloy, Inconel-625 (Special Metals Corp., Huntington, WV, USA). A line of alternating fused silica beads and fused silica tubes served as semi-flexible insulation between the outer conductor and the center conductor. The center conductor was a 0.5 mm diameter wire made of Ni-Cr. Each antenna was an extension of a center conductor that protruded 1 mm to 2 mm into the cavity and coupled to the non-degenerate microwave modes TMlmn with lmn = 010, 011, 012, 013 and 014. Both the source and the detector antennas were embedded in one endplate of the cylindrical resonator cavity. As shown in Figure 4, each antenna was insulated from the hole drilled through the endplate by a specially prepared fused silica tube. This tube had a ‘bump’ near one end that had been formed by a glass blower to act as a retention ring preventing the tube from sliding into the cavity. In contrast to the loop antenna reported by Underwood et al. [23], we chose to use the straight wire antenna by Feng et al. [19] because of its simplicity and ease of assembly. See Ref. [19] for more details.

Figure 4. Antenna and feedthrough in the resonator endplate. [19].

Figure 4

A standard S-type thermocouple (platinum/rhodium) was used to measure the temperature. This thermocouple was located approximately 3 mm from the top endplate of the cylindrical cavity. The pressure was measured with a digital pressure gauge (Druck: DPI 150).

3. Effect of waveguide at high temperature

AGTs rely on the prediction from kinetic theory that the speed of sound in a dilute gas in thermal equilibrium depends on its thermodynamic temperature. The pressure-dependence of the speed of sound is accounted for by the acoustic virial relation. AGTs accurately determine the thermodynamic temperature of a gas from measurements of the acoustic pressure response as a function of frequency near acoustic resonances in the cavity. Capacitive microphones are commonly used in AGTs because of their stability, sensitivity, and nearly flat frequency response over a wide frequency range and because they generate only small, calculable perturbations to the acoustic resonances in the cavities of typical AGTs. Additionally, PZTs have been used by the authors as both source and detector in c-AGTs. Both types of transducers have their useful temperature ranges. At temperatures above these ranges, acoustic waveguides must be used to bring sound from the source into the resonant cavity and to bring sound out of the cavity to the acoustic detector. Ripple et al. [18] investigated the performance of acoustic waveguides up to 600 K. Both the source and the detector were mounted in room temperature enclosures. The thermodynamic temperature measured by the AGT with the transducer-waveguide system [18] agreed well with the measurements by AGTs without such a system in the overlapping temperature range [12,15,17].

When sound propagates through a waveguide, it is attenuated by thermal and viscous dissipation in the gas and at the boundary between the gas and the walls of the waveguide. Gillis et al. [27] studied analytically and experimentally sound propagation inside circular, cylindrical waveguides of various lengths and diameters. Their model predicts that sound propagating inside a waveguide with radius ad is characterized by the attenuation length la,

la=1Re(Γ). (5)

Here, la is the distance over which the amplitude of sound pressure is attenuated to 1/e of its initial value and Re(Γ) stands for the real part of Γ. The propagation parameter Γ is a complex function of frequency, given by

Γ=iωu1+(γ1)Ft1Fv. (6)

Here, ω = 2πf; γ = Cp / Cv where Cp and Cv are the specific heat at constant pressure and volume, respectively. Ft and Fv are the thermal and viscous dissipation functions, respectively, which for a circular tube are given by

Ft2J1(ξt)ξtJ0(ξt),ξt=(1i)ad/δtandFv2J1(ξv)ξvJ0(ξv),ξv=(1i)ad/δv (7)

where δt=2λ/(ρCpω) is the thickness of the thermal boundary layer in the gas having thermal conductivity λ and mass density ρ; and, δv=2η/(ρω) is the thickness of the viscous boundary layer in the gas having shear viscosity η. The attenuation length increases as the duct diameter increases and decreases as the frequency increases.

When sound propagates a distance x through a duct, the acoustic pressure is attenuated by the factor ex/la. That is, when x = la the sound pressure will be attenuated to 1/e of its initial pressure (assuming no reflections). Therefore the length of the waveguide between the resonator and the transducer should be no longer than approximately la to avoid too much attenuation [18]. On the other hand, the waveguide used for the terminal impedance (connected to the side port of tee fitting sketched in Figure 1a) should be much longer than la to ensure that the sound in this waveguide is greatly attenuated to suppress resonances in the waveguide.

When resonances (standing waves) occur in the waveguide, they may couple to resonances in the cavity, thereby shifting the resonance frequencies of the cavity by an amount that is difficult to predict, and they may distort the measured frequency response profile. Standing waves occur at particular frequencies in sections of the waveguide that form quarter-wave or half-wave resonators, e.g., between the opening into the cavity and the transducer or between the transducer and the gas handling system. Standing waves in the latter section are suppressed by making the waveguide much longer than the attenuation length, as previously mentioned. Standing waves between the cavity and the transducer are suppressed by using tubes with the same ID for both sections of the waveguide and by designing the input impedance of the transducer + enclosure to be much higher than the characteristic impedance of the duct. The source and detector waveguides need not have the same ID.

The inner diameter of the waveguide is important, too. Increasing the inner diameter increases the attenuation length, which favors larger signals at the detector, but also increases the undesirable perturbations of the cavity modes. These competing effects may be optimized. We experimentally compared the uncertainty of several resonant frequency measurements at room temperature and ambient pressure in air with a cylindrical resonator with length of 80 mm and ID of 80 mm. Table 1 compares the results of waveguides with different inner diameters (all the waveguide were straight, 440 mm long).

Table 1.

Waveguide performance at room temperature and ambient pressure in air. The quality factor Q = fl/(2g), δfl/fl is the fractional uncertainty in the fitted acoustic resonance frequency fl, and g is the halfwidth from the fit.

ID of waveguide (mm) Longitudinal Mode (l00)

l = 2 3 4 5

Q 106 δfl/fl Q 106 δfl/fl Q 106 δfl/fl Q 106 δfl/fl
(s) 1.75 (d) 1.75 385 1.2 460 2.8 515 5.8 533 4.3
(s) 1.75 (d) 4.57 334 1.0 416 3.1 397 2.6 475 2.8
(s) 4.57 (d) 1.75 362 1.8 433 2.4 495 1.3 537 1.2
(s) 4.57 (d) 4.57 192 1.8 386 1.6 372 1.5 349 1.3

Note: (s) and (d) denote the source and detector waveguides, respectively.

The experimental result agrees with the prediction that a waveguide with a larger inner diameter decreases the attenuation and increases the signal-to-noise ratio, but lowers the quality factors. The waveguides with IDs of 1.75 mm and 4.57 mm represent two extreme situations. For the (200) and (500) modes, the quality factors Q = fr/(2g) of the cavity with the 4.57 mm ID were reduced to 50 % and 65 %, respectively, of the Q with the 1.75 mm ID waveguide, respectively. The perturbations caused by the waveguides of 4.57 mm ID were significantly larger.

At elevated temperatures, the pressure was chosen to keep the density of the gas in the resonator about the same as the density at the reference temperature (so that the perturbations would be the same), but there was an unavoidable temperature gradient (and a density gradient) in the waveguide to room temperature. The presence of the gradient did not adversely affect the perturbations due to the waveguides because the distance over which the gradient to room temperature occurred was greater than the wavelength of sound.

Gillis et al. [27] developed and tested a model that predicts the shifts of the resonance frequencies of a gas-filled cavity that results when a waveguide is attached into the wall of a cavity resonator. Table 2 lists the calculated perturbations to selected resonant frequencies at room temperature and ambient pressure in air caused by the 1.75 mm ID and 4.57 mm ID waveguides. For the 1.75 mm ID waveguide, the fractional perturbations are less than or equal to 3.5×10−6 for the modes (300), (400) and (500); the perturbation of the (200) mode is −12.0×10−6. The perturbations for the 4.57 mm ID waveguide are one order of magnitude larger.

Table 2. Calculation of the perturbation to the longitudinal modes of the cylindrical cavity due to a waveguide duct at room temperature in ambient pressure in air.

ID of waveguide (mm) Longitudinal Mode (l00)

l = 2 3 4 5

106Δf/f 106g/f 106Δf/f 106g/f 106Δf/f 106g/f 106Δf/f 106g/f
1.75 −12.0 94.9 −2.5 42.9 3.5 40.4 −2.7 30.0
4.57 −201.7 220.2 305.7 371.5 −130.3 173.9 121.0 183.7

The calculation of the waveguide-induced perturbations requires accurate values of the viscosity and thermal conductivity of the test gas. These properties are accurately known for argon and helium over a wide temperature range from recent advances in state-of-the-art measurements and from accurate, ab initio, quantum mechanical calculations. [28,29,30,31,32,33,34,35,36,37]. We estimate that the frequency shifts caused by 1.75 mm ID waveguides can be calculated with an uncertainty of 1×10−6 in the high temperature range. For the sake of larger signal-to-noise ratios, we used 4.57 mm ID waveguide ducts in this first exploration of c-AGT at high temperatures.

Figure 5 shows the detected in-phase (u) and quadrature (v) acoustic signals for the (300) mode at two state points: 295 K, 201 kPa (upper panel) and 797 K, 543 kPa (lower panel). These pressures were chosen so that the gas densities (and the Qs) in both state points were approximately equal. The results from the fits are given in Table 3. The signal-to-noise ratio at 797 K (approximately 370) is smaller than that at 295 K (approximately 870), and the deviations from fitting the resonance data at 797 K are approximately twofold larger than the deviations at 295 K. We now argue that temperature drift is responsible for some of the larger deviations from the fitted resonances at higher temperatures.

Figure 5.

Figure 5

Comparison of measurements of the (300) acoustic mode and deviations from fits as functions of the frequency at two state points. (a) 295 K, 202 kPa and (b) 797 K, 535 kPa. The data were acquired by stepping the frequency upward and then downward through the resonance. In both (a) and (b) the red circles are the in-phase signal, green triangles are the quadrature signal, and blue squares are the amplitude. The bottom panels are the deviations from fits summarized by lines 1 and 3 in Table 3.

Table 3.

Results from fits to measurements of mode (3,0,0) in argon. The 295 K and 797 K data in lines 1 and 3 are plotted in Figure 5. During the time that the measurements labeled (a) and (b) were acquired, the temperature drifted upward by 0.015 K. (a) Inconsistency between the up and down sweeps resulted in larger deviations and higher uncertainty than (b) for which the up and down sweeps were averaged before fitting. Figure 6 compares the deviations for the two fits.

T/K p/kPa Vpp/V fr/Hz g/Hz Q 106 δf/f 102 VppQ2 δf/f
295 202 0.3 5976.85 10.46 286 2.0 4.9
599 407 0.3 8491.61 14.52 292 2.9 7.4
797 535 0.2 9752.25 19.40 251 5.4 6.8
797 (a) 535 0.2 9753.23 19.54 250 13.0 16.3
797 (b) 535 0.2 9753.23 19.54 250 4.2 5.3

The frequency response for every resonance is measured by stepping the frequency both upward and downward through the resonance. This measurement method removes the effect of a temperature drift on the fitted resonance frequency fr and halfwidth g to first order. During one measurement of the (300) acoustic mode at 797 K and 535 kPa, the temperature drifted by 0.015 K. The deviations of the in-phase and quadrature signals from a fit, plotted in the top panel of Figure 6, show inconsistencies between the up and down data that are not random and result in a large uncertainty for fr. The fit results are labeled as (a) in Table 3. When the up and down data at each frequency are averaged before fitting, the systematic effects are removed, and the deviations are much smaller and more random (bottom panel of Figure 6). Although the two fits give the same value for fr and g, the estimated uncertainty is much smaller when the data are averaged. The right-most column in Table 3 shows that the values of δf/f at 295 K and 797 K are consistent when the changes in the drive voltage and Qs of the resonances are taken into account. No unexpected problem appeared when the waveguides were subjected to the temperature gradient from 295 K to 797 K.

Figure 6.

Figure 6

Comparison of fits to resonance data with a temperature drift. Deviations of the measured in-phase (red circles) and quadrature (green triangle) signals from two fits to data for the (300) mode at 797 K and 535 kPa while the temperature drifted by 0.015 K. Top: deviations from a fit when the up and down sweeps were not averaged first. The large systematic deviations give a poor estimate of the uncertainty in the fitted resonance frequency. Bottom: deviations from a fit to the same data after first averaging the up and down sweeps. The fit results are listed as (a) and (b), respectively, in Table 3.

4. Thermal expansion measurement by the microwave resonant method

As sketched in Figure 1, the resonator cavity has two built-in antennas used for the source and the detector of microwaves inside the cavity. Similar to Feng et al. [19], we used the TM 010, 011, 012 and 013 modes for the measurements of the thermal expansion of the cavity length from 377 K to 1000 K with flowing argon at the pressure of 200 kPa. To measure each resonance profile, we swept the frequency upward and downward in 5 MHz steps for a total of 201 points. The quality factors of the cavity modes ranged from 1600 to 2200. The signal-to-noise ratios varied from 6500 to 33500. We calculated the radius of the cavity at each temperature from the TM 010 mode, which does not depend on the cavity length. Using this radius, we obtained three measurements of the length of the cavity from the TM011, 012 and 013 modes at each temperature. Because AGT generally operates in the relative mode for measuring thermodynamic temperature, knowing the thermal expansion is necessary. The consistency among the thermal expansions extracted from the measurements of the modes is key for evaluating the microwave method. Table 4 lists the relative thermal expansion from 377 K at six temperatures and the fractional standard deviation of the thermal expansions among the modes. The measurements of the microwave frequencies were corrected from the perturbations using formulas from the literature [22,23].

Table 4. The relative thermal expansion LT/L0 from 377 K and the inconsistency indicated by the standard deviation of the thermal expansions obtained from three microwave modes.

T (K) LT/L0 106 std dev
478 1.001524 3.6
569 1.002949 2.9
673 1.004651 46.6
772 1.006368 11.8
871 1.008225 13.3
1000 1.010788 57.2

The extracted thermal expansions show only small inconsistencies at 478 K and 569 K, but the inconsistency increases to 57×10−6 at 1000 K. Except for the anomaly at 673 K, the increasing inconsistency with temperature is similar to that reported by Feng et al. [19], whose cavity and microwave feedthroughs were similar to the present design. The primary difference is that the new cavity is equipped with acoustic waveguides. We speculate that the anomalously large inconsistency at 673 K resulted from an undetected temperature change. Below, we will discuss the cause for the increasing standard deviation with increasing temperature.

We also measured the temperature dependence of the ratio of the cavity 's length to its radius (L/a). (See Figure 7.) Below900 K, L/a depends linearly on temperature with a negative slope that is consistent with the linear slope observed by Feng et al. [19]. However, when we heated the present cavity to 1000 K, we observed an increase in the slope in Figure 7 that was not reported in [19]. One difference between the present cavity and the previous cavity is the thermal history. The previous cavity had been thermally cycled between room temperature and 1350 K; the present cavity was never heated above 1000 K, where it appeared to “age” rapidly. A certain amount of annealing may take place at elevated temperature. We will test to see if the cavity stabilizes after thermal cycling multiple times to a temperature higher than the intended range for AGT.

Figure 7.

Figure 7

Temperature dependence of the ratio of the cylindrical cavity dimensions (length)/(radius) obtained from microwave measurements. Note that the vertical scale is parts per million.

5. Discussion and summary

We measured the performance of the acoustic waveguides at 295 K and 797 K near the argon density 82 mol/m3. The deviations from the fitted resonance function at 797 K were approximately five times larger than the deviations at 295 K. By sweeping the frequency both upward and downward to measure the resonance, we observed good reproducibility at 295 K but not at 797 K because the temperature drifted. Averaging the sweeps improves the fits, reduces the systematic deviations, and lowers the estimated uncertainty of the resonance frequency. Shorter data acquisition time also reduces the effects of temperature drifts at the expense of less averaging. We plan three improvements to accurately measure TT90 at high temperatures. First, we will use a heat pipe inside the furnace to reduce the temperature instabilities from ±5 K to approximately ±(0.1 to 0.5) K. Second, we will place the resonator inside a pressure vessel whose temperature will be controlled on the order of a few tens of millikelvin. Third, a heater on the surface of the resonator will control the temperature of the resonator to within a few millikelvin. We also will replace the 4.57 mm ID waveguides with 1.75 mm ID waveguides.

For argon-based AGT, Moldover et al. [24] recommend using argon densities in the range 40 mol m−3 < ρ/M < 200 mol m−3, where ρ is the mass density and M is the average molar mass of the gas. In order to maintain constant ρ/M, the gas pressure must increase with increasing temperature. However, increasing the gas pressures has the undesirable effect of increasing the perturbations caused by the recoil of the solid resonator shell in response to the oscillating gas within the resonator. References [25] and [26] discuss perturbations from four families of normal modes of a cylindrical shell with cylindrical endplates. These include longitudinal and radial oscillations of the cylindrical body of the shell, the bending of the endplates, and the recoil of the entire shell, considered as a rigid body. The frequency perturbations from the first three families of modes are predicted to have the functional form

(Δfl)shell,ifl(ρu2)gasGi,l1(fl/fshell,i)2, (8)

where u is the speed of sound of gas; fl is the gas resonant frequency at the eigenvalue l; fshell,i is a resonance frequency of the shell's motion at mode index i; Δfl is the shift of the resonant frequency fl; Gi,l is a compliance of the shell that depends upon the geometry and elastic properties of the shell and on the geometry of the gas mode l. The perturbations predicted by Eq. (8) increase with gas pressure because at constant density ρu2≈(5/3)p, where the factor (5/3) applies to monatomic gases. To counteract the undesired increase of these perturbations with temperature, we considered increasing the ratio (thickness of the shell)/(radius of cavity). However, increasing the outer dimensions of the shell or decreasing the radius of the cavity will degrade the performance of a c-AGT. Thus, a high-temperature c-AGT will rely on the general principle that perturbations from shell motions vanish linearly with the pressure at low pressures.

The acoustic resonance frequencies of the gas-filled cavity increase as T1/2. For each gas mode, the frequency at 1300 K is twice the frequency at room temperature. The elastic constants of metallic cavity walls are only weakly dependent on temperature, and the frequencies fshell,i in Eq. (8) decrease only slightly with increasing temperature. Therefore, the acoustic resonances of the gas may cross some of the frequencies fshell,i as the temperature increases. Near such crossings, Eq. (8) is a poor approximation and the crossing gas mode cannot be used to determine an accurate value of T. This is one of the challenges that acoustic thermometry at temperatures up to 1350 K will encounter.

We recall that the microwave frequency ratios (fmfm)p,T/(fmfm)p,Tref are not exactly equal to length ratios such as L(p,T)/L(p,Tref). When gas is admitted into a cavity at the temperature T, the microwave frequencies will decrease by the factor 1/n where n(p,T) is the refractive index of the gas at the temperature and pressure of interest. Furthermore, the microwave frequencies will increase because the cavity shrinks under the applied hydrostatic pressure. If the walls of the cavity are made of an isotropic material with a temperature-dependent isothermal bulk modulus BT(T) = −V/(∂p/∂V)T, each linear dimension shrinks by the factor [1+p/(3BT)]. If AGT is conducted at the copper point (1358 K) using argon at pressures up to 1.5 MPa, the product ℘ = n(p,T)[1+p/(3BT)] reaches a maximum value of approximately 1.0016. However, ℘ is very nearly a linear function of pressure. Therefore, when the pressure-dependent ratio data in Eq. (1) are fitted by a physically-motivated function of the pressure that includes a term proportional to p, the temperature and pressure dependences of ℘ will be accounted for and the correct value of u02(T)/u02(Tref) will be determined. Thus, accurate AGT does not require values of the difficult-to-measure quantity BT(T).

We used the TM modes to measure the thermal expansion of the cavity because they coupled to the cavity with the simplest, most-easily-modelled antennas: straight wires. Our coaxial cables had a small diameter; therefore, they made only very small perturbations to the acoustic modes of the cavity. Our thermal expansion measurements, except for the anomaly at 673 K, agree with the results reported by Feng et al. [19]. Both studies indicate that the inconsistency among the measured thermal expansions by different TM modes are at the level of 3×10−6 for temperatures below 600 K; 10×10−6 from 600 K to 900 K; and 50×10−6 at 1000 K and above. The data suggest three possible causes responsible for the phenomenon. First, the thermal expansion of the cavity may be anisotropic due to stresses in the metal from machining. If this is the case, we will use acoustic modes and TE modes with identical wavenumbers, so that the anisotropy of the thermal expansion exactly cancels out of the frequency ratios in Eq. (1). This choice completely circumvents the need to characterize the anisotropy of the thermal expansion. Second, there may be errors in the calculated corrections to the measured frequencies due to unknown current paths through the joints at the endplates. For example, a slit at this joint may open or close as the temperature changes. The sensitivity of the microwave frequency perturbations to such a slit vary from TM mode to TM mode. The effective resistance of the slit is sensitive to the slit's geometry; therefore it might explain the increasing dispersion of the TM modes with increasing temperature. We plan to eliminate this possible problem by using a new cylindrical cavity that has a single joint in the symmetry plane perpendicular to the cylinder's axis. (This joint is analogous to the joint at the “equator” of a quasi-sphere formed by joining two quasi-hemispheres.) Third, there may be a structural change in the metal above 1000 K. The possibility can be tested by measuring the thermal expansion of the alloy using a dilatometer.

Finally, we note that thermal cycling might reduce the stress in the alloy and slow the time-dependence of the cavity's dimensions at high temperatures.

Acknowledgments

We acknowledge our great appreciations for Robin Underwood from NPL and Laurent Pitre from LNE-CNAM.

Funding statement: This work was supported by the National Natural Science Foundation of China (Nos. 51476153, 51276175 and 61372041) and the Fundamental Research Program of NIM (No. AKY1401).

Footnotes

1

In order to describe materials and procedures adequately, it is occasionally necessary to identify commercial products by manufacturers' name or label. In no instance does such identification imply endorsement by the National Institute of Standards and Technology, nor does it imply that the particular product or equipment is necessarily the best available for the purpose.

Present address: Division of Thermophysics and Process Measurements, National Institute of Metrology, No.18, Bei San Huan Dong Lu, Beijing, 100029, China

Competing interests: We have no competing interests.

Authors' contributions: Zhang K. participated in the design and the assembling of the experimental facility, carried out the major part of the experiments. Feng X. J. designed the acoustic and microwave cavity for the experiments at high temperatures, directed the assembling of the facility and analyzed the experimental data. K. Gillis originated the design of the microwave antenna and acoustic waveguide operating at high temperatures, analyzed the experimental data and revised the peer-reviewed manuscript. M. Moldover originated the thought and the procedure for construction of acoustic and microwave resonant cavities, analyzed the experimental data and edited the revision of the peer-reviewed manuscript. Zhang J. T. coordinated the reported study, participated in the analysis of the experimental data and drafted the manuscript. Lin H., Qu J. F. and Duan Y. N. participated in the design of the experiment and the analysis of the experimental data. All authors gave final approval for publication.

References

  • 1.Fischer J, de Podesta M, Hill KD, Moldover M, Pitre L, Rusby R, Steur P, Tamura O, White R, Wolber L. Present Estimates of the Differences Between Thermodynamic Temperatures and the ITS-90. International Journal of Thermophysics. 2011;32:12–25. doi: 10.1007/s10765-011-0922-1. [DOI] [Google Scholar]
  • 2.White DR, Fischer J. The Boltzmann constant and the new kelvin. Metrologia. 2015;52:S213–216. doi: 10.1088/0026-1394/52/5/S213. [DOI] [Google Scholar]
  • 3.Colclough AR, Quinn TJ, Chandler TRD. An acoustic redetermination of the gas constant. Proceedings of the Royal Society A: Mathematical, Physical and Engineering Sciences. 1979;368:125–139. doi: 10.1098/rspa.1979.0119. [DOI] [Google Scholar]
  • 4.Moldover MR, Trusler JPM, Edwards TJ, Mehl JB, Davis RS. Measurement of R using a spherical acoustic resonator. Journal of Research of the National Institute of Standards and Technology. 1988;93:85–144. [Google Scholar]
  • 5.Fischer J. Fundamental Constants Meeting. Eltville, Germany: Feb, 2015. The International Temperature Scale and the new definition of the kelvin. [Google Scholar]
  • 6.de Podesta M, Yang I, Mark DF, Underwood R, Sutton G, Machin G. Correction of NPL-2013 estimate of the Boltzmann constant for argon isotopic composition and thermal conductivity. Metrologia. 2015;52:S353–S363. doi: 10.1088/0026-1394/52/5/s353. [DOI] [Google Scholar]
  • 7.Pitre L, Risegari L, Sparasci F, Plimmer MD, Himbert ME, Giuliano Albo PA. Determination of the Boltzmann constant from the speed of sound in helium gas at the triple point of water. Metrologia. 2015;52:S263–S273. doi: 10.1088/0026-1394/52/5/s263. [DOI] [Google Scholar]
  • 8.Gavioso RM, Madonna Ripa D, Steur PPM, Gaiser C, Truong D, Guianvarc'h C, Tarizzo P, Stuart FM, Dematteis R. A determination of the molar gas constant R by acoustic thermometry in helium. Metrologia. 2015;52:S274–S304. doi: 10.1088/0026-1394/52/5/s274. [DOI] [Google Scholar]
  • 9.Feng XJ, Lin H, Gillis KA, Moldover MR, Zhang JT. Test of a virtual cylindrical acoustic resonator for determining the Boltzmann constant. Metrologia. 2015;52:S343–S352. doi: 10.1088/0026-1394/52/5/s343. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Moldover MR, Boyes SJ, Meyer CW, Goodwin ARH. Thermodynamic temperatures of the triple points of mercury and gallium and in the interval 217 K to 303 K. Journal of Research of the National Institute of Standards and Technology. 1999;104:11–46. [Google Scholar]
  • 11.Ewing MB, Trusler JPM. Primary acoustic thermometry between T = 90 K and T = 300 K. The Journal of Chemical Thermodynamics. 2000;32:1229–1255. doi: 10.1006/jcht.1999.0606. [DOI] [Google Scholar]
  • 12.Strouse GF, Defibaugh DR, Moldover MR, Ripple DC. Progress in Primary Acoustic Thermometry at NIST: 273 K to 505 K. In: Ripple DC, editor. TEMPERATURE: Its Measurement and Control in Science and Industry vol VII; 8th Int Temperature Symp (Chicago IL, 21-24 October 2002. New York: AIP; 2003. pp. 31–6. [Google Scholar]
  • 13.Benedetto G, Gavioso RM, Spagnolo R, Marcarino P, Merlone A. Acoustic measurements of the thermodynamic temperature between the triple point of mercury and 380 K. Metrologia. 2004;41:74–98. doi: 10.1088/0026-1394/41/1/011. [DOI] [Google Scholar]
  • 14.Pitre L, Moldover MR, Tew WL. Acoustic thermometry: new results from 273 K to 77 K and progress towards 4 K. Metrologia. 2006;43:142–162. doi: 10.1088/0026-1394/43/1/020. [DOI] [Google Scholar]
  • 15.Ripple DC, Strouse GF, Moldover MR. Acoustic Thermometry Results from 271 to 552 K. International Journal of Thermophysics. 2007;28:1789–1799. doi: 10.1007/s10765-007-0255-2. [DOI] [Google Scholar]
  • 16.Underwood R, de Podesta M, Sutton G, Stanger L, Rusby R, Harris P, Morantz P, Machin G. Estimates of the difference between thermodynamic temperature and the ITS-90 in the range 118 K to 303 K. Phil Trans R Soc A. doi: 10.1098/rsta.2015.0048. submitted. [DOI] [PubMed] [Google Scholar]
  • 17.Ripple DC, Defibaugh DR, Moldover MR, Strouse GF. Techniques for primary acoustic thermometry to 800K. In: Ripple DC, editor. TEMPERATURE: Its Measurement and Control in Science and Industry vol VII; 8th Int Temperature Symp (Chicago IL, 21-24 October 2002) New York: AIP; 2003. pp. 25–30. [Google Scholar]
  • 18.Ripple DC, Murdock WE, Strouse GF, Gillis KA, Moldover MR. Room temperature acoustic transducers for high-temperature thermometry. In: Meyer CW, editor. TEMPERATURE: Its Measurement and Control in Science and Industry vol IX; 9th Int Temperature Symp (Anaheim, CA, 19-23 March 2012) Vol. 1552. New York: AIP; 2013. pp. 44–9. AIP Conf. Proc. [DOI] [Google Scholar]
  • 19.Feng X, Gillis KA, Moldover MR, Mehl JB. Microwave-cavity measurements for gas thermometry up to the copper point. Metrologia. 2013;50:219–226. doi: 10.1088/0026-1394/50/3/219. [DOI] [Google Scholar]
  • 20.Gillis KA, Moldover MR, Goodwin ARH. Accurate acoustic measurements in gases under difficult conditions. Rev Sci Instrum. 1991;62:2213–2217. doi: 10.1063/1.1142339. [DOI] [Google Scholar]
  • 21.Lin H, Gillis KA, Zhang JT. Characterization of piezoelectric ceramic transducer for accurate speed-of-sound measurement. International Journal of Thermophysics. 2010;31:1234–1247. doi: 10.1007/s10765-010-0795-8. [DOI] [Google Scholar]
  • 22.Underwood RJ, Mehl JB, Pitre L, Edwards G, Sutton G, de Podesta M. Waveguide effects on quasispherical microwave cavity resonators. Measurement Science and Technology. 2010;21:075103. doi: 10.1088/0957-0233/21/7/075103. [DOI] [Google Scholar]
  • 23.Underwood RJ, Edwards GJ. Microwave-Dimensional Measurements of Cylindrical Resonators for Primary Acoustic Thermometry. International Journal of Thermophysics. 2014;35:971–984. doi: 10.1007/s10765-014-1726-x. [DOI] [Google Scholar]
  • 24.Moldover MR, Gavioso RM, Mehl JB, Pitre L, de Podesta M, Zhang JT. Acoustic gas thermometry. Metrologia. 2014;51:R1–R19. doi: 10.1088/0026-1394/51/1/r1. [DOI] [Google Scholar]
  • 25.Zhang JT, Lin H, Feng XJ, Sun JP, Gillis KA, Moldover MR, Duan YY. Progress Toward Redetermining the Boltzmann Constant with a Fixed-Path-Length Cylindrical Resonator. International Journal of Thermophysics. 2011;32:1297–1329. doi: 10.1007/s10765-011-1001-3. [DOI] [Google Scholar]
  • 26.Lin H, Feng X, Gillis K, Moldover M, Zhang J, Sun J, Duan Y. Improved determination of the Boltzmann constant using a single, fixed-length cylindrical cavity. Metrologia. 2013;50:417–432. doi: 10.1088/0026-1394/50/5/417. [DOI] [Google Scholar]
  • 27.Gillis K, Lin H, Moldover M. Perturbations from ducts on the modes of acoustic thermometers. Journal of Research of the National Institute of Standards and Technology. 2009;114:263–285. doi: 10.6028/jres.114.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Berg RF. Simple flow meter and viscometer of high accuracy for gases. Metrologia. 2005;42:11–23. doi: 10.1088/0026-1394/42/1/002. [DOI] [Google Scholar]
  • 29.May EF, Moldover MR, Berg RF, Hurly JJ. Transport properties of argon at zero density from viscosity-ratio measurements. Metrologia. 2006;43:247–258. doi: 10.1088/0026-1394/43/3/007. [DOI] [Google Scholar]
  • 30.May EF, Berg RF, Moldover MR. Reference Viscosities of H2, CH4, Ar, and Xe at Low Densities. International Journal of Thermophysics. 2007;28:1085–1110. doi: 10.1007/s10765-007-0198-7. [DOI] [Google Scholar]
  • 31.Vogel E, Jäger B, Hellmann R, Bich E. Ab initio pair potential energy curve for the argon atom pair and thermophysical properties for the dilute argon gas. II. Thermophysical properties for low-density argon. Molecular Physics. 2010;108:3335–3352. doi: 10.1080/00268976.2010.507557. [DOI] [Google Scholar]
  • 32.Vogel E. Reference Viscosity of Argon at Low Density in the Temperature Range from 290 K to 680 K. International Journal of Thermophysics. 2010;31:447–461. doi: 10.1007/s10765-010-0760-6. [DOI] [Google Scholar]
  • 33.Cencek W, Przybytek M, Komasa J, Mehl JB, Jeziorski B, Szalewicz K. Effects of adiabatic, relativistic, and quantum electrodynamics interactions on the pair potential and thermophysical properties of helium. J Chem Phys. 2012;136:224303. doi: 10.1063/1.4712218. [DOI] [PubMed] [Google Scholar]
  • 34.Mehl JB. Ab initio properties of gaseous helium. Comptes Rendus Physique. 2009;10:859–865. doi: 10.1016/j.crhy.2009.10.009. [DOI] [Google Scholar]
  • 35.Berg RF, Moldover MR. Recommended Viscosities of 11 Dilute Gases at 25 °C. Journal of Physical and Chemical Reference Data. 2012;41:043104. doi: 10.1063/1.4765368. [DOI] [Google Scholar]
  • 36.Zhang JT, Lin H, Che J. Effects of connecting tubing on a two-capillary viscometer. Metrologia. 2013;50:377–384. doi: 10.1088/0026-1394/50/4/377. [DOI] [Google Scholar]
  • 37.Lin H, Feng XJ, Zhang JT, Liu C. Using a two-capillary viscometer with preheating to measure the viscosity of dilute argon from 298.15 K to 653.15 K. J Chem Phys. 2014;141:234311. doi: 10.1063/1.4903960. [DOI] [PubMed] [Google Scholar]

RESOURCES