A micro-patterned sample holder of single-crystalline silicon, loaded with multiple protein crystals which are exposed to a humidified gas stream, allows high-quality room-temperature data collection.
Keywords: crystallography on a chip, synchrotron serial crystallography, room-temperature crystallography, X-ray radiation damage
Abstract
Recent success at X-ray free-electron lasers has led to serial crystallography experiments staging a comeback at synchrotron sources as well. With crystal lifetimes typically in the millisecond range and the latest-generation detector technologies with high framing rates up to 1 kHz, fast sample exchange has become the bottleneck for such experiments. A micro-patterned chip has been developed from single-crystalline silicon, which acts as a sample holder for up to several thousand microcrystals at a very low background level. The crystals can be easily loaded onto the chip and excess mother liquor can be efficiently removed. Dehydration of the crystals is prevented by keeping them in a stream of humidified air during data collection. Further sealing of the sample holder, for example with Kapton, is not required. Room-temperature data collection from insulin crystals loaded onto the chip proves the applicability of the chip for macromolecular crystallography. Subsequent structure refinements reveal no radiation-damage-induced structural changes for insulin crystals up to a dose of 565.6 kGy, even though the total diffraction power of the crystals has on average decreased to 19.1% of its initial value for the same dose. A decay of the diffracting power by half is observed for a dose of D 1/2 = 147.5 ± 19.1 kGy, which is about 1/300 of the dose before crystals show a similar decay at cryogenic temperatures.
1. Introduction
Proteins are a fundamental building block of biological cells and perform a vast array of functions within living organisms. In the course of conducting their biological function, proteins in many cases undergo reversible structural changes. In order to understand these dynamics on the molecular level and their relation to molecular recognition, catalytic function and allostery, detailed descriptions of atomic coordinates and motions are desirable (Fraser et al., 2011 ▸; Fenwick et al., 2014 ▸). In structural biology, techniques such as X-ray crystallography, NMR spectroscopy and dual polarization interferometry are employed to investigate the atomic and molecular arrangement of proteins.
In X-ray crystallography, radiation damage limits the amount of diffraction data which can be obtained from one crystal. Radiation damage can generally be classified as either specific or global. Specific radiation damage arises from direct inelastic interaction of the X-ray photons with the sample via photoelectric absorption or from Compton scattering, often denoted primary damage. The subsequent radiolytic reactions are caused by the generated free electrons with energies between a few and several tens of electron volts, also referred to as secondary damage (Mozumder & Magee, 1966 ▸; Teng & Moffat, 2000 ▸; Ziaja et al., 2005 ▸). Possible repercussions of specific damage are redox processes, generation of free radicals and breakage of chemical bonds, which become visible in irreversible changes of the electron-density map (Helliwell, 1988 ▸; Weik et al., 2000 ▸; Meents et al., 2009 ▸, 2010 ▸). Global damage, also called tertiary damage, is not specific to particular atoms and becomes apparent because of the destabilization of the crystal lattice (Henderson, 1990 ▸). Possible indications of global damage have been reported as an overall decrease in the integrated Bragg reflection intensities (especially pronounced for the high-resolution reflections), an increase in the relative B factors and an expansion of the unit cell (Burmeister, 2000 ▸; Ravelli & McSweeney, 2000 ▸; Kmetko et al., 2006 ▸; Meents et al., 2009 ▸).
At the latest low-emittance synchrotron beamlines, which can provide a photon flux of up to 1013 photons s−1 on a small spot of a few square micrometres, the crystal lifetime in the beam at room temperature is typically limited to just a few milliseconds. In comparison, for data collection at cryogenic temperatures the crystal lifetime can be extended by about two orders of magnitude to about one second (Southworth-Davies et al., 2007 ▸; Burkhardt et al., 2013 ▸; Garman & Owen, 2006 ▸). For this reason X-ray crystallography is typically performed at cryogenic temperatures (Haas & Rossmann, 1970 ▸; Hope, 1988 ▸; Southworth-Davies et al., 2007 ▸).
However, the above-mentioned motion and dynamics in proteins can be quenched at cryogenic temperatures, meaning that the structural information obtained may not fully represent the structure of the protein in its natural environment (Rasmussen et al., 1992 ▸; Tilton et al., 1992 ▸). As stated by Halle (2004 ▸) in a theoretical analysis, the cryocooling process is typically too slow to fully trap the equilibrium distribution of the protein and the solvent configuration which exists at room temperature. In conformity with this perspective, Fraser et al. (2009 ▸) showed that catalytically essential conformations of the human proline isomerase, cyclophilin A, are populated at room temperature but not at cryogenic temperatures. Similar findings for different conformational distributions as a function of temperature were obtained for crystals of myoglobin (Frauenfelder et al., 1979 ▸, 1987 ▸) and RNase (Rasmussen et al., 1992 ▸; Tilton et al., 1992 ▸).
Besides these structural considerations, X-ray diffraction measurements at cryogenic temperatures are also in many cases associated with practical problems concerning crystal handling and cryo-protection. Cryo-protectants such as ethylene glycol and glycerol are normally added to protect the sample crystals from freezing damage, e.g. due to the formation of hexagonal ice. Upon cryo-protection, the native environment of the crystals is changed. In particular, membrane proteins, viruses and other large-unit-cell systems tend to yield small and fragile crystals which are very sensitive to temperature variations. Here, cryo-protection procedures can be tedious and often have a detrimental effect on crystal quality and their diffracting properties (Axford et al., 2012 ▸). One method which allows sample vitrification without cryo-protectants is high-pressure freezing, where samples are cryo-cooled while being subjected to high pressures of several hundreds of megapascals (Thomanek et al., 1973 ▸; Kim et al., 2005 ▸; Burkhardt et al., 2012 ▸, 2013 ▸).
With the development of X-ray free-electron lasers (XFELs) in recent years, crystallography experiments at room temperature are staging a comeback. At XFELs, a single diffraction pattern of a crystal is recorded before the crystal is vaporized by the intense X-ray pulse, an approach often denoted ‘diffraction before destruction’ (Chapman et al., 2011 ▸). With the availability of new high-frame-rate integrating detectors (Philipp et al., 2011 ▸; Herrmann et al., 2013 ▸; Jungmann-Smith et al., 2014 ▸), this has led to the emerging field of ‘serial crystallography’, where diffraction patterns of hundreds to hundreds of thousands of crystals are measured in a reasonable time and the few recorded reflections from each individual crystal are subsequently merged to give a complete data set (Boutet et al., 2012 ▸; Neutze & Moffat, 2012 ▸; Redecke et al., 2013 ▸; Hunter et al., 2014 ▸). Performing these experiments at room temperature allows, in addition to static structure determination, the investigation of protein kinetics, e.g. enzyme reactions (Kupitz et al., 2014 ▸; Tenboer et al., 2014 ▸).
Originally devised for crystallography experiments at XFELs, this serial approach has recently also been successfully employed at synchrotron sources. In particular, the development of high-frame-rate single-photon counting detectors has allowed the measurement of tens of thousands of diffraction patterns in a reasonable time at synchrotrons (Broennimann et al., 2006 ▸; Pennicard et al., 2012 ▸).
Different sample-delivery techniques such as liquid jets (Stellato et al., 2014 ▸; Botha et al., 2015 ▸), fixed targets (Zarrine-Afsar et al., 2012 ▸; Gati et al., 2014 ▸; Mueller et al., 2015 ▸; Murray et al., 2015 ▸; Roedig et al., 2015 ▸; Baxter et al., 2016 ▸), and array-like plates for in situ crystal growth and diffraction (McPherson, 2000 ▸; Kisselman et al., 2011 ▸; Axford et al., 2012 ▸; Pinker et al., 2013 ▸; Heymann et al., 2014 ▸; Huang et al., 2015 ▸) have been successfully used at synchrotrons. The challenges of the currently applied sample-delivery techniques are the high crystal consumption and the high background scattering levels caused by the surrounding mother liquor in the case of liquid jets. The use of ‘toothpaste’-like lipidic cubic phase jets can significantly reduce the crystal consumption (Weierstall et al., 2014 ▸) but is accompanied by a high background scattering level caused by the lipidic crystal matrix of the jet. Particularly in the case of smaller crystals with sizes of only a few micrometres, the weak diffraction signals are often buried in the background.
A promising approach is the use of fixed substrates which allow for efficient removal of the surrounding mother liquor and therefore generate almost no background (Roedig et al., 2015 ▸). The chips developed by our group are made of single-crystalline silicon so that, in a diffraction experiment, the substrate itself only contributes with distinct Bragg reflections from the silicon, which are typically at higher resolution shells than the protein diffraction and can be avoided by exact knowledge of the orientation of the substrate.
A significant challenge when performing macromolecular crystallography experiments with fixed substrates at room temperature is to prevent dehydration of the crystals after removal of the mother liquor. In order to ensure a constantly humid environment during data collection, a possible solution is to place the chip in a sealed container, using for example Mylar foil (Mueller et al., 2015 ▸; Sherrell et al., 2015 ▸). However, these approaches may be unfavourable with respect to the scattering background contribution of the Mylar foil and the additional tedious preparation step which becomes necessary.
In the following, we present a novel approach for room-temperature data collection from multiple macromolecular crystals mounted on a chip bathed in an open stream of humidified air. The humidified air stream, with adjustable relative humidity, is provided by a humidity control device (Sanchez-Weatherby et al., 2009 ▸). By this means the crystals can be kept stable for several hours without losing their diffracting properties owing to dehydration. Additionally, this technique enables experiments including controlled dehydration of protein crystals, which has been shown to improve the diffraction qualities of the crystals in many cases (Kiefersauer et al., 2000 ▸; Umena et al., 2011 ▸). Furthermore, reactions inside the protein crystals can be initiated by changing the chemical composition of the vapour stream (Sjögren et al., 2002 ▸).
In order to prove this concept and the applicability of our chip to room-temperature serial crystallography, we have performed room-temperature data collection from multiple cubic insulin crystals loaded onto our chip. It was a further goal of the present work to investigate specific and global radiation damage effects at room temperature as a function of dose and to draw conclusions from this for future room-temperature serial crystallography experiments at synchrotron sources.
2. Materials and methods
2.1. Sample preparation and chip handling
Porcine insulin was purchased from Sigma–Aldrich (catalogue No. I-5523). Cubic Zn-free crystals of porcine insulin were grown from alkaline sodium phosphate buffer by hanging-drop vapour diffusion (Jackson, 1974 ▸). A solution of 15 or 30 mg ml−1 insulin in Na2HPO4 buffer (c Phos = 0.05 M, 0.01 M EDTA, pH 9.8–10.4) was mixed with Na2HPO4 buffer (c Phos = 0.4–0.6 M, 0.01 M EDTA, pH 9.8–10.4) at different ratios (1:3, 1:1, 2:1, 3:1). The overall drop volume was adjusted to 6 µl. The mixture was equilibrated against 750 µl of Na2HPO4 buffer (c Phos = 0.4–0.6 M, 0.01 M EDTA, pH 9.8–10.4) at 298 K. Insulin microcrystals between 20 and 50 µm in size were obtained after 3–5 h.
A micro-patterned silicon chip was used as a sample holder, similar to the device recently described by our group for data collection at cryogenic temperatures (Roedig et al., 2015 ▸). The chip itself was attached to a standard magnetic base, which was directly mounted on the beamline goniometer as shown in Fig. 1 ▸. To prevent dehydration of the crystals during the experiment, the chip was continuously exposed to a stream of humidified air. For this, a humidity control device, HC1 (Arinax, France), was installed on the beamline goniometer, and it provided a continuous flow of 4 l min−1 humidified air with adjustable relative humidity (Sanchez-Weatherby et al., 2009 ▸). During data collection, the relative humidity was set to the highest possible value (∼100%) and was probably between 98 and 99% at the sample position. For crystal loading, a drop of 2–3 µl of the crystal suspension was pipetted onto the chip (Fig. 1 ▸ a). Subsequently, the mother liquor was blotted off the back of the chip through the micropores of the silicon membrane using a wedge of blotting paper (Fig. 1 ▸ b). The diameter of the pores was chosen to be ∼10 µm so that insulin crystals larger than the pore diameter were retained, while the surrounding mother liquor and other solid material smaller than the pore diameter were efficiently removed. A microscope image of insulin crystals loaded onto the chip is shown in Fig. 2 ▸. It can be clearly seen that the crystals are surrounded by only very small amounts of mother liquor.
2.2. Data collection
X-ray data collection was performed on the macromolecular beamline I03 at Diamond Light Source. An X-ray beam with a size of 50 × 50 µm (FWHM) and a photon flux of 5.7 × 1010 photons s−1 at an X-ray energy of 12.36 keV was used for the experiment. Individual crystals were selected and centred in the X-ray beam using the on-axis microscope installed on the beamline. From every crystal, 700 diffraction images, with a rotation increment of 0.1° and an exposure time of 0.1 s per image, were collected using a Pilatus 6M detector, resulting in a total rotation range of 70° and a total exposure time of 70 s per data set. Owing to the small thickness of the silicon membrane, typically less than 10 µm, the X-ray absorption by the chip at the given energy is less than 4% for normal incidence. With crystal dimensions of about 30 × 30 × 30 µm and a solvent content of 60%, the average absorbed X-ray dose for each recorded diffraction image was calculated as 0.808 kGy, using RADDOSE-3D (Zeldin et al., 2013 ▸).
2.3. Data processing and phasing
The recorded data sets were indexed and integrated using XDS (Kabsch, 2010b ▸). In order to study radiation damage effects, each data set from a crystal was divided into 14 subsets of 50 frames for data quality analysis and into seven subsets of 100 frames for structure refinement. Partial data sets from different crystals were merged using XSCALE (Kabsch, 2010b ▸) to obtain a complete data set for each subset. Subsets from five different insulin crystals with 100 frames each were sufficient for a complete data set, owing to the high symmetry of cubic insulin (space group I213). Since this space group is one of the 27 space groups for protein crystals in which the symmetry of the Bravais lattice is higher than the symmetry of the space group itself, an indexing ambiguity with two possible indexing modes exists (Brehm & Diederichs, 2014 ▸). As a consequence, data sets from about half of the crystals had to be reindexed before merging. By maximizing the Pearson correlation coefficient
between the individual partial data sets i and j with intensities Ii and Ij of common reflections h = (h,k,l), complete and consistent data sets with correlation coefficients rij ≃ 1 were obtained.
Structure solution was performed by molecular replacement with Phaser (McCoy et al., 2007 ▸), using PDB model 9ins as a template (Gursky et al., 1992 ▸), and the structure was refined isotropically with phenix.refine (Adams et al., 2010 ▸). This procedure was performed for each subset. Refinement statistics for the first subset (absorbed dose 80.8 kGy) and the last subset (absorbed dose 565.6 kGy) are summarized in Table 1 ▸. The refined structure for the first subset (absorbed dose 80.8 kGy) is available under PDB code 5fb6. In addition, raw images and processing results for all subsets can be downloaded from http://zenodo.org using the doi 10.1107/S1600576716006348.
Table 1. Diffraction data and refinement statistics for the first and last merged subsets, corresponding to absorbed doses of 80.8 and 565.6 kGy, respectively.
80.8 kGy | 565.6 kGy | |
---|---|---|
X-ray wavelength (Å) | 1.003 | 1.003 |
Resolution range (Å) | 30–1.9 (2.0–1.9) | 30–1.9 (2.0–1.9) |
Unit-cell lattice constant | 78.8 Å (cubic) | 78.8 Å (cubic) |
Space group | I213 | I213 |
Total reflections | 33764 (3372) | 34346 (3349) |
Unique reflections | 6542 (652) | 6480 (640) |
Multiplicity | 5.2 (5.2) | 5.3 (5.2) |
Completeness (%) | 99.21 (99.69) | 98.89 (98.77) |
Mean I/σ(I) | 18.04 (5.16) | 10.34 (1.25) |
CC1/2 | 0.999 (0.918) | 0.996 (0.375) |
CC* | 1.0 (0.978) | 0.999 (0.738) |
Wilson B factor (Å2) | 24.58 | 24.58 |
R merge | 0.05801 (0.372) | 0.1474 (1.327) |
R meas | 0.06457 | 0.1632 |
R work/R free | 15.71/17.35 | 17.71/20.01 |
No. of non-hydrogen atoms | 474 | 479 |
Macromolecules | 451 | 451 |
Water molecules | 23 | 28 |
Protein residues | 51 | 51 |
Average B factor (Å2) | 29.30 | 28.90 |
Protein | 28.70 | 28.20 |
Solvent | 41.80 | 40.30 |
R.m.s. deviation, bonds (Å) | 0.009 | 0.009 |
R.m.s. deviation, angles (°) | 0.93 | 0.94 |
Ramachandran plot (%) | ||
Favoured | 98 | 98 |
Allowed | 2 | 2 |
Outliers | 0 | 0 |
Clashscore | 4.51 | 6.76 |
3. Results and discussion
3.1. Integrated data statistics
The decay of different diffraction data quality indicators as a function of dose is shown in Fig. 3 ▸. The mean intensity over the whole resolution range (Fig. 3 ▸ a) and the I/σ(I) ratio of the high-resolution reflections (Fig. 3 ▸ b) show a distinct loss of signal as determined by XDS processing. While for data collection at cryo-temperatures a linear correlation model between damage rate and absorbed dose has been found to be appropriate in many cases (Teng & Moffat, 2000 ▸; Diederichs et al., 2003 ▸), both quadratic and exponential approaches have also been proposed (Burmeister, 2000 ▸; Diederichs, 2006 ▸). As shown in Fig. 3 ▸(a), the mean intensity for each crystal is well described by an exponential decay in the form
where D is the absorbed dose, βtot the damage rate constant, I 0 the fraction of the overall intensity which is directly affected by radiation damage effects and I 1 a constant term which accounts for an undamaged fraction in the crystal. For ease of comparability, the values of I mean were normalized to the values of the first subset, corresponding to an absorbed dose of D = 40.4 kGy. We define the specific dose as D* = 1/βtot, which is a measure of the amount of dose a crystal can absorb until the first term in equation (2) is reduced to 1/e times its initial value I 0. Fitted parameters for five chosen insulin crystals, obtained using the nonlinear least-squares method, are given in Table 2 ▸. By this means an average specific dose of D* = 153.1 ± 22.8 kGy is obtained for the insulin crystals in our experiment. A similar decay is observed for the intensities I/σ(I) of the high-resolution reflections [Fig. 3 ▸(b), also normalized to the value of the first subset]. The values for the mean intensity do not vanish completely, but converge to a lower boundary represented by the fitting parameter I 1 in the dose range under consideration. The values of this constant term I 1 are on average 19.1% of the theoretical initial mean intensity I 0 + I 1 for the undamaged crystal, which corresponds to D = 0 in equation (2). This behaviour is seen for all considered crystals. Since the beam was typically larger than the crystals and the crystals were therefore completely immersed in the X-ray beam, this observation cannot be attributed to new unexposed crystalline material which could have been rotated into the beam during oscillation. As suggested by Blake & Phillips, the observations indicate a fraction of crystalline structure which seemed to be rather unaffected by radiation damage processes, i.e with a damage rate constant much larger than the values obtained for βtot (Blake & Phillips, 1962 ▸; Hendrickson, 1976 ▸).
Table 2. Fit parameters corresponding to equation (2) for the five insulin crystals shown in Fig. 3 ▸ .
Crystal | I 0 | I 1 | βtot (kGy−1) | D* = 1/βtot (kGy) | I 1/(I 0 + I 1) | D 1/2 = D*ln[2I 0/(I 0 − I 1)] |
---|---|---|---|---|---|---|
Crystal 1 | 1.04 ± 0.04 | 0.236 ± 0.019 | 0.0067 ± 0.0006 | 149.1 ± 12.7 | 0.185 ± 0.013 | 141.7 ± 12.7 |
Crystal 2 | 1.06 ± 0.03 | 0.188 ± 0.018 | 0.0061 ± 0.0004 | 165.0 ± 11.4 | 0.150 ± 0.012 | 146.5 ± 10.7 |
Crystal 3 | 1.02 ± 0.05 | 0.249 ± 0.042 | 0.0055 ± 0.0008 | 182.0 ± 28.0 | 0.196 ± 0.028 | 177.0 ± 29.2 |
Crystal 4 | 1.03 ± 0.02 | 0.257 ± 0.006 | 0.0083 ± 0.0003 | 120.5 ± 4.2 | 0.199 ± 0.005 | 118.0 ± 4.3 |
Crystal 5 | 0.97 ± 0.03 | 0.280 ± 0.013 | 0.0067 ± 0.0004 | 149.0 ± 9.3 | 0.225 ± 0.009 | 154.3 ± 10.2 |
Besides the above-mentioned parameters, an interesting quantity is the dose required to reduce the mean intensity from its initial value I 0 + I 1 by half to (I 0 + I 1)/2, denoted D 1/2. Its relation to the above-mentioned fit parameters is given by D 1/2 = D*ln[2I 0/(I 0 − I 1)]. As shown in Table 2 ▸ for the insulin crystals under study, this value amounts to D 1/2 = 147.5 ± 19.1 kGy on average. This can be compared with the value of D 1/2 = 43 MGy for cryogenic temperatures, which was observed by Owen et al. (2006 ▸) from diffraction experiments with holoferritin and apoferritin crystals. From this, we conclude that the crystals, when measured at room temperature, can tolerate only 1/292 of the applied dose before showing a similar decay in their diffracting power in comparison to cryogenic temperatures.
It is noteworthy that the onset of the decay in the diffracting power occurs right from the beginning as the crystal is exposed to the applied X-ray radiation, which can be attributed to the relatively low dose rate of 16.2 kGy s−1. Our observation is consistent with the results shown by Owen et al. (2014 ▸), who described the loss in diffracting power as a function of dose with an initial slow decay, a so-called lag phase, followed by a faster exponential decay. However, those authors showed that the lag phase, lasting up to 500 kGy, is only observed for very high dose rates (>1 MGy s−1) and becomes absent for smaller dose rates. It turns out that, at room temperature and higher dose rates, dynamic effects become apparent which are governed by the diffusion of free radicals through the crystal lattice, particularly hydroxyl (OH−), the quenching of radicals by solvent molecules and free-radical recombination.
The dose dependence of the relative isotropic B factor can be taken as another indicator of global damage (Kmetko et al., 2006 ▸; Meents et al., 2009 ▸). Relative B factors as given by XSCALE [see Kabsch (2010a ▸) for details] are plotted in Fig. 3 ▸(c) for each crystal. The values were obtained by scaling all subsets from an individual crystal to the first subset. By this means, all subsets from the same crystal are placed on the same scale and the scaled intensities I scaled(h) are given by
where I meas(h) is the measured intensity, B the relative B factor and K an additional absolute scale factor. Since all subsets were scaled to the corresponding first subset of each crystal, the values are K = 1 and B = 0 for the first subset. As shown in Fig. 3 ▸(c), a linear increase in the relative B factor is observed in the dose range 40–200 kGy. At higher doses, the relative B factor reaches a saturation level before it decreases back to smaller values. As shown in Fig. 3 ▸(d), for the scale factor K a monotonic increase is observed which is consistent with the overall decrease in the diffraction intensity.
3.2. Structure refinement statistics
The refinement R values of the merged data sets are shown in Fig. 4 ▸. In general, an increase in the refinement R values is observed with respect to the X-ray dose absorbed by the crystal. For the first subset, corresponding to an absorbed dose of 80.8 kGy, high-quality electron-density maps with model R values of R work/R free = 15.71/17.35 could be obtained from the merged data sets. With increasing dose these values increase to R work/R free = 17.71/20.01, which were obtained from refinement of the last subset. Complete refinement statistics of the merged data from the first and last subsets are provided in Table 1 ▸.
3.3. Site-specific damage
The disulfide bridges in insulin crystals are known to be sensitive to radiation damage. In particular, the disulfide bond between residues CysA7 and CysB7 has been reported to be highly affected by specific radiation damage in the range 30–60 MGy at cryogenic temperatures between 5 and 100 K (Meents et al., 2010 ▸). However, such changes in the electron-density map could not be observed in the present experiments at room temperature. As shown in Fig. 5 ▸, no indications of specific radiation damage for this disulfide bond could be observed up to an absorbed dose of 565.6 kGy. Careful inspection of the electron-density maps and atomic B factors did not reveal any other sites of specific radiation damage for the overall dose applied.
4. Conclusions
Our chip is very well suited to room-temperature data collection from multiple crystals from a few micrometres up to several tens of micrometres in size. Loading of the crystals onto the chip is straightforward and takes less than a minute. The crystal suspension is pipetted onto the chip and the mother liquor is blotted from the reverse through the pores using a wedge of filter paper. Owing to the sieve-like structure of the chip, all crystals larger than the pore size are retained after blotting and can be utilized for data collection. Since single-crystalline silicon is used as the chip material, the background contribution of the chip itself is negligible. Dehydration of the crystals is prevented by exposing them to a stream of humidified air during sample loading and data collection. This approach requires no further sealing with additional material such as Kapton or Mylar foil which would lead to an increased background level. Using humidified helium gas instead of air would further decrease the path the scattered X-rays need to pass in air, thus further reducing the background level.
High-quality X-ray diffraction data sets merged from the measurement of several crystals could be obtained and were analysed with respect to potential radiation damage effects. The diffracting power of the insulin crystals decreases exponentially as a function of the applied X-ray dose and reaches 1/e of its initial value already after a dose of 153.1 ± 22.8 kGy. However, in contrast with the overall reduction in diffraction power, no signs of specific radiation damage could be observed from the electron-density maps for doses up to 565.6 kGy. An explanation for this different behaviour between global and specific radiation damage could be that specific radiation damage – and here in particular cleavage of disulfide bridges – is less temperature dependent than global radiation damage and generally occurs only at higher doses. This means that disulfide bond breakage is not the preferred damage pathway at room temperature, where global radiation damage to the lattice is clearly the dominating effect.
All five insulin crystals show a similar decay of the diffraction quality parameters versus dose, with small variations between the different crystals. This further highlights the applicability of our chip in combination with a humidified gas stream for a room-temperature crystallography setup which provides identical measurement conditions for all crystals. Previous room-temperature crystallography experiments were mostly carried out with crystals mounted in glass capillaries. For this, all crystals had to be individually soaked into the capillaries, which subsequently had to be sealed with hot wax to prevent solvent evaporation. Using this approach, crystals often suffered from dehydration and heating effects, which led to a large spread of the diffraction quality parameters between individual crystals.
In summary, our chip has been proven very useful for room-temperature data collection from multiple crystals. Sample preparation is straightforward, since the loading of thousands of crystals onto the chip takes less than a minute. With respect to macromolecular X-ray crystallography experiments at cryogenic temperatures, we observed that crystals can tolerate 1/292 of the dose before showing a similar decay in the diffracting power. This means that, at room temperature, about 300 times the number of crystals have to be measured in order to obtain a diffraction data set of similar quality to that at cryogenic temperature.
In the present experiment the crystals were relatively large, which allowed us to collect data in large rotation increments from every crystal. In the case of smaller crystals and aiming for the same resolution, because of radiation damage effects only data from smaller rotation increments can be collected. Hence, for structure determinations, data collection from a larger number of crystals will be required. For very small crystals in the range of a few micrometres, fast raster-scanning of the chip without any rotation might become the method of choice here.
We expect that such room-temperature measurements with microcrystals loaded onto our chip will also be well suited for measurements at XFELs and for future kinetic investigations of irreversible enzyme reactions.
Supplementary Material
Acknowledgments
The authors thank the staff at Diamond Light Source beamline I03 for their support. The research leading to these results has received funding from the European Community’s Seventh Framework Programme (grant No. FP7/2007–2013) under grant agreement No. 290605 (PSI-Fellow/Co-fund). This work was supported by the Virtual Institute VH-VI-403 of the Helmholtz Association.
References
- Adams, P. D. et al. (2010). Acta Cryst. D66, 213–221.
- Axford, D. et al. (2012). Acta Cryst. D68, 592–600. [DOI] [PMC free article] [PubMed]
- Baxter, E. L. et al. (2016). Acta Cryst. D72, 2–11.
- Blake, C. C. F. & Phillips, D. C. (1962). Biological Effects of Ionising Radiation at the Molecular Level, pp. 183–191. IAEA Symposium, Vienna.
- Botha, S., Nass, K., Barends, T. R. M., Kabsch, W., Latz, B., Dworkowski, F., Foucar, L., Panepucci, E., Wang, M., Shoeman, R. L., Schlichting, I. & Doak, R. B. (2015). Acta Cryst. D71, 387–397. [DOI] [PubMed]
- Boutet, S. et al. (2012). Science, 337, 362–364.
- Brehm, W. & Diederichs, K. (2014). Acta Cryst. D70, 101–109. [DOI] [PubMed]
- Broennimann, Ch., Eikenberry, E. F., Henrich, B., Horisberger, R., Huelsen, G., Pohl, E., Schmitt, B., Schulze-Briese, C., Suzuki, M., Tomizaki, T., Toyokawa, H. & Wagner, A. (2006). J. Synchrotron Rad. 13, 120–130. [DOI] [PubMed]
- Burkhardt, A., Wagner, A., Warmer, M., Reimer, R., Hohenberg, H., Ren, J., Fry, E. E., Stuart, D. I. & Meents, A. (2013). Acta Cryst. D69, 308–312. [DOI] [PubMed]
- Burkhardt, A., Warmer, M., Panneerselvam, S., Wagner, A., Zouni, A., Glöckner, C., Reimer, R., Hohenberg, H. & Meents, A. (2012). Acta Cryst. F68, 495–500. [DOI] [PMC free article] [PubMed]
- Burmeister, W. P. (2000). Acta Cryst. D56, 328–341. [DOI] [PubMed]
- Chapman, H. N. et al. (2011). Nature, 470, 73–77. [DOI] [PMC free article] [PubMed]
- Diederichs, K. (2006). Acta Cryst. D62, 96–101. [DOI] [PubMed]
- Diederichs, K., McSweeney, S. & Ravelli, R. B. G. (2003). Acta Cryst. D59, 903–909. [DOI] [PubMed]
- Fenwick, R. B., van den Bedem, H., Fraser, J. S. & Wright, P. E. (2014). Proc. Natl Acad. Sci. USA, 111, E445–E454. [DOI] [PMC free article] [PubMed]
- Fraser, J. S., van den Bedem, H., Samelson, A. J., Lang, P. T., Holton, J. M., Echols, N. & Alber, T. (2011). Proc. Natl Acad. Sci. USA, 108, 16247–16252. [DOI] [PMC free article] [PubMed]
- Fraser, J. S., Clarkson, M. W., Degnan, S. C., Erion, R., Kern, D. & Alber, T. (2009). Nature, 462, 669–673. [DOI] [PMC free article] [PubMed]
- Frauenfelder, H., Hartmann, H., Karplus, M., Kuntz, I. D., Kuriyan, J., Parak, F., Petsko, G. A., Ringe, D., Tilton, R. F. & Connolly, M. L. (1987). Biochemistry, 26, 254–261. [DOI] [PubMed]
- Frauenfelder, H., Petsko, G. A. & Tsernoglou, D. (1979). Nature, 280, 558–563. [DOI] [PubMed]
- Garman, E. F. & Owen, R. L. (2006). Acta Cryst. D62, 32–47. [DOI] [PubMed]
- Gati, C., Bourenkov, G., Klinge, M., Rehders, D., Stellato, F., Oberthür, D., Yefanov, O., Sommer, B. P., Mogk, S., Duszenko, M., Betzel, C., Schneider, T. R., Chapman, H. N. & Redecke, L. (2014). IUCrJ, 1, 87–94. [DOI] [PMC free article] [PubMed]
- Gursky, O., Li, Y., Badger, J. & Caspar, D. L. (1992). Biophys. J. 61, 604–611. [DOI] [PMC free article] [PubMed]
- Haas, D. J. & Rossmann, M. G. (1970). Acta Cryst. B26, 998–1004. [DOI] [PubMed]
- Halle, B. (2004). Proc. Natl Acad. Sci. USA, 101, 4793–4798. [DOI] [PMC free article] [PubMed]
- Helliwell, J. R. (1988). J. Cryst. Growth, 90, 259–272.
- Henderson, R. (1990). Proc. R. Soc. B Biol. Sci. 241, 6–8.
- Hendrickson, W. A. (1976). J. Mol. Biol. 106, 889–893. [DOI] [PubMed]
- Herrmann, S., Boutet, S., Duda, B., Fritz, D., Haller, G., Hart, P., Herbst, R., Kenney, C., Lemke, H., Messerschmidt, M., Pines, J., Robert, A., Sikorski, M. & Williams, G. (2013). Nucl. Instrum. Methods Phys. Res. A, 718, 550–553.
- Heymann, M., Opthalage, A., Wierman, J. L., Akella, S., Szebenyi, D. M. E., Gruner, S. M. & Fraden, S. (2014). IUCrJ, 1, 349–360. [DOI] [PMC free article] [PubMed]
- Hope, H. (1988). Acta Cryst. B44, 22–26. [DOI] [PubMed]
- Huang, C.-Y., Olieric, V., Ma, P., Panepucci, E., Diederichs, K., Wang, M. & Caffrey, M. (2015). Acta Cryst. D71, 1238–1256. [DOI] [PMC free article] [PubMed]
- Hunter, M. S. et al. (2014). Sci. Rep. 4, 6026. [DOI] [PMC free article] [PubMed]
- Jackson, R. L. (1974). US Patent No. 3856771.
- Jungmann-Smith, J. H., Bergamaschi, A., Cartier, S., Dinapoli, R., Greiffenberg, D., Johnson, I., Maliakal, D., Mezza, D., Mozzanica, A., Ruder, C., Schaedler, L., Schmitt, B., Shi, X. & Tinti, G. (2014). J. Instrum. 9, P12013.
- Kabsch, W. (2010a). Acta Cryst. D66, 133–144. [DOI] [PMC free article] [PubMed]
- Kabsch, W. (2010b). Acta Cryst. D66, 125–132. [DOI] [PMC free article] [PubMed]
- Kiefersauer, R., Than, M. E., Dobbek, H., Gremer, L., Melero, M., Strobl, S., Dias, J. M., Soulimane, T. & Huber, R. (2000). J. Appl. Cryst. 33, 1223–1230.
- Kim, C. U., Kapfer, R. & Gruner, S. M. (2005). Acta Cryst. D61, 881–890. [DOI] [PubMed]
- Kisselman, G., Qiu, W., Romanov, V., Thompson, C. M., Lam, R., Battaile, K. P., Pai, E. F. & Chirgadze, N. Y. (2011). Acta Cryst. D67, 533–539. [DOI] [PMC free article] [PubMed]
- Kmetko, J., Husseini, N. S., Naides, M., Kalinin, Y. & Thorne, R. E. (2006). Acta Cryst. D62, 1030–1038. [DOI] [PubMed]
- Kupitz, C. et al. (2014). Nature, 513, 261–265. [DOI] [PMC free article] [PubMed]
- McCoy, A. J., Grosse-Kunstleve, R. W., Adams, P. D., Winn, M. D., Storoni, L. C. & Read, R. J. (2007). J. Appl. Cryst. 40, 658–674. [DOI] [PMC free article] [PubMed]
- McPherson, A. (2000). J. Appl. Cryst. 33, 397–400.
- Meents, A., Dittrich, B. & Gutmann, S. (2009). J. Synchrotron Rad. 16, 183–190. [DOI] [PubMed]
- Meents, A., Gutmann, S., Wagner, A. & Schulze-Briese, C. (2010). Proc. Natl Acad. Sci. USA, 107, 1094–1099. [DOI] [PMC free article] [PubMed]
- Mozumder, A. & Magee, J. L. (1966). Radiat. Res. 28, 203–214. [PubMed]
- Mueller, C. et al. (2015). Struct. Dyn. 2, 054302. [DOI] [PMC free article] [PubMed]
- Murray, T. D., Lyubimov, A. Y., Ogata, C. M., Vo, H., Uervirojnangkoorn, M., Brunger, A. T. & Berger, J. M. (2015). Acta Cryst. D71, 1987–1997. [DOI] [PMC free article] [PubMed]
- Neutze, R. & Moffat, K. (2012). Curr. Opin. Struct. Biol. 22, 651–659. [DOI] [PMC free article] [PubMed]
- Owen, R. L., Paterson, N., Axford, D., Aishima, J., Schulze-Briese, C., Ren, J., Fry, E. E., Stuart, D. I. & Evans, G. (2014). Acta Cryst. D70, 1248–1256. [DOI] [PMC free article] [PubMed]
- Owen, R. L., Rudiño-Piñera, E. & Garman, E. F. (2006). Proc. Natl Acad. Sci. USA, 103, 4912–4917. [DOI] [PMC free article] [PubMed]
- Pennicard, D., Lange, S., Smoljanin, S., Hirsemann, H. & Graafsma, H. (2012). J. Instrum. 7, C11009.
- Philipp, H. T., Hromalik, M., Tate, M., Koerner, L. & Gruner, S. M. (2011). Nucl. Instrum. Methods Phys. Res. Sect. A, 649, 67–69.
- Pinker, F., Brun, M., Morin, P., Deman, A. L., Chateaux, J. F., Oliéric, V., Stirnimann, C., Lorber, B., Terrier, N., Ferrigno, R. & Sauter, C. (2013). Cryst. Growth Des. 13, 3333–3340.
- Rasmussen, B. F., Stock, A. M., Ringe, D. & Petsko, G. A. (1992). Nature, 357, 423–424. [DOI] [PubMed]
- Ravelli, R. B. & McSweeney, S. M. (2000). Structure, 8, 315–328. [DOI] [PubMed]
- Redecke, L. et al. (2013). Science, 339, 227–230.
- Roedig, P., Vartiainen, I., Duman, R., Panneerselvam, S., Stübe, N., Lorbeer, O., Warmer, M., Sutton, G., Stuart, D. I., Weckert, E., David, C., Wagner, A. & Meents, A. (2015). Sci. Rep. 5, 10451. [DOI] [PMC free article] [PubMed]
- Sanchez-Weatherby, J., Bowler, M. W., Huet, J., Gobbo, A., Felisaz, F., Lavault, B., Moya, R., Kadlec, J., Ravelli, R. B. G. & Cipriani, F. (2009). Acta Cryst. D65, 1237–1246. [DOI] [PubMed]
- Sherrell, D. A., Foster, A. J., Hudson, L., Nutter, B., O’Hea, J., Nelson, S., Paré-Labrosse, O., Oghbaey, S., Miller, R. J. D. & Owen, R. L. (2015). J. Synchrotron Rad. 22, 1372–1378. [DOI] [PMC free article] [PubMed]
- Sjögren, T., Carlsson, G., Larsson, G., Hajdu, A., Andersson, C., Pettersson, H. & Hajdu, J. (2002). J. Appl. Cryst. 35, 113–116.
- Southworth-Davies, R. J., Medina, M. A., Carmichael, I. & Garman, E. F. (2007). Structure, 15, 1531–1541. [DOI] [PubMed]
- Stellato, F. et al. (2014). IUCrJ, 1, 204–212. [DOI] [PMC free article] [PubMed]
- Tenboer, J. et al. (2014). Science, 346, 1242–1246. [DOI] [PMC free article] [PubMed]
- Teng, T. & Moffat, K. (2000). J. Synchrotron Rad. 7, 313–317. [DOI] [PubMed]
- Thomanek, U. F., Parak, F., Mössbauer, R. L., Formanek, H., Schwager, P. & Hoppe, W. (1973). Acta Cryst. A29, 263–265.
- Tilton, R. F., Dewan, J. C. & Petsko, G. A. (1992). Biochemistry, 31, 2469–2481. [DOI] [PubMed]
- Umena, Y., Kawakami, K., Shen, J.-R. & Kamiya, N. (2011). Nature, 473, 55–60. [DOI] [PubMed]
- Weierstall, U. et al. (2014). Nat. Commun. 5, 3309. [DOI] [PMC free article] [PubMed]
- Weik, M., Ravelli, R. B., Kryger, G., McSweeney, S., Raves, M. L., Harel, M., Gros, P., Silman, I., Kroon, J. & Sussman, J. L. (2000). Proc. Natl Acad. Sci. USA, 97, 623–628. [DOI] [PMC free article] [PubMed]
- Zarrine-Afsar, A., Barends, T. R. M., Müller, C., Fuchs, M. R., Lomb, L., Schlichting, I. & Miller, R. J. D. (2012). Acta Cryst. D68, 321–323. [DOI] [PubMed]
- Zeldin, O. B., Gerstel, M. & Garman, E. F. (2013). J. Appl. Cryst. 46, 1225–1230.
- Ziaja, B., London, R. A. & Hajdu, J. (2005). J. Appl. Phys. 97, 064905.
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.