Precision Tests of a Quantum Hall Effect Device DC Equivalent Circuit Using Double-Series and Triple-Series Connections

A Jeffery 1, R E Elmquist 1, M E Cage 1
PMCID: PMC4887258  PMID: 29151768

Abstract

Precision tests verify the dc equivalent circuit used by Ricketts and Kemeny to describe a quantum Hall effect device in terms of electrical circuit elements. The tests employ the use of cryogenic current comparators and the double-series and triple-series connection techniques of Delahaye. Verification of the dc equivalent circuit in double-series and triple-series connections is a necessary step in developing the ac quantum Hall effect as an intrinsic standard of resistance.

Keywords: ac quantum Hall effect, cryogenic current comparator, dc quantum Hall effect, equivalent electrical circuit, quantized Hall resistance, two-dimensional electron gas

1. Introduction

In the integer dc quantum Hall effect [13] the Hall resistance RH of the the plateau of a fully-quantized, two-dimensional electron gas (2DEG) is RH(i) = h/(e2i), where h is the Planck constant, e is the elementary charge, and i is an integer. We assume that RH(i) has the value of the von Klitzing constant, 25 812.807 Ω/i. The current flow within the 2DEG is nearly dissipationless in the quantum Hall plateau regions of high-quality devices, and the longitudinal voltage drops, Vx, and the longitudinal resistance drops, Rx, along the sides of the sample are very small.

It is important to remember that the conducting charges are electrons, and in the presence of a magnetic field, B, the sign and direction of the conducting charges determines the sign of the potentials around the periphery of the device. For the example shown in Fig. 1 the magnetic field is in the positive z direction, and the conducting electrons enter at the upper left-hand corner and exit at the lower right-hand corner [47], as indicated by the shaded curves. These corners remain the same on current reversal, but they interchange on magnetic field reversal. The potential probes 2, 4, and 6 are near the potential of the source S. Probes 1, 3, and 5 are near the potential of the drain D, and in this example have a positive potential relative to the source. The two positive potential sides of the device are indicated by thick lines. On current reversal, those two sides would have a negative potential relative to the source, and would be indicated by thin lines.

Fig. 1.

Fig. 1

Top view of a quantum Hall device showing the direction of the current flow (positive carriers) in and out of the sample. The shaded curves indicate the electron flow pattern for a magnetic field pointing into the sample in the positive z direction. The bold lines indicate the positive potentials along the periphery of the sample.

Electrons are moving from left to right in the example of Fig. 1, but electric circuit analyses traditionally assume currents composed of positive charges. Therefore the figure shows positively-charged currents ID and IS entering and leaving the device from right to left. If an external measurement system is connected to potential probes 3 and 4 to measure the quantum Hall voltage VHV3V4 = RHID, then additional positively-charged currents I3 and I4 can enter and leave the device. These two currents are arbitrarily assumed to enter potential probe 3 and exit probe 4 in Fig. 1. I3 and I4 are small compared to ID and IS.

2. The DC Equivalent Circuit

Ricketts and Kemeny [8] have described the electrical behavior of a quantum Hall device in terms of an equivalent circuit. Figure 2 shows this equivalent circuit for the wiring configuration, current, and magnetic field directions of Fig. 1 when the longitudinal resistance Rx is negligibly small. In ideal conditions, each arm of the circuit extends from the 2DEG to a source, drain, or potential contact, and has a large resistance RH/2 and a small contact resistance rcA for each arm A, where A represents contacts, 3, 4, S, or D. There is also a wire resistance rwA to each contact. The wire resistance is usually dominated by the sample probe leads, but it can also include wire-bonds to the sample header.

Fig. 2.

Fig. 2

Equivalent circuit for the quantum Hall device of Fig. 1 with external leads to two potential contacts (3 and 4) and a source and drain (S and D). The orientations of the voltage generators are for the indicated current and magnetic field directions. The positions of the intrinsic resistances RH/2, contact resistances rc, lead resistances rw, and probe positions (S, D, 3, and 4), are also indicated.

Between each pair of arms A and B there is a voltage generator VAB, where VAB is defined as

VABRH2|IA+IB|. (1)

The sign within the quantity |IA + IB| is positive if both IA and IB enter or leave the device, and negative if one current enters and the other leaves. VAB is zero if both IA and IB are zero.

The currents IA are assumed to be for positively-charged carriers, but the actual current is due to electrons traveling in the opposite direction, which in the presence of the magnetic field affects the signs of the potentials around the device periphery. Thus the orientations of the voltage generators are chosen in Fig. 2 to obtain the correct potentials. As shown in the simplified sketches of Fig. 3, the orientations (signs) of the voltage generators all reverse upon current reversal. They also all reverse on magnetic field reversal, but in that case the current directions remain the same.

Fig. 3.

Fig. 3

Diagrams of quantum Hall devices and the corresponding equivalent circuits to show how different magnetic field directions and different current flow directions affect the orientation of the voltage generators. (a) Positive current flow from D to S and a magnetic field in the positive z direction. (b) Current reversed and magnetic field unchanged. (c) Current unchanged from first case and magnetic field reversed. For simplicity, the contact resistances are not shown.

Let us, for the moment, neglect the contact resistances rcA and calculate the potentials V3, V4, and VD at the device contacts 3, 4, and D in Fig. 2 relative to the source potential VS:

V3VSISRH2VS3+I3RH2=RHIS=RH(ID+I3I4) (2a)
V4VSRHI4 (2b)
VDVSRH(ISI3). (2c)

By subtracting Eq. (2b) from (2a), this leads to a correction to the quantum Hall voltage VHRHID of order RHI3. Even though RH is large, the currents I3 and I4 can be made small enough in dc bridges to ensure that the correction to VH is negligible. That is not necessarily the case, however, for bridges used to measure the ac quantized Hall resistance. Therefore something must be done in ac resistance measurements to ensure that I3 and I4 are indeed small enough. A solution, that has been found by Delahaye [9,10], involves double-series or triple-series connections to quantum Hall devices.

3. Double-Series Connections

The total resistance in each arm, A, of the circuit can be split into the intrinsic resistance component RH/2 and the component rA, defined as

rArcA+rwA+rtest+rCCC, (3)

where A represents S, D, 3 or 4, rcA is the contact resistance, rwA is the wire resistance, rtest is a resistor that can be added to test the circuit equations, and rCCC is the resistance of a cryogenic current comparator that can be placed in arm A to measure the current IA. We display in Fig. 4 the equivalent circuit representation of two double-series connections to the quantum Hall device of Fig. 1, where rA is defined by Eq. (3). A total current IT enters point Y. It separates into currents I3, ID, I4, and IS, and then exits point Z as IT. By summing potentials around circuit loops one obtains the relations

I3=rD(RH+r3)ID=rD(RH+rD+r3)IT (4)

and

I4=rS(RH+r4)IS=rS(RH+rS+r4)IT. (5)

Therefore, I3 and I4 are small fractions of ID and IS.

Fig. 4.

Fig. 4

Double-series connection in an equivalent circuit representation of a quantum Hall device; rS, rD, r3, and r4 are defined in Eq. (3).

Four-terminal resistance measurements, RY,Z, can be made between points Y and Z by using the two current leads and the potential leads VY and VZ:

RY,Z=VYVZIT, (6)

or

RY,Z=RH[1+rDr3RH(RH+rD+r3)+rSr4RH(RH+rS+r4)]. (7)

RY,Z differs from RH by the two correction terms in Eq. (7), which we will label ΔRH(theory)/RH.

4. Triple-Series Connections

Figure 5 shows the equivalent circuit representation of two triple-series connections to a single device. The equations are

I3=rDr1[(RH+rD+r1)(RH+r3)+rDr1]IT, (8)
I1=rD(RH+rD+r1)(ITI3), (9)
I4=rSr6[(RH+rS+r6)(RH+r4)+rSr6]IT, (10)
I6=rS(RH+rS+r6)(ITI4), (11)

and

RY,Z=RH{1+rDr1r3RH[(RH+rD+r1)(RH+r3)+rDr1+rSr6r4RH[(RH+rS+r6)(RH+r4)+rS]}. (12)

RY,Z differs from RH by the two correction terms in Eq. (12), which we again label ΔRH(theory)/RH. Precision tests of these double-series and triple-series equations are described in Secs. 6 and 7.

Fig. 5.

Fig. 5

Triple-series connection in an equivalent circuit representation of a quantum Hall device.

5. Samples and Circuit Resistances

Two standards-quality samples were used. One sample, designated as GaAs(8), is a GaAs/AlxGa1–xAs heterostructure grown by molecular beam epitaxy at AT&T Bell Laboratories,1 with x = 0.29 being the fraction of aluminum atoms replacing gallium atoms in the crystal. It has a zero magnetic field mobility of about 100 000 cm2/(V·s) at 4.2 K, and a carrier density of 5.7×1011 cm−2. It has a length of 4.6 mm and a width of 0.4 mm, and the two outer Hall potential probe pairs are displaced from the central pair by 1 mm. Gold wires of diameter 25 µm were soldered onto alloyed indium dots to make electrical contact to the 2DEG.

The other sample is a BIPM/EUROMET GaAs/AlxGa1–xAs heterostructure grown by metalorganic vapor-phase epitaxy at the Laboratories d’Electronic Philips, France [11]. It is designated as E7C, has a Si3N4 protective coating, a zero magnetic field mobility of about 270 000 cm2/(V·s) at 4.2 K, a carrier density of 4.9×1011 cm−2, and an aluminum fraction of x = 0.29. It has a length of 2.2 mm and a width of 0.4 mm, and the two outer Hall potential probe pairs are displaced from the central pair by 0.5 mm. Gold wires of diameter 25 µm were wire-bonded onto enlarged Au/Cr contact pads evaporated onto the preexisting AuGe/Ni pads [12]. Both samples were mounted on 12-pin, T0-8 headers.

The samples were cooled in a He-3 refrigerator insert to a temperature of 0.3 K. The sample probe uses twisted-pairs of PTFE-covered copper thermocouple wire, whose diameters are only 76 µm to minimize heat loss. The resistances of these wires are about 10.0 Ω at room temperature and about 5.4 Ω when the sample is cold. There are two wires in the sample probe which are soldered together at the source connection of the T0-8 header, and two wires soldered together for the drain contact. The wire resistances rwS and rwD were determined by using one-half of the series resistances of the two source and the two drain wires, respectively. The wire resistances of the potential leads were then estimated by using the average values of rwS and rwD because all the wire resistances were equal to within 0.03 Ω. The wire resistances changed by as much as 0.2 Ω with liquid helium level, so it was important to monitor these resistances.

The contact resistances rcA were determined from three-terminal resistance measurements. They were negligibly small for GaAs(8), and averaged about (0.17±0.10) Ω for the E7C sample. (Here, and throughout this paper, all quoted uncertainties are one standard deviation estimates.) The experiments were all done on the i = 2 plateau, so RH = 12 906.403 5 Ω. The series resistance value of the 32-turn winding of a cryogenic current comparator, used in Sec. 6 to measure currents in selected leads of the sample probe, varied between 1.645 Ω and 1.650 Ω.

6. Current Measurements

Measurements were made of the current in selected potential leads of the GaAs(8) sample to determine if these currents agreed with the values predicted by Eqs. (4), (5), and (8) to (11) in Secs. 3 and 4. Both double-series and triple-series connections were used. The magnetic field direction was opposite to that for the equivalent circuits of Figs. 4 and 5, so the connections were on opposite sides of the sample. Both positive and negative currents were used for the applied current IT in order to eliminate the effects of thermally-induced voltages. The experimental current in the potential lead was found using a 32-turn cryogenic current comparator (CCC) winding coupled to a dc-SQUID by measuring the SQUID flux-locked-loop output voltage. The sensitivity of this measurement is (38.5±0.2) nA/V.

6.1 Double-Series Connections

Results of the current measurements in potential leads are summarized in Table 1 for the double-series configuration shown in Fig. 6 and IT = (39.914±0.001) µA. The agreement between measured and calculated values of the current are within the experimental uncertainty. No significant differences in currents measured (< 0.005 %) were observed for series-connections using different potential probe positions. Similar results were obtained if a double-series connection was made on only the side of the sample where the potential lead current was being measured.

Table 1.

Experimental and calculated values of currents in the potential lead of a double-series connection with and without added test resistances. Theoretical current values are found from Eq. (4), but with probe 2 rather than probe 3, and the magnetic field reversed

Test resistance
rtest (Ω)
Position of test resistance Measured current
I2 (nA)
Relative change in I2 from zero test resistance case
(%)
Theoretical current
I2 (nA)
Theoretical current ratio
I2/IT (%)
Theoretical current/measured current
  0 16.77   0.0 16.90 0.042 1.01±0.01
20 Lead 2 16.74 −0.2 16.87 0.042 1.01±0.01
10 Drain lead 47.67 184.2 47.73 0.120 1.00±0.01
20 Drain lead 78.43 408.0 78.59 0.197 1.00±0.01

Fig. 6.

Fig. 6

Double-series connection of a quantum Hall sample, and the position of a cryogenic current comparator (CCC) used to measure current in potential lead 2.

Resistances with nominal values of 10 Ω and 20 Ω, and actual values of (9.989±0.002) Ω and (20.008±0.002) Ω, were added to the potential lead or the drain lead to observe their effect. Resistance added to the potential lead had little effect on the current. For example, when 20 Ω was added to potential lead 2, the measured current I2 decreased by only a factor of 5. When 10 Ω resistance was added to the drain lead, however, I2 increased by a factor of 2.8, and by a factor of 4.7 if 20 Ω was added.

The measurements are sensitive to changes in lead resistance (which depends on the liquid helium level), especially in the source or drain leads. An increase of 0.1 Ω in the drain lead resistance increases the calculated current by 1.8 % of the current.

We also obtained similar results when leads 1 and 5 were used in place of leads 5 and S of Fig. 6, and leads 6 and 2 in place of 2 and D. Therefore it does not matter what contacts are used as the source and drain as long as their contact resistances are comparable.

6.2 Triple-Series Connections

Current measurements were made using the triple-series connection to the drain side of the sample, as shown in Fig. 7. The results are summarized in Table 2 for the current in potential lead 4 and IT = (39.914±0.001) µA. This example illustrates the extent to which the triple-series connection reduces the current in the potential leads. In fact, the current in lead 4 was so small (6×10−12 A) with no added resistance in the circuit that it was close to the detection limit of the measurement and could not be accurately compared with the theory, even when a 64-turn CCC winding was used to increase the sensitivity.

Fig. 7.

Fig. 7

Triple-series connection of a quantum Hall sample, and the position of a CCC used to measure current in potential lead 4.

Table 2.

Experimental and calculated values of currents in potential lead 4 of a triple-series connection with and without added test resistances. Theoretical current values are found from Eq. (8), but with probes 2 and 4 rather than probes 1 and 3, and the magnetic field reversed

Test resistance
rtest (Ω)
Position of test resistance Measured current
I4 (nA)
Theoretical current
I4(nA)
Theoretical current ratio
I4/IT(10−6)
Theoretical current/measured currenta
  0   0.006   0.007 0.18
250 Drain lead 0.32 0.33 8.24 1.03±0.01
500 Drain lead 0.62 0.64 16.00 1.02±0.01
250 Lead 2 0.32 0.33 8.16 1.02±0.01
500 Lead 2 0.63 0.63 15.84 1.01±0.01
a

The value for the theoretical/experimental current ratio with no added resistance is 1.187, which is in much poorer agreement than observed with the double-series connection because the current in potential lead 4 for the case with no added resistance is four orders of magnitude smaller than the value obtained for the double-series connection—and is therefore close to the experimental detection noise. By adding large test resistances, the current was increased so that it was only two orders of magnitude less than the current measured for the double-series connection, and better theoretical/experimental ratios were obtained.

A better test of agreement between the measured and theoretical current values was obtained when large resistances (250 Ω and 500 Ω) were added to the drain lead or to potential lead 2. The large added resistances increased the currents in lead 4 by about two orders of magnitude, which was still about two orders of magnitude smaller than the double-series case. As a result, these current measurements do not agree as well with the calculated values as for the double-series connections. As predicted by Eq. (8), there is very little difference in the values of I4 whether resistance is added to the drain lead or to lead 2.

7. Resistance Measurements

A cryogenic current comparator measurement system, which compares the value of a i = 2 quantized Hall resistor to a 100 Ω resistor, was used for resistance measurements of sample E7C. Double-series and triple-series connections were employed, and resistances were again added to the drain lead to observe their effect on the quantized Hall resistance. The applied current was IT = (30.702±0.001) µA, and the magnetic field direction was the same as in Figs. 4 and 5.

7.1 Double-Series Connections

The configuration for the double-series connection is shown in Fig. 8, and the results are in Table 3. Four-terminal double-series resistance measurements, RY,Z, between points Y and Z are compared with regular quantum Hall measurements, RH, for potential probes 3 and 4 to obtain a relative experimental difference, ΔRH(exp)/RH, between double-series and regular quantum Hall measurements. The relative change from the regular quantized Hall resistance when using the double-series connection was 3.30×10−7 without any added resistance, 9.39×10−7 when a (20.008±0.002) Ω resistor was added to the drain, and 1.463×10−6 when a (37.045±0.002) Ω resistor was added.

Fig. 8.

Fig. 8

Double-series connection used for the resistance measurements of a quantum Hall sample. Measurements were made between points Y and Z.

Table 3.

Experimental and calculated values of the relative change in quantized Hall resistance, ΔRH/RH, from the usual quantum Hall measurement between potential probes 3 and 4 for a measurement between points Y and Z with a double-series connection. (See Sec.7.1 and Eq. (13) for the definitions of ΔRH(exp)/RH, ΔRH(theory)/RH and δRH/RH.)

Test resistance
rtest (Ω)
Position of test resistance Experimental result
ΔRH(exp)/RH (10−6)
Theoretical result
ΔRH(theory)/RH(10−6)
Difference in results
δRH/RH (10−6)
  0 0.330 0.312 0.019
20 Drain lead 0.935 0.918 0.018
37 Drain lead 1.463 1.432 0.031

The calculated relative differences ΔRH(theory)/RH were obtained from Eq. (7). The relative differences

δRH/RH=ΔRH(exp)/RHΔRH(theory)/RH (13)

between the experimental and calculated values for ΔRH/RH in Table 3 range from 1.9×10−8 to 3.1×10−8, which is approximately four to six times the 5×10−9 relative experimental uncertainty of the 20 min duration CCC measurements. A likely explanation for these large values of δRH/RH is that the lead resistances, which were measured earlier in the day, increased significantly from those used in the calculations. For example, an increase in the lead resistances of 0.1 Ω, due to a decreasing liquid helium level, changes the predicted double-series-connected quantum Hall resistance values such that the three values of δRH/RH become 6×10−9, −3×10−9, and −3×10−9, respectively. To obtain reliable double-series measurements, the lead resistances need to be monitored frequently.

7.2 Triple-Series Connections

The triple-series connection is shown in Fig. 9, and the results of the measurements in Table 4. Four-terminal triple-series resistance measurements, RY,Z, between points Y and Z are compared with regular quantum Hall measurements, RH, for potential probes 3 and 4 to obtain an experimental relative difference, ΔRH(exp)/RH, between triple-series and regular quantum Hall measurements. The calculated differences ΔRH(theory)/RH were obtained from Eq. (12).

Fig. 9.

Fig. 9

Triple-series connection used for the resistance measurements of a quantum Hall sample. Measurements were made between points Y and Z.

Table 4.

Experimental and calculated values of the relative change in quantized Hall resistance, ΔRH/RH, from the usual quantum Hall measurement between potential probes 3 and 4 for a measurement between points Y and Z with a triple-series connection. (See Sec.7.1 and Eq. (13) for the definitions of ΔRH(exp)/RH, ΔRH(theory)/RH and δRH/RH)

Test resistance
rtest (Ω)
Position of test resistance Experimental result
ΔRH(exp)/RH (10−6)
Theoretical result
ΔRH(theory)/RH (10−6)
Difference in results
δRH/RH (10−6)
  37 Drain lead −0.003 0.001 −0.004
50000 Drain lead 0.124 0.124 0.000

ΔRH(exp)/RH and ΔRH(theory)/RH are very small for triple-series connections. A 50 kΩ resistor had to be added in order to observe a significant relative change (1.24×10−6) from the actual quantized Hall resistance RH. Experimental and theoretical values are in excellent agreement, and differ only by 4×10−9.

8. Conclusions

The equations in Secs. 3 and 4, which are based on an equivalent circuit model of the quantum Hall device, give very good predictions for the current in the potential leads of the device, and for the change in the quantized Hall resistance when double-series and triple-series connections are employed. The double and triple-series connections have proved to be an effective means of reducing the current in the potential leads of a quantum Hall device, which is required in precision ac quantized Hall resistance measurements [10,1315].

The triple-series connection is especially effective in reducing the current in the potential leads, even when very large resistances are added. Based on this, triple-series connections would be the better choice for use in high precision ac measurements.

The double-series and triple-series connections in this work were made outside of the liquid helium dewar. An alternative would be to make the series connections directly between the sample contacts. This would eliminate the effects of changes in the lead resistances due to variations in the liquid helium level and would reduce the number of leads required in the sample probe. It would, however, restrict one’s ability to make quantized Hall resistance measurements with reversed magnetic field, and to measure the longitudinal resistance.

It must be taken into consideration that the equivalent circuit model and the measurements are dc. It is not known for certain if this model will apply to high-accuracy ac measurements because no capacitive and inductive components are included in the model.

Four-terminal-pair connections [16] are used in high-precision ac measurements to ensure insensitivity to variations in series impedances and shunt admittances of the leads to the standard. A concern is that the resistance standard be composed of RH, and not a combination of RH and other resistances, capacitances, and inductors. In a four-terminal-pair definition, the resistance standard is composed of the quantized Hall resistance and two of the coaxial cables used in the four-terminal-pair connections [16]. The impedances and admittances of the current and potential coaxial leads in the sample probe would therefore be part of the quantized Hall resistance standard [17] and would have to be accounted for, whether or not the double-series or triple-series connections were made at the sample contacts or outside of the dewar. Otherwise the resistance standard would not have the intrinsic value RH.

Acknowledgments

We thank A. C. Gossard of the University of California at Santa Barbara who made the MBE-grown GaAs/AlGaAs heterostructure GaAs(8) while at AT&T Bell Laboratories, D.C. Tsui of Princeton University who defined the device geometry and made ohmic contacts to the 2DEG, and J. Q. Shields and K. C. Lee for useful discussions and comments. This work was supported in part by the Calibration Coordination Group of the Department of Defense.

Biography

About the authors: Anne-Marie Jeffery, Randolph E. Elmquist, and Marvin E. Cage are physicists in the Electricity Division at NIST. The National Institute of Standards and Technology is an agency of the Technology Administration, U.S. Department of Commerce.

Footnotes

1

Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

9. References

  • 1.von Klitzing K, Dorda G, Pepper M. New Method for High-Accuracy Determination of the Fine-Structure Constant Based on Quantized Hall Resistance. Phys Rev Lett. 1980;45:494–497. [Google Scholar]
  • 2.Prange RE, Girvin SM, editors. The Quantum Hall Effect. Springer-Verlag; New York: 1987. pp. 1–419. [Google Scholar]
  • 3.Van Degrift CT, Cage ME, Girvin SM, editors. The Integral and Fractional Quantum Hall Effects. American Association of Physics Teachers; College Park, Maryland: 1991. pp. 1–116. [Google Scholar]
  • 4.Dominguez D, von Klitzing K, Ploog K. Optimization of the Contact Geometry for Accurate Quantized Hall Resistance Measurements. Metrologia. 1989;26:197–201. [Google Scholar]
  • 5.Kent AJ. Imaging the Electron-Phonon Interaction in the 2D Electron Gas. Physica B. 1991;169:356–362. [Google Scholar]
  • 6.Fontein PF, Hendriks P, Blom FAP, Wolter JH, Giling LJ, Beenakker CWJ. Spatial Potential Distribution in GaAs/AlGaAs Heterostructures Under Quantum Hall Conditions Studied with the Linear Electro-Optic Effect. Surf Sci. 1992;263:91–96. doi: 10.1103/physrevb.43.12090. [DOI] [PubMed] [Google Scholar]
  • 7.Klass U, Dietsche W, von Klitzing K, Ploog K. Image of the Dissipation in Gated Quantum Hall Effect Samples. Surf Sci. 1992;263:97–99. [Google Scholar]
  • 8.Ricketts BW, Kemeny PC. Quantum Hall Effect Devices as Circuit Elements. J Phys D: Applied Phys. 1988;21:483–487. [Google Scholar]
  • 9.Delahaye F. Series and Parallel Connection of Multiterminal Quantum Hall Effect Devices. J Appl Phys. 1993;73:7915–7920. [Google Scholar]
  • 10.Delahaye F. Accurate ac Measurements of the Quantized Hall Resistance from 1 Hz to 1.6 kHz. Metrologia. 1995;31:367–373. [Google Scholar]
  • 11.Piquemal F, Geneves G, Delahaye F, Andre J-P, Patillon J-N, Frijink P. Report on a Joint BIPM-EUROMET Project for the Fabrication of QHE Samples by the LEP. IEEE Trans Instrum Meas. 1993;42:264–268. [Google Scholar]
  • 12.Lee KC. Bonding Wires to Quantized Hall Resistors. IEEE Trans Instrum Meas. 1995;44:249–253. [Google Scholar]
  • 13.Melcher J, Warnecke P, Hanke R. Comparison of Precision ac and dc Measurements with the Quantized Hall Resistance. IEEE Trans Instrum Meas. 1993;42:292–294. [Google Scholar]
  • 14.Hartland A, Kibble BP, Rodgers PJ, Bohacek J. ac Measurements of the Quantized Hall Resistance. IEEE Trans Instrum Meas. 1995;44:245–248. [Google Scholar]
  • 15.Piquemal F, Trapon G, Geneves G. ac Measurements of the Minimum Longitudinal Resistivity of a QHE Sample from 10 Hz to 10 kHz. IEEE Trans Instrum Meas. preprint. [Google Scholar]
  • 16.Cutkosky RD. Techniques for Comparing Four-Terminal-Pair Admittance Standards. J Res Natl Bur Stand (US) 1970;74C:63–78. [Google Scholar]
  • 17.Melcher J. Systematic Errors of Long Thin Coaxial Cables for the Connection of Four-Terminal Pair Devices. preprint. [Google Scholar]

Articles from Journal of Research of the National Institute of Standards and Technology are provided here courtesy of National Institute of Standards and Technology

RESOURCES