Where Are We Now?
For patients with a primary bone tumor of the lower extremity, limb-salvage surgery is the preferred approach when it is feasible to perform [4]. For patients undergoing distal-femur resections, endoprosthetic reconstructions allow immediate partial weight bearing and generally restore an acceptable level of limb function [9]. Traditional fixation approaches, including cement or press-fit intramedullary stems, depend in large measure on the amount of remaining bone into which the surgeon can anchor the device; the longer-term survivorship of these implants becomes critically dependent upon that sometimes-scarce resource. Among patients who survive for longer periods of time, aseptic loosening can result in additional disability and major surgery [6, 10].
Some patients simply do not have enough remaining host bone to allow secure fixation of the endoprosthesis. The Compress® Compliant Pre-stress implant (Biomet Inc, Warsaw, IN USA) was designed to promote osseointegration at the bone-prosthesis interface [2, 3, 8], particularly in patients with severe bone loss. In that setting, traditional cemented or cementless devices would have a high risk of failure. Compress® secures a titanium porous-coated spindle to the bone through a spring-loaded device anchored to a short-traction bar. A recent long-term outcome study demonstrated a survivorship of 80% using this fixation technique for distal femur implants [6]. However, little information has been reported analyzing the frequency of compressive osseointegration spindle failure in distal femoral reconstructions, nor have studies focused on the characteristics of rotational failure and the risk factors for this mechanical complication.
While we have some insight into this important complication—patients with less remaining host bone seem to be at risk for it [5, 7]—a recent study that sought to attribute these failures to rotational stresses was unable to prove this contention, perhaps because of insufficient sample size [6].
Where Do We Need To Go?
The present study showed that, while failures of compressive osseointegration implant do occur, and that spindle revision is sometimes called for, these events are uncommon and the overall survival rate from spindle failure is relatively high even at long-term followup. Injuries with a twisting mechanism produce the most rotational failures. It is recommended that we include antirotation pins at the fixation during the surgical procedure in order to prevent these failures. However, results from the current study showed that when omitting antirotational pins from surgery, the rate of rotational failures did not increase. More information is needed before we can properly determine whether to use antirotation pins.
How Do We Get There?
A future controlled study to determine the effect of these antirotation pins on the frequency of rotational failure is necessary in order to speak more definitively to this important unanswered question. Biomechanical studies should evaluate different stresses in heavy loading between stemmed implants (cemented and uncemented) and osseointegration implants. Results of such studies might also be applicable to the care of patients undergoing nononcological reconstructions, including revisions of traditional total hip, knee, or upper-extremity joint replacements.
Footnotes
This CORR Insights® is a commentary on the article “How Often Does Spindle Failure Occur in Compressive Osseointegration Endoprostheses for Oncologic Reconstruction?” by Goldman and colleagues available at: DOI: 10.1007/s11999-016-4839-7.
The author certifies that he, or a member of his immediate family, has no funding or commercial associations (eg, consultancies, stock ownership, equity interest, patent/licensing arrangements, etc.) that might pose a conflict of interest in connection with the submitted article.
All ICMJE Conflict of Interest Forms for authors and Clinical Orthopaedics and Related Research ® editors and board members are on file with the publication and can be viewed on request.
The opinions expressed are those of the writers, and do not reflect the opinion or policy of CORR ® or The Association of Bone and Joint Surgeons®.
This CORR Insights® comment refers to the article available at DOI: 10.1007/s11999-016-4839-7.
This comment refers to the article available at: http://dx.doi.org/10.1007/s11999-016-4839-7.
References
- 1.Avedian RS, Goldsby RE, Kramer MJ, O’Donnell RJ. Effect of chemotherapy on initial compressive osseointegration of tumor endoprostheses. Clin Orthop Relat Res. 2007;459:48–53. doi: 10.1097/BLO.0b013e3180514c66. [DOI] [PubMed] [Google Scholar]
- 2.Bini SA, Johnston JO, Martin DL. Compliant prestress fixation in tumor prostheses: interface retrieval data. Orthopedics. 2000;23:707–711. doi: 10.3928/0147-7447-20000701-18. [DOI] [PubMed] [Google Scholar]
- 3.Cristofolini L, Bini SA, Toni A. In vitro testing of a novel limb salvage prosthesis for the distal femur. Clin Biomech (Bristol, Avon). 1998;13:608–615. [DOI] [PubMed]
- 4.Grimer RJ, Carter SR, Pynsent PB. The cost-effectiveness of limb salvage for bone tumours. J Bone Joint Surg Br. 1997;79:558–561. doi: 10.1302/0301-620X.79B4.7687. [DOI] [PubMed] [Google Scholar]
- 5.Guo W, Ji T, Yang R, Tang X, Yang Y. Endoprosthetic replacement for primary tumours around the knee: experience from Peking University. J Bone Joint Surg Br. 2008;90:1084–1089. doi: 10.1302/0301-620X.90B8.20240. [DOI] [PubMed] [Google Scholar]
- 6.Healey JH, Morris CD, Athanasian EA, Boland BJ. Compress® knee arthroplasty has 80% 10-year survivorship and novel forms of bone failure. Clin Orthop Relat Res. 2013;471:774–783. doi: 10.1007/s11999-012-2635-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Kawai A, Muschler GF, Lane JM, Otis JC, Healey JH. Prosthetic knee replacement after resection of a malignant tumor of the distal part of the femur: medium to long-term results. J Bone Joint Surg Am. 1998;80:636–647. doi: 10.1302/0301-620X.80B4.8216. [DOI] [PubMed] [Google Scholar]
- 8.Kramer MJ, Tanner BJ, Horval AE, O’Donnell RJ. Compressive osseointegration promotes viable bone at the endoprosthetic interface: retrieval study of Compress® implants. Int Orthop. 2008;32:567–571. doi: 10.1007/s00264-007-0392-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Monument MJ, Bernthal NM, Bowles AJ, Jones KB, Randall RL. What are the 5-year survivorship outcomes of compressive endoprosthetic osseointegration fixation of the femur? Clin Orthop Relat Res. 2015;473:883–890. doi: 10.1007/s11999-014-3724-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Sluga M, Windhager R, Lang S, Heinzl H, Bielack S, Kotz R. Local and systemic control after ablative and limb sparing surgery in patients with osteosarcoma. Clin Orthop Relat Res. 1999;358:120–127. doi: 10.1097/00003086-199901000-00015. [DOI] [PubMed] [Google Scholar]