Abstract
Objectives
To overcome the challenges of B0 and RF excitation inhomogeneity at ultra-high field MRI, a workflow for volumetric B0 and flip-angle homogenisation was implemented on a human 9.4 T scanner.
Materials and methods
Imaging was performed with a 9.4 T human MR scanner (Siemens Medical Solutions, Erlangen, Germany) using a 16-channel parallel transmission system. B0- and B1-mapping were done using a dual-echo GRE and transmit phase-encoded DREAM, respectively. B0 shims and a small-tip-angle-approximation kT-points pulse were calculated with an off-line routine and applied to acquire T1- and T*2-weighted images with MPRAGE and 3D EPI, respectively.
Results
Over six in vivo acquisitions, the B0-distribution in a region-of-interest defined by a brain mask was reduced down to a full-width-half-maximum of 0.10 ± 0.01 ppm (39 ± 2 Hz). Utilising the kT-points pulses, the normalised RMSE of the excitation was decreased from CP-mode’s 30.5 ± 0.9 to 9.2 ± 0.7 % with all B+1 voids eliminated. The SNR inhomogeneities and contrast variations in the T1- and T*2-weighted volumetric images were greatly reduced which led to successful tissue segmentation of the T1-weighted image.
Conclusion
A 15-minute B0- and flip-angle homogenisation workflow, including the B0- and B1-map acquisitions, was successfully implemented and enabled us to reduce intensity and contrast variations as well as echo-planar image distortions in 9.4 T images.
Keywords: Ultra-high field MR, Parallel transmission, B0 shimming, Flip-angle homogenisation, MPRAGE, 3D EPI
Introduction
The increased signal-to-noise ratio (SNR) [1–3] and contrast-to-noise ratio (CNR) [4] of ultra-high field (UHF) MRI provide a platform for imaging at higher resolution [5], allowing both structural [6, 7], and functional [8–10] details to be observed down to cortical layers. The increased field-induced susceptibility contrast also opens up new opportunities in T*2- and susceptibility weighted imaging (SWI) as well as for quantitative susceptibility mapping (QSM) [11–14]. Novel contrast mechanisms also come into play, such as the anisotropic susceptibility of white matter [15–17].
However, in order to fully take advantage of the increased SNR and CNR in UHF MRI, especially for whole brain and body imaging, one must overcome the RF inhomogeneity at UHF. At field strengths of 3 T and beyond, the RF wavelength in vivo becomes comparable to, or smaller than, the dimension of the imaging object [18–25]. This leads to interferences in the transmitted RF (B+1) field and results in strong intensity and contrast variations in the final image [24]. The severity of the RF inhomogeneity increases as the RF wavelength reduces with field strength, e.g., from 7 T to 9.4 T [25].
These challenges can be addressed in several ways: RF coil [18], dielectric pads [26, 27], as well as RF pulse design [28] and parallel transmission with B1 shimming [25, 29–31] or transmit sensitivity encoding (Transmit-SENSE) [32–34] have been proposed as possible solutions to the problem of RF inhomogeneity at UHF. Small-tip-angle (STA) pulses can, for example, be designed using the spatial domain method [32] with magnitude least square optimisation [35] in combination with k-space trajectories such as kT-points [36] or spiral nonselective (SPINS) [37]. Both kT-points and SPINS pulses have been shown to be effective in ameliorating RF inhomogeneities and have been used in Magnetisation Prepared Rapid Acquisition Gradient Echo (MPRAGE) [38] for T1-weighted anatomical imaging [39, 40].
Benefits and challenges of UHF both include the increased effects of magnetic susceptibility and the change in relaxation times that improve various contrasts, but the downside is stronger image distortion [41], blurring and signal losses [42] due to B0-inhomogeneity. B0-inhomogeneities at UHF present a significant challenge especially in the context of echo planar imaging (EPI) [43] which is the sequence of choice in functional MRI and other rapid imaging applications. Furthermore, spatial variation in B0 influences the performance of the RF excitation. B0 variation can be alleviated with higher-order shim coils such as the 3rd and 4th orders, which have been shown to provide significant improvements in both global and local B0 homogeneity [44].
Here, we demonstrate a B0- and flip-angle-homogenisation workflow as implemented on a human 9.4 T scanner, to address the aforementioned challenges of UHF MRI. The B0-shimming workflow uses a field-map-based shimming technique that homogenises the static B0-field in a region defined by an interactively generated brain mask down to 0.12 ppm (48.0 Hz), utilising shim coils up to (partial) 3rd order. The flip-angle-homogenisation workflow takes in a B0 map from the shimmed field and channel-by-channel complex B+1 maps, both measured as part of the calibration routine. Parallel RF pulses are then calculated using STA approximation, which can then be applied on the scanner with modified sequences. The workflow is presented by the example of volumetric kT-points excitation pulses. Sample T1-weighted MPRAGE [38] and T*2-weighted 3D EPI [45] images collected in conjunction with the B0- and flip-angle-homogenisation routine are shown.
Materials and methods
All experiments were performed on a 9.4 T human MR scanner (Siemens Medical Solutions, Erlangen, Germany) using a head gradient set (AC84-mk2, maximum amplitude 80 mT/m, maximum slew rate 333 T/m/s, inner diameter 36 cm) in combination with a 16-channel parallel transmission system (1 kW per channel) and a dual-row 16-channel transmit/31-channel receive array coil [46]. Online local SAR monitoring was achieved by a vendor-provided system installed on the transmit chain [47]. The SAR matrices used for the online local SAR monitoring were derived from an EM simulation [30] on an adult male model (Hugo) with safety margins (a factor of 2) and limits set according to [48] and compressed according to the virtual observation points (VOPs) method [49]. A common model was used for all subjects. The same VOPs compressed SAR matrices were used in SAR prediction in the off-line pulse calculations and simulations. All in vivo experiments were approved by the local ethics committee and performed in accordance with internal safety guidelines and all participating volunteers gave written informed consent.
B0- and flip-angle-homogenisation workflow
The volumetric B0- and flip-angle-homogenisation workflow consists of calibration scans, B0-shim and RF calculations, and optional extra B0 and pTx RF pulse flip angle mappings for validation. The workflow is summarised in Fig. 1. All off-line calculations are carried out in MATLAB (MathWorks, Natick, MA US).
Calibration scans
B0 field maps were obtained from a dual-echo 3D GRE sequence (TR = 30 ms, TE1 = 1.00 ms, TE2 = 3.21 ms, nominal flip-angle = 8 degrees, nominal voxel size = 4 mm isotropic, matrix size 50 × 50 × 44, bandwidth 1560 Hz/pixel, total scan duration 1:49 min). B0 shimming including some 3rd order terms (Z3, Z2X, Z2Y, ZX2Y2) was accomplished with a custom software [50]. The B0 shim routine unwraps the measured fieldmap [51] and fits spherical harmonics basis functions up to all 4th orders to it. Including the extra harmonics terms, it allows us to take some of the deviation of the physical shim coils from the spherical harmonics basis functions into account. The spherical harmonics terms for the shim coils had previously been determined by one-off calibration measurement, and were used to calculate the required shim currents to achieve field optimisation in the target region typically by a least-square fit to the in-session measured fieldmap. In order to take the system’s shim current limits into account, a constrained optimisation routine provided by CVX [52, 53] was used in the shim current calculation. The target region for B0 shimming was defined by a brain mask generated from the magnitude images of the dual-echo GRE using FSL-BET [54]. A second B0 field map was acquired after applying the shim values in order to take the new B0 field distribution into account in the RF pulse calculation. This step is optional as the B0 shim routine can also generate a predicted B0 map after shimming for later pulse calculation.
Complex B1 maps from all the transmit channels were obtained using a transmit phase-encoded [55], T2 and T*2 compensated version of DREAM [56] (imaging train repetition time = 6.8 ms, TR = 7.5 s, TE1 = 2.22 ms, TE2 = 4.44 ms, nominal imaging flip-angle = 7 degrees, nominal preparation pulse FA = 55.5 degrees, imaging slice thickness 4 mm, slice separation 10 mm, preparation pulse slice thickness 8 mm, voxel size 4 mm isotropic, matrix size 64 × 56 × 15, bandwidth 690 Hz/pixel, 32 transmit phase encode steps, total scan duration 4:00 min). After the pulse calculation (see below), the flip angle of the RF pulse was mapped by an implementation of the pre-saturation turbo-flash (PreSat-TFL) sequence [57] to verify it against the prediction of the optimisation algorithm. The PreSat-TFL method parameters were: imaging train repetition time 5.9 ms, TR 10 s, TE 2.24 ms, nominal imaging FA 8 degrees, nominal preparation pulse FA 45 degrees, imaging slice thickness 4 mm, voxel size 4 mm isotropic, matrix size = 64 × 64 × 1, bandwidth 690 Hz/pixel, total scan duration 20 s.
RF pulse calculation
The parallel transmit excitation pulses utilised in the imaging sequences were designed using the spatial domain method [32] with magnitude least square optimisation [35] under the small tip-angle approximation. The optimisation was done using a conjugate-gradients-based algorithm [58] which includes a global SAR regularisation and a local SAR regularisation by means of a VOP-compressed SAR matrix [59]. Localised B+1 drop-out in the optimisation solution was avoided by using a region growing algorithm [60]. kT-points [36] were used as the k-space trajectory for the excitation pulses.
The kT-points trajectory consists of six equidistant points placed on the kx, ky and kz axes at ±6.33 m−1, which is roughly the inverse of the RF wavelength at 9.4 T [61], plus two additional points at the k-space centre at the beginning and the end of the trajectory. This trajectory equates to eight rectangular sub-pulses of duration 210 us interleaved with 70 us gradient blips. The RF waveform was divided into rectangular sub-pulses and played out between the gradient blips. The overall pulse duration was 2.24 ms. The trajectory, gradient blips and a RF pulse are shown in Fig. 2.
Imaging scans
MPRAGE scans were acquired using: TR = 3.75 s, TE = 3.64 ms, TI = 1.2 s, matrix size 384 × 384 × 256, voxel size 0.6 mm isotropic, sagittal slices and anterior-to-posterior phase encode direction, slice partial Fourier 7/8, echo train length = 2.1 s, nominal flip-angle = 5 degrees, GRAPPA parallel imaging using factor 3 along the in-plane phase encoding direction with 24 reference lines, bandwidth 180 Hz/pixel; total scan duration was 8:58 min. The T1-weighting in MPRAGE was achieved by an adiabatic inversion pulse. Here, the same complex RF waveform was applied in all transmit channels, i.e., only the channel relative phases were adjusted to achieve a CP-like mode of the coil. The nominal flip-angle of the CP-mode pulse was defined as the average flip angle in a central region around the bright CP-mode spot in the middle of the brain. The inversion pulse used in the experiment was a 13 ms TR-FOCI adiabatic pulse with a peak voltage of 79.8 V applied to each transmit channel (yielding a nominal peak B1 amplitude of 10 uT) [62]. An additional proton-density-weighted image was acquired for receive bias-field correction by switching off the inversion pulse and minimising TE and TR while keeping the other parameters unchanged. For one-off comparison, the same acquisitions were repeated using a standard non-selective RF excitation pulse of 0.10 ms duration in the same CP-like mode as used for the TR-FOCI inversion pulse. For this set of MPRAGE sequence parameters, the maximum 10 g averaged local SAR per TR estimated by using the VOPs compressed SAR matrices were 0.77 and 1.11 W/kg for the kT-points and the CP-like mode excitation pulses, respectively, and was 6.77 W/kg for the TR-FOCI inversion pulse.
The T1-weighted MPRAGE image was segmented into grey matter, white matter and cerebrospinal fluid maps with FSL-FAST [63] in a sequence of customised steps. First, the T1-weighted image was realigned rigidly to and divided by the proton-density image to remove any intensity bias introduced by receive field inhomogeneity [64]. Second, the co-registered proton-density image was corrected for bias fields with FSL-FAST and then passed to FSL-BET to create a brain mask. Third, the brain mask was then used to extract the brain region from the bias-corrected T1-weighted image. Finally, the brain extracted T1-weighted image was passed to FSL-FAST for tissue segmentation.
The same kT-points excitation was also applied in a 3D EPI sequence [45] with parameters as follows: TR = 61 ms, effective volume TR = 12.7 s, TE = 22 ms, matrix size 256 × 256 × 208, voxel size 0.75 mm isotropic, phase partial Fourier 6/8, nominal flip-angle 15 degrees, bandwidth 1396 Hz/pixel, and parallel imaging undersampling with GRAPPA factor 3 along the in-plane phase-encoding direction using 96/48 reference lines/partitions that were acquired in a segmented lines-in-partition order [65]. The total acquisition time was 1:28 min. This acquisition was also repeated in a CP-like mode transmission using a 1.0 ms duration rectangular pulse. With the given 3D EPI sequence parameters, the maximum 10 g averaged local SAR per TR were estimated to be 1.56 and 1.12 W/kg for the kT-points and the CP-like mode excitation pulses, respectively. The 3D EPI sequence was implemented such that each excitation yielded a kx-ky plane at a different kz increment, using linear ordering. Each readout was preceded by a three-lobed non-phase-encoded navigator readout at minimum TE, which served the dual purpose of EPI eddy current (Nyquist N/2 ghost) correction [66] and correction of B0-drift due to subject breathing during the acquisition [67].
Results
Figure 3 shows the B0 shimming results from an example set of in vivo measurement using the shim routine described above. Using only 1st and 2nd order shim coils, shimming reduces the standard deviation of the B0 field within the brain mask to 50.1 Hz with a 90th percentile range of 148.0 Hz; when adding the four available 3rd shim coils (Z3, Z2X, Z2Y, ZX2Y2), the standard deviation and the 90th percentile range were reduced to 46.1 and 138.6 Hz, respectively.
Illustrations of the flip-angle homogenisation from the same measurement session are shown in Fig. 4. The top row shows the B+1 magnitude distribution of the CP-like-mode, which has a normalised RMSE of 33.3 %. The middle row of Fig. 4 is the flip-angle prediction of the MLS optimised 8-point kT-points pulse, which has a normalised RMSE of 9.1 %, and the bottom row is the pulse’s corresponding single-slice flip-angle distribution measured by the PreSat-TFL in sagittal orientation. In both B0 and flip-angle homogenisation, the measured B0 and the flip angle distribution visually agrees very well with their predictions.
The top row of Fig. 5 shows the post hoc Bloch simulations of the TR-FOCI adiabatic pulse and the vendor’s standard 5.12 ms duration hyperbolic-secant adiabatic pulse [68]. They were simulated using the subject-specific B0 and B1 maps collected in the same session. The pulse amplitude of the TR-FOCI and the hyperbolic secant inversion pulses were 79.8 and 166.0 V, respectively. The lower adiabatic threshold of the TR-FOCI allowed it to achieve at least partial magnetisation inversion in regions with low field intensities in the CP-mode pattern. In contrast, large regions of brain, especially near the temporal lobes, were not inverted by the hyperbolic secant adiabatic pulse. The normalised RMSE of the TR-FOCI and hyperbolic secant inversions were 10.6 and 26.5 %, respectively. Given this superior inversion performance of the TR-FOCI pulse it was used in all the following MPRAGE measurements.
The B0 shim and RF pulses performances over a set of 6 in vivo measurements are summarised Fig. 6. On average, using only 1st and 2nd orders shim coils, the standard deviation and the 90th percentile range of the B0 distribution after shim were 45.5 ± 3.2 and 141.7 ± 12.0 Hz, respectively. Including the 3rd order coils reduced them to 38.8 ± 2.1 and 117.9 ± 6.1 Hz, respectively. The average normalised RMSE of the excitation pulses were 30.5 ± 0.9 % and 9.2 ± 0.7 %, respectively, for CP-mode and kT-points; and they were 25.6 ± 0.9 and 10.3 ± 0.6 %, respectively, for hyperbolic secant and TR-FOCI inversion pulses.
Figure 7a and b show the T1-weighted MPRAGE images using TR-FOCI inversion with CP-mode (left column) and kT-points (right column) excitation pulses for the imaging trains. The T1-weighted and the proton density images were co-registered and divided to remove the receive bias fields. The brain-extracted and receive-bias-free images are displayed in Fig. 7c). These images were segmented by FSL-FAST into grey matter, white matter and cerebrospinal fluid maps which are shown in Fig. 7d, e and f, respectively.
B+1 voids, SNR and contrast variations due to B+1 inhomogeneity can be seen in the CP-mode T1-weighted and proton density images. Consequently, in several regions proper segmentation of grey matter and CSF was not possible.
Results of the 3D EPI scan are shown in Fig. 8. The top and bottom rows are 3D EPI images obtained using CP-mode and kT-points excitations, respectively. Similar to the MPRAGE images, signal and contrast variations can be seen in the CP-mode image.
Discussion
B0- and B1-inhomogeneities are the well-known challenges in UHF MRI. Here, we presented the workflow used at our institution to homogenise both B0 and the excitation flip-angle, allowing us to acquire high quality volumetric images at 9.4 T and further explore the potential of UHF imaging. The entire B0 and B1 calibration workflow currently takes approximately 15 min in total, including the (~6 min) B0 and B1 maps acquisitions. With further developments such as incorporating the B0 and B1 mapping into the scanner’s on-line image reconstructions and further streamlining the data transfer between the scanner’s host computer and the PC on which the B0 and B1 shim calculations are performed, we expect that the total calibration time can be shortened to approximately 8 min. This is only a small proportion of time typically allocated for UHF MRI studies.
The first benefit that the homogenisation workflow can bring to single- and multi-channel transmission systems alike is the potential reduction in signal loss, T*2 blurring and distortion due to B0 inhomogeneity, especially for the case of EPI [69]. By using a ‘brain-only’ region of interest and precisely calibrated shim coils, the B0 linewidth in the brain was brought down to 39 ± 2 Hz (0.10 ± 0.01 ppm). Using the four additional 3rd order shim coils, the linewidth and the 90th percentile of the B0 field distribution in the brain were on average further reduced by 14.7 and 16.9 %, respectively. This improvement is most apparent in the frontal regions of the brain as shown in Fig. 3d, e.
The second benefit of the homogenisation workflow is the reduction in signal and contrast variations across the image due to B+1 inhomogeneity which is a bigger challenge than B0 homogenisation at 9.4 T. A previous simulation study on the same coil at 9.4 T has shown the limited capability of flip-angle homogenisation using static shim only [30]. By utilising the full 16-channel parallel RF transmission system dynamically, we were able to design kT-points pulse in the same brain region of interest to reduce the normalised RMSE of the excitation flip angle on average by 69.7 %, in comparison to CP-mode. All B+1 voids could be eliminated, as can be observed in Fig. 4. The PreSat-TFL validation scans confirmed the agreement between the predicted and achieved flip angle distribution, and implicitly also the correct delivery of the pulse (e.g., gradient axis orientation, channel ordering, and phase sign).
The results over 6 scan sessions demonstrated that the improvements on both the B0 and flip-angle homogenieties are consistent and reproducible. Various parts of the B0 and flip-angle homogenisation procedure work in synergy with each other. Apart from reducing artefacts, a homogenised and known B0 distribution can also ease the RF optimisation by narrowing the range of off-resonance in the problem. By excluding the subcutaneous tissue and the neck with the brain mask, the size of the RF optimisation problem is reduced and hence improves its speed and memory usage.
RF exposure to volunteers is a major concern in UHF MR, especially in combination with the use of parallel transmission. Using EM-simulations generated SAR matrices and VOPs compression, local SAR can be regularised during the pulse design process [59]. With the knowledge of the desired sequence parameters, the maximum local SAR can also be predicted in the pulse design workflow for a particular run of sequence [70]. The off-line maximum local SAR look-ahead allows the scanner operators to adjust the sequence parameters in order to keep the 10 s and 6 min exposure below the 20 W/kg and the 10 W/kg limits, respectively, as described in IEC 60601-2-33.
Several methods for B0 and B1 mapping have been shown in the literature for B0 shim and RF pulse optimisations. It is out of the scope of this article to review or evaluate them, and the choices here were made based on pragmatic criteria: ease of implementation and effectiveness for the experimental setting (whole brain B1 and B0 shim). Shim current calculation for B0 shims can, in principle, be achieved with a least-square fit expressed as a matrix inversion. However, the physical shim current limits need to be taken into account, for which we have chosen a constrained optimisation approach instead. For B0 mapping, we have found that a standard dual-echo 3D GRE sequence, with phase unwrapping as shown in Ref [51], was fully sufficient for application in the brain, without apparent artefacts. For the RF optimisation, the main goal was in homogenising a low flip angle excitation pulse and we have hence chosen the STA approximation approach with a magnitude least square cost, due to its ease of obtaining a magnitude homogenised solution with a potential reduction in global RF power [35].
Two B1 mapping sequences were use in this study, DREAM and PreSat-TFL. They were chosen for a couple of different aspects of B1 mapping based on their different desired properties. The individual transmit channel B1 maps were acquired with transmit phase encoded DREAM because of its speed, its insensitivity to B0 variation and more robust against motion in comparison to PreSat-TFL. However, due to the RF pulse timing and layout in DREAM, it is not straightforward to map the kT-points pulses with it. Hence, a 2D PreSat-TFL, which is more flexible in terms of the mapping RF pulse, was used as the optional fast (20 s) qualitative validation of the designed pTx pulse. A full 3D quantitative mapping of the kT-points pulse is also possible by using an AFI [71] sequence, but this is not included in the current workflow because of its long acquisition time (~3 min).
The B0 and B1 maps collected during the session were also used in a Bloch simulation to decide which type of adiabatic pulse and what voltage amplitude was suitable for inversion in our imaging experiments. Our simulations have shown that the TR-FOCI inversion pulse performed better than the hyperbolic secant pulse provided in the standard product sequence, despite only using half the maximum transmitter voltage. Hence, TR-FOCI was chosen as the inversion pulse in our T1-weighted MPRAGE for structural imaging and tissue segmentation. Despite its low adiabatic threshold, the TR-FOCI pulse in combination with a CP-like mode in this particular setup still could not provide a complete inversion throughout the region of interest defined by the brain mask. Using a large tip angle design as reported by [39, 72], the inversion efficiency can potentially be further improved. The feasibility of incorporating this approach into our B0- and flip-angle-homogenisation is being investigated.
The efficient inversion of TR-FOCI in combination with the homogenised excitation through a kT-points pulse have led to significant improvement in the quality of the T1-weighted MRRAGE images acquired at 9.4 T. As can be seen in the comparison of CP-mode and kT-points acquisitions of the T1- and proton density weighted images shown in Fig. 7, the variations in intensity and contrast across the image were largely eliminated by substituting the standard rectangular pulses by parallel transmit kT-points excitations. Successful segmentation of the receive-bias-corrected images was achieved without further adjustment or manual intervention; this was not the case for the CP-mode acquired image.
Similar improvements were observed for the 3D EPI when comparing the CP-mode and kT-points acquired images. Similar improvements were observed for the 3D EPI when comparing the CP-mode and kT-points acquired images. While a 0.1 ms rectangular pulse is typically used in an MPRAGE readout train, a 1 ms pulse was used for the 3D EPI CP mode acquisition, because of the higher required flip angle and the lack of a time constraint. This “stretching” of the pulse proportionately reduced the required pulse amplitude, which explains the inversion of the local SAR estimates between the CP-mode and kT-points protocols for MPRAGE and 3D EPI listed in the Materials and Methods section. The increased pulse duration reduces the pulse bandwidth, which however remained much higher than the B0 variations within the brain mask. Therefore most of the excitation inhomogeneity can be attributed to B1 inhomogeneity from the transmit coils at this particular field strength.
Application of 3D EPI at 3 T and above has recently been receiving increased attention not only for BOLD imaging [45, 73, 74] but has also been shown to be an attractive choice for rapid acquisition of T*2-weighted structural data [75]. This can readily be extended to the application of SWI and QSM, as has recently been demonstrated at 3 T [76], 7 T and 9.4 T [77]. The benefits of 3D EPI are greatest at high-resolution which is particularly beneficial at UHF. A tremendous advantage of 3D sequences with two phase encoding directions is that they lend themselves to efficient parallel imaging undersampling strategies such as CAIPIRINHA [78] as has been shown also in the context of 3D EPI [79, 80].
The B0 shim and RF pulse calculated by the B0- and flip-angle-homogenisation workflow can be applied to other 3D scans such as GRE. The volumetric workflow has been extended to design slice-specific spokes pulses for 2D excitations as demonstrated in [81] for high in-plane resolution GRE imaging.
Conclusion
In summary, we demonstrated a B0- and flip-angle-homogenisation procedure that currently takes approximately 15 min to run, and homogenises B0 linewidth down to 0.10 ± 0.01 ppm and flip-angle normalised RMSE down to 9.2 ± 0.7 % in the same whole brain region of interest. This enables us to reduce intensity and contrast variations as well as distortions in our 9.4 T images.
Acknowledgments
The authors thank Dr. Michael S. Poole for his help with the B0 and flip-angle optimisation codes. Scan time was funded under Scannexus/BrainsUnlimited Project D0112, intramural Grants MBIC F0006B03 and FHML M0108A06, and Dutch Science Foundation NWO Grant VIDI 452-11-00 to Kâmil Uludağ.
Compliance with ethical standards
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
References
- 1.Edelstein WA, Glover GH, Hardy CJ, Redington RW. The intrinsic signal-to-noise ratio in NMR imaging. Magn Reson Med. 1986;3(4):604–618. doi: 10.1002/mrm.1910030413. [DOI] [PubMed] [Google Scholar]
- 2.Hoult DI, Richards RE. The signal-to-noise ratio of the nuclear magnetic resonance experiment. J Magn Reson. 1976;24(1):71–85. doi: 10.1016/j.jmr.2011.09.018. [DOI] [PubMed] [Google Scholar]
- 3.Pohmann R, Speck O, Scheffler K (2015) Signal-to-noise ratio and MR tissue parameters in human brain imaging at 3, 7, and 9.4 tesla using current receive coil arrays. Magn Reson Med doi:10.1002/mrm.25677 [DOI] [PubMed]
- 4.Duyn JH. The future of ultra-high field MRI and fMRI for study of the human brain. Neuroimage. 2012;62(2):1241–1248. doi: 10.1016/j.neuroimage.2011.10.065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Budde J, Shajan G, Scheffler K, Pohmann R. Ultra-high resolution imaging of the human brain using acquisition-weighted imaging at 9.4T. Neuroimage. 2014;86:592–598. doi: 10.1016/j.neuroimage.2013.08.013. [DOI] [PubMed] [Google Scholar]
- 6.Duyn JH, van Gelderen P, Li TQ, de Zwart JA, Koretsky AP, Fukunaga M. High-field MRI of brain cortical substructure based on signal phase. Proc Natl Acad Sci USA. 2007;104(28):11796–11801. doi: 10.1073/pnas.0610821104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Marques JP, van der Zwaag W, Granziera C, Krueger G, Gruetter R. Cerebellar cortical layers: in vivo visualization with structural high-field-strength MR imaging. Radiology. 2010;254(3):942–948. doi: 10.1148/radiol.09091136. [DOI] [PubMed] [Google Scholar]
- 8.Budde J, Shajan G, Zaitsev M, Scheffler K, Pohmann R. Functional MRI in human subjects with gradient-echo and spin-echo EPI at 9.4 T. Magn Reson Med. 2014;71(1):209–218. doi: 10.1002/mrm.24656. [DOI] [PubMed] [Google Scholar]
- 9.van der Zwaag W, Francis S, Head K, Peters A, Gowland P, Morris P, Bowtell R. fMRI at 1.5, 3 and 7 T: characterising BOLD signal changes. Neuroimage. 2009;47(4):1425–1434. doi: 10.1016/j.neuroimage.2009.05.015. [DOI] [PubMed] [Google Scholar]
- 10.Yacoub E, Shmuel A, Pfeuffer J, Van De Moortele PF, Adriany G, Andersen P, Vaughan JT, Merkle H, Ugurbil K, Hu X. Imaging brain function in humans at 7 Tesla. Magn Reson Med. 2001;45(4):588–594. doi: 10.1002/mrm.1080. [DOI] [PubMed] [Google Scholar]
- 11.Bilgic B, Fan AP, Polimeni JR, Cauley SF, Bianciardi M, Adalsteinsson E, Wald LL, Setsompop K. Fast quantitative susceptibility mapping with L1-regularization and automatic parameter selection. Magn Reson Med. 2014;72(5):1444–1459. doi: 10.1002/mrm.25029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Budde J, Shajan G, Hoffmann J, Ugurbil K, Pohmann R. Human imaging at 9.4 T using T(2) *-, phase-, and susceptibility-weighted contrast. Magn Reson Med. 2011;65(2):544–550. doi: 10.1002/mrm.22632. [DOI] [PubMed] [Google Scholar]
- 13.Shmueli K, de Zwart JA, van Gelderen P, Li TQ, Dodd SJ, Duyn JH. Magnetic susceptibility mapping of brain tissue in vivo using MRI phase data. Magn Reson Med. 2009;62(6):1510–1522. doi: 10.1002/mrm.22135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Wharton S, Bowtell R. Whole-brain susceptibility mapping at high field: a comparison of multiple- and single-orientation methods. Neuroimage. 2010;53(2):515–525. doi: 10.1016/j.neuroimage.2010.06.070. [DOI] [PubMed] [Google Scholar]
- 15.Denk C, Hernandez Torres E, MacKay A, Rauscher A. The influence of white matter fibre orientation on MR signal phase and decay. NMR Biomed. 2011;24(3):246–252. doi: 10.1002/nbm.1581. [DOI] [PubMed] [Google Scholar]
- 16.van Gelderen P, Mandelkow H, de Zwart JA, Duyn JH. A torque balance measurement of anisotropy of the magnetic susceptibility in white matter. Magn Reson Med. 2015;74(5):1388–1396. doi: 10.1002/mrm.25524. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Wiggins CJ, Gudmundsdottir V, Le Bihan D, V L MC (2015) Orientation dependence of white matter T2* contrast at 7T: a direct demonstration. In: Proceedings of the 23rd scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto
- 18.Adriany G, Van de Moortele P-F, Ritter J, Moeller S, Auerbach EJ, Akgün C, Snyder CJ, Vaughan T, Ugurbil K. A geometrically adjustable 16-channel transmit/receive transmission line array for improved RF efficiency and parallel imaging performance at 7 Tesla. Magn Reson Med. 2008;59(3):590–597. doi: 10.1002/mrm.21488. [DOI] [PubMed] [Google Scholar]
- 19.Barfuss H, Fischer H, Hentschel D, Ladebeck R, Oppelt A, Wittig R, Duerr W, Oppelt R. In vivo magnetic resonance imaging and spectroscopy of humans with a 4 T whole-body magnet. NMR Biomed. 1990;3(1):31–45. doi: 10.1002/nbm.1940030106. [DOI] [PubMed] [Google Scholar]
- 20.Bomsdorf H, Helzel T, Kunz D, Röschmann P, Tschendel O, Wieland J. Spectroscopy and imaging with a 4 tesla whole-body mr system. NMR Biomed. 1988;1(3):151–158. doi: 10.1002/nbm.1940010308. [DOI] [PubMed] [Google Scholar]
- 21.Bottomley PA, Andrew ER. RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imaging. Phys Med Biol. 1978;23(4):630. doi: 10.1088/0031-9155/23/4/006. [DOI] [PubMed] [Google Scholar]
- 22.Glover GH, Hayes CE, Pelc NJ, Edelstein WA, Mueller OM, Hart HR, Hardy CJ. O?Donnell M, Barber WD (1985) Comparison of linear and circular polarization for magnetic resonance imaging. J Magn Reson. 1969;64(2):255–270. [Google Scholar]
- 23.Keltner JR, Carlson JW, Roos MS, Wong STS, Wong TL, Budinger TF. Electromagnetic fields of surface coil in vivo NMR at high frequencies. Magn Reson Med. 1991;22(2):467–480. doi: 10.1002/mrm.1910220254. [DOI] [PubMed] [Google Scholar]
- 24.Van de Moortele P-F, Akgun C, Adriany G, Moeller S, Ritter J, Collins CM, Smith MB, Vaughan JT, Ugurbil K. B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil. Magn Reson Med. 2005;54(6):1503–1518. doi: 10.1002/mrm.20708. [DOI] [PubMed] [Google Scholar]
- 25.Vaughan T, DelaBarre L, Snyder C, Tian J, Akgun C, Shrivastava D, Liu W, Olson C, Adriany G, Strupp J, Andersen P, Gopinath A, van de Moortele P-F, Garwood M, Ugurbil K. 9.4T human MRI: preliminary results. Magn Reson Med. 2006;56(6):1274–1282. doi: 10.1002/mrm.21073. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.O’Brien KR, Magill AW, Delacoste J, Marques JP, Kober T, Fautz HP, Lazeyras F, Krueger G. Dielectric pads and low- B1+ adiabatic pulses: complementary techniques to optimize structural T1 w whole-brain MP2RAGE scans at 7 tesla. J Magn Reson Imaging. 2014;40(4):804–812. doi: 10.1002/jmri.24435. [DOI] [PubMed] [Google Scholar]
- 27.Teeuwisse WM, Brink WM, Webb AG. Quantitative assessment of the effects of high-permittivity pads in 7 Tesla MRI of the brain. Magn Reson Med. 2012;67(5):1285–1293. doi: 10.1002/mrm.23108. [DOI] [PubMed] [Google Scholar]
- 28.Pauly J, Nishimura D. Macovski A (1989) A k-space analysis of small-tip-angle excitation. J Magn Reson. 1969;81(1):43–56. doi: 10.1016/j.jmr.2011.09.023. [DOI] [PubMed] [Google Scholar]
- 29.Curtis AT, Gilbert KM, Klassen LM, Gati JS, Menon RS. Slice-by-slice B1 + shimming at 7 T. Magn Reson Med. 2012;68(4):1109–1116. doi: 10.1002/mrm.23319. [DOI] [PubMed] [Google Scholar]
- 30.Hoffmann J, Shajan G, Scheffler K, Pohmann R (2014) Numerical and experimental evaluation of RF shimming in the human brain at 9.4 T using a dual-row transmit array. Magn Reson Mater Phy 27 (5):373–386 [DOI] [PubMed]
- 31.Mao W, Smith MB, Collins CM. Exploring the limits of RF shimming for high-field MRI of the human head. Magn Reson Med. 2006;56(4):918–922. doi: 10.1002/mrm.21013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Grissom W, Yip C-Y, Zhang Z, Stenger VA, Fessler JA, Noll DC. Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magn Reson Med. 2006;56(3):620–629. doi: 10.1002/mrm.20978. [DOI] [PubMed] [Google Scholar]
- 33.Katscher U, Börnert P, Leussler C, van den Brink JS. Transmit SENSE. Magn Reson Med. 2003;49(1):144–150. doi: 10.1002/mrm.10353. [DOI] [PubMed] [Google Scholar]
- 34.Zhu Y. Parallel excitation with an array of transmit coils. Magn Reson Med. 2004;51(4):775–784. doi: 10.1002/mrm.20011. [DOI] [PubMed] [Google Scholar]
- 35.Setsompop K, Wald LL, Alagappan V, Gagoski BA, Adalsteinsson E. Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channels. Magn Reson Med. 2008;59(4):908–915. doi: 10.1002/mrm.21513. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Cloos MA, Boulant N, Luong M, Ferrand G, Giacomini E, Le Bihan D, Amadon A. kT -points: short three-dimensional tailored RF pulses for flip-angle homogenization over an extended volume. Magn Reson Med. 2012;67(1):72–80. doi: 10.1002/mrm.22978. [DOI] [PubMed] [Google Scholar]
- 37.Malik SJ, Keihaninejad S, Hammers A, Hajnal JV. Tailored excitation in 3D with spiral nonselective (SPINS) RF pulses. Magn Reson Med. 2012;67(5):1303–1315. doi: 10.1002/mrm.23118. [DOI] [PubMed] [Google Scholar]
- 38.Mugler JP, 3rd, Brookeman JR. Rapid three-dimensional T1-weighted MR imaging with the MP-RAGE sequence. J Magn Reson Imaging. 1991;1(5):561–567. doi: 10.1002/jmri.1880010509. [DOI] [PubMed] [Google Scholar]
- 39.Cloos MA, Boulant N, Luong M, Ferrand G, Giacomini E, Hang MF, Wiggins CJ, Le Bihan D, Amadon A. Parallel-transmission-enabled magnetization-prepared rapid gradient-echo T1-weighted imaging of the human brain at 7 T. Neuroimage. 2012;62(3):2140–2150. doi: 10.1016/j.neuroimage.2012.05.068. [DOI] [PubMed] [Google Scholar]
- 40.Hans Hoogduin RMGMJVHPL . Proceedings of the 22nd scientific meeting. Milan: International Society for Magnetic Resonance in Medicine; 2014. Malik SJ Initial experience with SPIral Non Selective (SPINS) RF pulses for homogeneous excitation at 7T; p. 4914. [Google Scholar]
- 41.Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med. 1995;34(1):65–73. doi: 10.1002/mrm.1910340111. [DOI] [PubMed] [Google Scholar]
- 42.Frahm J, Merboldt KD, Hanicke W. The influence of the slice-selection gradient on functional MRI of human brain activation. J Magn Reson B. 1994;103(1):91–93. doi: 10.1006/jmrb.1994.1015. [DOI] [PubMed] [Google Scholar]
- 43.Speck O, Stadler J, Zaitsev M. High resolution single-shot EPI at 7T. Magn Reson Mater Phy. 2008;21(1–2):73–86. doi: 10.1007/s10334-007-0087-x. [DOI] [PubMed] [Google Scholar]
- 44.Pan JW, Lo K-M, Hetherington HP. Role of very high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 tesla. Magn Reson Med. 2012;68(4):1007–1017. doi: 10.1002/mrm.24122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Poser BA, Koopmans PJ, Witzel T, Wald LL, Barth M. Three dimensional echo-planar imaging at 7 Tesla. Neuroimage. 2010;51(1):261–266. doi: 10.1016/j.neuroimage.2010.01.108. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Shajan G, Kozlov M, Hoffmann J, Turner R, Scheffler K, Pohmann R. A 16-channel dual-row transmit array in combination with a 31-element receive array for human brain imaging at 9.4 T. Magn Reson Med. 2014;71(2):870–879. doi: 10.1002/mrm.24726. [DOI] [PubMed] [Google Scholar]
- 47.Gumbrecht R, Fontius U, Adolf H, Benner T, Schmitt F, Adalsteinsson E, Wald LL, H.-P. F (2013) Online Local SAR Supervision for Transmit Arrays at 7T In: Proceedings of the 21st scientific meeting, International Society for Magnetic Resonance in Medicine, Salt Lake City
- 48.Hoffmann J, Henning A, Giapitzakis IA, Scheffler K, Shajan G, Pohmann R, Avdievich NI. Safety testing and operational procedures for self-developed radiofrequency coils. NMR Biomed. 2015 doi: 10.1002/nbm.3290. [DOI] [PubMed] [Google Scholar]
- 49.Eichfelder G, Gebhardt M. Local specific absorption rate control for parallel transmission by virtual observation points. Magn Reson Med. 2011;66(5):1468–1476. doi: 10.1002/mrm.22927. [DOI] [PubMed] [Google Scholar]
- 50.Poole MS, Panchuelo RS, Panel R, Peters A. Bowtell R. ISMRM Scientific Workshop: Ultra-High Field Systems and Applications: 7T and Beyond: Progress. Pitfalls and Potential: Lake Louise; 2011. Benefits of Brain Segmentation and 3D Field Mapping in B0 Shimming at 7T; p. 34. [Google Scholar]
- 51.Cusack R, Papadakis N. New robust 3-D phase unwrapping algorithms: application to magnetic field mapping and undistorting echoplanar images. Neuroimage. 2002;16(3 Pt 1):754–764. doi: 10.1006/nimg.2002.1092. [DOI] [PubMed] [Google Scholar]
- 52.Grant M, Boyd S (2008) Graph implementations for nonsmooth convex programs. In: Blondel V, Boyd S, Kimura H (eds) Recent Advances in Learning and Control. Springer-Verlag Limited
- 53.Grant M, Boyd S (2014) CVX: Matlab Software for Disciplined Convex Programming, version 2.1
- 54.Smith SM. Fast robust automated brain extraction. Hum Brain Mapp. 2002;17(3):143–155. doi: 10.1002/hbm.10062. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Tse DHY, Poole MS, Magill AW, Felder J, Brenner D, Jon Shah N. Encoding methods for B1(+) mapping in parallel transmit systems at ultra high field. J Magn Reson. 2014;245:125–132. doi: 10.1016/j.jmr.2014.06.006. [DOI] [PubMed] [Google Scholar]
- 56.Nehrke K, Versluis MJ, Webb A, Bornert P. Volumetric B1 (+) mapping of the brain at 7T using DREAM. Magn Reson Med. 2014;71(1):246–256. doi: 10.1002/mrm.24667. [DOI] [PubMed] [Google Scholar]
- 57.Chung S, Kim D, Breton E, Axel L. Rapid B1 + mapping using a preconditioning RF pulse with TurboFLASH readout. Magn Reson Med. 2010;64(2):439–446. doi: 10.1002/mrm.22423. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Sbrizzi A, Hoogduin H, Lagendijk JJ, Luijten P, SGL G, van den Berg CAT. Time efficient design of multi dimensional RF pulses: application of a multi shift CGLS algorithm. Magn Reson Med. 2011;66(3):879–885. doi: 10.1002/mrm.22863. [DOI] [PubMed] [Google Scholar]
- 59.Sbrizzi A, Hoogduin H, Lagendijk JJ, Luijten P, Sleijpen GLG, van den Berg CAT. Fast design of local N-gram-specific absorption rate?optimized radiofrequency pulses for parallel transmit systems. Magn Reson Med. 2012;67(3):824–834. doi: 10.1002/mrm.23049. [DOI] [PubMed] [Google Scholar]
- 60.Poole MS, Tse DHY, Vahedipour K, Shah NJ (2014) A region growing algorithm for robust kt-points B1 + homogenisation at 9.4T. In: Proceedings of the 22nd scientific meeting, International Society for Magnetic Resonance in Medicine, Milan
- 61.Amadon A, Dupas L, Vignaud A. Proceedings of the 23rd scientific meeting. Toronto: International Society for Magnetic Resonance in Medicine; 2015. Boulant N Does the best distance beween 2 spokes match the inverse RF wavelength? p. 2388. [Google Scholar]
- 62.Hurley AC, Al-Radaideh A, Bai L, Aickelin U, Coxon R, Glover P, Gowland PA. Tailored RF pulse for magnetization inversion at ultrahigh field. Magn Reson Med. 2010;63(1):51–58. doi: 10.1002/mrm.22167. [DOI] [PubMed] [Google Scholar]
- 63.Zhang Y, Brady M, Smith S. Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithm. IEEE Trans Med Imaging. 2001;20(1):45–57. doi: 10.1109/42.906424. [DOI] [PubMed] [Google Scholar]
- 64.Van de Moortele PF, Auerbach EJ, Olman C, Yacoub E, Ugurbil K, Moeller S. T1 weighted brain images at 7 Tesla unbiased for Proton Density, T2* contrast and RF coil receive B1 sensitivity with simultaneous vessel visualization. Neuroimage. 2009;46(2):432–446. doi: 10.1016/j.neuroimage.2009.02.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Ivanov D, Barth M, Uludag K. Proceedings of the 23rd scientific meeting. Toronto: International Society for Magnetic Resonance in Medicine; 2015. Poser BA Robust ACS acquisition for 3D echo planar imaging; p. 2059. [Google Scholar]
- 66.Jesmanowicz A, Wong E, J. H (1993) Phase correction for EPI using internal reference lines. In: Proceedings of the 12th scientific meeting, International Society for Magnetic Resonance in Medicine, Kyoto
- 67.Pfeuffer J, Van de Moortele PF, Ugurbil K, Hu X, Glover GH. Correction of physiologically induced global off-resonance effects in dynamic echo-planar and spiral functional imaging. Magn Reson Med. 2002;47(2):344–353. doi: 10.1002/mrm.10065. [DOI] [PubMed] [Google Scholar]
- 68.Silver MS, Joseph RI, Hoult DI. Selective spin inversion in nuclear magnetic resonance and coherent optics through an exact solution of the Bloch-Riccati equation. Phys Rev A. 1985;31(4):2753–2755. doi: 10.1103/PhysRevA.31.2753. [DOI] [PubMed] [Google Scholar]
- 69.Jezzard P. Correction of geometric distortion in fMRI data. Neuroimage. 2012;62(2):648–651. doi: 10.1016/j.neuroimage.2011.09.010. [DOI] [PubMed] [Google Scholar]
- 70.Graesslin I, Homann H, Biederer S, Bornert P, Nehrke K, Vernickel P, Mens G, Harvey P, Katscher U. A specific absorption rate prediction concept for parallel transmission MR. Magn Reson Med. 2012;68(5):1664–1674. doi: 10.1002/mrm.24138. [DOI] [PubMed] [Google Scholar]
- 71.Nehrke K. On the steady-state properties of actual flip angle imaging (AFI) Magn Reson Med. 2009;61(1):84–92. doi: 10.1002/mrm.21592. [DOI] [PubMed] [Google Scholar]
- 72.Cao Z, Donahue MJ, Ma J, Grissom WA. Joint design of large-tip-angle parallel RF pulses and blipped gradient trajectories. Magn Reson Med. 2015 doi: 10.1002/mrm.25739. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Lutti A, Thomas DL, Hutton C, Weiskopf N. High-resolution functional MRI at 3 T: 3D/2D echo-planar imaging with optimized physiological noise correction. Magn Reson Med. 2013;69(6):1657–1664. doi: 10.1002/mrm.24398. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.van der Zwaag W, Marques JP, Kober T, Glover G, Gruetter R, Krueger G. Temporal SNR characteristics in segmented 3D-EPI at 7T. Magn Reson Med. 2012;67(2):344–352. doi: 10.1002/mrm.23007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Zwanenburg JJ, Versluis MJ, Luijten PR, Petridou N. Fast high resolution whole brain T2* weighted imaging using echo planar imaging at 7T. Neuroimage. 2011;56(4):1902–1907. doi: 10.1016/j.neuroimage.2011.03.046. [DOI] [PubMed] [Google Scholar]
- 76.Langkammer C, Bredies K, Poser BA, Barth M, Reishofer G, Fan AP, Bilgic B, Fazekas F, Mainero C, Ropele S. Fast quantitative susceptibility mapping using 3D EPI and total generalized variation. Neuroimage. 2015;111:622–630. doi: 10.1016/j.neuroimage.2015.02.041. [DOI] [PubMed] [Google Scholar]
- 77.Poser BA, Ivanov D, Barth M, Uludag K (2015) High-resolution 3D EPI at 7 and 9.4T and its application to quantitative susceptibility mapping. In: Proceedings of 19th Annual Meeting, Organization for Human Brain Mapping, Honolulu
- 78.Breuer FA, Blaimer M, Mueller MF, Seiberlich N, Heidemann RM, Griswold MA, Jakob PM. Controlled aliasing in volumetric parallel imaging (2D CAIPIRINHA) Magn Reson Med. 2006;55(3):549–556. doi: 10.1002/mrm.20787. [DOI] [PubMed] [Google Scholar]
- 79.Poser BA, Kemper VG, Ivanov D, Kannengiesser SA, Uludag K. Proceedings of 30th Annual Scientific Meeting. Toulouse: European Society for Magnetic Resonance in Medicine and Biology; 2013. Barth M CAIPIRINHA-accelerated 3D EPI for high temporal and/or spatial resolution EPI acquisitions; p. 287. [Google Scholar]
- 80.Zahneisen B, Ernst T, Poser BA. SENSE and simultaneous multislice imaging. Magn Reson Med. 2015;74(5):1356–1362. doi: 10.1002/mrm.25519. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Tse DHY, Brenner D, Guerin B, Poser BA (2015) High resolution GRE at 9.4T using spokes pulses. In: Proceedings of the 23rd scientific meeting, International Society for Magnetic Resonance in Medicine, Toronto