Abstract
Background: The implantable loop recorder (ILR) is an electronic cardiac device indicated to aid diagnosis in patients with recurrent unexplained episodes of syncope or palpitations. At Leeds General Infirmary there are two surgical approaches for ILR implantation. The first utilises electrocardiographic mapping to obtain a position of optimal QRS signal characteristics; this is usually at the left sternal edge (LSE). The second approach focuses on cosmetic outcome, placing the ILR where pacemakers are typically fitted, at the left infra-clavicular (LIC) site. Patients referred for an ILR implant are assigned consecutively and randomly to the lists of operators applying either of the surgical approaches.
Diagnostic utility, forming a symptom-rhythm correlation leading to a diagnosis, is a clinically relevant way of evaluating these approaches. The aim of this project was to investigate whether the use of electrocardiographic mapping to determine an ILR implant position can influence diagnostic utility.
Methods: The current report concerns two studies. Initially, a retrospective study was carried out, involving 77 patients (46% male) implanted with an ILR during 2010. Data regarding ILR position at implantation and diagnostic outcomes of the patients were collected. Subsequently, we collected electrical and demographic data prospectively from 95 patients (46% male) with standard indications for ILR implantation: recurrent unexplained syncope, pre-syncope or palpitations, at implant and at the 6-week follow-up check. Statistical analysis was undertaken using SPSS statistical software. A p value < 0.05 was considered statistically significant.
Results: Groups in both studies were well matched for basic demographics and indications. Diagnostic utility did not differ according to implant site; the rate of diagnosis was 82% at LSE and 83% at LIC. The prospective study showed that although at implant, the QRS signal was significantly weaker at LIC (p <0.001); at the 6-week check the QRS signal became analogous between the groups (p=0.21) (Figure 1).
Conclusion: Electrocardiographic mapping does not improve QRS signals at follow-up and therefore does not affect diagnostic utility of loop recorders implanted for standard indications. Hence a pragmatic approach that considers cosmetic factors and the potential requirement for future device therapy is a reasonable way to determine position at the time of the initial implantation.