Abstract
Minimal invasive extracorporeal circulation (MiECC) systems have initiated important efforts within science and technology to further improve the biocompatibility of cardiopulmonary bypass components to minimize the adverse effects and improve end-organ protection. The Minimal invasive Extra-Corporeal Technologies international Society was founded to create an international forum for the exchange of ideas on clinical application and research of minimal invasive extracorporeal circulation technology. The present work is a consensus document developed to standardize the terminology and the definition of minimal invasive extracorporeal circulation technology as well as to provide recommendations for the clinical practice. The goal of this manuscript is to promote the use of MiECC systems into clinical practice as a multidisciplinary strategy involving cardiac surgeons, anaesthesiologists and perfusionists.
Keywords: Extracorporeal circulation, Minimally invasive extracorporeal circulation, Cardiopulmonary bypass, Modular systems, Systemic inflammation reaction syndrome, Complications
INTRODUCTION
Substantial experience has been accumulated with cardiac procedures performed using extracorporeal circulation (ECC) over the last decades. Several technological improvements have been realized, thus making cardiopulmonary bypass (CPB) the gold standard equipment for the majority of cardiac surgical procedures. This has contributed to improved perioperative and long-term results, despite an increasing prevalence of elderly and high-risk patients [1]. For the most frequent procedure, coronary artery bypass grafting (CABG), CPB provides optimal conditions (bloodless field and arrested heart) to allow the most complete myocardial revascularization and additionally offers for the possibility to perform other procedures such as valve repair or replacement and aortic surgery [2].
Major drawbacks of CPB are the adverse systemic effects triggered by a systemic inflammatory response syndrome (SIRS), which is mainly caused by the contact of blood with air and foreign surfaces [3, 4]. Trials have shown that the inflammatory response to CPB adversely influences clinical outcome [5, 6] although CPB cannot be considered as the main cause of postoperative morbidity.
Since the beginning of extracorporeal perfusion, the main inputs have been focused on one objective—to reduce the adverse effects of CPB. Perfusionists and bioengineers have developed optimized ‘CPB systems’ that combined the best features derived from perfusion science. The idea was to create a system that integrates all modifications into one combined set-up, known as the minimally invasive extracorporeal circulation (MiECC) system [7]. This concept has further initiated important new efforts to improve the biocompatibility of CPB components and minimize the side-effects.
Despite clinical advantages that have been reported in several papers [8], the penetration of MiECC technology into clinical practice remains extremely low. There is also significant heterogeneity between the various systems. Low implementation of MiECC may be due to the inability to predict which aspects of MiECC are beneficial, because several elements may act both interactively and/or independently, e.g. coated surfaces, closed systems, anticoagulation strategies, shed blood separation and reduced priming volumes.
The Minimal invasive Extra-Corporeal Technologies international Society (MiECTiS) was founded to create an international forum to exchange ideas on clinical practice and research in the field of minimally invasive extracorporeal circulation technology (www.miectis.org). The Society brings together, under a scientific interdisciplinary association, cardiac surgeons, anaesthesiologists, perfusionists and basic researchers.
The present work is a consensus document developed to standardize the terminology around MiECC technology and to provide recommendations for clinical practice. The authors have graded the levels of evidence and classified the findings listed below using the criteria recommended by the American Heart Association and the American College of Cardiology Task Force on Practice Guidelines (Table 1). The authors represent a multidisciplinary group to promote evidence-based perfusion practice to improve clinical outcomes.
Table 1:
Methodology and policy from the American College of Cardiology/American Heart Association Task Force on Practice Guidelines
| Classification of recommendations | Level of evidence |
|---|---|
| Class I: Conditions for which there is evidence, general agreement or both that a given procedure or treatment is useful and effective | Level A: Data derived from multiple randomized clinical trials or meta-analyses |
| Class II: Procedure treatment should be performed–administered | Level B: Data derived from a single randomized trial or non-randomized studies |
| Class IIA: Additional studies with focused objective needed | |
| Class IIB: Additional studies with broad objective needed; additional registry data would be helpful | Level C: Consensus opinion of experts |
| Class III: Procedure treatment should not be performed–administered because it is not helpful or might be harmful |
ACCF/AHA Task Force on Practice Guidelines. Methodology Manual and Policies from the ACCF/AHA Task Force on Practice Guidelines [9].
METHODS
The initiative to analyse the current practice was based on a questionnaire that was written by the Steering Committee of MiECTiS (Kyriakos Anastasiadis, Adrian Bauer, Thierry Carrel, Erich Gygax, John Murkin, Marco Ranucci, Jan Schaarschmidt). During an Expert Consensus Meeting, the statements were discussed and, subsequently, this consensus paper was developed. For each statement, the best available published evidence derived from meta-analyses of peer-reviewed literature, randomized controlled trials (RCTs) and data coming from large cohort studies were considered. Relevant studies were searched in PubMed (1975–present), Embase (January 1980– present) and Cochrane review of aggregate data for reports written in any language. The full PubMed search strategy is available in Supplementary Table 1. Moreover, hand or computerized search involving the recent (1999–2014) conference proceedings from the Society of Thoracic Surgeons, European Association for Cardiothoracic Surgery and European Society for Cardiovascular Surgery and the American Association for Thoracic Surgery Annual Meetings was performed; ClinicalTrials.gov was explored in order to identify any ongoing or unpublished trials (Table 2).
Table 2:
Summary of the studies used for the consensus document
| Author, journal, date [Ref.] | Study type | Type of procedure | Patient groups | Type of MiECC circuit | Key results | Comments |
|---|---|---|---|---|---|---|
| Wiesenack et al., Artif Organs, 2004 [10] | Retrospective analysis | CABG | 485 MiECC/485 CCPB | Type I |
|
First reported large series showing improved perfusion characteristics and clinical results |
| Yilmaz et al., Interact CardioVasc Thorac Surg, 2010 [11] | Prospective cohort study | CABG + AVR | 65 MiECC/135 CCPB | Type III |
|
Feasibility study |
| Anastasiadis et al., Perfusion, 2015 [12] | Prospective cohort study | Various cardiac case-mix | 50 consecutive patients | Type IV |
|
Clinical study on modular type IV MiECC in all types of cardiac surgery (feasibility and safety study) |
| El-Essawi et al., Perfusion, 2011 [13] | Multicentre RCT (six centres) | CABG and/or AVR | 252 MiECC/248 CCPB | Type IV |
|
Focus on modular type IV MIECC in CABG and/or AVR |
| Fromes et al., Anaesthesia, 2011 [14] | Retrospective analysis | CABG | 100 patients 300 IU/kg heparin/68 patients 145 IU/kg heparin | Type II |
|
Implementation of low-dose heparin protocol |
| Nilsson et al., Interact CardioVasc Thorac Surg, 2012 [15] | RCT | CABG | 27 low-dose heparin/29 regular dose | Type II |
|
Feasibility of low-dose heparin |
| Anastasiadis et al., J Cardiothorac Vasc Anesth, 2013 [16] | RCT | CABG | 60 MiECC/60 CCPB | Type II |
|
Focus on fast-track protocols |
| Anastasiadis et al., Perfusion, 2010 [17] | RCT | CABG | 50 MiECC/49 CCPB | Type I |
|
Focus on haematological effects |
| Haneya et al., ASAIO J, 2013 [18] | Retrospective cohort analysis | CABG | 1073 MiECC/872 CCPB | Type I |
|
Focus on patients with preoperative anaemia |
| Zangrillo et al., J Thorac Cardiovasc Surg, 2010 [19] | Meta-analysis (16 RCTs) | CABG or AVR | 803 MiECC/816 CCPB |
|
Meta-analysis | |
| Anastasiadis et al., Int J Cardiol, 2013 [20] | Meta-analysis (24 RCTs) | CABG or AVR | 1387 MiECC/1383 CCPB |
|
The largest meta-analysis | |
| Rahe-Meyer et al., Artif Organs, 2010 [21] | Prospective cohort study | CABG | 44 MiECC/44 CCPB | Type I |
|
Focus on coagulation |
| El-Essawi et al., Perfusion, 2013 [22] | Cohort study (Jehovah's Witnesses) | Various cardiac case-mix | 29 patients 22CABG ± AVR 7 various case-mix |
Type IV |
|
Feasibility study on Jehovah's Witnesses |
| Fromes et al., Eur J Cardiothorac Surg, 2002 [23] | RCT | CABG | 30 MiECC/30 CCPB | Type I |
|
Focus on SIRS |
| Immer et al., Ann Thorac Surg, 2007 [24] | Comparative cohort study | CABG | 1053 MiECC/353 CCPB | Type I + smart suction device |
|
Feasibility/safety study |
| Abdel-Rahman et al., Ann Thorac Surg, 2005 [25] | RCT | CABG | 101 MiECC/103 CCPB | Type II |
|
Feasibility/safety study |
| Ohata et al., J Artif Organs, 2007 [26] | RCT | CABG | 15 MiECC/15 CCPB | Type I |
|
Focus on SIRS |
| Puehler et al., Ann Thorac Surg, 2009 [27] | Comparative cohort study | CABG | 558 MiECC/558 CCPB/558 OPCAB | Type I |
|
Feasibility/safety study |
| Biancari, Heart, 2009 [28] | Meta-analysis (13 RCTs) | CABG or AVR | 562 MiECC/599 CCPB |
|
Meta-analysis | |
| Liebold, J Thorac Cardiovasc Surg, 2006 [29] | RCT | CABG | 20 MiECC/20 CCPB | Type I |
|
Focus on cerebral protection |
| Zanatta, J Cardiothorac Vasc Anesth, 2013 [30] | Retrospective cohort | CABG | 19 MiECC (CABG)/18 CCPB (AVR or MVR)/18 port-access MVR | Type I |
|
Focus on cerebral protection |
| Camboni, ASAIO J, 2009 [31] | RCT | CABG | 42 MiECC type I 10 MiECC type II 41 CCPB |
Type I and II |
|
Focus on cerebral protection |
| Anastasiadis et al., Heart, 2011 [32] | RCT | CABG | 29 MiECC /31 CCPB | Type I |
|
Focus on neurocognitive outcome |
| Reineke et al., Interact CardioVasc Thorac Surg, 2014 [33] | Cohort study | CABG | 31 MiECC | Type I + smart suction device |
|
Focus on neurocognitive outcome |
| Gynaydin, Perfusion, 2009 [34] | RCT | CABG | 20 MiECC/20 CCPB | Type IV |
|
Focus on SIRS and haemodilution |
| Bennett et al., Perfusion, 2014 [35] | Cohort study | CABG and/or AVR | 39 MiECC 41 CCPB |
Type II |
|
Focus on cerebral protection |
| Panday et al., Interact Cardiovasc Thorac Surg, 2009 [36] | Prospective cohort study | CABG | 220 MiECC 1143 CCPB 109 OPCAB |
Type II |
|
Focus on blood transfusion |
| Remadi, Am Heart J, 2006 [37] | RCT | CABG | 200 MiECC/200 CCPB | Type I + suction device |
|
Feasibility/safety study |
| Diez et al., ASAIO J, 2009 [38] | Retrospective observational study | CABG | 1685 MiECC /3046 CCPB | Type I |
|
Focus on renal function |
| Huybregts et al., Ann Thorac Surg, 2007 [39] | RCT | CABG | 25 MiECC/24 CCPB | Type II |
|
Focus on renal and intestinal function |
| Capuano et al., Interact CardioVasc Thorac Surg, 2009 [40] | Prospective cohort study | CABG | 30 MiECC/30 CCPB | Type II |
|
Focus on renal injury |
| Benedetto et al., Ann Thorac Surg, 2009 [41] | Prospective cohort study | CABG | 104 MiECC/601 CCPB | Type II |
|
Focus on renal injury |
| Bauer et al., J Extra Corpor Technol, 2010 [42] | RCT | CABG | 18 MiECC/22 CCPB | Type II |
|
Focus on perfusion characteristics |
| Skrabal, ASAIO J, 2007 [43] | RCT | CABG | 30 MiECC/30 CCPB | Type I |
|
Focus on myocardial protection |
| Van Boven et al., Eur J Cardiothorac Surg, 2008 [44] | RCT | CABG | 10 MiECC 10 CCP 10 OPCAB |
Type I |
|
Focus on myocardial protection |
| Nguyen, Mol Cell Biochem, 2014 [45] | RCT | CABG | 13 MiECC/13 CCPB (intermittent cross-clamp fibrillation) | Type III |
|
Focus on myocardial protection |
| Van Boven, Eur J Anaesthesiol, 2013 [46] | RCT | CABG | 20 MiECC 20 CCP 20 OPCAB |
Type I |
|
Focus on end-organ protection |
| Prasser et al., Perfusion, 2007 [47] | RCT | CABG | 10 MiECC/10 CCPB | Type I |
|
Focus on liver function |
| Donndorf et al., J Thorac Cardiovasc Surg, 2012 [48] | RCT | CABG | 20 MiECC/20 CCPB | Type I |
|
Focus on microvascular perfusion |
| Haneya et al., Eur J Cardiothorac Surg, 2009 [49] | Retrospective cohort study | CABG | 105 MiECC /139 CCPB (high-risk patients) | Type I |
|
Focus on high-risk patients |
| Kolat et al., J Cardiothorac Surg, 2014 [50] | Retrospective cohort analysis | CABG | 1137 MiECC /1137 CCPB | Type I |
|
Focus on clinical outcome |
| Ried et al., J Cardiothorac Surg, 2013 [51] | Propensity score analysis | Emergency CABG | 146 MiECC /175 CCPB | Type I |
|
Focus on emergency CABG |
| Koivisto et al., Perfusion, 2010 [52] | Propensity score analysis | CABG | 89 MiECC /147 CCPB | Type II |
|
Focus on high-risk patients |
| Anastasiadis et al., Int J Cardiol, 2013 [53] | Cost-analysis | CABG | 1026 MiECC/1023 CCPB |
|
Focus on cost-effectiveness | |
| Fernandes, Perfusion, 2010 [54] | Retrospective cohort study | CABG | 15 MiECC | Type II |
|
Focus on perfusion characteristics |
| Puehler et al., Thorac Cardiovasc Surg, 2010 [55] | Retrospective comparative cohort study | CABG | 119 MiECC /119 CCPB | Type I |
|
Focus on high-risk patients |
a-GST: a-glutathione S-Transferase AF: atrial fibrillation; AKI: acute kidney injury; AVR: aortic valve replacement; CABG: coronary artery bypass grafting; CCPB: conventional cardiopulmonary bypass; CPB: cardiopulmonary bypass; HFABP: heart type fatty acid binding protein; ICU: intensive care unit; IFABP: intestinal type fatty acid binding protein; IL: interleukin; MAP: mean arterial pressure; MiECC: minimal invasive extracorporeal circulation; MVR: mitral valve replacement; NGAL: neutrophil gelatinase-associated lipocalin; OPCAB: off-pump coronary artery bypass grafting; POD: postoperative day; RCT: randomized controlled trial; SIRS: systemic inflammatory response syndrome; TNF: tumour necrosis factor; TnT: troponin-T; TnI: troponin I.
Recommendations and evidence-based practice guidelines
Expert Committee statements are presented in Table 3. Evidence-based clinical practice guidelines are presented in Table 4.
Table 3:
Summary of statements endorsed by the Expert Committee
| Recommendation |
|---|
| MiECC refers to a combined strategy of surgical approach, anaesthesiological and perfusion management and should not be limited to the CPB circuit alone. |
| In order to be characterized as MiECC, the main components of the system must include closed circuit; biologically inert blood contact surfaces; reduced priming volume; centrifugal pump; membrane oxygenator; heat exchanger; cardioplegia system; venous bubble trap/venous air removing device; shed blood management system. Additional components that can be integrated to a MiECC system are pulmonary artery vent; pulmonary vein vent; aortic root vent; soft bag/soft-shell reservoir; hard-shell reservoir (modular systems); regulated smart suction device; arterial line filtration. |
MiECC: minimal invasive extracorporeal circulation; CPB: cardiopulmonary bypass.
Table 4:
Summary of evidence-based practice guidelines
| Recommendation | Level of evidence | References |
|---|---|---|
| Class I | ||
| MiECC systems reduce haemodilution and better preserve haematocrit as well as reduce postoperative bleeding and the need for RBC transfusion | A | [17, 19, 20] |
| MiECC systems reduce the incidence of postoperative atrial fibrillation | A | [13, 16, 19, 20] |
| MiECC systems preserve renal function | A | [20, 39] |
| MiECC is associated with improved myocardial protection | A | [20, 43–45] |
| Class IIA | ||
| Inflammatory response assessed by specific inflammatory markers is attenuated with use of MiECC | B | [23–26] |
| MiECC systems can reduce cerebral gaseous microembolism and preserve neurocognitive function | B | [20, 29–33] |
| MiECC exerts a subclinical protective effect on end-organ function (lung, liver, intestine) which is related to enhanced recovery of microvascular organ perfusion | B | [39, 46–48] |
| Class IIB | ||
| Within a MiECC strategy, less thrombin generation may permit reduced heparin dose targeted to shorter ACT times. When such a strategy is followed, individual heparin dose should be determined using heparin dose–response monitoring systems | B | [14, 15, 56, 57] |
| MiECC appears to offer survival benefit in terms of lower 30-day mortality after CABG procedures | B | [20, 49, 50, 51] |
| The use of short-acting opioids in combination with propofol or volatile anaesthetics, and hypnotic effect monitoring by processed EEG, is recommended for induction and maintenance of anaesthesia for MiECC-based surgery. TOE findings pertinent to institutional management of MiECC should be communicated during the preoperative surgical safety time out | C | [16, 58–61] |
ACT: activated clotting time; CABG: coronary artery bypass grafting; EEG: electroencephalogram; MiECC: minimal invasive extracorporeal circulation; RBC: red blood cells; TOE: transoesophageal echocardiography; rScO2: regional cerebral oxygen saturation.
Terminology
MiECC refers to a combined strategy of surgical approach, anaesthesiological and perfusion management and is not limited to the CPB circuit alone.
Several terms have been used to describe a MiECC circuit: miniaturized extracorporeal circulation (MECC), mini-extracorporeal circulation (mECC), minimized extracorporeal circulation, mini-cardiopulmonary bypass (mCPB, mini-CPB), minimally invasive cardiopulmonary bypass (MICPB), miniaturized cardiopulmonary bypass (MCPB), veno-arterial extracorporeal membrane oxygenation, minimized perfusion circuit, minimized extracorporeal life support system, minimized CPB, MiECC. This divergent terminology creates confusion and disagreement between centres. But the major problem is the fact that the focus is made only on the priming volume of the circuit and not on the reduction of the adverse effects of ECC.
The Steering Committee of MiECTiS uses the term ‘minimal invasive’ to describe a procedure which involves not only the CPB circuit, but the global approach to the procedure. This concept strives to render the procedure minimally invasive as opposed to the widely employed misnomer ‘minimal invasive’ when a limited surgical access is performed. The term ‘minimal invasive’ is misleading since the patient is often a longer period on CPB, cross-clamping and duration of the anaesthesia are prolonged. In this sense, the term ‘minimal invasive’ relates only to the size of the scar [62]. Hence, we believe that the term ‘minimal invasive extracorporeal circulation’ corresponds better to the above mentioned concept and should be used to describe this technology with the abbreviation: ‘MiECC’.
Components of MiECC system
In order to be characterized as MiECC, the main components of the system must include a closed CPB circuit; biologically inert blood contact surfaces; reduced priming volume; a centrifugal pump; a membrane oxygenator; a heat exchanger; a cardioplegia system; a venous bubble trap/venous air removing device and a shed blood management system.
Because different groups have utilized either commercially available or customized CPB circuits with a variety of components, the Consensus Meeting defined the main components of the CPB circuit when it should correspond to a MiECC system. The Steering Committee of MiECTiS emphasizes that a MiECC system should comprise all necessary elements to obtain a maximal benefit.
Originally, MiECC system was an extracorporeal life support (ECLS) circuit with the possibility to administrate cardioplegia (Type I) and used mainly to perform CABG procedures [10]. However, the safety concerns regarding air entrapment/air lock into the venous line prompted the integration of venous bubble trap/venous air removing devices into the system (Type II). This design increased safety for CABG procedures and enabled aortic valve surgery [11]. The need for blood volume management during valvular procedures required the addition of a soft-bag/soft-shell reservoir integrated into the system (Type III). This enabled safe performance of aortic valve surgery and other intracardiac procedures. Initiation of modular MiECC (hybrid) systems that integrate a second open circuit with a venous reservoir and cardiotomy suction as a stand-by component (Type IV) enabled performance of complex procedures that pertain a high possibility of unexpected perfusion scenario [12, 13]. Classification of MiECC types is illustrated in Fig. 1. The Consensus Meeting defined as a prerequisite for a system to be considered as MiECC to have at least Type II circuit characteristics.
Figure 1:
Classification of MiECC circuits [12]. [Note that the modular type IV circuit is literally type III with a standing-by component, used only when necessary] (X: pump; O: oxygenator; C: cardioplegia; T: bubble trap/air-removing device; V: vent (aortic/pulmonary); S: soft-bag/reservoir; H: hard-shell/reservoir). MiECC: minimal invasive extracorporeal circulation.
Additional components to be integrated into a MiECC system are (i) pulmonary artery vent, (ii) aortic root vent, (iii) pulmonary vein vent, (iv) soft-bag/soft-shell reservoir, (v) hard-shell reservoir (modular systems), (vi) regulated smart suction device and (vii) arterial line filtration.
Modular systems
The major reticence to limit expansion of MiECC is due to the thoughts about safety in case of massive air entrance into the system or significant blood loss. Although CABG and valve surgery are feasible with the standard type II MiECC circuit, a modular configuration is welcome to expand MiECC for the majority of cardiac procedures and to create a ‘safety net’ for unexpected intraoperative scenarios. Recently published results from a single-centre indicate that a modular circuit design offers 100% technical success rate in high-risk patients, even in those undergoing complex procedures including reoperations, valve and aortic surgery as well as emergency cases [12].
Anticoagulation management
During perfusion with MiECC, less thrombin generation may allow reduced heparin dose targeted by shorter ACT (Class of Recommendation IIB, Level of Evidence B). In this case, individual heparin dosage should be determined using heparin dose–response monitoring systems.
A number of factors including better biocompatible surfaces, elimination of blood–air interaction and exclusion of unprocessed shed blood reinfusion favourably influence thrombin generation under MiECC system compared with the standard CPB [56]. A patient-adjusted and/or a procedure-adjusted coagulation management based on unfractionated heparin (UFH) can be adopted [14, 15, 63]. Thus, a low-dose anticoagulation protocol for CABG with a targeted activated clotting time (ACT) of 300–350 s and 400–450 s for valve surgery and complex cardiac procedures is safe, respectively [20]. Serial assessment of ACT during CPB is mandatory. Point-of-care (POC) coagulation monitoring (for instance, the Hepcon system) to optimize heparin and protamine dosage during CPB is recommended if a low-dose heparin protocol is adopted. Appropriate protamine reversal should be used post-CPB to normalize ACT. Continuous infusion of UFH may result in less consumptive coagulopathy and transfusion requirements [57, 64].
Anaesthesia for surgery on MiECC
The use of short-acting opioids in combination with propofol or volatile anaesthetics, and monitoring of the depth of anaesthesia by processed electroencephalogram (EEG), is recommended for all patients undergoing cardiac surgery with MiECC (Class of Recommendation IIB, Level of Evidence C). Transoesophageal echocardiography (TOE) findings pertinent to institutional management of MiECC should be communicated during the preoperative surgical safety time out (Class of Recommendation IIB, Level of Evidence C).
Anaesthetic management of patients undergoing cardiac surgery with the aid of a MiECC system follows the international recommendations, especially regarding the use of TOE [58, 59]. Following anaesthesia induction, TOE may provide additional information that may influence the site and/or the type of cannulation or perfusion strategy (e.g. patent foramen ovale, significant mitral or aortic valve pathology or severe aortic atheromatosis). This information is important when Type I or II MiECC systems are used, whereas any modifications can be accommodated when Type III or modular type IV configuration are available.
Specifically, the absence of venous reservoir in MiECC systems renders the patient’s own venous capacitance compartment critical for haemodynamics as well as for optimal volume management. Positioning of the patient (Trendelenburg or anti-Trendelenburg) and low-dose vasoactive agents are useful to control intraoperative haemodynamics. Excessive fluid administration should be avoided to reduce haemodilution and avoid transfusion [63].
Beneficial effects of MiECC include attenuation of inflammatory response, higher haematocrit, less coagulation disorders and improved end-organ function (brain, kidneys, lungs). It facilitates implementation of fast-track protocols [16]. Hence, perioperative use of short-acting intravenous and/or volatile anaesthetic agents is recommended. Moreover, the titration of anaesthetic agents using processed EEG ensures adequate anaesthesia depth [60]. Microporous capillary membrane oxygenators enable volatile anaesthetics to be used for anaesthesia maintenance, which is not feasible with diffusion membrane oxygenators [61]. To date, RCTs comparing different anaesthetic protocols for MiECC-based surgery are still missing.
Haemodilution: haematocrit—transfusion
MiECC systems reduce haemodilution, better preserve haematocrit and reduce postoperative bleeding and the need for RBC transfusion (Class of Recommendation I, Level of Evidence A).
There is compelling evidence that MiECC—mainly because of the significantly reduced priming volume of the circuit—reduces haemodilution and results in a higher haematocrit at the end of the perfusion period [17, 18]. This significantly reduces need for red blood cells transfusion and improves oxygen delivery during perfusion [13, 17, 19, 20, 65]. Coagulation disorders are reduced [17] and platelet count and function are better preserved following perfusion with MiECC systems [21]. Postoperative bleeding and incidence of re-exploration are significantly lower in patients operated with MiECC [20]. As it reduces haemodilution, MiECC fulfils Class of Recommendation I, Level of Evidence A indication for blood conservation according to the STS guidelines, especially in patients at high risk for adverse effects of haemodilution (paediatric patients and small-sized adults) [8]. Patients refusing transfusion of allogeneic blood products, e.g. Jehovah's Witnesses, are optimal candidates for this strategy [22].
Attenuation of the inflammatory response
Inflammatory response is attenuated with use of MiECC (Class of Recommendation IIA, Level of Evidence B).
Several studies have investigated the inflammatory response triggered conventional CPB and compared it with MiECC systems. MiECC components are designed to limit the severity of SIRS. Coating and reduction of the size of the circuit reduce the amount of foreign surfaces, which is the main trigger of SIRS, but multicentre studies still have to confirm this observation [66]. Assessment of the inflammatory response is complex and clinical presentation is highly variable [67]. Nevertheless, some studies provide evidence of the beneficial effects of MiECC. Moreover, Fromes et al. [23] described a less pronounced intraoperative decrease of monocytes as well as during the first 24 h in patients with MiECC than in those with conventional CPB. Others demonstrated significantly lower peak levels of IL-6 under MiECC [23, 24, 68]. Finally, several studies demonstrated that perfusion with MiECC resulted in significantly lower levels of neutrophil elastase—a specific marker of neutrophil activation—than with conventional CPB [23, 25, 26].
Neurological function
MiECC systems reduce cerebral gaseous microembolism and better preserve neurocognitive function (Class of Recommendation IIA, Level of Evidence B).
Several prospective studies and meta-analyses have reported reduced incidence of stroke following MiECC when compared with conventional CPB [19, 27, 28]. A recent meta-analysis found a trend to reduction of neurological damage in favour of MiECC [20]. Of course, stroke is multifactorial and the perfusion system is only one of the issues besides aortic manipulations and other patient's specific factors [69]. A possible explanation for the neuroprotective effect of MiECC is the significant reduction of gaseous microemboli [29–33]. MiECC also offers improved cerebral perfusion during CPB, as indicated by the lower reduction in near infrared spectroscopy (NIRS)-derived regional cerebral oxygen saturation values and cerebral desaturation episodes [29, 32, 34, 35]. Reduced incidence of cerebral desaturation episodes favourably affects neurocognitive outcome [70–72].
Atrial fibrillation
MiECC reduces the incidence of postoperative atrial fibrillation (AF) (Class of Recommendation I, Level of Evidence A).
Several randomized studies have demonstrated that postoperative AF is significantly reduced following MiECC when compared with conventional CPB [13, 16, 24, 36]. Moreover, there is strong evidence of a lower incidence of AF in all meta-analyses regarding MiECC systems [19, 20, 28]. Attenuated inflammatory reaction and less volume shifts associated with MiECC may be an explanation for this beneficial effect [37].
Renal function
MiECC preserves renal function (Class of Recommendation I, Level of Evidence A).
Several studies have shown that the use of MiECC systems was associated with better preservation of renal function [38–40]. This was confirmed by a meta-analysis of 24 RCTs but this meta-analysis and other studies failed to demonstrate a reduced incidence of postoperative renal failure [20, 38, 41]. More stable haemodynamics together with higher perfusion pressure and a reduced need for vasopressors during MiECC perfusion may explain this observation [10, 42]. A significant independent association was found between the lowest haematocrit value during bypass and acute renal injury, with significant benefits on renal function seen after reduction of the priming volume. This may be due to a higher DO2 associated with a higher haematocrit on CPB [65]. In addition, different markers to evaluate renal function (i.e. glomerular filtration rate, levels of neutrophil gelatinase-associated lipocalin) confirm better renal protection under MiECC. Larger studies are required to investigate if this protective effect is sufficient to prevent development of acute renal failure.
Myocardial protection
MiECC is associated with improved myocardial protection (Class of Recommendation I, Level of Evidence A).
Several studies have demonstrated a beneficial effect of MiECC on intraoperative myocardial protection [10, 20, 43, 44]. Reduced cardioplegia volumes with less crystalloids and attenuation of SIRS may explain this beneficial effect [23]. Studies with MiECC and intermittent cross-clamping show a similar effect on myocardial protection [45]. However, myocardial protection is not related only to the duration of ischaemia, but also to the reperfusion phase. Increased arterial pressure during CPB as well as the volume-constant perfusion with a closed system may also contribute to improved myocardial protection [38, 42].
End-organ protection
MiECC has a subclinical protective effect on end-organ function (lung, liver, intestine) caused by improved microvascular organ perfusion (Class of Recommendation IIA, Level of Evidence B).
MiECC is a closed system that allows a better peripheral perfusion with higher arterial pressure and systemic vascular resistance close to normal values [38]. This is associated with reduced requirement for vasoactive support [10, 42]. Data from randomized studies suggest improved lung protection [46], attenuated liver and intestinal dysfunction [39, 46, 47]. These studies evaluated only surrogate markers of end-organ dysfunction that may beneficiate from MiECC, whereas the effects remain subclinical. However, it may become clinically perceptible in high-risk patients and in those with longer procedures since MiECC would lead to fewer alterations of microperfusion [48].
Mortality
MiECC appears to offer survival benefit in terms of lower 30-day mortality after CABG procedures (Class of Recommendation IIB, Level of Evidence B).
A number of studies have demonstrated a trend towards reduced mortality in CABG performed on MiECC. A recent meta-analysis of 24 studies involving 2770 patients showed that MiECC was associated with a significant decrease in mortality, compared with conventional CPB (0.5 vs 1.7%; P = 0.02) [20]. This finding has also confirmed by other studies [49, 50, 51]. A trend towards decreased mortality in favour of MiECC has also been found in meta-analyses [19, 28] and in a propensity score analysis [52]. This survival benefit may be the result of the cumulative beneficial effects of MiECC on end-organ protection but it calls for a multicentre RCT sufficiently powered to prospectively investigate this survival benefit.
Cost-effectiveness
Data from a cost-analysis study indicate a cost-effectiveness of MiECC systems that offer economic advantages in various healthcare settings [53]. Nevertheless, these results have to be considered in the context of the local conditions. A more detailed analysis together with an analysis from a payer's perspective is necessary. Better standardization should be achieved to allow comparison of costs and economic benefits.
DISCUSSION
MiECC systems have been developed to integrate all advances in CPB technology in one closed circuit: the goal is to improve biocompatibility and minimize side-effects of CPB. MiECC is associated with more stable haemodynamics during and early after perfusion and better end-organ protection. This concept provides comparable or better outcomes in terms of morbidity and mortality in CABG and valve procedures, as shown in prospective randomized studies and meta-analyses. However, despite several clinical advantages, the implementation of MiECC technology remains weak probably there are still some concerns regarding air handling as well as blood and volume management during perfusion [12]. This Consensus paper primarily serves to summarize the available information about this technology and to clarify some of the open issues. We have made substantial efforts to provide the best available actual evidence and strongly encourage readers to consider the technology as a multidisciplinary strategy.
There is still debate about the optimal handling of air during the perfusion, as well as volume and blood management when a MiECC system is used. Mean arterial pressure (MAP) is usually higher during MiECC: this raises the question of optimal pump flow rate during MiECC perfusion [10, 42]. A reference blood flow based on body surface area is not a guarantee of adequate body perfusion during CPB. Modern protocols adjust pump flow to achieve adequate DO2. In this area, it is still unclear if the use of MiECC may allow lower than traditional cardiac index without end-organ damage as has been suggested by recent studies [54, 73]. The use of NIRS and other parameters to monitor cerebral blood flow may lead to greater individualization of perfusion index for adequate end-organ perfusion [35, 74]. Lower heparin requirement and reduced haemodilution offered by MiECC facilitate the management of postoperative bleeding. The prophylactic use of low-dose antifibrinolytics [75] and POC coagulation management based on thromboelastometry and aggregometry is generally advised [76]. In patients with higher perioperative risk [52], those with low ejection fraction and emergencies [51, 52, 55], MiECC has proved to be safe.
In general, MiECC can be considered as the ‘circuit-of-choice’ to replace conventional CPB at least for CABG surgery. Novel modular systems (Type IV MiECC) may be utilized for all cardiac procedures. We believe that the terms ‘circuit’ which refers to the CPB, the ‘MiECC system’ which integrates certain components to a CPB circuit and the ‘MiECC strategy’ that represents the multidisciplinary approach to MiECC should be differentiated. The MiECTiS advocates this strategy to obtain the maximal benefits for the patients. The authors believe that MiECC should be understood as an additional tool in the chapter of minimal invasiveness. The latter should not be restricted to ‘minimal-access’ surgery, but should also incorporate a strategy towards a ‘more physiologic CPB’. The use of MiECC should be integrated within fast-track algorithms, POC management of coagulation disorders together with any initiatives that improve aortic assessment (epiaortic ultrasound), novel anti-inflammatory strategies, low shear-stress cannula design and implementation of contemporary biofiltration techniques.
Lack of high-volume data requires the creation of a registry to further evaluate this technology. Moreover, the variation in extent of miniaturization/complexity of MiECC systems should be analysed. Additional RCTs, focusing on valve and other cardiac procedures, as well as large cohorts of patients will provide more evidence regarding clinical effectiveness. Adequately powered prospective multicentre studies are required in order to prove the superiority of the MiECC over the conventional CPB.
Concerns in the literature have been raised regarding decreased safety, ventricular dilatation during perfusion using the MiECC circuit, loss of a bloodless field and the risk of air embolism [77, 78]; however, these reports are anecdotal and are not supported by large-scale studies. Loss of safety during perfusion with a modern MiECC circuit is easily addressed with integration of a venous bubble trap/air removing device into the circuit. Moreover, significant air entrainment that blocks the circuit could be resolved immediately by a skilled perfusionist. Ventricular dilatation, attributed to poor off-loading of the heart, is anticipated with the use of aortic root and/or pulmonary artery/vein venting from Type II MiECC onwards. The same applies to creation of a full bloodless field. Special patient populations, such as patients with a higher body surface area requiring higher circulatory flows, are easily managed with kinetic-assisted venous drainage and increased flow through the centrifugal pump. Regarding air embolism, contemporary evidence suggests that there is significantly reduced amount of gaseous microemboli in the arterial line of MiECC systems compared with conventional CPB [79].
Nevertheless, it should be emphasized that MiECC is a demanding system that should be implemented in cardiac surgery as a strategy and not as a simple circuit. Real teamwork from all disciplines of the surgical team, meticulous surgery, a skilful perfusionist and optimal anaesthetic management are mandatory for a more physiological perfusion that could lead to improved clinical outcomes. MiECTiS supports initiatives that promote research and clinical application of MiECC systems as a strategy through multidisciplinary training programmes (dry labs/hands-on simulators, wet labs, peer-to-peer workshops). Integration of specific training programmes under the accreditation of MiECTiS will stimulate and improve the collaboration between clinicians while the industry will get important information to further improve the systems. MiECTiS is planning to endorse a comprehensive and structured programme that contributes to the advancement of patient care.
In conclusion, the authors consider MiECC as a physiologically based strategy and not just a CPB circuit or a particular product. For this reason, multidisciplinary approach is mandatory. Collaboration between surgeons, anaesthesiologists and perfusionists is of paramount importance to emphasize the key tenets of MiECTiS.
SUPPLEMENTARY MATERIAL
Supplementary material is available at ICVTS online.
REFERENCES
- 1.Song HK, Diggs BS, Slater MS, Guyton SW, Ungerleider RM, Welke KF. Improved quality and cost-effectiveness of coronary artery bypass grafting in United States from 1988 to 2005. J Thorac Cardiovasc Surg 2009;137:65–9. [DOI] [PubMed] [Google Scholar]
- 2.Anastasiadis K, Antonitsis P, Argiriadou H (eds.) Principles of Miniaturized Extra-Corporeal Circulation. Berlin Heidelberg: Springer-Verlag, 2013, 1–8. [Google Scholar]
- 3.Schonberger JP, Everts PA, Hoffmann JJ. Systemic blood activation with open and closed venous reservoir. Ann Thorac Surg 1995;59:1549–55. [DOI] [PubMed] [Google Scholar]
- 4.Butler J, Rocker GM, Westaby S. Inflammatory response to cardiopulmonary bypass. Ann Thorac Surg 1993;55:552–9. [DOI] [PubMed] [Google Scholar]
- 5.Kirklin JK, Westaby S, Blackstone EH, Kirklin JW, Chenoweth DE, Pacifico AD. Complement and the damaging effects of cardiopulmonary bypass. J Thorac Cardiovasc Surg 1983;86:845–57. [PubMed] [Google Scholar]
- 6.Speir AM, Kasirajan V, Barnett SD, Fonner E Jr. Additive costs of postoperative complications for isolated coronary artery bypass grafting patients in Virginia. Ann Thorac Surg 2009;88:40–5. [DOI] [PubMed] [Google Scholar]
- 7.Anastasiadis K, Bauer A, Antonitsis P, Gygax E, Schaarschmidt J, Carrel T. Minimal invasive Extra-Corporeal Circulation (MiECC): a revolutionary evolution in perfusion. Interact CardioVasc Thorac Surg 2014;19:541–2. [DOI] [PubMed] [Google Scholar]
- 8.Society of Thoracic Surgeons Blood Conservation Guideline Task Force Ferraris VA, Brown JR, Despotis GJ, Hammon JW, Reece TB et al. 2011 update to the Society of Thoracic Surgeons and the Society of Cardiovascular Anesthesiologists blood conservation clinical practice guidelines. Ann Thorac Surg 2011;91:944–82. [DOI] [PubMed] [Google Scholar]
- 9.American College of Cardiology Foundation and American Heart Association, Inc. cardiosource.org 2010. Available at: https://my.americanheart.org/idc/groups/ahamah-public/@wcm/@sop/documents/downloadable/ucm_319826.pdf.
- 10.Wiesenack C, Liebold A, Philipp A, Ritzka M, Koppenberg J, Birnbaum DE et al. Four years’ experience with a miniaturized extracorporeal circulation system and its influence on clinical outcome. Artif Organs 2004;28:1082–8. [DOI] [PubMed] [Google Scholar]
- 11.Yilmaz A, Sjatskig J, van Boven WJ, Waanders FG, Kelder JC, Sonker U et al. Combined coronary artery bypass grafting and aortic valve replacement with minimal extracorporeal closed circuit circulation versus standard cardiopulmonary bypass. Interact CardioVasc Thorac Surg 2010;11:754–7. [DOI] [PubMed] [Google Scholar]
- 12.Anastasiadis K, Antonitsis P, Argiriadou H, Deliopoulos A, Grosomanidis V, Tossios P. Modular minimally invasive extracorporeal circulation systems; can they become the standard practice for performing cardiac surgery? Perfusion 2015;30:195–200. [DOI] [PubMed] [Google Scholar]
- 13.El-Essawi A, Hajek T, Skorpil J, Böning A, Sabol F, Ostrovsky Y et al. Are minimized perfusion circuits the better heart lung machines? Final results of a prospective randomized multicentre study. Perfusion 2011;26:470–8. [DOI] [PubMed] [Google Scholar]
- 14.Fromes Y, Daghildjian K, Caumartin L, Fischer M, Rouquette I, Deleuze P et al. A comparison of low vs conventional-dose heparin for minimal cardiopulmonary bypass in coronary artery bypass grafting surgery. Anaesthesia 2011;66:488–92. [DOI] [PubMed] [Google Scholar]
- 15.Nilsson J, Scicluna S, Malmkvist G, Pierre L, Algotsson L, Paulsson P et al. A randomized study of coronary artery bypass surgery performed with the Resting Heart System utilizing a low vs a standard dosage of heparin. Interact CardioVasc Thorac Surg 2012;15:834–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Anastasiadis K, Asteriou C, Antonitsis P, Argiriadou H, Grosomanidis V, Kyparissa M et al. Enhanced recovery after elective coronary revascularization surgery with minimal versus conventional extracorporeal circulation: a prospective randomized study. J Cardiothorac Vasc Anesth 2013;27:859–64. [DOI] [PubMed] [Google Scholar]
- 17.Anastasiadis K, Asteriou C, Deliopoulos A, Argiriadou H, Karapanagiotidis G, Antonitsis P et al. Haematological effects of minimized compared to conventional extracorporeal circulation after coronary revascularization procedures. Perfusion 2010;25:197–203. [DOI] [PubMed] [Google Scholar]
- 18.Haneya A, Philipp A, Von Suesskind-Schwendi M, Diez C, Hirt SW, Kolat P et al. Impact of minimized extracorporeal circulation on outcome in patients with preoperative anemia undergoing coronary artery bypass surgery. ASAIO J 2013;59:269–74. [DOI] [PubMed] [Google Scholar]
- 19.Zangrillo A, Garozzo FA, Biondi-Zoccai G, Pappalardo F, Monaco F, Crivellari M et al. Miniaturized cardiopulmonary bypass improves short-term outcome in cardiac surgery: a meta-analysis of randomized controlled studies. J Thorac Cardiovasc Surg 2010;139:1162–9. [DOI] [PubMed] [Google Scholar]
- 20.Anastasiadis K, Antonitsis P, Haidich AB, Argiriadou H, Deliopoulos A, Papakonstantinou C. Use of minimal extracorporeal circulation improves outcome after heart surgery; a systematic review and meta-analysis of randomized controlled trials. Int J Cardiol 2013;164:158–69. [DOI] [PubMed] [Google Scholar]
- 21.Rahe-Meyer N, Solomon C, Tokuno ML, Winterhalter M, Shrestha M, Hahn A et al. Comparative assessment of coagulation changes induced by two different types of heart-lung machine. Artif Organs 2010;34:3–12. [DOI] [PubMed] [Google Scholar]
- 22.El-Essawi A, Breitenbach I, Ali K, Jungebluth P, Brouwer R, Anssar M et al. Minimized perfusion circuits: an alternative in the surgical treatment of Jehovah's Witnesses. Perfusion 2013;28:47–53. [DOI] [PubMed] [Google Scholar]
- 23.Fromes Y, Gaillard D, Ponzio O, Chauffert M, Gerhardt MF, Deleuze P et al. Reduction of the inflammatory response following coronary bypass grafting with total minimal extracorporeal circulation. Eur J Cardiothorac Surg 2002;22:527–33. [DOI] [PubMed] [Google Scholar]
- 24.Immer FF, Ackermann A, Gygax E, Stalder M, Englberger L, Eckstein FS et al. Minimal extracorporeal circulation is a promising techniques for coronary artery bypass grafting. Ann Thorac Surg 2007;84:1515–21. [DOI] [PubMed] [Google Scholar]
- 25.Rahman UA, Ozaslan F, Risteski PS, Martens S, Moritz A, Daraghmeh AA et al. Initial experience with a minimized extracorporeal bypass system: is there a clinical benefit? Ann Thorac Surg 2005;80:238–44. [DOI] [PubMed] [Google Scholar]
- 26.Ohata T, Mitsuno M, Yamamura M, Tanaka H, Kobayashi Y, Ryomoto M et al. Minimal cardiopulmonary bypass attenuates neutrophil activation and cytokine release in coronary artery bypass grafting. J Artif Organs 2007;10:92–5. [DOI] [PubMed] [Google Scholar]
- 27.Puehler T, Haneya A, Philipp A, Wiebe K, Keyser A, Rupprecht L et al. Minimal extracorporeal circulation: an alternative for on-pump and off-pump coronary revascularization. Ann Thorac Surg 2009;87:766–72. [DOI] [PubMed] [Google Scholar]
- 28.Biancari F, Rimpilainen R. Meta-analysis of randomised trials comparing the effectiveness of miniaturised versus conventional cardiopulmonary bypass in adult cardiac surgery. Heart 2009;95:964–9. [DOI] [PubMed] [Google Scholar]
- 29.Liebold A, Khosravi A, Westphal B, Skrabal C, Choi YH, Stamm C et al. Effect of closed minimized cardiopulmonary bypass on cerebral tissue oxygenation and microembolization. J Thorac Cardiovasc Surg 2006;131:268–76. [DOI] [PubMed] [Google Scholar]
- 30.Zanatta P, Forti A, Minniti G, Comin A, Mazzarolo AP, Chilufya M et al. Brain emboli distribution and differentiation during cardiopulmonary bypass. J Cardiothorac Vasc Anesth 2013;27:865–75. [DOI] [PubMed] [Google Scholar]
- 31.Camboni D, Schmidt S, Philipp A, Rupprecht L, Haneya A, Puehler T et al. Microbubble activity in miniaturized and in conventional extracorporeal circulation. ASAIO J 2009;55:58–62. [DOI] [PubMed] [Google Scholar]
- 32.Anastasiadis K, Argiriadou H, Kosmidis MH, Megari K, Antonitsis P, Thomaidou E et al. Neurocognitive outcome after coronary artery bypass surgery using minimal versus conventional extracorporeal circulation: a randomised controlled pilot study. Heart 2011;97:1082–8. [DOI] [PubMed] [Google Scholar]
- 33.Reineke D, Winkler B, Konig T, Meszaros K, Sodeck G, Schönhoff F et al. Minimized extracorporeal circulation does not impair cognitive brain function after coronary artery bypass grafting. Interact CardioVasc Thorac Surg 2015;20:68–73. [DOI] [PubMed] [Google Scholar]
- 34.Gunaydin S, Sari T, McCusker K, Schonrock U, Zorlutuna Y. Clinical evaluation of minimized extracorporeal circulation in high-risk coronary revascularization: impact on air handling, inflammation, hemodilution and myocardial function. Perfusion 2009;24:153–62. [DOI] [PubMed] [Google Scholar]
- 35.Bennett M, Weatherall M, Webb G, Dudnikov S, Lloyd C. The impact of haemodilution and bypass pump flow on cerebral oxygen desaturation during cardiopulmonary bypass—a comparison of two systems of cardiopulmonary bypass. Perfusion 2014;30:389–94. [DOI] [PubMed] [Google Scholar]
- 36.Panday GF, Fischer S, Bauer A, Metz D, Schubel J, El Shouki N et al. Minimal extracorporeal circulation and off-pump compared to conventional cardiopulmonary bypass in coronary surgery. Interact CardioVasc Thorac Surg 2009;9:832–6. [DOI] [PubMed] [Google Scholar]
- 37.Remadi JP, Rakotoarivelo Z, Marticho P, Benamar A. Prospective randomized study comparing coronary artery bypass grafting with the new mini-extracorporeal circulation Jostra System or with a standard cardiopulmonary bypass. Am Heart J 2006;151:e1–198. [DOI] [PubMed] [Google Scholar]
- 38.Diez C, Haneya A, Brunger F, Philipp A, Hirt S, Ruppecht L et al. Minimized extracorporeal circulation cannot prevent acute kidney injury but attenuates early renal dysfunction after coronary bypass grafting. ASAIO J 2009;55:602–7. [DOI] [PubMed] [Google Scholar]
- 39.Huybregts RA, Morariu AM, Rakhorst G, Spiegelenberg SR, Romijn HW, de Vroege R et al. Attenuated renal and intestinal injury after use of a mini-cardiopulmonary bypass system. Ann Thorac Surg 2007;83:1760–6. [DOI] [PubMed] [Google Scholar]
- 40.Capuano F, Goracci M, Luciani R, Gentile G, Roscitano A, Benedetto U et al. Neutrophil gelatinase-associated lipocalin levels after use of mini-cardiopulmonary bypass system. Interact CardioVasc Thorac Surg 2009;9:797–801. [DOI] [PubMed] [Google Scholar]
- 41.Benedetto U, Luciani R, Goracci M, Capuano F, Refice S, Angeloni E et al. Miniaturized cardiopulmonary bypass and acute kidney injury in coronary artery bypass graft surgery. Ann Thorac Surg 2009;88:529–35. [DOI] [PubMed] [Google Scholar]
- 42.Bauer A, Diez C, Schubel J, El-Shouki N, Metz D, Eberle T et al. Evaluation of hemodynamic and regional tissue perfusion effects of minimized extracorporeal circulation (MECC). J Extra Corpor Technol 2010;42:30–9. [PMC free article] [PubMed] [Google Scholar]
- 43.Skrabal CA, Steinhoff G, Liebold A. Minimizing cardiopulmonary bypass attenuates myocardial damage after cardiac surgery. ASAIO J 2007;53:32–5. [DOI] [PubMed] [Google Scholar]
- 44.van Boven WJ, Gerritsen WB, Driessen AH, Morshuis WJ, Waanders FG, Haas FJ et al. Myocardial oxidative stress, and cell injury comparing three different techniques for coronary artery bypass grafting. Eur J Cardiothorac Surg 2008;34:969–75. [DOI] [PubMed] [Google Scholar]
- 45.Nguyen BA, Suleiman MS, Anderson JR, Evans PC, Fiorentino F, Reeves BC et al. Metabolic derangement and cardiac injury early after reperfusion following intermittent cross-clamp fibrillation in patients undergoing coronary artery bypass graft surgery using conventional or miniaturized cardiopulmonary bypass. Mol Cell Biochem 2014;395:167–75. [DOI] [PubMed] [Google Scholar]
- 46.van Boven WJ, Gerritsen WB, Driessen AH, van Dongen EP, Klautz RJ, Aarts LP. Minimised closed circuit coronary artery bypass grafting in the elderly is associated with lower levels of organ-specific biomarkers: a prospective randomised study. Eur J Anaesthesiol 2013;30:685–94. [DOI] [PubMed] [Google Scholar]
- 47.Prasser C, Abbady M, Keyl C, Liebold A, Tenderich M, Philipp A et al. Effect of a miniaturized extracorporeal circulation (MECC System) on liver function. Perfusion 2007;22:245–50. [DOI] [PubMed] [Google Scholar]
- 48.Donndorf P, Kuhn F, Vollmar B, Rösner J, Liebold A, Gierer P et al. Comparing microvascular alterations during minimal extracorporeal circulation and conventional cardiopulmonary bypass in coronary artery bypass graft surgery: a prospective, randomized study. J Thorac Cardiovasc Surg 2012;144:677–83. [DOI] [PubMed] [Google Scholar]
- 49.Haneya A, Philipp A, Schmid C, Diez C, Kobuch R, Hirt S et al. Minimised versus conventional cardiopulmonary bypass: outcome of high-risk patients. Eur J Cardiothorac Surg 2009;36:844–8. [DOI] [PubMed] [Google Scholar]
- 50.Kolat P, Ried M, Haneya A, Philipp A, Kobuch R, Hirt S et al. Impact of age on early outcome after coronary bypass graft surgery using minimized versus conventional extracorporeal circulation. J Cardiothorac Surg 2014;9:143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Ried M, Haneya A, Kolat P, Philipp A, Kobuch R, Hilker M et al. Emergency coronary artery bypass grafting using minimized versus standard extracorporeal circulation—a propensity score analysis. J Cardiothorac Surg 2013;8:59. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Koivisto SP, Wistbacka JO, Rimpilainen R, Nissinen J, Loponen P, Teittinen K et al. Miniaturized versus conventional cardiopulmonary bypass in high-risk patients undergoing coronary artery bypass surgery. Perfusion 2010;25:65–70. [DOI] [PubMed] [Google Scholar]
- 53.Anastasiadis K, Fragoulakis V, Antonitsis P, Maniadakis N. Coronary artery bypass grafting with minimal versus conventional extracorporeal circulation; an economic analysis. Int J Cardiol 2013;168:5336–43. [DOI] [PubMed] [Google Scholar]
- 54.Fernandes P, MacDonald J, Cleland A, Walsh G, Mayer R. What is optimal flow using a mini-bypass system? Perfusion 2010;25:133–7. [DOI] [PubMed] [Google Scholar]
- 55.Puehler T, Haneya A, Philipp A, Camboni D, Hirt S, Zink W et al. Minimized extracorporeal circulation in coronary artery bypass surgery is equivalent to standard extracorporeal circulation in patients with reduced left ventricular function. Thorac Cardiovasc Surg 2010;58:204–9. [DOI] [PubMed] [Google Scholar]
- 56.Ovrum E, Holen EA, Tangen G, Brosstad F, Abdelnoor M, Ringdal MA et al. Completely heparinized cardiopulmonary bypass and reduced systemic heparin: clinical and hemostatic effects. Ann Thorac Surg 1995;60:365–71. [DOI] [PubMed] [Google Scholar]
- 57.Despotis GJ, Joist JH, Hogue CW Jr, Alsoufiev A, Kater K, Goodnough LT et al. The impact of heparin concentration and activated clotting time monitoring on blood conservation. A prospective, randomized evaluation in patients undergoing cardiac operation. J Thorac Cardiovasc Surg 1995;110:46–54. [DOI] [PubMed] [Google Scholar]
- 58.Flachskampf FA, Badano L, Daniel WG, Feneck RO, Fox KF, Fraser AG et al. Recommendations for transoesophageal echocardiography: update 2010. Eur J Echocardiogr 2010;11:557–76. [DOI] [PubMed] [Google Scholar]
- 59.Hahn RT, Abraham T, Adams MS, Bruce CJ, Glas KE, Lang RM et al. Guidelines for performing a comprehensive transesophageal echocardiographic examination: recommendations from the American Society of Echocardiography and the Society of Cardiovascular Anesthesiologists. Anesth Analg 2014;118:21–68. [DOI] [PubMed] [Google Scholar]
- 60.Barry AE, Chaney MA, London MJ. Anesthetic management during cardiopulmonary bypass: a systematic review. Anesth Analg 2015;120:749–69. [DOI] [PubMed] [Google Scholar]
- 61.Pagel PS. Myocardial protection by volatile anesthetics in patients undergoing cardiac surgery: a critical review of the laboratory and clinical evidence. J Cardiothorac Vasc Anesth 2013;27:972–82. [DOI] [PubMed] [Google Scholar]
- 62.Tutschka MP, Bainbridge D, Chu MW, Kiaii B, Jones PM. Unilateral postoperative pulmonary edema after minimally invasive cardiac surgical procedures: a case-control study. Ann Thorac Surg 2015;99:115–22. [DOI] [PubMed] [Google Scholar]
- 63.Anastasiadis K, Antonitsis P, Argiriadou H (eds.) Principles of Miniaturized ExtraCorporeal Circulation. Berlin Heidelberg: Springer-Verlag, 2013, 63–71. [Google Scholar]
- 64.Hofmann B, Bushnaq H, Kraus FB, Raspe C, Simm A, Silber RE et al. Immediate effects of individualized heparin and protamine management on hemostatic activation and platelet function in adult patients undergoing cardiac surgery with tranexamic acid antifibrinolytic therapy. Perfusion 2013;28:412–8. [DOI] [PubMed] [Google Scholar]
- 65.Ranucci M, Romitti F, Isgro G, Cotza M, Brozzi S, Boncilli A et al. Oxygen delivery during cardiopulmonary bypass and acute renal failure after coronary operations. Ann Thorac Surg 2005;80:2213–20. [DOI] [PubMed] [Google Scholar]
- 66.Anastasiadis K, Antonitsis P, Argiriadou H (eds.) Principles of Miniaturized ExtraCorporeal Circulation. Berlin Heidelberg: Springer-Verlag Eds, 2013, 73–99. [Google Scholar]
- 67.Clive Landis R, Murkin JM, Stump DA, Baker RA, Arrowsmith JE, De Somer F et al. Consensus statement: minimal criteria for reporting the systemic inflammatory response to cardiopulmonary bypass. Heart Surg Forum 2010;13:E116–23. [DOI] [PubMed] [Google Scholar]
- 68.Liebold A, Reisinger S, Lehle K, Rupprecht L, Philipp A, Birnbaum DE. Reduced invasiveness of perfusion with a minimized extracorporeal circuit (the Jostra MECC System). Thorac Cardiovasc Surg 2002;50(Supp 1):S75. [Google Scholar]
- 69.Murkin JM. Attenuation of neurologic injury during cardiac surgery. Ann Thorac Surg 2001;72:S1838–44. [DOI] [PubMed] [Google Scholar]
- 70.Murkin JM. Cerebral oximetry: monitoring the brain as the index organ. Anesthesiology 2011;114:12–3. [DOI] [PubMed] [Google Scholar]
- 71.Slater JP, Guarino T, Stack J, Vinod K, Bustami RT, Brown JM III et al. Cerebral oxygen desaturation predicts cognitive decline and longer hospital stay after cardiac surgery. Ann Thorac Surg 2009;87:36–44. [DOI] [PubMed] [Google Scholar]
- 72.Murkin JM, Adams SJ, Novick RJ, Quantz M, Bainbridge D, Iglesias I et al. Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study. Anesth Analg 2007;104:51–8. [DOI] [PubMed] [Google Scholar]
- 73.de Somer F, Mulholland JW, Bryan MR, Aloisio T, Van Nooten GJ, Ranucci M. O2 delivery and CO2 production during cardiopulmonary bypass as determinants of acute kidney injury: time for a goal-directed perfusion management? Crit Care 2011;15:R192. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Murkin JM. Is it better to shine a light, or rather to curse the darkness? Cerebral near-infrared spectroscopy and cardiac surgery. Eur J Cardiothorac Surg 2013;43:1081–3. [DOI] [PubMed] [Google Scholar]
- 75.Hutton B, Joseph L, Fergusson D, Mazer CD, Shapiro S, Tinmouth A. Risks of harms using antifibrinolytics in cardiac surgery: systematic review and network meta-analysis of randomised and observational studies. BMJ 2012;345:e5798. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Gorlinger K, Shore-Lesserson L, Dirkmann D, Hanke AA, Rahe-Meyer N, Tanaka KA. Management of hemorrhage in cardiothoracic surgery. J Cardiothorac Vasc Anesth 2013;27(Supp 1):S20–34. [DOI] [PubMed] [Google Scholar]
- 77.Nollert G, Schwabenland I, Maktav D, Kur F, Christ F, Fraunberger P et al. Miniaturized cardiopulmonary bypass in coronary artery bypass surgery: marginal impact on inflammation and coagulation but loss of safety margins. Ann Thorac Surg 2005;80:2326–32. [DOI] [PubMed] [Google Scholar]
- 78.Aboud A, Liebing K, Börgermann J, Ensminger S, Zittermann A, Renner A et al. Excessive negative venous line pressures and increased arterial air bubble counts during miniaturized cardiopulmonary bypass: an experimental study comparing miniaturized with conventional perfusion systems. Eur J Cardiothorac Surg 2014;45:69–74. [DOI] [PubMed] [Google Scholar]
- 79.Bauer A, Schaarschmidt J, Anastasiadis K, Carrel T. Reduced amount of gaseous microemboli in the arterial line of minimized extracorporeal circulation systems compared with conventional extracorporeal circulation. Eur J Cardiothorac Surg 2014;46:152. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.

