Abstract
An improved unroofing method enabled the cantilever of an atomic force microscope (AFM) to reach directly into a cell to visualize the intracellular cytoskeletal actin filaments, microtubules, clathrin coats, and caveolae in phosphate-buffered saline (PBS) at a higher resolution than conventional electron microscopy. All of the actin filaments clearly exhibited a short periodicity of approximately 5–6 nm, which was derived from globular actins linked to each other to form filaments, as well as a long helical periodicity. The polarity of the actin filaments appeared to be determined by the shape of the periodic striations. Microtubules were identified based on their thickness. Clathrin coats and caveolae were observed on the cytoplasmic surface of cell membranes. The area containing clathrin molecules and their terminal domains was directly visualized. Characteristic ridge structures located at the surface of the caveolae were observed at high resolution, similar to those observed with electron microscopy (EM). Overall, unroofing allowed intracellular AFM imaging in a liquid environment with a level of quality equivalent or superior to that of EM. Thus, AFMs are anticipated to provide cutting-edge findings in cell biology and histology.
Binnig et al.1 first outlined the principle of the atomic force microscope (AFM) as a type of scanning probe microscopy. The AFM imaging method is fundamentally different from that of other microscopes. A sharp needle attached to a cantilever (the AFM probe) scans over the surface of a sample, and the topological structure is displayed based on measurements of the surface undulation. Subsequent technical developments by many investigators have revealed a wide range of biological applications for the AFM (see reviews2,3). Thus, the AFM has become a powerful tool to detect the surface structure of biological samples in liquid environments at a theoretical resolution comparable to that of electron microscopy (EM).
Unfortunately, AFM cantilevers cannot stably follow sudden changes in vertical undulations. Furthermore, several scanning instabilities originating from the surrounding liquid environment have limited the general use of atomic force microscopy in cell biology. In addition, an AFM has not been able to image intracellular structures because the cantilever cannot scan the inside of the cell membrane. Therefore, in vivo structural analysis using an AFM has been restricted to the outer surface of cell membranes. Indeed, even in their present state, AFM measurements in cells have not progressed very much, despite remarkable improvements in scanning speed and stability. The use of a high-resolution AFM has allowed the structural analysis of isolated and purified proteins and their complexes in vitro4,5,6,7,8,9, but not the in vivo or in situ analysis of intracellular structures. Most related atomic force microscopy studies have only detected the structure of the outer surface of cell membranes in a living or fixed state.
Recently, an unroofing technique has provided a way to view intracellular structures with an AFM10. Unroofing refers to breaking the cell membrane. When the cell membrane is broken, the intracellular region is exposed by removing the cytoplasmic-soluble components, which allows the AFM cantilever to access intracellular structures such as the cytoskeleton and organelles. This preparation technique, which uses sonication, was originally developed to observe membrane undercoats in freeze-etching EM11,12. Although rupture of the apical cell membrane is caused by the collision of micro air bubbles (cavitation) induced by sonication, the specimen manipulation for this technique requires meticulous care to prevent the detachment of whole cells from the coverslips. The current study presents a new unroofing apparatus consisting of an improved sonicator and light stereomicroscope that improves reproducibility and facilitates sample handling. The improved (custom-made) sonicator is able to generate cavitation at a much lower power (0.3 − 1 W) than that of the commonly used sonicator (50 W or higher). We used this method in an atomic force microscopy analysis of the intracellular cytoskeleton in phosphate-buffered saline (PBS) at a molecular resolution. Furthermore, to improve the signal-to-noise (S/N) ratio and thereby refine the images obtained with the AFM, serial scanned images of a target organelle were averaged using mathematical calculations.
Results
Unroofing is a very useful method to expose the inside of cells and thereby enable intracellular observations via various microscopic techniques, but there are currently few methods that are easy to use with good reproducibility. Although chemical unroofing methods using detergents such as Triton X 100 are easy to use, interactions between the cytoskeleton and membrane are never observed because the membrane and organelles are completely dissolved. A membrane-breaking method using changes in osmotic pressure was inefficient and did not work in some cells in our experience. In an earlier study13, the authors were not able to visualize the cytoskeleton and organelles inside the cell at high resolution. In contrast, the unroofing method using sonication is particularly good for freeze-etching EM11,12. However, the reproducibility is not good because whole cells can frequently be detached from the glass surface during sonication under the original conditions (10 to 50 W). Even if the cell membrane remains on the glass after unroofing, many of the cytoskeletal components and organelles are completely removed. Therefore, we developed a reliable unroofing device. First, an ultrasonic generator was improved to precisely regulate the output power in the range of 0 − 1 W. This custom-made low-power sonicator was further combined with a stereomicroscope equipped with a position controller for the sonication probe and an LED light-sheet illumination system to observe the unroofing procedure in detail (Fig. 1B). After these modifications, this apparatus enabled us to unroof cells to varying degrees, from the partial to entire removal of the cytoplasm (Fig. 1C).
The cantilever was thus able to directly reach the cell interior. Unlike conventional AFM measurements from outside the cell membrane, single actin filaments, bundles of actin filaments (stress fibres), clathrin coats, caveolae and so on were observed at the cytoplasmic surface of the cell membrane (Fig. 2A), yielding images that were similar to the those obtained via freeze-etching EM. Upon increasing the magnification, each actin filament in the stress fibre was discernible, revealing a short periodicity derived from globular actin (G-actin) that assembled to form filaments (Fig. 2B). This periodicity varied from 4.5 to 6 nm along the long axis of a single actin filament, with an average interval of 5.4 nm, as obtained by Fourier transformation. As depicted in Fig. 2A, it should be noted that these structures are displayed at an obviously high resolution that is comparable to EM, although the measurements were performed in a liquid environment (PBS). With careful observation, the periodic striation appeared to be slightly curved because the two G-actin molecules that formed each striation were oriented at different angles. The striation became more evident upon averaging the images (Fig. 2C,D). Based on the direction of the curve, the polarity of the actin filaments could be determined. To assess whether the slightly curved shape of the short periodic striations in the actin filaments is a real structure and whether the curved form virtually reflects the molecular structure of the actin filament, we first observed actin filaments that were polymerized from purified G-actin14 and then simulated how the actin filaments should be displayed by an AFM15,16. The in vitro polymerized actin filaments had the same structure as those observed in situ (Fig. 3A). This result confirms that the AFM measurements in this study provided accurate structural information about the actin filaments. In addition, the simulated image was quite similar to that obtained experimentally, as shown in Fig. 3B. Namely, it is evident that the curved shape of the periodic striation reflects the molecular structure of the actin filaments. When comparing the simulated image with the molecular model, the concave side of the curved striation appears to correspond to the barbed end. Therefore, the polarity of the actin filament could be determined in situ when the shape of the striation appeared curvilinear at high magnification. We tentatively marked the polarity in Fig. 2C,D according to this judgment. Further experiments may be required to confirm the polarity in a more concrete way. However, one of the co-authors, Narita, and colleagues17 succeeded in determining the polarity of actin filaments in a similar manner using EM.
In contrast, the thickness of the actin filaments detected with the AFM was variable and ranged from 5 nm to 15 nm. This may be partly because the thickness of the filaments is modified by the diameter of the tip of the scanning probe and reduced by the position of the probe relative to the sample. Therefore, an AFM may be not as reliable in measuring actin filament thickness.
Microtubules with a diameter of approximately 24 nm were observed to accompany actin filaments or stress fibres in the cytoplasm (Fig. 4). Unfortunately, proto-filaments were not easily detected in situ. This is partly because the microtubules in the cytoplasm are modified with several microtubule-associated proteins (Fig. 4). When the surface of a microtubule was scratched mechanically with the scanning tip, longitudinal striations resembling proto-filaments became visible (Fig. 4). However, the actual structure of microtubules in a liquid environment requires future study.
Clathrin coats and caveolae were frequently detected on the cytoplasmic surface of cell membranes, as revealed in Fig. 2A. Clathrin coats were relatively stable against repeated probe scanning in PBS due to the characteristic hexagonal lattice structure of these features. Therefore, the portion of the clathrin coats that contained the triskelion shapes of the clathrin molecules and their terminal domains could be observed at higher magnification (Fig. 5A). The visible region that could be traced seems to be the proximal segment of the heavy chain based on the molecular model produced by Fotin et al.18. Because the terminal regions are located in the centre of the hexagonal meshwork in this model, some blobs observed in the centre of the hexagonal mesh are thought to be the terminal regions of the clathrin molecules. The quality of the AFM images seems to be better than images of freeze-etched clathrin coats that have been taken to date using EM. We also averaged three consecutively scanned images (Fig. 5B1, 2, 3) to enhance the details of the clathrin coats, as described in Fig. 5C. The cytoplasmic surface of the cell membrane also contained caveolae and sometimes clusters of caveolae (Fig. 6A), which are easily identified by their characteristic ridge structures on the surface19,20 (see supplemental Fig. S1). Clearly, the ridge structures on the surface of the caveolae visualized with the AFM at high magnification (Fig. 6B) are similar or superior to those visualized using freeze-etching EM, even though the images were obtained in a liquid environment.
As mentioned above, conventional sonication (10 W or more) results in the loss of many whole cells. Even if cells are unroofed successfully, neither the organelles nor nuclei would be attached. Only a few cytoskeletons are attached on the ventral cell membrane that remains on the glass. Although the ablation of the cytoplasm by unroofing is not constant even at the same sonication power, sonication affects not only the amount of remaining cytoskeleton but also appears to change the types of remaining organelles. Indeed, mild unroofing revealed the existence of membranous structure just beneath the cell membrane (Figs 2A and 6A). It should be noted that this membrane system was located over the caveolar region where tubular networks are formed (Fig. 6A). This is very interesting when considering the function of the membrane structure. However, the precise identification of this structure remains for future work.
Discussion
In cell biology, in vivo AFM imaging has long been limited to surface observations of living or fixed cells at a slightly higher resolution than that of light microscopes. There has been a large gap between the theoretical resolution and biological applications of AFM, partly because there have been no appropriate sample preparation methods for atomic force microscopy. Many studies have investigated cortical actin filaments with an AFM, but unfortunately, these studies involved indirect observations through cell membranes21,22,23,24,25,26,27,28,29,30,31,32,33,34. In these studies, bundles of actin filaments (stress fibres) were only observed as several linear disturbances or ridges on the surface of cells. Therefore, stress fibres were suggested to be producing the images based on comparisons with fluorescence microscopy data and/or the measurement of the width of the linear disturbances. Considering that the cortical cytoskeleton consists of actin filaments, microtubules and intermediate filaments, it is impossible to identify individual actin filaments through the cell membrane. In this regard, several studies used a detergent such as Triton X to remove the cell membrane and successfully observed intracellular structures35. However, the detergent completely dissolved the membrane system, including the organelles; therefore, the membrane undercoat, clathrin coats, caveolae, and the interactions between the cytoskeleton and cell membrane were never observed. The unroofing method makes it possible for the AFM cantilever to directly access intracellular organelles, while maintaining the ventral cell membrane in its native state. Indeed, actin filaments, clathrin coats and other structures could be detected at the expected resolution. In practice, the short periodicity of the actin filaments along the long axis was observed more clearly with the AFM than by freeze-etching EM because the sharp tip of the cantilever traces the surfaces of the actin filaments while in contact with these structures during atomic force microscopy. In contrast, the surfaces of actin filaments are evaporated with platinum and carbon in a freeze-etching replica in EM. This is why an AFM can clearly describe the short periodicity of actin filaments. However, determination of the thickness of the filaments may be not reliable because the tip diameter and the position of the cantilever needle in the z axis changes the apparent thickness of the filaments (see the illustration in Fig. 7). With a freeze-etching replica in EM, optimal shadowing is necessary to visualize the periodicity of actin filaments. Even if the periodicity of the actin filaments became apparent, their polarity could not be detected without myosin S1 decoration because metal shadowing covered the slight curvature of the periodic striation. This is also true for the triskelion shape of the clathrin molecules. Clearly, the ability of the AFM to observe this fine structure is an advantage over EM.
Atomic force microscopy has not been able to depict intracellular structures at theoretical resolution until the removal of the apical cell membrane by unroofing techniques. In practice, in situ images at the same resolution as those in this study have not previously been taken with any type of microscope. Because the current images were obtained in PBS, the structures that contain water molecules can be imaged. Therefore, the current results are valuable for understanding the actual structure of cytoskeletons and organelles in cells, revealing the utility of the developed unroofing technique for future studies in cell biology.
However, it is not simple to regulate the degree of unroofing to visualize the surface of intracellular organelles. The unroofing technique was originally developed to observe the cytoplasmic surface of membranes (the membrane undercoat) in freeze-etching replica EM10,11. Therefore, relatively strong sonication might be acceptable in certain cases because the complete removal of soluble proteins was necessary to reveal membrane-attached structures such as clathrin coats. Commercial sonicators are typically manufactured for homogenization or cleaning; therefore, their output power is usually approximately 50 W or higher at a frequency of 27 KHz. Under such conditions, whole cells are detached from the glass surfaces and fragmented. Therefore, it is necessary to decrease the power considerably to promote effective unroofing. For this purpose, we developed with the assistance of Tomy Seiko (Tokyo, Japan) a sonication device that is capable of precisely regulating the sonication power from 0 to 1 W at a frequency of 27 KHz (custom-made, Patent pending). During our study, it became clear that membrane rupture is induced by micro bubbles (cavitation), particularly at lower power. The tip of the sonication probe was scratched to form a shallow groove on the surface to produce sufficient cavitation.
A reduction of the output power also seemed to decrease the artefacts of unroofing, including breaking the organelle membranes, removing too much cytoskeleton and fragmentation. Moreover, the unroofing artefacts produced by low-power sonication appear to be less than those produced by chemical unroofing, such as Triton X treatment or biochemical isolation procedures. Low-power sonication is similar to sectioning, although the soluble substances are washed away upon the disruption of the cell membrane. This type of unroofing could be used as a preparation method in freeze-etching EM, cryo-EM, SEM and light microscopic immunocytochemistry. Whereas unroofing has been successfully used for these microscopy techniques, we will continue to validate its use for AFM studies to avoid any potential artifacts. Low-power unroofing allowed us to image fragile membrane structures in this study, although the structures were still fragmented, as indicated in Fig. 2A (asterisk and area painted yellow) and Fig. 6A (asterisks). We assumed that the fragmented membrane structures possible represented smooth endoplasmic reticulum (smooth ER). Their precise identification requires future study. Although the smooth ER is well developed in muscle cells (Ca2+ regulation) and adrenal cells (steroid synthesis), the spatial localization of the smooth ER in non-specialized cells is not well known at present. In these tissues, the smooth ER is characterized by a very flat, thin and tubular meshwork. Sometimes, this membrane structure was observed just beneath the cell membrane in cryo-EM and freeze-etching EM (see supplemental Fig. S2).
Recently, mathematical averaging of images has become popular in single-particle analyses of purified proteins by negative staining or cryo-EM. In the case of in vitro imaging using purified proteins, there are no problems with averaging the images of different particles (proteins). However, many proteins, particularly filaments, are modified in vivo with several proteins to function properly at different locations in cells. Therefore, averaging the images of target proteins localized at different places within a cell results in the loss of minor structures or the actual structure. In this study, we applied an averaging method to consecutive, scanned images of the same filament or structure. This improved the S/N ratio of the images such that the short periodicity of the actin filaments was refined and useful to determine the polarity of the filaments. The images averaged from three consecutively scanned images of clathrin coats were also well refined compared with the raw images. Therefore, averaging scanned images seems to be effective for structural analysis.
Although freeze-etching EM is a unique method for describing three-dimensional cytoskeletons, the images reveal a surface topology similar to atomic force microscopy because of the necessary shadowing with platinum and carbon. The image resolution is limited by the amount of shadowing and does not equal the resolution of EM. In contrast, the scanning tip of an AFM is able to trace the surface undulation laterally at the theoretical resolution, which is limited only by vertical instability. If the samples are prepared appropriately, for example via unroofing as shown in this study, an AFM is able to describe intracellular structures at molecular resolution. However, measured values, such as the filament thickness, depend on the relative position of the scanning tip to the sample in addition to the tip radius, as illustrated in Fig. 7. Filaments located in the uppermost layer appeared thicker than their true thickness, while filaments deeper in the sample appeared thinner because of partial tracing. However, individual actin filaments were discernible in stress fibres, as illustrated in Fig. 2B. Another major disadvantage of the AFM is that its findings are not applicable to those of immunocytochemistry. Therefore, we cannot identify the constituent substances in the structure. In freeze-etching EM, some actin filaments appeared to be fused firmly with each other in the stress fibre, which is attributed to shadowing and/or the differences in the observation environments: freeze-drying for EM and a liquid environment for AFM. Overall, the developed unroofing sample preparation approach allows different imaging techniques to provide complementary data on biological samples.
Methods
Cell culture
Glass slides containing hydrophilic circular windows with a diameter of 15 mm surrounded by a framework of printed hydrophobic black ink (TF0215; Matsunami Glass Industry, Ltd., Osaka, Japan) were used for both cell culture and the AFM measurements. When culture medium (D-MEM: Sigma-Aldrich Co. St. Louise, MO, USA) was added to this hydrophilic circular window, the medium produced droplet because of the hydrophobicity of the surrounding framework, as depicted in Fig. 1A. Normal rat kidney (NRK) cells were cultured in this swelled medium for 2 days in a CO2 incubator and then used in the subsequent experiments. To move the cantilever exactly onto the target area in the cells, a thin sapphire disc (3 mm in diameter, 0.05 mm thick; Rudolf Brügger, Swiss Micro Technology, Lugano, Switzerland) with coordinates was glued in the centre of each hydrophilic circle prior to cell culture. The coordinates on the sapphire discs were printed by carbon shadowing after an EM finder grid was placed on the disc.
Unroofing
To observe the intracellular cytoskeleton and the other organelles, the following unroofing method that used sonication was performed. Cells cultured on glass slides were sequentially washed with Ringer’s solution consisting of 155 mM NaCl, 3 mM KCl, 2 mM CaCl2, 1 mM MgCl2, 3 mM NaH2PO4, and 10 mM glucose in 5 mM HEPES buffer (pH 7.4) followed by Ca2+-free Ringer’s solution (lacking CaCl2). The cells were soaked for approximately 10 s in a polylysine solution (Mw 30,000–50,000; Sigma-Aldrich, St. Louis, MO, USA; 0.5 mg/ml dissolved in Ca2+-free Ringer’s solution) and then washed three times for a few seconds each in hypotonic Ringer’s solution, which was prepared by mixing one part Ca2+-free Ringer’s solution with two parts distilled water. This step swelled the cells, enabling them to burst easily during sonication. Immediately after washing with the hypotonic solution, the samples were placed approximately 2 mm from the tip of the sonicator probe and exposed for 5–10 s to fine air bubbles (cavitation) generated by weak sonication (0.5 W, 27 KHz) in isotonic buffer11,12 consisting of 30 mM HEPES, pH 7.4, 70 mM KCl, 3 mM MgCl2, 1 mM EGTA, 1 mM DTT and 0.1 mM 4-(2-aminoethyl) benzene sulfonyl fluoride hydrochloride (AEBSF, Pefabloc SC: Roche Diagnostics Gmbh, Mannheim Germany) as a protease inhibitor. The unroofed cells were washed briefly in the same fresh buffer and immediately fixed with 2% glutaraldehyde in the same buffer for 20 min. The fixed samples were washed twice with PBS and then examined with the AFM. The ultrasonic generator (Tomy Seiko Co., Tokyo, Japan) was custom-made to regulate the power in the range between 0 to 5 W and assembled with a stereo-light-microscope that was specifically designed (Olympus Co., Tokyo, Japan) to observe the unroofing process with light-sheet illumination. Light-sheet or low-angle illumination (nearly 0°) is important to clearly view both the micro bubbles generated by sonication and the cells attached to the glass slide. The final instrument is described as an unroofing apparatus and is depicted in Fig. 1B (Patent pending). In this study, the sonication power was 0.3 to 0.5 W.
This apparatus enabled us to observe various intracellular organelles by changing the degree of unroofing. The condition of the unroofed cells was checked with a phase-contrast light microscope. Partially and/or well-unroofed cells are presented in Fig. 1C.
AFM measurements
The glass slide containing unroofed cells soaked in PBS was assembled in a tip-scan high-speed AFM imaging system that was improved based on a previously developed AFM38. This AFM unit was combined with an inverted fluorescence microscope (IX73 Olympus Corporation, Tokyo) and named BIXAM. Therefore, the probe cantilever was depressed slowly onto the target region of the cells while observing the cells under a light microscope. The AFM was used in the phase-modulation mode for high-resolution imaging39. A characteristic of this AFM is to adopt small and soft cantilevers (2 μm wide, 9 μm long, and 0.1 μm thick in size) with a spring constant of 0.1 N/m (BL-AC10DS, BL-AC10EGS, Olympus Corporation). The resonance frequency of the probe tip is 1.2 MHz in air and approximately 400 kHz in water. The diameter of the scanning tip was approximately 8–10 nm. This AFM (BIXAM) acquires a sample image with a characteristic cantilever in the phase modulation mode in an amplitude domain of approximately 1–5 nm. When image measurements were performed in the phase modulation mode, it is unreasonable to estimate the set force. However, provided that the probe amplitude is approximately 5 nm in water, and it is decreased by approximately 10% upon the interaction between the probe and sample, the force against samples is estimated to be 50 pN. The maximum scanning range of the device was 4 μm (x axis) × 3 μm (y axis), but the range of 200–600 nm was primarily scanned in the present study. The scanning resolution was 320 pixels (x axis) × 240 lines (y axis).
Image analysis
To improve the S/N ratio of some images, computer-assisted averaging was examined for the first time in atomic force microscopy. More than three sequential frames taken at the region of interest were averaged, while considering drifting and rotation. EOS software was used for the averaging15. To obtain detailed information from the AFM images by enhancing the high-resolution signal by more than 18 nm, some images were multiplied by a filter in Fourier space (Fig. 2D Inset).
AFM observation of purified actin
Actin was purified from rabbit skeletal muscle as previously described14. Actin (0.1 mg/ml) was incubated for 30 min at room temperature in a solution containing 50 mM NaCl, 10 mM sodium phosphate buffer (pH 7.4), 1 mM MgCl2 and 1 mM DTT. To increase its affinity for the actin filament, a glass slide (TF0215; Matsunami Glass Ind., Ltd., Osaka, Japan) was coated with polylysine as described below. The glass slide was washed twice with ethanol, a polylysine solution (Mw 30,000–50,000; Sigma-Aldrich, St. Louis, MO, USA; 0.4 mg/ml dissolved in water) was applied, and then the slide was washed three times with the buffer. Polymerized actin was diluted 10 times with buffer containing 1 μM phalloidin and applied to the polylysine-coated glass slide. The AFM measurement was performed as described above.
Generating the simulated images
EOS software was used again for the image simulation15. The actin filament model16 was placed along the y axis in the computer. As a model for the AFM tip, a sphere with a diameter of 8 nm was generated at an (x, y, z) coordinate. The initial z value was larger than the maximum z position in the actin filament model by 8 nm. Then, z was decreased until the sphere touched the actin filament model. The final z was plotted as a pixel value at (x, y) in the simulated image. When 0.5% of the volume of the sphere overlapped with the actin filament model, we considered that the sphere touched the actin filament. The radius of the sphere was determined such that the simulated images resembled the observed images.
Additional Information
How to cite this article: Usukura, E. et al. An Unroofing Method to Observe the Cytoskeleton Directly at Molecular Resolution Using Atomic Force Microscopy. Sci. Rep. 6, 27472; doi: 10.1038/srep27472 (2016).
Supplementary Material
Acknowledgments
The reported results were obtained as part of the JST-SENTAN program (J.U.). Beginning in April 2015, the AMED-SENTAN program has promoted this development project (15hm0102003h0004) (J.U.), and it was also supported in part by a Grant-in-Aid for Challenging Exploratory Research (26650049) from JSPS (J.U.) and JST PREST (A.N.). We also thank Mr. Nobuaki Sakai, Mr. Kiyohiko Tateyama, Mr. Yoshitsugu Uekusa and Ms. Yuka Imaoka (Olympus Corporation) for their assistance with the experiments.
Footnotes
The authors declare no competing financial interests.
Author Contributions E.U. prepared most of the samples and made the figures. A.N. prepared the purified actin samples and oversaw the image analysis. S.I. and A.Y. skilfully operated the AFM apparatus. J.U. managed the research projects and largely wrote the paper. All the authors observed samples in practice and reviewed the paper critically.
References
- Binnig G., Quate C. F. & Gerber C. Atomic force microscope . Phys. Rev. Let. 56, 930–934 (1986). [DOI] [PubMed] [Google Scholar]
- Goldsbury C. & Scheuring S. Introduction to atomic force microscopy (AFM) in biology. Curr. Protoc. Protein Sci. Chapter 17, Unit 17.7 1–19 (2009). [DOI] [PubMed] [Google Scholar]
- Hamon L., Curmi P. A. & Pastre D. High-resolution imaging of microtubules and cytoskeleton structures by atomic force microscopy. Methods Cell Biol. 95, 157–174 (2010). [DOI] [PubMed] [Google Scholar]
- Yamamoto S. et al. Observation of human corneal and scleral collagen fibrils by atomic force microscopy. Jap. J. Ophth. 46, 496–501 (2002). [DOI] [PubMed] [Google Scholar]
- Schmitz S. et al. Malaria parasite actin polymerization and filament structure. J. Biol. Chem. 285, 36577–36585 (2010). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kodera N., Yamamoto D., Ishikawa R. & Ando T. Video imaging of walking myosin V by high-speed atomic force microscopy. Nature 468, 72–76 (2010). [DOI] [PubMed] [Google Scholar]
- Sharma S. et al. Atomic force microscopy reveals drebrin induced remodeling of F-actin with subnanometer resolution. Nano Lett. 11, 825–827 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schaap I. A., Carrasco C., de Pablo P. J. & Schmidt C. F. Kinesin walks the line: single motors observed by atomic force microscopy. Biophys. J. 100, 2450–2456 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ando T. High speed atomic force microscopy coming of age. Nanotechnol. 23, 1–27 (2012). [DOI] [PubMed] [Google Scholar]
- Usukura J. et al. Use of the unroofing technique for atomic force microscopic imaging of the intra-cellular cytoskeleton under aqueous conditions. J. Electron Microsc. 61, 321–326 (2012). [DOI] [PubMed] [Google Scholar]
- Heuser J. E. & Anderson R. G. Hypertonic media inhibit receptor-mediated endocytosis by blocking clathrin-coated pit formation. J. Cell Biol. 108, 389–400 (1989). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Heuser J. E. The origins and evolution of freeze-etch electron microscopy. J. Electron Microsc. 60, S3–29 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dunlap D., Cattelino A., de Curtis I. & Vallorta F. Cytoplasmic topography of focal contacts. FEBS Lett. 382, 65–72 (1996). [DOI] [PubMed] [Google Scholar]
- Spudich J. A. & Watt S. The regulation of rabbit skeletal muscle contraction. I. Biochemical studies of the interaction of the tropomyosin-troponin complex with actin and the proteolytic fragments of myosin. J. Biol. Chem. 246, 4866–4871 (1971). [PubMed] [Google Scholar]
- Yasunaga T. & Wakabayashi T. Extensible and object-oriented system Eos supplies a new environment for image analysis of electron micrographs of macromolecules. J. Struct. Biol. 116, 155–60 (1996). [DOI] [PubMed] [Google Scholar]
- Oda T. et al. The nature of the globular- to fibrous-actin transition. Nature 457, 441–445 (2009). [DOI] [PubMed] [Google Scholar]
- Narita A. et al. Direct determination of actin polarity in the cell. J. Mol. Biol. 419, 359–368 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fotin A. et al. Molecular model for a complete clathrin lattice from electron cryomicroscopy. Nature 432, 573–579 (2004). [DOI] [PubMed] [Google Scholar]
- Anderson R. G. The caveolae membrane system. Annu. Rev. Biochem. 67, 199–225 (1998). [DOI] [PubMed] [Google Scholar]
- Stan R. V. Structure of caveolae. Biochim. Biophy. Acta 1746, 334–348 (2005). [DOI] [PubMed] [Google Scholar]
- Henderson E., Haydon P. G. & Sakaguchi D. S. Actin filament dynamics in living glial cells imaged by atomic force microscopy. Science 257, 1944–1946 (1992). [DOI] [PubMed] [Google Scholar]
- Braet F., Seynaeve C., De Zanger R. & Wisse E. Imaging surface and submembranous structures with the atomic force microscope: a study on living cancer cells, fibroblasts and macrophages. J. Microsc. 190, 328–338 (1998). [DOI] [PubMed] [Google Scholar]
- Nagao E. & Dvorak J. A. An integrated approach to the study of living cells by atomic force microscopy. J. Microsc. 191, 8–19 (1998). [DOI] [PubMed] [Google Scholar]
- Rotsch C. & Radmacher M. Drug-induced changes of cytoskeletal structure and mechanics in fibroblasts: an atomic force microscopy study. Biophys. J. 78, 520–535 (2000). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mathur A. B., Truskey G. A. & Reichert W. M. Atomic force and total internal reflection fluorescence microscopy for the study of force transmission in endothelial cells. Biophys. J. 78, 1725–1735 (2000). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dvorak J. A. The application of atomic force microscopy to the study of living vertebrate cells in culture. Methods 29, 86–96 (2003). [DOI] [PubMed] [Google Scholar]
- Madl J. et al. A combined optical and atomic force microscope for live cell investigations. Ultramicroscopy 106, 645–651 (2006). [DOI] [PubMed] [Google Scholar]
- Kellermayer M. S. Z. et al. Spatially and temporally synchronized atomic force and total internal reflection fluorescence microscopy for imaging and manipulating cells and biomolecules. Biophys. J. 91, 2665–2677 (2006). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Xiong Y. et al. Topography and nanomechanics of live neuronal growth cones analyzed by atomic force microscopy. Biophys. J. 96, 5060–5072 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Doak S. H. et al. High-resolution imaging using a novel atomic force microscope and confocal laser scanning microscope hybrid instrument: essential sample preparation aspects. Histochem. Cell Biol. 130, 909–916 (2008). [DOI] [PubMed] [Google Scholar]
- Iscru D. F., Anghelina M., Agarwal S. & Agarwal G. Changes in surface topologies of chondrocytes subjected to mechanical forces: an AFM analysis. J. Struct. Biol. 162, 397–403 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chtcheglova L. A. et al. AFM functional imaging on vascular endothelial cells. J. Mol. Recognit. 23, 589–596 (2010). [DOI] [PubMed] [Google Scholar]
- Pogoda K. et al. Depth-sensing analysis of cytoskeleton organization based on AFM data. Eur. Biophys. J. 41, 79–87 (2012). [DOI] [PubMed] [Google Scholar]
- Chacko J. V., Zanacchi F. C. & Diaspro A. Probing cytoskeletal structures by coupling optical superresolution and AFM techniques for a correlative approach. Cytoskeleton 70, 729–740 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Berdyyeva T., Woodworth C. D. & Sokolov I. Visualization of cytoskeletal elements by the atomic force microscope. Ultramicroscopy 102, 189–198 (2005). [DOI] [PubMed] [Google Scholar]
- Santacroce M. et al. Atomic force microscopy imaging of actin cortical cytoskeleton of Xenopus laevis oocyte. J. Microsc. 223, 57–65 (2006). [DOI] [PubMed] [Google Scholar]
- Meller K. & Theiss C. Atomic force microscopy and confocal laser scanning microscopy on the cytoskeleton of permeabilised and embedded cells. Ultramicroscopy 106, 320–325 (2006). [DOI] [PubMed] [Google Scholar]
- Suzuki Y. et al. High-speed atomic force microscopy combined with inverted optical microscopy for studying cellular events. Sci. Rep. 3, 1–7 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yoshida A. et al. Probing in vivo dynamics of mitochondria and cortical actin networks using high-speed atomic force/fluorescence microscopy. Genes Cells 20, 85–94 (2015). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.