Skip to main content
Annals of Cardiac Anaesthesia logoLink to Annals of Cardiac Anaesthesia
letter
. 2015 Jan-Mar;18(1):105. doi: 10.4103/0971-9784.148333

Author's reply for comment on “positive end-expiratory pressure valve malfunctioning detected by capnographic and airway pressure waveform”

Sohan Lal Solanki 1,, Jeson R Doctor 1, Vijaya P Patil 1, Meenal Rana 1
PMCID: PMC4900304  PMID: 25566723

The Editor,

We thank for that valuable comment on our article “Positive end-expiratory pressure valve malfunctioning detected by capnographic and airway pressure waveform.”

We would like to point out that we use an Heat Moisture Exchanger (HME) filter at the inspiratory (filter on Y piece) and the expiratory port for all our patients. According to the company (Drager Medical) for disinfection/cleaning/autoclaving for adjustable pressure limiting valves it is mentioned (as per the instruction manual)[1] that in noninfectious patients when a filter is used for inspiratory and expiratory port, the washing with detergent and wet pasteurization/immersion in cidex/autoclaving should be done on a weekly basis[1] and not after each patient use as per the response.

As per this recommendation the cleaning was carried out according to schedule. Furthermore, the servicing by the company was as per the annual maintenance contract.

REFERENCE

  • 1.Dräger Medical Operator's Instruction Manual, Fabius GS Software Version 2.n; June, 2004. Drager Medical AG and Co. :125. [Google Scholar]

Articles from Annals of Cardiac Anaesthesia are provided here courtesy of Wolters Kluwer -- Medknow Publications

RESOURCES