Skip to main content
Neurophotonics logoLink to Neurophotonics
. 2016 Jun 13;3(3):031412. doi: 10.1117/1.NPh.3.3.031412

Establishing the diffuse correlation spectroscopy signal relationship with blood flow

David A Boas 1,*, Sava Sakadžić 1, Juliette Selb 1, Parisa Farzam 1, Maria Angela Franceschini 1, Stefan A Carp 1
PMCID: PMC4904065  PMID: 27335889

Abstract.

Diffuse correlation spectroscopy (DCS) measurements of blood flow rely on the sensitivity of the temporal autocorrelation function of diffusively scattered light to red blood cell (RBC) mean square displacement (MSD). For RBCs flowing with convective velocity vRBC, the autocorrelation is expected to decay exponentially with (vRBCτ)2, where τ is the delay time. RBCs also experience shear-induced diffusion with a diffusion coefficient Dshear and an MSD of 6Dshearτ. Surprisingly, experimental data primarily reflect diffusive behavior. To provide quantitative estimates of the relative contributions of convective and diffusive movements, we performed Monte Carlo simulations of light scattering through tissue of varying vessel densities. We assumed laminar vessel flow profiles and accounted for shear-induced diffusion effects. In agreement with experimental data, we found that diffusive motion dominates the correlation decay for typical DCS measurement parameters. Furthermore, our model offers a quantitative relationship between the RBC diffusion coefficient and absolute tissue blood flow. We thus offer, for the first time, theoretical support for the empirically accepted ability of the DCS blood flow index (BFi) to quantify tissue perfusion. We find BFi to be linearly proportional to blood flow, but with a proportionality modulated by the hemoglobin concentration and the average blood vessel diameter.

Keywords: diffuse correlation spectroscopy, blood flow, Monte Carlo, photon migration

1. Introduction

Diffuse correlation spectroscopy (DCS) is increasingly being used as a noninvasive method to quantify tissue blood flow, especially for neurological applications.1,2 DCS measurements quantify the temporal autocorrelation of fluctuations in the intensity of diffusely reflected light that has traveled through tissue. These fluctuations are almost entirely driven by the motion of red blood cells (RBCs) and thus by blood flow.24 Although DCS blood flow measurements have been validated against several alternative perfusion quantification methods,511 a discrepancy remains between the expected nature of RBC displacement and experimental results. Specifically, instead of an expected ballistic random flow with a uniform spatial velocity distribution,12 RBC motions measured by DCS have the characteristics of diffusive motion.4,13

Previous publications from our group13 as well as Ninck et al.14 hypothesized that DCS measurements are also sensitive to the shear-induced diffusion of RBCs in blood flow, a phenomenon previously observed through particle tracking experiments.15 Shear-induced diffusion refers to the random walk-like motion of particles in a shear flow due to interparticle hydrodynamic interactions. Although it was noted that the effective RBC diffusion coefficient is proportional to the flow shear rate, the question of what are the relative contributions of diffusive and convective motions to the observed intensity autocorrelation decay has not been addressed previously.

In this publication, we seek to answer this question using Monte Carlo simulations of light scattering through tissues with varying densities and diameters of vessels. We explicitly model the location of each light scattering event within the vessel flow profiles and consider both convective motion and the corresponding shear-induced diffusion. Through these simulations, we obtain quantitative estimates of the contributions of both processes to the decay of the intensity temporal autocorrelation function. We present definitive evidence that diffusive motion dominates the decay for typical DCS experimental conditions and derive explicit expressions for the relationship between the measured RBC diffusion coefficient and absolute blood flow. Furthermore, we show that the commonly used DCS blood flow index BFi derived using the correlation diffusion equation (CDE) for a semi-infinite geometry is directly proportional to absolute blood flow, but this proportionality is modulated by the hematocrit and the average vessel diameter in the volume of tissue probed by light.

2. Theory and Methods

2.1. Correlation Diffusion Equation

DCS measurements of blood flow are generally analyzed using the CDE.2,3 The CDE is derived from the correlation transfer equation (CTE)3,16 under the assumption that the probability of light scattering is much greater than the probability of light absorption, i.e., reduced scattering coefficient (μs) is much greater than the absorption coefficient (μa). The CTE17 permits calculation of the electric field temporal autocorrelation function G1(τ) under more general conditions of photon migration. The CTE is similar to the traditional radiative transfer equation (RTE), which describes the propagation of light intensity through scattering media. The difference between the CTE and the RTE is that the CTE describes the propagation of the time-varying specific intensity, which represents an angular spectrum of the mutual coherence function. The temporal autocorrelation function is obtained as an integral of the time-varying specific intensity over all solid angles.

Generally, an analytic solution of the CDE for a semi-infinite medium is used to fit the experimental measurements, i.e.,

G1(ρ,τ)=3μs4π[exp(Kr1)r1exp(Kr2)r2], (1)

where

K2=3μaμs+μs2k02αΔr2(τ), (2)

where μs is the reduced scattering coefficient, μa is the absorption coefficient, k0 is the wavenumber of light in the scattering medium, α is the probability of scattering from a moving scatterer, ρ is the source–detector separation, r1=(ρ2+zo2)1/2, r2=(ρ2+(zo+2zb)2)1/2, zo=μs1, and zb=1.76/μs given an index of refraction for tissue of 1.37.18

It is customarily expected that the scatterer motion measured by DCS arises almost exclusively from RBCs. This is justified by experimental studies in animals, which showed that cessation of blood flow results in a nearly 100-fold reduction in the observed dynamics.4 In this paper, we will assume that RBCs are the only source of dynamic scattering events, and we will use PRBC instead of α to denote the probability of scattering from an RBC.

The mean square displacement (MSD) Δr2(τ) of the RBCs is generally given by

Δr2(τ)=6Dτ+vRBC2τ2, (3)

where τ is the delay time, D is the diffusion coefficient, and vRBC is the speed of the RBCs. Note that Eq. (3) can be expanded to use the Langevin formulation for diffusion,13,19,20 accounting for the diffusive motion velocity vector randomization that occurs at short-time delays.

2.2. Monte Carlo Simulations

Investigations with DCS have generally assumed that the motion of tissue scatterers is uncorrelated. Specifically, the assumption is that the phase shifts accumulated by the motions of scattering particles is uncorrelated, which depends on the motion of the scattering particle and the scattering angle. This assumption is generally valid when photons are scattering from particles undergoing Brownian motion. In the case of moving RBCs, this assumption is valid if the photon direction is randomized between scattering events from RBCs. This is true if the scattering is isotropic (i.e., the average cosine of the scattering angle is g=0) or if the photons scatter no more than once inside a vessel and the direction is randomized before scattering from another RBC in another vessel. The optical properties of blood at 800 nm indicate,21 however, that for a typical hematocrit of 40%, g0.98 and the scattering length ls is 12  μm, whereas the absorption length la is over 250 times larger, at around 3.3 mm. Thus, photons entering any blood vessel larger than a capillary will most likely undergo multiple consecutive scattering events within the vessel with correlated linear motions of the RBCs. This breakdown of the uncorrelated motion assumption raises the possibility that the typically used solutions of the CTE, and thus the CDE, are not good models for analyzing DCS measurements of blood flow.

To properly account for the effect of correlated motions on the decay of G1(τ), we use Monte Carlo simulations of photon migration through highly scattering dynamic media and record details of the scattering angles for consecutive scattering events within vessels. Our Monte Carlo code is derived from what Boas originally utilized in Ref. 16 based on earlier work by Middleton and Fisher,22 Durian,23 and Koelink et al.24 The original Monte Carlo simulation propagates photons according to the scattering length and scattering anisotropy of the tissue, which we describe in more detail in Ref. 25. Photons are launched from a source position and recorded at discrete detector locations. For each detected photon, the total path length L is recorded as well as the total dimensionless momentum transfer accumulated over all scattering events Y. The momentum transfer for a photon scattering event is given by q=koutkin, where kout and kin are the photon wavevectors scattered from and incident on the scattering center, respectively, and the total dimensionless momentum transfer is given by Y=q2/2k02, where the sum is over scattering events. The temporal field autocorrelation function is then calculated by

G1(τ)=1Npn=1Npexp(13Ynk02Δr2(τ))exp(μaLn), (4)

where Np is the number of photons detected, Yn is the total dimensionless momentum transfer for photon n, Ln is the total path length of photon n, μa is the absorption coefficient, and Δr2(τ) is the MSD of the scattering particles at a time delay of τ.

The principal advance in the Monte Carlo simulations used in this paper is the introduction of blood vessels with different optical and dynamic properties than the surrounding tissue and with specified internal flow profiles, as shown in Figs. 1 and 2. Photon migration is simulated in a semi-infinite medium with the air tissue interface in the (x,y) plane at z=0. The blood vessels share a common radius α and are all oriented along the y-axis with an even spacing in x and z of hspace. The resulting translational symmetry allows for significant savings in the amount of memory needed to represent the tissue structure for the Monte Carlo simulation. We have verified that this preferred orientation of the vessels only introduces a small bias in our results compared to a truly random orientation of vessels, as expected, given that we typically simulate photons that have propagated a net-distance of several centimeters and have scattered hundreds of times.

Fig. 1.

Fig. 1

Simulation geometry used in this study. (a) Three-dimensional view and (b) (x,z) plane cross section. Vessels of radius R are aligned with the y-axis and are uniformly spaced in the (x,z) plane every hspace μm. Photons are injected at a location within the (x,y) plane and detected a distance ρ away.

Fig. 2.

Fig. 2

Photon propagation tracking. For each scattering event, we record the photon location and the magnitude of the momentum transfer |q|. If the scattering event occurs inside a vessel, we also record the radial position r and the component of q along the vessel direction. In conjunction with a specified flow profile, the radial position r determines the RBC convection velocity vRBC(r) and the shear rate experienced by the RBC s(r).

Photons are launched at a point on the surface along the z-axis and are detected through a 2-mm-diameter circular aperture, a distance ρ from the source point. Figure 2 shows a segment of a photon path and the parameters we record during the photon propagation. For each scattering step during the simulation, we explicitly test for the photon crossing the boundary of a cylindrical vessel. If a crossing occurs, the remaining propagation length for the scattering step is renormalized based on the new scattering coefficient as we describe in Ref. 25. For each detected photon, we record for every scattering event inside a vessel the index of the vessel, the radial position of the scattering event, the magnitude of the momentum transfer |q|, and the component of q along the vessel direction (y-axis). Recording the vessel index enables us to determine consecutive scattering events within a vessel. Knowing the radial position of the scattering event permits us to consider a radial distribution for the convective and diffusive dynamics of the RBCs. We record the y component of q as it is needed to consider the correlated convective motion of the RBCs along the y-axis. All Monte Carlo simulations in this paper use a source–detector separation of 2 cm, an (x,z) cross section of 6×3  cm, and make use of translational symmetry to represent an infinite extent in y. Each simulation run launches 108 photons.

To calculate the temporal field autocorrelation function, we start with the definition of G1(τ) and expand it to sum over all detected photons

G1(τ)=E(0)E*(τ)=1Npn=1Npexp(im=1Ns,nqn,m·Δrn,m(τ))exp(p=1Ntisμa,pLn,p). (5)

Note that one generally reports the normalized temporal field autocorrelation function g1(τ), which is just G1(τ) normalized by the average intensity. The angle brackets () indicate the ensemble average over scattering angles and particle displacements. The number of photons received at the specific detector is Np. The number of scattering events for the n’th detected photon is Ns,n, where qn,m is the momentum transfer for the m’th scattering event of the n’th photon and Δrn,m(τ) is the displacement of the corresponding scattering particle at time lag τ. The total path length of the n’th photon in the p’th tissue type is given by Ln,p and μa,p is the absorption coefficient of the p’th tissue type.

We are considering the convective and shear-induced diffusive movement of RBCs such that the MSD of the RBCs is given by Eq. (3). Although the strength of the diffusive motion correlates with the magnitude of the local shear, the actual diffusive displacements are spatially uncorrelated with the convective displacements. Therefore, note that qn,m·Δrn,m(τ) is generally much smaller than 1, and Eq. (5) can be rewritten as

G1(τ)=n=1Npexp[12l=1Nv,n(m=1Ns,n,lqn,l,m·vn,l,m)2τ2]×exp(m=1Ns,nqn,m2Dn,mτ)exp(p=1Ntisμa,pLn,p). (6)

This formulation reveals the expected decay of G1(τ) due to the RBC displacements arising from diffusion and convective flow. Note that the decay arising from flow speed accounts for the correlated RBC displacements as the sum (l=1Ns,n,lqn,l,m·vn,l,m)2 is performed over the Ns,n,l consecutive scattering events of the n’th photon in the l’th vessel. The sum over the Nv,n vessels entered by the n’th photon is outside the ()2 because the phase shifts qn,l,m·vn,l,m are assumed uncorrelated between vessels, as the scattering angle and flow speed vector are generally randomized between photon interactions with different vessels. All of the information for calculating Eq. (6) is recorded by our Monte Carlo simulation. Given details about the RBC speed and shear-induced diffusion, we can now explicitly test which dominates the decay of g1(τ) and whether the CDE provides a good model fit and allows estimation of blood flow.

2.3. Red Blood Cell Laminar Flow Profile and Shear-Induced Diffusion

Our Monte Carlo simulations permit us to incorporate radial distributions for the flow speed and diffusion coefficient. A general form of the flow speed profile in vessels has been proposed to be26

vRBC(r)=vmax{1[a(rR)]k}, (7)

where vmax is the maximum speed at the center of the vessel, R is the radius of the vessel, a is a scale factor that allows for a nonzero velocity at the wall, and k determines the degree of “bluntness” of the profile. In general, the smaller the vessel the more blunt the profile.26 For sufficiently large vessels, the flow speed profile can be considered to be laminar; thus, in our simulations, we only consider parabolic flow profiles with a=1 and k=2. We believe this assumption captures the essence of the influence of the flow profile on RBC dynamic scattering events. If needed, other a and k values can be used in the future to account for the various degrees of flow profile blunting that have been observed in blood microrheology investigations.

RBC dynamics, while flowing within a vessel, have been studied in numerous rheological studies including video microscopy to track the motion of RBCs ex vivo15 and whole blood in microchannels,27 as well as in vivo in rat venules.28 These studies have revealed that shear induces a diffusive behavior in the movement of the RBCs with the diffusion coefficient proportional to the shear rate, i.e.,

D(r)=αshears(r)=αshear|vRBCr|=2αshearvmaxrR2. (8)

For our simulations, we use the value of αshear=106  mm2 from Goldsmith and Marlow.15

Of note, a prior study by Wu et al.29 has shown that the gradient in the speed of scattering particles flowing within a channel can be the leading factor in the decay of the autocorrelation function. This is an informative study, but the parameters used are quite different from those encountered in biological tissue, leading to negligible levels of shear-induced diffusion. This is due to the use of a single flow channel with a much smaller shear rate than that experienced by RBCs in blood vessels, as well as a much smaller particle concentration and scattering coefficient. Furthermore, this study did not consider multiple flow channels in an otherwise static scattering medium.

2.4. Relating the Diffuse Correlation Spectroscopy Blood Flow Index to Absolute Blood Flow

We are now in a position to perform Monte Carlo simulations to calculate g1,MC(τ) for RBCs flowing at varying speeds in vessels of varying diameters and spacings. We can then fit the result with g1,CDE(τ) obtained from the CDE to determine if the CDE remains a good model and to establish a quantitative relationship between the CDE blood flow index BFi and the absolute blood flow BFabs. For a semi-infinite medium, we use g1,CDE(τ) given in Eq. (1) and estimate K from Eq. (2). The optical properties and wavenumber of light are generally known in Eq. (2), leaving us to estimate PRBCΔr2(τ). We define the blood flow indices BFiconv=PRBCv2 and BFidif=PRBCD to be estimated by fitting either the convective or diffusive MSD model to g1,MC(τ), where either BFi is expected to be linearly proportional to flow speed. The angle brackets for v and D are an explicit recognition that the BFi measures the average speed v or diffusion coefficient D. Note that the probability of scattering from an RBC is not generally known a priori. In the context of photon diffusion theory, we would expect that PRBC is given by the product of the volume fraction of blood, fblood=πR2/hspace2 and the ratio of the blood to average tissue reduced scattering coefficient, μs,blood/μs,avg, i.e.,

PRBC=πR2hspace2μs,bloodμs,avg. (9)

The average tissue-reduced scattering coefficient is given by the volume fraction weighted average of the tissue and blood coefficients μs,avg=fbloodμs,blood+(1fblood)μs,tis.

We present results below for fitting g1,MC(τ). To establish the quantitative relationship between BFi and BFabs for convective and diffusive RBC displacements independently, we obtain fitting results alternately setting D=0 and vRBC=0, respectively. We will distinguish these temporal field autocorrelation functions as g1conv(τ) and g1dif(τ) for convective and diffusive RBC displacements, respectively.

In the case of convective motion only, we expect v to be the spatially weighted average of RBC speed within the blood vessel. As the absorption length and photon transport length inside the vessel are generally larger than the vessel diameter, we expect uniform spatial sampling of the vessel and thus expect v=vmax/2 for a parabolic laminar flow profile. We thus arrive at an equation for BFiconv

BFiconv=12(μs,bloodμs,avg)0.5π0.5Rhspacevmax. (10)

In the case of diffusive motion only, we expect D to be the spatially weighted average of the RBC diffusion coefficient within the blood vessel, i.e., D=4αshearvmax/3R given a parabolic laminar profile and uniform photon sampling of the blood vessel. We thus arrive at an equation for BFidif

BFidif=4αshear3μs,bloodμs,avgπRhspace2vmax. (11)

We define absolute blood flow as the volume of blood transiting through a unit volume of tissue per second, which is given by the flow in a vessel (cross-sectional area times average RBC speed) times the number of vessels per unit volume of tissue, i.e.,

BFabs=12πR2vmaxnves, (12)

where the number of vessels per unit volume of tissue in our geometry is

nves=1hspace2. (13)

We thus see that

BFiconv=(μs,bloodμs,avg)0.5(hspaceπ0.5R)BFabs (14)

and

BFidif=8αshear3μs,bloodμs,avg1RBFabs. (15)

Thus, the prediction from the CDE is that the blood flow index measured by DCS is proportional to absolute blood flow, but the proportionality is modulated by μs,blood and R. Additionally, the proportionality between BFiconv and BFabs is further modulated by the number density of vessels. Note that we assume an estimate of μs,avg from, for instance, an NIRS measurement or from a multidistance DCS measurement.30

Finally, we present results below considering the combination of convective and diffusive RBC displacements to establish which dominates the decay of g1total(τ). Given the overwhelming indication from experimental studies,4 we expect to observe that the diffusive motion dominates the decay.

3. Results

3.1. Impact of Scattering Length Within a Vessel

Table 1 summarizes representative optical properties of blood and extravascular tissue at 800 nm for a blood hematocrit of 40%.21 We use g and the scattering coefficient μs=1/ls in our Monte Carlo simulations. For the analytical solution provided by the CDE, we use the reduced scattering coefficient μs=μs(1g).

Table 1.

Optical properties for blood and extravascular tissue at 800 nm.

  μa (mm1) la (mm) μs (mm1) ls (mm) g μs (mm1)
Blood 0.3 3.3 82 0.012 0.977 1.9
Tissue 0.002 500 10 0.1 0.9 1.0

As a preamble, we explicitly test the prediction offered by the CDE that g1(τ) is invariant to changes in μs and g provided that μs is constant. To this end, we ran Monte Carlo simulations for three different combinations of μs,blood and gblood maintaining a constant μs,blood=1.9  mm1. Specifically, we use μs,blood=1.9, 19, and 190  mm1, and gblood=0,0.9, and 0.99, respectively. The rest of the optical parameters match the values in Table 1, except that we use μs,tis=1.0  mm1 and gtis=0, which are typical values we use in our Monte Carlo simulations for the human head to accelerate the computation.25 We show results for 80-μm-diameter vessels, spaced every 300  μm, with vRBC,max=2  mm/s.

The results for g1conv(τ) and g1dif(τ) are shown in Fig. 3. These results confirm the prediction from the CDE that, at least for early decay times, g1(τ) is invariant as long as μs(1g) remains constant. The behavior at long decay times does exhibit some dependence on μs of the blood [see enlargement in panel b)]. Likely, as μs of the blood is decreased, the probability increases that a photon traverses a vessel without scattering from a moving RBC. Importantly, we observe that g1dif(τ) decays more quickly than g1conv(τ), indicating that the shear-induced diffusion of the RBCs produces more dynamic fluctuations in the electromagnetic field than the flow speed convective displacements. Based on these results, for the remainder of the paper, we use gblood=0.9 since this gives us effectively the same results we would obtain with gblood=0.977 and leads to a more efficient execution of the Monte Carlo simulations.

Fig. 3.

Fig. 3

(a) The decay of the temporal field autocorrelation function at a 2-cm source–detector separation is shown for μs,blood=1.9  mm1 with g=0,0.9, and 0.99. We used 80-μm-diameter vessels, spaced every 300  μm, with vRBC,max=2  mm/s. The dotted, dashed, and solid lines indicate g1conv(τ), g1dif(τ), and g1tot(τ), respectively. (b) An enlargement of the 105<τ<2×104 range in (a).

3.2. Blood Flow Indices Estimated from Fitting Monte Carlo Data Using the Correlation Diffusion Equation

In this section, we present results to validate the relations for the blood flow indices BFiconv and BFidif given by Eqs. (10) and (11), respectively. We do so by fitting the solution of the CDE, given by Eq. (1), to g1,MC(τ), obtained from the Monte Carlo simulations. As detailed in Table 2, we varied RBC speed, vessel radius, vessel spacing, and hematocrit in these simulations to fully test the dependencies of BFiconv and BFidif on each parameter. Note that we fit g1conv(τ) to estimate BFiconv and we fit g1tot(τ) to estimate BFidif. We only fit g1tot(τ) with the diffusion term because our simulations indicate that the τ2 dependence arising from the convective displacement of the RBCs has negligible impact on g1tot(τ) for our parameters of interest. For all cases presented in this section, the blood optical properties were μs=1.9  mm1, g=0.9, and μa=0.30  mm1. For the tissue, we set μs=1.0  mm1, g=0.0, and μa=0.002  mm1.

Table 2.

Vascular parameter ranges explored in this study. Each parameter was varied separately while the others were held at a fixed value.

Vascular parameter Range
RBC speed (vRBC) 1, 2, 4  mm/s
Vessel radius (R) 30, 40, 50, 60, 70, 80, 90, 100  μm
Vessel spacing (hspace) 200, 300  μm
Hematocrit (Hct) 32%, 36%, 40%, 44%, 48%

We first present results increasing the speed vmax of the RBCs. Figures 4(a) and 4(b) show g1,MC(τ) and the CDE fit for vmax=1,2, and 4  mm/s for g1tot(τ) and g1conv(τ), respectively. In this example, the vessel diameter is set to 80  μm, the vessels are spaced every 300 μm and the hematocrit is set to 40%. These results indicate that the CDE fits g1tot(τ) and g1conv(τ) well. Figures 4(c) and 4(d) show the fitted values of BFidif and BFiconv versus speed, compared with those predicted using Eqs. (10) and (11). These results confirm that the BFis increase linearly with RBC speed and confirm the accuracy of Eqs. (10) and (11).

Fig. 4.

Fig. 4

(a) and (b) The CDE fit to the Monte Carlo results for g1tot(τ) and g1conv(τ). The Monte Carlo results for different RBC speeds are indicated by the symbols and the diffuse equation fits are given by the colored solid lines. The symbols in (c) and (d) show the blood flow index BFi determined from fitting g1tot(τ) with a diffusive motion model and from fitting g1conv(τ) with a convective motion model, respectively, versus the true RBC speed. The solid lines in (c) and (d) are the theoretical predictions for the BFis based on Eqs. (10) and (11). Here, we used 80 μm vessels spaced every 300 μm and set the hematocrit to 40%.

Figure 5 shows the BFi fitting results for variations in the absolute volumetric blood flow achieved by varying the vessel diameter at a constant maximum RBC speed vmax. The vessel diameter was varied from 30 to 100  μm and results are shown for vessel spacings of 200 and 300 μm, respectively. As with the previous results, we fit g1tot(τ) and g1conv(τ) to estimate BFidif and BFiconv, respectively. We then used Eqs. (14) and (15) to determine the expected BFi values and find that this predicts well the measured BFi.

Fig. 5.

Fig. 5

The symbols in (a) and (b) show the diffusive and convective blood flow indices, respectively, estimated using the CDE versus the true blood flow. The blood flow variation was accomplished by increasing the vessel diameter from 30 to 100  μm with a fixed RBC flow speed of 1  mm/s. The solid lines indicate the expected blood flow index values predicted by Eqs. (15) and (14), respectively. These results were obtained for a hematocrit of 40%

In Fig. 6, we plot fitted BFi values for changing the blood hematocrit by ±20% from our baseline value of 40%. In this range, the reduced scattering coefficient of blood is changing linearly with hematocrit.21 We also show the expected BFi values and find excellent agreement.

Fig. 6.

Fig. 6

The symbols in (a) and (b) show the diffusive and convective blood flow indices, respectively, estimated using the CDE for vessel scattering coefficients corresponding to hematocrit values ranging from 20% to +20% of our 40% baseline for a vessel diameter of 80  μm, with 300-μm vessel spacing, and an RBC flow speed of 2  mm/s. The solid lines indicate the expected blood flow index values predicted by Eqs. (15) and (14), respectively.

4. Discussion

The agreement between the prediction by the CDE and the fits to the Monte Carlo results is striking. Despite the heterogeneous optical and dynamic properties of the vascular structure and the correlated consecutive scattering events within a vessel, the homogeneous solution of the CDE accurately describes the decay of g1(τ). We have shown results fitting g1,MCconv(τ) and g1,MCtot(τ), allowing us to test the convective and diffusive RBC displacement models for the CDE and demonstrate that the diffusive motion of RBCs dominates the decay of g1,MCtot(τ). Our fitting results confirm that Eqs. (14) and (15), which relate BFiconv and BFidif to blood flow, are accurate within 5%. Importantly, while the CDE allows us to write a linear relation between the estimated blood flow index and true absolute blood flow, we see that the proportionality is modulated by the radius of the blood vessels and the reduced scattering coefficient of the blood (which is modulated by hematocrit). In the case of BFiconv, the proportionality is further dependent on the number density of vessels. In this sense, it is fortunate that the diffusive motion of RBCs dominates the decay of g1,MCtot(τ); thus, we do not need knowledge of the number density of vessels to estimate BFabs. Recall that the BFis depend on hematocrit even when absolute blood flow is constant because the BFis depend on the probability of scattering from an RBC, Eq. (9), which increases linearly with the reduced scattering coefficient of blood.

Given that the diffusive motion of RBCs dominates the decay of g1,MCtot(τ), Eq. (15) thus stands as the quantitative relationship between the blood flow index measured by DCS and the true absolute blood flow. Recall that in order to get BFi, we first need to know the average optical properties of the tissue μs,avg and μa,avg, which would generally be obtained from an independent NIRS measurement31,32 and possibly also estimated with multidistance DCS measurements.30 To then convert the BFidif to BFabs, we need to know the proportionality αshear between shear flow and the RBC diffusion coefficient, the reduced scattering coefficient of the blood μs,blood, and the radius of the blood vessels. In this paper, we used a value of αshear=106  mm2 obtained from Goldsmith and Marlow.15 We note that this is a value obtained for a hematocrit of 40% to 47% and that αshear was found to linearly increase with hematocrit from 0% to 45% and then to plateau and reverse.33 The reduced scattering coefficient of blood μs,blood is linearly proportional to hematocrit from 0% to 40%, the highest value measured by Meinke et al.21 In principle, if one has an independent measure of hematocrit, from a blood draw for instance, one can then determine the appropriate value to use for αshear and μs,blood. The remaining factor needed to estimate BFabs is the vessel radius R. Although we performed simulations using vessels with a common radius, in reality DCS will measure vessels with a distribution of radii. We anticipate that Eq. (15) will still be valid when measuring a distribution of vessel radii, but that the effective vessel radius will represent a complex nonlinear dependence on the distribution of vessel radii and the corresponding RBC speed distribution.

We note that in our previous publication,13 we postulated that the diffusional dependence of the DCS measurements in tissue arose because of sequential multiple scattering within a single vessel. As demonstrated in this paper, this factor does not play a major role in the interplay between diffusive and convective influences on the autocorrelation of multiple scattered light. Rather, it is simply that at source–detector separations commonly used for DCS measurements, the shear-induced diffusion dominates the signal. We estimate that the convective contribution only becomes significant when the detected photon path lengths become less than a millimeter, but this requires further investigation. A potential implication is that the convective contribution should be considered in analyzing laser speckle contrast imaging data.34 It is worth noting at this point that the concept of a well-defined absolute blood flow has been questioned as the value of absolute blood flow obtained has been theoretically demonstrated to depend on vascular geometry, measurement procedure, and spatial scale sampled.35,36 Further research is needed to better understand the implications of these theoretical arguments.

5. Summary

We have presented a set of Monte Carlo simulations that quantify the contribution of diffusive and convective RBC motion to the decay of the autocorrelation function of light that has scattered through tissue. Through this modeling, we demonstrate that diffusive motion dominates the autocorrelation decay for typical experimental parameters used in DCS measurements. We have also provided expressions for the expected dependence of the measured DCS blood flow index on absolute tissue blood flow in terms of the vascular volume fraction, average vessel diameter, and blood optical properties. Our results indicate that the DCS blood flow index commonly obtained using the established CDE indeed offers a direct measure of tissue blood flow, but the proportionality is also sensitive to changes in hematocrit and in the average vessel diameter in the region probed by light. The immediate implication is that care must be taken to control for variations in these additional parameters when reporting DCS blood flow measurements. We expect further studies to explore these sources of variability and help define the range of effective DCS applications.

Acknowledgments

We acknowledge financial support from the National Institutes of Health (NIH) Grants Nos. P41-EB015896, R01-NS091230, R21-NS093259, and R01-CA187595.

Biographies

David A. Boas is the director of the Martinos Optics Division in the Department of Radiology at Massachusetts General Hospital, and a professor of radiology at Harvard Medical School. He received his BS degree in physics at RPI and his PhD degree in physics at the University of Pennsylvania. He is the founding president of the Society for Functional Near-Infrared Spectroscopy, founding editor-in-chief of Neurophotonics, and the recipient of the 2016 Britton Chance Award in biomedical optics.

Sava Sakadžić is a faculty member in the Optics Division of the Martinos Center for Biomedical Imaging at Massachusetts General Hospital and an assistant professor at Harvard Medical School. His research is focused on the development of optical microscopy technologies and their applications to study cortical oxygen delivery and consumption in normal brain and in various brain conditions.

Juliette Selb is an instructor in the Optics Division of the Athinoula A. Martinos Center for Biomedical Imaging at Massachusetts General Hospital, Harvard Medical School. She received her PhD in 2002 from the Université Paris Sud in France for her work on acousto-optic imaging. Her current research focuses on diffuse optical modalities for brain imaging in humans.

Parisa Farzam is a research fellow at Martinos Center at Massachusetts General Hospital, Harvard Medical School. She received her PhD in biomedical photonics from ICFO-The Institute of Photonics Sciences, Spain. Her research is focused on developing novel diffuse optical instrumentation, algorithms and protocols and applying them for preclinical and clinical investigations.

Maria Angela Franceschini is a faculty member in the Optics Division of the Martinos Center for Biomedical Imaging at Massachusetts General Hospital and an associate professor at Harvard Medical School. Her research focuses broadly on the development and application of noninvasive optical techniques to studies of the human brain. She is a pioneer in the field of near infrared spectroscopy (NIRS) and has made substantial contributions to the modeling, testing, and clinical and neuroscience translation of NIRS and diffuse correlation spectroscopy.

Stefan A. Carp is an assistant professor of radiology at Harvard Medical School and a member of Massachusetts General Hospital Martinos Center Optics Division. He received his BS degrees in chemistry and chemical engineering from MIT and a PhD from the University of California, Irvine, in biomedical optics. He focuses on the development of novel techniques for tissue hemodynamics and oxygen metabolism monitoring, as well as their translation to clinical use.

References

  • 1.Torricelli A., et al. , “Neurophotonics: non-invasive optical techniques for monitoring brain functions,” Funct. Neurol. 29(4), 223–230 (2014). 10.11138/FNeur/2014.29.4.223 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Durduran T., Yodh A. G., “Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement,” NeuroImage 85(1), 51–63 (2014). 10.1016/j.neuroimage.2013.06.017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Boas D., Campbell L., Yodh A., “Scattering and imaging with diffusing temporal field correlations,” Phys. Rev. Lett. 75(9), 1855–1858 (1995). 10.1103/PhysRevLett.75.1855 [DOI] [PubMed] [Google Scholar]
  • 4.Durduran T., et al. , “Diffuse optics for tissue monitoring and tomography,” Rep. Prog. Phys. 73(7), 076701 (2010). 10.1088/0034-4885/73/7/076701 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Durduran T., et al. , “Optical measurement of cerebral hemodynamics and oxygen metabolism in neonates with congenital heart defects,” J. Biomed. Opt. 15(3), 037004 (2010). 10.1117/1.3425884 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Carp S. A., et al. , “Validation of diffuse correlation spectroscopy measurements of rodent cerebral blood flow with simultaneous arterial spin labeling MRI; towards MRI-optical continuous cerebral metabolic monitoring,” Biomed. Opt. Express 1(2), 553–565 (2010). 10.1364/BOE.1.000553 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Yu G., et al. , “Validation of diffuse correlation spectroscopy for muscle blood flow with concurrent arterial spin labeled perfusion MRI,” Opt. Express 15(3), 1064–1075 (2007). 10.1364/OE.15.001064 [DOI] [PubMed] [Google Scholar]
  • 8.Buckley E. M., et al. , “Cerebral hemodynamics in preterm infants during positional intervention measured with diffuse correlation spectroscopy and transcranial Doppler ultrasound,” Opt. Express 17(15), 12571–12581 (2009). 10.1364/OE.17.012571 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Roche-Labarbe N., et al. , “Noninvasive optical measures of CBV, StO(2), CBF index, and rCMRO(2) in human premature neonates’ brains in the first six weeks of life,” Hum. Brain Mapp. 31(3), 341–352 (2010). 10.1002/hbm.v31:3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Kim M. N., et al. , “Noninvasive measurement of cerebral blood flow and blood oxygenation using near-infrared and diffuse correlation spectroscopies in critically brain-injured adults,” Neurocrit. Care 12(2), 173–180 (2010). 10.1007/s12028-009-9305-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Zhou C., et al. , “Diffuse optical monitoring of hemodynamic changes in piglet brain with closed head injury,” J. Biomed. Opt. 14(3), 034015 (2009). 10.1117/1.3146814 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Bonner R., Nossal R., “Model for laser Doppler measurements of blood flow in tissue,” Appl. Opt. 20(12), 2097–2107 (1981). 10.1364/AO.20.002097 [DOI] [PubMed] [Google Scholar]
  • 13.Carp S. A., et al. , “Due to intravascular multiple sequential scattering, diffuse correlation spectroscopy of tissue primarily measures relative red blood cell motion within vessels,” Biomed. Opt. Express 2(7), 2047–2054 (2011). 10.1364/BOE.2.002047 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Ninck M., Untenberger M., Gisler T., “Diffusing-wave spectroscopy with dynamic contrast variation: disentangling the effects of blood flow and extravascular tissue shearing on signals from deep tissue,” Biomed. Opt. Express 1(5), 1502–1513 (2010). 10.1364/BOE.1.001502 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Goldsmith H. L., Marlow J. C., “Flow behavior of erythrocytes. II. Particle motions in concentrated suspensions of ghost cells,” J. Colloid Interface Sci. 71(2), 383–407 (1979). 10.1016/0021-9797(79)90248-0 [DOI] [Google Scholar]
  • 16.Boas D. A., Yodh A. G., “Spatially varying dynamical properties of turbid media probed with diffusing temporal light correlation,” J. Opt. Soc. Am. A 14(1), 192–215 (1997). 10.1364/JOSAA.14.000192 [DOI] [Google Scholar]
  • 17.Ackerson B. J., et al. , “Correlation transfer—application of radiative transfer solution methods to photon correlation problems,” J. Thermophys. Heat Transfer 6(4), 577–588 (1992). 10.2514/3.11537 [DOI] [Google Scholar]
  • 18.Haskell R. C., et al. , “Boundary conditions for the diffusion equation in radiative transfer,” J. Opt. Soc. Am. A 11(10), 2727–2741 (1994). 10.1364/JOSAA.11.002727 [DOI] [PubMed] [Google Scholar]
  • 19.Berne B., Pecora R., Dynamic Light Scattering: With Applications to Chemistry, Biology, and Physics, Dover Publications, Mineola, New York: (2000). [Google Scholar]
  • 20.Verdecchia K., et al. , “Assessment of the best flow model to characterize diffuse correlation spectroscopy data acquired directly on the brain,” Biomed. Opt. Express 6(11), 4288–4301 (2015). 10.1364/BOE.6.004288 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Meinke M., et al. , “Empirical model functions to calculate hematocrit-dependent optical properties of human blood,” Appl. Opt. 46(10), 1742–1753 (2007). 10.1364/AO.46.001742 [DOI] [PubMed] [Google Scholar]
  • 22.Middleton A., Fisher D., “Discrete scatterers and autocorrelations of multiply scattered light,” Phys. Rev. B 43(7), 5934–5938 (1991). 10.1103/PhysRevB.43.5934 [DOI] [PubMed] [Google Scholar]
  • 23.Durian D., “Accuracy of diffusing-wave spectroscopy theories,” Phys. Rev. E 51(4), 3350–3358 (1995). 10.1103/PhysRevE.51.3350 [DOI] [PubMed] [Google Scholar]
  • 24.Koelink M. H., et al. , “Laser Doppler blood flowmetry using two wavelengths: Monte Carlo simulations and measurements,” Appl. Opt. 33(16), 3549–3558 (1994). 10.1364/AO.33.003549 [DOI] [PubMed] [Google Scholar]
  • 25.Boas D., et al. , “Three dimensional Monte Carlo code for photon migration through complex heterogeneous media including the adult human head,” Opt. Express 10(3), 159–170 (2002). 10.1364/OE.10.000159 [DOI] [PubMed] [Google Scholar]
  • 26.Tangelder G. J., et al. , “Velocity profiles of blood platelets and red blood cells flowing in arterioles of the rabbit mesentery,” Circ. Res. 59(5), 505–514 (1986). 10.1161/01.RES.59.5.505 [DOI] [PubMed] [Google Scholar]
  • 27.Higgins J. M., et al. , “Statistical dynamics of flowing red blood cells by morphological image processing,” PLoS Comput. Biol. 5(2), e1000288 (2009). 10.1371/journal.pcbi.1000288 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Bishop J. J., et al. , “Effect of aggregation and shear rate on the dispersion of red blood cells flowing in venules,” Am. J. Physiol. Heart Circ. Physiol. 283(5), H1985–H1996 (2002). 10.1152/ajpheart.00888.2001 [DOI] [PubMed] [Google Scholar]
  • 29.Wu X. L., et al. , “Diffusing-wave spectroscopy in a shear-flow,” J. Opt. Soc. Am. B 7(1), 15–20 (1990). 10.1364/JOSAB.7.000015 [DOI] [Google Scholar]
  • 30.Farzam P., Durduran T., “Multidistance diffuse correlation spectroscopy for simultaneous estimation of blood flow index and optical properties,” J. Biomed. Opt. 20(5), 055001 (2015). 10.1117/1.JBO.20.5.055001 [DOI] [PubMed] [Google Scholar]
  • 31.Franceschini M. A., et al. , “Assessment of infant brain development with frequency-domain near-infrared spectroscopy,” Pediatr. Res. 61(5 Pt1), 546–551 (2007). 10.1203/pdr.0b013e318045be99 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Martelli F., et al. , “Phantom validation and in vivo application of an inversion procedure for retrieving the optical properties of diffusive layered media from time-resolved reflectance measurements,” Opt. Lett. 29(17), 2037–2039 (2004). 10.1364/OL.29.002037 [DOI] [PubMed] [Google Scholar]
  • 33.Cha W., Beissinger R., “Evaluation of shear-induced particle diffusivity in red cell ghosts suspensions,” Korean J. Chem. Eng. 18(4), 479–485 (2001). 10.1007/BF02698294 [DOI] [Google Scholar]
  • 34.Boas D. A., Dunn A. K., “Laser speckle contrast imaging in biomedical optics,” J. Biomed. Opt. 15(1), 011109 (2010). 10.1117/1.3285504 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Guibert R., et al. , “On the normalization of cerebral blood flow,” J. Cereb. Blood Flow Metab. 33(5), 669–672 (2013). 10.1038/jcbfm.2013.39 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Kudo K., et al. , “Quantitative cerebral blood flow measurement with dynamic perfusion ct using the vascular-pixel elimination method: comparison with h2(15)o positron emission tomography,” AJNR Am. J. Neuroradiol. 24(3), 419–26 (2003). [PMC free article] [PubMed] [Google Scholar]

Articles from Neurophotonics are provided here courtesy of Society of Photo-Optical Instrumentation Engineers

RESOURCES