Description
Gore-Tex is a commonly used material for external thyroplasty in patients with unilateral vocal cord palsy.1 It medialises the paralysed cord to enable the normal contralateral cord to appose with it in the midline, thereby improving voice quality and preventing aspiration.
Potential complications from its use, although rare, include extrusion and chronic inflammation at the site of insertion, with evidence of foreign body giant cell reaction.
We present a case of a patient presenting with shortness of breath and worsening voice quality. The patient had suffered a right vocal cord palsy following thyroidectomy. The patient subsequently underwent gel foam injections followed by a Gore-Tex medialisation thyroplasty. Nasendoscopy demonstrated exuberant right false cord granulation (figure 1).
Figure 1.

Flexible nasendoscopic view of the glottis. Arrow indicating granulation tissue arising from the right vocal cord.
The decision was made to remove the Gore-Tex, and an attempt was made through a standard external thyroplasty approach, although this was unsuccessful as the implant could not be located. A repeat CT was organised with high-resolution laryngeal reconstructions (figure 2), which demonstrated the implant spiralling into a position more superiorly than expected. At re-exploration, the glottis was approached from the superior edge of the thyroid ala with dissection of the mucosa. The implant was located in the right false vocal cord and removed without further complication.
Figure 2.

CT scan (coronal view) of the neck, demonstrating high-density material (arrow) in the right supra-glottic space, which at surgery was confirmed to be the Gore-Tex implant.
Extrusion of Gore-Tex implants, though uncommon, has been documented; and with migration to other laryngeal sub sites, it is even less common. Extruded implants are more obvious and can be managed endoscopically.2 This case also demonstrates the value of high-resolution CT scans with laryngeal reconstructions as part of preoperative planning. Standard thyroplasty approaches may not be sufficient to retrieve migrated implants. Dissecting the laryngeal mucosa off the thyroid cartilage from superior to inferior is a good technique for accessing superiorly migrated implants.
Learning points.
Migration of a Gore-Tex implant to another sub site within the larynx is possible.
High-resolution CT scans with laryngeal reconstructions are useful for preoperative planning.
Approaching the larynx superiorly over the thyroid cartilage is a successful technique to remove migrated implants.
Footnotes
Contributors: JH prepared the manuscript for submission. mab acquired the clinical photos and reviewed the manuscript. FR was the consultant surgeon.
Competing interests: None declared.
Patient consent: Obtained.
Provenance and peer review: Not commissioned; externally peer reviewed.
References
- 1.Misono S, Merati AL. Evidence-based practice: evaluation and management of unilateral vocal fold paralysis. Otolaryngol Clin North Am 2012; 45:1083–108. doi:10.1016/j.otc.2012.06.011 [DOI] [PubMed] [Google Scholar]
- 2.Halum SL, Postma GN, Koufman JA. Endoscopic management of extruding medialization laryngoplasty implants. Laryngoscope 2005;115:1051–4. doi:10.1097/01.MLG.0000163342.10412.EB [DOI] [PubMed] [Google Scholar]
