Abstract
We have developed a range of disposable and colorless tear glucose sensing contact lenses, using off-the-shelf lenses embedded with new water soluble, highly fluorescent and glucose sensitive boronic acid containing fluorophores. The new lenses are readily able to track tear glucose levels and therefore blood glucose levels, which are ideally suited for potential use by diabetics. The fluorescence responses from the lenses can be monitored using simple excitation and emission detection devices. The novelty of our approach is two fold. Firstly, the notion of sensing extremely low glucose concentrations in tears, which track blood levels, by our contact lens approach, and secondly, the unique compatibility of our new glucose signaling probes with the internal mildly acidic contact lens environment. The new lenses are therefore ideal for the non-invasive and continuous monitoring of tear glucose, with about 15-min response time, and a measured shelf life in excess of 3 months. In this review article, we show that fluorescence based signaling using plastic disposable lenses, which have already been industrially optimized with regard to vision correction and oxygen/analyte permeability etc, may a notable alternative to invasive and random finger pricking, the most widely used glucose monitoring technology by diabetics.
Keywords: Ophthalmic glucose monitoring; tear glucose; glucose sensing; monosaccharides, contact lens; non-invasive sensing; continuous glucose monitoring; clinical assessment; drug compliance; diabetes
INTRODUCTION
As a common medical condition that produces excessive thirst, continuous urination and severe weight loss, Diabetes4 has interested medical researchers for over three millennia. Unfortunately it wasn’t until the early 20th century that the prognosis for this condition became any better than it was 3000 years ago. Today, approximately 150 million people worldwide are affected by diabetes. With its prevalence still rising, diabetes still continues to fascinate, practitioners and researchers alike by its elusive cause and its many manifestations.
While the Ebers Papyrus, which was written around 1500 BC, excavated in 1862 AD from an ancient grave in Thebes, Egypt, described the first reference to Diabetes Mellitus, it was physicians in India at around the same time that developed the first crude test for diabetes. They observed that the urine from people with diabetes attracted ants and flies. They subsequently named the condition “madhumeha” or “honey urine” [1].
The important elements of our current understanding of diabetes can be traced to the early to mid 19th century. In 1815 Eugene Chevreul in Paris concluded that the sugar in urine was indeed Glucose, the first quantitative test for glucose in urine being developed by Von Fehling some years later in 1848 [2].
Over 150 years later, significant attention is still given to the development of physiological glucose monitoring [3–36]. This is because one important aspect for diabetes management, involves the tight control of blood glucose levels, so as to manage food intake and the dosage and timing of insulin injection. Tests for determining serum glucose concentration typically require blood collection by some invasive technique, usually a needle or other device causing arterial or venous puncture. Currently, millions of diabetics are left with very few alternatives, except to invasively draw blood many times daily to determine their blood sugar levels. To this end, many technologies have been developed over the past 20 years in an attempt to provide a technology which promises both non-invasive and continuous physiological glucose monitoring. These include near infrared spectroscopy [3,4], optical rotation [5,6], colorimetric [7,8] and fluorescence detection [9–13], to name but just a very few. Recently we have seen the launch of the new GlucoWatch, which approved by the FDA in 2001, is the first step towards both the continuous and non-invasive monitoring of physiological glucose. However, in addition to wearing this wrist watch based glucose sensor, it is also recommended that glucose monitoring by another blood sampling technique be additionally used from time to time. Other technologies which are presently emerging include, Glucose monitoring skin Patches, implantable glucose sensors coupled insulin pumps, and laser blood drawing, which is deemed less painful than finger pricking with a lancet or needle. Yet with all these emerging technologies, there is still a need for new technologies, which are truly non-invasive and continuous. To this end our laboratories have been developing glucose sensing contact lenses which when worn by diabetics, who often require vision correction in any case, can potentially monitor tear glucose levels, which are known to directly relate and track blood glucose levels [14–17].
In this review article of work from our laboratories, we employ the notion of elevated tear glucose levels during hyperglycemia to investigate, for the first time, the possibility of monitoring tear glucose and therefore blood glucose, using a disposable, off-the-shelf, contact lens. By incorporating new monosaccharide signaling fluorescent probes within such a lens (Fig. 1), we can indeed make progress towards this non-invasive and continuous approach for glucose monitoring (Fig. 2).
As with any sensors, there are several issues that have to be addressed. The first is to identify suitable transduction elements, which in the presence of glucose, can report/produce suitable signals. The second is the design of the matrix to incorporate the transduction elements. For this, we have chosen an off-the-shelf disposable plastic contact lens, primarily because its physiological compatibility has already been assessed, and finally, the optimization of the sensor, with regard to sensitivity, response time, reversibility and shelf-life etc. The later two issues will be discussed throughout much of this review and indeed in past papers by the authors [14–17]. For the identification of suitable transduction elements, boronic acid has been known to have high affinity for diol-containing compounds such as carbohydrates [18–20] (Scheme 1), where the strong complexation has been used for the construction of carbohydrate sensors [21–28], transporters [29] and chromatographic materials [30]. Boronic acid compounds have also been used for the synthesis of glucose sensors [31–36], where we note the work of Shinkai [31,32] and Lakowicz [33–36] to name but just a few workers in this field.
Boronic acids are weak Lewis Acids composed of an electron deficient boron atom and two hydroxyl groups, (1 in scheme 1), which can interact with strong bases like OH− to from the anionic borate form (2 in scheme 1), showing typically high pKa around 9 [37,38]. Boronic acids couple with diols to form a boronic acid diester group (3 in scheme 1). The diol is linked covalently, and the reaction is fast and completely reversible [38]. In comparison to the boronic acid group, the boronic acid ester group shows higher acidity (pKa ≈ 6) due to a higher electrophilic boron atom. The monophenylboronic acid group shows higher affinity for d-fructose with a smaller affinity for d-glucose [38], with binding constants of ≈0.5 and 10 mM respectively [38]. The use of the boronic acid groups for sensing sugars is strongly dependent on the molecular geometry and the aromatic species where the boronic acid group is present, hence glucose sensitive probes can be made with a variety of affinities, in the mM range for blood glucose [34–36], and in the μM range for tear glucose [39,40].
LENS FEASIBILITY STUDY
To determine the usefulness of glucose signaling probes based on the boronic acid glucose chelating moiety, we initially compared the response towards glucose of a wide variety of published BAFs when doped within the contact lenses [14].
In most cases the response towards glucose in the lenses was significantly different, with drastically reduced glucose responses [14]. Figure 3 shows the response of one boronic acid containing fluorophore, DDPBBA, towards glucose within the lens, as compared to the response typically observed at physiological pH (Fig. 4). As with the other probes studied, we observed that the response towards glucose in the contact lens was similar to the response observed in ≈pH 6.0 bulk solution (cf. Figs. 3 and 4 bottom right). We subsequently doped the well-known pH sensitive probe fluorescein [41] within the lenses, and determined that the lenses had an unbufferable lens pH of ≈ 6.1 (Fig. 5 top). However, the influence of an acidic lens pH does not solely explain the spectral shifts observed in Figs. 3 and 4, with DDPBBA additionally showing a 50 nm hyposchromic shift. An investigation of the local polarity within the contact lens, using pyrene [14,42,43] (Fig. 5 bottom), revealed a lens polarity similar to that of methanol, which in hindsight was not too surprising, given the nature of the polymer, i.e. it is PVA based.
RATIONALE FOR THE DESIGN OF NEW GLUCOSE SIGNALING PROBES
Feasibility studies of doped lenses with published boronic acid containing fluorophores had produced poor glucose responses, rationaled as due to the mildly acidic pH and methanol like polarity within the lens and the subsequent effect on the CT mechanism. This was also not surprising given the fact that these probes were designed for sensing at a physiological pH of ≈7.2, the probes typically having pKa around 9. Hence to obtain a notable glucose response in the contact lens polymer, it was deemed necessary to design new probes with significantly reduced sugar-bound pKa. In addition to the environmental parameters and constraints of pH and polarity, the probes also have to be sensitive to the very low concentrations of tear glucose, ≈ 500 μM for a healthy person, increasing up to several mM in diabetics, recalling that the blood glucose levels for a healthy person are ≈10-fold higher [14–17].
The pKa of phenyl boronic acid is known to be tunable with the appropriate substituents [44], for example, an electron withdrawing group reduces the pKa while an electron donating group increases the pKa of the sugar bound form. We therefore considered the interaction between the quaternary nitrogen of the 6-methyl and methoxy quinolinium moieties, and the boronic acid group, which reduces the pKa of the probe. In this regard we have synthesized 2 new classes of isomeric boronic acid containing probes (8-probes in total) (Fig. 1), where the spacing between the interacting moieties, quaternary nitrogen of the 6-methyl or methoxy quinolinium and boronic acid groups, enables both an understanding of the sensing mechanism to be realized, and the selection of the most suitable isomer based on its glucose binding affinity. In addition, control compounds (BMQ and BMOQ), which do not contain the boronic acid moiety, and are therefore insensitive towards sugar, were synthesized to understand the spectral properties and responses of the probes (Fig. 1).
EXPERIMENTAL
Materials and Lens Doping
All chemicals were purchased from Sigma. The contact lenses were supplied by CIBA Vision, Atlanta, USA, part of their daily disposable contact lens range, and were stirred in 500 mL water, 20°C for 24 hr before post-doping. The contact lens is a Polyvinyl alcohol type photocured polymer which swells slightly in water. Its hydrophilic character readily allows for the diffusion of the aqueous analytes in tears.
Doping was undertaken by incubating the lenses in a high concentration of the respective BAFs solution for 24 hr before being rinsed in Millipore water. Lenses were used directly after being prepared.
The preparation of the BAFs (Fig. 1) was undertaken as briefly follows, and are described in more detail elsewhere [39,40].
Methyl Quinolinium BAFs
The Boronic acid containing probes (o-BMQBA–N-(2-boronobenzyl)-6-methylquinolinium bromide, m-BMQBA–N -(3-boronobenzyl)-6-methylquinolinium bromide, p-BMQBA–N-(4-boronobenzyl)-6-methylquinolinium bromide) and the control compound (BMQ–N-benzyl-6-methylquinolinium bromide), were conveniently prepared using the following generic one step synthetic procedure, described below for the control compound BMQ. The corresponding o-, m-, or p-boronobenzyl bromides are employed instead of benzyl bromide to obtain the isomeric boronic acid derivatives o-, m- and p-BMQBA, respectively (Fig. 1).
Equimolar amounts of 6-methylquinoline and benzylbromide were dissolved in 10 mL dry acetonitrile in a 25 mL round bottomed flask equipped with a magnetic stirrer. The reaction mixture was allowed to stir under an inert atmosphere for 24 hr at room temperature. During this time a quantitative amount of quaternized salt was precipitated as a colorless solid. The solid product was recovered by filtration, washed several times with dry acetonitrile, and then dried under vacuum for 12 hr. The NMR and HRMS data obtained for all 8 compounds was consistence with the structure of the probes.
Methoxy Quinolinium BAFs
The glucose signaling probes based on the 6-methoxy quinoline nucleus; o-, m- and p-BMOQBA and a control compound BMOQ (Fig. 1) were prepared in an analogous manner to the methyl quinolinium probes described above, using 6-methoxy quinoline instead of 6-methyl quinoline as the starting reagent.
The Contact Lens Holder
The quartz lens holder which was used in all the lens studies, has dimensions of 4 × 2.5 × 0.8 cm, all four sides being of optical quality. The contact lens is mounted onto a stainless steel mount of dimensions 4 × 2 × 0.4 cm which fits tightly within the quartz outer holder. A circular hole in the center of the mount with a 2.5 cm ID, has a raised quartz lip, which enables the lens to be mounted. The mount and holder readily allow for ≈1.5 cm3 of solution to be in contact with the front and back sides of the lens for the sugar sensing experiments (Fig. 6 left).
Methods
All steady-state fluorescence measurements were undertaken in 4 × 1 × 1 cm fluorometric plastic cuvettes, using a Varian Cary Eclipse fluorometer, and all absorption measurements were performed using a Varian UV/VIS 50 spectrophotometer.
Time-resolved intensity decays were measured using reverse start-stop time-correlated single-photon timing (TCSPC), with a Becker and Hickl Gmbh 630 SPC PC card and unamplified MCP-PMT. Vertically polarized excitation at ≈372 nm was obtained using a pulsed LED source (1 MHz repetition rate) and a dichroic sheet polarizer. The instrumental response function was ≈1.1 ns fwhm. The emission was collected at the magic angle (54.7°), using a long pass filter (Edmund Scientific) which cut-off the excitation wavelengths.
Doped contact lenses were mounted in a custom made (CIBA Vision) lens holder (Fig. 6), which was itself inserted into a quartz holder for fluorescence sensing measurements. Excitation and emission was performed using a Varian Fluorometer, where the geometry shown in Fig. 6 (right), was employed to reduce any scattering of the excitation light, the concave edge of the lens facing towards the excitation source. We additionally tested the lens excited from the convex edge, just as would be used in the eye, and encouragingly found identical results.
Probe Leaching
Leaching of the probes from the contact lens polymer was observed using the sample holder shown in Fig. 6, which contained ≈1.5 cm3 buffer, 20°C. A Varian fluorometer measured the intensity change as a function of time to determine the percentage signal change, corresponding to dye leaching. It should be noted that with no sample present, no intensity fluctuations or drifts were observed, indicating stability of the fluorometer Xenonarc source.
Data Analysis
Titration curves with pH were determined in buffer solution: pH 3 and 4 acetate buffer; pH 5–9 phosphate buffer and pH 10 and 11 carbonate buffer. Titration curves were fitted and pKa (pKa = −Log10Ka) values were obtained using the relation:
(1) |
where Iacid and Ibase are the intensity limits in the acid and base regions respectively.
Stability (KS) and dissociation (KD) constants were obtained by fitting the titration curves, with sugar, using the relation:
(2) |
where Imin and Imax are the initial (no sugar) and final (plateau) fluorescence intensities of the titration curves, where KD = (1/KS).
The fluorescence intensity decays were analyzed in terms of the multi-exponential model:
(3) |
where αI are the amplitudes and τI the decay times, ΣαI = 1.0. The fractional contribution of each component to the steady-state intensity is given by:
(4) |
The mean lifetime of the excited state is given by:
(5) |
and the amplitude-weighted lifetime is given by:
(6) |
The values of αI and τI were determined by non-linear least squares impulse reconvolution with a goodness-of-fit criterion [41].
RESULTS AND DISCUSSION
Photophysical Characterization of the Quinolinium Probes
Figure 7 shows representative absorption and emission spectra for the ortho-isomers of the BMOQBA and BMQBA probes. Photophysical data of the probes are also shown in Table I. Typical absorption and emission band maximum of the probes can be seen at 318 and 450 nm, and 319 and 427 nm for the BMOQBA and BMQBA probes respectively. The additional absorption band at ≈350 nm for BMOQBA is attributed to the n → π* transition of the oxygen [45]. The excitation independent emission band at ≈450 nm indicates only one ground-state species is present for both classes of probes. The large Stokes-shifted fluorescence emission band of ≈100 nm is ideal for fluorescence sensing, allowing easy discrimination of the excitation wavelengths [41,45]. All probes were found to be readily water soluble.
Table I.
o-BMOQBA | m-BMOQBA | p-BMOQBA | BMOQ | |
---|---|---|---|---|
λabs(max)/nm | 318, 346 | 318, 347 | 318, 346 | 318, 347 |
λem(max)/nm | 450 | 450 | 451 | 453 |
ϕ f | 0.46 | 0.51 | 0.49 | 0.54 |
τf/ns | 26.7 | 25.9 | 24.9 | 27.3 |
o-BMQBA | m-BMQBA | p-BMQBA | BMQ | |
| ||||
λabs(max)/nm | 319 | 322 | 322 | 322 |
λem(max)/nm | 427 | 427 | 427 | 427 |
ϕ f | 0.043 | 0.025 | 0.023 | 0.045 |
τf/ns | 4.0a | 3.72a | 2.10a | 2.59 |
Mean fluorescence lifetime, Eq. 5.
Table I shows the quantum yield values for the probes in water, obtained from a spectral comparison with N-(3-Sulfopropyl)-6-methoxyquinolinium [(SPQ) (Φf = 0.53 in water [46])], where we can see that the BMOQBA probes have significantly larger quantum yields as compared to the BMQBA probes. Another reference compound, N-methyl-6-methylquinolinium bromide (MMQ) previously published by the authors [47] exhibits very similar spectral properties, except for a noticeable quantum yield and mean lifetime difference, approximately 10-fold higher than its methyl quinolinium counterpart, namely BMQBA. This indicates an interaction between the phenyl ring and quinolinium moiety for the BMQBA and BMQ probes, Table I, which is not present, or present to a much lesser extent, for the BMOQBA probes. We have attributed the relatively shorter lifetime and quantum yields of the new BMQBA probes and control compound to a photo-induced electron transfer mechanism, where the phenyl ring of BMQBA is the donor, and the quaternary nitrogen heterocyclic center is the acceptor. On this note, the BMOQBA probes were found to have monoexponential lifetimes (≈24.9 → 26.7 ns) as compared to the BMQBA probes which were biexponential in water, with significantly reduced lifetimes of 2.18 (46% amplitude) and 4.74 ns (54% amplitude), for the ortho-isomer. Both control compounds were found to have monoexponential lifetimes in water, with the BMOQ over ten times longer, 27.3 ns, as compared to BMQ, 2.59 ns. These lifetime changes further support the electron transfer hypothesis, the B−(OH)3 present at neutral pH further reducing the lifetime of the boronic acid probes, i.e. 2 in scheme 1.
In addition to the quantum yield and fluorescence lifetime differences between the phenyl ring containing BMQBA and BMQ probes, and the MMQ probe, we can also see lifetime differences between the BMQBA isomers themselves. We have attributed these changes due to the changes in electron donating ability of the different phenyl isomers, and additionally to their different through-space/through-bond interactions [48,49] with the positively charged quaternary nitrogen center, noting again that some B−(OH)3 is likely to be present at neutral pH.
The quantum yield values of the BMOQBA isomers are slightly lower than that of the control compound (BMOQ), the quantum yield values increasing in the order ortho-, para- and meta-. In contrast the monoexponential fluorescence lifetimes of the isomers increased in the order para-, meta- and then ortho-, which was slightly surprising as the quantum yields and lifetimes usually change in unison. Similar to having the highest quantum yield, the control compound also had as expected, the longest lifetime, some 27.3 ns. One explanation for these differences between the isomers, lies in the interaction between the boronate ester form present in solution, B−(OH)3, 2 in Scheme 1, and the positively charged nitrogen center at neutral pH, the extent of which being determined by the isomer spacing.
The emission spectra of the new boronic acid containing probes, as the pH increases from 3 to 11, typically show a steady decrease in fluorescence intensity (data not shown). In contrast the control compounds, BMOQ and BMQ, having no boronic acid group, shows no change in fluorescence intensity. Subsequently the corresponding titration curves in the absence and in the presence of 100 mM glucose and fructose, obtained by plotting the normalized intensities at band maximum versus pH, are shown in Fig. 8.
The pKa values obtained using Eq. 1 and the titration curves shown in Fig. 8 are shown in Table II. We can see considerably reduced pKa values for the new phenylboronic acid containing fluorophores in buffered media, as compared to the typical boronic acid probes reported in the literature [31–36], which are in the range 8–9. The quaternary nitrogen of the quinolinium nucleus not only reduces the pKa of the probes, but also serves to stabilize the boronatediester, formed upon sugar complexation. This in turn increases the affinity of the probes for sugar as shown in Table III. Hence the reduced sugar bound pKa of these new probes, coupled with their increased glucose affinity, is most attractive for our glucose sensing contact lens application, noting our previous findings of a lens pH around 6.1 [14].
Table II.
Medium | o-BMOQBA | m-BMOQBA | p-BMOQBA |
---|---|---|---|
7.90 | 7.70 | 7.90 | |
Buffer | |||
+100 mM Glucose | 6.62 | 6.90 | 6.90 |
+100 mM Fructose | 4.80 | 5.00 | 5.45 |
o-BMQBA | m-BMQBA | p-BMOQBA | |
| |||
6.70 | 7.75 | 7.80 | |
Buffer | |||
+100 mM Glucose | 6.10 | 6.85 | 6.95 |
+100 mM Fructose | 5.00 | 5.05 | 5.45 |
Table III.
Glucose |
Fructose |
|||
---|---|---|---|---|
Probe | Buffer | Lens | Buffer | Lens |
o-BMOQBA | 49.5 | 322.6 | 0.65 | 84.7 |
M-BMOQBA | 1000 | 54.6 | 1.8 | 4.9 |
P-BMOQBA | 430.0 | 111.1 | 9.1 | 34.7 |
o-BMOQBA | 100 | 17.9 | 4.7 | 34.8 |
M-BMOQBA | 476 | 58.1 | 13.2 | 21.6 |
P-BMOQBA | 370 | 128.2 | 13.8 | 12.9 |
To understand this new fluorescence signaling base mechanism and the response of the probes towards both pH and monosaccharides, it is informative to consider Fig. 9. The boronic acid group is an electron-deficient Lewis acid having an sp2-hybridized boron atom with a triangular conformation. The anionic form of the boronic acid, formed in high pH solutions, is characterized by a more electron rich sp3-hybridized boron atom with a tetrahedral geometry. The change in the electronic properties and the geometry at the boron atom induces the fluorescence spectral changes of the probes. It is well-known that the quinine/quinoline compounds exhibit high quantum yields in acidic media, from the corresponding quaternized salt [46,47]. Similarly here, the boronic acid probes are more fluorescent in acidic solutions. However, when the pH of the medium is increased the electron density on the boron atom is increased, facilitating the partial neutralization of the positively charged quaternary nitrogen of the quinolinium moiety. We have termed this interaction as a charge neutralization–stabilization mechanism, and a schematic representation of this mechanism with regard to glucose binding/sensing is illustrated in Fig. 9. In contrast, the control compounds BMQ and BMOQ are unperturbed.
The Monosaccharide induced spectral changes of the probes are shown in Fig. 10. In an analogous manner to that described for increasing pH above, we observed a systematic decrease in fluorescence intensity of the boronic acid containing probes in pH 7.5 phosphate buffer, for increasing glucose concentrations. The corresponding titration curves obtained by plotting I′ divided by I, where I′ and I are fluorescence intensities at 427 nm for BMQBA and 450 nm for BMOQBA, in the absence and presence of sugar respectively, versus glucose concentration, are also shown in Fig. 11. The right-hand column of Fig. 11 shows the response of the probes in the tear glucose concentration range.
For the BMOQBA probes we typically see a greater response for the para- isomer, with a 2.4-fold change in signal for 50 mM glucose. Interestingly, a ≈13% change in signal is observed up to 2 mM glucose, noting that glucose levels in tears can change from ≈500 μM for a healthy person, up to 5 mM for diabetics. For ortho-BMQBA, then a ≈16% change in fluorescence signal can be observed over a similar glucose range. Interestingly, for both classes of probes, the para-isomers show the weakest response towards glucose (Fig. 11). Given that the through-bond mechanism is expected to be prevalent for the para-isomers, and the ortho-isomer is expected to show both through bound and through space interactions, then a greater response was indeed expected, as is observed, for the ortho-isomer. However, the much weaker response of the para-isomer suggests that steric hindrance, with regard to glucose binding, is not present here.
The dissociation constants of the probes with both glucose and fructose in pH 7.5 phosphate buffer are presented in Table III, calculated using Eq. 2. As expected, a higher affinity for fructose is observed (lower KD value), which is a general observation for monophenyl boronic acid derivatives [31–36], but it should be noted that the concentration of fructose in tears is substantially lower than for glucose [14–17]. A comparison of the trends in glucose response observed in Fig. 11, and the recovered KD values in Table III, show some differences, which we have attributed to the difficulties encountered during data fitting. While beyond the scope of this text, these fitting difficulties clearly reflect the need for a new kinetic sugar binding function with our new probes. Further studies are underway in this regard.
Response of the Glucose Signaling Probes in the Contact Lens
Doped contact lenses, which were previously washed and allowed to leach dye for 1 hr were tested with both glucose and fructose. Buffered solutions of sugars were added to the lens, pH 7.5 phosphate buffer, in an analogous manner to ocular conditions. Fluorescence spectra were typically taken 15 min after each sugar addition to allow the lens to reach equilibrium. The 90% response time, the time for the fluorescence signal to change by 90% of the initial value, was ≈10 min.
Figure 12 shows the response of o-BMOQBA and o-BMQBA, Top and Bottom left respectively, for increasing concentrations of glucose injected into the 1.5 cm3 contact lens volume. Similar to the solution based measurements, the probes show a decrease in fluorescence intensity, which we attribute to the complexation of glucose with boronic acid and the subsequent charge neutralization mechanism described earlier. We were again able to construct the I′ /I plots, where I′ is the intensity in the absence of sugar (Fig. 12 right). As was observed in solution, Fructose had a greater response, reflecting the greater affinity of mono phenyl boronic acid derivatives for fructose. However, in the low sugar concentration ranges, <2 mM sugar, the response towards both sugars was comparable [15–17]. Differences in the response towards glucose for the isomers could also be observed in the lenses (Fig. 13), where m-BMOQBA was found to have a greatest response amongst this class of probes. From Fig. 13 we can clearly we see a greater response in lens towards sugars than in our solution based studies at pH 7.5, with p-BMQBA showing a greater than 20% fluorescence signal change with as little as 2 mM glucose. This wasn’t unexpected, and is simply explained by the pKa of the probes being <2 7, the probes being compatible with the mildly acidic environment. Figure 14 directly compares the response of the probes in both contact lens and buffer, where we see a comparable if not better response towards glucose in the lens. We are uncertain at this time however, as to the nature of the smaller fluorescence response at higher glucose concentrations in the lens, but we speculate that it may be due to the greater leaching of the glucose-bound (boronate diester) form of the probes and/or their displaced solubility within the contact lens polymer. This may also account for some of the complex binding kinetics we have observed, evident in some I′ /I Vs [sugar] plots. In this regard the response of doped lenses also show complex behavior towards mM fructose concentrations, with much simpler kinetics observed in the tear glucose concentration range (data not shown).
For all our glucose contact lens sensing studies we repeated these doped lens experiments several times, and in all cases the trends were reproducible. It is difficult to assess the effect of the PVA hydroxyl groups of the contact lens polymer on the response of boronic acid to sugar, but our studies with solutions of glycerol indicated that sugar had much higher binding affinities than glycerol hydroxyl groups. We therefore speculate that sugars will preferentially bind boronic acid groups in the PVA lens polymer. In any event the boronic acid probes function well (reversibly) towards sugars in the lens, in what is likely to be an environment saturated with PVA hydroxyl groups.
Probe Leaching, Interferents, and Shelf Life
Leaching studies of the probes from the contact lens polymer were undertaken using the lens holder, which contained ≈1.5 cm3buffer, 20°C. A Varian fluorometer measured the intensity change as a function of time to determine the percentage signal change, corresponding to dye leaching. It should be noted that with no sample present, no intensity fluctuations or drifts were observed, indicating stability of the fluorometer Xenon-arc source.
We were able to observe up to an 8% change in fluorescence intensity, attributed due to dye leaching, for the BMOQBA class of probes, with very little change after about 25 min (Fig. 15). In contrast, the BMQBA probes show a much greater extent of leaching over the same time period and under identical conditions. Given that the BMQBA probes typically showed a greater response towards glucose in the lens, then this suggests that the BMQBA probes may be more accessible in the lens, than the BMOQBA probes. In addition, similar results were obtained at 38°C but with a different leaching rate.
In all our lens response studies described here, lenses were pre-leached to a steady-state fluorescence intensity before use. After glucose measurements were undertaken the outer lens fluid volume surrounding the contact lens was found to be non-fluorescent indicating that dye had not leached from the lens during actual glucose sensing measurements. It should be noted that while chemistries are available to covalently label our probes within the contact lens polymer, which would eliminate any leaching, it is an important design concern for our approach that the lenses remain unmodified, so that their physiological characteristics and compatibilities remain unchanged. In fact our approach is targeted at reducing future lens redesign costs for industry, by using simple probe doping.
As with all sensors it is important to consider the effects of potential interferents and sensor shelf-life on the working response of the device. Throughout much of this paper we have shown the response of the probes towards fructose, primarily because of its well-known greater affinity for the boronic acid moiety [31–36]. However, the concentration of fructose in blood is ≈10 times lower than glucose [14], a relationship which is also thought to occur in tears [14]. Hence fructose is not thought to be a major interferent in tears, simply shown here to place the binding trends in context. However, tears are a complex mixture of proteins and other analtyes, such as sodium (120–170 mM), potassium (6–26 mM) and chloride (100 mM) [50]. We subsequently tested these new contact lens probes with various aqueous halides, given that Na+ and K+ are unlikely to perturb our fluorophores (Table IV). As expected, the BMOQBA probes are modestly quenched by chloride, with steady-state Stern-Volmer constants in the range 170–182 M−1, the BMQBA probes having significantly smaller quenching constants in the range 17–44 M−1. This result is simply explained by the shorter lifetime of the BMQBA probes as compared to the BMOQBA probes, and the probability of an excited-state chloride ion encounter [41,45]. Encouragingly, the BMQBA probes typically showed a greater response towards glucose as well as being the least perturbed by aqueous chloride. In any event, simple corrections in the fluorescence signal can readily account for chloride interference on the glucose response. This is shown in Fig. 2. Sensors spots on the surface of the lens could contain a reference chloride compound or indeed another probe sensitive to both glucose and chloride. By employing the extended Stern-Volmer equation for multiple analytes [45], one can easily correct for background interferences.
Table IV.
Probe | CI− (Ksv/M−1) | Br− (Ksv/M−1) | I− (Ksv/M−1) |
---|---|---|---|
o-BMOQBA | 170 | 332a | 471a |
M-BMOQBA | 182 | 413 | 540 |
P-BMOQBA | 177 | 370 | 959 |
BMOQB | 222 | 384 | 520 |
o-BMQBA | 44.0 | 55.0 | 97.0 |
M-BMQBA | 20.0 | 32.0 | 48.0 |
P-BMQBA | 17.0 | 26.5 | 42.0 |
BMQB | 35.0 | 55.0 | 71.0 |
The concentrations of Br− and I− in tears is extremely low and is unlikely to be an interferent in our glucose measurements.
It is also informative to consider the pH of tears as a potential interferent, given the response of these probes to pH as shown in Fig. 8. It is known that unstimulated tear pH levels can vary in the range 7.14–7.82 measured from healthy subjects at different times of the day, with a typical mean value around pH 7.45 [50]. However, a more acidic pH of less than 7.3 is found following prolonged lid closure, e.g. after sleep, which is thought to result from carbon dioxide produced by the cornea and trapped in the tear pool under the eye lids. While solutions of these new probes would be susceptible to these changes in pH, we have found that the doped lenses we studied were indeed unbufferable, hence external changes in pH are most unlikely to affect the glucose response.
With regard to the glucose-sensing contact lens shelf-life, lenses that had been doped, leached and stored for several months both wet and dry, gave identical sugar sensing results, indicating no lens polymer–fluorophore interactions over this time period, or indeed probe degradation.
FUTURE DEVELOPMENTS BASED ON THIS TECHNOLOGY
Continuous and Non-Invasive Glucose Monitoring
In this review article of work from our laboratories, we have shown that fluorescent probes can be fabricated to be compatible with commercially available contact lenses, which have already been assessed and optimized with regard to vision correction and oxygen permeability. This has enabled the first prototype based on this new approach to be realized. With regard to glucose monitoring by this approach, we speculate on several future improvements to this technology.
Clear or colored contact lenses
Many boronic containing fluorophores have a visible absorption [31–36], which apart from their lack of glucose sensitivity in the lens as discussed earlier [14], would introduce color into a doped lens. While colored lenses are attractive to a few people as sports or even fashion accessories, the majority of contact lenses worn today are clear, hence our colorless lenses described here are ideal in this regard. One disadvantage however with our lenses is the requirement for an excitation and detection device as shown in Fig. 2. One improvement to our technology could be the use of colored contact lenses, which change color due to the concentration of tear and therefore blood glucose. This can be achieved by the ground-state binding of glucose to boronic acid and the subsequent changes in fluorophore absorption spectrum. A patient wearing the lenses could simply look in the mirror, or the color even determined by an on-looker, and compared to a precalibrated color strip, one could assess the extent of hyperglycemia. This technology would be most attractive to parents of young diabetic children or for care workers of the elderly. Work in currently underway in our laboratories in this regard.
Sensor spots or doped lenses
As briefly mentioned earlier and shown in Fig. 2, sensor spots on the surface of contact lenses could correct signal responses for interferents such as aqueous chloride. Indeed the spots could either be visible to the wearer (self-readout) or readable by an external monitoring device.
Detection methods
While simple colorimetric methods are likely to be the most simple to introduce to the market place, other fluorescence sensing methodologies, such as polarization, lifetime and ratiometric sensing, offer many spectroscopic sensing advantages over the simple intensity measurements described for our lenses [41]. For example, fluorescence lifetime and ratiometric measurements are independent on total light intensity or indeed fluctuations in ambient room light.
Clinical Condition and Diagnosis from Tears
Given that tear glucose can be continuously and noninvasively monitored using our lens approach, then it may be possible for both clinical condition assessment and disease diagnosis using the contact lens sensing platform, and suitably designed fluorophores. In fact, as compared to saliva, tears represent a more stable body fluid of low protein concentration and with only modest variations in pH.
For example, for clinical condition assessment, Na+, K+, Ca2+, Mg2+, pH, Histamine, Urea, Lactate, Cholesterol and Glucose in tears are known to directly track or relate to the serum levels [50]. Indeed, one could potentially even track body core temperature using thermochromic type probes [51], embedded within contact lenses or even sensor spots.
For disease diagnosis, the possibility of diagnosing; Glaucoma, Sjogrens disease, Lysosomal storage diseases, corneal ulceration and bacterial infections could be realized by designing lenses to detect; Catecholamines, Lysozyme, Lysosomal enzymes, Collagenase and α-Antitrypsin respectively. It is possible that many other clinical conditions and diseases could also be either monitored or diagnosed via this approach, although relatively little tear biochemistry is known [50].
Drug Testing, Compliance, and Screening
Saliva is and has been used for therapeutic drug monitoring, by predicting the free fraction of drugs in blood from that determined in saliva. Changes in the free drug levels can have important clinical consequences, since either toxic or sub-therapeutic levels may exist, even when the total drug concentration is in the normal range. Coupled with the fact that saliva concentrations varying greatly in both pH and composition, obviates the need for a novel clinical sensing platform for drug testing, compliance and screening. As mentioned earlier, tear fluid has a relatively lower protein concentration with only slight changes in pH, where the passage of drugs from plasma to tears takes place by diffusion of the non-protein bound fraction [50]. Subsequently, tears have already been used to assess the concentration of antibiotics such as Ampicillin and the Anticonvulsants, Phenobarbital and Carbamazepine [50]. However, these drug concentration measurements are inherently difficult due to the 5–10 μL total tear volume to be sampled [50]. Indeed, while tear glucose levels have been known to be elevated during hyperglycemia for nearly 70 years, it is the difficulties associated with tear collection, which has limited the practical use of tears for diabetes mellitus assessment. However, similar to our glucose sensing contact lenses, it may be possible to develop lenses for drug testing, based on either colorimetric or other fluorescence spectroscopic based methodologies.
CONCLUSIONS
We have developed a range of new glucose sensing contact lenses, by doping strategically designed fluorescent probes into commercially available contact lenses. The probes are completely compatible with the new lenses and can readily detect glucose changes up to several mM glucose, appropriate for the tear glucose concentration range for diabetics, i.e. ≈ 500 μM → 5 mM [14].
The lenses have a 90% response time of about 10 min, allowing the continuous and noninvasive monitoring of ocular glucose. This is a significant improvement over enzymatic methods based on blood sampling by finger pricking, with many diabetics begrudgingly testing between 4 and 6 times daily.
With diabetes being widely recognized as one of the leading causes of death and disability in the western world, we believe our boronic acid doped contact lens approach and findings, are a notable step forward towards the continuous and non-invasive monitoring of physiological glucose.
ACKNOWLEDGMENTS
The authors would like to thank the University of Maryland Biotechnology Institute and the NIH, National Center for Research Resources, RR-08119, for financial support. Drs Angelika Domschke and Dawn Smith of Cibavision are also acknowledged for supplying the contact lens holder shown in Fig. 6.
Abbreviations
- BA
Boronic Acid
- BAFs
Boronic acid containing fluorophores
- BMQ
N-Benzyl-6-methylquinolinium bromide
- BMOQ
N-Benzyl-6-methoxyquinolinium bromide
- BMQBA
N-(Boronobenzyl)- 6-methylquinolinium bromide
- BMOQBA
N-(Boronobenzyl)-6-methoxyquinolinium bromide
- CT
Excited-state Charge Transfer
- DDPBBA
1-(p-Boronophenyl)-4-(p-dimethylaminophenyl)buta-1,2-deine
- HRMS
High-Resolution Mass Spectroscopy
- LD
Laser Diode
- LED
Light Emitting Diode
- PVA
Polyvinyl alcohol
- TCSPC
Time-correlated Single Photon Counting
Footnotes
The term “Diabetes” was first used around 230 BC by Apollonius of Mephis, which in Greek means “to pass through” (Dia– through, betes – to go) [1].
REFERENCES
- 1.Poretsky Leonid., editor. Principles of Diabetes Mellitus. Kluwer Academic/Plenum; Norwell, MA: 2002. [Google Scholar]
- 2.Medvei VC. In The History of Clinical Endocrinology: A Comprehensive Account of Endocrinology from Earliest Times to Present Day. Parthenon, New York: 1993. [Google Scholar]
- 3.Robinson MR, Eaton RP, Haaland DM, Koepp GW, Thomas EV, Stallard BR, Robinson PL. Non-invasive glucose monitoring in diabetic patients: A preliminary evaluation. Clin. Chem. 1992;38:1618–1622. [PubMed] [Google Scholar]
- 4.Heise HM, Marbach R, Koschinsky TH, Gries FA. Non-invasive blood glucose sensors based on near-infrared spectroscopy. Ann. Occup. Hyg. 1994;18:439–447. doi: 10.1111/j.1525-1594.1994.tb02230.x. [DOI] [PubMed] [Google Scholar]
- 5.March WF, Rabinovitch B, Adams R, Wise JR, Melton M. Ocular Glucose sensor. Trans. Am. Soc. Artif. Intern. Organs. 1982;28:232–235. [PubMed] [Google Scholar]
- 6.Rabinovitch B, March WF, Adams RL. Non-invasive glucose monitoring of the aqueous humor of the eye, Part 1: Measurement of very small optical rotations. Diabetes Care. 1982;5:254–258. doi: 10.2337/diacare.5.3.254. [DOI] [PubMed] [Google Scholar]
- 7.Schier GM, Moses RG, Gan IET, Blair SC. An evaluation and comparison of reflolux Iiand Glucometer II, two new portable reflectance meters for capillary blood glucose determination. Diabetes Res. Clin. Pract. 1988;4:177–181. doi: 10.1016/s0168-8227(88)80015-9. [DOI] [PubMed] [Google Scholar]
- 8.Clarke W, Becker DJ, Cox D, Santiago JV, White NH, Betschart J, Eckenrode K, Levandoski LA, Prusinki EA, Simineiro LM, Snyder AL, Tideman AM, Yaegar T. Evaluation of a new system for self blood glucose monitoring. Diabetes Res. Clin. Pract. 1988;4:209–214. doi: 10.1016/s0168-8227(88)80020-2. [DOI] [PubMed] [Google Scholar]
- 9.Trettnak W, Wolfbeis OS. Fully reversible fiber-optic glucose biosensor based on the intrinsic fluorescence of glucose-oxidase. Anal. Chim. Acta. 1989;221:195–203. [Google Scholar]
- 10.Meadows D, Schultz JS. Fiber optic biosensor based on fluorescence energy transfer. Talanta. 1988;35:145–150. doi: 10.1016/0039-9140(88)80053-5. [DOI] [PubMed] [Google Scholar]
- 11.Tolosa L, Malak H, Rao G, Lakowicz JR. Optical assay for glucose based on the luminescence decay time of the long wavelength dye Cy5. Sens. Actuators B. 1997;45:93–99. doi: 10.1016/S0925-4005(97)00275-X. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Tolosa L, Gryczynski I, Eichorn LR, Dattelbaum JD, Castellano FN, Rao G, Lakowicz JR. Glucose sensors for low cost lifetime-based sensing using a genetically engineered protein. Anal. Biochem. 1999;267:114–120. doi: 10.1006/abio.1998.2974. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.D’Auria S, Dicesare N, Gryczynski Z, Gryczynski I, Rossi M, Lakowicz JR. A thermophilic apoglucose dehydrogenase as a nonconsuming glucose sensor. Biochem. Biophys. Res. Commun. 2000;274:727–731. doi: 10.1006/bbrc.2000.3172. [DOI] [PubMed] [Google Scholar]
- 14.Badugu R, Lakowicz JR, Geddes CD. The noninvasive continuous monitoring of physiological glucose using a novel monosaccharide-sensing contact lens. Anal. Chem. 2004;76:610–618. doi: 10.1021/ac0303721. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Badugu R, Lakowicz JR, Geddes CD. A glucose sensing contact lens: A non-invasive technique for continuous physiological glucose monitoring. J. Fluorescence. 2003;13:371–374. doi: 10.1023/A:1026103804104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Geddes CD, Badugu R, Lakowicz JR. Contact lenses may provide window to blood glucose. Biophotoincs int. 2004 Feb;(2):50–53. [Google Scholar]
- 17.Badugu R, Lakowicz JR, Geddes CD. Ophthalmic glucose sensing: A novel monosaccharide sensing disposable and colorless contact lens. The Analyst. 2004;129:516–521. doi: 10.1039/b314463c. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Sugihara JM, Bowman CM. Cyclic benzeneboronate esters. J. Am. Chem. Soc. 1958;80(10):2443–2446. [Google Scholar]
- 19.Lorand JP, Edwards JO. Polyol complexes and structure of the benzeneboronate ion. J. Org. Chem. 1959;24(6):769–774. [Google Scholar]
- 20.Springsteen G, Wang B. A detailed examination of boronic acid–diol complexation. Tetrahedron. 2002;58(26):5291–5300. [Google Scholar]
- 21.James TD, Sandanayake KRAS, Shinkai S. Chiral discrimination of monosaccharides using a fluorescent molecular sensor. Nature. 1995;374:345. [Google Scholar]
- 22.Norrild JC, Eggert H. Evidence for monodentate and bidentate boronate complexes of glucose in the furanose form—Application of (1)J(C–C)-coupling-constants as a structural probe. J. Am. Chem. Soc. 1995;117(5):1479–1484. [Google Scholar]
- 23.Eggert H, Frederiksen J, Morin C, Norrild JC. A new glucose-selective fluorescent bisboronic acid. First report of strong alpha–furanose complexation in aqueous solution at physiological pH. J. Org. Chem. 1999;64(11):3846–3852. [Google Scholar]
- 24.Yang W, He H, Drueckhammer DG. Computer-guided design in molecular recognition: Design and synthesis of a glucopyranose receptor. Angew. Chem. Int. Ed. 2001;40(9):1714–1718. [PubMed] [Google Scholar]
- 25.Wang W, Gao S, Wang B. Building fluorescent sensors by template polymerization: The preparation of a fluorescent sensor for D-fructose. Org. Lett. 1999;1(8):1209–1212. doi: 10.1021/ol9908732. [DOI] [PubMed] [Google Scholar]
- 26.Gao S, Wang W, Wang B. Building fluorescent sensors for carbohydrates using template-directed polymerizations. Bioorg. Chem. 2001;29:308–320. doi: 10.1006/bioo.2001.1219. [DOI] [PubMed] [Google Scholar]
- 27.Lavigne JJ, Anslyn EV. Teaching old indicators new tricks: A colorimetric chemosensing ensemble for tartrate/malate in beverages. Angew. Chem. Int. Ed. 1999;38(24):3666–3669. [PubMed] [Google Scholar]
- 28.Yoon J, Czarnik AW. Fluorescent chemosensors of carbohydrates. A means of chemically communicating the binding of polyols in water based on chelation-enhanced quenching. J. Am. Chem. Soc. 1992;114:5874–5875. [Google Scholar]
- 29.Smith BD, Gardiner SJ, Munro TA, Paugam MF, Riggs JA. Facilitated transport of carbohydrates, catecholamines, and amino acids through liquid and plasticized organic membranes. J. Incl. Phenom. Mol. Recogn. Chem. 1998;32:121–131. [Google Scholar]
- 30.Soundararajan S, Badawi M, Kohlrust CM, Hagerman JH. Boronic acids for affinity–chromatography–spectral methods for determinations of ionization and diol-binding constants. Anal. Biochem. 1989;178:125–134. doi: 10.1016/0003-2697(89)90367-9. [DOI] [PubMed] [Google Scholar]
- 31.James TD, K. R. A. S., Shinkai S. A glucose-selective molecular fluorescence sensor. Angew, Chem. Int. Ed. 1994;33(21):2207–2209. [Google Scholar]
- 32.James TD, Sandanayake KRAS, Iguchi R, Shinkai S. Novel saccharide-photoinduced electron-transfer sensors based on the interaction of boronic acid and amine. J. Am. Chem. Soc. 1995;117(35):8982–8987. [Google Scholar]
- 33.Dicesare N, Lakowicz JR. Evaluation of two synthetic glucose probes for fluorescence-lifetime based sensing. Anal. Biochem. 2001;294:154–160. doi: 10.1006/abio.2001.5170. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Dicesare N, Lakowicz JR. Wavelength-ratiometric probes for saccharides based on donor-acceptor diphenylpolyenes. J. Photochem. Photobiol. A: Chem. 2001;143:39–47. doi: 10.1016/S1010-6030(01)00471-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Dicesare N, Lakowicz JR. New color chemosensors for monosaccharides based on Azo dyes. Org. Lett. 2001;3(24):3891–3893. doi: 10.1021/ol016813p. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Dicesare N, Lakowicz JR. Chalcone-analogue fluorescent probes for saccharides signaling using the boronic acid group. Tet. Lett. 2002;43:2615–2618. doi: 10.1016/s0040-4039(02)00312-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Karnati VV, Gao X, Gao S, Yang W, Ni W, Sankar S, Wang B. A glucose-selective fluorescence sensor based on boronic acid-diol recognition. Bioorg. Med. Chem. Lett. 2002;12:3373–3377. doi: 10.1016/s0960-894x(02)00767-9. [DOI] [PubMed] [Google Scholar]
- 38.Dicesare N, Lakowicz JR. Charge transfer fluorescent probes using boronic acids for monosaccharide signaling. J. Biomed. Opt. 2002;7(4):538–545. doi: 10.1117/1.1502263. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Badugu R, Lakowicz JR, Geddes CD. Fluorescence sensors for monosaccharides based on the 6-methylquinolinium nucleus and boronic acid moiety: Application to ophthalmic diagnostics. Talanta. doi: 10.1016/j.talanta.2004.08.003. (manuscript submitted for publication) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Badugu R, Lakowicz JR, Geddes CD. Boronic acid fluorescent sensors for monosaccharide signaling based on the 6-methoxyquinolinium heterocyclic nucleus: Progress towards noninvasive and continuous glucose monitoring. Bioorg. Med. Chem. doi: 10.1016/j.bmc.2004.09.058. (manuscript submitted for publication) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Lakowicz JR. Principles of Fluorescence Spectroscopy. 2nd ed Kluwer/Academic Plenum; New York: 1997. [Google Scholar]
- 42.Turro NJ, Baretz BH, Kuo PI. Photoluminescence probes for the investigation of interactions between sodium dodecylsulfate and water-soluble polymers. Macromolecules. 1984;17(7):1321–1324. [Google Scholar]
- 43.Kalyanasundaram K, Thomas JK. Environmental effects on vibronic band intensities in pyrene monomer fluorescence and their application in studies of micellar systems. J. Am. Chem. Soc. 1977;99(7):2039–2044. [Google Scholar]
- 44.Dicesare N, Lakowicz JR. Spectral properties of fluorophores combining the boronic acid group with electron donor or withdrawing groups. Implication in the development of fluorescence probes for saccharides. J. Phys. Chem. A. 2001;105(28):6834–6840. doi: 10.1021/jp010076x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Geddes CD. Optical halide sensing using fluorescence quenching: Theory, simulations and applications—A review. Meas. Sci. Tech. 2001;12(9):R53. [Google Scholar]
- 46.Wolfbeis OS, Urbano E. J. Heterocyclic Chem. 1982;19:841–843. [Google Scholar]
- 47.Geddes CD, Apperson K, Karolin J, Birch DJS. Chloride sensitive probes for biological applications. Dyes Pigments. 2001;48:227–231. [Google Scholar]
- 48.Fox MA, Chanon M, editors. Photoinduced Electron Transfer Parts A–D. Elsevier; New York: 1998. [Google Scholar]
- 49.Kavarnos GJ. Fundamentals of Photoinduced Electron Transfer. VCH; New York: 1993. [Google Scholar]
- 50.Van Haeringen NJ. Climical biochemistry in tears. Survey Ophthalmol. 1981;26(2):84–96. doi: 10.1016/0039-6257(81)90145-4. [DOI] [PubMed] [Google Scholar]
- 51.Chandrasekharan N, Kelly L. In: Reviews in Fluorescence 2003. Geddes CD, Lakowicz JR, editors. Kluwer Academic/Plenum; New York: 2004. [Google Scholar]