Skip to main content
Journal of Research of the National Institute of Standards and Technology logoLink to Journal of Research of the National Institute of Standards and Technology
. 1991 Sep-Oct;96(5):529–540. doi: 10.6028/jres.096.029

Summary of the Intercomparison of the Force Standard Machines of the National Institute of Standards and Technology, USA, and the Physikalisch-Technische Bundesanstalt, Germany

Simone L Yaniv 1, A Sawla 2, M Peters 2
PMCID: PMC4927229  PMID: 28184127

Abstract

A comparison of force measurements performed at the National Institute of Standards and Technology, USA, and at the Physikalisch-Technische Bundesanstalt, Germany is reported. The focus of the study was the intercomparison of the forces realized by the two Institutes rather than the measurement process. The transfer standards used in the comparison consisted of force transducers and associated readout instrumentation. The results of the intercomparison reveal that over a range of 50 kN to 4.5 MN, the forces realized at NIST and at PTB compare favorably. For forces up to 900 kN the agreement is within ±40 ppm; above 900 kN the agreement is within ± 100 ppm.

Keywords: deadweight machines, force measurements, force standards, force transducers

1. Introduction

This paper summarizes the results of a comparison of force measurements performed at the National Institute of Standards and Technology (NIST), USA, and at the Physikalisch-Technische Bundesanstalt (PTB), Germany. A detailed description of the study can be found in Ref. [1].

The objective of the study was to determine the comparability of the forces realized by the two Institutes over a range of 50 kN to 4.5 MN, so that the results of force transducer calibrations performed at one Institute would be more readily accepted by the other. The need for the comparison was acute as there are significant differences in the force machines used by the two Institutes. At NIST the forces applied are generated by deadweights over the entire range included in the study. At PTB forces up to 1 MN are applied by deadweights but higher forces are generated by means of a hydraulic force-multiplication system.

The overall program for the intercomparison was developed jointly by NIST and PTB. The protocol used during the measurements was developed by PTB. Planning for the program began late in 1988. Initial measurements of all the force transducers involved in the comparison were carried out first at PTB in September 1989. These initial measurements were followed by a set of similar measurements at NIST in October 1989. To verify the stability of the force transducers used during the comparison, a final set of measurements was obtained at PTB in November 1989.

2. Force Standard Machines

The force standard machines of both Institutes have been described in details in Refs. [14]. Accordingly, only a brief description of the machines is given here.

2.1 The NIST Deadweight Force Standard Machines

Only the three largest NIST deadweight machines were included in the intercomparison. These machines, known as the 1 Mlbf, the 300 klbf and the 112 klbf machines have been described in Refs. [1,2]. The most important features of these machines are summarized in Table 1. Drawings of the two largest machines are given in Figs. 1 and 2, and a photograph of the 112 klbf machine is given in Fig. 3.

Table 1.

Characteristics of the three largest NIST deadweight machines

Capacity: (nominal)
 kN 4448  1334  498 
  (klbf) (1000) (300) (112)
Minimum load: (nominal)
 kN 222  44  13 
  (klbf) (50) (10) (3)
Minimum increment: (nominal)
 kN 222  44  4.4
  (klbf) (50) (10) (1)
Time to capacity, s 345  292  274 
Compression setup space:
 Vertical, m 1.98  1.65  1.02 
 Horizontal, m 0.86  0.91  0.71 
Tension setup space:
 Vertical, m 4.45  2.49  2.16 
 Horizontal, m 1.17  0.91  0.71 

Figure 1.

Figure 1

The NIST 1 Mlbf deadweight machine.

Figure 2.

Figure 2

The NIST 300 klbf deadweight machine.

Figure 3.

Figure 3

The NIST 112 klbf deadweight machine.

The estimated total uncertainty of the vertical component of force applied by any weight is 0.002 percent. If corrections are made for the actual air density during a calibration and for the actual adjusted mass of each weight, the uncertainty of the applied force can be reduced to 0.001 percent.

The schematic drawing in Fig. 4 illustrates the operating principles of the deadweight machines. Each of the three machines has a stack of large weights, represented by the lower stack of larger weights in Fig. 4. Only the 112 klbf machine also has a second stack of smaller weights that are operated by screw jacks. To apply a deadweight load, the hydraulic jack raises the lifting frame and the loading frame, acting through the device being calibrated in either compression or tension.

Figure 4.

Figure 4

Schematic drawing of the three largest NIST dead-weight machines.

As the loading frame is raised, the large weights are picked up in sequence, beginning with the top weight of the stack. In the 112 klbf machine, the smaller weights are lowered onto the loading frame by the screw jacks, in sequence, beginning with the bottom weight of the stack. The time required to apply the capacity load in each machine is given in Table 1. The vertical positions of the compression and tension platens of the lifting frame are adjusted as required to fit each calibration setup. The maximum setup space in each machine is given in Table 1. The safety nuts are adjusted so that they will support the deadweight load if the calibration setup becomes unstable. The vertical members of the loading frames are not clearly visible in Figs. 1 and 2 because the widths of the loading frames are approximately equal to the widths of the respective lifting frames.

2.2 The PTB Force Standard Machines

To cover the force range included in the inter-comparison, two PTB force standard machines were utilized. In the first machine, the 1 MN machine, forces are generated by deadweights. In the second machine, the 15 MN machine, forces are generated by means of a hydraulic force multiplication system. These machines have been described in detail in Refs. [1,3,4].

A schematic diagram of the 1 MN deadweight machine is shown in Fig. 5. A schematic view of the 15 MN hydraulic force standard machine is shown in Fig. 6.

Figure 5.

Figure 5

Schematic diagram of the PTB 1MN deadweight machine.

Figure 6.

Figure 6

Schematic diagram of the PTB 15 MN hydraulie foree standard maehine.

The overall uncertainty of the force realized in the hydraulic standard force machine has been estimated to be about 100 ppm [5]. The uncertainty of the 1 MN machine is comparable to that of the NIST deadweight machines.

3. Force Transfer Standards

A total of six force transducers were used in the intercomparison. Among these, five force transducers, having capacities of 100 kN, 200 kN, 500 kN, 1 MN, and 5 MN, have been in use at PTB over a substantial period of time. Accordingly, the behavior of these five force transducers is well known as is their long-term stability [1]. The history of the sixth transducer used, the 4.7 MN transducer, is not well documented. A complete description of each transducer can be found in Ref. [1].

To minimize the uncertainty associated with the indicating instrument a high-resolution indicator having a good stability was chosen. The resolution of the selected indicator is 1 ppm.

Prior to the start of the intercomparison, measurements were conducted at PTB, using a calibrating standard bridge to ascertain the reliability of the indicator when different main frequencies (50 and 60 Hz in Germany and USA, respectively) and different voltages (220 and 115 V in Germany and USA, respectively) are used. No significant differences were found.

4. Measurement Procedure

In developing the procedure utilized to perform the intercomparison, great care was given to minimize the effects of those parameters that are known to contribute to the measurement uncertainty. The following subsections describes these parameters and the ways in which their effects were minimized.

4.1 Time Interval

The difference between the output of a transducer at a load and its output when no load is applied represents the response of the transducer to that load. When a load is applied to a force transducer or when the force transducer is unloaded, there are initial mechanical, thermal and electrical responses in the various interconnected elements, followed by a delayed creep response or drift in the output of the transducer as the elements approach a new equilibrium condition. The process may be further complicated by local heating due to electrical-power dissipation by the strain-measuring bridge.

Although different force transducers exhibit different creep patterns [6], in general, the creep rate decreases greatly during the first few minutes following loading or unloading. To minimize the effect of creep, for each force transducer included in the intercomparison, the time required to achieve a stable response following loading and unloading was determined prior to the start of the intercomparison. In most instances it was found that a 3 min time delay between the initiation of the loading (or unloading) and the actual reading was adequate. When tests were conducted in the NIST 300 klbf and 1 Mlbf standard machines an additional 1.5 min was allowed to account for the longer loading and unloading times of these very-large deadweight machines.

In each instance, each set of measurements was duplicated once. In all cases a 3 min delay time was introduced between the completion of the initial set of measurements and the initiation of the duplicate set of measurements.

4.2 Machine-Transducer Interaction

Machine-transducer interactions can significantly influence measurement accuracy. Normal imperfections in the alignment of loading machines and force transducers can result in significant bending, shear, and twist components of deformation in the force transducer. To minimize the errors due to these nonaxial components of deformation, it is desirable to sample the response of the force transducer at several symmetrically distributed positions [7,8]. For this reason, the response of each force transducer was obtained at five positions relative to the axis of the machine (0°, 90°, 180°, 270°, 360°).

While at the 0° position and prior to the start of a measurement cycle, the force transducer was exercised by applying the maximum test load three times, returning to zero after each maximum load application. After a 3 min delay, with the force transducer still in the same position, two sets of measurements were obtained, each separated by a 3 min interval. Then, the force transducer was rotated by 90° and two new sets of measurements, each separated by a 3 min interval, were obtained, and so on. A special rotating mechanism, installed underneath the force transducer, allowed the transducer to be rotated rapidly through all the positions. During a measurement cycle (two sets of measurements at five positions) the force transducer was exercised only once at the beginning of the cycle at the 0° position.

4.3 Ambient Conditions

The measurements were carried out at (23 ± 0.5)°C, the usual laboratory conditions at NIST. Normally at PTB the laboratories are maintained at (20 ± 1) °C. However, a week prior to the initiation of the measurements, the temperature of the PTB laboratories was increased to (23 ± 1)°C. Both the force transducers and the indicator were kept at this temperature for a week prior to the initiation of measurements.

4.4 Force Steps

The forces realized in the PTB 15 MN standard machine were compared to the forces realized in the NIST 1 Mlbf deadweight machine. The forces realized in the PTB 1 MN deadweight machine were compared to those realized in the NIST 112 klbf, 300 klbf, and 1 Mlbf deadweight machines. The loads selected for machine intercomparison are listed in Tables 2, 3, and 4. Load selection was dictated by the constraints of the force standard machines intercompared, and the following criteria:

  1. Limit the measurement range so that no measurements are made below 40% of the force transducer capacity. However, because of the limitation of the machines in some instances data were taken with loads as low as 20%.

  2. Use only load sequences that can be applied monotonically;

  3. For each force standard machine intercompared, select the same loads.

Table 2.

Load chosen to intercompare the PTB 1 MN and the NIST 112 klbf standard machines

Force transducer capacity, kN Selected loads in PTB machine, kN Selected loads in NIST machine
kN (klbf)
100   50   48.9304 (11)
100   97.8609 (22)
200 100 102.3091 (23)
150 151.2395 (34)
200 200.1700 (45)
500 200 200.1700 (45)
300 298.0309 (67)
400 400.3400 (90)
500 498.2009 (112)

Table 3.

Loads chosen to intercompare the PTB 1 MN and the NIST 300 klbf standard machines

Force transducer capacity, kN Selected loads in PTB machine, kN Selected loads in NIST machine
kN (klbf)
200   90   88.9644 (20)
130 133.4467 (30)
180 177.9289 (40)
500 220 222.4111 (50)
310 311.3755 (70)
400 400.3400 (90)
1000 410 400.3400 (90)
540 533.7866 (120)
670 667.2333 (150)
800 800.6800 (180)
930 934.1266 (210)

Table 4.

Load chosen to intercompare the PTB 1 MN, 15 MN and the NIST 1 Mlbf standard machines

Force transducer capacity, kN Selected loads in PTB machine, kN Selected loads in NIST machine
kN (klbf)
  1000   220   222.4111 (50)
  440   444.8222 (100)
  660   667.2333 (150)
  880   889.6444 (200)
  4700   900   889.6444 (200)
1800 1779.2888 (400)
and 2700 2668.9332 (600)
3600 3558.5776 (800)
  5000 4500 4448.2220 (1000)

The following relationship was used to convert pound force to newtons:

1lbf=4.448222N.

5. Measurements Results

An effort was made to select for inclusion in the intercomparison similar loads at both PTB and NIST. However, because of machine limitations, the loads actually used, while similar, were significantly different as can be seen in Tables 2 through 4. For this reason, the NIST readings were normalized to correspond to the PTB applied force steps in accordance with the following equation:

NISTnormalizedindicatorreading=PTBappliedforcekNNISTappliedforcekN×NISTindicatorreading,

where the NIST indicator reading is the net indicator reading obtained by subtracting the zero indicator reading from the reading at load.

5.1 Comparison of Forces in the Range of 50 to 500 kN

Three force transducers having nominal capacities of 100, 200, and 500 kN were used to intercom-pare, over a range of 50 to 500 kN, the forces realized in the NIST 112 klbf deadweight machine and the forces realized at PTB. The force steps selected for this intercomparison were 50, 100, 150, 200, 300, 400, and 500 kN.

The measurement variability in each series of measurements, at each force step and at each force transducer position, expressed as the relative data spread between runs, is given in Tables 5, 6, and 7 for the 100, 200, and 500 kN force transducers, respectively. The relative spread between runs was calculated by taking the difference between the first and second readings and dividing the result by the initial reading. Tables 5 through 7 show that, for each force step, the spread in the data, for all measurements performed at both NIST and PTB, is below 50 ppm.

Table 5.

Relative data spread between runs for the 100 kN force transducer as a function of rotational position: 50 to 500 kN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial   50 kN   0   6 24 6
100 kN   2   0   1 4
NIST    11 klbf   4   9   9 8
   22 klbf   4 17 11 6
PTB final   50 kN 14 29 31 5
100 kN   4   6   2 9

Table 6.

Relative data spread between runs for the 200 kN force transducer as a function of rotational position: 50 to 500 kN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial 100 kN   9   3   7   6
150 kN   9   3   2   6
200 kN 15   3 23   5
NIST    23 klbf 21 14   4 47
   34 klbf   8   4   1   5
   45 klbf   9   2   2 14
PTB final 100 kN 12   2   5   5
150 kN 12   2 18   3
200 kN 44 11   1   6

Table 7.

Relative data spread between runs for the 500 kN force transducer as a function of rotational position: 50 to 500 kN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial 200 kN   1   4   1 10
300 kN   8   2   3   7
400 kN 10   2   1   3
500 kN   9   1   0   1
NIST    45 klbf 13   1   9 21
   67 klbf   8 18   2   0
   90 klbf   5   7   4   4
 112 klbf   7   3   1   3
PTB final 200 kN   9   6   6   1
300 kN   5   3   0   4
400 kN   9   0   1   1
500 kN   2   2   0   1

The net mean force transducer outputs measured during the NIST measurements, the initial and final PTB measurements, and those obtained by averaging the initial and final PTB measurements are presented in Table 8. The values in Table 8 are in indicator units.

Table 8.

Force transducer outputs measured at NIST and PTB: 50 to 500 kN range

Transducer kN Force step kN PTB initial PTB final PTB mean NIST
100   50 1.069592 1.069593 1.069593 1.069559
100 2.137819 2.137841 2.137830 2.137778
200 100 0.999438 0.999449 0.999444 0.999440
150 1.499222 1.499257 1.499240 1.499246
200 1.999049 1.999081 1.999065 1.999087
500 200 0.799180 0.799230 0.799205 0.799214
300 1.198875 1.198944 1.198909 1.198908
400 1.598594 1.598667 1.598630 1.598628
500 1.998344 1.998420 1.998382 1.998382

The relative differences between the average indicator readings at NIST and the corresponding average indicator readings at PTB are listed in Table 9 as a function of force transducer and force step. The values shown in Table 9 were obtained by taking the mean reading at NIST, subtracting from it the corresponding mean reading at PTB, and then dividing the result by the corresponding mean value of initial and final PTB readings. The values shown are rounded to the nearest ppm.

Table 9.

Relative differences between the NIST and PTB mean readings for the force transducer tested in the 112 klbf deadweight machine

Transducer kN Force step kN PTB initial ppm PTB final ppm PTB mean ppm
100   50 −31 −32 −31
100 −19 −29 −24
200 100     2 −10   −4
150   16   −8     4
200   19     3   11
500 200   43 −19   12
300   28 −30   −1
400   21 −24   −1
500   19 −19     0

The average deviations in the mean PTB data relative to the mean NIST data for all force transducers and force steps examined (i.e., last column in Table 9) are given in Fig. 7. Figure 7 shows that the agreement between the data obtained at NIST and PTB is excellent. The deviations fall within a band ranging from −31 to 12 ppm.

Figure 7.

Figure 7

Deviations in the mean PTB data relative to the mean NIST data for the 50 to 500 kN range.

5.2 Comparison of Forces in the Range of 90 kN to 1MN

Three force transducers having nominal capacities of 200 kN, 500 kN, and 1 MN were used to intercompare, over a range of 90 kN to 1 MN, the forces realized in the NIST 300 klbf deadweight machine and the forces realized at PTB. The force steps selected for this intercomparison were 90, 130, 180, 220, 310, 400, 410, 540, 670, 800, and 930 kN. When the program was planned originally, it was not anticipated that the 90, 130, and 180 kN force steps would be included in this subset of intercomparison. Accordingly, the first series of measurements performed at PTB did not include measurements for these force steps.

The measurement variability in each series of measurements, at each force step and each force transducer position, expressed as the relative data spread between runs, is given in Tables 10, 11, and 12 for the 200 kN, 500 kN, and 1 MN force transducers, respectively. The relative spread between runs was calculated by taking the difference between the first and second readings and dividing the result by the initial reading. Tables 10 through 12 show that, for each force step, the spread of the data is below 35 ppm for all measurements performed at both PTB and NIST.

Table 10.

Relative data spread between runs for the 200 kN force transducer as a function of rotational position: 90 kN to 1 MN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB final   90 kN   3 11 6 23
130 kN 21   8 3   8
180 kN   1 13 3   2
NIST    20 klbf   0   1 3   3
   30 klbf   6   4 1   4
   40 klbf 11   4 2   1

Table 11.

Relative data spread between runs for the 500 kN force transducer as a function of rotational position: 90 kN to 1 MN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial 220 kN 5   5   1 16
310 kN 9   2 10 13
400 kN 1   1 13   9
NIST    50 klbf 0   4   4   4
   70 klbf 2   3   5   5
   90 klbf 5 10   2   4
PTB final 220 kN 2   1   3   6
310 kN 2   6   2   2
400 kN 6   1   3   1

Table 12.

Relative data spread between runs for the 1 MN force transducer as a function of rotational position: 90 kN to 1 MN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial 410 kN   2 17   0   0
540 kN   3 25 25 31
670 kN   3   9   2 10
800 kN   1 10   2   4
930 kN   3   2   2   6
NIST    90 klbf   5 27   5 20
 120 klbf   0   7   2   2
 150 klbf   4   0   3   2
 180 klbf   5   1   1   4
 210 klbf   2   4   4   5
PTB final 410 kN   2   0   1   6
540 kN 12   8   1   4
670 kN   2   8   7   4
800 kN 10   4   2   1
930 kN 13   8   4   2

The net mean force transducer outputs measured during the NIST measurements, the initial and final PTB measurements, and those obtained by averaging the initial and final PTB measurements are presented in Table 13. The values shown are in indicator units.

Table 13.

Force transducer outputs measured at NIST and PTB: 90 kN to 1 MN range

Transducer kN Force step kN PTB initial PTB final PTB mean NIST
200a   90 0.899500 0.899500
130 1.299331 1.299366
180 1.799171 1.799204
500b 220 0.879081 0.879165 0.879123 0.879184
310 1.238808 1.238898 1.238853 1.238927
400 1.598538 1.598644 1.598591 1.598666
1000 410 0.835596 0.835609 0.835602 0.835623
540 1.100592 1.100621 1.100607 1.100618
670 1.365630 1.365669 1.365649 1.365654
800 1.630647 1.630694 1.630670 1.630684
930 1.895624 1.895674 1.895649 1.895656
a

No initial PTB measurements were taken with this transducer.

b

Initial PTB measurements contained systematic errors and should not be used to intercompare the forces achieved in the NIST 300 klbf deadweight machine with those achieved at PTB.

The relative differences between the average indicator readings at NIST and the corresponding readings at PTB are listed in Table 14 as a function of force transducer and force step. The values shown in Table 14 were obtained as described in Sec. 5.1 except that, for the 200 kN force transducer, the values shown were obtained by taking the mean reading at NIST, subtracting from it the corresponding mean final PTB reading, and dividing the result by the mean final PTB reading. All the values in Table 14 were rounded to the nearest ppm.

Table 14.

Relative differences between NIST and PTB readings for the force transducers tested in the 300 klbf deadweight machine

Transducer kN Force step kN PTB initial ppm PTB final ppm PTB mean ppm
 200a   90     0
130   27
180   18
 500b 220 117   22 69
310   96   23 60
400   80   14 47
1000 410   32   17 25
540   24   −3 10
670   18 −11   4
800   23   −6   9
930   17   −9   4
a

No initial PTB measurements were taken with this transducer.

b

Initial PTB measurements contained systematic errors and should not be used to intercompare the forces achieved in the NIST 300 klbf deadweight machine with those achieved at PTB.

When evaluating the differences in the forces realized at PTB and NIST in the range between 220 and 400 kN, that is the forces determined using the 500 kN force transducer, only the relative differences between NIST and the final PTB data are meaningful because the initial set of measurements obtained at PTB were found to contain systematic errors. This is the reason why the decision was made to include the 200 kN force transducer in the set of force transducers measured in NIST 300 klbf deadweight machine.

The average deviations in the mean PTB data relative to the mean NIST data for all force steps and all force transducers included in the 90 kN through the 1 MN range are shown in Fig. 8. For both the 200 and 500 kN force transducers, the values shown are based on the final PTB measurements only. The data shown for the 1 MN transducer are based on both the initial and final PTB measurements. Figure 8 shows that the agreement between the data obtained at NIST and PTB is very good. The deviations fall within a band from 0 to 30 ppm.

Figure 8.

Figure 8

Deviations in the mean PTB data relative to the mean NIST data for the 90 to 1 MN range.

5.3 Comparison of Forces in the Range of 220 to 4500 kN

Three force transducers having nominal capacities of 1, 4.7, and 5 MN were used to intercompare, over a range of 220 to 4500 kN, the forces realized in the NIST 1 Mlbf machine and (a) the forces realized in the 1 MN PTB deadweight machine, and, (b) for forces above 1 MN, those achieved in the 15 MN PTB hydraulic multiplication standard machine. The force steps selected for this intercom-parison were 220, 440, 660, 880, 900, 1800, 2700, 3600, and 4500 kN.

The measurement variability in each series of measurements, at each force step and at each force transducer position, expressed as the relative data spread between runs, is given in Tables 15, 16, and 17 for the 1, 4.7, and 5 MN force transducers, respectively. The relative spread between runs was calculated in accordance with the procedure described in Sec. 5.1. Tables 15 through 17 show that for the 1 MN force transducer the largest spread, 57 ppm, was obtained at NIST at the 50 klbf force step. With the other two transducers, the largest spread was obtained at PTB with the maximum, 240 ppm, occurring with the 5 MN transducer at the 900 kN force step.

Table 15.

Relative data spread between runs for the 1 MN force transducer as a function of rotational position: 220 to 880 kN range

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial 220 kN   4   2 13   7
440 kN   3 11 10   8
660 kN 16   5   8   5
880 kN   3   3   2   1
NIST    50 klbf 31 51 53 57
 100 klbf 17 19 19 13
 150 klbf   4   1   9 25
 200 klbf   4   0 14   9
PTB final 220 kN   0   4   9 27
440 kN   2   4   4   7
660 kN   1 11   9   1
880 kN   5   6   7   0

Table 16.

Relative data spread between runs for the 4.7 MN force transducer as a function of rotational position

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial   900 kN 95   54   27 27
1800 kN 48   34   27 14
2700 kN 64 130 140 27
3600 kN 37   58   51 20
4500 kN 35   38   44 14
NIST    200 klbf 25   33   39 33
   400 klbf 17   29   25 18
   600 klbf 34     2     6   6
   800 klbf 28     6     2   2
 1000 klbf 32     8     0   5
PTB final   900 kN 27 110   68 95
1800 kN   0   14   48   0
2700 kN   0   59   23 32
3600 kN   7   37   10 24
4500 kN   3   16   22 38

Table 17.

Relative data spread between runs for the 5 MN force transducer as a function of rotational position

Institute Applied force Rotational position
0° ppm 90° ppm 180° ppm 270° ppm
PTB initial   900 kN   81   90   18   27
1800 kN     4   27   13   27
2700 kN     6   42   15   18
3600 kN   20     2   20   16
4500 kN   14   29     7   13
NIST    200 klbf   11   38   42   82
   400 klbf   33   25   25   25
   600 klbf     1   51     1   16
   800 klbf     0   37   28     1
 1000 klbf   12   29   29   14
PTB final   900 kN 110   54 170 240
1800 kN     0 130     9 130
2700 kN   99   27 110   99
3600 kN   36   72   83 130
4500 kN   23   23   63   22

The net mean force transducer outputs measured at NIST and at PTB, and those obtained by averaging the initial and final PTB values are shown in Table 18. The values are in indicator units.

Table 18.

Force transducer outputs measured at NIST and PTB: 220 to 4500 kN range

Transducer MN Force step kN PTB initial PTB final PTB mean NIST
1   220 0.448320 0.448304 0.448312 0.448302
  440 0.896749 0.896739 0.896744 0.896757
  660 1.345272 1.345269 1.345270 1.345285
  880 1.793741 1.793745 1.793743 1.793748
4.7   900 0.557929 0.557866 0.557898 0.557926
1800 1.115960 1.115896 1.115928 1.115971
2700 1.673944 1.674003 1.673974 1.673927
3600 2.231629 2.231759 2.231694 2.231717
4500 2.789248 2.789319 2.789284 2.789406
5   900 0.367344 0.367306 0.367325 0.367351
1800 0.734166 0.734051 0.734108 0.734212
2700 1.101155 1.100933 1.101044 1.101248
3600 1.468967 1.468721 1.468844 1.469005
4500 1.837706 1.837474 1.837590 1.837671

The relative differences between the average indicator readings at NIST and the corresponding average indicator readings at PTB are listed in Table 19 as function of force transducer and force step. The values shown were obtained in accordance with the procedure given in Sec. 5.1. All values shown in the table were rounded to the nearest ppm. Table 19 shows that overall there was good agreement between the data obtained at PTB and NIST. However, the response of the largest two transducers drifted somewhat during the course of the intercomparison. This drift was most significant with the 5 MN force transducer.

Table 19.

Relative differences between the NIST and PTB readings for the force transducers tested in the 1 Mlbf deadweight machine

Transducer MN Force step kN PTB initial ppm PTB final ppm PTB mean ppm
1   220 −40   −5 −23
  440     9   20   14
  660   10   12   11
  880     4     2     3
4.7   900   −5 110   51
1800   10   68   39
2700 −11 −46 −28
3600   40 −19   10
4500   57   31   44
5   900   17 120   69
1800   63 220 140
2700   84 290 190
3600   26 190 110
4500 −19 110   44

The average deviations in the mean PTB data relative to the mean NIST data for the 1 and 4.7 MN force transducers and force steps examined are shown in Fig. 9. Because of the drift observed in the 5 MN force transducer, the results obtained with this transducer were not included in the generation of Fig. 9. The average deviations fall within a band from −30 to 50 ppm.

Figure 9.

Figure 9

Deviations in the mean PTB data relative to the mean NIST data for the 220 kN to 4.5 MN range.

6. Conclusions

Over a range of 50 to 4500 kN, the forces realized in the NIST deadweight machines compare favorably with those realized at PTB. Comparison of forces up to 900 kN indicate that the forces realized in the NIST deadweight machines and the forces realized in the PTB 1 MN deadweight machines agree within ±40 ppm. Peters et al. [7] have reported that theoretically the uncertainty in the forces realized by deadweight force machines is on the order of ±20 ppm. Accordingly, for the 50 to 900 kN force range, the agreement between the forces realized at NIST and PTB is close to what is theoretically achievable.

Comparison of forces in the range of 1 to 4.5 MN indicate that the forces realized in the PTB hydraulic force multiplication system compare favorably with the forces realized in the NIST deadweight machines. In this range, the agreement between the PTB data and the NIST data is within ±100 ppm, a remarkable agreement considering that the uncertainty of hydraulic force standard machine is inherently greater than that associated with deadweight machines.

Biography

About the authors: Simone L. Yaniv is a physical scientist in charge of the Force Group in the NIST Manufacturing Engineering Laboratory. M. Peters is in charge of the PTB Force Laboratory in Braunschweig, Germany where A. Sawla is a senior researcher.

7. References

  • 1.Sawla A, Peters M, Yaniv SL, Mitchell RA. Inter-comparison of Force Standard Machines of the National Institute of Standards and Technology, USA, and the Physikalisch-Technische Bundcsanstalt, Braunschweig, Germany. 1991. (PTB-Bericht MA-20). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Mitchell RA. Force Calibration at the National Bureau of Standards. U.S. Department of Commerce; Gaithersburg MD: 1986. (NBS Technical Note 1227). [Google Scholar]
  • 3.Weiler W, Peters M, Gassmann H, Fricke H, Ackerschott W. Die 1-MN-Kraft-NormalmeBeinrichtung der PTB Braunsehweig. VDI-Z. 1978;120:1–6. [Google Scholar]
  • 4.Weiler W, Sawla A, Peters M. Design and Calibration Problems of the 15 MN Hydraulic Force Standard Machine. VDI-Bericht. 1974:212. [Google Scholar]
  • 5.Peters M, Weiler W. Zur Kalibrierung der 15-MN-Kraft-NormalmeBeinrichtung der Physikalisch-Technischen bundesanstalt. PTB-Mitt. 1973:83. [Google Scholar]
  • 6.Mitchell RA, Baker SM. Characterizing the Creep Response of Load Cells. VDI-Berichte. 1978:312. [Google Scholar]
  • 7.Peters M, Sawla A, Peschel D. Uncertainty in Force Measurement, Report of the CCM Working Group Force. 1990. (PTB-Bericht MA-17). [Google Scholar]
  • 8.Pontius PE, Mitchell RA. Inherent Problems in Force Measurement. Exper Mech. 1982;22:81–88. [Google Scholar]

Articles from Journal of Research of the National Institute of Standards and Technology are provided here courtesy of National Institute of Standards and Technology

RESOURCES