Skip to main content
HHS Author Manuscripts logoLink to HHS Author Manuscripts
. Author manuscript; available in PMC: 2016 Jul 1.
Published in final edited form as: J Radioanal Nucl Chem. 2016 Mar 21;301(1):285–291. doi: 10.1007/s10967-014-3103-4

Determination of 241Am in Urine Using Sector Field Inductively Coupled Plasma Mass Spectrometry (SF-ICP-MS)

Ge Xiao 1,*, David Saunders 1, Robert L Jones 1, Kathleen L Caldwell 1
PMCID: PMC4930159  NIHMSID: NIHMS788634  PMID: 27375308

Abstract

Quantification of 241Am in urine at low levels is important for assessment of individuals’ or populations’ accidental, environmental, or terrorism-related internal contamination, but no convenient, precise method has been established to rapidly determine these low levels. Here we report a new analytical method to measure 241Am as developed and validated at the Centers for Disease Control and Prevention (CDC) by means of the selective retention of Am from urine directly on DGA resin, followed by SF-ICP-MS detection. The method provides rapid results with a Limit of Detection (LOD) of 0.22 pg/L (0.028 Bq/L), which is lower than 1/3 of the C/P CDG for 241Am at 5 days post-exposure. The results obtained by this method closely agree with CDC values as measured by Liquid Scintillation Counting, and with National Institute of Standards Technology (NIST) Certified Reference Materials (CRM) target values.

Introduction

Americium is a man-made, radioactive, metallic element produced when plutonium atoms undergo successive neutron capture events in nuclear reactors, in nuclear weapons, and during nuclear weapons’ detonations. Americium has several different isotopes, all of which are radioactive. The most important and prevalent americium isotope is 241Am, with a half-life of 432.7 years. As it decays, 241Am releases alpha particles at 5.44 MeV (13%) and 5.49 MeV (84.5%), becoming 237Np, which (in 35.9% of the decays) immediately emits gamma radiation at 59.5 keV. The 241Am decay chain ends with 209Bi, a nonradioactive element. 241Am in the environment originated from atmospheric testing of nuclear weapons during the 1950s and 1960s, as well as reprocessing plants and nuclear accidents. Facilities that are involved with nuclear weapons, well logging sources and manufacture smoke detectors are minor sources of 241Am contamination [1].

The critical 241Am exposure pathways are inhalation and ingestion. 241Am poses significant health hazards, even in small concentrations, if it is taken into the body in a soluble form. Once in the body, 241Am concentrates in the skeleton, liver, and muscle. It can stay in the body for decades and continue to expose the surrounding tissues to both alpha and gamma radiation. Long-term internal exposure to 241Am may create an increased risk of developing cancer. Exposure to any significant amount of 241Am is unlikely under normal circumstances [1, 2].

Several techniques exist for the determination of 241Am concentration in environmental and human samples [3, 4]. Gamma spectrometry, using High Purity Germanium detectors, is the primary tool used to determine 241Am at levels of 0.1−1 Bq/kg or higher; but to obtain accurate results, it requires that the user correct for the attenuation of gamma rays in the samples [5, 6]. Alpha spectrometry is the most commonly applied technique for determination of low-level 241Am content. Its principal advantages are relatively low equipment costs, high sensitivity due to low background, and high selectivity for alpha particles against other types of radiation [3]. A limit of detection (LOD) of 10−20 mBq/kg has been reported for various sample matrices, depending on the counting time and count rate of the procedure blank [79]. However, tedious, time-consuming sample preparation procedures (e.g., precipitation, evaporation, elution, filtration, electroplating, etc.) and long measurement times limit throughput. Such preparation procedures are due to possible interference from other radionuclides with close alpha energies, and long counting times are necessary because of 241Am’s relatively low specific activity.

SF-ICP-MS offers substantial advantages over conventional radiometric techniques and has recently been used for analysis of many long-lived radionuclides in various sample matrices. It is one of the fastest methods for 241Am analysis [1013]. An LOD of 1 pg/L for 241Am has been reported on SF-ICP-MS [14]. This LOD is comparable to that of alpha spectrometry, assuming no interferences exist for SF-ICP-MS. However, the main analytical issue in SF-ICP-MS originates from isobaric and polyatomic interferences such as 241Pu+, 240PuH+, 209Bi32S+, 209BiO2+, 206Pb35Cl+, 204Pb37Cl+, 205Tl36Ar+, 207Pb34S+, and 201Hg40Ar+ [11]. The major isobaric interference with 241Am is 241Pu. Since 241Am is the decay product of 241Pu (half-life is 14.33 years), in some samples of reactor origin the concentration of 241Am is comparable to that of 241Pu. Therefore, the method includes a thorough chemical separation of 241Pu (which also removes most of the other interfering molecular ions) from the samples [11, 15].

Developing methods to determine exposure to 241Am is within CDC’s public health mission. Quantitative analysis of 241Am in urine is considered a useful, noninvasive way to assess levels of internal contamination. Our Emergency Response Analytical goal is to be able to detect threat-radionuclides in urine at levels well below (i.e., 1/3 of or lower) the levels for a general population or for special subgroups such as children or pregnant woman (C/P) at the National Council on Radiation Protection & Measurements (NCRP) Report No. 161 Clinical Decision Guide (CDG) based action level of 0.73 pg/L (0.093 Bq/L) for 241Am (urine output expected at 5 days post intake) [16]. CDC’s IRATB recently developed a urine Gross Alpha/Gross Beta method using Liquid Scintillation Counting (LSC) [17]. However, the LOD of this method for 241Am is 4.2 Bq/L, equivalent to 32.3 pg/L, which is much higher than the C/P CDG of 0.73 pg/L for 241Am.

In this study, we report a novel and rapid analytical method for determination of 241Am in urine samples. The Solid Phase Extraction (SPE) part of the method is based on preliminary studies carried out by Horwitz, et al. [18], Li, et al. [19] and Sadi, et al. [20] using a single DGA resin cartridge to separate Am from other actinides such as U and Pu. We further optimized the method to isolate 241Am from a 10-mL volume of urine using simple extraction steps and used SF-ICP-MS for detection instead of LSC. The study’s purpose was to develop a rapid, simple method to address and respond to public health or other accidental, environmental, or terrorism-related exposures to 241Am. This method is not designed to characterize the normal background level of 241Am in the non-occupationally exposed population, but it does have a detection limit below the suggested CDG action levels. Thus it can serve as a means of rapidly identifying both adults and children who have been exposed to 241Am and who might require medical intervention.

Experimental

Reagents and solutions

DGA Cartridges (normal, 1 mL) and a polycarbonate vacuum box (24 holes) were purchased from Eichrom Technologies (Darien, IL, USA). All nitric (HNO3) and hydrochloric (HCl) acid solutions were prepared from double-distilled (DD) acids (GFS Chemicals Inc. Columbus, OH). Deionized water was used for all solutions (≥ 18 MΩ·cm, from an Aqua Solutions Ultrapure Water System, Aqua Solutions, Inc., Jasper, GA). “Base urine” was collected through anonymous human donations (CDC protocol 3994) and acidified to 1% v/v HNO3. All radioactivity solution sources were traceable to the National Institute for Standards and Technology (NIST, Gaithersburg, MD, USA). Both low and high quality control (QC) solutions and other urine pools for LOD were prepared for determination by spiking base urine with dilutions of an 241Am isotope standard (Eckert & Ziegler Analytics, Inc., Atlanta, GA). A series of aqueous 241Am Certified Reference Materials (CRM) solutions were prepared by dilution of 241Am radioactive source solutions from NIST. 243Am (Eckert & Ziegler Analytics Inc., Atlanta, GA) was used as an internal standard (tracer). Sodium nitrite (Sigma-Aldrich, St. Louis, MO) was used to adjust the oxidation states. Serial dilutions of uranium, lead, thallium, mercury, bismuth single-element stock standards (SPEX Industries, Inc., Edison, NJ) and a 242Pu radioactivity solution (U.S. Department of Energy, New Brunswick Laboratory, Argonne, IL) were spiked into the urine samples to verify that high separation factors for U, Pb, Tl, Hg, Bi and Pu were obtained using this SPE procedure.

Sample preparation

The urine sample volume for a single analysis is 10 mL. Allow urine specimens to reach ambient temperature, shake or vortex them to mix for 5 seconds before pipetting. Spike 400 µL of 1 ng/L 243Am solution as an internal standard (tracer) to every 10 mL of urine patient sample or QC sample. Add 4.76 mL of concentrated HNO3 (68–70%, the final concentration in the sample is 5M) and then 0.13g of sodium nitrite to each sample as a valence adjuster to convert Pu to the tetravalent state. Shake or vortex to mix the samples for 5 seconds and let reaction occur at room temperature for at least 10 minutes. Load each sample on a DGA resin cartridge of 1 mL bed volume (cartridge preconditioned with 15 mL of 5 M HNO3 using a vacuum box). Rinse the cartridge again with 15 mL of 5M HNO3 followed by 15 mL × 3 of 0.5M HNO3 using a vacuum box. Strip 241Am from the column with 5 mL of 0.5M HCl. Transfer 1 mL of the purified samples into 4 mL polystyrene conical bottom sample cups for analysis (Figure 1). Prepare external, aqueous-based stock calibration standards by spiking 0.5M HCl with dilutions of 241Am isotope standard, and then add 40 µL of internal standard solution (1 ng/L 243Am) to every 1 mL of standards to reach the same tracer concentration as the patient and QC samples. Prepare both calibration standards and sample blanks as 0.5 M HCl solutions, which match the elute solutions for the column of this method.

Figure 1.

Figure 1

Sequential sample preparation procedure for 241Am determination

* Samples containing U concentrations greater than 10 µg/L, add steps 5 – 6.

Instrumentation

This method measures 241Am concentrations using an extended dynamic range, high-resolution ICP-MS model Element XR (Thermo Fisher Scientific, Bremen, Germany), which is a double-focusing, magnetic sector, inductively-coupled-plasma mass spectrometer with a single discrete dynode detector (Mascom, Bremen, Germany). It uses the ICP-MS, equipped with nickel sampler and skimmer cones and a CD-2 guard electrode, in triple mode. The sample introduction system consists of a computer-controlled ASX-112 (Cetac, Omaha, NE) autosampler and an Aridus II™ (Cetac, Omaha, NE) desolvation unit. As discussed in our lab’s previous report [21], the Aridus II™ setup increases the sensitivity of the SF-ICP-MS by more than 10 times, enabling the measurement of 241Am at the low level of < 1 pg/L. Samples self-aspirate from the autosampler into the desolvation unit through an Apex perfluoroalkoxy (PFA) 100 μL/minute nebulizer (ESI, Omaha, NE, or equivalent). The desolvation unit, equipped with an upgraded PFA spray chamber, operates at 110 °C. With the aid of argon sweep gas and nitrogen gas for sensitivity enhancement, the sample passes through a semi-permeable membrane coil in the unit that operates at 160°C. Optimize flow rates as needed, with argon sweep gas at ~ 3−7 L/min and nitrogen gas at ~ 3−7 mL/min. The desolvated sample exits the unit into a 1.8 mm I.D. sapphire injector and a standard quartz torch, and then into the mass spectrometer. All experimental parameters are optimized for 241Am concentrations determination by SF-ICP-MS with respect to maximum ion intensity of 238U and minimum uranium oxide formation rate using a 5 ng/L natural uranium tuning solution. Table 1 contains a summary of our optimized operating conditions.

Table 1.

Instrumental conditions and data acquisition settings for SF-ICP-MS measurements

RF Power (KW) 1.2 – 1.3
Cooling Gas flow (L/min) 16
Auxiliary Gas flow (L/min) 0.9
Sample Gas flow (L/min) 0.7 – 0.8
Lenses (V) Optimized as needed

Sample Take up time (min) 2.1
Wash (min) 3
Pump Speed During Wash (rpm) 1

LR Runs/Passes 3* 60
Detection Mode Triple
Measurement Units CPS
Scan Type ESCAN
Scan Optimization Speed
Number of Pre-Scans 5
Integration Type Average
Res. Switch Delay (s) 2

Resolution 300
Mass Window (%) 15
Setting Time (s) 0.001
Sample Time (s) 0.001
Samples Per Peak 200
Search Window (%) 20
Integration Window (%) 15
Measured Isotopes 241Am, 243Am

Results and discussion

Removal of potential spectral interferences

Potential interferences for analysis of 241Am include isobaric overlaps with anthropogenic 241Pu and polyatomic overlaps with 240PuH+, 209Bi32S+, 209BiO2+, 206Pb35Cl+, 204Pb37Cl+, 205Tl36Ar+, 207Pb34S+, and 201Hg40Ar+. To test for complete removal of 241Pu, a 50 pg/L solution of 242Pu isotope spike in base urine was prepared and tested. Experiments showed more than 99% of 242Pu is removed by the SPE portion of sample preparation. Using SPE sample preparation as described above, Pb, Tl, and Hg, spiked in base urine at concentrations of 3 µg/L, 0.5 µg/L, and 5 µg/L respectively, did not result in apparent (> 0.1 pg/L) 241Am concentrations. These spiked urine samples’ concentrations were above the National Health and Nutrition Examination Survey (NHANES) 95th percentile of urine Pb, Tl, and Hg concentrations [22]. Although no NHANES survey data was available for bismuth, analysis of what was otherwise determined [23, 24] to be a high urine concentration (5 µg/L) of Bi, produced no apparent 241Am concentration.

Performance of a natural U-spike experiment determined that due to peak tailing, small interferences remained at m/z = 241 when the separated sample solutions contain high levels of U (> 0.5 µg/L). Analysis of urine samples with U =1.0 µg/L with this SPE method as part of the sample preparation procedure removed more than 99% of the U that might cause tailing into the m/z=241 region. Samples having U concentrations higher than 10.0 µg/L (the NHANES 95th percentile of U concentration in urine of normal U. S. residents is 0.031 µg/L) [22] should be treated by the modified sample preparation procedure as shown in Figure 1, which will be described in more detail below.

Limit of detection

The LOD for 241Am in urine specimens is based on 60 analytical runs of 4 different low-concentration samples close to the LOD (a first approximation of LOD is the measured blank concentration plus 3 times the Standard Deviation (SD) of the measured blank concentration) and was calculated according to the formula:

ConcLOD= [meanb + 1.645(Sb + int)]/[1-1.645(slope)], where mean b = blank average, Sb = standard deviation of blank average, int = intercept of the equation of SD versus concentration for LOD samples analyzed at least 60 times, Slope = slope of the equation of SD versus concentration for LOD samples analyzed at least 60 times.

The LOD of this method is 0.22 pg/L (Figure 2). This LOD is < 1/3 of the C/P CDG (~0.734 pg/L), and is therefore acceptable for an emergency radiobioassay method for determining the concentration of 241Am in urine collected at 5 days post-exposure.

Figure 2.

Figure 2

Plot for 241Am LOD determination (60 runs per point).

Linearity

A linearity study determined the linear reportable range for this method. The method exhibits good linear signal response between concentrations of 0.3 pg/L and 1000 pg/L of 241Am with a Coefficient of Determination of 1.000. The normal calibration range is from 0.3 pg/L to 30 pg/L, and the extended calibration range is from 30 pg/L to 1000 pg/L. If a urine 241Am value is above the highest calibrator, the urine sample is diluted with 5% HNO3 to bring the concentration within the validated calibration range.

Internal methods comparison study

A comparison of urine sample analysis results was performed between this method and our CLIA validated LSC method. The two samples LU-077203 and HU-077201 were prepared as QC material and, using LSC, analyzed for 241Am at relatively high concentrations. They then were diluted 1:1000 to get within the desired 241Am concentration range for the present method, purified and analyzed using SF-ICP-MS. The difference between the described methods is 2.1% to 3.0% (Table 2),

Table 2.

Observed 241Am concentrations (pg/L) and among-run precision for reference materials and internal quality control materials

Sample N 241Am
Average SD Target Value Bias (%)
LU-077203a 12 8,220 310 8,050d 2.1
HU-077201a 12 20,700 600 20,100d 3.0
Pool1b 59 0.335 0.058 0.3e 12
Low QCc 120 0.708 0.068 0.7e 1.2
Pool2b 60 1.44 0.093 1.4e 3.0
High QCc 120 10.1 0.56 10.0e 0.9
Extended High QCc 40 783 33.6 800e −2.1
Dilution 1f 1 100 100 0.0
Dilution 2f 1 203 200 1.7
Dilution 3f 1 300 300 −0.1
Dilution 4f 1 399 400 −0.3
Dilution 5f 1 605 600 0.8
Dilution 6f 1 1007 1000 0.7
a

1:1000 dilution of urine QC materials used for the LSC Gross Alpha/Beta method at CDC.

b

Urine materials made at CDC by spiking certified reference material in pooled urine collected anonymously.

c

Internal quality control materials made at CDC by spiking certified reference material in pooled urine collected anonymously.

d

Characterized results of co-worker by using the LSC Gross Alpha/Beta method at CDC[17].

e

Target values of spiked urine pools using certified reference material.

f

Aqueous dilutions of CRM from NIST.

Precision and accuracy

Analysis of serial aqueous dilutions of a Certified Reference Material (CRM) from NIST was also used to verify method accuracy. The observed 241Am concentrations were in close agreement with the target values, with an analytical bias from −0.3% to 1.7% (Table 2). Table 2 also shows the typical precision observed at different concentrations of daily quality control materials analyzed at the beginning, in the middle, and at the end of each analytical run. Accuracy and precision of the reported results was assured based on adherence to the quality control/quality assurance program of the Division of Laboratory Sciences, NCEH, CDC [25].

Analysis of samples from the NIST Radiochemistry Intercomparison Program (NRIP)

NRIP is a performance evaluation program which provides high quality, traceable radionuclide materials to support low-level radioanalytical laboratories conducting environmental and radiobioassay radioactivity measurements. 241Am is among the radionuclides used for testing. However, we found that the extraordinarily high concentrations of uranium present in these samples (intended for evaluation of environmental levels of uranium by alpha spectrometry) significantly affects the accuracy of trace level 241Am determination by SF-ICP-MS. Further, these uranium concentrations would possibly produce significant, troublesome instrument contamination. To address this problem we developed and recommend a modified sample preparation procedure that is further optimized for samples with extremely high U content (usually higher than 10 µg/L).

In this procedure, after rinsing the cartridge with 15 mL of 5M HNO3 followed by 15 mL × 3 of 0.5M HNO3, replace both the cartridge reservoirs and tips to eliminate possible U deposits and rinse the cartridges with more 0.5M HNO3 (15 mL × (3 - 6) of 0.5M HNO3, see Figure 1). Table 3 and Table 4 show the results observed for 241Am analysis of the NRIP samples. These samples were from two radiobioassay preparedness exercises during 2012 with different turnaround times (TATs): one 60 days, and one 8 hours. These synthetic urine samples typically have U concentrations ranging from 140 µg/L to 450 µg/L. After the more aggressive rinsing procedure, U concentrations in the elution solutions were under 0.20 µg/L, and did not result in apparent 241Am signal contribution for these samples. All but one result had slight negative bias (average −2.1 +/− 2.4% at a 95% Confidence Level) compared with the NIST target values. Most of the observed results for the NRIP samples show a small negative bias compared to the NIST target values, indicating a slight negative systematic uncertainty. One result had a positive bias of 12.7%. We noted that analyses of this sample for other radionuclides yielded a similar positive bias, indicating an external sample preparation error, as opposed to method bias.

Table 3.

Comparison of CDC 241Am results with NIST target values for NRIP12 60 days samples*

Sample ID Massic Activity
(NIST Target Value)
Massic Activity
(CDC Observed Results)
Relative Expanded
Uncertainty (k=2)
Bias
(Bq/g spike) (Bq/g spike) (%) (%)
207 1.784 1.74 11.5 −2.52
212 1.784 1.74 11.2 −2.41
220 1.784 1.76 12.8 −1.57
224 1.784 1.74 12.5 −2.52
227 1.784 1.76 12.1 −1.63
214 1.784 1.74 12.3 −2.47
216 1.784 1.71 12.1 −4.09
228 1.784 1.75 12.2 −2.02
231 1.784 1.77 11.5 −0.95
232 1.784 1.77 11.2 −0.73
208 1.784 1.74 13.1 −2.75
211 1.784 1.74 13.0 −2.58
219 1.784 1.68 13.8 −5.61
223 1.784 1.76 11.7 −1.57
226 1.784 1.77 11.8 −1.01
*

All samples were diluted 1:2 before DGA (Eichrom’s extraction chromatographic materials in which the extractant system is N,N,N’,N’-tetra-n-octyldiglycolamide resin) separation

Table 4.

Comparison of CDC 241Am results with NIST target values for NRIP12 8 hours samples*

Sample ID Massic Activity
(NIST Target Values)
Massic Activity
(CDC Observed Results)
Relative Expanded
Uncertainty (k=2)
Bias
(Bq/sample) (Bq/sample) (%) (%)
215 0.146 0.14 11.0 −1.93
218 0.292 0.29 10.9 −0.62
222 0.149 0.15 11.1 −1.47
230 0.297 0.29 11.1 −1.99
234** 0.372 0.42 11.2 12.7
*

Samples 218 and 230 were diluted 1:2 and sample 234 was diluted 1:4 before DGA (Eichrom’s extraction chromatographic materials in which the extractant system is N,N,N’,N’-tetra-n-octyldiglycolamide resin) separation

**

Analyses for other radionuclides also produced unusually high results for this sample.

Sample turnaround time (TAT)

While maintaining high quality results, sample TAT is one of the important considerations in a radiological emergency. For this method, ~ 2.5 hours are required to pretreat the urine samples for a batch of 20 patient urine specimens plus QC samples. An additional 3.5 hours are required for final analysis of 20 patient samples by SF-ICP-MS, including calibrators, blanks, and QC samples. Samples may be pretreated concurrently with final SF-ICP-MS analysis, resulting in a daily throughput of approximately 120 samples per day (24 hours) per instrument.

Conclusions

We introduced a method for rapidly determining ultra-low levels of 241Am in urine samples using a Solid Phase Extraction purification procedure and a high-sensitivity sample introduction system (Aridus II™), coupled with SF-ICP-MS. This method provides for analysis of 241Am at very low levels, with a LOD of 0.22 pg/L (well below the C/P CDG level) and allows rapid throughput of samples. It attained good agreement, with a bias of 2.1%−3.0%, for urine samples in an internal comparison with a CDC LSC method. It also produced recoveries from 99.7% to 101.7% in analysis of aqueous dilutions of 241Am SRM from NIST.

This method’s efficient urine sample separation scheme effectively eliminates most molecular ion interferences. However, if urine samples contain more than 10 µg/L of U, more aggressive rinsing procedures are required to eliminate U from the elution solutions. The results obtained by this method for NIST/NRIP reference materials with high-U levels are in close agreement with the NIST target values, with biases ranging from −0.62% to −5.61%.

A major advantage of this method over alpha spectrometry and other methods is that only a small, 10 mL volume of each urine sample is needed to perform the analysis, making successful analysis more likely, especially for young children and infants.

This procedure is appropriate for rapid identification and quantification of 241Am in urine for emergency response involving accidental or terrorism-related elevated exposures, or for evaluating chronic environmental or other non-occupational exposures.

Acknowledgments

The authors thank Baki B. Sadi at the Radiation Protection Bureau, Health Canada and Ted Zateslo at the Thermo Fisher Scientific technical support group (U.S.) for their help and technical assistance.

The opinions expressed in this article are the author's own and do not reflect the view of the Centers for Disease Control and Prevention, the Department of Health and Human Services, or the United States government.

Footnotes

The Authors declare that they have no competing financial interest.

References

  • 1. [Last Accessed on 9/10/2013]; http://www.epa.gov/radiation/radionuclides/americium.html.
  • 2. [Last Accessed on 9/10/2013]; http://www.atsdr.cdc.gov/phs/phs.asp?id=809&tid=158.
  • 3.Vajda N, Kim C-K. Journal of Radioanalytical and Nuclear Chemistry. 2010;284:341–366. [Google Scholar]
  • 4.Hou X, Roos P. Analytica Chimica Acta. 2008;608:105–139. doi: 10.1016/j.aca.2007.12.012. [DOI] [PubMed] [Google Scholar]
  • 5.Appleby PG, Richardson N, Nolan PJ. Nuclear Instruments & Methods in Physics Research Section B-Beam Interactions with Materials and Atoms. 1992;71:228–233. [Google Scholar]
  • 6.Canberra Genie (2000) software package – self-absorption correction. [Google Scholar]
  • 7.Kumar R, Yadav JR, Rao DD. Journal of Radioanalytical and Nuclear Chemistry. 2011;289:451–454. [Google Scholar]
  • 8.Dai X, Kramer-Tremblay S. Journal of Radioanalytical and Nuclear Chemistry. 2011;289:461–466. [Google Scholar]
  • 9.Warwick PE, Croudace IW, Oh JS. Analytical chemistry. 2001;73:3410–3416. doi: 10.1021/ac001510e. [DOI] [PubMed] [Google Scholar]
  • 10.Varga Z, Suranyi G, Vajda N, Stefanka Z. Microchemical Journal. 2007;85:39–45. [Google Scholar]
  • 11.Varga Z. Analytica Chimica Acta. 2007;587:165–169. doi: 10.1016/j.aca.2007.01.051. [DOI] [PubMed] [Google Scholar]
  • 12.Agarande M, Benzoubir S, Bouisset P, Calmet D. Applied Radiation and Isotopes. 2001;55:161–165. doi: 10.1016/s0969-8043(01)00045-8. [DOI] [PubMed] [Google Scholar]
  • 13.Horwitz EP, Chiarizia R, Dietz ML. Reactive & Functional Polymers. 1997;33:25–36. [Google Scholar]
  • 14.La Rosa JJ, Burnett W, Lee SH, Levy I, Gastaud J, Povinec PP. Journal of Radioanalytical and Nuclear Chemistry. 2001;248:765–770. [Google Scholar]
  • 15.Pourmand A, Dauphas N. Talanta. 2010;81:741–753. doi: 10.1016/j.talanta.2010.01.008. [DOI] [PubMed] [Google Scholar]
  • 16.Management of Persons Contaminated With Radionuclides: Handbook. 2008. Report No. 161 I, p. 158, ISBN-13: 978-0-929600-99-4, Executive Director: DA Schauer. [DOI] [PubMed] [Google Scholar]
  • 17.Piraner O. CDC CLIA Urine Gross Alpha-Beta. 2009. Feburary. 3011.1. [Google Scholar]
  • 18.Horwitz EP, McAlister DR, Bond AH, Barrans RE. Solvent Extraction and Ion Exchange. 2005;23:319–344. [Google Scholar]
  • 19.Li C, Sadi B, Benkhedda K, St-Amant N, Moodie G, Ko R, DiNardo A, Kramer G. Radiation Protection Dosimetry. 2010;141:228–232. doi: 10.1093/rpd/ncq175. [DOI] [PubMed] [Google Scholar]
  • 20.Sadi BB, Li C, Masoud A, Ko R, Kramer GH. Radiation Protection Dosimetry. 2010;141:134–139. doi: 10.1093/rpd/ncq157. [DOI] [PubMed] [Google Scholar]
  • 21.Xiao G, Jones RL, Saunders D, Caldwell KL. Radiation Protection Dosimetry. 2014 doi: 10.1093/rpd/ncu023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. [Last accessed on 02/25/2014];The Fourth National Report on Human Exposure to Environmental Chemicals. 2009 http://www.cdc.gov/exposurereport/pdf/FourthReport.pdf.
  • 23.Serfontein WJ, Mekel R, Bank S, Barbezat G, Novis B. Res Commun Chem Pathol Pharmacol. 1979;26(2):383–389. [PubMed] [Google Scholar]
  • 24.Carson BL, Ellis HV, McCann JL. Toxicology and Biological Monitoring of Metals in Humans. Lewis Publishers, Inc.; 1986. pp. 44–47. [Google Scholar]
  • 25.Caudill SP, Schleicher RL, Pirkle JL. Stat Med. 2008;27(20):4094–4106. doi: 10.1002/sim.3222. [DOI] [PubMed] [Google Scholar]

RESOURCES