. 2016 Jun 27;8(2):6445. doi: 10.4081/or.2016.6445

Biomechanical Properties of a Novel Biodegradable Magnesium-Based Interference Screw

Marco Ezechieli 1,, Hanna Meyer 1, Arne Lucas 2, Patrick Helmecke 3, Christoph Becher 4, Tilman Calliess 1, Henning Windhagen 1, Max Ettinger 1
PMCID: PMC4933821  PMID: 27433303

Abstract

Magnesium-based interference screws may be an alternative in anterior/posterior cruciate ligament reconstruction. The well-known osteoconductive effects of biodegradable magnesium alloys may be useful. It was the purpose of this study to evaluate the biomechanical properties of a magnesium based interference screw and compare it to a standard implant. A MgYREZr-alloy interference screw and a standard implant (Milagro®; De Puy Mitek, Raynham, MA, USA) were used for graft fixation. Specimens were placed into a tensile loading fixation of a servohydraulic testing machine. Biomechanical analysis included pretensioning of the constructs at 20 N for 1 min following cyclic pretensioning of 20 cycles between 20 and 60 N. Biomechanical elongation was evaluated with cyclic loading of 1000 cycles between 50 and 200 N at 0.5 Hz. Maximum load to failure was 511.3±66.5 N for the Milagro® screw and 529.0±63.3 N for magnesium-based screw (ns, P=0.57). Elongations after preload, during cyclical loading and during failure load were not different between the groups (ns, P>0.05). Stiffness was 121.1±13.8 N/mm for the magnesium-based screw and 144.1±18.4 for the Milagro® screw (ns, P=0.32). MgYREZr alloy interference screws show comparable results in biomechanical testing to standard implants and may be an alternative for anterior cruciate reconstruction in the future.

Key words: Degradable magnesium screw, anterior cruciate ligament reconstruction, biomechanics

Introduction

The use of interference screws for anterior cruciate (ACL) or posterior cruciate (PCL) reconstruction is widely spread.1 Degradable screws (DS) manufactured from various polymers such as polylactic acid (PLLA), polylactide carbonate (PLC), and polylactic-co-glycolic acid (PLGA) are the most commonly used DS. Compared to metallic screws (MS) DS show a higher incidence of complications such as implant breakage,2,3 tunnel widening due to foreign body reaction,4 adverse immune reactions5 and prolonged joint effusion.1 Despite the potential advantages, the use of MS is compromised by artifacts in postoperative magnetic resonance imaging and the necessity of implant removal in case of revision surgery.6

Magnesium (Mg) alloys are of high interest for future orthopedic implants. The related modules of magnesium are comparable to natural bone and lead to less stress-shielding when Mg is basis of an orthopedic implant.7,8 Besides that, Mg has an osteoinductive effect and stimulates the creation of extracellular matrix.9-11 Promising animal studies have shown no negative effect, such as cartilage toxicity or synovitis, using Mg or Mg-alloys.12-14 The MgYREZr alloy is similar to the WE43-alloy, which has shown a good biological response in both, animal studies and clinical applications.12,15 A MgYREZr interference screw has been analyzed before.16 In this study different threads in an all-synthetic in vitro setup have been tested. The aim of this study was to evaluate the biomechanical properties of the first MgYREZr interference screw in ACL reconstruction and compare it to a standard implant. We hypothesized that both have similar properties and, therefore, the MgYREZr interference screw may be an alternative for ACL reconstruction in the future.

Materials and Methods

The interference screws used in this study are made of MAGNEZIX™ (Syntellix AG, Hannover, Germany), a powder metallurgical processed magnesium alloy. This aluminum-free material is based on the MgYREZr-alloy system and contains more than 90% magnesium. With an average grain size <5 µm, this high-performance alloy exhibits an offset elastic limit Rp0.2>250 MPa, a tensile strength (TS) >275 MPa, and an elongation at break of more than 10%. Compared with steel or titanium implants that are machined under oil or emulsion, these screws are machined from extruded MAGNEZIX™ bar stocks by dry cutting processes.

A total of 18 flexor digitorum profundus tendons and femora from skeletally mature 10-month-old female or castrated male pigs were used in this study. Hind limbs were obtained from an industrial slaughterhouse and stored at -20° C. Before use, specimens were thawed at room temperature for 24 hours. Flexor digitorum profundus tendons were harvested from the hinder limbs. The graft diameter was measured with an ACL graft sizer tube. Only tendons passing through the 8 mm sizer were included.

Porcine tendons were randomly assigned to one of the following groups: i) Interference screw fixation with a standard Implant (Milagro® 8×23 mm) or ii) custom made magnesium-based (MgYREZr-alloy 8×23 mm interference screw) (Figure 1). The potted femurs were rigidly fixed in a base platform at 50°, with the bone tunnel force direction angle at 0°. There was a distance of 30 mm between the graft and the clamp. Photo optical markers were placed on the tendons in order to measure the relative construct elongation within the tendons and from the fixation. During the entire procedure, tendons were kept moist with saline spray. For biomechanical evaluation, specimens were placed into a tensile loading fixation of a servohydraulic testing machine (Mini Bionix 858; MTS Systems Co, Minneapolis, MN, USA). The tendons were rigidly fixed in a clamp holding 2 cm of the distal end of the tendon.

Figure 1.

Figure 1.

Interference screws: 1) Milagro® 8×23 mm, DePuy Mitek, Raynham, MA, USA and 2) custom made magnesium-based (MgYREZr-alloy) 8×23 mm interference screw.

Biomechanical analysis included pretensioning of the constructs at 20 N for 1 min following cyclic pretensioning of 20 cycles between 20 and 60 N. Biomechanical elongation was evaluated with cyclic loading of 1000 cycles between 50 and 200 N at 0.5 Hz. The increase in construct length was recorded at a frequency of 20 Hz and a measurement accuracy of 0.1 mm. Ultimate failure load test was performed with a ramp speed of 20 mm/min. The maximum failure load and failure mode of the constructs were analyzed. A video analysis was performed in order to exclude a fixation bias. Elongation was measured evaluated after preloading, after cycles 200, 400, 600, 800 and 1000 as well as after failure. Stiffness (N/mm) was analyzed during failure testing.

Statistical analysis

Sample size determination was performed using an online tool (www.clinical-trials.de). The sample size calculation advised 9 specimens per group for a power of 0.8 at a significance level of alpha 0.05. All statistical analyses were performed using Graph Pad Prism (GraphPad Software Inc., La Jolla, CA, USA). Normal distribution was analyzed by the D’Agostino & Pearson omnibus normality test. All values are presented in the form of mean ± standard deviation (SD). Analysis of variance was used for parametric data and Kruskal-Wallis test for nonparametric data. A P value less than 0.05 was considered to be statistically significant.

Results

Maximum load to failure was 511.3±66.5 N for the Milagro® screw and 529.0±63.3 N for magnesium-based screw (ns, P=0.57; Figure 2). Elongations after preload, during cyclical loading and during failure load were not different between the groups (ns, P>0.05; Figure 3). Total elongation after preload averaged from 5.1±0.5 mm in the Milagro group and 5.1±0.4 in the Mg group (P=0.92). Stiffness was 121.1±13.8 N/mm for the magnesium-based screw and 144.1±18.4 for the Milagro® screw (ns, P=0.32). In the Mg group fixations failed by a tendon rupture in 2/10 cases and by a tendon pull out of the bony tunnel in 8/10 cases. In the Milagro® group fixations failed by a tendon rupture in 3/10 cases and by a tendon pull out of the bony tunnel in 7/10 cases. In both group the video analysis indicated no significant elongation between tendon and clamp.

Figure 2.

Figure 2.

Maximum load to failure was 511.3±66.5 N for the Milagro® screw and 529.0±63.3 N for magnesium-based screw (ns, P=0.57).

Figure 3.

Figure 3.

Elongation after preload, during cyclical loading and during failure load were not different between the groups (ns, P>0.05).

Discussion

The most important finding of this study is that the MgYREZr alloy interference screw showed similar biomechanical properties compared to a standard implant. Consequently, our null hypothesis is to accept.

Several study limitations have to be addressed: i) we only applied one force-angle of 0°. This gives no conclusion about the biomechanical properties in flexion-extension motion. ii) This is a time zero scenario. No conclusion concerning the characteristics of the degradation process as well as the changing biomechanical properties is possible. The use of degradable magnesium interference screws may be controversial. Kuhlmann and colleagues described how hydrogen gas is produced and gas cavities are seen during the degradation process of magnesium-based implants. The source of the gas cavities remains an issue of debate.17 The MgYREZr alloy showed no bone erosion resulting from gas formation after 1 year of follow up in a recently published animal study.18 iii) We used a whole porcine model for our investigation. Compared to human hamstring tendons, porcine flexor digitorum profundus tendons have a lower maximum load to failure, but comparable elongation and lower standard deviations (unpublished data from our study group). Taken this into consideration, porcine flexor digitorum profundus tendons can be used as suitable graft substitutes for human hamstring tendons in biomechanical in vitro studies. Due to their high availability and constant size they have advantages over human tissue. The uniformity of the porcine tendons may furthermore help to reduce specimen-related bias in comparative studies for ligamentous fixations techniques. To achieve adequate early postoperative rehabilitation after ACL surgery, primary stability of the graft fixation is mandatory. The maximum failure loads observed in our study are comparable to those of other soft tissue interference screw fixations.19,20 Our results demonstrate for both screws in this biomechanical setup that maximum load to failure rates are as high as needed for early rehabilitation.21,22 It is well known that cyclic displacement has an influence on the long-term graft function. Daniel and colleagues reported that a 3 mm side-to-side laxity after unilateral ACL reconstruction leads to a high rate of ACL failures.23,24 Morrison et al.21 described an estimated force of 350 N achieved in normal walking; however, the peak cyclic loading force in our study was 250 N. We believe that this is not a critical success factor for this study as it was during the initial postoperative period when lower graft forces are expected. The total elongation detected in this study was distinctly higher compared to a screw fixation recently published by Mayr and colleagues18 (mean: 1.87 mm) but comparable results to Zantop and colleagues25 (mean: 4.81 mm) and Weimann and colleagues26 (mean: 4.1 mm). This might be due to the different test setups (human hamstring graft, porcine bone, calf bone or porcine flexor tendons). The elongation after preload, cyclical loading and the total elongation of the constructs were similar in both of our groups. This reflects the equivalence of the Mg screw.

The overall construct stiffness is a crucial part of ACL reconstruction procedure. It is known that a high stiffness graft construct better restores the anterior and posterior limit to normal compared to a low stiffness graft.27 The stiffness at failure mode evaluated in this study was similar in both groups. Our results are comparable to other studies focusing on the biomechanics of interference screws18,25 and to extra cortical fixation techniques.28,29

Conclusions

The present in vitro biomechanical examination investigated the biomechanical properties of a novel magnesium alloy (MgYREZr) screw compared to a control, consisting of a commercial polymer screw. All biomechanical parameters were similar. This study shows the first promising results of a degradable material, which may be a clinical alternative in the future. Further in vivo studies, with focus on hydrogen gas is production and gas cavities during the degradation process of magnesium, are necessary to validate these biomechanical results.

Funding Statement

Funding: The authors acknowledge the financial support given by the German Research Foundation (DFG) within the collaborative research project SFB 599.

References

  • 1.Mascarenhas R, Saltzman BM, Sayegh ET, et al. Bioabsorbable versus metallic interference screws in anterior cruciate ligament reconstruction: a systematic review of overlapping meta-analyses. Arthroscopy 2015;31:561-8. [DOI] [PubMed] [Google Scholar]
  • 2.Lembeck B, Wulker N. Severe cartilage damage by broken poly-L-lactic acid (PLLA) interference screw after ACL reconstruction. Knee Surg Sports Traumatol Arthroscopy 2005;13:283-6. [DOI] [PubMed] [Google Scholar]
  • 3.Watson JN, McQueen P, Kim W, Hutchinson MR. Bioabsorbable interference screw failure in anterior cruciate ligament reconstruction: a case series and review of the literature. Knee 2015;22:256-61. [DOI] [PubMed] [Google Scholar]
  • 4.Bach FD, Carlier RY, Elis JB, et al. Anterior cruciate ligament reconstruction with bioabsorbable polyglycolic acid interference screws: MR imaging follow-up. Radiology 2002;225:541-50. [DOI] [PubMed] [Google Scholar]
  • 5.Mastrokalos DS, Paessler HH. Allergic reaction to biodegradable interference poly- L-lactic acid screws after anterior cruciate ligament reconstruction with bone- patellar tendon-bone graft. Arthroscopy 2008;24:732-3. [DOI] [PubMed] [Google Scholar]
  • 6.Noyes FR, Barber-Westin SD. Revision anterior cruciate surgery with use of bone-patellar tendon-bone autogenous grafts. J Bone Joint Surg Am 2001;83:1131-43. [DOI] [PubMed] [Google Scholar]
  • 7.Witte F, Calliess T, Windhagen H. [Biodegradable synthetic implant materials: clinical applications and immunological aspects]. Orthopade 2008;37:125-30. [Article in German]. [DOI] [PubMed] [Google Scholar]
  • 8.Staiger MP, Pietak AM, Huadmai J, Dias G. Magnesium and its alloys as orthopedic biomaterials: a review. Biomaterials 2006;27:1728-34. [DOI] [PubMed] [Google Scholar]
  • 9.Pietak A, Mahoney P, Dias GJ, Staiger MP. Bone-like matrix formation on magnesium and magnesium alloys. J Mater Sci Mater Med 2008;19:407-15. [DOI] [PubMed] [Google Scholar]
  • 10.Witte F, Kaese V, Haferkamp H, et al. In vivo corrosion of four magnesium alloys and the associated bone response. Biomaterials 2005;26:3557-63. [DOI] [PubMed] [Google Scholar]
  • 11.Castellani C, Lindtner RA, Hausbrandt P, et al. Bone-implant interface strength and osseointegration: biodegradable magnesium alloy versus standard titanium control. Acta Biomater 2011;7:432-40. [DOI] [PubMed] [Google Scholar]
  • 12.Waizy H, Diekmann J, Weizbauer A, et al. In vivo study of a biodegradable orthopedic screw (MgYREZr-alloy) in a rabbit model for up to 12 months. J Biomater Appl 2014;28:667-75. [DOI] [PubMed] [Google Scholar]
  • 13.Thormann U, Alt V, Heimann L, et al. The biocompatibility of degradable magnesium interference screws: an experimental study with sheep. Biomed Res Int 2015;2015:943603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Liao Y, Xu Q, Zhang J, et al. Cellular response of chondrocytes to magnesium alloys for orthopedic applications. Int J Mol Med 2015;36:73-82. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Windhagen H, Radtke K, Weizbauer A, et al. Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study. Biomed Eng Online 2013;12:62. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Ezechieli M, Ettinger M, Konig C, et al. Biomechanical characteristics of bioabsorbable magnesium-based (MgYREZr-alloy) interference screws with different threads. Knee Surg Sports Traumatol Arthrosc. 2014. [Epub ahead of print] [DOI] [PubMed] [Google Scholar]
  • 17.Kuhlmann J, Bartsch I, Willbold E, et al. Fast escape of hydrogen from gas cavities around corroding magnesium implants. Acta Biomater 2013;9:8714-21. [DOI] [PubMed] [Google Scholar]
  • 18.Mayr R, Heinrichs CH, Eichinger M, et al. Biomechanical comparison of 2 anterior cruciate ligament graft preparation techniques for tibial fixation: adjustable-length loop cortical button or interference screw. Am J Sports Med 2015;43:1380-5. [DOI] [PubMed] [Google Scholar]
  • 19.Adam F, Pape D, Schiel K, et al. Biomechanical properties of patellar and hamstring graft tibial fixation techniques in anterior cruciate ligament reconstruction: experimental study with roentgen stereometric analysis. Am J Sports Med 2006;32:71-8. [DOI] [PubMed] [Google Scholar]
  • 20.Ahmad CS, Gardner TR, Groh M, et al. Mechanical properties of soft tissue femoral fixation devices for anterior cruciate ligament reconstruction. Am J Sports Med 2004;32:635-40. [DOI] [PubMed] [Google Scholar]
  • 21.Morrison JB. The mechanics of the knee joint in relation to normal walking. J Biomech 1970;3:51-61. [DOI] [PubMed] [Google Scholar]
  • 22.Shelburne KB, Pandy MG, Anderson FC, Torry MR. Pattern of anterior cruciate ligament force in normal walking. J Biomech 2004;37:797-805. [DOI] [PubMed] [Google Scholar]
  • 23.Daniel DM, Stone ML, Dobson BE, et al. Fate of the ACL-injured patient. A prospective outcome study. Am J Sports Med 1994;22:632-44. [DOI] [PubMed] [Google Scholar]
  • 24.Daniel DM, Stone ML, Sachs R, Malcom L. Instrumented measurement of anterior knee laxity in patients with acute anterior cruciate ligament disruption. Am J Sports Med 1985;13:401-7. [DOI] [PubMed] [Google Scholar]
  • 25.Zantop T, Weimann A, Rummler M, et al. Initial fixation strength of two bioabsorbable pins for the fixation of hamstring grafts compared to interference screw fixation: single cycle and cyclic loading. Am J Sports Med 2004;32:641-9. [DOI] [PubMed] [Google Scholar]
  • 26.Weimann A, Rodieck M, Zantop T, et al. Primary stability of hamstring graft fixation with biodegradable suspension versus interference screws. Arthroscopy 2005;21:266-74. [DOI] [PubMed] [Google Scholar]
  • 27.Karchin A, Hull ML, Howell SM. Initial tension and anterior load- displacement behavior of high-stiffness anterior cruciate ligament graft constructs. J Bone Joint Surg Am 2004;86-A:1675-83. [DOI] [PubMed] [Google Scholar]
  • 28.Barrow AE, Pilia M, Guda T, et al. Femoral suspension devices for anterior cruciate ligament reconstruction: do adjustable loops lengthen? Am J Sports Med 2014;42:343-9. [DOI] [PubMed] [Google Scholar]
  • 29.Petre BM, Smith SD, Jansson KS, et al. Femoral cortical suspension devices for soft tissue anterior cruciate ligament reconstruction: a comparative biomechanical study. Am J Sports Med 2013;41:416-22. [DOI] [PubMed] [Google Scholar]

Articles from Orthopedic Reviews are provided here courtesy of Open Medical Publishing

RESOURCES