Abstract
Pharmacokinetic parameters derived from dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) have been increasingly used to evaluate the permeability of tumor vessel. Histogram metrics are a recognized promising method of quantitative MR imaging that has been recently introduced in analysis of DCE-MRI pharmacokinetic parameters in oncology due to tumor heterogeneity. In this study, 21 patients with renal cell carcinoma (RCC) underwent paired DCE-MRI studies on a 3.0 T MR system. Extended Tofts model and population-based arterial input function were used to calculate kinetic parameters of RCC tumors. Mean value and histogram metrics (Mode, Skewness and Kurtosis) of each pharmacokinetic parameter were generated automatically using ImageJ software. Intra- and inter-observer reproducibility and scan–rescan reproducibility were evaluated using intra-class correlation coefficients (ICCs) and coefficient of variation (CoV). Our results demonstrated that the histogram method (Mode, Skewness and Kurtosis) was not superior to the conventional Mean value method in reproducibility evaluation on DCE-MRI pharmacokinetic parameters (K trans & Ve) in renal cell carcinoma, especially for Skewness and Kurtosis which showed lower intra-, inter-observer and scan-rescan reproducibility than Mean value. Our findings suggest that additional studies are necessary before wide incorporation of histogram metrics in quantitative analysis of DCE-MRI pharmacokinetic parameters.
Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI), as a very common MRI technique, not only can subjectively judge the enhancement of a target area on a visual basis, semi-quantitatively characterize tumors using curvology1,2, but also can quantitatively evaluate parameters generated using pharmacokinetic models3,4 which reflected the dynamic distribution of Ga-related contrast agent in the different compartments of the tissue. The two-compartment model of DCE-MRI assumes the contrast agent exchanges between the plasma space and the extravascular-extracellular space (EES)5, and the forward and backward transfer rate could reflect the permeability of the microvasculature. It is used extensively in measuring tumor angiogenesis and blood brain barrier (BBB) disruption.
Pharmacokinetic DCE-MRI in oncology has been increasingly applied in quantitative scientific research and clinical practice. Zahra et al. recently summarized studies that have utilized DCE-MRI parameters to predict the efficacy of chemotherapy and concluded that DCE-MRI was a reasonably accurate and non-invasive method6.
Traditionally, many researchers utilize the mean value of the targeted region of interest (ROI) to perform analysis of tumors and made comparisons in the intra-observer, inter-observer, or test-retest analyses7,8,9,10. As a promising quantitative tool, the reliability and reproducibility of DCE-MRI suggests it will be widely used in future oncology analyses. Previously, we showed that the pharmacokinetic parameters of DCE-MRI in renal cell carcinoma (RCC) using Mean value of pharmacokinetic parameters demonstrated good reproducibility11.
However, beyond the tumor itself, much attention has been rightfully paid to tumor heterogeneity that exists in the tumor cell population due to the surrounding extracellular matrix, angiogenesis, and other tumor microenvironment features, all of which influence tumor characterization and therapeutic effect to a certain degree. Indeed, there is increasing interest in analyzing lesion heterogeneity by way of histogram analysis to characterize tumor subtypes12,13,14,15, tumor histological grades16,17,18,19, tumor aggressiveness20 and evaluate treatment effects21,22,23,24. This methodology has showed its utility in investigating the distributions of various tumor parameters such as permeability in dynamic contrast-enhanced MRI (DCE-MRI)17,25.
With the expected increase in use of heterogeneity analysis with DCE-MRI, it is therefore important to analyze its reproducibility capability before adopting its widespread use in performing analysis of tumor characterization or prediction of therapeutic effect. To the best of our knowledge, with the exception of a study by Heyes et al.26 that presented a histogram analysis approach combined with a semi-automatic lesion segmentation to show a decrease in inter-observer variability in the K trans parameter in DCE-MRI, no other studies have examined the reproducibility of histogram analysis. Herein, we evaluated the intra- and inter-observer, as well as scan–rescan reproducibility of histogram metrics in regard to DCE-MRI pharmacokinetic parameters in RCC.
Methods
Patients
Institutional Review Board of Chinese PLA General Hospital approved this prospective study. The methods used in this study were carried out in accordance with the Declaration of Helsinki. Written informed consent was obtained from each subject prior to study initiation. Patients with suspected renal cell carcinoma (RCC) during the imaging examinations were recruited from the urological clinic at our hospital from September 2012 to November 2012. Inclusion criteria were as follows: age ≥18 years old, glomerular filtration rate >60 mL/min, size of lesions >1.0 cm in diameter to avoid partial volume artifact concerns, and clear cell RCCs – as the most common pathologic subtype. Exclusion criteria included the following: common contraindication for MRI scans and the use of Ga-related contrast (such as metal implants, heart pacemaker, severe claustrophobia etc.), age <18 years old, glomerular filtration rate of <60 mL/min, size of lesions ≤1.0 cm in diameter, lesions with complete necrosis or cystic degeneration confirmed in MR examination, and patients with unacceptable DCE-MR imaging quality such as severe motion artifacts.
Sample size in this study was estimated using Power Analysis & Sample Size Software, PASS 11.0 (NCSS, LLC. Kaysville, Utah, USA). Due to usage of Intra-class Correlation Coefficient (ICC) as statistical tool and three observers in this study, we assumed the expected ICC of 0.9 (R1) and acceptable lowest ICC of 0.75 (R0), thus we set α = 0.05 and β = 0.20. Finally, through automatic calculation of PASS, the least acceptable number of subject (k) was 19.
MRI technique
MRI scans were performed on a 3.0 T platform (GE Discovery MR 750, GE Healthcare, Milwaukee, WI) with an 8-channel surface phased-array coil. Patients were scanned twice with the first scan within 48 h of the initial diagnosis and the second scan at 48–72 h after the first scan, where the same lying position and scanning location were utilized. Breathing training was conducted before each scan. Besides routine scanning sequence (i.e., axial and coronal T2-weighted imaging), DCE-MRI was performed, which consisted of a pre-contrast T1 mapping sequence and a dynamic sequence. T1 mapping included multi-flip angles (3°, 6°, 9°, 12°, and 15°) pre-contrast scan with three-dimensional (3D) spoiled-gradient recalled-echo sequences for liver acquisition with volume acceleration (LAVA) in breath-hold mode. Dynamic sequence was performed with the same parameters as T1 mapping but with flip angle 12°, which resulted in a tempo resolution of 6 s. During dynamic scan, two successive phases for 12 s in a breath-holding mode and an interval for 6 s in a free-breathing mode were performed alternatively. The entire dynamic process lasted for 4.4 minutes. Scanning parameters were as follows: repetition time (TR) 2.8 ms, echo time (TE) 1.3 ms, matrix 288 × 180, field of view (FOV) 38 × 38 cm, slice thickness 6 mm, number of excitations (NEX) 1, bandwidth 125 kHz, and parallel imaging acceleration factor 3. When the scan for the third phase was started, the contrast media (0.1 mmol/kg, Omniscan, GE Healthcare) was administered intravenously as a bolus injection at a rate of 2 mL/s using a power injector (Spectris; MedRad, Warrendale, PA), followed with 20 mL normal saline flush at the same rate.
Image post-processing and analysis
All images were transferred to an Omni-Kinetics workstation (GE Healthcare, LifeScience, China) for analysis. Non-rigid registration method suggested in literature27,28,29 was used to assess and correct medical image alignment within dynamic scans. The workstation used a framework (a free-form deformation algorithm) as previously described30,31,32 to help remove any error of misalignment between consecutive MRI scans, thus making our results more accurate than the non-processed images.
Calculation of Pharmacokinetic Parameters
Multiple flip angles method33,34 was used to perform T1 mapping to obtain both the T1 value of the tissue before and after contrast agent injection using Equation 1, where m0 is the equilibrium signal intensity, θ is the flip angle, TR is the repetition time, T1 is the tissue T1 value, S(θ) is the T1 signal intensity. Then the contrast agent concentration in the tissue was computed using Equation 234, where T1 is the T1 value after contrast injection, T10 is T1 value before contrast injection, and r (mM−1s−1) represents the longitudinal CA relaxation coefficient; thus, signal intensity of the tissue is converted to tissue CA concentration (Ct(t)). The widely used two-compartment extended-Tofts model35 (Equation 3) with population averaged arterial input function (AIF)33,34 (Equation 4) was used to calculate the kinetic parameters. Where in Equation 3, K trans represented the transfer constant from plasma to the extracellular extravascular space (EES); Ve represented the ratio of the EES volume to tissue volume; Vp represented the ratio of blood plasma volume to tissue volume;
Kep was the efflux rate constant from EES to plasma and equaled K trans/Ve; Ct(t) and Cp(t) represented the contrast agent concentrations in the tissue and blood plasma, respectively. In Equation 4, D = 1.0 mmol/kg, a1 = 2.4 kg/l, a2 = 0.62 kg/l, m1 = 3.0 and m2 = 0.016.
ROI selection
Using reference information from anatomic axial and coronal T2-weighted images and post-contrast T1 images, the slice with the maximum diameter of the tumor was selected in the ImageJ software (National Institutes of Health, Bethesda, MD). Three radiologists (Z.S., F.D., Y.S., all board-certified radiologists engaged in abdominal imaging for 8, 10, 9 years, respectively) outlined ROIs around the edges of the tumors on the DCE-MRI map (Fig. 1a). Parameter outlines covered the whole tumor as much as possible and excluded pulsatile artifacts from blood vessels and susceptibility artifacts from adjacent bowels. Then the same ROI was copied to parametric maps (Fig. 1b,c).
Commonly, values of K trans greater than 1.2 min−1 are considered pseudo-permeability in large blood vessels or errors in fitting36,37; therefore any pixels with K trans larger than 1.2 min−1 or with Ve beyond the range of 0–100% were excluded from parametric maps. Based on this situation, histogram function in ImageJ was utilized and threshold value of kinetic parameters were set respectively such as K trans (0, 1.2 min−1), and Ve (0, 1). Then the traditional Mean values of K trans, and Ve and heterogeneity analysis (i.e., Mode, Skewness, and Kurtosis) were automatically calculated. Kurtosis described how sharply peaked a histogram was compared with the histogram of a normal distribution. Accordingly, whereas a normal distribution had a Kurtosis of 0, a more peaked histogram had a positive Kurtosis value. Skewness described the degree of asymmetry of a histogram: a perfectly symmetric histogram had a Skewness of 0, a histogram with a long right tail had a positive Skewness, whereas a negative Skewness was due to the presence of a long left tail. The histogram graphs were plotted with the parametric values on the x-axis with a bin size of 0.024 min−1 for K trans, and 0.02 for Ve (with a bin number of 50) (Fig. 2a,b).
The first observer (Z.S.) measured parameters for the first MRI examination twice (for intra-observer reproducibility) and observers 2 (F.D.) and 3 (Y.S.) measured the parameters of the first examination once (to examine inter-observer reproducibility). Then the first observer measured parameters of the second examination once (for scan–rescan reproducibility), carefully choosing the same slice as in the first scan or as close as possible.
Statistical Analyses
Intra-, inter-observer, and scan–rescan differences in histogram metrics of kinetic parameters
Intra-observer and inter-scan differences were assessed using paired t tests. Inter-observer differences were evaluated using ANOVA.
Intra-, inter-observer, and scan–rescan agreement analyses in histogram metrics of kinetic parameters
Intra-observer, inter-observer, and scan–rescan agreements of histogram metrics of pharmacokinetic parameters were evaluated using the inter-class correlation coefficient (ICC). The agreement was defined as good (ICC > 0.75), moderate (ICC = 0.5–0.75), or poor (ICC < 0.5).
Intra-, inter-observer, and scan–rescan variability in histogram metrics of kinetic parameters
Coefficients of variation (CoV) were computed as the proportion of the standard deviation of the mean (standard deviation/mean, expressed as percentage). For CoVs describing the inter-observer variability, standard deviation was computed over each parameter obtained by all three observers. For CoVs concerning the intra-observer variability, standard deviation was computed over two measurements by each observer. For scan-rescan variability, the CoV for each subject was first computed and then averaged to obtain mean between patients’ CoVs for each parameter.
All statistical analyses were performed with SPSS software (IBM SPSS Statistics for Macintosh, Version 22.0. Armonk, NY: IBM Corp.) and GraphPad Prism (ver. 6.0; GraphPad Software, Inc., La Jolla, CA). P values <0 .05 were considered to indicate a statistically significant difference.
Results
Patients and lesions characteristics
A total of 28 patients with renal lesions underwent DCE-MRI scanning. After reviewing imaging quality and histopathologic results, two cases were excluded due to poor imaging quality and five cases due to other tumor types (1 papillary RCC, 3 chromophobic RCC, and a renal angiomyolipoma). Thus, 21 effective paired data sets of clear cell RCC cases (17 male, 4 female; age range 37–69 years, mean age 54.6 years; mean tumor size, 5.0 ± 2.2 cm) were included in this study.
Histogram metrics of pharmacokinetic parameters of renal cell carcinoma
Mean, Mode, Skewness, Kurtosis of K trans and Ve of each ROI of 21 patients were automatically calculated and recorded. Then all Mean, Mode, Skewness, Kurtosis were documented for intra-observer, inter-observer and scan-rescan comparison in Table 1.
Table 1. Histogram metrics of pharmacokinetic parameters of DCE-MRI and analysis on difference.
Kinetic Parameters | Histogram Metrics | Intra-observer |
Inter-observer |
Scan-rescan |
|||||||
---|---|---|---|---|---|---|---|---|---|---|---|
1st Measurement | 2nd Measurement | P value | Observer 1 | Observer 2 | Observer 3 | P value | 1st Scan | 2nd Scan | P value | ||
Ktrans | Mean (min−1) | 0.466 ± 0.140 | 0.465 ± 0.145 | 0.878 | 0.466 ± 0.140 | 0.457 ± 0.132 | 0.461 ± 0.137 | 0.986 | 0.466 ± 0.140 | 0.450 ± 0.092 | 0.581 |
Mode (min−1) | 0.370 ± 0.194 | 0.372 ± 0.189 | 0.754 | 0.370 ± 0.194 | 0.374 ± 0.196 | 0.402 ± 0.170 | 0.899 | 0.370 ± 0.194 | 0.325 ± 0.128 | 0.306 | |
Skewness | 0.622 ± 0.396 | 0.613 ± 0.374 | 0.638 | 0.622 ± 0.396 | 0.575 ± 0.281 | 0.654 ± 0.336 | 0.850 | 0.622 ± 0.396 | 0.653 ± 0.389 | 0.870 | |
Kurtosis | −0.752 ± 0.494 | −0.758 ± 0.512 | 0.927 | −0.752 ± 0.494 | −0.532 ± 0.658 | −0.878 ± 0.362 | 0.268 | −0.752 ± 0.494 | −0.708 ± 0.543 | 0.854 | |
Kep | Mean (min−1) | 0.822 ± 0.353 | 0.835 ± 0.352 | 0.339 | 0.823 ± 0.353 | 0.833 ± 0.358 | 0.839 ± 0.368 | 0.972 | 0.823 ± 0.353 | 0.760 ± 0.347 | 0.160 |
Mode (min−1) | 0.600 ± 0.300 | 0.550 ± 0.266 | 0.339 | 0.600 ± 0.300 | 0.525 ± 0.186 | 0.599 ± 0.300 | 0.732 | 0.600 ± 0.300 | 0.550 ± 0.266 | 438 | |
Skewness | 4.634 ± 1.398 | 4.671 ± 1.370 | 0.083 | 4.634 ± 1.398 | 4.490 ± 1.720 | 4.532 ± 1.215 | 0.971 | 4.634 ± 1.398 | 5.121 ± 1.206 | 0.097 | |
Kurtosis | 23.593 ± 13.392 | 23.954 ± 13.166 | 0.111 | 23.593 ± 13.392 | 23.126 ± 14.327 | 22.271 ± 12.777 | 0.971 | 23.593 ± 13.392 | 28.691 ± 12.979 | 0.104 | |
Ve | Mean (min−1) | 0.559 ± 0.107 | 0.558 ± 0.105 | 0.651 | 0.559 ± 0.107 | 0.551 ± 0.116 | 0.553 ± 0.118 | 0.985 | 0.559 ± 0.107 | 0.576 ± 0.107 | .0423 |
Mode (min−1) | 0.508 ± 0.231 | 0.511 ± 0.230 | 0.491 | 0.508 ± 0.231 | 0.517 ± 0.229 | 0.517 ± 0.216 | 0.995 | 0.508 ± 0.231 | 0.578 ± 0.224 | 0.116 | |
Skewness | 0.330 ± 0.370 | 0.386 ± 0.476 | 0.253 | 0.330 ± 0.370 | 0.290 ± 0.467 | 0.290 ± 0.425 | 0.730 | 0.330 ± 0.370 | 0.231 ± 0.572 | 0.322 | |
Kurtosis | −0.692 ± 0.485 | −0.577 ± 0.619 | 0.245 | −0.692 ± 0.485 | −0.712 ± 0.581 | −0.623 ± 0.640 | 0.928 | −0.692 ± 0.485 | −0.722 ± 0.746 | 0.816 |
Analysis of differences in kinetic parameters
There were no statistically significant intra-observer or inter-observer differences in any histogram metrics of each kinetic parameter examined, nor between MRI scan (all P > 0.05) (Table 1).
Agreement analysis
Intra- and inter-observer agreement
The intra-observer ICCs of histogram parameters and Mean of kinetic parameters were all greater than 0.80, which indicated good-to-excellent agreements (range, 0.824–0.999; P < 0.001) (Table 2). The inter-observer ICCs of Mean, Mode and Skewness of K trans demonstrated excellent agreement while Kurtosis of K trans showed moderate agreement (ICC, 0.728; 95%CI, 0.454~0.902). The inter-observer ICCs of histogram parameters and Mean of Ve showed good-to-excellent agreement (range, 0.828~0.968; P < 0.001). The ICCs details are listed in Table 2. Moreover, in both intra- and inter-observer agreement analyses, Mode, Skewness, and Kurtosis showed slightly lower ICCs than Mean.
Table 2. ICC analysis on histogram metrics of pharmacokinetic parameters of DCE-MRI.
Kinetic Parameters | Histogram Metrics | Intra-observer |
Inter-observer |
Scan-rescan |
|||
---|---|---|---|---|---|---|---|
ICC(95%CI) | P value | ICC(95%CI) | P value | ICC(95%CI) | P value | ||
K trans | Mean | 0.999 (0.996, 1.000) | <0.001 | 0.993 (0.981, 0.998) | <0.001 | 0.686 (0.212, 0.898) | 0.006 |
Mode | 0.994 (0.980, 0.998) | <0.001 | 0.923 (0.816, 0.975) | <0.001 | 0.616 (0.121, 0.870) | 0.001 | |
Skewness | 0.985 (0.951, 0.996) | <0.001 | 0.898 (0.761, 0.966) | <0.001 | 0.352 (−0.288, 0.762) | 0.863 | |
Kurtosis | 0.929 (0.770, 0.979) | <0.001 | 0.728 (0.454, 0.902) | <0.001 | 0.308 (−0.346, 0.743) | 0.767 | |
Ve | Mean | 0.998 (0.993, 0.999) | <0.001 | 0.991 (0.976, 0.997) | <0.001 | 0.764 (0.378, 0.925) | 0.001 |
Mode | 0.999 (0.998, 1.000) | <0.001 | 0.934 (0.837, 0.979) | <0.001 | 0.758 (0.370, 0.923) | 0.001 | |
Skewness | 0.925 (0.769, 0.977) | <0.001 | 0.945 (0.950, 0.994) | <0.001 | 0.766 (0.390, 0.926) | 0.001 | |
Kurtosis | 0.824 (0.517, 0.945) | <0.001 | 0.895 (0.755, 0.965) | <0.001 | 0.780 (0.562, 0.932) | 0.001 |
Scan-rescan agreement
ICC of all histogram parameters of Ve showed good agreement (range, 0.758~0.798, P < 0.001) and showed similar ICCs with Mean. However, Mean, Mode of K trans showed moderate agreement, Skewness and Kurtosis of Ktrans showed poor agreement (0.352, 0.308, respectively). The ICCs in details was listed in Table 2.
Variability analysis
Intra- and inter-observer variability
In both intra- and inter-observer analysis, Mean of Ktrans and Ve showed small variation (<=2.31%), Mode showed a larger variation (up to 10.54%), and Skewness and Kurtosis showed much higher CoVs than Mean (Fig. 3a,b) except for Skewness of Ktrans in intra-observer analysis.
Scan-rescan variability
In scan-rescan analysis, Mean of K trans and Ve showed small variation (10.82% and 6.88% respectively), Mode of K trans and Ve showed relatively larger variation (25.44% and 15.43% respectively); however, Mode, Skewness and Kurtosis demonstrated larger variation, especially for Skewness and Kurtosis (>30%) (Fig. 3c).
In addition, when comparing scan-rescan performance with intra- and inter-observer performance, the former variation was greater than the latter (Table 3) for nearly all histogram metrics of both K trans and Ve. In scan-rescan analysis, Mean value of pharmacokinetic parameters was similar between the two scans, and Skewness and Kurtosis showed obvious difference (Fig. 4a,b).
Table 3. Variability analysis on histogram metrics of pharmacokinetic parameters of DCE-MRI.
Kinetic Parameters | Histogram Metrics | Coefficient of variation (%) |
||
---|---|---|---|---|
Intra-observer | Inter-observer | Scan-rescan | ||
K trans | Mean | 0.98 | 2.31 | 10.82 |
Mode | 2.10 | 10.54 | 25.44 | |
Skewness | 0.73 | 23.84 | 32.29 | |
Kurtosis | 42.22 | 66.72 | 85.84 | |
Ve | Mean | 0.72 | 1.84 | 6.88 |
Mode | 0.66 | 9.21 | 15.43 | |
Skewness | 47.87 | 114.86 | 109.42 | |
Kurtosis | 40.92 | 87.36 | 40.53 |
Discussion
In this study, we found that scan-rescan performance had a relatively poorer reproducibility than intra- and inter-observer analysis regarding to histogram metrics of DCE-MRI pharmacokinetic parameters (K trans & Ve) in RCC. As for agreement analysis, scan-rescan ICCs of all histogram parameters were lower than intra- and inter-observer ICCs and intra-observer performance showed the highest ICCs. This suggested that although we attempted to ensure the situations were identical between the 1st and 2nd scan, it was unavoidable that minute differences in biological elements and/or hardware situation persisted between two scans, which likely resulted in more variation than difference of observers or drawing ROI.
In analyzing the variability results, scan-rescan variation for most of parameters was higher than intra- and inter-observer variation. However, Skewness and Kurtosis of Ve in inter-observer analysis showed the largest variation, which probably indicated that the observers exerted relatively great influence on measurement of these two values. In another aspect, when making comparison among the four histogram metrics of pharmacokinetic parameters regarding to reproducibility, we found that Mean and Mode presented better reproducibility than Skewness and Kurtosis in intra-, inter-observer and scan-rescan performance. These results showed that although heterogeneity analysis has been a trend in quantitative image analysis, it may not be as reproducible as standard Mean value analysis.
In examining intra- and inter-observer agreement, Mean of K trans and Ve demonstrated good agreement (all ICC values >0.75). Similar results were previously reported by Davenport et al.38 (i.e., inter-observer agreement: 0.88 and 0.87 ICCs for K trans and Ve, respectively) and a study by Braunagel et al. also on RCC (ICC ranging from 0.79~0.97 K trans, Kep, and Vp in both intra- and inter-observer agreement)39. In scan-rescan agreement analysis, Mean of Ve showed good agreement (ICC, 0.764), which was in accordance with previous studies in gliomas37 and uterine fibroids40.
However, for Ktrans alone, Skewness and Kurtosis demonstrated markedly lower ICCs and higher variation than Mean and Mode except for Skewness in intra-observer analysis. Additionally, for Ve alone, although ICC analysis showed similar result, variation of Skewness and Kurtosis were much higher than Mean and Mode. It is not clear why Skewness and Kurtosis were relatively poorly reproducible than Mean and Mode. We posit that the former was more sensitive to human interference (intra-observer), experience (inter-observer), and change of situation (scan-rescan) than the latter. However, we cannot rule out the likelihood that Skewness and Kurtosis were probably more sensitive to minute tumor changes.
Furthermore, we demonstrated that when comparing Ktrans with Ve, Mean of Ve had better reproducibility than Ktrans, which we also observed in our prior study study11. However for Skewness and Kurtosis, Ve and Ktrans showed poor reproducibility except for Skewness in intra-observer analysis.
During parameter extraction, the most sensitive method to a dynamic scan’s temporal resolution is AIF. Personal or individual AIF if calculated accurately can improve performance of pharmacokinetic parameters, however, personal AIF requires a high temporal resolution and may be influenced by patients’ physiological condition, ROI placement, partial volume effect and inflow effect etc. So it is almost impossible to have an identical AIF when performing scans twice in the same patient. Due to non-continuous scanning mode of DCE-MRI (See “MRI technique” in Methods) for balancing the needs of clinical practice and scientific research, the temporal resolution of DCE-MRI was limited in this study. These facts led us to use a population-based AIF method, rather than a personal AIF. Population-based AIF not only helped address temporal resolution difficulties but also reduced AIF ROI location and sizing errors that have been reported previously41. In addition, the population-based AIF works equally well as the individual AIF for estimating pharmacokinetic parameters, as confirmed by several investigators42,43,44.
In our study we performed the DCE-MRI scan on a 3.0-Tesla MRI system. When compared with 1.5- or 1.0-Tesla, 3.0-Tesla DCE-MRI presented higher SNR and faster scan speed (potentially increasing temporal resolution) which therefore benefit DCE-MRI performance. However, 3.0-Tesla DCE-MRI increased potency of magnetic susceptibility and chemical shift, especially susceptibility to air artifacts. Hence, it is not recommended that 3.0-Tesla DCE-MRI was used to evaluate tumors adjacent to air or gas45.
This study has a few limitations. Firstly, we analyzed only single slices of tumor. Although it is reported that the efficacy was similar with whole tumor analysis, this method will likely exclude some information reflecting on the whole tumor characteristics. However, whole tumor analysis is very time-consuming and manual ROI allocation on all slices may increase measurement error. Secondly, besides the histogram parameters we used, histogram metrics covers many more aspects. In this study, we only analyzed a portion of histogram metrics, Median, Percentiles, and Texture parameters (uniformity and entropy) were not taken into consideration; but we included the descriptive parameters and distribution parameters such as Skewness and Kurtosis, which can adequately analyze the average value and heterogeneity to a certain degree. Thirdly, we used renal tumor as an example to compare histogram metrics to conventional Mean value analysis. Potentially, these results cannot be generally extended to other types of tumors derived from other anatomical sites. Further studies and exploration of other tumors are therefore required.
In conclusion, histogram method (Mode, Skewness and Kurtosis) was inferior to the conventional Mean value method in reproducibility evaluation on DCE-MRI pharmacokinetic parameters (Ktrans & Ve) in renal cell carcinoma, which suggests that histogram analysis may not be appropriate for quantitative evaluation of DCE-MRI pharmacokinetic parameters in renal cell carcinoma at present.
Additional Information
How to cite this article: Wang, H.-y. et al. Dynamic Contrast-enhanced MR Imaging in Renal Cell Carcinoma: Reproducibility of Histogram Analysis on Pharmacokinetic Parameters. Sci. Rep. 6, 29146; doi: 10.1038/srep29146 (2016).
Acknowledgments
This study was supported by National Natural Science Foundation of China (Grant No. 81471641). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. We would like to express our gratitude for the technical support and assistance from Ning Huang Ph.D. of Life Science, GE Healthcare China, Zhenyu Zhou Ph.D. and Dandan Zheng Ph.D. of MR Research GE Healthcare China.
Footnotes
Author Contributions H.-y.W., Z.-h.S. and X.X. contributed equally to this work. H.-y.W., Z.-h.S. and H.-y.Y. designed this study. H.-y.W. and Z.-h.S. wrote the main manuscript; L.M. revised the manuscript. H.-y.W. did statistical analysis. H.-y.W. and X.X. prepared all figures; Z.-p.S., F.-x.D. and Y.-y.S. perform the ROI drawing. X.X. did the imaging analysis. L.L. and Y.-w.W. performed DCE-MRI scanning. X.M. enrolled patients with renal cell carcinoma. A.-t.G. did the pathologic analysis. H.-y.Y. supervised all the procedures of this study.
References
- El Khouli R. H. et al. Dynamic contrast-enhanced MRI of the breast: quantitative method for kinetic curve type assessment. AJR. American journal of roentgenology 193, W295–300, 10.2214/AJR.09.2483 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Engelbrecht M. R. et al. Discrimination of prostate cancer from normal peripheral zone and central gland tissue by using dynamic contrast-enhanced MR imaging. Radiology 229, 248–254, 10.1148/radiol.2291020200 (2003). [DOI] [PubMed] [Google Scholar]
- Jackson A., O’Connor J. P., Parker G. J. & Jayson G. C. Imaging tumor vascular heterogeneity and angiogenesis using dynamic contrast-enhanced magnetic resonance imaging. Clinical cancer research: an official journal of the American Association for Cancer Research 13, 3449–3459, 10.1158/1078-0432.CCR-07-0238 (2007). [DOI] [PubMed] [Google Scholar]
- Oostendorp M., Post M. J. & Backes W. H. Vessel growth and function: depiction with contrast-enhanced MR imaging. Radiology 251, 317–335, 10.1148/radiol.2512080485 (2009). [DOI] [PubMed] [Google Scholar]
- Yankeelov T. E. & Gore J. C. Dynamic Contrast Enhanced Magnetic Resonance Imaging in Oncology: Theory, Data Acquisition, Analysis, and Examples. Current medical imaging reviews 3, 91–107, 10.2174/157340507780619179 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zahra M. A., Hollingsworth K. G., Sala E., Lomas D. J. & Tan L. T. Dynamic contrast-enhanced MRI as a predictor of tumour response to radiotherapy. Lancet Oncol 8, 63–74, 10.1016/S1470-2045(06)71012-9 (2007). [DOI] [PubMed] [Google Scholar]
- Hotker A. M., Schmidtmann I., Oberholzer K. & Duber C. Dynamic contrast enhanced-MRI in rectal cancer: Inter- and intraobserver reproducibility and the effect of slice selection on pharmacokinetic analysis. Journal of magnetic resonance imaging: JMRI 40, 715–722, 10.1002/jmri.24385 (2014). [DOI] [PubMed] [Google Scholar]
- Gaens M. E. et al. Dynamic contrast-enhanced MR imaging of carotid atherosclerotic plaque: model selection, reproducibility, and validation. Radiology 266, 271–279, 10.1148/radiol.12120499 (2013). [DOI] [PubMed] [Google Scholar]
- Donekal S. et al. Inter-study reproducibility of cardiovascular magnetic resonance tagging. Journal of cardiovascular magnetic resonance: official journal of the Society for Cardiovascular Magnetic Resonance 15, 37, 10.1186/1532-429X-15-37 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bauknecht H. C. et al. Intra- and interobserver variability of linear and volumetric measurements of brain metastases using contrast-enhanced magnetic resonance imaging. Investigative radiology 45, 49–56, 10.1097/RLI.0b013e3181c02ed5 (2010). [DOI] [PubMed] [Google Scholar]
- Wang H. et al. Reproducibility of Dynamic Contrast-Enhanced MRI in Renal Cell Carcinoma: A Prospective Analysis on Intra- and Interobserver and Scan–Rescan Performance of Pharmacokinetic Parameters. Medicine 94, e1529, 10.1097/md.0000000000001529 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chaudhry H. S., Davenport M. S., Nieman C. M., Ho L. M. & Neville A. M. Histogram analysis of small solid renal masses: differentiating minimal fat angiomyolipoma from renal cell carcinoma. AJR. American journal of roentgenology 198, 377–383, 10.2214/AJR.11.6887 (2012). [DOI] [PubMed] [Google Scholar]
- Rodriguez Gutierrez D. et al. Metrics and textural features of MRI diffusion to improve classification of pediatric posterior fossa tumors. AJNR. American journal of neuroradiology 35, 1009–1015, 10.3174/ajnr.A3784 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gaing B. et al. Subtype differentiation of renal tumors using voxel-based histogram analysis of intravoxel incoherent motion parameters. Investigative radiology 50, 144–152, 10.1097/RLI.0000000000000111 (2015). [DOI] [PubMed] [Google Scholar]
- Tozer D. J. et al. Apparent diffusion coefficient histograms may predict low-grade glioma subtype. NMR in biomedicine 20, 49–57, 10.1002/nbm.1091 (2007). [DOI] [PubMed] [Google Scholar]
- Woo S., Cho J. Y., Kim S. Y. & Kim S. H. Histogram analysis of apparent diffusion coefficient map of diffusion-weighted MRI in endometrial cancer: a preliminary correlation study with histological grade. Acta radiologica 55, 1270–1277, 10.1177/0284185113514967 (2014). [DOI] [PubMed] [Google Scholar]
- Jung S. C. et al. Glioma: Application of histogram analysis of pharmacokinetic parameters from T1-weighted dynamic contrast-enhanced MR imaging to tumor grading. AJNR. American journal of neuroradiology 35, 1103–1110, 10.3174/ajnr.A3825 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Downey K. et al. Relationship between imaging biomarkers of stage I cervical cancer and poor-prognosis histologic features: quantitative histogram analysis of diffusion-weighted MR images. AJR. American journal of roentgenology 200, 314–320, 10.2214/AJR.12.9545 (2013). [DOI] [PubMed] [Google Scholar]
- Zhang Y. D. et al. The Histogram Analysis of Diffusion-Weighted Intravoxel Incoherent Motion (IVIM) Imaging for Differentiating the Gleason grade of Prostate Cancer. European radiology 25, 994–1004, 10.1007/s00330-014-3511-4 (2015). [DOI] [PubMed] [Google Scholar]
- Rosenkrantz A. B. et al. Whole-lesion diffusion metrics for assessment of bladder cancer aggressiveness. Abdominal imaging 40, 327–332, 10.1007/s00261-014-0213-y (2015). [DOI] [PubMed] [Google Scholar]
- Kim H. S., Suh C. H., Kim N., Choi C. G. & Kim S. J. Histogram analysis of intravoxel incoherent motion for differentiating recurrent tumor from treatment effect in patients with glioblastoma: initial clinical experience. AJNR. American journal of neuroradiology 35, 490–497, 10.3174/ajnr.A3719 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Steffen-Smith E. A. et al. Diffusion tensor histogram analysis of pediatric diffuse intrinsic pontine glioma. BioMed research international 2014, 647356, 10.1155/2014/647356 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yuh W. T. et al. Predicting control of primary tumor and survival by DCE MRI during early therapy in cervical cancer. Investigative radiology 44, 343–350, 10.1097/RLI.0b013e3181a64ce9 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Johansen R. et al. Predicting survival and early clinical response to primary chemotherapy for patients with locally advanced breast cancer using DCE-MRI. Journal of magnetic resonance imaging: JMRI 29, 1300–1307, 10.1002/jmri.21778 (2009). [DOI] [PubMed] [Google Scholar]
- Peng S. L. et al. Analysis of parametric histogram from dynamic contrast-enhanced MRI: application in evaluating brain tumor response to radiotherapy. NMR in biomedicine 26, 443–450, 10.1002/nbm.2882 (2013). [DOI] [PubMed] [Google Scholar]
- Heye T. et al. Reproducibility of dynamic contrast-enhanced MR imaging. Part II. Comparison of intra- and interobserver variability with manual region of interest placement versus semiautomatic lesion segmentation and histogram analysis. Radiology 266, 812–821, 10.1148/radiol.12120255 (2013). [DOI] [PubMed] [Google Scholar]
- Ruthotto L., Hodneland E. & Modersitzki J. In Biomedical Image Registration Vol. 7359 Lecture Notes in Computer Science (eds Dawant BenoîtM, Christensen GaryE., Fitzpatrick J. Michael & Rueckert Daniel) Ch. 20, 190–198 (Springer Berlin Heidelberg, 2012). [Google Scholar]
- Rosen M. A. & Schnall M. D. Dynamic contrast-enhanced magnetic resonance imaging for assessing tumor vascularity and vascular effects of targeted therapies in renal cell carcinoma. Clinical cancer research: an official journal of the American Association for Cancer Research 13, 770s–776s, 10.1158/1078-0432.CCR-06-1921 (2007). [DOI] [PubMed] [Google Scholar]
- Zollner F. G. et al. Assessment of 3D DCE-MRI of the kidneys using non-rigid image registration and segmentation of voxel time courses. Computerized medical imaging and graphics: the official journal of the Computerized Medical Imaging Society 33, 171–181, 10.1016/j.compmedimag.2008.11.004 (2009). [DOI] [PubMed] [Google Scholar]
- Klein A. et al. Evaluation of 14 nonlinear deformation algorithms applied to human brain MRI registration. NeuroImage 46, 786–802, 10.1016/j.neuroimage.2008.12.037 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rueckert D. et al. Nonrigid registration using free-form deformations: application to breast MR images. IEEE transactions on medical imaging 18, 712–721, 10.1109/42.796284 (1999). [DOI] [PubMed] [Google Scholar]
- Pluim J. P., Maintz J. B. & Viergever M. A. Mutual-information-based registration of medical images: a survey. IEEE transactions on medical imaging 22, 986–1004, 10.1109/tmi.2003.815867 (2003). [DOI] [PubMed] [Google Scholar]
- Khalifa F. et al. Models and methods for analyzing DCE-MRI: A review. Medical physics 41, 124301, 10.1118/1.4898202 (2014). [DOI] [PubMed] [Google Scholar]
- Whitcher B. & Schmid V. J. Quantitative Analysis of Dynamic Contrast-Enhanced and Diffusion-Weighted Magnetic Resonance Imaging for Oncology in R. 2011 44, 29, 10.18637/jss.v044.i05 (2011). [DOI]
- Tofts P. S. & Kermode A. G. Measurement of the blood-brain barrier permeability and leakage space using dynamic MR imaging. 1. Fundamental concepts. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 17, 357–367 (1991). [DOI] [PubMed] [Google Scholar]
- Parker G. J. et al. Probing tumor microvascularity by measurement, analysis and display of contrast agent uptake kinetics. Journal of magnetic resonance imaging: JMRI 7, 564–574 (1997). [DOI] [PubMed] [Google Scholar]
- Jackson A. et al. Reproducibility of quantitative dynamic contrast-enhanced MRI in newly presenting glioma. The British journal of radiology 76, 153–162 (2003). [DOI] [PubMed] [Google Scholar]
- Davenport M. S. et al. Inter- and intra-rater reproducibility of quantitative dynamic contrast enhanced MRI using TWIST perfusion data in a uterine fibroid model. Journal of magnetic resonance imaging: JMRI 38, 329–335, 10.1002/jmri.23974 (2013). [DOI] [PubMed] [Google Scholar]
- Braunagel M. et al. Dynamic contrast-enhanced magnetic resonance imaging measurements in renal cell carcinoma: effect of region of interest size and positioning on interobserver and intraobserver variability. Investigative radiology 50, 57–66, 10.1097/rli.0000000000000096 (2015). [DOI] [PubMed] [Google Scholar]
- Heye T. et al. Reproducibility of dynamic contrast-enhanced MR imaging. Part I. Perfusion characteristics in the female pelvis by using multiple computer-aided diagnosis perfusion analysis solutions. Radiology 266, 801–811, 10.1148/radiol.12120278 (2013). [DOI] [PubMed] [Google Scholar]
- Cutajar M., Mendichovszky I. A., Tofts P. S. & Gordon I. The importance of AIF ROI selection in DCE-MRI renography: reproducibility and variability of renal perfusion and filtration. European journal of radiology 74, e154–160, 10.1016/j.ejrad.2009.05.041 (2010). [DOI] [PubMed] [Google Scholar]
- Wang Y., Huang W., Panicek D. M., Schwartz L. H. & Koutcher J. A. Feasibility of using limited-population-based arterial input function for pharmacokinetic modeling of osteosarcoma dynamic contrast-enhanced MRI data. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 59, 1183–1189, 10.1002/mrm.21432 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Parker G. J. et al. Experimentally-derived functional form for a population-averaged high-temporal-resolution arterial input function for dynamic contrast-enhanced MRI. Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine/Society of Magnetic Resonance in Medicine 56, 993–1000, 10.1002/mrm.21066 (2006). [DOI] [PubMed] [Google Scholar]
- Li X. et al. A novel AIF tracking method and comparison of DCE-MRI parameters using individual and population-based AIFs in human breast cancer. Physics in medicine and biology 56, 5753–5769, 10.1088/0031-9155/56/17/018 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hirashima Y. et al. Pharmacokinetic parameters from 3-Tesla DCE-MRI as surrogate biomarkers of antitumor effects of bevacizumab plus FOLFIRI in colorectal cancer with liver metastasis. Int J Cancer 130, 2359–2365, 10.1002/ijc.26282 (2012). [DOI] [PubMed] [Google Scholar]