. 2016 Jul 7;11(7):e0159169. doi: 10.1371/journal.pone.0159169

Correction: Correlation between Serum Levels of 3,3ʹ,5ʹ-Triiodothyronine and Thyroid Hormones Measured by Liquid Chromatography-Tandem Mass Spectrometry and Immunoassay

Hiroyuki Sakai, Hidenori Nagao, Mamoru Sakurai, Takako Okumura, Yoshiyuki Nagai, Junpei Shikuma, Rokuro Ito, Tetsuya Imazu, Takashi Miwa, Masato Odawara
PMCID: PMC4936695  PMID: 27388908

There are numerical errors in the rT3 columns of Table 2 of the published article. Please see the correct Table 2 here.

Table 2. Precision and Accuracy of intra-day assay.

Concentration Accuracy (%)
(ng/mL) T4 T3 rT3
0.05 104.0 ± 8.0 (7.7) 110.0 ± 4.0 (3.6) 108.0 ± 6.0 (5.6)
0.1 105.0 ±6.0 (5.7) 104.0 ± 7.0 (6.7) 102.0 ± 11.0 (10.8)
2.5 108.1 ± 3.1 (2.9) 100.3 ± 5.5 (5.5) 101.8 ± 4.5 (4.4)
80 102.5 ± 1.3 (1.3) 96.2 ± 6.6 (6.9) 98.5 ± 6.4 (6.5)

Data are expressed as the mean ± S.D. (n = 5).

Values in parentheses are coefficients of variance (%).

Reference

  • 1.Sakai H, Nagao H, Sakurai M, Okumura T, Nagai Y, Shikuma J, et al. (2015) Correlation between Serum Levels of 3,3′,5′-Triiodothyronine and Thyroid Hormones Measured by Liquid Chromatography-Tandem Mass Spectrometry and Immunoassay. PLoS ONE 10(10): e0138864 doi: 10.1371/journal.pone.0138864 [DOI] [PMC free article] [PubMed] [Google Scholar]