Skip to main content
Journal of Neurology, Neurosurgery, and Psychiatry logoLink to Journal of Neurology, Neurosurgery, and Psychiatry
. 1972 Apr;35(2):216–220. doi: 10.1136/jnnp.35.2.216

Quantitative electromyography: a new analogue technique for detecting changes in action potential duration

A Moosa 1,2, B H Brown 1,2
PMCID: PMC494039  PMID: 4556578

Abstract

An automatic method using an analogue analyser is described for obtaining an index which is based on the mean phase duration of an EMG signal. This method gives similar values for motor unit action potential duration to an alternative digital method of analysis based on the distribution of time intervals between zero crossings of the EMG signal. The analysis has been found to give good discrimination of normal and myopathic EMG signals.

Full text

PDF
216

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Gardner-Medwin D. Studies of the carrier state in the Duchenne type of muscular dystrophy. 2. Quantitative electromyography as a method of carrier detection. J Neurol Neurosurg Psychiatry. 1968 Apr;31(2):124–134. doi: 10.1136/jnnp.31.2.124. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Rose A. L., Willison R. G. Quantitative electromyography using automatic analysis: studies in healthy subjects and patients with primary muscle disease. J Neurol Neurosurg Psychiatry. 1967 Oct;30(5):403–410. doi: 10.1136/jnnp.30.5.403. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Neurology, Neurosurgery, and Psychiatry are provided here courtesy of BMJ Publishing Group