Abstract
Background
Patient-specific absorbed dose calculations for molecular radiotherapy require accurate activity quantification. This is commonly derived from Single-Photon Emission Computed Tomography (SPECT) imaging using a calibration factor relating detected counts to known activity in a phantom insert.
Methods
A series of phantom inserts, based on the mathematical models underlying many clinical dosimetry calculations, have been produced using 3D printing techniques. SPECT/CT data for the phantom inserts has been used to calculate new organ-specific calibration factors for 99mTc and 177Lu. The measured calibration factors are compared to predicted values from calculations using a Gaussian kernel.
Results
Measured SPECT calibration factors for 3D printed organs display a clear dependence on organ shape for 99mTc and 177Lu. The observed variation in calibration factor is reproduced using Gaussian kernel-based calculation over two orders of magnitude change in insert volume for 99mTc and 177Lu. These new organ-specific calibration factors show a 24, 11 and 8 % reduction in absorbed dose for the liver, spleen and kidneys, respectively.
Conclusions
Non-spherical calibration factors from 3D printed phantom inserts can significantly improve the accuracy of whole organ activity quantification for molecular radiotherapy, providing a crucial step towards individualised activity quantification and patient-specific dosimetry. 3D printed inserts are found to provide a cost effective and efficient way for clinical centres to access more realistic phantom data.
Keywords: Activity quantification, SPECT, Molecular radiotherapy, 3D printing, Absorbed dose
Background
Accurate activity quantification is essential for patient-specific absorbed dose calculations for Molecular Radiotherapy (MRT). Current activity quantification for Single-Photon Emission Computed Tomography (SPECT), as commonly used in MRT, often relies on camera-specific calibration factors, determined from phantom data. These camera-specific calibrations are used to compensate for the many quantitative inaccuracies in SPECT imaging [1]. Calibration factors determined from physical phantom data are highly dependent on the specific geometry and activity distribution being considered. The geometry of phantom inserts used for SPECT activity calibration is often far removed from both the clinical situation and the mathematical models which provide the basis of the most common MRT dosimetry calculations. The OLINDA/EXM software [2] applies the MIRD schema for whole organ dosimetry [3], using mathematical models defined in [4–7], and is currently the only FDA-approved general MRT dosimetry code. Whilst these generic organ models do not reflect the range of MRT patient morphologies, they do allow dose calculations to be adapted to consider individual organ masses and patient age and sex. As such, the ready availability of phantoms providing activity distributions which better approximate clinical MRT distributions will provide a significant step forward towards accurate patient-specific dosimetry for all clinical centres performing MRT.
In this work, a series of novel phantom inserts corresponding to organ models used in common MRT dosimetry systems have been produced, based on the models described in [4]. The inserts have been printed using fused filament modelling (3D printing) with acrylonitrile butadiene styrene (ABS) plastic. This paper presents SPECT/CT data for the printed inserts containing 99mTc and 177Lu in a water-filled phantom. The suitability of ABS plastic as a tissue analogue for MRT phantom production is demonstrated for the first time. SPECT calibration factors for 99mTc and 177Lu, corresponding to the MIRD organ geometries defined in [4], are determined. A significant dependence on organ shape for calibration factors is demonstrated and compared to calculated values. Finally, the influence of improved organ-specific calibration factors on clinically representative absorbed dose calculations (using the OLINDA/EXM package) is evaluated. 3D printing production allows the phantom inserts to be made available [8] in stereolithography (STL) format, which can be printed using a commercial 3D printer or printing service. Providing access to the phantom inserts described in this work will allow any clinical centre to accurately calibrate clinical SPECT activity quantification for input to whole organ MRT dosimetry calculations, resulting in improved dose calculations.
Methods
3D printed phantom inserts
The designs and final printed inserts are shown in Fig. 1. The inserts consist of shells with 2-mm thick walls (3 mm for the larger liver model) which can be filled with a radioisotope solution. The inserts have integrated attachment and filling ports which can be fitted with standard M5 screw threads, allowing them to be used in a variety of commercial phantoms. Inserts were designed based on the adult male models for MRT dose-limiting organs as described in [4] (see Table 1 for details). Additional inserts corresponding to age-5 and age-10 kidneys were also produced. The adult male liver model has been extended to include two separate lobes and a small (16 ml) tumour. The variation of the insert volumes from those described in [4] (shown in Table 1) reflects the physical constraints of the 3D printing technique. Binary STL files of the models, compatible with most common models of printer, are available online [8]. The inserts used in this work were printed from ABS plastic (ABSplus - Stratasys Ltd) with a 0.178-mm layer thickness (solid wall interior) using a Dimension Elite printer (Stratasys Ltd).
Table 1.
Organ a | Model volume b | Printed insert volume | Activity 99mTc | Activity 177Lu |
---|---|---|---|---|
(cm 3) | (cm 3) | (MBq) | (MBq) | |
Spleen | 176.0 | 195.26 ± 0.02 | 122.3 ± 0.3 | 639.1 ± 1.2 |
Pancreas | 83.8 | 103.26 ± 0.02 | 110.4 ± 0.3 | 397.4 ± 0.7 |
Liver | 1830.0 | 1577.49 ± 0.02 | 515.3 ± 1.2 | 2976.8 ± 5.6 |
Liver (left) c | – | 479.00 ± 0.02 | 172.4 ± 0.8 | 924.2 ± 4.0 |
Liver (right) c | – | 1098.49 ± 0.02 | 342.9 ± 0.9 | 2052.6 ± 3.9 |
Liver (tumour) c | – | 16.49 ± 0.02 | 6.4 ± 0.2 | 43.0 ± 0.5 |
Kidney | 144.0 | 169.81 ± 0.02 | 125.0 ± 0.3 | 1024.6 ± 1.2 |
Kidney age 10 | 80.3 | 99.54 ± 0.02 | 92.5 ± 0.2 | 589.5 ± 0.2 |
Kidney age 5 | 55.5 | 64.96 ± 0.02 | 32.6 ± 0.1 | 383.9 ± 0.2 |
SPECT phantom data
The phantom inserts were filled with saline containing either 99mTc pertechnetate or 177Lu-Dotatate (with the activities given in Table 1) and placed individually in the centre of a water-filled elliptical phantom without activity (9.5 l, 186-mm high with 305 and 221 mm radii). The phantom was imaged in the same central position on the bed with a fixed head radius of 25 cm for each insert to ensure a constant SPECT resolution. Positioning both the phantom and insert centre-of-mass in the centre of the SPECT field-of-view ensures optimal performance of CT-based attenuation correction for SPECT reconstruction [9]. SPECT/CT data were acquired for each insert and radioisotope using an Infinia-Hawkeye-4 SPECT/CT camera (GE Healthcare). LEHR and MEGP collimators were used for 99mTc and 177Lu, respectively. Scans were performed using 60 views in a step-and-shoot circular orbit (20 s per view for 99mTc and 40 s per view for 177Lu). Projection data was acquired in a matrix of 128 × 128 4.42-mm pixels for the 140 ±10% keV photopeak of 99mTc and the 208 ±10% of 177Lu. Data was also acquired for energy windows adjacent to the photopeak (99mTc 120 ±3% keV, 177Lu 181 ±3% and 236 ±3% keV) allowing dual or triple energy windows scatter correction to be applied [10]. The phantom data were reconstructed using a clinical OSEM algorithm (GE OSEM) with ten subsets and four iterations. This reconstruction algorithm does not include collimator-detector response (CDR) compensation, which can improve spatial resolution and reduce spill-out of counts. It should be noted however that CDR compensation may create ringing artefacts which change the distribution of counts within a volume of interest (VOI) [11]. The reconstruction parameters were found to provide stable quantification for all organ VOIs for both 99mTc and 177Lu, with the total number of reconstructed counts increasing by <0.7 % for higher numbers of reconstruction iterations. Attenuation correction (AC) was applied using Hawkeye CT data (140 kV, 2.5 mA) acquired as part of the scan protocol. The central slice through the SPECT reconstructions of the inserts, demonstrating the activity distribution, is shown in Fig. 1.
Organ-specific SPECT calibration factors
Generally, a calibration factor, cf, is defined which relates the count rate (counts per second, cps) in a reconstructed SPECT image volume-of-interest (VOI) to the total activity (MBq) in the VOI in order to provide activity quantification of SPECT imaging for MRT patients,
1 |
A VOI defining each organ insert was manually outlined using the Hawkeye CT data and transferred to the SPECT image. A corresponding calibration factor (cfvoi) for 99mTc and 177Lu was calculated for each VOI. The total number of isotropic voxels, of size 4.42 mm (as used for the SPECT images), defining the inner volume for each VOI was also measured. The sensitivity of a SPECT camera is system specific. Experimentally determined SPECT calibration factors are highly dependent on the choice of reconstruction parameters, including scatter correction, attenuation correction and partial volume corrections. In addition, when applying any phantom-based SPECT calibration factors to patient data, the activity concentration outside the VOI, which varies for each patient and organ, must also be accounted for. In order to isolate the influence of VOI shape on calibration factor from these effects, identical reconstruction parameters and phantom position were used for all scans with no background activity. This ensures that any variation in measured calibration factors are due to the choice of phantom insert alone, highlighting the benefits of using more representative organ geometries.
Measuring true camera sensitivity calibration factor (cftrue)
The true camera sensitivity calibration factor, cftrue, gives the SPECT camera sensitivity (cps/MBq) when no partial volume effects are present. To determine cftrue, additional SPECT/CT data was acquired for a 7.3-l cylindrical phantom (height 200 mm and radius 108 mm) containing 712(1) MBq and 3551(7) MBq of 99mTc and 177Lu, respectively. The phantom was imaged using the same acquisition and reconstruction protocols previously described for the phantom inserts. The true calibration factor was calculated (equation 1) from the uniform phantom data using the reconstructed total number of counts in the SPECT field of view.
VOI calibration factor
The finite resolution of a reconstructed SPECT image causes the measured calibration factor for a specific VOI (cfvoi) to deviate from the “true” camera sensitivity calibration factor, cftrue due to spatial blurring of the count distribution. The measured counts detected within a VOI (countsvoi) can therefore be defined by,
2 |
where countstrue is the “true” number of counts in the VOI and countsoutside represents counts originating from the VOI which are detected outside the VOI. For a simple geometry, described by a known mathematical function, the effects of finite spatial resolution and discritization of counts in a voxelised VOI may be considered individually. However, obtaining an analytical solution for spatial blurring of an arbitrarily shaped volume is often not possible. In this work, a voxelised approach, which can be applied to any VOI and combines the effect of discretization and limited spatial resolution, is presented. For a VOI with nVoxels, total voxels countsoutside can be defined as,
3 |
where αi is the fraction of true events in voxel i which are located outside of the VOI and ci is the number of true events in the voxel. The value of αi can be calculated for an individual voxel within a VOI, centred at xi,yi and zi, by considering the convolution with a 3D Gaussian kernel, g(x,y,z,σ), integrated over the volume of the VOI.
4 |
where xa,xb,ya,yb,za and zb are the closest perpendicular distances to the VOI boundary (for a voxel centred at xi,yi and zi) in the x, y and z dimensions, respectively. g(x,y,z,σ) is defined as,
5 |
and σ is related to the full width at half maximum (FWHM) of the SPECT camera,
6 |
In the case of a phantom insert with a uniform activity distribution (constant ci), combining Eqs. 1–3 gives the variation in measured calibration factor (cfvoi) from (cftrue) for a specific VOI, with αi defined by Eq. 4,
7 |
The value of αi, and therefore cfvoi, is highly dependent on both the specific three-dimensional geometry of a VOI and the resolution of the SPECT camera. In addition, for voxels located on the edge of a VOI, the voxel size will define the closest perpendicular distance to the VOI boundary. In general, the value of αi for a voxel on the perimeter of a VOI will vary significantly according to position on the surface, whilst for centrally located voxels, αi∼0 (when the VOI is sufficiently large compared to FWHM).
For example, with a spatial resolution of 10 mm, an isotropic 4.42-mm voxel size and no activity outside the VOI, Eq. 4 predicts αi = 0.30 (a 30 % loss in counts) for a voxel in the centre of a face of a cubic VOI of length 300 mm. For a voxel on the centre of one edge or corner of the VOI perimeter, αi = 0.51 (51 % loss) and 0.66 (66 % loss of counts), respectively, and for a single cubic voxel, αi = 0.94. The corresponding values of αi for an infinitesimal voxel size (point source) are 0.5 (centre of a face), 0.75 (edge of cube) and 0.875 (corner). In general, for a voxel on the edge of a VOI, the value of αi can be expressed as αi(x,σ)n where n is the number of VOI edges intersected by the voxel. The spill-out fraction from a VOI can be approximated as being proportional to areavoi/volumevoi.
When activity is present outside the VOI, then Eq. 7 has an additional spill-in component, which is analogous to Eq. 3 but summed over all voxels containing activity outside the VOI, increasing the value of cfvoi/cftrue.
Absorbed dose calculations
Absorbed dose calculations were performed on VOIs corresponding to the liver, spleen and kidneys for co-registered sequential SPECT/CT images of a patient receiving 177Lu-Dotatate therapy with the OLINDA/EXM package [2]. VOIs for these organs were manually defined at each imaging time point on the CT images fused with the SPECT data. The patient CT images showed no evidence of motion between SPECT and CT acquisition or imaging artefacts at the organ edges which may affect VOI definition. Separate time-activity curves (for imaging times, 1-, 24-, 71- and 143-hr post injection) were generated, using a calibration factor for each organ, or with standard sphere (113 ml, 27-mm radius) and cylinder calibration factors (278 ml, 44-mm diameter and 183-mm height). The residence time (Acumulative/Ainjected), as defined by [2], for each organ was calculated using a fit to a single exponential decay and an initial injection activity of 7448 MBq. MRT time-activity data is highly patient-specific and varies significantly for different organs and tumour burdens, often requiring detailed pharmacokinetic modelling [12]. For the patient time-activity data chosen in this study, the difference in calculated residence time for a bi-exponential fit, compared to a single exponential, was found to be <1 %, significantly smaller than the uncertainty in Acumulative (∼12 %). Masses were calculated from CT imaging for the liver, spleen and kidneys; otherwise, organ masses were scaled to the total body weight of the patient.
Results
ABS material properties
CT data from SPECT/CT scans of the phantom inserts were first used to measure the CT attenuation of the ABS material. The ABS material in the inserts was found to have a mean value of −54±13 HU. The effective density of an ABS print depends on the model geometry and properties of the printer. The inserts produced in this work have a mean effective density of 1.01 ±0.02 g/cm 3, corresponding to a packing density of 0.97 ±0.02. ABS printed material was found to provide a closer analogue to water (−7.4±8.7 HU) than perspex (116 ±9 HU) reducing the artificial boundary effect between regions of activity in the phantom. This result demonstrates that ABS printing, the most commonly available and affordable form of 3D printing, is a suitable technique for MRT phantom production.
Whole organ calibration factors
The measured true camera sensitivity calibration factors (cftrue) for 99mTc and 177Lu are given in Table 2. These values were calculated from uniform phantom data using the reconstructed total number of counts in the SPECT field of view. The ratio of the measured calibration factor (cfvoi) to cftrue for 99mTc and 177Lu, calculated using CT-defined VOIs outlining the phantom inserts, is shown in Fig. 2 (black squares). Data is presented for non-spherical inserts consisting of 3D printed organ inserts, for a 278-ml perspex cylinder and for a series of spherical perspex inserts (2 to 113 ml). The plotted standard uncertainties include a ±1 voxel (4.42 mm) uncertainty in positioning the VOIs.
Table 2.
Isotope | Phantom activity | cf true |
---|---|---|
(MBq) | (cps/MBq) | |
99mTc | 712 ± 1 | 150.49 ± 0.34 |
177Lu | 3551 ± 7 | 11.74 ± 0.02 |
As predicted (by Eqs. 4 and 7), the value of cfvoi/cftrue approaches 1 for the larger organ inserts and is significantly reduced for the smaller spherical inserts. The theoretical value of cfvoi/cftrue for a uniform activity distribution, calculated using Eq. 7 and a FWHM of 15 mm, is shown for each VOI in Fig. 2 (red circles). There is a good agreement with the measured values of cfvoi/cftrue for the larger inserts, including the 113-ml sphere. However, the measured values of cfvoi/cftrue is systemically larger for the smaller spherical inserts (2–16 ml). In this paper, data is presented without scatter correction to ensure the influence of organ shape is isolated. It should be noted however that the same tendency in cfvoi (with a corresponding reduction in counts per MBq) is observed for both 99mTc and 177Lu when consistent scatter correction is applied to these data (using the previously defined scatter windows).
Absorbed dose calculations
Table 3 shows 177Lu cumulative activity and absorbed dose calculations for the liver, spleen and kidneys using organ-specific activity calibration factors from 3D inserts (as previously described). Values are also given for a standard spherical insert (113 ml) and cylindrical insert (278 ml) calibration factor for comparison, which approximate the average organ volumes in Table 1 and are commonly available to clinical centres.
Table 3.
Organ factor | Sphere factor a | Cylinder factor b | ||||||
---|---|---|---|---|---|---|---|---|
A cumulative | Abs. Dose | A cumulative | Abs. dose | Δ Dose c | A cumulative | Abs. dose | Δ Dose c | |
(GBq.Hrs) | (mGy) | (GBq.Hrs) | (mGy) | (%) | (GBq.Hrs) | (mGy) | (%) | |
Spleen | 48.8 ± 9.2 | 10500 | 54.9 ± 10.1 | 11800 | 11.0 | 43.4 ± 0.8 | 9370 | −12.1 |
Liver | 59.8 ± 4.2 | 2690 | 78.6 ± 5.5 | 3540 | 24.0 | 62.1 ± 4.4 | 2790 | 3.6 |
Kidneys | 13.9 ± 1.6 | 6040 | 15.0 ± 1.1 | 6570 | 8.1 | 11.9 ± 0.8 | 5190 | −16.4 |
OLINDA/EXM calculations [2] using representative organ-specific time-activity curve data from clinical patient
a113-ml volume
b278-ml volume
cChanged in absorbed dose compared to using organ calibration factor
Discussion
Organ-specific calibration factors
SPECT activity calibration factors measured for 3D printed organ inserts display a clear dependence on organ shape for both 99mTc and 177Lu, see Fig. 2. For each insert, the ratio of cfvoi to cftrue is constant for 99mTc and 177Lu, demonstrating that the shape and volume of the VOI are the main influences on relative calibration factor. The measured values of cftrue (Table 2) reflect the gamma emission probabilities for the 140 keV emission from the decay of 99mTc and the 208 keV emission following the decay of 177Lu (89 and 10.4 %, respectively) and the system sensitivities at these energies.
Figure 2 shows close agreement between the measured VOI calibration factors and theoretical values, as defined in Eq. 7 with a FWHM of 15 mm (σ= 6.4 mm), for insert volumes >16 ml and for 99mTc and 177Lu. The converging value of FWHM (which will vary with the choice of collimator, orbital distance and emission energy) for 99mTc (LEHR collimators) and 177Lu (MEGP collimators) at a 25-cm orbital radius is consistent with previously reported data [13].
For all insert data, discretization effects will give counts in voxels adjacent to the VOI, analogous to background activity. This effect is negligible for large volume inserts resulting in good agreement with calculated values from Eq. 7. For the smaller inserts, the CT-defined VOIs correspond to a relatively small number of voxels (20–180 voxels in 2–16 ml) and so the effect is enhanced, resulting in the underestimation of the calculated values (see Fig. 2). The inclusion of an additional component, as previously discussed in the case of background activity outside the VOI, is required to fully reproduce the measured values of cfvoi/cftrue below 16 ml. Compensation for this effect is particularly important for accurate tumour activity quantification.
This work has demonstrated a clear relationship between measured whole organ calibration factors and VOI shape. The observed variation in calibration factor has been reproduced with a simple voxel level calculation over two orders of magnitude change in insert volume. These results highlight the potential for individualised whole organ activity quantification, corresponding to patient organ VOIs. An alternative method for deriving whole organ calibration factors based on small VOIs placed in areas of homogeneous activity within organs is reported in [14]. It should be noted however that this method does not account for any inhomogeneous activity distribution within an VOI or for the the shape of the VOI, as presented in this paper.
The close agreement between measured VOI calibration factors and theoretical values calculated using Eq. 7 shown in this current work highlights the potential to use theoretical VOI specific calibration factors. It should be noted, however, that activity quantification from SPECT imaging relies on a complex chain of related corrections and calibration [1] determinations. The presence of background activity surrounding a VOI, as seen in clinical situations, will result in variable amounts of spill-in which must be accounted for when applying any SPECT calibration. In addition, changes to the specific SPECT reconstruction, including scatter correction and VOI definition techniques [15], will result in different activity calibration factors. Consequently, whilst the analysis presented in this paper has been constrained to focus on the influence of insert geometry on SPECT calibration, the trends in calibration factor resulting from VOI shapes must be determined for each individual SPECT system and choice of reconstruction parameters. The use of 3D printed inserts provides a cost-effective and efficient way for clinical centres to produce phantom data for non-spherical activity distributions. Access to this data is essential for the validation of whole organ calibration factors (and corrections based on them) and has the potential to improve the accuracy of any SPECT quantification.
Impact on absorbed dose calculations
The representative absorbed dose calculations for 177Lu shown in Table 3 demonstrate a significant variation in calculated dose to all organs with changing calibration factor. The observed variations in organ-specific calibration factors from a spherical factor (as shown in Fig 2) translate directly into a corresponding change in residence time and subsequent absorbed dose calculation.
In particular, a 24 % reduction in calculated absorbed dose for the liver is seen when using calibration factors from a 3D printed organ insert in comparison to a factor from a standard 113-ml spherical insert. It should be noted that the measured value cfvoi for the liver insert is still 20 % smaller than the true camera sensitivity factor cftrue. A calibration factor (cfvoi/cftrue = 0.86) for a theoretical 1577-ml sphere, matching the volume of the liver insert, can be calculated using Eq. 7. Applying this larger spherical factor reduces the calculated absorbed dose for the liver to 2880 mGy. The calculated dose is still 7 % higher than when an organ-specific factor is applied (see Table 3) showing that this approach does not fully account for the shape of a VOI. This highlights the importance of using appropriate calibration factors which match both the shape and volume of a VOI. For the spleen and kidneys, a 113-ml sphere is a closer approximation to the organ geometry; however, a reduction in organ dose of 11 and 8.1 %, respectively, is observed when using organ-specific calibration factors compared to a 113-ml spherical calibration factor. These results suggest that the accuracy of absorbed dose calculations may be profoundly influenced by using these generic organ models. There is a potential for further improvement by extending the technique presented in this paper to patient-specific CT-based phantoms [16], providing either individualised phantom insert measurements or a general validation for patient-specific calculations. These techniques have the potential to provide fully individualised activity quantification which can be further extended to provide improved sub-organ volume dosimetry. The most efficient clinical implementation of these techniques on an individual patient basis remains to be established.
Conclusions
This work demonstrates a novel technique which delivers improved whole organ SPECT activity quantification and absorbed dose calculations for MRT. The inserts developed in this work provide unique SPECT imaging data, with activity distributions matching the underlying mathematical models for MRT dosimetry, which can be used to validate all approaches to SPECT activity quantification. Organ-specific SPECT calibration factors for 99mTc and 177Lu demonstrate a clear dependence on organ shape for both 99mTc and 177Lu, with a reduction in calculated absorbed dose compared to simple spherical inserts. The results presented in this work are accurately reproduced with Gaussian kernel-based calculations, over two orders of magnitude change in insert volume. These results highlight the potential for individualised whole organ activity quantification, corresponding to patient organ VOIs, to improve accuracy in whole organ dosimetry. The ability to produce MRT phantom inserts, with a reduced boundary effect between regions of activity in the phantom, using commonly available ABS printers allows complex phantom inserts to be made available to all clinical centres. The ability for clinical centres to test SPECT systems with activity distributions which represent the mathematical models underlying MIRD schema dose calculations is a crucial step in establishing validated MRT dosimetry.
Acknowledgements
This work was supported by the Science and Technology Facilities Council [ST/M004589/1].
Footnotes
Authors’ contributions
APR conceived the study design, performed the data collection and data analysis and drafted the manuscript. JT help conceive the study design and performed the data collection and revised the manuscript. DMC revised the manuscript and is the research funding grant holder. DH revised the manuscript and advised on the study design. RB contributed to the design of 3D printed phantom inserts and data acquisition. AF contributed to the design of 3D printed phantom inserts and data acquisition. CO contributed to the theoretical model of partial volume effects presented in this paper and the data acquisition. EP contributed to the absorbed dose calculations presented in this paper. EP contributed to the 177Lu data acquisition and phantom insert production. AS provided the major contribution to phantom insert design and production. RS contributed to the 177Lu data acquisition and phantom insert production. All authors read and approved the final manuscript.
Competing interests
The authors declare that they have no competing interests.
References
- 1.Dewaraja Y, Frey E, Sgouros G, Brill A, Roberson P, Zanzonico P, Ljungberg M. MIRD pamphlet no. 23: quantitative SPECT for patient-specific 3-dimensional dosimetry in internal radionuclide therapy. J Nucl Med. 2012;53(8):1310–25. doi: 10.2967/jnumed.111.100123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Stabin M, Sparks R, Crowe E. OLINDA/EXM: the second-generation personal computer software for internal dose assessment in nuclear medicine. J Nucl Med. 2005;46(6):1023–7. [PubMed] [Google Scholar]
- 3.Snyder W, Ford M, Warner G, Watson S. Mird pamphlet no. 11: S, absorbed dose per unit cumulated activity for selected radionuclides and organs (part 1–3). Society of Nuclear Medicine, Reston, VA. 1975:1–257.
- 4.Cristy M, Eckerman K. Specific Absorbed fractions of energy at various ages from internal photon sources: 1, Methods. Oak Ridge, TN: Oak Ridge National Laboratory. 1987.
- 5.Stabin M, Watson E, Cristy M, Ryman J, Eckerman K, Davis J, Marshall D, Gehlen M. Mathematical models and specific absorbed fractions of photon energy in the nonpregnant adult female and at the end of each trimester of pregnancy. Oak Ridge, TN: Oak Ridge National Laboratory. 1995.
- 6.Siegel J, Stabin M. Absorbed fractions for electrons and beta particles in spheres of various sizes. J Nucl Med. 1994;35(1):152–6. [PubMed] [Google Scholar]
- 7.Stabin M, Konijnenberg M. Re-evaluation of absorbed fractions for photons and electrons in spheres of various sizes. J Nucl Med. 2000;41:149–60. [PubMed] [Google Scholar]
- 8.STL Nuclear Medicine Phantoms. http://nucpc92.ph.man.ac.uk/phantoms/. Updated Feb 01 2016, Accessed Feb 01, 2016.
- 9.Shcherbinin S, Celler A, Belhocine T, Vanderwerf R, Driedger A. Accuracy of quantitative reconstructions in SPECT/CT imaging. Phys Med Biol. 2008;53(17):4595–604. doi: 10.1088/0031-9155/53/17/009. [DOI] [PubMed] [Google Scholar]
- 10.Hutton B, Buvat I, Beekman F. Review and current status of SPECT scatter correction. Phys Med Biol. 2011;56(14):85–112. doi: 10.1088/0031-9155/56/14/R01. [DOI] [PubMed] [Google Scholar]
- 11.Erlandsson K, Buvat I, Pretorius PH, Thomas B. a., Hutton BF. A review of partial volume correction techniques for emission tomography and their applications in neurology, cardiology and oncology. Phys Med Biol. 2012;57(21):119–59. doi: 10.1088/0031-9155/57/21/R119. [DOI] [PubMed] [Google Scholar]
- 12.Brolin G, Gustafsson J, Ljungberg M, Sjögreen Gleisner K. Pharmacokinetic digital phantoms for accuracy assessment of image-based dosimetry in 177 Lu-DOTATATE peptide receptor radionuclide therapy. Phys Med Biol. 2015;60:6131–49. doi: 10.1088/0031-9155/60/15/6131. [DOI] [PubMed] [Google Scholar]
- 13.Sandström M, Tolmachev V, Kairemo K, Lundqvist H, Lubberink M. Performance of coincidence imaging with long-lived positron emitters as an alternative to dedicated PET and SPECT. Phys Med Biol. 2004;49(24):5419. doi: 10.1088/0031-9155/49/24/002. [DOI] [PubMed] [Google Scholar]
- 14.Sandström M, Garske U, Granberg D, Sundin A, Lundqvist H. Individualized dosimetry in patients undergoing therapy with (177)Lu-DOTA-D-Phe (1)-Tyr (3)-octreotate. Eur J Nucl Med. 2010;37(2):212–5. doi: 10.1007/s00259-009-1216-8. [DOI] [PubMed] [Google Scholar]
- 15.Grimes J, Celler A, Birkenfeld B, Shcherbinin S, Listewnik MH, Piwowarska-Bilska H, Mikolajczak R, Zorga P. Patient-specific radiation dosimetry of 99mTc-HYNIC-Tyr3-Octreotide in neuroendocrine tumors. J Nucl Med. 2011;52(9):1474–81. doi: 10.2967/jnumed.111.088203. [DOI] [PubMed] [Google Scholar]
- 16.Gear J, Long C, Rushforth D, Chittenden S, Cummings C, Flux G. Development of patient-specific molecular imaging phantoms using a 3D printer. Med Phys. 2014;41:082501–13. doi: 10.1118/1.4887854. [DOI] [PubMed] [Google Scholar]