. 2016 Apr 15;75(6):2227–2234. doi: 10.1002/mrm.26241

High resolution spectroscopy and chemical shift imaging of hyperpolarized 129Xe dissolved in the human brain in vivo at 1.5 tesla

Madhwesha Rao 1, Neil J Stewart 1, Graham Norquay 1, Paul D Griffiths 1, Jim M Wild 1,
PMCID: PMC4950000  PMID: 27080441

Abstract

Purpose

Upon inhalation, xenon diffuses into the bloodstream and is transported to the brain, where it dissolves in various compartments of the brain. Although up to five chemically distinct peaks have been previously observed in 129Xe rat head spectra, to date only three peaks have been reported in the human head. This study demonstrates high resolution spectroscopy and chemical shift imaging (CSI) of 129Xe dissolved in the human head at 1.5 Tesla.

Methods

A 129Xe radiofrequency coil was built in‐house and 129Xe gas was polarized using spin‐exchange optical pumping. Following the inhalation of 129Xe gas, NMR spectroscopy was performed with spectral resolution of 0.033 ppm. Two‐dimensional CSI in all three anatomical planes was performed with spectral resolution of 2.1 ppm and voxel size 20 mm × 20 mm.

Results

Spectra of hyperpolarized 129Xe dissolved in the human head showed five distinct peaks at 188 ppm, 192 ppm, 196 ppm, 200 ppm, and 217 ppm. Assignment of these peaks was consistent with earlier studies.

Conclusion

High resolution spectroscopy and CSI of hyperpolarized 129Xe dissolved in the human head has been demonstrated. For the first time, five distinct NMR peaks have been observed in 129Xe spectra from the human head in vivo. Magn Reson Med 75:2227–2234, 2016. © 2016 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

Keywords: hyperpolarized gas, xenon spectroscopy, chemical shift imaging, human brain spectroscopy

INTRODUCTION

Hyperpolarized (HP) 129Xe MRI has gained interest ever since the first image was acquired from a biological sample 1. Xenon, when inhaled into the lungs, dissolves into the blood by diffusive gas exchange across the alveolar–capillary membrane 2, 3, 4, 5, 6. The fraction that dissolves in the blood is determined by the xenon blood–gas partition coefficient 7, 8, ratio of ventilated airspace volume to pulmonary blood volume 9, capillary perfusion and alveolar surface‐area‐to‐volume ratio. In the bloodstream, dissolved HP 129Xe is transported to distal organs by the systemic blood circulation. The longitudinal relaxation time T1 of HP 129Xe in the human blood is sufficiently long (6 s to 8 s for arterial‐oxygenated blood and 3 s to 4 s for venous‐deoxygenated blood) for it to be detected in the distal organs 10, 11, 12, 13.

Recently, studies involving NMR spectroscopy and chemical shift imaging (CSI) of HP 129Xe dissolved in the brain have been conducted in rats 14, 15, 16, 17 and in humans 4, 18. Kershaw et al and Nakamura et al 19, 20 observed five peaks in 129Xe spectra from the rat head, which they attributed to jaw muscle (187–191 ppm), white matter (191–194 ppm), gray matter (193–197 ppm), fat tissue outside the brain (197–201 ppm), and red blood cells (210 ppm). The feasibility of imaging rat brain using HP 129Xe CSI was demonstrated by Swanson et al 16. Using HP 129Xe, Mazzanti et al 14 explored the possibility of detecting increased brain perfusion and functional activity following sensory stimulation in rats and Zhou et al 15 demonstrated the detection of ischemic stroke in rats. To our understanding, the only two studies of HP 129Xe in the human head to date are from Mugler et al 4 and Kilian et al 18, 21 who presented time resolved spectra depicting the uptake of HP 129Xe into the human brain in vivo. Subsequently, Kilian et al 18, 21 reported preliminary results of HP 129Xe CSI in the human brain at 3.0 Tesla (T) and estimated T1 relaxation of HP 129Xe dissolved in gray matter and white matter to be 14 s and 8 s, respectively.

Unlike the HP 129Xe studies conducted in rats 19, 20, the preliminary human studies have revealed two dissolved peaks which were attributed to gray matter (196 ppm) and white matter (193 ppm) 18, 21, and one report showed evidence of a third peak at 186 ppm 21. Thus, the motivation of our study at 1.5T was to perform (i) high resolution 129Xe spectroscopy of the human head in vivo and (ii) CSI to spatially resolve the compartments of the brain responsible for the peaks observed.

METHODS

RF Coil and Calibration

For the construction of a custom 129Xe radiofrequency (RF) coil for the human head, a rigid mechanical former was recycled from a disused 1H head birdcage coil. An eight‐leg birdcage coil of band pass topology was constructed with a capacitance of 289.5 pF (270 pF + 18 pF + 1.5 pF) and 1000 pF on the legs and on the end ring, respectively, as shown in Figure 1b. A photograph of the RF birdcage coil is shown in Figure 1a. The end ring diameter was 305 mm and the leg length was 300 mm as shown in Figure 1a. The capacitors used were of 10C package (Dalian Dalicap Technology Co., Ltd, Dalian, China). The birdcage coil was driven in quadrature transmit–receive mode. RF measurements were performed using an Agilent 5061B Network Analyzer (Keysight Technologies, Santa Rosa, CA).

Figure 1.

Figure 1

RF birdcage coil for 129Xe brain MR spectroscopy and chemical shift imaging. a: Photograph of the RF birdcage coil. b: Schematic of the RF birdcage coil. c: Picture of the experimental setup for calculation of flip angle profile. d: 129Xe NMR signal measured after the 5th, 15th, 25th, and 35th RF pulses (black lines) indicating depletion in polarization and the flip angle profile for a 90° RF pulse (red line).

The RF power required to achieve a desired flip angle was calculated based on a standard calibration procedure, whereby the rate of depletion of polarization of HP 129Xe resulting from repeated acquisitions using hard RF pulse within a short time‐interval ( TR T1) was fit to  cosn1α, where n is the number of repeated acquisitions and α is the flip angle for the transmit RF power. To calculate the variation in flip angle across the coil, a plastic tube with 6 mm inner diameter was filled with HP 129Xe and positioned radially within the RF coil as shown in the photograph in Figure 1c, and the coil was loaded with a standard nickel chloride (NiCl2·6H20:H20) phantoms. The length of the tube was 36 cm, which covered the diameter of the RF coil.

Using a fast spoiled gradient echo sequence with a soft (Gaussian) RF pulse, with frequency encoding applied only along one direction (anterior to posterior as shown in Figure 1c), the flip angle profile along the diameter of the RF coil was calculated by measuring the depletion of polarization of the HP 129Xe in each voxel along the length of the tube as shown in Figure 1d. Acquisition parameters were: echo time (TE) = 3.6 ms, repetition time (TR) = 18.9 ms, matrix = 1 (phase) × 44 (frequency), slice thickness = 200 mm, number of slices = 1, field of view (FOV) = 44 cm, RF pulse duration = 800 µs, and the acquisition lasted 800 ms. A total of 39 repeated acquisitions were performed. T1 relaxation and gas diffusion were both neglected for the calculation.

MRI Methods

All in vivo NMR with 129Xe experiments were performed with approval from the UK National Research Ethics Committee under a protocol for the evaluation of novel MRI methods for imaging hyperpolarized xenon in the lungs, brain, and heart of healthy human volunteers. MRI experiments were performed on a GE whole‐body 1.5T Signa® HDx scanner. All in vivo experiments were conducted following inhalation of isotopically enriched xenon (87% 129Xe) from 1 L Tedlar® bags. 129Xe gas was polarized using spin exchange optical pumping to around 25% polarization 22. Chemical shift values were calibrated with respect to the resonance of 129Xe in the gas‐phase measured in a Tedlar® bag positioned at the isocenter of the scanner. In vivo experiments were conducted on three healthy male volunteers with the mean age of 29 years and standard deviation of 3 years.

Wideband Spectroscopy

The xenon gas dose for the acquisition was 1 L. The acquisition was performed under breath‐hold commencing immediately after the inhalation of the gas and lasted for 20 s thereafter. After inhalation, the residual gas in the Tedlar® bag was placed on the neck of the subject to obtain a reference HP 129Xe gas‐phase signal. A total of 20 discrete free induction decays (FIDs) were acquired with a flip angle of 20° and TR of 1 s. The transmit RF pulse was a hard pulse with a pulse duration of 500 µs. The center frequency was set to 197.93 ppm downfield (17,660,800 Hz) from the 129Xe gas‐phase resonance and the receiver bandwidth was 674.16 ppm (11.9 kHz). The spectral resolution was 0.16 ppm (2.9 Hz) with 4096 sampling points in the spectrum. All 20 acquired complex spectra were averaged to improve the signal‐to‐noise ratio (SNR). No spectral line broadening filters were applied.

High Resolution Spectroscopy

For high resolution spectroscopy, data acquisition was performed during a 20 s breath‐hold, commencing 4 s after inhalation of a 1 L dose of xenon gas. Ten discrete FIDs were acquired with a TR of 2 s and a flip angle of 45°. The transmit RF pulse was a hard pulse with a pulse duration of 500 µs. The center frequency was set to 197.93 ppm downfield (17,660,800 Hz) from the 129Xe gas‐phase resonance and the receiver bandwidth was 136.09 ppm (2403 Hz). The spectral resolution was 0.033 ppm (0.58 Hz) with 4096 sampling points in the spectrum. All 10 acquired complex spectra were averaged and no spectral line broadening filters were applied.

Chemical Shift Imaging

To spatially localize the 129Xe spectral peaks anatomically in the human brain, 2D CSI was conducted by applying phase‐encoding in the anterior–posterior and right–left directions for axial‐plane CSI, anterior–posterior and superior–inferior directions for sagittal‐plane CSI, and right–left and superior–inferior directions for coronal‐plane CSI. FIDs were acquired for each phase‐encoding step using a pulse‐acquire sequence without slice selection. The transmit RF pulse was as described earlier. The center frequency was set to 197.93 ppm downfield (17,660,800 Hz) from the 129Xe gas‐phase resonance and the receiver bandwidth was 136.05 ppm (2403 Hz). The spectral resolution was 2.1 ppm (37 Hz) with 64 sampling points per spectrum. The flip angle was 10°, and the TR was 45 ms. For a matrix size of 12 × 12 and a FOV of 24 cm, 144 phase‐encoded spectra were acquired with an effective in‐plane spatial resolution of 20 mm × 20 mm. The acquisition was performed under breath‐hold, commencing 16 s after the inhalation of a 1 L dose of xenon gas and lasted for 6.5 s thereafter. No spectral line broadening filters were applied.

For anatomical reference, 1H MR images were acquired (in each of the three anatomical planes) using an inversion recovery gradient echo pulse sequence in a separate scan with the same FOV as the 129Xe CSI acquisition. The imaging parameters were: TE = 5.2 ms, inversion time = 450 ms, TR = 12.4 ms, flip angle = 20°, bandwidth = 15.6 kHz, slice thickness = 1.6 mm and matrix size = 512 × 512. For the purpose of assignment of a particular resonance to a particular brain compartment, the 129Xe chemical shift images were superimposed on the corresponding anatomical reference 1H images. Each pixel of the HP 129Xe CSI represents the maximum value of a particular chemical shift peak at that spatial location. An image intensity threshold determined by visual inspection was applied to the 2D CSI images of dissolved HP 129Xe to eliminate background noise. The threshold for each of the 2D CSI images were determined and applied individually.

RESULTS

RF Coil and Calibration

The quality factor (Q) of the birdcage RF coil in the unloaded condition was 270 and in the loaded condition was 86. The ratio of the Q factor in unloaded to loaded condition was, therefore, 3.1. The isolation (transfer coefficient,  Sij, (ij)) between the in‐phase and the quadrature‐phase ports of the RF birdcage coil was lower than −15 dB. Figure 1d shows the measured 129Xe NMR signal after the 5th, 15th, 25th, and 35th RF pulse with frequency encoding applied only along one direction (anterior–posterior), indicating depletion of polarization. For a nominal 90° RF pulse, the standard deviation of the flip angle measured across the diameter of the coil was 5.9° (6.56%) and the flip angle profile is shown in Figure 1d.

Spectroscopy

An averaged spectrum of HP 129Xe dissolved in the human head acquired with a wide receiver bandwidth is shown in Figure 2a. The spectrum exhibits two distinct peaks near 0 ppm assigned to 129Xe in the gaseous phase. Near the center frequency of the spectrum, five distinct dissolved 129Xe peaks from the human head can be observed, at 188.4 ppm, 192.7 ppm, 195.6 ppm, 199.6 ppm, and 217.2 ppm.

Figure 2.

Figure 2

In vivo magnitude spectra of HP 129Xe dissolved in the human head. a: Spectrum with wide bandwidth (674.16 ppm) and a spectral resolution of 0.16 ppm. b: Spectrum with a narrow bandwidth (136.09 ppm) and a higher spectral resolution of 0.033 ppm. Note: (a) and (b) represent two separate acquisitions. Subject: male 31 years old.

An averaged whole head spectrum acquired from HP 129Xe dissolved in the human head with high spectral resolution (0.033 ppm) is shown in Figure 2b. The spectrum exhibits five distinct peaks: 187.2 ppm, 192.5 ppm, 195.4 ppm, 199.0 ppm, and 216.2 ppm.

Chemical Shift Imaging

To demonstrate that the SNR of spatially resolved spectra from the CSI experiment of the human head are sufficient, example spectra (at arbitrary locations) for each of the anatomical planes are shown in Figure 3. It can be observed, that the SNR of these spatially resolved spectra are sufficient to enable distinction of the peaks at 188 ppm, 192 ppm, 196 ppm, 200 ppm, and 217 ppm, and that the signal amplitudes are clearly prominent above the noise floor. Spatially resolved spectra in all the three imaging planes superimposed on corresponding anatomical 1H images are shown in Figure 4.

Figure 3.

Figure 3

Individual spectra acquired in a 2D chemical shift imaging experiment at arbitrary locations in the head. a: Axial plane spectrum from the left center of the head, showing peaks at 188 ppm, 194 ppm, and 217 ppm. b: Sagittal plane spectrum from the anterior inferior part of the head, showing peaks at 189 ppm, 196 ppm, and 200 ppm. c: Coronal plane spectrum from the superior center of the head, showing peaks at 192 ppm and 196 ppm. The peaks have been labeled to indicate the compartment of the head with which they are assigned (see the Discussion section). Subject: male 31 years old.

Figure 4.

Figure 4

Spatially resolved 129Xe spectra of the head in all the three imaging planes superimposed on corresponding anatomical 1H images. a: Axial plane. B: Sagittal plane. C: Coronal plane. Subject: male 31 years old.

Chemical shift images corresponding to 129Xe NMR peak from the head at 188 ppm, 192 ppm, 196 ppm, 200 ppm, and 217 ppm, superimposed (in color) on corresponding anatomical 1H images are shown in Figures 5a–c,e,f. Figure 5d shows the spectrum from Figure 2b split into its constituent real and imaginary parts after zeroth order phase‐correction using the 195.4 ppm (gray matter) peak as a reference. The full width half maximum of the peak at 195.4 ppm (gray matter) in Figure 5d is 36 Hz.

Figure 5.

Figure 5

The 2D chemical shift images of spatially resolved peaks from 129Xe in the human head, superimposed (in color) on 1H reference images (gray scale) in all the three anatomical planes. a: 188 ppm ‐ tissue in the cheek muscle and the midbrain/brainstem. b: 192 ppm ‐ white matter and cartilaginous soft‐tissue. c: 196 ppm ‐ gray matter. e: 200 ppm ‐ body interstitial fluid/plasma, fat tissue outside the brain and cerebrospinal fluid. f: 217 ppm ‐ red blood cells. The threshold for each of the 2D CSI images were determined and applied individually. Color bars represent arbitrary units normalized to the maximum NMR signal intensity of hyperpolarized 129Xe dissolved in the gray matter. d: Phase‐corrected real and imaginary parts of the spectrum from Figure 2b, showing HP 129Xe dissolved in the human head in vivo with a spectral resolution of 0.033 ppm and a bandwidth of 136.09 ppm. Subject: male 31 years old.

DISCUSSION

In Figure 2a, the peak at −0.7 ppm can be attributed to residual gas in the Tedlar® bag placed on the subject's neck. The 129Xe gas‐phase resonance being shifted by −0.7 ppm (instead of 0 ppm) suggests B0 variation of 1 μT due to the Tedlar® bag not being positioned at the isocenter of the magnet combined with bulk susceptibility shifts related to the subject's head. Similarly, we postulate that the peak at −2.8 ppm is related to detection of 129Xe gas in the smaller airways of the lungs in the stray field of the RF coil. The up‐field (negative) chemical shift of this 129Xe gas in the smaller airways of the lungs could be due to a combination of effects, including; bulk magnetic susceptibility shift effects 5, the chemical shift associated with mixing with oxygen 23 and the fact that while an average adult head is positioned at the isocenter, the lungs lie further away from the isocenter at a lower static magnetic field strength than the isocenter and would experience a lower Larmor frequency.

The chemical shift values reported in Figure 3 are marginally different from the values reported in Figure 2, most likely due to the considerable difference in spectral resolution between the two acquisitions. In addition, it can be observed that some of the five peaks (from Figure 2) are absent in some of the spectra in Figure 3. This indicates that the particular biochemical compartment giving rise to a specific chemical shift value is not present at that particular voxel location.

From the anatomical location of the 129Xe NMR peaks in the chemical shift images superimposed on corresponding 1H images in Figure 5, we propose the assignment of each individual peak to a particular compartment of the head as follows:

188 ppm

From Figure 5a, the peak at 188 ppm appears to be most intense in the anterior region of the axial plane image. The sagittal plane image indicates that most of the signal from this peak arises from regions below the eyes, above the jawline and between the nose and ears. In the coronal plane image, it is evident that the signal intensity is approximately symmetric on both right and left sides of the head. In addition, this image shows that some contribution arises from a location at the midbrain. We allocate the peak at 188 ppm to HP 129Xe dissolved in soft muscular tissue in cheek and midbrain.

192 ppm

From Figure 5b, it can be observed that the peak at 192 ppm appears to predominantly arise from 129Xe dissolved in the brain. In the axial plane, the peak at 192 ppm is most intense in the brain with some evidence of tracts. Similarly, in the coronal plane image, the peak at 192 ppm appears to be most intense in the central brain, again with some evidence of tracts. In the sagittal plane image, a contribution to the 192 ppm peak can be observed from 129Xe in the region of the nasal cavity. Thus, it can be postulated that the peak at 192 ppm corresponds to HP 129Xe dissolved in white matter (Commissural and Projection tracts) and cartilaginous soft‐tissue.

196 ppm

Figure 5c shows that the 129Xe peak at 196 ppm is predominantly localized to the brain; it can be clearly seen in images from the all three anatomical planes that the peak not only emanates from the brain, but does not appear anywhere else. Thus it can be concluded that the peak at 196 ppm corresponds to HP 129Xe dissolved in the gray matter.

Unlike the case of xenon gas in an isolated bag, in the human brain in vivo, 129Xe will undergo diffusive chemical exchange with other compartments of the brain that come in to contact with each other (such as cerebral blood, interstitial fluid, and cerebrospinal fluid). Existence of any such chemical exchange will broaden the line‐width of the NMR peaks of the compartments under exchange. Nevertheless, under these conditions the full width half maximum of a particular spectral peak will provide a lower limit estimate of the  T2* relaxation time from a given voxel. The full width half maximum of 36 Hz for the gray matter peak in Figure 5d corresponds to  T2* 8.8 ms.

200 ppm

From Figure 5e, the peak at 200 ppm appears to originate from 129Xe located in the anterior region of the head, as seen from the axial plane image. In the sagittal plane image, most of the signal appears to be from regions among the eyes, nose, and ears, and extends below toward the jaw. In the coronal plane image, the signal appears to arise from the center of the head in the right–left direction. In all the three anatomical planes, the signal appears to coincide with the location of the ventricles. Thus, the peak at 200 ppm is likely from HP 129Xe dissolved in body interstitial fluid/plasma, fat tissue outside the brain, and cerebrospinal fluid.

217 ppm

From Figure 5f, in all three anatomical planes, the only location where the peak at 217 ppm shows considerable signal intensity is at the location of the Circle of Willis. This peak at 217 ppm can, therefore, be allocated to HP 129Xe dissolved in the red blood cells. It is worth noting that this peak is broader than the others, most probably due to the diffusive chemical exchange between 129Xe dissolved in various compartments, and to the fact that blood oxygenation will vary across the head, resulting in a dispersion of the chemical shift due to difference in the oxygenation state of the hemoglobin 24.

The assignment of the chemical shift values with the particular biochemical environments as described above is supported by earlier studies, which include in vivo reports of 129Xe in the rat and human head and in vitro studies of 129Xe dissolved in human blood. These earlier studies suggested the following peak assignments: 187.2 ppm, muscle 19, 20, 21; 192.7 ppm, white matter (and soft‐tissue in this study) 18, 19, 20, 21; 195.6 ppm, gray matter 18, 19, 20, 21; 199.6 ppm, aqueous solution (cerebrospinal fluid, plasma, and interstitial fluid in this study) 10, 11, 24, 25, 26 and fat/lipid tissue outside the brain 19, 20; and 217.2 ppm, red blood cells 4, 10, 24, 26.

The peaks seen in whole brain spectroscopy and their anatomical distribution from CSI, were found to be consistent between the subjects as shown in Figure 6. The SNR of the obtained spectra was found to be somewhat variable between the subjects. As SNR is limited by the quantity and polarization of 129Xe delivered to the brain by the systemic blood circulation, any variability in the polarization of the SEOP process, gas dose inhaled or  T1 losses related to different lung‐to‐brain transit times will affect the observed SNR. Assessment of these factors, their variability and their effect on measured spectroscopic SNR of dissolved 129Xe in the brain is the subject of future work.

Figure 6.

Figure 6

Chemical shift images of the spectral peak at 195.5 ppm assigned to 129Xe dissolved in gray matter: male subject 25 years old (a) and male subject 30 years old (c). Spectra (real part) of HP 129Xe dissolved in the human brain in vivo acquired with same acquisition parameter as high resolution spectrum: male subject 25 years old (b) and male subject 30 years old (d).

On closer inspection of Figure 2b, it can be seen that the peak at 187.2 ppm appears to be comprised of two closely spaced peaks at 186.7 ppm and 187.7 ppm which requires further investigation. In addition, a peak at 190.6 ppm can be observed in both Figure 2a inset and 2b which also requires further investigation in future work.

CONCLUSIONS

This is the first time that five distinct 129Xe NMR peaks have been observed in the human head in vivo and we believe this to be the first report of 129Xe CSI in the human head in vivo at 1.5T. Although earlier studies evidenced the presence of 5 129Xe NMR peaks in the rat head 19, 20, only two or three peaks have been previously detected from 129Xe in the human head from spectra of significantly lower SNR and spectral resolution than those in this study 4, 18. Our results should provide a useful reference for future 129Xe NMR spectroscopy studies of the human brain. Furthermore, the methods established here may be directly applied to quantify the dynamics of xenon uptake into the human brain and probe the potential pathophysiological changes induced by functional abnormalities.

ACKNOWLEDGMENTS

This report is an independent research supported by the NIHR Research Professorship. The views expressed in this publication are those of the author(s) and not necessarily those of the NHS, the NIHR or the Department of Health. The authors thank Felix Horn, Helen McQuillan, Guilhem Collier, and Helen Marshall assistance with some of the experiments.

Correction added after online publication 2 May 2016. The author renumbered their references and updated the term “cerebrum” to “brain” under subsections 192 ppm and 196 ppm in the Discussion. The authors also removed the term “brainstem” from “midbrain/brainstem” under Discussion subsection 188 ppm, “white matter” under subsection 196 ppm, and the phrase “(internal carotid artery and middle cerebral artery)” under subsection 217 ppm.

REFERENCES

  • 1. Albert MS, Cates GD, Driehuys B, Happer W, Saam B, Springer CS Jr, Wishnia A. Biological magnetic resonance imaging using laser‐polarized 129Xe. Nature 1994;370:199–201. [DOI] [PubMed] [Google Scholar]
  • 2. Fain SB, Korosec FR, Holmes JH, O'Halloran R, Sorkness RL, Grist TM. Functional lung imaging using hyperpolarized gas MRI. J Magn Reson Imaging 2007;25:910–923. [DOI] [PubMed] [Google Scholar]
  • 3. Möller HE, Chen XJ, Saam B, Hagspiel KD, Johnson GA, Altes TA, De Lange EE, Kauczor HU. MRI of the lungs using hyperpolarized noble gases. Magn Reson Med 2002;47:1029–1051. [DOI] [PubMed] [Google Scholar]
  • 4. Mugler JP III, Driehuys B, Brookeman JR, et al. MR imaging and spectroscopy using hyperpolarized 129Xe gas: preliminary human results. Magn Reson Med 1997;37:809–815. [DOI] [PubMed] [Google Scholar]
  • 5. Wagshul ME, Button TM, Li HFF, Liang ZR, Springer CS, Zhong K, Wishnia A. In vivo MR imaging and spectroscopy using hyperpolarized Xe‐129. Magn Reson Med 1996;36:183–191. [DOI] [PubMed] [Google Scholar]
  • 6. Ruppert K, Mata JF, Brookeman JR, Hagspiel KD, Mugler JP. Exploring lung function with hyperpolarized Xe‐129 nuclear magnetic resonance. Magn Reson Med 2004;51:676–687. [DOI] [PubMed] [Google Scholar]
  • 7. Goto T, Suwa K, Uezono S, Ichinose F, Uchiyama M, Morita S. The blood‐gas partition coefficient of xenon may be lower than generally accepted. Br J Anaesth 1998;80:255–256. [DOI] [PubMed] [Google Scholar]
  • 8. Kety SS. The theory and applications of the exchange of inert gas at the lungs and tissues. Pharmacol Rev 1951;3:1–41. [PubMed] [Google Scholar]
  • 9. West JB. Respiratory physiology: the essentials. Philadelphia: Lippincott Williams & Wilkins; 2012. [Google Scholar]
  • 10. Norquay G, Leung G, Stewart NJ, Tozer GM, Wolber J, Wild JM. Relaxation and exchange dynamics of hyperpolarized (129) Xe in human blood. Magn Reson Med 2015;74:303–311. [DOI] [PubMed] [Google Scholar]
  • 11. Wolber J, Cherubini A, Dzik‐Jurasz AS, Leach MO, Bifone A. Spin‐lattice relaxation of laser‐polarized xenon in human blood. Proc Natl Acad Sci U S A 1999;96:3664–3669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Albert MS, Kacher DF, Balamore D, Venkatesh AK, Jolesz FA. T(1) of (129)Xe in blood and the role of oxygenation. J Magn Reson 1999;140:264–273. [DOI] [PubMed] [Google Scholar]
  • 13. Albert MS, Balamore D, Kacher DF, Venkatesh AK, Jolesz FA. Hyperpolarized 129Xe T1 in oxygenated and deoxygenated blood. NMR Biomed 2000;13:407–414. [DOI] [PubMed] [Google Scholar]
  • 14. Mazzanti ML, Walvick RP, Zhou X, Sun Y, Shah N, Mansour J, Gereige J, Albert MS. Distribution of hyperpolarized xenon in the brain following sensory stimulation: preliminary MRI findings. PloS One 2011;6:e21607. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Zhou X, Sun Y, Mazzanti M, Henninger N, Mansour J, Fisher M, Albert M. MRI of stroke using hyperpolarized 129Xe. NMR Biomed 2011;24:170–175. [DOI] [PubMed] [Google Scholar]
  • 16. Swanson SD, Rosen MS, Agranoff BW, Coulter KP, Welsh RC, Chupp TE. Brain MRI with laser‐polarized Xe‐129. Magn Reson Med 1997;38:695–698. [DOI] [PubMed] [Google Scholar]
  • 17. Swanson SD, Rosen MS, Coulter KP, Welsh RC, Chupp TE. Distribution and dynamics of laser‐polarized Xe‐129 magnetization in vivo. Magn Reson Med 1999;42:1137–1145. [DOI] [PubMed] [Google Scholar]
  • 18. Kilian W, Seifert F, Rinneberg H. Dynamic NMR Spectroscopy of Hyperpolarized 129Xe in Human Brain Analyzed by an Uptake Model. Magn Reson Med 2004;51:843–847. [DOI] [PubMed] [Google Scholar]
  • 19. Kershaw J, Nakamura K, Kondoh Y, Wakai A, Suzuki N, Kanno I. Confirming the existence of five peaks in 129Xe rat head spectra. Magn Reson Med 2007;57:791–797. [DOI] [PubMed] [Google Scholar]
  • 20. Nakamura K, Kondoh Y, Wakai A, Kershaw J, Wright D, Kanno I. 129Xe spectra from the heads of rats with and without ligation of the external carotid and pterygopalatine arteries. Magn Reson Med 2005;53:528–534. [DOI] [PubMed] [Google Scholar]
  • 21. Kilian W, Seifert F. 2D Chemical Shift Imaging of hyperpolarized isotopically enriched 129Xe within human brain. In Proceedings of the 13th Annul Meeting of ISMRM, Miami Beach, Florida, USA, 2005. Abstract 1163.
  • 22. Norquay G, Parnell SR, Xu X, Parra‐Robles J, Wild JM. Optimized production of hyperpolarized 129Xe at 2 bars for in vivo lung magnetic resonance imaging. J Appl Phys 2013;113. [Google Scholar]
  • 23. CJ Jameson, JA Keith. Density dependence of 129Xe N.M.R. chemical shifts in O2 and NO. Mol Phys 1971;20(5):957–959. [Google Scholar]
  • 24. Wolber J, Cherubini A, Leach MO, Bifone A. Hyperpolarized 129Xe NMR as a probe for blood oxygenation. Magn Reson Med 2000;43:491–496. [DOI] [PubMed] [Google Scholar]
  • 25. Miller KW, Reo NV, Schoot Uiterkamp AJ, Stengle DP, Stengle TR, Williamson KL. Xenon NMR: chemical shifts of a general anesthetic in common solvents, proteins, and membranes. Proc Natl Acad Sci U S A 1981;78:4946–4949. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Bifone A, Song Y‐Q, Seydoux R, Taylor RE, Goodson BM, Pietrass T, Budinger TF, Navon G, Pines A. NMR of laser‐polarized xenon in human blood. Proc Natl Acad Sci U S A 1996;93:12932–12936. [DOI] [PMC free article] [PubMed] [Google Scholar]