. 2016 Jul 18;9(7):973–978. doi: 10.18240/ijo.2016.07.06

Comparison of corneal biomechanical properties in normal tension glaucoma patients with different visual field progression speed

Ying Hong 1, Nobuyuki Shoji 2,3, Tetsuya Morita 4, Kazunori Hirasawa 2, Kazuhiro Matsumura 4, Masayuki Kasahara 3, Kimiya Shimizu 4
PMCID: PMC4951676  PMID: 27500103

Abstract

AIM

To compare the corneal biomechanical properties difference by ocular response analyzer (ORA) in normal tension glaucoma (NTG) patients with different visual field (VF) progression speed.

METHODS

NTG patients with well-controlled Goldmann applanation tonometer (GAT) who routinely consulted Kitasato University Hospital Glaucoma Department between January 2010 and February 2014 were enrolled. GAT and ORA parameters including corneal compensated intraocular pressure (IOPcc), Goldmann estimated intraocular pressure (IOPg), corneal hysteresis (CH), corneal resistance factor (CRF) were recorded. VF was tested by Swedish interactive threshold algorithm (SITA)-standard 30-2 fields. All patients underwent VF measurement regularly and GAT did not exceed 15 mm Hg at any time during the 3y follow up. Patients were divided into four groups according to VF change over 3y, and ORA findings were compared between the upper 25th percentile group (slow progression group) and the lower 25th percentile group (rapid progression group).

RESULTS

Eighty-two eyes of 56 patients were studied. There were 21 eyes (21 patients) each in rapid and slow progression groups respectively. GAT, IOPcc, IOPg, CH, CRF were 12.1±1.4 mm Hg, 15.8±1.8 mm Hg, 12.8±2.0 mm Hg, 8.4±1.1 mm Hg, 7.9±1.3 mm Hg respectively in rapid progression group and 11.5±1.3 mm Hg, 13.5±2.1 mm Hg, 11.2±1.6 mm Hg, 9.3±1.1 mm Hg, 8.2±0.9 mm Hg respectively in slow progression group (P=0.214, <0.001, 0.007, 0.017, 0.413, respectively). In bivariate correlation analysis, IOPcc, IOPcc-GAT and CH were significant correlated with mΔMD (r=-0.292, -0.312, 0.228 respectively, P=0.008, 0.004, 0.039 respectively).

CONCLUSION

Relatively rapid VF progression occurred in NTG patients whose IOPcc are rather high, CH are rather low and the difference between IOPcc and GAT are relatively large. Higher IOPcc and lower CH are associated with VF progression in NTG patients. This study suggests that GAT measures might underestimate the IOP in such patients.

Keywords: ocular response analyzer, intraocular pressure, corneal biochemical properties, visual field, normal tension glaucoma

INTRODUCTION

The prevalence of primary open angle glaucoma (POAG) in Japan is 3.9% with 92% of POAG patients whose intraocular pressure (IOP) did not exceed 21 mm Hg, which is defined as normal-tension glaucoma (NTG)[1]. In some patients with well-controlled IOP, the visual field (VF) remains stable or progresses very slowly, while in others the condition is quite different: IOP is well controlled but VF progresses rapidly[2]. There have been a number of studies focusing on the pressure-independent pathway of glaucoma[3]–[7], but the accuracy of IOP measurement may play an important role in such cases.

Goldmann applanation tonometer (GAT) is currently the gold standard for IOP measurement, but these values (GAT-IOP) may be affected by central corneal thickness (CCT), corneal curvature, corneal astigmatism and other corneal biomechanical properties[8]–[11]. Therefore, it is very important to find a new method to determine the true IOP.

Ocular response analyzer (ORA) (Reichert©; Reichert Technologies, Buffalo, NY, USA) is a new device that is described as a non-contact tonometer. ORA can measure corneal hysteresis (CH), corneal resistance factor (CRF) and determine a specific corneal compensated intraocular pressure (IOPcc) which is less influenced by corneal viscoelasticity. There have been studies discussing the relationship between ORA measurements and structural or functional changes in glaucoma patients, but most of them have focused on patients with POAG or suspected glaucoma[2],[12]–[15].

There are limited data on ORA measurements in NTG patients. Moreover, the relationship between corneal viscoelasticity, corneal thickness and VF progression in NTG patients remains unclear. In the current study, we obtained ORA parameters including IOPcc, Goldmann estimated intraocular pressure (IOPg), GAT, CH and CRF in NTG patients with well-controlled GAT-IOP. We compare ORA data in patients with different VF progression speed to see relatively rapid VF progression might occur in what kind of NTG patients.

SUBJECTS AND METHODS

This retrospective study was approved by the Kitasato University Hospital Review Board and followed the tenets of the Declaration of Helsinki. Informed consent was obtained from all patients included in this study. Patients routinely consulted Kitasato University Hospital glaucoma department between January 2010 and February 2014 and were usually seen at intervals of 3-6mo.

The inclusion criteria were as follows: NTG was defined by the presence of glaucomatous optic neuropathy[16] associated with reproducible VF abnormalities. GAT-IOP did not exceed 15 mm Hg including diurnal variation during the recent 3y, because NTG patients whose IOP exceed 16 mm Hg in the daytime are likely to have an IOP that exceeds 21 mm Hg at night[17].

The ocular examination included visual acuity, corneal parameters such as corneal curvature and corneal refractive power, GAT, ORA parameters including IOPcc, IOPg, CH, CRF and CCT. VF was examined by SITA-Standard 30-2 fields (Carl Zeiss Meditec, Inc., Dublin, CA, USA). The parameters of VF using for the current study were the mean deviation (MD) and the pattern standard deviation (PSD). The minimum criterion for a VF abnormality was a glaucoma hemifield test outside normal limits or a PSD result <5% on 2 consecutive reliable examinations. All enrolled patients underwent at least 5 VF tests to analyze VF progression statistically. We analyzed VF change by global trend analysis by MD slope. The MD slope, MD change per year (dB/y) was obtained from linear regression analysis of the HFA II glaucoma progression analysis software. MD change was calculated by MD slope multiply 3 because the follow-up period was 3y. And the PSD change was the difference between the VF results at last follow-up and the baseline ones.

All eyes had best corrected visual acuity ≥20/30. Refractive error of the patients was between -8.0 diopters and 8.0 diopters spherical equivalent and corneal astigmatism between -3.0 diopters and 3.0 diopters.

We excluded patients with an insufficient number of VF results because we could not analyze whether VF had progressed in such cases. Patients with other diseases that may affect the VF test and/or ORA test were also excluded. Patients who had undergone any type of intraocular surgery within the past 3mo before the participation in this study were also excluded[13]. Patients underwent VF test first and then IOP measurements were obtained in a random sequence in order to minimize the potential for a statistical effect of applanation on lowering IOP.

Corneal biomechanical properties were measured with ORA once at the patient's study visit. All patients were tested by one experienced doctor. The device obtains 2 measurements of the corneal response to the air pulse. The major outcomes are CH, CRF and IOPcc (mm Hg). The difference between the 2 pressures is CH (mm Hg). CRF is thought to be one of the indices of corneal elasticity based on CH, IOPg is the average of P1 and P2. IOPcc is a pressure measurement based on CH, which is thought little to be affected by corneal biomechanical properties. ORA can also provide CCT results[9],[11],[18]. A good quality reading was defined as one with symmetrical height of force-in and force-out waveform peaks and a waveform score >7 on a software-generated scale of 0 to 10[18]–[19].

Statistical Analysis

Data were analyzed by statistical software (SPSS version 20.0 SPSS Inc., Chicago, IL, USA). Categorical data were compared by χ2 tests. Continuous variables were tested for normal distribution. Variables with normal distribution are presented as mean±SD and were compared by independent Student's t-test. Variables with skewed distribution are presented as median with interquartile range (IR) and were compared by Mann-Whitney U test. Bivariate correlation analysis was constructed to determine variables associated with VF damage. A two-tailed P<0.05 was considered significant.

RESULTS

There were 142 NTG patients with good follow-up during the study period, but only 82 eyes of 56 patients met the inclusion criterion of GAT-IOP (GATmax) not exceeding 15 mm Hg during the 3y. Therefore, 82 eyes of 56 patients were enrolled in this study. These 26 males and 30 females had an average age of 62.6y (range from 37-84y). One eye of 30 patients and both eyes of 26 patients were included. All patients were Asian. The included eyes underwent a median of 7.1 (range from 5-12) VF tests during follow-up. There were 16 eyes (19.5%) without any kind of antiglaucoma eye drops, 43 eyes (52.4%) receiving one kind of eye drops, 17 eyes (20.7%) receiving 2 kinds of eye drops and 6 eyes (7.3%) receiving 3 kinds of eye drops. Details are showed in Table 1.

Table 1. Antiglaucoma eye drops of the patients.

Treatment Number (eyes) Percentage (%)
None 16 19.5
PG 31 37.8
β-Blocker 11 13.4
CAI 1 1.2
PG+β-Blocker 11 13.4
PG+CAI 6 7.3
PG+β-Blocker+CAI 6 7.3

PG: Prostaglandin; CAI: Carbonic anhydrase inhibitor.

Patient general information, their ORA parameters and VF change over the 3y are listed in Table 2. In brief, the GAT on the day ORA obtained was 12.0 mm Hg, as the same as the median GAT (GATavg, 12 mm Hg) over the 3y. Both of such measurements were approximately 3 mm Hg lower than IOPcc (15.0 mm Hg). The average CH was 8.9 mm Hg and CRF was 8.1 mm Hg, which were both below normal limits[18],[20]. The median MD change (mΔMD) was -0.8 dB over the 3y, indicating approximately 0.3 dB VF loss per year.

Table 2. General and clinical characteristics of the patients.

Characteristics x±s (median) Range
Age 62.6±11.8 37-84
Gender (M/F) 26/30
Eyedrops 1.2±0.8 0-3
Corneal curvature 7.73±0.23 7.27-8.49
Corneal refractive power 43.7±1.3 39.75-46.50
GAT (mm Hg) 12.0±1.5 9-15
GATmax 14 (14, 15) 10-15
GATavg 12 (12, 13) 9-14
IOPcc (mm Hg) 15.0±2.5 9.3-20.1
IOPg (mm Hg) 12.3±2.2 7.4-17.7
IOPcc-IOPg (mm Hg) 2.6±1.4 -0.9-5.6
IOPcc-GAT (mm Hg) 2.9±2.2 -2.0-9.1
CRF (mm Hg) 8.1±1.2 5.6-11.5
CH (mm Hg) 8.9±1.3 6.1-11.7
CCT (µm) 520.0±26.8 474-580
MD (dB) -5.2 (-9.6, -2.5) -27.86-1.91
PSD (dB) 9.3±4.4 1.66-18.34
Baseline MD (dB) -4.9 (-7.8, -1.3) -28.15-1.48
Baseline PSD (dB) 8.5 (3.7, 12.5) 1.72-18.31
ΔMD (dB) -0.8 (-1.8, -0.1) -5.58-1.58
ΔPSD (dB) 0.9 (-0.2, 1.8) -2.75-6.67

GAT: Goldmann applanation tonometer; IOP: Intraocular pressure; IOPcc: Corneal compensated intraocular pressure; IOPg: Goldmann estimated intraocular pressure; CRF: Corneal resistance factor; CH: Corneal hysteresis; CCT: Central corneal thickness; MD: Mean deviation; PSD: Pattern standard deviation. Data are skewed distribution and presented as median (IR).

Patients were divided into four groups according to the mΔMD over 3y (Figure 1). The median, P25 and P75 value of the mΔMD in the four groups are -2.1 (-3.2, -1.9) dB; -1.1 (-1.3, -0.9) dB; -0.5 (-0.6, -0.3) dB and 0.4 (0.1, 0.8) dB respectively. The data-box plots of CH, CRF, IOPcc and IOPg for 4 groups were shown in Figure 2. And findings were compared between the upper 25th percentile group (slow progression group, 21 eyes in total) and the lower 25th percentile group (rapid progression group, 21 eyes in total). Patient age, gender and the numbers of antiglaucoma eyedrops did not significantly differ between the two groups (P=0.484, 1.000, 0.396 respectively). There were no statistically significant differences of the parameters that may affect ORA results such as corneal curvature or corneal refractive power between the two groups (P=0.106, 0.101, respectively). And there were no significant differences in GAT, GATmax or GATavg between two groups (P=0.142, 0.890, 0.966, respectively).

Figure 1. Delta MD of all patients over 3y.

Figure 1

Figure 2. The data-box plots of CH, CRF, IOPcc and IOPg for 4 groups.

Figure 2

ORA results showed that the average IOPcc in the rapid progression group was 15.8 mm Hg, significantly higher than that in the slow group (13.5 mm Hg) (P<0.001). The average IOPg in the rapid group was 12.8 mm Hg, significantly higher than that in the slow group (11.2 mm Hg) (P=0.006). The difference between IOPcc and IOPg was 3.0 mm Hg in the rapid group, significantly higher than 2.3 mm Hg in the slow group (P=0.035). The difference between IOPcc and GAT was 3.8 mm Hg in the rapid group, significantly higher than 1.9 mm Hg in the slow group (P=0.004).

The average CH in the rapid group was 8.4 mm Hg, significantly lower than 9.3 mm Hg in the slow group (P=0.009). The average CRF in the rapid group was 7.9 mm Hg, which was lower than 8.2 mm Hg in the slow group, but the difference did not reach significance (P=0.310). There was no significant difference in CCT between the two groups (P=0.849).

At baseline, MD did not significantly differ between the rapid and slow group (P=0.134) but PSD significantly differ between the two groups (P=0.037). After 3y, there was a significant difference. MD and PSD were -8.8 dB and 10.4 dB in the rapid group, but -2.1 dB and 6.4 dB in the slow group (P=0.001, 0.002, respectively). Therefore, the VF change was also significantly different between the two groups, mΔMD and mΔPSD were -2.1 dB and 1.9 dB in the rapid group, but 0.4 dB and -1 dB in the slow group (P<0.001, 0.036 respectively) (Table 3).

Table 3. The comparison of the worst and best 25th percentile group.

Characteristics Rapid (n=21) Slow (n=21) P
Age 60.9±12.2 62.2±12.9 0.484
Gender (M/F) 10/11 11/10 1.000
Eyedrops 1.4±0.9 1.2±0.9 0.396
Corneal curvature 7.73±0.23 7.65±0.16 0.106
Corneal refractive power 43.67±1.29 44.13±0.91 0.101
GAT (mm Hg) 12.1±1.4 11.5±1.3 0.142
GATmax 14 (14, 14.5) 14 (13.5, 14) 0.890
GATavg 12 (12, 12) 12 (11.5, 12.5) 0.966
IOPcc (mm Hg) 15.8±1.8 13.5±2.1 <0.001
IOPg (mm Hg) 12.8±2.0 11.2±1.6 0.006
IOPcc-IOPg (mm Hg) 3.0±1.3 2.3±1.1 0.035
IOPcc-GAT (mm Hg) 3.8±2.1 1.9±1.8 0.004
CRF (mm Hg) 7.9±1.3 8.2±0.9 0.310
CH (mm Hg) 8.4±1.1 9.3±1.1 0.009
CCT (µm) 516.1±28.2 515.3±25.0 0.849
MD (dB) -8.8 (-11.4, -3.4) -2.1 (-5.5, 0.1) 0.001
PSD (dB) 10.4±4.6 6.4±3.7 0.002
Baseline MD (dB) -5.4 (-9.3, -1.1) -2.9 (-5.7, -0.6) 0.134
Baseline PSD (dB) 8.9±5.2 6.1±3.8 0.037
ΔMD (dB) -2.1 (-3.2, -1.9) 0.4 (0.1, 0.8) <0.001
ΔPSD (dB) 1.9 (0.3, 2.6) -1 (-0.7, 1.4) 0.036

Data are skewed distribution and presented as median (IR).

In bivariate correlation analysis, the Pearson correlation coefficient of CH, CRF etc. were listed in Table 4. Briefly, IOPcc, IOPcc-GAT and CH were significant correlated with mΔMD (r=-0.292, -0.312, 0.228 respectively and P=0.008, 0.004 and 0.039 respectively).

Table 4. The Pearson correlation coefficient of bivariate correlation analysis.

Characteristics r P
Age -0.044 0.693
Corneal curvature -0.058 0.606
Corneal refractive power 0.056 0.619
GAT (mm Hg) -0.020 0.860
GATmax 0.006 0.960
GATavg 0.109 0.330
IOPcc (mm Hg) -0.292 0.008
IOPg (mm Hg) -0.206 0.063
IOPcc-IOPg (mm Hg) -0.191 0.085
IOPcc-GAT (mm Hg) -0.312 0.004
CRF (mm Hg) 0.106 0.344
CH (mm Hg) 0.228 0.039
CCT (µm) -0.042 0.709

DISCUSSION

The current study investigated whether there was difference of corneal biomechanical properties in NTG patients with different VF progression speed. The results demonstrated that in NTG patients whose GAT-IOP did not exceed the mid-teens during the recent 3y, relatively rapid VF progression occurred in patients with rather high IOPcc, rather low CH and relatively large difference between IOPcc and GAT. IOPcc and CH were significantly correlated with VF progression. These findings indicated that IOPcc and CH might be associated with VF progression in NTG patients.

Recently, there have been some studies focusing on corneal biomechanical properties, the accuracy of IOP measurement and VF progression. The study by Congdon et al[13] in 2006 included 230 subjects (POAG, suspected POAG and ocular hypertension) and showed that neither CCT nor CH was related to VF progression. However, other studies suggested that there was relationship between corneal biomechanical properties and VF progression[2],[12],[14]–[15]. Anand et al[12] compared CH and VF asymmetry in open angle glaucoma. Their findings demonstrated that CH, CRF and IOPcc were risk factors for worse VF. Mansouri et al[14] compared corneal biomechanical properties and VF between glaucoma patients and suspected glaucoma patients and found worse CH, CRF and CCT values in the glaucoma group. De Moraes et al[2] categorized different types of glaucoma patients based on whether VF progressed and concluded that CH and CCT are associated with VF progression. Most recently, a prospective longitudinal study by Medeiros et al[15] claimed that baseline CH and baseline GAT were associated with the risk of glaucoma progression. A summary of recent research is presented in Table 5. Our study differed from the previous studies are all of the patients included in the current study were Asian and NTG patients. A group comprised of a single race might demonstrate fewer anatomic differences than a group comprised of different races[21].

Table 5. Summary of recent similar researches.

Authors Eye/Patient Diagnosis MD (dB) PSD (dB) GAT (mm Hg) CH (mm Hg) CRF IOPcc (mm Hg) Associated with worse VF
Congdon et al[13] N/A/230 POAG; POAG suspect;OH N/A N/A N/A N/A N/A N/A Neither nor CH
Anand et al[12] 234/117 POAG with asymmetric VF Worse eye Worse eye 14 Worse eye Worse eye Worse eye CH; CRF; IOPcc
-11.2±6.4 9.6±3.2 8.2±1.9 8.6±2.0 17.4
Better eye Better eye Better eye Better eye Better eye
-2.1±2.5 3.2±2.3 8.9±1.9 8.8±2.1 16.9
Mansouri et al[14] 299/191 Glaucoma; glaucoma suspect Glaucoma Glaucoma Glaucoma Glaucoma Glaucoma Glaucoma CH; CRF
-3.3±3.3 4.0±3.0 15.0±5.6 9.4±1.7 9.4±2.0 16.6±5.4
Suspect -0.38±1.6 Suspect 1.6±0.9 Suspect 16.6±4.5 (IOPg) Suspect 10.4±1.7 Suspect 10.7±2.1 Suspect 6.9±4.1
De Moraes et al[2] 153/153 POAG; NTG; XFG; ACG; JOAG; PG Progress Progress Progress Progress Progress Progress CH
-5.3±4.1 4.7±3.0 15.3±3.7 7.5±1.4 7.6±1.3 18.0±5.3
Non-prog Non-prog Non-prog Non-prog Non-prog Non-prog
-6.5±6.8 5.4±4.3 14.7±3.9 9.0±1.8 8.9±2.0 16.5±5.0
Medeiros et al[15] 114/68 POAG Baseline -2.45±3.22 Baseline 3.32±2.84 Baseline 16.1±3.8 Baseline 9.5±1.7 N/A N/A Baseline; CH; baseline GAT

The findings of the current study concurred with recent articles suggesting that corneal biomechanical properties were associated with VF change[2],[12],[14]–[15]. This difference between IOPcc and GAT was similar to what was found in other studies[2],[12]. The VF loss of approximately 0.3 dB per year which did not reach the progression standard of 1 dB per year[22], indicating that the patients included in the current study were relatively well controlled. These results of the comparison of rapid progression group and slow progression group indicate that patients showing rapid progression had rather high IOPcc, rather low CH which along with relatively large difference between IOPcc and GAT. This finding suggests that the IOP values obtained of such patients during follow up were underestimated. In such cases, the optic nerve might be chronically exposed to relatively high IOP resulting in obvious progression of VF.

There are lots of experimental and clinical evidence that the biomechanical properties of the eyeball may be related to those of the optic nerve complex[23]–[27]. Scleral stiffness and collagen fiber organization influence IOP-induced deformation of the optic nerve head in a computer model[27]. Downs et al[24] reported a change in the viscoelastic properties of peripapillary sclera on exposure to chronic IOP elevations in monkey eyes with glaucoma. Another study reported that monkey eyes with stiff or thick sclera seemed to be less prone to biomechanical changes in response to chronic IOP elevation[26]. Another experimental study found an association between higher CH and greater optic nerve deformation when IOP was artificially elevated in glaucoma eyes[25]. We think the biomechanical properties of the eyeball in NTG patients may also be related to those of the optic nerve complex. So it may explain why the VF progress rapidly of NTG patients with rather high IOPcc and rather low CH. But it remains unclear whether there is a causal relationship between CH and VF progression or not[2],[12]–[15]. It may be that the corneal biomechanical properties change first, then compliance of the eyeball to IOP decreases and pressure on the optic nerve head increases, finally causing retina nerve fiber layer defects (RNFLD) and glaucomatous VF change. Another possibility is that lower CH presents as a result of chronic IOP elevation, similar to optic disc cupping and RNFLD. A third possibility is that these are simultaneous but independent changes. Further research is needed to clarify the nature of the association.

As far as was concerned, unlike ocular hypertension treatment study[28], we did not find was associated with VF progression in our study. This may be because the patients in their study had hypertension, while our study investigated NTG. And other studies did not find any relationship between and VF progression too[12]–[13],[15],[25],[29].

There are several limitations in current study. First, it is a small sample study. It is because the inclusion criteria were very strict. Although this choice reduced the number of patients, it increased the homogeneity and reduced the influence of other confounding factors. Second, both eyes of some patients were included in this study, but only one eye of each patient was compared in the two groups. Third, because of the retrospective nature of the study, the baseline corneal biomechanical properties of the patients were not available.

The current study demonstrated that relatively rapid VF progression occurred in NTG patients with rather high IOPcc, rather low CH and relatively large difference between IOPcc and GAT. These findings indicated that IOPcc and CH were associated with VF progression in NTG patients.

Since treatment to decrease IOP is the only therapy confirmed by evidence-based medicine for controlling the progression of VF in NTG patients, the “target” IOP should take corneal biomechanical properties into consideration. IOPcc is significantly higher than GAT in those who appear to progress faster. So IOPcc may be a better method of monitoring IOP in NTG patients and patients with low CH should undergo more thorough investigation and careful monitoring.

Acknowledgments

Conflicts of Interest: Hong Y, None; Shoji N, None; Morita T, None; Hirasawa K, None; Matsumura K, None; Kasahara M, None; Shimizu K, None.

REFERENCES

  • 1.Iwase A, Suzuki Y, Araie M, Yamamoto T, Abe H, Shirato S, Kuwayama Y, Mishima HK, Shimizu H, Tomita G, Inoue Y, Kitazawa Y, Tajimi Study Group, Japan Glaucoma Society The prevalence of primary open-angle glaucoma in Japanese: the Tajimi Study. Ophthalmology. 2004;111(9):1641–1648. doi: 10.1016/j.ophtha.2004.03.029. [DOI] [PubMed] [Google Scholar]
  • 2.De Moraes CV, Hill V, Tello C, Liebmann JM, Ritch R. Lower corneal hysteresis is associated with more rapid glaucomatous visual field progression. J Glaucoma. 2012;21(4):209–213. doi: 10.1097/IJG.0b013e3182071b92. [DOI] [PubMed] [Google Scholar]
  • 3.Bell K, Gramlich OW, Von Thun Und Hohenstein-Blaul N, Beck S, Funke S, Wilding C, Pfeiffer N, Grus FH. Does autoimmunity play a part in the pathogenesis of glaucoma? Prog Retin Eye Res. 2013;36:199–216. doi: 10.1016/j.preteyeres.2013.02.003. [DOI] [PubMed] [Google Scholar]
  • 4.Joachim SC, Reinehr S, Kuehn S, Laspas P, Gramlich OW, Kuehn M, Tischoff I, VON Pein HD, Dick HB, Grus FH. Immune response against ocular tissues after immunization with optic nerve antigens in a model of autoimmune glaucoma. Mol Vis. 2013;19:1804–1814. [PMC free article] [PubMed] [Google Scholar]
  • 5.Krizaj D, Ryskamp DA, Tian N, Tezel G, Mitchell CH, Slepak VZ, Shestopalov VI. From mechanosensitivity to inflammatory responses: new players in the pathology of glaucoma. Curr Eye Res. 2014;39(2):105–119. doi: 10.3109/02713683.2013.836541. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Shazly TA, Aljajeh M, Latina MA. Autoimmune basis of glaucoma. Semin Ophthalmol. 2011;26(4-5):278–281. doi: 10.3109/08820538.2011.588649. [DOI] [PubMed] [Google Scholar]
  • 7.Huang W, Fan Q, Wang W, Zhou M, Laties AM, Zhang X. Collagen: a potential factor involved in the pathogenesis of glaucoma. Med Sci Monit Basic Res. 2013;19:237–240. doi: 10.12659/MSMBR.889061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Kniestedt C, Lin S, Choe J, Bostrom A, Nee M, Stamper RL. Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol. 2005;123(11):1532–1537. doi: 10.1001/archopht.123.11.1532. [DOI] [PubMed] [Google Scholar]
  • 9.Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma. 2006;15(5):364–370. doi: 10.1097/01.ijg.0000212268.42606.97. [DOI] [PubMed] [Google Scholar]
  • 10.Morita T, Shoji N, Kamiya K, Fujimura F, Shimizu K. Corneal biomechanical properties in normal-tension glaucoma. Acta Ophthalmol. 2012;90(1):e48–e53. doi: 10.1111/j.1755-3768.2011.02242.x. [DOI] [PubMed] [Google Scholar]
  • 11.Morita T, Shoji N, Kamiya K, Hagishima M, Fujimura F, Shimizu K. Intraocular pressure measured by dynamic contour tonometer and ocular response analyzer in normal tension glaucoma. Graefes Arch Clin Exp Ophthalmol. 2010;248(1):73–77. doi: 10.1007/s00417-009-1169-4. [DOI] [PubMed] [Google Scholar]
  • 12.Anand A, De Moraes CG, Teng CC, Tello C, Liebmann JM, Ritch R. Corneal hysteresis and visual field asymmetry in open angle glaucoma. Invest Ophthalmol Vis Sci. 2010;51(12):6514–6518. doi: 10.1167/iovs.10-5580. [DOI] [PubMed] [Google Scholar]
  • 13.Congdon NG, Broman AT, Bandeen-Roche K, Grover D, Quigley HA. Central corneal thickness and corneal hysteresis associated with glaucoma damage. Am J Ophthalmol. 2006;141(5):868–875. doi: 10.1016/j.ajo.2005.12.007. [DOI] [PubMed] [Google Scholar]
  • 14.Mansouri K, Leite MT, Weinreb RN, Tafreshi A, Zangwill LM, Medeiros FA. Association between corneal biomechanical properties and glaucoma severity. Am J Ophthalmol. 2012;153(3):419–427. doi: 10.1016/j.ajo.2011.08.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Medeiros FA, Meira-Freitas D, Lisboa R, Kuang TM, Zangwill LM, Weinreb RN. Corneal hysteresis as a risk factor for glaucoma progression: a prospective longitudinal study. Ophthalmology. 2013;120(8):1533–1540. doi: 10.1016/j.ophtha.2013.01.032. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Foster PJ, Buhrmann R, Quigley HA, Johnson GJ. The definition and classification of glaucoma in prevalence surveys. Br J Ophthalmol. 2002;86(2):238–242. doi: 10.1136/bjo.86.2.238. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Aihara M, Yamagami J, Araie M, Yamamoto S. Relationship of the office intraocular pressure (IOP) to diurnal fluctuation of IOP in low-tension glaucoma: a multivariate analysis. Nippon Ganka Gakkai Zasshi. 1992;96(8):1007–1013. [PubMed] [Google Scholar]
  • 18.Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156–162. doi: 10.1016/j.jcrs.2004.10.044. [DOI] [PubMed] [Google Scholar]
  • 19.Lam AK, Chen D, Tse J. The usefulness of waveform score from the ocular response analyzer. Optom Vis Sci. 2010;87(3):195–199. doi: 10.1097/OPX.0b013e3181d1d940. [DOI] [PubMed] [Google Scholar]
  • 20.Laiquzzaman M, Bhojwani R, Cunliffe I, Shah S. Diurnal variation of ocular hysteresis in normal subjects: relevance in clinical context. Clin Experiment Ophthalmol. 2006;34(2):114–118. doi: 10.1111/j.1442-9071.2006.01185.x. [DOI] [PubMed] [Google Scholar]
  • 21.Leite MT, Alencar LM, Gore C, Weinreb RN, Sample PA, Zangwill LM, Medeiros FA. Comparison of corneal biomechanical properties between healthy blacks and whites using the Ocular Response Analyzer. Am J Ophthalmol. 2010;150(2):163–168. doi: 10.1016/j.ajo.2010.02.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Smith SD, Katz J, Quigley HA. Analysis of progressive change in automated visual fields in glaucoma. Invest Ophthalmol Vis Sci. 1996;37(7):1419–1428. [PubMed] [Google Scholar]
  • 23.Bochmann F, Ang GS, Azuara-Blanco A. Lower corneal hysteresis in glaucoma patients with acquired pit of the optic nerve (APON) Graefes Arch Clin Exp Ophthalmol. 2008;246(5):735–738. doi: 10.1007/s00417-007-0756-5. [DOI] [PubMed] [Google Scholar]
  • 24.Downs JC, Suh JK, Thomas KA, Bellezza AJ, Hart RT, Burgoyne CF. Viscoelastic material properties of the peripapillary sclera in normal and early-glaucoma monkey eyes. Invest Ophthalmol Vis Sci. 2005;46(2):540–546. doi: 10.1167/iovs.04-0114. [DOI] [PubMed] [Google Scholar]
  • 25.Wells AP, Garway-Heath DF, Poostchi A, Wong T, Chan KC, Sachdev N. Corneal hysteresis but not corneal thickness correlates with optic nerve surface compliance in glaucoma patients. Invest Ophthalmol Vis Sci. 2008;49(8):3262–3268. doi: 10.1167/iovs.07-1556. [DOI] [PubMed] [Google Scholar]
  • 26.Girard MJ, Suh JK, Bottlang M, Burgoyne CF, Downs JC. Biomechanical changes in the sclera of monkey eyes exposed to chronic IOP elevations. Invest Ophthalmol Vis Sci. 2011;52(8):5656–5669. doi: 10.1167/iovs.10-6927. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Sigal IA, Yang H, Roberts MD, Burgoyne CF, Downs JC. IOP-induced lamina cribrosa displacement and scleral canal expansion: an analysis of factor interactions using parameterized eye-specific models. Invest Ophthalmol Vis Sci. 2011;52(3):1896–1907. doi: 10.1167/iovs.10-5500. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, Parrish RK, 2nd, Wilson MP, Kass MA. The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002;120(6):714–720. doi: 10.1001/archopht.120.6.714. [DOI] [PubMed] [Google Scholar]
  • 29.Nemesure B, Wu SY, Hennis A, Leske MC, Barbados Eye Study Group Corneal thickness and intraocular pressure in the Barbados eye studies. Arch Ophthalmol. 2003;121(2):240–244. doi: 10.1001/archopht.121.2.240. [DOI] [PubMed] [Google Scholar]

Articles from International Journal of Ophthalmology are provided here courtesy of Press of International Journal of Ophthalmology