. 2015 Mar 6;68(4):1063–1080. doi: 10.1007/s10616-015-9862-0

Table 4.

LOD (limit of detection) evaluation of Vero cell line infected with mycoplasma (positive control cell line)

Vero cell line mycoplasma positive Reading A Reading B B/A Test results
Undiluted positive control Vero cell line 9/514 267/860 28/154 Positive > 1
Dilution 1/2 13/673 132/044 9/657 Positive > 1
Dilution 1/4 17/407 95/361 5/478 Positive > 1
Dilution 1/8 11/005 74//011 6/725 Positive > 1
Dilution 1/16 7/262 48/519 6/681 Positive > 1
Dilution 1/32 5/025 15/477 3/080 Positive > 1
Dilution 1/64 5/555 14/522 2/614 Positive > 1
Dilution 1/128 3/660 7/194 1/956 Positive > 1
Dilution 1/256 3/119 4/375 1/402 Positive > 1
Dilution 1/512 3/390 2/685 0/792 Negative < 1
Dilution 1/1024 4/686 2/121 0/452 Negative < 1
Dilution 1/2048 2/279 0/931 0/408 Negative < 1
Dilution 1/4096 1/892 0/682 0/360 Negative < 1
Control positive kit 6/051 216/435 35/768 Positive > 1
Control negative kit 3/075 0/459 0/149 Negative < 1
Vero cell line mycoplasma free (negative) 2/235 0/585 0/261 Negative < 1

Different serial dilutions in the range of 1/2 to 1/4,096 were prepared