Abstract
This article provides data and a method related to a research paper entitled “Assessing vitamin D nutritional status: is capillary blood adequate?” (Jensen et al., 2016) [1]. Circulating 25OHD, the accepted biomarker of the vitamin D nutritional status, is routinely measured by automated immunoassays, that although may be performed in hospital central laboratories, often suffer from a lack of specificity with regards to the different vitamin D metabolites, “Measurement of circulating 25-hydroxyvitamin D: a historical review” (Le Goff et al., 2015) [2]. Mass spectrometry offers this specificity. This article describes the performance of an in-house tandem mass spectrometry method for the individual measurement of 25OHD3, 25OHD2 and 3-épi-25OHD3.
Keywords: Vitamin D, 25-hydroxyvitamin D, Mass spectrometry
Specifications Table
| Subject area | Laboratory medicine |
| More specific subject area | Clinical Chemistry |
| Type of data | Tables, figures |
| How data was acquired | Mass spectrometry, Agilent 6460 triple quadrupole mass spectrometer equipped with a JetStream™ interface coupled to |
| Data format | Mass spectral analysis |
| Experimental factors | Charcoal-stripped serum served as a blank. Deuterated [25OHD3 (26,26,26,27,27,27-d6, IS1), 25OHD2 (26,26,26,27,27,27-d6, IS2) and 3-epi-25OHD3 (6,19,19-d3, IS3) served as internal standards for each vitamin D metabolite quantitation. |
| Experimental features | The mass spectral analysis was performed using a MassHunter workstation software, version B.04.00 (Agilent Technologies Canada Inc., Mississauga, ON, Canada). |
| Data source location | Montreal, Québec, Canada |
| Data accessibility | The data is available with this article |
Value of the data
-
•
The data describes a tandem mass spectrometry method for the measurement of serum 25OHD.
-
•
The details given enable other researchers to reproduce this method.
-
•
These data will be useful for implementing tandem mass spectrometric methods for the quantification of vitamin D metabolites in future clinical studies.
1. Data
Data Capillary blood has been shown to be an adequate matrix for measuring 25OHD [1[. Automated immunoassays routinely used by hospital central laboratories suffer from a lack of specificity that mass spectrometry offers [2].
The data shared include the description of the extraction, chromatography and mass spectrometry protocols as well sample mass spectra obtained from standards and extracted serum. The validation procedure of the method and results are also described.
2. Experimental design, materials and methods
100 μL of a blank, consisting of charcoal-stripped plasma (cat. #1131-00) purchased from Biocell (Rancho Dominguez, CA, USA), sample, calibrator or control were transferred to glass tubes, and spiked with 250 μL of a mixture of deuterated internal standards 25OHD3 (26,26,26,27,27,27-d6, IS1), 25OHD2 (26,26,26,27,27,27-d6, IS2) from Chemaphor Inc., (Ottawa, ON, Can) and 3-epi-25OHD3 (6,19,19-d3, IS3) from Sigma-Aldrich Canada (Oakville, ON, Can) mixed gently for 10 s, incubated for 1 h at room temperature with periodic short mixing, before the addition of 1.0 mL of 2-methoxy-2-methylpropane/hexane (50/50 v/v). After a further 10-min incubation at room temperature, the mixture was centrifuged at 3000 RPM for 5 min. 800 μL of the supernatant was transferred to a 12/75-glass tube, evaporated to dryness at 37 °C under a stream of N2. The crude extract was reconstituted in 100 μL of the initial HPLC mobile phase (see below) and placed on the auto-sampler.
The detailed chromatography and optimized+ion mode ESI-MS/MS conditions and ionic transition masses used are described in Table 1. HPLC/ESI MS-MS (high performance liquid chromatography/electrospray ionization tandem mass spectrometry) was performed on an Agilent 1200 HPLC system, consisting of degassers, binary and quaternary pumps, an auto-sampler equipped with a micro Rheodyne valve, and a temperature controlled column compartment with a column-switching valve coupled to an Agilent 6460 triple quadrupole mass spectrometer equipped with a JetStream™ interface (Agilent Technologies Canada Inc., Mississauga, ON, Canada). For the chromatography, 10 μl of the extracts were injected on a jacket-heated (50 °C) column [100×2.1 mm Kinetex™ 2.6 µm PentaFluoroPhenyl Core shell Silica 100 Å (Phenomenex, Torrance, CA, USA)] preceded by a 2.1 mm internal diameter (ID) SecurityGuard™ ULTRA cartridge (Phenomenex). The MassHunter workstation software, version B.04.00 (Agilent Technologies) was used for the management of the HPLC/ESI MS-MS system and data acquisition. The spectral analysis was done in the Multiple Reaction Mode (MRM). Fig. 1 illustrates representative selected ion LC-MS/MS chromatographic profiles of spiked charcoal-stripped plasma spiked with vitamin D standards and deuterated internal standards, and a patient serum extract analyzed in the conditions described above.
Table 1.
Analytical data.
| Chromatography | ||||
| The mobile phases were: (A) H2O/methanol (50/50 v/v)+0.1% formic acid, and (B) methanol+0.1% formic acid. The needle rinse solvent was acetonitrile/methanol/2-propanol (50/25/25). The binary pump flow rate was set at 0.4 mL/min with 44% B for the first 6 min (flow was switched away from cartridge at 3 min for a backwash step with the quaternary pump) followed by 100% B for a 1 min wash step before returning to 44% B for at least a 2 min post-run. The quaternary pump flow rate was set at 0.4 mL/min with 100% B from 0 to 6 min (the flow was directed to waste for the first 3 min and in a back-flush mode for the cartridge in the last 3 min) followed by 44% B for 1 min and 100% B for at least a 2 min post-run. | ||||
| Mass spectrometer settings | ||||
| The optimized+ion mode ESI-MS/MS conditions were as follows: gas temperature 275 °C, gas flow 5 L/min, sheath gas heater 325 °C, sheath gas flow 11 L/min: nebulizer 45 psi, and capillary voltage 5000 V. Nitrogen was used as desolvation and collision gas. The multiplier voltage was set at 0 V except between 2.9 and 6 min where it was set at 635 V, the interval during which the flow is directed to the nebulizer and not diverted to waste. | ||||
| Vitamin D metabolite | Transition ion m/z | Dwell time (ms) | Fragmentor (V) | CE (V) |
|---|---|---|---|---|
| 25OHD3 | 401.3→383.2 | 100 | 106 | 35 |
| [6d2]-25OHD3 | 407.3→389.3 | 100 | 106 | 35 |
| 25OHD2 | 413.3→395.3 | 100 | 100 | 32 |
| [6d2]-25OHD2 | 419.3→401.3 | 100 | 106 | 32 |
| 3-epi-25OHD3 | 401.3→365.2 | 100 | 106 | 35 |
| [3d2]-3-epi-25OHD3 | 404.6→386.2 | 100 | 106 | 35 |
Fig. 1.
Representative selected ion LC-MS/MS chromatograms of charcoal-stripped blank plasma (lower solid line) and charcoal-stripped blank plasma spiked with vitamin D metabolites (dashed lines), and a patient plasma (upper solid line) using a PentaFluoroPhenyl Core shell Silica column. Panels A, C and E: Arrows indicate 25OHD3, 3-epi-25OHD3, 25OHD2 respectively. All plasma samples were spiked with deuterated internal standards (ion chromatograms shown in panels B, D and F). Deuterated and non-deuterated compounds had same retention times.
2.1. Analytical performance
The method validation was performed with modified CLSI Guidelines [3], [4]. Briefly, the Lower Limit of quantification (LLoQ) was estimated by the serial dilution of the standard solution (n=5 per dilution) and was defined as the concentration at which precision was ≤20%. Linearity was evaluated by serially diluting a pool of high 25OHD3 concentration samples with charcoal-stripped serum to generate 8 samples of intermediate concentrations that were measured in duplicate.
The MassCheck® controls for the 3 analytes (Chromsystems Instruments and Chemicals GmbH, Gräfelfing, Germany) traceable to NIST 972a standard reference material, served for precision and bias assessment. For within-assay imprecision, 5 sequential injections of each concentration of the MassCheck® controls were performed. Between-assay imprecision was assessed by analyzing 1 reference sample at 2 different concentrations in each batch over 14 months.
Linear responses were observed up to 312 nmol/L for 25OHD3, and 25OHD2 and 128 nmol/L for 3-epi-25OHD3. Table 2 summarizes the performance characteristics of the method. The LLOQ for 25OHD3, 25OHD2 and 3-epi-25OHD3 were 7.0 nmol/L, 5.0 nmol/L and 4.0 nmol/L, respectively. The intra- and inter-assay imprecisions were respectively ≤2.5% and ≤6.0% for all metabolite measured. The mean bias for 25OHD3 was 1.4% and 4.8% at 37.9 and 87.5 nmol/L respectively, within the limits set by the Vitamin D Standardization Program [5].
Table 2.
Method performance.
| Vitamin D metabolite | LLoQ nmol/L | Intra-assay CV% (nmol/L)a | Inter-assay CV% (nmol/L)b | Bias % (nmol/L)b |
|---|---|---|---|---|
| 25OHD3 | 7.0 | 3.34 (40.6) | 5.3 (38.4) | 1.4 (37.9) |
| 0.79 (95.2) | 4.3 (91.7) | 4.8 (87.5) | ||
| 25OHD2 | 5.0 | 4.02 (40.4) | 5.9 (46.8) | 11.2 (42.1) |
| 1.09 (93.6) | 3.8 (113.1) | 4.7 (108.0) | ||
| 3-epi-25OHD3 | 4.0 | 2.07 (36.2) | 6.0 (39.3) | 14.9 (34.2) |
| 2.32 (61.0) | 3.7 (67.5) | 23.8 (54.5) |
LLoQ: Lower limit of quantification.
5 sequential injections of each concentration of the MassCheck® controls were performed. Values in parentheses are the means.
1 aliquot of each concentration of the MassCheck® controls was analyzed over a period of 14 consecutive months [n=90]. For the inter-assay repeatability, values in parentheses are those obtained by our method. For the bias, values in parentheses are those assigned by the manufacturer.
Acknowledgments
The authors acknowledge the contribution of Samuel Marcotte, Medical Technologist, for the analysis of vitamin D metabolites.
Footnotes
Supplementary data associated with this article can be found in the online version at doi:10.1016/j.dib.2016.07.017.
Transparency document. Supplementary material
Supplementary material
References
- 1.Jensen M.E., Ducharme F.M., Théorêt Y., Bélanger A.-S., Delvin E. Assessing vitamin D nutritional status: is capillary blood adequate? Clin. Chim. Acta. 2016;457:59–62. doi: 10.1016/j.cca.2016.03.020. [DOI] [PubMed] [Google Scholar]
- 2.Le Goff C., Cavalier E., Souberbielle J.-C., Gonzáles-Antuña, Delvin E. Measurement of circulating 25-hydroxyvitamin D: a historical review. Pract. Lab. Med. 2015;2:1–14. doi: 10.1016/j.plabm.2015.04.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.CLSI, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline, Second Edition, CLSI document EP17-A2, Clinical Laboratory Standards Institute: Wayne, PA, 2012
- 4.CLSI, Evaluation of the Linearity of Quantitative Measurement Procedures, A Statistical Approach, Approved Guideline, CLSI document EP6-A, Clinical Laboratory Standards Institute: Wayne, PA, 2003.
- 5.Binkley N., Sempos C.T. for the Vitamin D Standardization Program (VDSP). Standardizing vitamin D assays: the way forward. J. Bone Miner. Res. 2014;29:1709–1714. doi: 10.1002/jbmr.2252. [DOI] [PMC free article] [PubMed] [Google Scholar]