Abstract
Direct visualization and assessment of the iridocorneal angle (ICA) region with high resolution is important for the clinical evaluation of glaucoma. However, the current clinical imaging systems for ICA do not provide sufficient structural details due to their poor resolution. The key challenges in achieving high quality ICA imaging are its location in the anterior region of the eye and the occurrence of total internal reflection due to refractive index difference between cornea and air. Here, we report an indirect axicon assisted gonioscopy imaging probe with white light illumination. The illustrated results with this probe shows significantly improved visualization of structures in the ICA including TM region, compared to the current available tools. It could reveal critical details of ICA and expected to aid management by providing information that is complementary to angle photography and gonioscopy.
The trabecular meshwork (TM) in the iridocorneal angle (ICA) of the eye, plays a vital role in aqueous outflow. Aqueous humor flows continuously through the TM, at a rate of 2.75 μl/min approximately, to exit the anterior chamber of eye1. Abnormalities in TM can alter the aqueous humor flow rate, thereby increasing the intraocular pressure (IOP) which could subsequently lead to glaucoma. These abnormalities include marked loss of TM cells, fusion and thickening of the trabecular lamellae and deposition of sheath-derived plaques in comparison to age-matched normas. For this reason, direct high quality visualization of the TM is of potential diagnostic value.
There has been a lot of rapid advancement in the last decade in ophthalmology in the area of ICA imaging2,3,4,5. However, gonioscopic angle examination and current clinical imaging modalities such as EyeCam6,7, optical coherence tomography (OCT)8,9,10, Scheimpflug imaging (Pentacam-Scheimpflug)11, ultrasound biomicroscopy (UBM)12,13,14 and Orbscan scanning slit topography15 do not have the spatial resolution sufficient enough to resolve TM structures16. OCT is an effective tool to analyze quantitative parameters of the ICA. However, the OCT is not extensively used by clinicians as it is expensive tool, difficult to interpret and due to the fact that it only provides a single cross-sectional view of the eye. Pentacam using Scheimpflug principle allows noncontact quantification of the anterior chamber parameters11. The anterior curvature and posterior surface of the cornea can be analyzed by Orbscan scanning-slit topography15. However, direct visualization of ICA is not possible using either the Pentacam or Orbscan scanning slit topography17. UBM can be used for direct angle visualization, however, it is a contact procedure, expensive and has not been established as a common clinical procedure2.
The EyeCamTM is a portable hand held photographic device used to image retina and the ICA. EyeCam has shown results comparable to gonioscopy6. EyeCam being a contact procedure and requiring supine positioning for imaging, could not find its use in routine clinical practice. Till date, gonioscopy remain the gold standard for assessing ICA in a clinical setting. It is popular among clinicians due to its low cost and the ability to indent the eye to visualize dynamic changes. However, neither gonioscopy nor EyeCam has sufficient imaging resolution to visualize TM region with high clarity.
There are also different indirect imaging systems developed with integration of gonioscopy with other techniques. In a notable study by Masihzadeh et al., multiphoton gonioscopy was utilized to image TM structures16,18. The advantage of multiphoton gonioscopic imaging system lies in its ability to perform two photon autofluorescence and second harmonic generation to image TM region through gonioscope with better resolution and without using any fluorescence stains. However, multiphoton sources (lasers) are very expensive and are not used in routine clinical procedures. In 2015, McNabb et al. demonstrated complete 360° circumferential gonioscopic optical coherence tomography imaging of the iridocorneal angle8. However, high resolution structural information of the TM region was not possible by this method.
In this article, we report the first demonstration of an integrated gonioscopy concept named as axicon assisted gonioscopy. The proposed concept and method uses gonioscopy imaging approach by integrating Bessel Beam Microscopy (BBM) concept19,20. The proposed concept is explained through optical simulation which is followed by experimental demonstration using the developed system. The imaged ICA region of porcine eyes indicates the efficiency of the system and its clinical significance. A Comparison of ICA images taken by the newly developed system with gonioscopy and EyeCam, demonstrate improved resolution in documenting the TM, pigmentation and ciliary body base.
Results
Optical design and simulation of gonioscopic system
Figure 1 shows the optical setup of the proposed simple indirect white light gonioscopic imaging system (the schematic diagram is provided in the Supplementary Fig. S4). Here, the indirect gonioscopy imaging system is designed using a conventional gonioscopic contact lens (Hoskins-Barkan Goniotomy lens, adult 11 mm lens, Ocular instruments, WA, United States of America), as the imaging head. For optical simulation, a 3D model of gonioscopic lens is made using Zemax software by using the design specifications of Hoskins-Barkan Goniotomy lens. Figure 2 shows the 3D model of gonioscopic lens and eye. In Fig. 2, only the chief ray is illustrated. The parameters for the eye model and ray tracing of the imaging system are provided in the Supplementary material.
The results of 3D Huygen point spread function (PSF) and the PSF grid for gonioscopy imaging system integrated with and without axicon lens unit are shown in Figs 3 and 4. We performed the simulations with two point sources separated by 3 μm. For the gonioscopy imaging system integrated with axicon lens unit, improvement in resolution is evident from Huygen PSF, which showed a distinct separation of the 3 μm point sources. However, it is also evident from Fig. 3 that when axicon is not used there is clear overlap of the two point sources. The difference between the gonioscopy imaging system with and without the axicon lens unit was further compared with result of PSF grid simulation, as shown in Fig. 4.
For theoretical modelling, we used a plano convex lens (F = 25 mm) in combination with a gonio lens as the imaging head. However, for performing the experiments, we utilized LSM03 (4.6X, VIS OCT scan lens, Thorlabs, USA) instead of plano convex lens. Therefore, we further performed an experimental comparison of both systems by imaging a USAF chart through gonio lens. The result is shown in Fig. 4c. The USAF resolution chart 1951, group 7 element 4 (2.76 μm) was resolved for gonioscopy system integrated with the axicon lens unit. On the other hand, the gonioscopic system without the axicon lens unit only resolved group 6 element 2 (16 μm) (Fig. 4c). Thus, it is clearly evident from these results that the performance of gonioscopy system in terms of spatial resolution is improved with the integration of axicon lens unit.
ICA image of demo eye
Functionality of the imaging system was first tested on an ocular imaging eye model (OEM1-7, Ocular Instruments Inc., Bellevue, WA, USA). Gonioscopic lens was placed onto the eye model interfacing with coupling gel (Vidisic gel). The result is shown in Fig. 5a. Figure 5b shows the image of the demo eye model.
TM region of porcine eye visualized by the developed probe
The ICA of the porcine eyes imaged with our system is shown in Fig. 6. Figure 6a is the acquired raw image. Raw image has structures slightly false colored since the light source used to illuminate has yellow tint and not perfectly white. The processed image is shown in Fig. 6b. It is evident that the different regions in the ICA of the porcine eye are distinctly visible in the image. Close to the iris, there is a simple organization of ciliary body base circumferentially distinct in a discontinuous bundle like pattern. In this zone, fibers are longer and reach the sclera and the scleral spur. It is also evident that the ciliary body is highly pigmented. This observation is corroborated by the earlier studies which indicated higher pigmentations in porcine ciliary body as compared to ciliary body of human eyes21.
The TM region is very distinct in the image (Fig. 6b). However, TM meshwork structures are further highlighted by performing edge detection on the image. This result is shown in Fig. 6c. Pigmentation in the angular region is also clearly evident in Fig. 6b. It is important to note that the imaging was performed on enucleated porcine eyes, and we therefore assume the image quality will be even better for live animal due to less or no corneal haze.
Comparison with other direct ICA imaging methods
Figure 7 shows the comparison of ICA of Porcine eyes imaged using the newly developed system with the state of the art imaging systems such as EyeCam and Gonioscopy. The amount of clouding (haze) of the cornea is known to affect the image quality drastically. Due to this reason, imaging was performed on clear and mildly cloudy porcine eyes with all the imaging systems, for comparative purposes. It is evident from the Fig. 7 that, irrespective of the different imaging systems used, the structures of the ICA imaged for clear eyes were more distinct than cloudy ones. The details of the EyeCam images were better than gonioscopic images. Wide field camera of the EyeCam also allows to get a larger view of iridocorneal angle region as compared to gonioscopy images (Fig. 7). However, it is clear ICA imaged with our system is much better than EyeCam and gonioscopy. Notably, even for cloudy eyes our system showed better image clarity (Fig. 7j). Moreover, fine structural details of pigmentation, TM, ciliary body can be visualized by our system.
Further, we also compared the results with an integrated flexible handheld wide angle probe which was recently developed in our research centre for imaging and documenting ICA17 (Fig. 7g,h). Though the probe is flexible and can image large area using an endoscopic CCD camera as compared to other imaging systems, the resolution is poor compared to the proposed new axicon assisted gonioscopy. All the above mentioned results confirms the improved quality of ICA imaging which can be achieved by our system as compared to the state of the art systems as well as those reported in literatures.
Illustration of pigmentation in the ICA with clarity
Assessment of pigmentation in the TM is of major significance to glaucoma diagnosis22,23,24. (Porcine eyes are generally known to have higher pigmentation than human eyes25.) The pigmentation pattern of an individual angle is dynamic over time. Heavy pigmentation of the TM should indicate pigment dispersion or exfoliation syndrome. In addition, a line of pigment deposition anterior to the Schwalbe line is often present in exfoliation syndrome (Sampaolesi line). Moreover, studies have also revealed that pigmentation can affect or introduce challenges for glaucoma treatment such as laser trabeculoplasty26. Other conditions that cause increased anterior chamber angle pigmentation include malignant melanoma, trauma, surgery, inflammation, angle closure, and hyphema.
However, there are only few relevant studies which statistically analyzed the amount of pigmentation in the TM region27,28,29. Currently available imaging systems can neither provide new insight nor quantify the pigment level of TM27. For this reason, we have demonstrated the potential of our system by illustrating better quality porcine eye images with different pattern of pigmentation. The result is shown in Fig. 8. It is evident from Fig. 8, that the pigmentation have better contrast. We believe the high contrast imaging of ICA region would be useful to perform further statistical studies to get quantitative information regarding pigmentation.
Discussion
In vision sciences research, porcine eyes were often used as an ex vivo animal model because of similarity in its morphology to the human eye30,31. In addition, porcine eyes are also generally used as animal model eyes for validation in a variety of studies such as R & D on glaucoma32, in neuroretinal studies33, cataract surgery34, corneal transplant studies35 and in aberrometry studies36. In our study, we have used the porcine eye as the test eye samples because of its resemblance to the human eye and due to easy availability.
ICA imaging systems must have resolution in the range of 1 to 5 μm16 to resolve TM region. However, the existing clinical white light gonioscopic systems do not have this above mentioned spatial resolution to resolve the structural details of TM. This key drawback limits their efficiency to visualize critical details of ICA.
In summary, we have demonstrated for the first time to the best of our knowledge, an axicon assisted white light indirect gonioscopic system which can image the ICA with spatial resolution down to 3 μm. This imaging system is thus capable of direct imaging of ICA with better structural clarity as compared to current imaging system such as gonioscopy, OCT and EyeCam. Our gonioscopic system may pave the way for future ICA imaging.
Methods
Preparation of Sample
Enucleated porcine eyes were obtained from abattoir. To avoid corneal haze which would reduce image quality, porcine eyes were imaged within 5 to 6 hours of death. The ex vivo samples were transported on ice (approximately −4 °C) to the laboratory to maintain their freshness. Before imaging, the extraocular tissues such as conjunctiva and lacrimal gland were removed from the samples. Each sample was fixed on to a custom eye holder, which was mounted on a translational stage with micrometer accuracy. A sample size of 50 porcine eyes were used for the study. The experiment was conducted in accordance to the regulations of Agri-Food & Veterinary Authority of Singapore and Nanyang Technological University’s regulations on biosafety.
Gonioscopy
Hoskins-Barkan Goniotomy lens (adult 11 mm lens, Ocular instruments, WA, USA) was used in this study. The thickness of the gonioscopic lens is 1 cm. The gonioscopic lens was fixed to an optical post and secured to optical table. A thin layer of ophthalmic gel (Vidisic Gel, Bausch & Lomb, NY, USA) was applied to the gonioscopic lens before placing it on the eye, to provide lubrication and to reduce the refractive index mismatch. The porcine eye (with cornea in centre) was then placed gently on to the developed gonioscopic lens based imaging system to image the TM region in the ICA.
Additionally using Latina 5 Bar SLT, (single mirror, Ocular instruments, WA, USA) static gonioscopy was also performed on porcine eyes under white light. Vidisic gel was used as the lubricant. Gonioscopy was performed on clear as well as mildly cloudy porcine eyes to compare the image quality. The images were recorded using a colour CCD camera.
EyeCam
EyeCam (Clarity Medical systems, Pleasanton, CA) imaging was also performed on porcine samples (n = 4). The images were taken by a trained technician at Singapore Eye Research Institute, who is specialized on EyeCam imaging and had basic knowledge of angle anatomy and structures. Using foot pedal, illumination was adjusted to avoid overexposure. Clear still images of the ICA were then recorded with the probe being placed at the opposite limbal region.
In-house developed handheld imaging probe for iridocorneal angle
The flexible handheld probe has an imaging sensor located at the central axis of the probe and has a variable resolution at different depths which is optimized for recording the ICA of the eye17. The imaging probe was placed near the limbal region of the porcine cornea to image the opposite ICA through vidisic gel. The imaging probe has LEDs integrated to provide illumination.
Optical Setup
In the gonioscopic setup shown in Fig. 1, a white light source (Correct Shimadzu FA-150EN Fiber Illuminator, Japan) was used for illumination. (3D model and the schematic diagram is also provided in the Supplementary Fig. S4.) The imaging head is a gonioscopic lens (Hoskins-Barkan Goniotomy adult 11 mm lens; Ocular Instruments, WA, USA). Through the gonioscopic lens, ICA region of the eye was imaged by a microscopic objective lens (LSM03 - 4.6X, VIS OCT scan lens, Thorlabs, USA). The long working distance of the objective lens (25 mm) and it’s large field of view enabled the integration of the microscopic lens with a gonioscopic lens. The light from the microscopic objective was then collected via a tube lens (ITL 200, Thorlabs). A convex lens (LB-1844-A, Thorlabs) was placed at a focal length distance from the focal plane of the tube lens. An axicon lens was placed in close proximity of convex lens. The combination of convex lens and axicon lens was known to transform the wavefront of a point source into a Bessel beam, which has the special property to propagate without diffracting. A beam passing through axicon lens was captured by the CCD camera (EXi Aqua Bio-Imaging Microscopy Camera, QImaging, Canada). Lateral resolution of the imaging system is around 3 μm. Further, optical ray tracing of the imaging system is detailed in the Supplementary Fig. 3. Additionally information about basic BBM configuration is further detailed in the Supplementary material.
Acquired images were post processed using ImageJ software. In order to enhance the image quality, noise reduction filtering and thresholding was applied to the captured images. Additionally, edge detection was also performed to highlight the structures in the images.
Additional Information
How to cite this article: Perinchery, S. M. et al. High resolution iridocorneal angle imaging system by axicon lens assisted gonioscopy. Sci. Rep. 6, 30844; doi: 10.1038/srep30844 (2016).
Supplementary Material
Acknowledgments
The authors acknowledge the financial support received through A*STAR-SERC grant (No. 112148 0003), COLE-EDB and MOE (RG162/15). One of the authors, A.S. thanks NTU for the research student scholarship award.
Footnotes
Author Contributions S.M.P. designed and developed the research along with V.M.M. presented in this manuscript. S.M.P. performed experiments, modelling and analysed the data. C.Y.F. performed the optical modelling. A.S. performed experiments and image processing. S.M.P. and V.M.M. discussed and analysed the research. X.J.J.H. performed imaging with hand held probe. M.B. and T.A. gave clinical inputs and facilitated experiments with clinical instruments. All authors prepared, read and approved the final manuscript. V.M.M. advised and supervised the research as principal investigator of the project.
References
- Brubaker R. F. Flow of aqueous humor in humans [The Friedenwald Lecture]. Invest Ophthalmol Vis Sci. 32, 3145–3166 (1991). [PubMed] [Google Scholar]
- Konstantopoulos A., Hossain P. & Anderson D. F. Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis? Br J Ophthalmol. 91, 551–557 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wolffsohn J. S. & Davies L. N. Advances in anterior segment imaging. Curr Opin Ophthalmol. 18, 32–38 (2007). [DOI] [PubMed] [Google Scholar]
- Guyomard J.-L. et al. A Low-Cost and Simple Imaging Technique of the Anterior and Posterior Segments: Eye Fundus, Ciliary Bodies, Iridocorneal Angle. Invest Ophthalmol Vis Sci. 49, 5168–5174 (2008). [DOI] [PubMed] [Google Scholar]
- Quek D. T., Nongpiur M. E., Perera S. A. & Aung T. Angle imaging: Advances and challenges. Indian J Ophthalmol. 59, S69 (2011). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Perera S. A. et al. Use of EyeCam for imaging the anterior chamber angle. Invest Ophthalmol Vis Sci. 51, 2993–2997 (2010). [DOI] [PubMed] [Google Scholar]
- Baskaran M., Aung T., Friedman D. S., Tun T. A. & Perera S. A. Comparison of EyeCam and anterior segment optical coherence tomography in detecting angle closure. Acta Ophthalmol. 90, e621–e625 (2012). [DOI] [PubMed] [Google Scholar]
- McNabb R. P., Challa P., Kuo A. N. & Izatt J. A. Complete 360° circumferential gonioscopic optical coherence tomography imaging of the iridocorneal angle. Biomed Opt Express. 6, 1376–1391 (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wojtkowski M., Marcos S., Ortiz S. & Grulkowski I. Application of Fourier Domain OCT Imaging Technology to the Anterior Segment of the Human Eye. Optical Coherence Tomography: Technology and Applications, 1617–1648 (2015). [Google Scholar]
- Kalev-Landoy M., Day A. C., Cordeiro M. F. & Migdal C. Optical coherence tomography in anterior segment imaging. Acta Ophthalmol Scand. 85, 427–430 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith S. D. et al. Evaluation of the anterior chamber angle in glaucoma: a report by the american academy of ophthalmology. Ophthalmology 120, 1985–1997 (2013). [DOI] [PubMed] [Google Scholar]
- Ishikawa H. & Schuman J. S. Anterior segment imaging: ultrasound biomicroscopy. Ophthalmology Clinics of North America 17, 7 (2004). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Narayanaswamy A. et al. Anterior Chamber Angle Assessment Using Gonioscopy and Ultrasound Biomicroscopy. Jpn J Ophthalmol. 48, 44–49 (2004). [DOI] [PubMed] [Google Scholar]
- Radhakrishnan S. et al. Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 123, 1053–1059 (2005). [DOI] [PubMed] [Google Scholar]
- Friedman D. S. & He M. Anterior chamber angle assessment techniques. Surv Ophthalmol. 53, 250–273 (2008). [DOI] [PubMed] [Google Scholar]
- Masihzadeh O., Ammar D. A., Kahook M. Y., Gibson E. A. & Lei T. C. Direct trabecular meshwork imaging in porcine eyes through multiphoton gonioscopy. J Biomed Opt. 18, 036009-036009-5 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shinoj V. K., Murukeshan V. M., Baskaran M. & Aung T. Integrated flexible handheld probe for imaging and evaluation of iridocorneal angle. J Biomed Opt. 20, 016014-016014-5 (2015). [DOI] [PubMed] [Google Scholar]
- Masihzadeh O., Ammar D. A., Kahook M. Y., Gibson E. A. & Lei T. C. Multiphoton gonioscopy to image the trabecular meshwork of porcine eyes. Proc. of SPIE. 8567, 85671E-85671E-8 (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Snoeyink C. & Wereley S. Single-image far-field subdiffraction limit imaging with axicon. Opt Lett. 38, 625–627 (2013). [DOI] [PubMed] [Google Scholar]
- Snoeyink C. Imaging performance of Bessel beam microscopy. Opt Lett. 38, 2550–2553 (2013). [DOI] [PubMed] [Google Scholar]
- McMenamin P. G. & Steptoe R. J. Normal anatomy of the aqueous humour outflow system in the domestic pig eye. J Anat. 178, 65–77 (1991). [PMC free article] [PubMed] [Google Scholar]
- Richardson T. M., Hutchinson B. T. & Grant W. M. The outflow tract in pigmentary glaucoma: a light and electron microscopic study. Arch Ophthalmol. 95, 1015–1025 (1977). [DOI] [PubMed] [Google Scholar]
- Kupfer C., Kuwabara T. & Kaiser-Kupfer M. The histopathology of pigmentary dispersion syndrome with glaucoma. Am J Ophthalmol. 80, 857–862 (1975). [DOI] [PubMed] [Google Scholar]
- Sugar H. S. & Barbour F. A. Pigmentary glaucoma; a rare clinical entity. Am J Ophthalmol. 32, 90–92 (1949). [PubMed] [Google Scholar]
- Voss Kyhn M. et al. The multifocal electroretinogram (mfERG) in the pig. Acta Ophthalmol Scand. 85, 438–444 (2007). [DOI] [PubMed] [Google Scholar]
- Heijl A., Peters D., Leske M. C. & Bengtsson B. Effects of Argon Laser Trabeculoplasty in the Early Manifest Glaucoma Trial. Am J Ophthalmol. 152, 842–848 (2011). [DOI] [PubMed] [Google Scholar]
- Kinori M., Hostovsky A., Skaat A., Schwartsman J. & Melamed S. A novel method for quantifying the amount of trabecular meshwork pigment in glaucomatous and nonglaucomatous eyes. J Glaucoma. 23, e13–17 (2014). [DOI] [PubMed] [Google Scholar]
- Nakamura Y., Nakamura Y., Morine-Shinjo S., Sakai H. & Sawaguchi S. Assessment of chamber angle pigmentation during longterm latanoprost treatment for open-angle glaucoma. Acta Ophthalmol Scand. 82, 158–160 (2004). [DOI] [PubMed] [Google Scholar]
- Chun Y. S., Park I. K., Lee H. I., Lee J. H. & Kim J. C. Iris and trabecular meshwork pigment changes after posterior chamber phakic intraocular lens implantation. J Cataract Refract Surg. 32, 1452–1458 (2006). [DOI] [PubMed] [Google Scholar]
- Middleton S. Porcine Ophthalmology. Veterinary Clinics of North America: Food Animal Practice 26, 557–572 (2010). [DOI] [PubMed] [Google Scholar]
- Sanchez I., Martin R., Ussa F. & Fernandez-Bueno I. The parameters of the porcine eyeball. Graefes Arch Clin Exp Ophthalmol. 249, 475–482 (2011). [DOI] [PubMed] [Google Scholar]
- Ruiz-Ederra J. et al. The pig eye as a novel model of glaucoma. Exp Eye Res. 81, 561–569 (2005). [DOI] [PubMed] [Google Scholar]
- Fernandez-Bueno I., Pastor J. C., Gayoso M. J., Alcalde I. & Garcia M. T. Müller and macrophage-like cell interactions in an organotypic culture of porcine neuroretina. Mol Vis. 14, 2148 (2008). [PMC free article] [PubMed] [Google Scholar]
- Nishi O., Nishi K., Nishi Y. & Chang S. Capsular bag refilling using a new accommodating intraocular lens. J Cataract Refract Surg. 34, 302–309 (2008). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Faber C. et al. Orthotopic porcine corneal xenotransplantation using a human graft. Acta Ophthalmol. 87, 917–919 (2009). [DOI] [PubMed] [Google Scholar]
- Acosta E., Vázquez D. & Castillo L. R. Analysis of the optical properties of crystalline lenses by point‐diffraction interferometry. Ophthalmic and Physiological Optics. 29, 235–246 (2009). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.