Table 2.
2DLC recovery evaluation. Nac-linker-DSEA reference standard was prepared at 4 concentrations throughout the experimentally determined dynamic range and evaluate for recovery efficiency using the 2DLC configuration illustrated in Figure 4A. Identical separations were performed in a 1DLC configuration using the same system with a union in lieu of the SPE column (1st dimension) as a reference. Comparison of peak area across triplicate injections of the NAc-linker-DSEA reference standard indicate sample recovery was equivalent between both 1DLC and 2DLC configurations.
Conc. (ng/mL) |
1DLC |
2DLC |
||||
---|---|---|---|---|---|---|
Area | R.SD (%) | R.E. (%) | Area | R.SD (%) | R.E. (%) | |
2 | 458 | 5.2 | 95.4 | 482 | 1.7 | 100.4 |
17 | 4117 | 0.2 | 99.6 | 4146 | 1.0 | 100.3 |
74 | 17776 | 1.7 | 99.2 | 18063 | 2.7 | 100.8 |
146 | 35556 | 1.9 | 101.0 | 34839 | 0.2 | 99.0 |