. 2015 Dec 14;8(2):306–317. doi: 10.1080/19420862.2015.1116659

Table 2.

2DLC recovery evaluation. Nac-linker-DSEA reference standard was prepared at 4 concentrations throughout the experimentally determined dynamic range and evaluate for recovery efficiency using the 2DLC configuration illustrated in Figure 4A. Identical separations were performed in a 1DLC configuration using the same system with a union in lieu of the SPE column (1st dimension) as a reference. Comparison of peak area across triplicate injections of the NAc-linker-DSEA reference standard indicate sample recovery was equivalent between both 1DLC and 2DLC configurations.

Conc. (ng/mL)
1DLC
2DLC
  Area R.SD (%) R.E. (%) Area R.SD (%) R.E. (%)
2 458 5.2 95.4 482 1.7 100.4
17 4117 0.2 99.6 4146 1.0 100.3
74 17776 1.7 99.2 18063 2.7 100.8
146 35556 1.9 101.0 34839 0.2 99.0