Abstract
Myo-inositol (MI) is one of the stereoisomers of hexahydroxycyclohexane, which plays an important role in intracellular signal pathway. Derivatization is an indispensable step in both external and internal standard method during the chromatography-mass spectrometer (GC–MS) detection, as MI can’t be ionized directly. It is valuable to optimize the derivative process and the detection volume for clinical detection. This article contains optimization data related to research publication “Quantification of plasma myo-inositol using gas chromatography–mass spectrometry” [1]. Here we introduce the data on the optimized derivatization volume, temperature, duration and the detection volume.
Keywords: Gas chromatography-mass spectrometry, Myo-inositol, Derivatives
Specifications Table
| Subject area | Chemistry |
| More specific subject area | Analytical Chemistry, Chromatography-Mass Spectrometer |
| Type of data | Table, graph |
| How data was acquired | GC–MS |
| Data format | Raw |
| Experimental factors | Extraction reagents were added to the plasma and evaporated to be dry before derivatization; experimental factors include the derivatization volume, temperature, duration and the detection volume. |
| Experimental features | 7890A Gas Chromatography equipment a fused silica HP- 5 MS capillary column (Agilent Technologies, USA) was used for the GC separation. |
| Data source location | Beijing, China |
| Data accessibility | Data with this article |
Value of the data
-
•
The data for the optimized derivatization volume, temperature, duration and the detection volume is presented;
-
•
Periphery blood is enough for the detection of plasma MI;
-
•
The optimized derivatization condition for MI analysis could be used not only for biological specimens, but also for food and others.
1. Data
This data consist of the optimal derivatization condition of MI, including amount of derivatization reagent, derivative temperature and derivative time (Fig. 1, Fig. 2). Furthermore, the detection volume of the plasma was minimized (Table 1).
Fig. 1.
Total ion chromatogram (TIC) profiles of GC–MS results.
Fig. 2.
Optimal derivatization conditions for plasma MI detection. The derivatization reagent amount (A) derivative temperature (B) and derivative time (C) were optimized.
Table 1.
Optimize the plasma volume for MI detection.
| Injection volume (μL) | Peak area | MI concentrion (μg/L) | Ratio (MI concentrion/Injection volume) |
|---|---|---|---|
| 10 | 1142317 | 5 | 0.50 |
| 30 | 2205462 | 87 | 2.90 |
| 50 | 2556501 | 136 | 2.72 |
| 70 | 3211173 | 181 | 2.59 |
2. Experimental design, materials and methods
2.1. Gas chromatography instrument and reagents
The gas chromatography instrument and reagents were used as our previous study [1].
2.2. Optimal derivatization condition of MI
Derivatization is an important and vital step in pretreatment process in GC–MS analysis [2], [3]. Derivatization conditions were optimized by using a 5 mg/l MI working solution, which was placed into 30 μl human plasma. The various amount of derivatization reagent (1 ml, 3 ml, 5 ml, 8 ml, 10 ml, 13 ml and 15 ml), derivative temperature (65 °C, 70 °C, 75 °C, 80 °C and 85 °C) and derivative time (15 min, 30 min, 45 min, 60 min, 75 min, 90 min and 105 min) were investigated. The peak area of various derivatization conditions was analyzed and results were shown in Fig. 1. It was found that the derivatization yield increased as the increase of reaction volume less than 5 ml. 5 ml reagent was observed to be sufficient for derivatization. So did the situation when the derivative temperature above 70 °C and the derivative duration longer than 60 min (Fig. 2). Therefore, it was determined that the optimized derivatization conditions were using a 5 ml mixture of TMCS/HMDS/N, N - DMF at 70 °C for 60 min and shaking at 10 min intervals.
2.3. Optimal the volumes of plasma for detection
To minimize the detection volume of the plasma, various volumes of plasma samples (10 μl, 30 μl, 50 μl and 70 μl) were pretreated and evaluated by GC–MS method. The result showed that the plasma volume was proportional to the concentration of the plasma MI from 70 μl to 30 μl of the detection volume. Therefore, 30 μl was the minimal volume for detection (Table 1).
Acknowledgments
This study was supported by National Science and Technology Support Program (2013BAI12B01); the Ministry of Science and Technology of China, National “973” Project on Population and Health (2013CB945404); and the National Natural Science Foundation of China (81571443).
Footnotes
Transparency data associated with this article can be found in the online version at 10.1016/j.dib.2016.07.024.
Contributor Information
Jianhua Wang, Email: fywjh@163.com.
Ting Zhang, Email: zhangtingcv@126.com.
Transparency document. Supplementary material
Supplementary material
References
- 1.Guo Jin, Shi Yingfei, Xu Chengbao, Zhong Rugang, Zhang Feng, Zhang Ting, Wang Jianhua, Niu Bo. Quantification of plasma myo-inositol using gas chromatography-mass spectrometry. Clin. Chim. Acta. 2016;460:88–92. doi: 10.1016/j.cca.2016.06.022. [DOI] [PubMed] [Google Scholar]
- 2.Marcos J., Pozo O.J. Derivatization of steroids in biological samples for GC–MS and LC-MS analyses. Bioanalysis. 2015;7(19):2515–2536. doi: 10.4155/bio.15.176. [DOI] [PubMed] [Google Scholar]
- 3.Emwas A.H., Al-Talla Z.A., Kharbatia N.M. Sample collection and preparation of biofluids and extracts for gas chromatography mass spectrometry. Methods Mol. Biol. 2015;1277:75–90. doi: 10.1007/978-1-4939-2377-9_7. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supplementary material