. 1991 Feb;44(2):130–133. doi: 10.1136/jcp.44.2.130

Comparison of a modified thiazole orange technique with a fully automated analyser for reticulocyte counting.

D Bowen 1, N Bentley 1, T Hoy 1, I Cavill 1
PMCID: PMC496974  PMID: 1864984

Abstract

Two independent methods for quantitating reticulocyte counts were compared. One used a modified thiazole orange technique and a flow cytometer (Becton Dickinson FACS); the other was a fully automated whole blood analyser (Sysmex R1000). Both methods gave comparable results with a coefficient of variation of less than 5%. Samples measured using the R1000 showed a negligible decrease in the reticulocyte count over five days at room temperature, although there was evidence of continuing intracellular maturation: with thiazole orange there was an apparent increase. A practical reference range of 20-70 x 10(9)/l was established from 89 normal subjects. The close correlation between the two independent estimates indicates the validity of the quantitation of the reticulocyte count and shows that automation allows significant changes within and below the normal range to be detected with a degree of reliability which was not previously possible.

Full text

PDF
130

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Carter J. M., McSweeney P. A., Wakem P. J., Nemet A. M. Counting reticulocytes by flow cytometry: use of thiazole orange. Clin Lab Haematol. 1989;11(3):267–271. doi: 10.1111/j.1365-2257.1989.tb00218.x. [DOI] [PubMed] [Google Scholar]
  2. Ferguson D. J., Lee S. F., Gordon P. A. Evaluation of reticulocyte counts by flow cytometry in a routine laboratory. Am J Hematol. 1990 Jan;33(1):13–17. doi: 10.1002/ajh.2830330104. [DOI] [PubMed] [Google Scholar]
  3. Hillman R. S. Characteristics of marrow production and reticulocyte maturation in normal man in response to anemia. J Clin Invest. 1969 Mar;48(3):443–453. doi: 10.1172/JCI106001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Kojima K., Niri M., Setoguchi K., Tsuda I., Tatsumi N. An automated optoelectronic reticulocyte counter. Am J Clin Pathol. 1989 Jul;92(1):57–61. doi: 10.1093/ajcp/92.1.57. [DOI] [PubMed] [Google Scholar]
  5. Lee L. G., Chen C. H., Chiu L. A. Thiazole orange: a new dye for reticulocyte analysis. Cytometry. 1986 Nov;7(6):508–517. doi: 10.1002/cyto.990070603. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES