Skip to main content
Journal of Clinical Pathology logoLink to Journal of Clinical Pathology
. 1991 Feb;44(2):165–167. doi: 10.1136/jcp.44.2.165

The need for primary quality control of commercially available immunoassay kits.

T Hashimoto 1, F Matsubara 1
PMCID: PMC496983  PMID: 1864991

Abstract

Discrepancies among three commercially available luteinising hormone (LH) radioimmunoassay RIA kits, all calibrated against the second-IRP-HMG (World Health Organisation International Laboratory for Biological Standards), were observed. Even though the Japan Radioisotope Association carries out external quality control in Japan every year, inaccuracies still exist. This problem also exists elsewhere, and it is proposed that a global monitoring system of primary quality control (PQC) should be introduced by a body such as the World Health Organisation.

Full text

PDF
165

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Banfi G., Casari E., Murone M., Bonini P. Discrepancies among commercial kits for assaying growth hormone. Clin Chem. 1990 Feb;36(2):402–402. [PubMed] [Google Scholar]
  2. Hashimoto T., Ohba N., Matsubara F. Discordant CA 125 measurements in normal healthy women. Clin Chem. 1990 Jul;36(7):1377–1378. [PubMed] [Google Scholar]
  3. Klug T. L., Green P. J., Zurawski V. R., Jr, Davis H. M. Confirmation of a false-positive result in CA 125 immunoradiometric assay caused by human anti-idiotypic immunoglobulin. Clin Chem. 1988 Jun;34(6):1071–1076. [PubMed] [Google Scholar]
  4. Mahadevan M. S., Ooi D. S. Changes in free thyroxin (FT4) values of Clinical Assays Gammacoat two-step assay. Clin Chem. 1988 May;34(5):987–987. [PubMed] [Google Scholar]
  5. Monro J., Brostoff J., Carini C., Zilkha K. Food allergy in migraine. Study of dietary exclusion and RAST. Lancet. 1980 Jul 5;2(8184):1–4. doi: 10.1016/s0140-6736(80)92887-1. [DOI] [PubMed] [Google Scholar]
  6. Shishiba Y., Irie M., Yamada H., Kinoshita F. An inter- and intralaboratory quality-control survey of radioimmunoassay of insulin, thyroxin, thyrotropin, cortisol, digoxin, gastrin, beta 2-microglobulin, and IgE in Japan with commercially available kits. Clin Chem. 1983 Aug;29(8):1501–1507. [PubMed] [Google Scholar]
  7. Smith M. C., Brown M. E., Morel J. H., Woodward S. C. Erroneous test results obtained with the Tandem-E TSH kit. Clin Chem. 1988 Sep;34(9):1946–1947. [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES