Comparison of implant quality between intraoperatively built custom-linked seeds and loose seeds in permanent prostate brachytherapy using sector analysis

Norihisa Katayama 1,*, Mitsuhiro Takemoto 2, Atsushi Takamoto 3, Hiroki Ihara 1, Kuniaki Katsui 1, Shin Ebara 3, Yasutomo Nasu 3, Susumu Kanazawa 1
PMCID: PMC4973643  PMID: 26976125

Abstract

We compared the implant quality of intraoperatively built custom-linked (IBCL) seeds with loose seeds in permanent prostate brachytherapy. Between June 2012 and January 2015, 64 consecutive prostate cancer patients underwent brachytherapy with IBCL seeds (n = 32) or loose seeds (n = 32). All the patients were treated with 144 Gy of brachytherapy alone. Brachytherapy was performed using a dynamic dose calculation technique. Computed tomography/magnetic resonance imaging fusion-based dosimetry was performed 1 month after brachytherapy. Post-implant dose–volume histogram (DVH) parameters, prostate sector dosimetry, operation time, seed migration, and toxicities were compared between the IBCL seed group and the loose seed group. A sector analysis tool was used to divide the prostate into six sectors (anterior and posterior sectors at the base, mid-gland, and apex). V100 (95.3% vs 89.7%; P = 0.014) and D90 (169.7 Gy vs 152.6 Gy; P = 0.013) in the anterior base sector were significantly higher in the IBCL seed group than in the loose seed group. The seed migration rate was significantly lower in the IBCL seed group than in the loose seed group (6% vs 66%; P < 0.001). Operation time per seed was significantly longer in the IBCL seed group than in the loose seed group (1.31 min vs 1.13 min; P = 0.003). Other post-implant DVH parameters and toxicities did not differ significantly between the two groups. Our study showed more dose coverage post-operatively in the anterior base prostate sector and less seed migration in IBCL seed implantation compared with loose seed implantation.

Keywords: prostate cancer, brachytherapy, intraoperatively built custom-linked seeds, dosimetry, sector analysis, seed migration

INTRODUCTION

Permanent prostate brachytherapy (PPB) has become a standard treatment option for patients with localized prostate cancer, with long-term local and biochemical control similar to outcomes observed after radical prostatectomy and external beam radiation therapy (EBRT) [13]. In Japan, PPB was first adopted in 2003, and 109 institutions have used this treatment for 27 000 cases throughout the country up to the end of 2013 [4]. However, because stranded or linked seeds had not been introduced until 2012, only loose seeds had been used for PPB until then.

An intraoperatively built custom-linked (IBCL) seeds system is a push-button seed delivery system that allows the user to create intraoperatively customized linked seeds, using a combination of seeds, connectors and spacers. Zauls et al. first reported this system in 2010 [5]. It was introduced into Japan in 2012.

Sector analysis was developed by Bice et al., and is a method in which the organ is divided into different sectors according to anatomic locations [6]. Sector analysis of the prostate allows dose calculations not only to the whole prostate but also to specific parts of it [68].

To date, only three studies have compared the implant quality of IBCL seeds with that of loose seeds for use in PPB [5, 9, 10]. Although the three studies compared post-implant dose–volume histogram (DVH) parameters, seed migration rates, and operation times between the IBCL seed group and the loose seed group, they did not use sector analysis. Therefore, we compared the implant quality of IBCL seeds with that of loose seeds in PPB using sector analysis.

MATERIALS AND METHODS

Study design

This retrospective study was approved by the Institutional Review Board. Between June 2012 and January 2015, 64 consecutive patients with low-risk or intermediate-risk prostate cancer (prostate-specific antigen level ≤20 ng/ml; Gleason score 6–7; Union Internationale Contre le Cancer 2009 clinical stage T1–T2) were treated with PPB at Okayama University Hospital (Table 1). All patients were treated with brachytherapy alone using 125I radioactive seeds. IBCL and loose seeds were alternately used basically. All patients were treated by the same radiation oncologist (N.K.) and urologist (A.T.). Both had been well trained in PPB.

Table 1.

Patient characteristics

IBCL groupa Loose groupa P value
n 32 32
Age (years) 66.3 ± 4.8 67.6 ± 6.3 0.11
PSA (ng/ml) 6.14 ± 2.09 7.41 ± 3.21 0.81
T stage (1c/2a/2b/2c) 15/9/5/3 18/11/2/1 0.43
Gleason score (3+3/3+4) 24/8 20/12 0.28
Hormonal therapy 20 21 0.79

 aValues are given as number or mean ± SD. IBCL = intraoperatively built custom-linked, PSA = prostate-specific antigen.

All patients underwent an ultrasonography volume study 2–6 weeks before implantation to determine the number of seeds to order. Loose seeds (Oncoseed; GE Healthcare, Medi-Physics, Arlington Heights, IL) were implanted using a Mick applicator (Mick Radio Nuclear Instruments, Mount Vernon, NY). The seed activity of the loose seed was 0.35 mCi/seed. The IBCL seeds were constructed using a Quicklink device (CR Bard, Covington, GA) and implanted. The seed activity of the IBCL seeds was 0.367 mCi/seed. Zauls et al. [5] have described the detailed procedure of constructing IBCL seeds, and we used a procedure similar to theirs. The radiation oncologist constructed the IBCL seeds.

Transrectal ultrasonography images in the axial plane were imported into the Variseed (Varian Medical Systems, Palo Alto, CA) brachytherapy planning system. The prostate, urethra, and rectal wall were contoured by the radiation oncologist. Needles were placed in the periphery of the gland 0.5–1 cm apart at the largest cross-section of the gland by the urologist. The radiation oncologist determined the positions of the centrally located needles. The planning software determined the first plan of seed number and location combinations for both the peripheral and centrally located needles. Modifications to the plan were made by the radiation oncologist, and the software recalculated the DVH and isodose lines in real time. The urologist placed the seeds, and the radiation oncologist performed a dynamic dose calculation [11]. For the IBCL seeds, the radiation oncologist both constructed the IBCL seeds and performed the dynamic dose calculation.

The prescribed dose was set at 144 Gy. Dose–volume targets were as follows: V100 (the percentage of the prostate volume that receives 100% of the prescribed dose) >99%; UD1 (the dose irradiating 1% of the urethral volume) <200 Gy; and RV100 (the rectal volume that receives 100% of the prescribed dose) <0.5 ml. We placed a few seeds inside the seminal vesicle, if necessary, but only with the IBCL seed. For both the IBCL and loose seed we placed some seeds outside the prostate capsule to achieve the urethral dose–volume target.

Post-implant dosimetry was performed using computed tomography (CT)/magnetic resonance imaging (MRI) fusion 1 month after implant. The same radiation oncologist (N.K.) performed the post-implant dosimetry for all patients, and the results were confirmed by the second radiation oncologist (M.T.). The second radiation oncologist (M.T.) had also been well trained in PPB. No urinary catheter was put in place. A CT scan was obtained using a 128-channel, dual-source multidetector-row CT scanner (SOMATOM Definition Flash, Siemens AG, Forchheim, Germany). Axial 0.6-mm-thick CT images of the prostate at 0.6-mm intervals were obtained. For these scans, a field of view (FOV) of 15 cm and a 512-square matrix were used. An MRI scan was obtained immediately after the CT using a 3-tesla (T) MRI scanner (Magnetom Verio 3 T, Siemens AG, Forchheim, Germany). Axial T2-weighted images of the prostate were obtained using a 16-channel body array anterior and posterior coil. Technical parameters were as follows: repetition time (TR)/echo time (TE) in milliseconds, 4000/99; FOV, 20 cm; slice thickness, 3 mm without a gap; matrix size, 224 × 320. The CT and MRI were electronically fused using the manual-fusion procedure of the Variseed fusion system. We identified nine or more corresponding seed pairs in the CT and MRI and then used the Variseed program to calculate the transformation. This fusion procedure was performed until the positions of the seeds in the CT image corresponded with those in the MRI. On CT, the seeds were detected automatically using the Variseed system, and the rectum was manually contoured. On MRI, the prostate and urethra were manually contoured. The urethra and rectum were contoured using the same slices as the prostate contour.

A sector analysis tool was used to divide the contoured prostate into six sectors: anterior and posterior sectors at the base, mid-gland, and apex (Fig. 1). Evaluation of radiation coverage in each of the six sectors was performed for both the intraoperative plan and the post-implant dosimetry. V100 and D90, the dose irradiating 90% of the prostate volume, were calculated and compared between the two groups.

Fig. 1.

Fig. 1.

The sector analysis tool was used to divide the prostate into six sectors: anterior and posterior sectors at the base, mid-gland, and apex.

On the same day as the CT/MRI, a chest radiograph (anteroposterior view) and an abdominal radiograph were taken to check for any possible seed migration. Seed migration was defined as a seed separated from the main seed cluster. The definition of ‘distal seed migration’ was one or more seeds located distally toward the perineum due to needle drag.

The urinary and rectal toxicities were evaluated in accordance with the Japanese version of the National Cancer Institute's Common Terminology Criteria for Adverse Events, version 4.0.

Statistical analysis

We compared the baseline characteristics and the outcomes of interest between the IBCL seed group and the loose seed group using the 2-sample t test for continuous data that followed a normal distribution, the Mann–Whitney test for continuous data that did not follow a normal distribution, and the Chi-squared test for categorical data. Probability (P) values of <0.05 were considered significant. Data processing and statistical analyses were carried out with SPSS Statistics 22 software (IBM, Chicago, IL).

RESULTS

No differences in patient characteristics were seen between the two groups (Table 1).

The number of seeds was significantly larger in the loose seed group than in the IBCL seed group; however, there was no significant difference in total activity between the two groups (Table 2). Although the same predetermined dosimetric parameters were used, D90 and D90 in the posterior base sector were significantly higher in the loose seed group than in the IBCL seed group in the planning phase. There was no significant difference in operation time between the two groups; however, operation time per seed was significantly longer in the IBCL seed group than in the loose seed group (1.31 min vs 1.13 min; P = 0.003).

Table 2.

Planning phase

IBCL groupa Loose groupa P value
n 32 32
Prostate volume (ml) 31.4 ± 4.1 32.7 ± 4.8 0.27
No. of seeds 83.3 ± 7.0 89.2 ± 7.4 <0.001
No. of needles 23.9 ± 2.2 24.0 ± 1.8 0.81
Total activity (mCi) 30.6 ± 2.6 31.2 ± 2.6 0.18
D90 (Gy) 175.6 ± 6.7 179.6 ± 7.3 0.024
V100 (%) 99.4 ± 0.6 99.5 ± 0.7 0.96
V150 (%) 56.3 ± 6.3 59.6 ± 7.1 0.051
RV100 (ml) 0.40 ± 0.24 0.34 ± 0.22 0.38
RV150 (ml) 0.01 ± 0.02 0.01 ± 0.01 0.29
UD90 (Gy) 146.8 ± 8.5 151.7 ± 13.0 0.077
UD5 (Gy) 195.4 ± 13.8 194.2 ± 9.4 0.69
Prostate sector dosimetry
 D90 (Gy) Anterior base 175.8 ± 10.6 180.1 ± 11.7 0.13
 Posterior base 177.7 ± 11.8 186.3 ± 12.1 0.005
 Anterior mid-gland 185.2 ± 9.4 186.0 ± 10.4 0.76
 Posterior mid-gland 179.3 ± 12.1 182.4 ± 8.4 0.23
 Anterior apex 175.6 ± 8.8 176.5 ± 9.8 0.69
 Posterior apex 161.8 ± 14.8 165.8 ± 12.8 0.26
 V100 (%) Anterior base 99.5 ± 1.0 99.1 ± 1.7 0.33
 Posterior base 99.7 ± 0.5 99.7 ± 0.5 0.26
 Anterior mid-gland 99.9 ± 0.2 99.9 ± 0.3 0.57
 Posterior mid-gland 99.4 ± 1.2 99.7 ± 0.4 0.93
 Anterior apex 99.6 ± 0.7 99.5 ± 1.9 0.32
 Posterior apex 96.3 ± 5.4 97.9 ± 2.3 0.80
Operation time (min) 108.7 ± 16.6 102.0 ± 15.2 0.13
Operation time/seed (min) 1.31 ± 0.21 1.13 ± 0.21 0.003

aValues are given as mean ± SD. D90 = the dose irradiating 90% of the prostate volume, V100 = the percentage of the prostate volume that receives 100% of the prescribed dose, V150 = the percentage of the prostate volume that receives 150% of the prescribed dose, RV100 = the rectal volume that receives 100% of the prescribed dose, RV150 = the rectal volume that receives 150% of the prescribed dose, UD90 = the dose irradiating 90% of the urethral volume, UD5 = the dose irradiating 5% of the urethral volume.

Table 3 shows the dosimetric results at 1 month and the number of patients with seed migration. In prostate sector dosimetry, V100 (95.3% vs 89.7%; P = 0.014) and D90 (169.7 Gy vs 152.6 Gy; P = 0.013) in the anterior base sector were significantly higher in the IBCL seed group than in the loose seed group. The seed migration rate was significantly lower in the IBCL seed group than in the loose seed group (6% vs 66%; P < 0.001).

Table 3.

Post-implantation phase at 1 month

IBCL groupa Loose groupa P value
n 32 32
Prostate volume (ml) 29.0 ± 4.6 29.8 ± 4.0 0.47
D90 (Gy) 180.7 ± 12.7 178.1 ± 15.4 0.29
V100 (%) 98.2 ± 1.4 97.0 ± 2.4 0.057
V150 (%) 69.2 ± 9.9 68.8 ± 11.3 0.88
RV100 (ml) 0.97 ± 0.69 1.00 ± 0.71 0.78
RV150 (ml) 0.07 ± 0.09 0.11 ± 0.17 0.34
UD90 (Gy) 165.4 ± 19.4 154.6 ± 24.3 0.056
UD5 (Gy) 251.7 ± 28.3 246.4 ± 31.3 0.48
Prostate sector dosimetry
 D90 (Gy) Anterior base 169.7 ± 25.0 152.6 ± 28.3 0.013
 Posterior base 180.1 ± 20.7 179.6 ± 23.1 0.93
 Anterior mid-gland 196.5 ± 22.6 198.4 ± 27.9 0.77
 Posterior mid-gland 189.9 ± 18.7 200.1 ± 27.1 0.08
 Anterior apex 211.0 ± 26.1 198.3 ± 28.1 0.065
 Posterior apex 180.1 ± 29.8 171.9 ± 24.8 0.23
 V100 (%) Anterior base 95.3 ± 5.4 89.7 ± 10.1 0.014
 Posterior base 98.3 ± 2.6 97.2 ± 5.9 0.35
 Anterior mid-gland 99.2 ± 1.9 98.8 ± 2.8 0.86
 Posterior mid-gland 99.4 ± 1.3 99.5 ± 1.3 0.54
 Anterior apex 99.4 ± 1.6 99.0 ± 2.5 0.12
 Posterior apex 95.8 ± 11.7 96.5 ± 6.0 0.39
Patient with seed migration 2 21 <0.001
 Chest 1 15
 Abdominopelvic region 0 15
 Seminal vesicle 0 6
 Distal 1 1

aValues are given as number (%) or mean ± SD. Abbreviations are as in Table 2.

Table 4 shows toxicities after brachytherapy. The median follow-up was 18 months (range, 1–36 months). No significant differences in toxicities were seen between the two groups.

Table 4.

Crude rate of toxicity

Grade IBCL groupa Loose groupa P value
n 32 32
GU 0 4 (13%) 4 (13%) 0.99
1 27 (84%) 27 (84%)
2 1 (3%) 0 (0%)
3 0 (0%) 1 (3%)
GI 0 30 (93%) 31 (94%) 0.56
1 2 (6%) 1 (3%)

 aValues are given as number (%). GI = gastrointestinal, GU = genitourinary.

DISCUSSION

There are theoretical advantages in using linked seeds over using loose seed for PPB, including less seed migration and possible improved dosimetry if the implanted seeds do not migrate away from the prostate and stabilize due to linking [5, 10, 12]. Furthermore, the IBCL seeds system allows for combining seeds and connectors into seed trains of variable length and seed-to-seed spacing in the operating room, so it can meet the intraoperative planning method. The intraoperative planning method has a number of advantages over the preplanning method and some authors have reported it gives better dosimetric outcomes [1315]. However, to date, there are only three studies that have compared IBCL with loose seeds in PPB [5, 9, 10]. Our analysis of 64 patients (32 treated with IBCL seeds and 32 treated with loose seeds) has shown that IBCL seeds provide more dose coverage post-operatively in the anterior base prostate sector and less seed migration than loose seed implantation.

Based on these results, we recommend that prostate cancer derived from the anterior fibromuscular stroma and transition zone be treated with IBCL seeds. Additionally, it is usually reasonable that seminal vesicle invasion suspected disease should be treated with IBCL seeds. Therefore, we believe that indications for the use of IBCL seeds include cases of prostate cancer derived from the anterior fibromuscular stroma and transition zone and cases in which seminal vesicle invasion is suspected. However, a higher dose in the anterior base sector may be correlated with a higher bladder neck dose, and might cause more urinary toxicity—Hathout et al. reported that, among the standard dosimetric variables, the dose to the bladder neck was the strongest predictor for acute and late urinary toxicity [16].

Table 5 shows the previous data comparing post-implant dosimetry between IBCL seeds and loose seeds. Although Jarusevicius et al. reported some DVH parameters differed significantly between the two groups [9], DVH parameters did not differ significantly in the other two reports [5, 10]. In our study, post-implant DVH parameters did not differ significantly between the two groups, consistent with most prior data.

Table 5.

Previously published data comparing post-implant dosimetry between IBCL seeds and loose seeds

Year [Ref] n DVH parameter IBCL seeda Loose seeda P value
2010 [5] 91 D90 (Gy) Pd: 104.0 Pd: 98.2 0.42
I: 165.1 I: 164.5 0.36
% of RV100 > 1.3 ml 16.7 23.3 NS
2012 [9] 230 D90 (Gy) 177.9 184.7 0.002
V100 (%) 94.9 95.5 0.21
UD30 (Gy) 197.4 218.6 0.001
RV100 (ml) 0.3 0.6 <0.001
2014  [10] 140 D90 (Gy) 174.4 170.7 NS
V100 (%) 96.6 95.7 NS
UD30 (Gy) 203.2 206.8 NS
RV100 (ml) 0.47 0.51 NS
Present study 64 D90 (Gy) 180.7 178.1 0.29
V100 (%) 98.2 97.0 0.057
UD5 (Gy) 251.7 246.4 0.48
RV100 (ml) 0.97 1.00 0.78

aValues are given as mean or %. DVH = dose–volume histogram, UD30 = the dose irradiating 30% of the urethral volume. Other abbreviations are as in Table 2.

Regarding prostate sector dosimetry, many studies have reported that the anterior base sector and the base sector received a lower dose [8, 1721]. Seed migration, needle drag, needle splay, the proximity of the prostate base to the bladder, and the difficulty in accurately contouring the prostate base are proposed as possible explanations for the lower dose in the anterior base and base sectors [1721]. Seed migration is thought to be the reason why the dose in the anterior base sector decreased more in the loose seed group than in the IBCL seed group in our study. Because of the proximity of the venous plexus to the anterior base sector, it is reasonable to speculate that the majority of migrating seeds were intended for implant in the anterior base sector [17, 19]. Indeed, in our study, V100 (88.1% vs 92.4%; P = 0.043) and D90 (145.5 Gy vs 164.4 Gy; P = 0.068) in the anterior base sector were lower in the loose seed patients with seed migration into the chest and/or abdominopelvis (n = 20) than the loose seed patients without it (n = 12). Therefore, the dose in the anterior base sector might have decreased more in the loose seed group due to more seed migration than in the IBCL seed group.

Ishiyama et al. reported that the seed migration rate was significantly lower in the IBCL seed group (0%) than in the loose seed group (55%; P < 0.001) [10]. Several other studies have also reported that the seed migration rate was significantly lower in the stranded seed group than in the loose seed group [12, 22, 23]. Our study had similar result to theirs.

Zauls et al. [5] and Ishiyama et al. [10] reported the operation time was significantly longer in the IBCL seed group than in the loose seed group. In both their reports, there was no significant difference in the number of seeds between the IBCL seed group and the loose seed group. In the present study, there was no significant difference in operation time between the IBCL seed group and the loose seed group. However, the number of seeds was significantly larger in the loose seed group than in the IBCL seed group, and the operation time per seed was significantly longer in the IBCL seed group than in the loose seed group. We believe this was because we were accustomed to loose seed brachytherapy and constructing and placing IBCL seeds took a longer time than placing loose seeds.

The number of seeds was significantly larger in the loose seed group than in the IBCL seed group in our study. However, no significant differences were seen in prostate volume and total activity between the two groups, probably because the activity of the loose seed was 0.35 mCi/seed and that of the IBCL seed was 0.367 mCi/seed. Our study showed that D90 and D90 in the posterior base sector were significantly higher in the loose seed group than in the IBCL seed group in the planning phase. This may be attributable to unintentional bias. For the IBCL seed, we placed a few seeds inside the seminal vesicle, if necessary; therefore, we seem to have placed more seeds inside the prostate capsule in the posterior base sector for the loose seed to achieve V100 > 99% than for the IBCL seed. This may have contributed to the differences in D90 and D90 in the posterior base sector.

Our study has some limitations. This is a retrospective study. However, all patients were treated by the same radiation oncologist (N.K.) and urologist (A.T.). Other than placing a few seeds inside the seminal vesicle, if necessary, in the IBCL seed group, all other procedures and dose–volume targets remained the same between the two groups. No differences in patient characteristics were seen between the two groups. It is possible that intra-observer variability in post-implant dosimetry occurred [24]. However, CT/MRI fusion, which we used, is regarded as the best method for post-implant dosimetry [2527]. The radiation oncologist (N.K.) had performed more than 100 post-implant dosimetries using CT/MRI fusion before our study. Therefore, intra-observer variability in post-implant dosimetry should be limited in our study.

In conclusion, our study showed more dose coverage postoperatively in the anterior base prostate sector and less seed migration in IBCL seed implantation than in loose seed implantation.

CONFLICT OF INTEREST

Yasutomo Nasu owns the stock of Momotaro-Gene Inc., a company relevant to prostate cancer gene therapy, outside the submitted work.

ACKNOWLEDGEMENTS

The results of this study were presented at the 74th Annual Meeting of the Japan Radiological Society, Yokohama, Kanagawa, 16–19 April 2015.

REFERENCES

  • 1.Battermann JJ, Boon TA, Moerland MA. Results of permanent prostate brachytherapy, 13 years of experience at a single institution. Radiother Oncol 2004;71:23–8. [DOI] [PubMed] [Google Scholar]
  • 2.Stone NN, Stock RG. 15-year cause specific and all-cause survival following brachytherapy for prostate cancer: negative impact of long-term hormonal therapy. J Urol 2014;192:754–9. [DOI] [PubMed] [Google Scholar]
  • 3.Yorozu A, Kuroiwa N, Takahashi A et al. Permanent prostate brachytherapy with or without supplemental external beam radiotherapy as practiced in Japan: outcomes of 1300 patients. Brachytherapy 2015;14:111–7. [DOI] [PubMed] [Google Scholar]
  • 4.Saito S, Ito K, Yorozu A et al. Nationwide Japanese Prostate Cancer Outcome Study of Permanent Iodine-125 Seed Implantation (J-POPS). Int J Clin Oncol 2015;20:375–85. [DOI] [PubMed] [Google Scholar]
  • 5.Zauls AJ, Ashenafi MS, Onicescu G et al. Comparison of intraoperatively built custom linked seeds versus loose seed gun applicator technique using real-time intraoperative planning for permanent prostate brachytherapy. Int J Radiat Oncol Biol Phys 2010;81:1010–6. [DOI] [PubMed] [Google Scholar]
  • 6.Bice WS Jr, Prestidge BR, Sarosdy MF. Sector analysis of prostate implants. Med Phys 2001;28:2561–7. [DOI] [PubMed] [Google Scholar]
  • 7.Nasser NJ, Wang Y, Borg J et al. Sector analysis of dosimetry of prostate cancer patients treated with low-dose-rate brachytherapy. Brachytherapy 2014;13:369–74. [DOI] [PubMed] [Google Scholar]
  • 8.Merrick GS, Butler WM, Grimm P et al. Multisector prostate dosimetric quality: analysis of a large community database. Brachytherapy 2014;13:146–51. [DOI] [PubMed] [Google Scholar]
  • 9.Jarusevicius L, Inciura A, Juozaityte E et al. Comparison of implant quality between loose and intra-operatively linked iodine-125 seeds in prostate cancer brachytherapy. J Radiat Res 2012;53:439–46. [DOI] [PubMed] [Google Scholar]
  • 10.Ishiyama H, Satoh T, Kawakami S et al. A prospective quasi-randomized comparison of intraoperatively built custom-linked seeds versus loose seeds for prostate brachytherapy. Int J Radiat Oncol Biol Phys 2014;90:134–9. [DOI] [PubMed] [Google Scholar]
  • 11.Polo A, Salembier C, Venselaar J et al. Review of intraoperative imaging and planning techniques in permanent seed prostate brachytherapy. Radiother Oncol 2010;94:12–23. [DOI] [PubMed] [Google Scholar]
  • 12.Reed DR, Wallner KE, Merrick GS et al. A prospective randomized comparison of stranded vs. loose 125I seeds for prostate brachytherapy. Brachytherapy 2007;6:129–34. [DOI] [PubMed] [Google Scholar]
  • 13.Zelefsky MJ, Yamada Y, Marion C et al. Improved conformality and decreased toxicity with intraoperative computer-optimized transperineal ultrasound-guided prostate brachytherapy. Int J Radiat Oncol Biol Phys 2003;55:956–63. [DOI] [PubMed] [Google Scholar]
  • 14.Blasko JC. Low-dose-rate brachytherapy for prostate cancer: preplanning vs. intraoperative planning-preplanning is best. Brachytherapy 2006;5:139–42. [DOI] [PubMed] [Google Scholar]
  • 15.Yoshida K, Ohashi T, Yorozu A et al. Comparison of preplanning and intraoperative planning for I-125 prostate brachytherapy. Jpn J Clin Oncol 2013;43:383–9. [DOI] [PubMed] [Google Scholar]
  • 16.Hathout L, Folkert MR, Kollmeier MA et al. Dose to the bladder neck is the most important predictor for acute and late toxicity after low-dose-rate prostate brachytherapy: implications for establishing new dose constraints for treatment planning. Int J Radiat Oncol Biol Phys 2014;90:312–19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Sidhu S, Morris WJ, Spadinger I et al. Prostate brachytherapy postimplant dosimetry: a comparison of prostate quadrants. Int J Radiat Oncol Biol Phys 2002;52:544–52. [DOI] [PubMed] [Google Scholar]
  • 18.Takiar V, Pugh TJ, Swanson D et al. MRI-based sector analysis enhances prostate palladium-103 brachytherapy quality assurance in a phase II prospective trial of men with intermediate-risk localized prostate cancer. Brachytherapy 2014;13:68–74. [DOI] [PubMed] [Google Scholar]
  • 19.Mohamed Yoosuf AB, Workman G, O'Toole MM et al. Sector analysis of 125I permanent prostate brachytherapy provides a rapid and effective method of evaluating and comparing pre- and post-implant dosimetry. Brachytherapy 2013;12:254–9. [DOI] [PubMed] [Google Scholar]
  • 20.Brown AP, Pugh TJ, Swanson DA et al. Improving prostate brachytherapy quality assurance with MRI-CT fusion-based sector analysis in a phase II prospective trial of men with intermediate-risk prostate cancer. Brachytherapy 2013;12:401–7. [DOI] [PubMed] [Google Scholar]
  • 21.Nasser NJ, Wang Y, Borg J et al. Sector analysis of dosimetry of prostate cancer patients treated with low-dose-rate brachytherapy. Brachytherapy 2014;13:369–74. [DOI] [PubMed] [Google Scholar]
  • 22.Tapen EM, Blasko JC, Grimm PD et al. Reduction of radioactive seed embolization to the lung following prostate brachytherapy. Int J Radiat Oncol Biol Phys 1998;42:1063–7. [DOI] [PubMed] [Google Scholar]
  • 23.Fuller DB, Koziol JA, Feng AC. Prostate brachytherapy seed migration and dosimetry: analysis of stranded sources and other potential predictive factors. Brachytherapy 2004;3:10–19. [DOI] [PubMed] [Google Scholar]
  • 24.Dubois DF, Prestidge BR, Hotchkiss LA et al. Intraobserver and interobserver variability of MR imaging- and CT-derived prostate volumes after transperineal interstitial permanent prostate brachytherapy. Radiology 1998;207:785–9. [DOI] [PubMed] [Google Scholar]
  • 25.Aoki M, Yorozu A, Dokiya T. Results of a dummy run of postimplant dosimetry between multi-institutional centers in prostate brachytherapy with 125I seeds. Jpn J Radiol 2009;27:410–5. [DOI] [PubMed] [Google Scholar]
  • 26.Tanaka O, Hayashi S, Matsuo M et al. Comparison of MRI-based and CT/MRI fusion-based postimplant dosimetric analysis of prostate brachytherapy. Int J Radiat Oncol Biol Phys 2006;66:597–602. [DOI] [PubMed] [Google Scholar]
  • 27.Moerland MA, van Deursen MJ, Elias SG et al. Decline of dose coverage between intraoperative planning and post implant dosimetry for I-125 permanent prostate brachytherapy: comparison between loose and stranded seed implants. Radiother Oncol 2009;91:202–6. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Radiation Research are provided here courtesy of Oxford University Press