Abstract
In recent years there has been considerable interest in developing photonic temperature sensors such as the Fiber Bragg gratings (FBG) as an alternative to resistance thermometry. In this study we examine the thermal response of FBGs over the temperature range of 233 K to 393 K. We demonstrate, in hermetically sealed dry Argon environment, FBG devices show a quadratic dependence on temperature with expanded uncertainties (k=2) of ≈500 mK. Our measurements indicate that the combined measurement uncertainty is dominated by uncertainty in determining peak center fitting and thermal ageing of polyimide coated fibers.
1. Introduction
Temperature is one of the most measured quantities in the world. Despite the ubiquity of thermometers, the underlying technology has been slow to advance over the last century[1, 2]. For the past century, industrial temperature measurements have relied on resistance measurement of a thin metal wire or filament whose resistance varies with temperature [1, 3]. Though resistance thermometers can routinely measure temperatures with uncertainties of ≤10 mK (US industry requirement), they are sensitive to mechanical shock induced strain which causes the sensor resistance to drift over time [3]. Consequently, resistance thermometers require frequent off-line, expensive, and time consuming calibrations resulting in high ownership cost. Additionally, these devices suffer from electromagnetic interference, limiting their utility in extreme environments. These fundamental limitations of resistance devices have produced considerable interest in the development of photonic temperature sensors as an alternative to resistance thermometers as they have the potential to leverage advances in frequency metrology to provide greater temperature sensitivity while being robust against mechanical shock and electromagnetic interference [2, 4–10].
The vast and varied application landscape for temperature measurement has spawned a host of photonic temperature sensing solutions. The proposed sensor technologies range from temperature sensitive dyes [11], polymers [9, 12, 13] to silicon photonics such as ring resonators[2, 5, 6]. Fiber Bragg grating (FBG) based temperature sensors have already been commercially introduced as photonic alternative to resistance thermometry [14, 15]. FBG are a narrow band filter commonly used in the telecommunications industry for routing information. FBGs are commonly fabricated using photo-sensitive optical fibers that are exposed to spatially varying light source such as a deep UV laser in a Michelson interferometer [14, 15]. The periodic varying light induces photo-chemical reactions which modifies the local structure of the fiber to create a periodic variation in the local refractive index that acts like a Bragg grating. Wavelength of light resonant with the Bragg period is reflected back, while non-resonant wavelengths pass through the grating. Change in surrounding temperature impacts the effective grating period by either linear thermal expansion of the material and/or a change in the fiber’s refractive index due to temperature (thermo-optic effect) [10, 14, 15]. Existing literature indicates the FBG show a temperature dependent shift of ≈10 pm/K around 293 K.[10] There is some disagreement over the temperature dependent behavior and measurement uncertainties with different reports suggesting the thermal response may be linear[16, 17] or quadratic[18].
In this study we have examined the temperature dependent response of polyimide coated FBGs over the temperature range of 233 K to 393 K against a calibrated Platinum resistance thermometer (PRT). Our results indicate that the FBG show a quadratic dependence on temperature with combined expanded measurement uncertainties of ≈500 mK (k=2).
2. Experimental
1. Fiber Bragg Gratings
In this study we have utilized commercially available polyimide coated silica fiber based FBG with Bragg resonance set at 1540 nm, 1550 nm (3 fibers), and 1560 nm. Fibers where stored in a humidity controlled environment (20 % RH) prior to use. Each fiber was cleaved such as to leave 5 mm of excess fiber on one side of the sensor, with the other side, 0.5 m long terminated in a fiber optic coupler. The sensor was then guided through a 23 gauge needle (inserted through a cork) into a 240 mm long (6.5 mm diameter) glass tube. The bottom 50 mm of glass tube was filled finely ground, dry MgO power, completely immersing the sensor, to ensure excellent thermal contact between the temperature bath and FBG. We find that packing the sensor in MgO powder improves the temperature repeatability by a factor of four (data not shown) likely by minimizing thermal gradients. A small amount of desiccant was added on top of MgO powder to keep the assembly dry. The glass tube was then backfilled with Ar gas, following which the glass tube was sealed by inserting the cork and sealing the cork/needle/glass interface with quick-dry epoxy.
2. FBG interrogation system
We have custom built a laser based FBG interrogation system for interrogating the FBG (Fig 1). Briefly, the assembled FBG is placed in a cylindrical Al block (25 mm diameter, 170 mm length). The cylinder has two 150 mm long blind holes (2.5 mm and 6.5 mm diameter) for accommodating a calibrated Platinum resistance thermometer (PRT) and the assembled FBG sensor, respectively. The metal block is placed inside the dry-well calibrator (Fluke 91701). The dry- well calibrator’s temperature is controlled by an automated LabVIEW program which cycles the temperature between 233 K to 393 K at 5 K intervals. Once the set temperature is achieved, the program allows 5 mins of equilibration time following which the laser (New Focus TLB-6700 series, 1520 nm to 1565 nm) scan is initiated. A small amount of laser power was immediately picked up from the laser output for wavelength monitoring (HighFinesse WS/7) while the rest, after passing through the photonic device via an optical circulator (ThorLabs), was detected by a large sensing-area power meter (Newport, model 1936-R). Five consecutive scans were recorded at each temperature and each sensor was thermally cycled four to five times in each run unless noted otherwise. The recorded data was fitted using a polynomial fitting routine to extract peak center, peak height and peak width as a function of temperature.
Figure 1.
a Schematic of experimental setup used to measure the temperature response of FBG
3. Temperature dependence of FBG
The grating equation is given by:
| (1) |
Where λB is the Bragg wavelength, L is the grating period and ne is effective refractive index. The temperature dependence of the FBG resonance derives from changes in the refractive index (ne) due to the thermo-optic effect and to a lesser extent changes in grating period due to thermal expansion of the fiber. The thermal behavior of FBG can be adequately captured using a Taylor expansion of wavelength as function of temperature which yields the following expression:
| (2) |
where, T is the sensor temperature, R is the reference temperature (273.15 K) and λT and λR are the Bragg wavelength at temperatures T and R, respectively. Using the wavelength-period relationship (eq 1), eq 2 can be re-written as:
| (3) |
Examination of eq 3 shows that significant departure from linearity in the wavelength-temperature relationship would indicate that the thermo-optic coefficient is varying with temperature. Studies of thermal dependence of thermo-optic coefficient in fused silica[19] indicate a monotonic dependence on temperature.
2. Results and Discussion
The thermal response of FBG was investigated over the range of 233 K to 393 K. As shown in Fig 2 the wavelength-normalized temperature response of the FBG shows a similar profile over a wide range of Bragg wavelengths. Fitting the thermal response to linear function (Δλ/ΔT ≈ 10.9 pm/K) consistently yields parabolic residuals with deviations as large as ≈6 K. In agreement with Flockhart et al[18] we find that fitting to a quadratic function adequately minimizes the residuals to the point where measurement repeatability dominates the residuals. We do not observe any significant correlation between bandwidth or amplitude and temperature (data not shown).
Figure 2.
a) Wavelength normalized (λT/λ273.15 K) temperature- dependent response of four different sensors. b) FBG show a quadratic dependence on temperature
Thermal cycling reveals the FBG sensors undergo a temperature related aging or annealing process that results in significant hysteresis. As shown in Fig 3a, the thermal hysteresis is largest in the first cycle and steadily decreases over subsequent cycles. We estimate the contribution of annealing to measurement uncertainty by taking standard deviation of the mean of residual in each subsequent thermal cycle excluding the first cycle. Annealing or aging of the fibers is only observed when fiber is heated to temperatures above 373 K (data not shown). Examination of the temperature induced drift indicates that annealing at temperatures of 373 K or above results in a clear downshift of the Bragg resonance at 293 K indicating the effective path length of the grating has decreased. The 7.6 pm downshift corresponds to temperature drift of ≈ 0.7 K. This change likely derives from changes in the poly-crystallinity of polyimide which is known to vary with environmental variables such as temperature and humidity.[20, 21]
Figure 3.
a) polyimide coated FBG sensors show a thermal induced hysteresis that reduces with increasing runs b) heating sensor to 393 K results in a 6 K drift.
Once the FBG have been properly annealed, the measurement uncertainty in our measurements is dominated by the uncertainty in thermal hysteresis and best fit uncertainty which includes contributions from peak center determination (Table 1). The 170 μW of laser power used in the study avoids induced self-heating eliminating a common source of uncertainty (data not shown); similarly the use of sealed, low humidity environment allows us to eliminate another potential source of measurement uncertainty[22]. As shown in Table 1, for five thermally annealed fibers, our combined expanded measurement uncertainty for humidity controlled, strain-free sensor is ≈ 500 mK.
Table 1.
Uncertainty in FBG temperature measurement (all values in degrees Kelvin)
| Temperature (PRT) | 0.002 |
| Wavelength | 0.01 |
| Fit Residual | 0.14 |
| Ageing Effect | 0.2 |
| Combined Uncertainty (k=1) | 0.24 |
| Expanded Uncertainty (k=2) | 0.49 |
3. Summary
Using careful, traceable measurements of FBG temperature sensors in controlled environment we have demonstrated that FBG sensors can be used to make accurate temperature measurements over the range of 233 K to 393 K. Our results for polyimide coated sensors indicate the polyimide cladded fibers undergo temperature induced structural changes resulting in significant hysteresis that reduces over subsequent cycles. Hermetically sealed, annealed sensors, in low humidity and strain-free environment show expanded measurement uncertainties of ≈500 mK.
Acknowledgments
The authors would like to acknowledge Arec Jamgochian and Wyatt Millar for help in setting up measurements and Kevin Douglass and Justin Bohn for helpful discussions
Footnotes
Disclaimer: Certain equipment or materials are identified in this paper in order to specify the experimental procedure adequately. Such identification is not intended to imply endorsement by the National Institute of Standards and Technology, nor is it intended to imply that the materials or equipment identified are necessarily the best available.
References
- 1.Price R. The Platinum resistance Thermometer. Platinum Metals Rev. 1959;3:78–87. [Google Scholar]
- 2.Xu H, Hafezi M, Fan J, Taylor J, Strouse GF, Ahmed Z. Ultra-Sensitive Chip-Based Photonic Temperature Sensor Using Ring Resonator Structures. Optics Express. 2014;22:3098–3104. doi: 10.1364/OE.22.003098. [DOI] [PubMed] [Google Scholar]
- 3.Strouse GF. Standard Platinum Resistance Thermometer Calibrations fromthe Ar TP to the Ag FP. NIST Special Publication. 2008:250–281. [Google Scholar]
- 4.Kersey AD, Berkoff TA. Fiber-optic Bragg-grating differential-temperature sensor. Photonics Technology Letters, IEEE. 1992;4:1183–1185. [Google Scholar]
- 5.Kim GD, Lee HS, Park CH, Lee SS, Lim BT, Bae HK, Lee WG. Silicon photonic temperature sensor employing a ring resonator manufactured using a standard CMOS process. Opt Express. 2010;18:22215–22221. doi: 10.1364/OE.18.022215. [DOI] [PubMed] [Google Scholar]
- 6.Klimov NN, Berger M, Ahmed Z. Characterization of Ring Resonator Structures for Applications in Photonic Thermometry. 2015 manuscript in reveiw. [Google Scholar]
- 7.Peng Y, Hou J, Huang Z, Lu Q. Temperature sensor based on surface plasmon resonance within selectively coated photonic crystal fiber. Applied Optics. 2012;51:6361–6367. doi: 10.1364/ao.51.006361. [DOI] [PubMed] [Google Scholar]
- 8.Zou L, Bao X, Chen L. Simultaneous distributed Brillouin strain and temperature sensor with photonic crystal fiber. Proceedings of SPIE 5384. 2004:13–17. [Google Scholar]
- 9.Ahmed Z, Gooding EA, Pimenov KV, Wang L, Asher SA. UV Resonance Raman Determination of Molecular Mechanism of Poly(N-isopropylacrylamide) Volume Phase Transition. The Journal of Physical Chemistry B. 2009;113:4248–4256. doi: 10.1021/jp810685g. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Mihailov SJ. Fiber Bragg Grating Sensors for Harsh Environments. Sensors. 2012;12:1898–1918. doi: 10.3390/s120201898. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Donner JS, Thompson SA, Kreuzer MP, Baffou G, Quidant R. Mapping Intracellular Temperatrure Using Green Flurorescent Protein. Nano Letters. 2012;12:2107–2111. doi: 10.1021/nl300389y. [DOI] [PubMed] [Google Scholar]
- 12.Reese CE, Mikhonin AV, Kamenjicki M, Tikhonov A, Asher SA. Nanogel nanosecond photonic crystal optical switching. J Am Chem Soc. 2004;126:1493–1496. doi: 10.1021/ja037118a. [DOI] [PubMed] [Google Scholar]
- 13.Weissman JM, Sunkara HB, Tse AS, Asher SA. Thermally Switchable Periodicities and Diffraction from Mesoscopically Ordered Materials. Science. 1996;274:959–960. doi: 10.1126/science.274.5289.959. [DOI] [PubMed] [Google Scholar]
- 14.Hecht J. Understanding Fiber Optics. 4. Prentice Hall; 2002. [Google Scholar]
- 15.Krohn DA. Fiber Optic Sensors: Fundamentals and Applications. 3. ISA; 2000. [Google Scholar]
- 16.Yun-Jiang R, Webb DJ, Jackson DA, Lin Z, Bennion I. In-fiber Bragg-grating temperature sensor system for medical applications. Lightwave Technology, Journal of. 1997;15:779–785. [Google Scholar]
- 17.Hidayat A, Wang Q, Niay P, Douay M, Poumellec B, Kherbouche F, Riant I. Temperature-induced reversible changes in teh spectral characteristics of fiber Bragg gratings. Applied Optics. 2001;40:2632–2642. doi: 10.1364/ao.40.002632. [DOI] [PubMed] [Google Scholar]
- 18.Flockhart GMH, Maier RRJ, Barton JS, MacPherson WN, Jones JDC, Chisholm KE, Zhang L, Bennion I, Read I, Foote PD. Quadratic Behavior of Fiber Bragg Grating Temperature Coefficients. Applied Optics. 2004;43:2744–2751. doi: 10.1364/ao.43.002744. [DOI] [PubMed] [Google Scholar]
- 19.Leviton DB, Frey BJ. Temperature-dependent absolute refractive index measurements of synthetic fused silica. Proc SPIE, Optomechanical Technologies for Astronomy. 2006;6273:6273K–6271. [Google Scholar]
- 20.Heberer DP, Cheng SZD, Barley JS, Lien SHS, Bryant RG, Harris FW. Crystallization and Morphology of Semicrystalline Polyimides. Macromolecules. 1991;24:1890–1898. [Google Scholar]
- 21.Jwo-Huei Jou RH, Huang Pier-Teh, Shen Wen-Ping. Structure effect on water diffusion and hygroscopic stress in polyimide films. Journal of Applied Polymer Science. 1991;43:857–875. [Google Scholar]
- 22.Kronenberg P, Rastogi PK, Giaccari P, Limberger HG. Relative humidity sensor with optical fiber Bragg gratings. Optics Letters. 2002;27:1385–1387. doi: 10.1364/ol.27.001385. [DOI] [PubMed] [Google Scholar]